Tri-Tech Medical Inc.

Similar documents
Med Touch Combination Alarm Conversion Kits

Med Touch Master Alarm

Tri-Tech Medical Inc.

EZT Series Integra Med Touch Area Alarm & Zone Valve Box with EZ Backfeed

Installation & Operating Instructions for Med Touch Series Area, Master & Combination Alarm Systems

Installation & Operating Instructions for ZT and EZT Integra Touch Series Area Alarm Zone Valve Box Systems

DC Series Area, Master and Combination Alarm Systems Opreator s Manual

Integra Med Touch Series LCD Area Alarm & Zone Valve Box Systems

Installation, Operation, and Maintenance Instructions

Installation & Operating Instructions for Area, Master & Combination Alarm Systems

T a b l e o f C o n t e n t s

Alert-3 LCD Alarm v2.0

Installation & Operating Instructions for Integrated Series Area Alarm Zone Valve Box Systems

Tri-Tech Medical Inc.

Zone Valve Alarm Combination Systems Installation and Operating Manual

MEGA. cal electronic gas alarm. Installation, Operation, and Maintenance Instructions. Part No Rev. D Pg.

SPECIFICATION MODEL 5200 GAS MONITOR SYSTEM. User Instructions For The Model 5200 Gas Monitor System

Alert-3 LCD Alarm v2.5

Alert-3 LCD Alarm v2.4

CRIOTEL. The complete solution for local and remote monitoring of cryogenic tanks under pressure

Self-contained and intrinsically safe wireless stand-alone multi-point gas detector.

Installation Instructions for Med-Touch Series LCD Medical Gas Area, Master & Combination Alarms Conversion Kits

SCAN200E USER S MANUAL

Installation and Maintenance Manual. Alarm Valve Combo Unit Alert-1 v1.6

Series 9000 Total Hydrocarbon Analyzer

FTA550 JOCKEY PUMP CONTROLLERS

RPM1600 Series Room Pressure Monitors

Fire control panel detect 3004 plus and accessories

Liquid Helium Level Instruments

User Manual. Dryer Controller M720

OJD ITB No Attachment E Duress Alarm Equipment Specifications

FTA550F JOCKEY PUMP CONTROLLERS STANDARD SUBMITTAL PACKAGE

Across-the-Line SERIES MP300 Combined Manual and Automatic

Carbon Monoxide Transmitter

UltraLITE Model ELU Centralized Emergency Lighting Inverter 4.2 KW- 5 KW

Dräger Polytron 7000 Fixed gas detector

Operating & Maintenance Manual. Alert-4 Ethernet LCD Master Alarm

Dryer Controller M720

Variable Frequency Drive SERIES MP800 VFD

APPLICATION DATA SHEET

MODEL A-316 CARBON MONOXIDE (CO) MONITOR AND ALARM FOR COMPRESSED AIR TESTING

PROCESS ANALYSERS. SERVOFLEX MiniMP (5200) Gas Analyser. Operator Manual. Part Number: A Revision: 8 Language: UK English

Laboratory Enclosed Scroll Air Compressor System with Premium NFPA Controls

THERMAX Efficiency Monitor

Soft Start Series MP700 Solid State, Reduced Voltage

1.1 Remote Annunciator Controller (RAC) using the one ATC-300+ connected via RS-485 network.

CSP-204 CSP-208 CSP-104 CSP-108

opti-visor ultra-efficient chiller plant automation Typical specification

DIESEL Engine Fire Pump Controllers Features. FD120 Diesel Engine Controllers

Nitrogen Dioxide (NO2) Single-Point Gas Detection System

Power Flame Incorporated

Halton SAFE / 7.14 user guide and installation instructions

ROYCE SUSPENDED SOLIDS SYSTEM MODEL 7011A AND SENSORS MODELS 72A, 72P, 73B, 74A, 75A & 76A

D8024, D9024, D10024 Analog Fire Alarm Control Panels Programming Guide

Bacharach Bodyguard 4 User's Guide

Laptop / PC Programming Manual

Replaceable LED modules. Sleep or unattended mode. Auto-silence and auto-acknowledge

The EN54 Part 2 & 4 Fire System

Two-Channel Gas Controller

EasyTronic III MANUAL SERVICE

Centro-Matic Automated Lubrication Systems System Controls

NexSysLink. 2 CAN Display Operation Manual. CAN Instruments Product Family

Integrated Security Solutions

Model D12-IR. Digital Gas Transmitter...

LC150 VIEW-AT-A-GLANCE PUMP CONTROL-TELEMETRY UNIT

Wilo CO-Helix Booster Systems. Product Presentation July, 2017

APPLICATION DATA SHEET

STEAMPAK Series BurnerPAK BMS Burner Management Systems

Air Comfort. Installation instructions (English) A

for Drinking Water Systems

Contents. Outline of Product. Features of Product. Specifications of Product. Name and central function of section. Illustration of Terminal

1.1 Remote Annunciator Controller (RAC) and 1 Switch (ATC-600/800) via Modbus TCP/IP Ethernet Gateway

DFM. flow pressure temperature MULTI PARAMETER DIGITAL MASS FLOW METERS. Totalizer. Interface

IntelliDoX Operator Manual

Generator Gas Analyzer (GGA)

P R O D U C T S P E C I F I C A T I O N MSA Ultima X Series Sensor/Transmitter Specification

HALT & HASS Systems.

Series. NX-4-EUR Control Panel. Installation manual

NU2 Smart Power Supply (SPS) Manual

Alarm module for leak detection with webserver

Installation Instructions for Med Touch Master Alarms - Wireless

70E Pressure Booster. Bell & Gossett. Bulletin D-215D. Efficient Operation. Easy to Select and Install. Quick Set-up through EZ Start

SmarTemp Control fx 2.0

Control/Communicator Installation Manual

Stability Chambers & Rooms

RE6100 Series Helix Security and Automation Platform

SECTION PACKAGED HEAT TRANSFER SYSTEM

SECTION NURSE CALL SYSTEM

4-20mA CYBER Cyber Transmitter for flammable, toxic and IR gas detection Cyber Head Increased security in ATEX certified head

halt & hass systems

USER MANUAL FOR OPERATING SYSTEM

Control Panel. 1.0 GENERAL SCOPE OF WORK Introduction... 2

ModBus DE-1 INSTALLATION AND USER MANUAL

1.1 Ethernet Remote Annunciator Controller (RAC) using the up to 4 ATC-300+ connected via RS-485 to TCP/IP gateway

Watchguard WGAP864 User Manual

SECTION Process and Immersion Turbidity and Total Suspended Solids Sensor

Flopurge TS. Operation Manual

Ax60+ Multi-Gas. Service Manual

DCA4. Digitally controlled Audio Matrix. Current Thinking. Product Specification March Document Number DDC DCA ACS

Danfoss gas detection units

Transcription:

Submittal Data Sheet Project Information Project Number Approval Features The Tri-Tech Area and Master Alarm Panel digitally displays gas pressure (1 psi increments) monitors and displays normal and alarm conditions from up to 128 remote medical gas source signals and provides alarm conditions as required by the latest edition of NFPA 99 for up to 16 medical gases. The alarm is ETL listed to UL 1069 and CSA C22.2 # 205 Signal equipment*. Transducers & DISS risers are included for area alarm gas boards. Transducers may be installed remotely or in the back box. Pressure switches, transducers for master alarms and DISS union check valve connectors are sold separately. Five year parts and one year labor warranty 2.85 LCD touch screen displays up to 8 normal and alarm conditions for pressure switch sensor inputs 2.85 LCD touch screen displays up to 8 normal and alarm conditions and provides a digital pressure display for up to 4 transducer sensor inputs 2.85 LCD touch screen displays normal and high/low alarm conditions for one 4 20 ma sensor input The LCD touch screen allows all alarm programming and set up to be done without the use of tools A green normal or red alarm condition for each gas services confirms the condition for each individual gas service Emergency preparedness instructions - Med Touch Alarm Panels allow users to set up customized instructions for each alarm signal, to appear on the screen when the signal is in alarm Up to 128 signal points in a single panel Wireless (minimal low voltage wiring) models available (wireless feature for master signal points only) Last event history (per gas board /signal point) Made in the U.S.A. NFPA and ISO pre-loaded gas labels English, and English/French pre-loaded languages Editable text and alarm labels Self-contained unit - Designed for ease of installation and service Self-diagnostic and error message display for ease of maintenance Audio and visual alarm indicators Bright easy to read LCD displays clearly visible in both day and night lighting conditions Constant display and monitoring of each gas User programmable high/low set points on transducer and 4 20 ma sensor input boards Dry contacts provided on the CPU module for general fault condition covering entire panel Hinged frame for easy accessibility Color coded displays Additional features: Individual user programmable remote signal alarm points to accept NO or NC signals, or may be disabled. Factory preset to accept Normally Closed signals LCD indicators (Green) confirms normal status, (Red) indicates abnormal condition Menu of NFPA 99 master alarm signals for quick, easy selection Gas specific transducer master signal boards. An error message will be displayed if incorrect sensor or no sensor is attached. Transducer master signal boards shall be capable of displaying gas readouts in PSI, kpa, BAR, inhg or mmhg. Alarm repeat feature is factory set at 10 minutes, and is adjustable or may be turned off User programmable high/low set points on transducer master boards In addition, each Area Alarm Module shall incorporate the following features: Does not require re-calibration Gas specific sensor with DISS nut & nipple. An error message will be displayed if incorrect sensor or no sensor is attached Shall be capable of displaying gas readouts in PSI, kpa, BAR, inhg or mmhg. Gas alarm repeat feature is factory set at 10 minutes, and is adjustable or may be turned off * ETL Listing in Process Page 1 of 5

As options, the panel shall be able to; The following options shall be able to be added to the panel at the time of order or after installation; Optional Text / e mail notification of alarm events sent to up to 5 addresses Optional Ethernet module. This module may be added to any alarm panel. It will provide the following features: Webserver allows a remote user access to the alarm s webpage viewing a graphic image of the alarm with all signal conditions, pressure & event / history log information Text notification of alarm conditions (up to 5 addresses) Event / history log maintains a rolling list of the 100 most recent alarm condition events and a file of the 1,000 most recent alarm events and is downloadable through embedded web pages Optional Ethernet connection for BACnet or Modbus transmission of signals to a building automation system Optional Ethernet connectivity with embedded web page Interface with building automation system via Ethernet Wireless (minimal low voltage wiring) models available Specification The alarm shall be the Tri-Tech Medical. The panel shall be microprocessor controlled and designed to comply with NFPA 99. The panel shall incorporate an individual 2.85 LCD touch screen display for each gas service and a 2.85 LCD touch screen display(s) for groups of up to 8 master signal points in communication with either dry contacts, transducers or 4 20 ma sensors which shall allow all alarm programming and set up to be done without the use of tools. The unit shall be enclosed in a steel box and shall be designed to accept an electrical input range of 120-240 volts AC 50-60 hertz. The source voltage shall be stepped down with a self-contained transformer. The panel shall contain audible and visual alarm indicators. The audible alarm may be silenced by pressing the alarm silence button, but the visual alarm indicator can only be cancelled by fault correction. The alarm shall detect and filter out transient (less than 0.6 seconds) signals created by R.F.I. The alarm shall be capable of displaying alarm history for all possible alarm conditions. A security pin code shall be required and must be entered in order to access the programming software. A security pin code shall be required and must be entered in order to access the programming software. Gas / Area Display Shown in alarm condition. Display blinks on and off when in alarm condition drawing attention and alternates displaying rooms/area covered with user programmed emergency instructions. Display (both dry and transducer sensors) Shown in normal condition. User selects signal labels from database of NFPA signals included in the boards memory. Individual signal points are pre-programmed NC but may be re-programmed as Disabled or NO by user. Display (dry contact sensors) Shown in alarm condition. Display blinks on and off when in alarm condition drawing attention and alternates displaying signal identification with user programmed emergency instructions. Display (transducer senors) Shown in alarm condition. The standard display looks and acts the same as both of the above Master/Remote displays. By touching the signal point you wish to view, the display changes to the view shown here providing the additional actual line pressure of the gas being monitored. This may be viewed at any time, in Normal or Alarm condition. Page 2 of 5

How to Order: Easy to use modular ordering system. Fill in the blanks to specify the Alarm that meets your needs. Model Numbering: T N C G G/S G/S G/S G/S G/S G/S G/S G/S G/S G/S G/S G/S G/S G/S G/S 2 Slot Front 3 Slot Front 5 Slot Front Note: a 35-3032 kit may be ordered separately and installed in the field to provide; Ethernet webserver, event log, alarm notifications may be sent up to 5 users via text or e-mail and Modbus Interface. The 35-3033 Ethernet programming kit is required and must be ordered when the 35-3032 kit is ordered. Only one is needed per facility. Note: a 35-3035 kit may be ordered separately and installed in the field to provide a BACnet interface. The 35-3032 and 35-3033 kits are also required whenever the 35-3035 kits is ordered. Only one 35-3033 kit is needed per facility. S = Master Signal Input Type Non-Latching Alarm Relays M P Q Description 4-20 ma Dry Contact ( 8 Signal Points ) Transducer ( 4 Gases, 8 Signal Points ) Latching Alarm Relays % + # Description 4-20 ma Dry Contact ( 8 Signal Points ) Transducer ( 4 Gases, 8 Signal Points ) B Blank Slot - 4 Vertical Future Expansion (Blank) Positions in One slot F Future Expansion (Blank) Position List Signal Input Type Choices from Left to Right G = Gas Service A Medical Air B Blank Slot C Carbon Dioxide 50 psig D Carbon Dioxide 80 psig F Future Expansion H Hyperbaric Oxygen I Instrument Air J Carbon Dioxide-Oxygen Mixt DISS 1080 (CO2>7%) K Helium-Oxygen Mixt DISS 1060 (He>80%) L - Helium N Nitrous Oxide O - Oxygen R Tri-Gas S - AGSS T - Nitrogen V Medical Vacuum W- WAGD/EVAC C = Gas Set U Tri-Tech Labeled NFPA Color Code English I Tri-Tech Labeled ISO Color Code English/French N = Number of Slots on Front Panel 2 2 Slot Front Panel 3 3 Slot Front Panel 5-5 Slot Front Panel T = Type of Alarm T - Touch Screen Digital Page 3 of 5

Part No. Examples T2UOVPF = 2 Slot Combination Alarm Tri-Tech labeled, set for NFPA Color Code English gas set. Oxygen, Medical Vacuum, 8 Dry Contact Signal Points w/ Non-Latching Alarm Relays and one Future Expansion. T3IOAVF+++F = 3 Slot Combination Alarm Tri-Tech labeled, set for ISO Color Code English/French gas set. Oxygen, Medical Air and Medical Vacuum, one Future Expansion {Blank) position at bottom, of 2 nd slot 24 Dry Contact Signal Points w/ Latching Alarm Relays and one Future Expansion (Blank) position at bottom of 3 rd slot. T5UOAVFPPPPQQQQB = 5 Slot Combination Alarm Tri-Tech labeled, set for NFPA Color Code English gas set. Oxygen, Medical Air, Medical Vacuum and one Future Expansion (Blank) position at bottom of 2 nd slot. 64 Dry Contact Signal Points w/ Non-Latching Alarm Relays in 3 rd slot, 32 Transducer Signal Points W/ Non- Latching Alarm Relays in 4 th slot, 5 th slot is Blank. Page 4 of 5

Page 5 of 5