Product Specifications MaxLumen MRI LED Diffused Lighting

Similar documents
Product Specifications MaxLumen MRI LED Diffused Lighting

Product Specifications MaxLumen MRI LED Diffused Lighting

Product Specifications Linear Accelerator Relay Junction Box

Product Specifications 480 VAC, 175A Main Power Distribution Panel

Product Specifications 480 VAC, 70A Main Power Distribution Panel

Product Specifications Auto Restart Main Power Distribution Panel

Product Specifications Warning Light Control Cabinet

Product Specifications Warning Light Control Cabinet

Product Specifications 3 Breaker Main Disconnect Panel with Undervoltage Trip

Product Specifications Fail Safe Emergency Power Off Control Panel 480V 3 Phase

Product Specifications 4 Breaker Main Disconnect Panel with Undervoltage Trip

Vaya Linear LP, White & Mono

Basics Series Surface Mount LED Fixtures. Affordable and easy-to-install, 4" and 6" fixtures. It s a. Snap

Millenium Os. Luminaire LED. For Stairwells and Public Spaces. Designed and manufactured in the USA. High Abuse Motion-Activated Lighting Product

LUXCOMFO TM LED Flood Light Datasheet

SECTION INTERIOR LIGHTING

BAS SFP22 Edge-lit LED Panel

E10 & L9 Low Wattage Light Cone Data

4 LED Recessed Adjustable. Energy efficient, stylish downlighting for residential and commercial applications

Hazardous Location LED Pivoting Light - C1D2 - Dimmable - Modbus TCP/IP Network - Meets USCG Specs

TROFFER RETROFIT KIT (TRK)

Class 1 Div 2 LED Pivoting Light - 2 foot 2 lamp - LED T series Style V Dimmable

LED ECO PANEL.

MAR2 LED SERIES M A R 2 L E D. Linear LED Lighting for Hazardous Locations

WARRANTY LED ECO PANEL.

Product Specifications

POINT HELIDECK LIGHTS PRL-LSM LED LOW SURFACE MOUNT LIGHT

DRD5S & SurfaceFrame Alt/EM Surface Mount LED Downlight

LED RETROFIT SOLUTIONS

LED ROUND MINI PANELS

LUMINAIRE ID SIZE CENTER OPTICS SIDE OPTICS LIGHT SOURCE CRI LUMEN PACKAGES COLOR TEMP. VOLTAGE. LED - high performance LED 80-80CRI 90-90CRI

Dimmable LED Retrofit Kit

LED retrofits! 4 T8 Specification Guide. Retrofit your existing Fluorescent Lights in Seconds No Electrician Needed! Pure Bright

Albeo LED Luminaire. Heavy Industrial Low Bay Lighting. (ALR1 - Series)

50W Explosion Proof Trunnion Mount Compact LED Flood Light - C1D2, C2D1 - Paint Spray Booth Approved

Classix 6" 6" Round Lens Wall Wash

Classix 6" 6" Round Lens Wall Wash

LED High Bay. 150W LED High Bay GLC-ANHB-150W-1. PERFORMANCE LED Chip Lumen Output (IES) Power Consumption Efficacy (IES) Beam Angle

Classix 6" 6" Round Lens Wall Wash

PRODUCT DATA SHEET LIGHTING PRODUCTS

VISAGE ENVIRONMENTAL LIGHTING SOLUTION. INDOOR Lighting SERIES

C2D1/C1D2 Explosion Proof 50 Watt LED Light Fixture - Paint Spray Booth Approved - 6,250 Lumens

POINT ROLLOVER LIGHTS PRL LED HELIDECK SEMIFLUSH LIGHT

CoreLine Waterproof the clear choice for LED

LED LUMINAIRE SERIES LED Luminaire Series Improves Your Illumination Systems

80W C1D2 Hazardous Location Integrated LED - Corrosion Resistant for Marine - Emergency Backup

Modular & linear architectural lighting for residential and commercial environments.

DHL-LP-LED Low Profile LED Linear High/Low Bay

C1D1 Explosion Proof 50W LED Flood Light Fixture - Paint Spray Booth Approved - Adj Surface Mount

150W Explosion Proof High Bay LED Light Fixture - C1D1&2 - C2D1&2-17,500 Lumens - 60 Beam

THE HIGH STRIP FIXTURE SERIES PERFORMANCE KEY FEATURES INCLUDE: LED Fixture Series for General, Direct and Indirect Illumination

Beautiful Light Technology Corp,.

Evolution EXWW RECESSED LINEAR WALL WASH LUMINAIRE

LED Edge-Lit Flat Panels

Product Specifications

Lumination LED Luminaires

CLM 422 Series LED Ceiling Luminaire Module - 45Watt

LED RETROFIT CONVERSION KITS

LED Linear Design Highlights

MASTERSPEC TECHNICAL SPECIFICATIONS DIVISION 26 ELECTRICAL

Explosion Proof LED Strobe Light - 4Hz Flash Rate - 10 Watts - Class I, Div. I - IP67

Dimmable General Area Use High Bay 320 Watt LED Light Fixture - Low Profile - High Efficiency

Project: Comments: Prepared By: Control. Code Fixture & Trim D3 Wall Wash, Round. Platform & Control Wide Gamut, Tunable

Capella 2. Project Name. Construction. Optical System. Product Applications. Housing and lens frame are constructed from heavy-duty low

POINT ROLLOVER LIGHTS PRL LED HELIPORT INSET LIGHT

48W Fluorescent T8 Light w/ Emergency Backup Dimmable Ballast - (2) 2 Foot T8 Lamps - C1D2

SYSTEM. field cuttable. lipleds 8200 Series

Suite: Foyer/Hallway/Closet PROJECT: ZIGG CONDOS TYPES ' S1 ' & ' S6 ' Description: Flush Mount lass. Finish: Diameter: " Height: 4" Lamp: x

Explosion Proof C1D2 0-10V Dimmable LED Fixture - 4ft 2 Lamp - Emergency Ballast - Terminal Strips

OLUX LED Flood Light OSRAM. Datasheet APPLICATIONS.

Inset Kit LED. Inset Trend Swing REF+ V2 5W 40 WH WW PRODUCT OVERVIEW

CPPG ENERGY EFFICIENT CANOPY LIGHTS FOR PARKING AREAS

The Commercial LED Lighting Specialist

FPLHB - Fidelux Premier Linear Highbay

Ultra Energy Efficient

DRD5S & SurfaceFrame Alt/EM 18" Linear Surface Mount LED Area Light

Fortimo LED Line Gen2 Module

RECESSED - WITH StepLens OPTION

Classix 6" 6" Round Lens Downlight

300W Low Profile Clean Room LED Light - LED Signal Light - C1D1 - C2D2 - ISO 14644/FS-209E Rated

Light. can be. LED Edge-Lit Flat Panel AWESOME

SECTION LIGHTING

ULTRA RT6 and RT6 Gimbal LED Recessed Downlight Kit

Low Profile Clean Room 150W LED Light - 2x2 Lay-In Troffer C1D1 - C2D2 - ISO 14644/FS-209E

THE HIGH STAIRWELL AND PASSAGEWAY FIXTURE PERFORMANCE KEY FEATURES INCLUDE:

SenScape Luminaires Ideal for Covered Parking, Canopy and Low Bay Applications

33 Watt Low Voltage LED Fixture for Hazardous Location Lighting - Low Profile Linear Fixture - C1D2

ZipTwo LED 707. Micro-profile interior direct lighting for open office, wall wash and wall graze applications. Spec Guide

56W Hazardous Location LED Fixture - C1D2 - Stainless Steel - Dimmable - DALI Network

MALIBU FM /2" N-Brushed Nickel. Nickel **ESTIMATED FROM LED SOURCE

Grüv Flare Regressed. Recessed Linear LED

Etch 2' x 2' Recessed LED

Explosion Proof 70 Watt Low Profile LED Light Fixture Lumens - C1D1 Group B - Hydrogen Rated

LED ECO PANEL.

Technical Application LED Pendant Light LUZ-PL W-W

ELLA ID. High CRI, High lm/w OptoDrive LED-module

Grüv Flare Regressed. Recessed Linear LED

DIMMING 0-10V Continuous

LED ECO PANEL.

THINTEK G2 PANEL.

Transcription:

NOVA Energy & Automation Product Specifications MaxLumen MRI LED Diffused Lighting Model MXLumen22-X 03 January 2012

Product Description 1. Standard Applications The Nova Max Lumen LED 2x2 light fixture was developed specifically for use inside MRI imaging suites to provide entire room illumination while emitting no electrical magnetic interference and shielded to prevent distortion from the high frequency noise emitted from the MRI machine. The LED system provides maintenance free operation with a rated LED life of 10 years. The Max Lumen LED 2x2 is fully dimmable and provides a smooth linear dimming when used in conjunction with the Nova Dimming controls. Figure 1 Nova Max Lumen LED 2x2 fixture 2. System Compatibility The Nova Max Lumen LED 2x2 light fixtures are designed, shielded, and tested specific for MRI applications including MRI installations including 3T. Author: JF 2

3. Features/ Benefits Features Benefits ETL, cetl, CE listed, tested to UL standards to conform to NEC. Labeled to conform with NFPA99, NFPA-70. MRI tested Multiple configurations to fit the budget and architecture of the room Factory wired and tested. Shielded Electronics Non Ferrous Construction- Constructed with light weight aluminum extrusions and commercial grade stainless steel hardware Photography grade acrylic diffuser White Powder Coated Finish Remote Modular Power Supply Smooth Linear Dimming Control High Efficient LED technology 91 lm/w Multiple color temperatures available from warm, natural and cool white LM-79 IES tested RoHS certified Low Watt LED array dissipates heat without the need for heavy heat sinks Maintenance free LEDs provide a 10 year illumination life span No UV or IR Radiation 4 corner earthquake bracing Designed, tested, and installed on several projects world wide Manufactured using the highest quality components for high reliability and long life Available to ship in 3 weeks Thin, Light Weight, Durable Panels Even and diffused Illumination Simplified Wiring 500% reduction in energy consumption as compared to MRI DC Lighting Cool to the touch, reduces building heat load Reduces Building Carbon Footprint No Bulbs to replace Save on initial cost- The Nova MaxLumen LED system provides substantially more lumens per fixture than comparable products, requiring fewer fixtures per room. The MaxLumen LEDs do not require a large, expensive DC power supply that would be needed on an incandescent installation. Save On Maintenance Costs- MRI lighting with traditional incandescent DC power require lamp maintenance at ½ the rated life of the lamp- requiring a maintenance technician to enter the room with the proper non ferrous tools every 4 months and change out the lamps. Reduce Insurance Liability- the risk of injury to a maintenance technician changing lamps is completely eliminated. The room will always be properly lit to the desired foot candle which will reduce the risks to patients and staff associated with a poorly lit procedure room. Author: JF 3

4. Specifications Model MxLumen22- X (3 rd generation) Application Input Voltage Current Power Classification Power Consumption MRI LED Light Fixture 24-28V DC (See Power Supply Specifications) 1400mA Class 2 power limited circuit 42W ± 2W Power Factor >0.95 LED Count 729 Luminous Flux Luminous Efficacy 3850± 50 lm 91lm/W Color Temperature Warm White: 2800-3200 K Part Number MxLumen22-W Natural White 4250-4750 K Cool White: 5500-6500 K Part Number MxLumen22-N Part Number MxLumen22-C CRI >85 Beam Angle 150 0 Lighting Source LED Array Approvals ETL, cetl, CE, RoHS, IES LM 79, California title 24 Author: JF 4

5. Weight and Dimensions Mounting Dimensions Depth Weight Enclosure Finish Lens Type Environment Surface or hard lid or suspended ceiling grid installation 23.74 (603mm) x 23.74 (603mm) 1.57 (40mm) 9 lbs Aluminium Extrusion White housing with brushed aluminium trim 1/8 Frosted Acrylic diffuser Indoor use only Operating Temperature -4~104 0 F (-20~40 0 C) Figure 2 Dimensions: Outside, Mounting, Depth Author: JF 5

6. IESNA LM 79 Test Results IES file available for download. Figure 3: Polar Graph Maximum Candela = 1395.85 Located at Horizon Angle =0, Vertical Angle = 0 #1 Vertical Plane Through Horizontal Angles (0-180) Red #2 Vertical Plane Through Horizontal Angles (90-270) Black #3 Horizontal Cone Through Vertical Angle (0) (Through Max. Cd.) Green Zonal Lumen Summary Luminance Data (cd/sq.m) Zone Lumens % Lamp % Fixture Angle Average 0-deg Average 45-deg Average 90-deg 0-30 1071.91 27.4 27.4 45 4324 4381 4428 0-40 1744.93 44.6 44.6 55 4070 4155 4241 0-60 3055.28 78.1 78.1 65 3838 3922 3982 0-90 3894.93 99.6 99.6 75 3453 3527 3629 90-120 7.34.2.2 85 2662 2704 2808 90-130 9.77.2.2 90-150 113.32.3.3 90-180 16.68.4.4 0-180 3911.61 100 100 Total Luminaire Efficiency = 100% Author: JF 6

Candela Tabulation 0 30 45 75 90 0 1395.85 1395.85 1395.85 1395.85 1395.85 5 1388.55 1388.16 1388.95 1388.87 1389.03 15 1329.66 1330.76 1332.05 1333.99 1334.62 25 1218.87 1221.11 1224.97 1230.33 1230.25 35 1064.74 1071.27 1075.81 1084 1085.11 45 885.38 891.06 896.93 905.56 906.67 55 675.99 683.76 690.09 702.6 704.43 65 469.64 474.77 479.93 488.21 487.31 75 258.77 261.49 264.3 269.85 271.95 85 67.19 67.37 68.23 70.48 70.87 90 7.97 8.43 8.44 8.17 6.09 7. Power Supply The Nova Max Lumen LED Power Supply is the ideal LED power supply to meet the hospital-grade lighting power requirements of RF-shielded MRI suites. The voltage regulation and performance characteristics of the LED Lighting Control Unit offer a significant advantage over competing products. As a single phase or 3-phase (optional), solid state DC power supply, the LED Lighting Control Unit assures a highly-filtered, regulated, dimmable, DC output, and provides illumination without image interference. Figure 1 Typical LED Power Supply One-Line 7

8. System Compatibility The Nova 24-28V LED Power Supply are designed, filtered, and tested specific for MRI applications including MRI installations up to 3T. The Nova LED Lighting System has been successfully installed on MRI installations from OEMs including: Toshiba, GE, Siemens, and Philips. 9. Features/ Benefits Features Benefits ETL, cetl, CE listed, tested to UL standards to conform to NEC. Labeled to conform with NFPA99, NFPA-70. MRI tested Multiple configurations to fit the budget and architecture of the room Factory wired and tested. Filtered Power Smooth Linear Dimming Control RoHS certified Pre wired Pre-terminated Power Distribution Box Includes MRI Shielded Cables Designed, tested, and installed on several projects world wide Manufactured using the highest quality components for high reliability and long life Available to ship in 3 weeks Thin, Light Weight, Durable Panels Simplified Wiring 500% reduction in energy consumption as compared to MRI DC Lighting Cool to the touch, reduces building heat load Reduces Building Carbon Footprint Save on initial cost Fast Installation Figure 2 MRI LED Power Supply 8

10. Power Supply Specifications Dimming LED Power Supply Model NEAS-DLPS-24 Application Output Voltage Output Power Input Voltage AC Input Current Output Current Dimming 24-28V DC MRI LED Power Supply 24-28V DC 240W 320W 480W 120/ 277 120/ 277 120 2.1A 2.8 A 4.3A 10A 15A 20A Efficiency 95% 95% 92.7% Number of Fixtures Part Number 6 8 10 NEAS-DLPS-24-240 NEAS-DLPS-24-320 NEAS-DLPS-24-1-480 Lifetime Expectancy Noise Voltage max Output Over Voltage Protection Over Temp Protection Input Transient Voltage Protection Power Classification > 130k Hrs <50mVpp Typ 30.5VDC max 32Vdc Shutdown & Auto Restart MOV (Metal Oxide Varistor) Class 1 power limited circuit Dimming Mode Pulse Width Modulation Operating Frequency 390 Hz Dimming Range 0-100% 9

11. Power Supply Weight and Dimensions The controller may be semi flush mounted within a 6 deep wall. This will leave exposed approximately 2.27 extending past the finish drywall surface. Model Dimensions Weight Mounting Enclosure Enclosure Sealing Finish Latching Mechanism Interior Panel Protection Category NEAS-DLPS-24 Height: 15.75 (400mm) Width:13.77 (350mm) Depth: 5.99 (150mm) 30 lbs. (14 kg) Rear wall mounting holes (4). Spaced 0.79 (20mm) from enclosure edge. Optional mounting brackets are available 16 Gauge Carbon Steel, all rounded corners 1 gland plate in the enclosure base. Foamed-in place polyurethane door gasket Dip coat primed, powder coated in textured RAL 7035 Two ¼ turn latches with double-bit inserts (cam lock). Door is hinged on left side but can be swapped if specified when order is placed. Zinc-Plated subpanel IP 66 to EN60 529/09:2000, complies with NEMA 4 Enclosure Approvals: UL, CSA, TUV, GL, Lloyds, VDE, Bureau Veritas Figure 3 LED Power Supply Mounting Diagram 10

12. Class 2 Power Distribution Panel Specifications The Class 2 Power Distribution Panel allows intrinsically safe (NEC Article 504) class 2 low voltage wiring to be completed using a single class 1 RF Filter Model NEAS-PD8UL, NEAS-PD16UL Application Class 1 to Class 2 Power Distribution Panel Incoming Voltage 24VDC Incoming Current 10A/ 20A Lighting Load 8/16 Circuit, Class 2 Power Limited 3.5A Dimensions 10 Hx 6 W x 4 D Figure 3: MRI LED Class 2 Power Distribution 13. MRI Shielded Cable Specifications The 50' Pre-terminated MRI Shielded Cable with quick connect termination comes installed and connected to the Power Distribution Panel (1 cable/ fixture) and allows each light to be installed with no additional tools. Model Application Cable Size Termination at Power Distribution Box Termination at Light Fixture Length MXLumen-Cable MRI Shielded cable with quick connect termination 20 AWG Pre Wired and terminated Quick Connect Plug (no tools required) 50 feet Figure 4: 50' MRI Shielded Cable- Pre-terminated with a quick connect for fast installation- no tools or conduit needed 11

Dimming Controls The LED Dimming switch is proprietary to the power supply, and it is not recommended to use dimming controls from other manufacturers Model Application LED-DSW 0-10V Dimmer Switch for LED lighting Figure 7: MRI LED Dimming Control 14. Additional requirements RF Filter The RF filter is typically provided and installed by the shielding contractor. There are no special requirements for this item, a typical off the shelf 30A filter will meet the needs of the LED lighting circuit. Model Description Quantity Suggested Manufacturer Suggested Part Number RF Filter supplied by shielding contractor Typical 30A MRI filter, no special requirements Minimum 1 Filcoil or equal FC-3147 or equal 12

15. Example Projects 29ft L x 17ft W MRI room 1.5T magnet 7 LED 2x2 fixtures 26ft L x16ft W MRI room 1.5T magnet 6 LED 2x2 fixtures 4 LED 2x2 sky ceiling mural 13

26ft L x 19ft W MRI room 3T magnet 8 LED 2x2 fixtures 6 LED 2x2 Sky Mural 24ft L x15ft W MRI room 3T magnet 6 LED 2x2 fixtures 14

16. Product Service and Technical Assistance Nova Automation, LLC 2722 N. Avondale Blvd Milwaukee WI 53210 PH# 414-779-NOVA Website: www.novaautomation.net 17. Warranty This limited warranty set forth below is given by Nova Automation ( Seller ) with respect to the electrical equipment ( Product ) packaged with this limited warranty. The Product, when delivered to you in new condition in its original packaging, is warranted against defects in materials or workmanship as follows: For a period of five (5) years from the date of original purchase, defective parts or a defective Product returned to a Seller, or its authorized service providers, as applicable, and proven to be defective upon inspection, will be repaired, or exchanged for a new Product, as determined by the Seller, or the authorized service provider. This limited warranty covers all defects encountered in normal use of the Product, and does not apply in the following cases: Loss of or damage to the Product due to abuse, mishandling, alteration, accident, electrical current fluctuations, failure to follow operating, maintenance or environmental instructions prescribed by Seller, failure to follow Sellers installation instructions, or service performed by someone other than Seller or its authorized service provider. Nova assumes no responsibility for labor or freight costs incurred in connection with the installation, removal, or replacement of products determined to be defective or any consequential or incidental damages arising from the use of the product. Nova Automations entire liability on any claim of loss or damage resulting from a defective product is limited to the replacement of the product. WARRANTY IS VOID IF PRODUCT IS NOT USED FOR THE PURPOSE FOR WHICH IT WAS MANUFACTURED. 15