Gaseous detectors. 1

Similar documents
Experimental Particle Physics PHYS6011 Joel Goldstein, RAL

February 12, PHY357 Lecture 11. Experimental Methods. Accelerators. Particle Interactions. Particle Detectors. Full experiment (eg.

Today s Outline - September 12, C. Segre (IIT) PHYS Fall 2016 September 12, / 21

Technical Manual Ionisation Chamber MUSIC80

Particle Detectors. Summer Student Lectures 2007 Werner Riegler, CERN, History of Instrumentation History of Particle Physics

Aging tests with GEM-MSGCs

The LHCb Outer Tracker: Production & Ageing studies

LECTURE 11. Dr. Teresa D. Golden University of North Texas Department of Chemistry

New spark-protected GEMlike detectors with resistive electrodes

Solid State Detectors. T. Bowcock

Aging measurements on triple-gem detectors operated with CF 4 based gas mixtures

RECENT DEVELOPMENTS ON MICRO-PATTERN GASEOUS DETECTORS TECHNOLOGY

Chemistry Instrumental Analysis Lecture 4. Chem 4631

TPC BASED IMP. Limin Duan, Gas Detector Group. Institute of Modern Physics, CAS TPC tracker detector technology mini-workshop

Status Report about the TPC Detector and Module at CEPC

Micromegas in a Bulk

A GEM-based detector for detection and imaging sparks and flames G. Volpe 1 and V. Peskov 2. Abstract

Application of gaseous detectors for full-field EDXRF imaging of works of art

arxiv: v1 [physics.ins-det] 10 May 2017

Antoine Drouart DSM/DAPNIA/SPhN. Köln, 5 th of December,

Instrumental technique (MCP detectors) - Kamalesh ( )

IFIN-HH, Hadron Physics Department infrastructure for ALICE TPC upgrade

Aging Analysis of Micromegas Detectors for ATLAS New Small Wheel

Commissioning of the Silicon Drift Detector of the ALICE experiment at the LHC

50(394) IEC. ore content meter ore sorting equipment overvoltage (of a Geiger-Müller county tube)...

Tracking Detectors 2 silicon for HEP

GEM detectors activity at the Laboratori Nazionali di Frascati of INFN

Competency: Critically evaluate the function of the ionisation chamber in the Linear Accelerator and its importance for correct treatment delivery

LARGE VOLUME HEMISPHERICAL NUCLEAR RADIATION DETECTOR CZT/500(S)

A Fast Multi-GEM based Detector for High-Rate Charged-Particle Triggering

Portable Survey Instruments NISP-RP-01

I Seminario Nazionale Rivelatori Innovativi

Identify the following features and specifications for the ESP1

Discharge reduction technologies for Micromegas detectors in high hadron flux environments

NUCLEAR INDUSTRY STANDARD PROCESS Radiological Protection. Level 3 Information Use

Temperature dependent Components within the Pyroelectric Detector

A Triple GEM Detector for the central Region of Muon Station 1

Measurement system for detection of Radon in building

The Effect of Different Fill Gases on Bragg Curve Spectrometer Detectors.

Semiconductor Radiation Detectors

Electron detectors. The most used signals in a scanning electron microscope in standard conditions are the electrons and in particular the SE.

Pyroelectric Infrared Sensor PIR

The Effect of Different Fill Gases on Bragg Curve Spectrometer Detectors.

ESTB Proposal from the Stony Brook University: Test of a RICH-Prototype Based on CsI-GEMs

Freiburger Materialforschungszentrum SG Material Characterization & Detector Technology. Albert-Ludwigs-Universität Freiburg

Saintech Ion Beam Systems - Ion Current Monitor

Customized MCPs for Analytical Instruments. An MCP for every application

Development of a Novel Contamination Resistant Ionchamber for Process Tritium Measurement and use in the JET First Trace Tritium Experiment

dewar or an electrically powered cooler. The sensitive detector surfaces are thus protected from moisture and condensable contaminants.

Lecture 5 Particulate emission control by electrostatic precipitation

Aging Tests of of Full Scale CMS Cathode Strip Muon Chambers

R&D Fast and lightweight EIC integrated tracking system Barrel MicroMegas & Forward Triple-GEM

818 Series. Low-Power Detectors ABSOLUTE SPECTRAL RESPONSIVITY (A/W)

Single-sided CZT strip detectors

arxiv: v1 [physics.ins-det] 2 Oct 2013

Silicon Sensors S. MIDDELHOEK S.A. AUDET

Vacuum Systems & Thin Films (Jaeger 6, Campbell 10&12, Ruska 7) Deposition of thin films involves vacuum system Direct Deposition from Source

EVAPORATION NEEDS ENERGY

Technical Description. Ionisation Chambers

DAWN Electronics, Inc.

Diamond Windows & MCP Detectors for Synchrotron Applications

DR.RUPNATHJI( DR.RUPAK NATH )

Alphasense Application Note AAN 106 HUMIDITY EXTREMES: DRYING OUT AND WATER ABSORPTION

Micro Pattern Gas Detectors (MPGD) fabrication processes

High Throughput Large Area Silicon Drift Detectors

Texas active target (TexAT) detector part 1: Design and construction progress

MOXTEK MULTI-LAYER OPTICS AND X-RAY TECHNOLOGY

DEVELOPMENT OF LARGE AREA PLASTIC SCINTILLATION DETECTOR FOR RADIOACTIVE CONTAMINATION MONITOR

LCTPC Setup at the DESY Testbeam. AWLC 14, Fermilab , R. Diener, DESY

M O D U L E IV C O N T R O L T E C H N I Q U E S

Chemistry Instrumental Analysis Lecture 14. Chem 4631

Unique Solutions for Your Research. Introduction to Velocitas VMI. Contents.

CONTAMINATION MONITORING INSTRUMENTATION RCT STUDY GUIDE LEARNING OBJECTIVES:

Sensors and. Transducers. D. Patranabis. Second Edition

Using downhole probing for real-time grade estimation for uranium exploration and mining

UVC Detection as a Potential for Alpha Particle Induced Air Fluorescence Localisation

Optical Sensor (2A) Optical Sensor Type Optical Sensor Characteristics. Young Won Lim 9/12/09

Diamond Detectors Ltd. Fabrication and Packaging Capabilities. Kevin Oliver

Low-Power Detectors. Key Features. 818 Series. Power levels from pw to 2W. Wavelengths from 190 nm to 1800 nm. Proprietary detector optics

2008 JINST 3 S The CMS experiment at the CERN LHC THE CERN LARGE HADRON COLLIDER: ACCELERATOR AND EXPERIMENTS.

Performance Requirements for Monitoring Pulsed, Mixed Radiation Fields Around High-Energy Accelerators

HYDROGEN DETECTION INSTRUMENTATION FOR THE SIAC HYDROGEN RECOMBINERS. Summary

Vacuum Systems & Thin Films (Jaeger 6, Campbell 10&12, Ruska 7) Deposition of thin films involves vacuum system Direct Deposition from Source

Table of Contents. Page Quick Set-up Instructions 1. Theory of Operation 2. Specifications 3. TM 1003W Controls and Connections 4

818 Series Low-Power Detectors

Recent Advances in Large Area Micro-channel Plates and LAPPD TM

Status of the PRad Experiment (E )

A Beam radiation monitor for the SVT based on CVD diamond sensors!

The CDF Muon Detectors

Instruction Manual. gray Mounted Detectors. for. greenteg AG Technoparkstrasse Zürich, Switzerland T: F:

UV-Ozone Technology and Applications

SILICON SENSOR ACTIVITIES AT SINTEF MINALAB. Marco Povoli, Angela Kok, Ozhan Koybasi

Radiation Safety issues for the PF-AR in KEK

Distributed Temperature Sensing

SILICON DRIFT DETECTORS

CHAPTER 2 EXPERIMENTAL APPARATUS AND PROCEDURES

MODEL-BASED OPTIMIZATION OF AN INFRARED GAS SENSOR

Measuring Couch Motion Accuracy with X-ray Beam Fluence

MD456: Monitoring of abort gap population with diamond particle detectors at the BGI in IP 4

Division III Process Gas Treatment Equipment

Transcription:

Gaseous detectors Very important class of detectors with many applications charged particle, x- & gamma rays, visible light detection applications in many areas of research, & commercial particle & nuclear physics space-borne astro-particle physics medical imaging x-ray crystallography environmental monitoring long history Investigations of ionisation of gases and spark discharges Crookes (?), Townsend (fluorescent light) Geiger-Muller tube Very large systems in particle physics experiments Position sensitive detectors most useful kind rapid progress following developments in 1960 s [Nobel: G. Charpak 1992] g.hall@ic.ac.uk www.hep.ph.ic.ac.uk/~hallg/ 1

Principle of gas detector Ionisation of gas -> electron-ion pair Drift of charges in electric field Avalanche multiplication of charges by electron-atom collision in high E field applying voltage across gas - geometry dependent Signal induction via motion of charges Typical properties signal sizes are for a high energy particle crossing the detector for photons use absorption length (photoionisation cross-section) Gas E average per ion pair (ev) Ionisations per cm Free electrons per cm He 28 5 16 Ne 36 12 42 Ar 25 25 103 Xe 46 340 CH 4 28 27 62 CO 2 33 35 107 g.hall@ic.ac.uk www.hep.ph.ic.ac.uk/~hallg/ 2

Signal vs applied field E range I II III IV Behaviour recombination of part or all of signal constant signal = total ionisation deposited no recombination impact ionisation during charge collection signal proportional to initial ionisation linear response V = Anq/C A > 10 4 - depends on gas impact ionisation but gain depends on initial signal size A = A(n) For high fields ion (space) charge builds up around anode reduces gain - eventually saturates Geiger region - photon emission spreads avalanche throughout chamber longer recovery time between pulses Ionisation chamber Proportional region Limited proportional region V Pulse size independent of initial ionisation Geiger region VI Continuous discharge C = chamber capacitance n = initial no e-ion pairs g.hall@ic.ac.uk www.hep.ph.ic.ac.uk/~hallg/ 3

NB Signal vs applied field Although the figure shows behaviour vs voltage, it is actually the electric field which determines the behaviour, ie. these results are for a particular geometry: cylindrical (?) It is the general form which counts, not the absolute voltage g.hall@ic.ac.uk www.hep.ph.ic.ac.uk/~hallg/ 4

Amplification process Electrons acquire kinetic energy from E field, so gain velocity but subject to collisions with gas molecules where energy lost mean free path between collisions typically ~ few µm If KE gained > ionisation energy -> impact ionisation = amplification dn = αndx α = f(e, pressure) ~ Ape -Bp/E first Townsend coefficient N ~ N 0 e αx Impact ionisation also often produces photons can spread and produce further ionisation, far from original site organic "quenchers" absorb photons well molecules have many excited states (vibration/rotation) Choice of gas for detector - wide choice & big subject low working voltage, high gain operation, good proportionality, drift speed,... noble gases + organic quencher some elements have strong affinity for electrons (O 2, Cl, F) - avoid g.hall@ic.ac.uk www.hep.ph.ic.ac.uk/~hallg/ 5

Cylindrical chamber Simplest design - long wire (r=a), with surrounding conducting cylinder (r=b) q = charge per unit length of wire Gauss law E.n.dS=q/ε 0 E(r) = q/2πε 0 r = -dφ/dr solve for q, with φ(b) = 0 φ(r) = V 0 - (q/2πε 0 )ln(r/a) possible values V 0 = 4000V b = 1cm a = 50µm Argon mixture with E ion = 30eV E(r) = 755V/r if mfp 10µm then ee(r) x mfp > 30eV at r = 250µm although ionisation only begins close to the wire there are still ~ 20 mfp so each electron can multiply many times, in ~ nsec dn = αndx N ~ N 0 e 20 α ~ 1 A ~ 10 8 Many variants on design possible, including multi-wire designs g.hall@ic.ac.uk www.hep.ph.ic.ac.uk/~hallg/ 6

Multi-Wire Proportional Counter MWPC anode wires equally spaced at few mm spatial resolution of few x100µm operated in proportional region Cathode Anode wires E field lines each anode acts as independent counter measure many particles simultaneously confinement of charge by field energy measurements possible Cathode = wire spacing very often used as binary position detector => σ meas = / 12 measure signal amplitudes on each wire and form weighted sum centre of gravity method care in construction needed to place wires accurately displacements lead to field distortions multiple (y/z/u) layers in single chamber give coordinate measurements g.hall@ic.ac.uk www.hep.ph.ic.ac.uk/~hallg/ 7

Drift chamber constant (low) electric field in region to ensure constant velocity avalanche begins only close to single wire time measurement -> distance t avalanche -t 0 = L/v +V drift region external "trigger" defines t 0 charged particle -V 1 for accurate measurement anode in-situ calibration of drift velocity typical v e ~ 5 cm/µs v ion ~ v e /1000 field shaping wires scintillation counter to provide trigger -V 1 measurements σ < 100µm achieved over few cm drift path Large chambers wires configured in electrostatic "cells" small cells maximise capability to operate in high intensity conditions g.hall@ic.ac.uk www.hep.ph.ic.ac.uk/~hallg/ 8

Actual chambers Manufacture under clean conditions wide range of designs and geometries g.hall@ic.ac.uk www.hep.ph.ic.ac.uk/~hallg/ 9

Time Projection chamber chamber is large cylindrical volume with high voltage at central plane electrons drift to ends - up to few m diffusion via molecular collisions would spread signal over wide area apply B-field parallel to E-field - 3-d measurement wanted anyway for momentum measurements limits diffusion, e - s follow helical paths achieved in very large volumes σ T < 200µm σ L < 500µm Negative high voltage electrode Wire chambers Electron drift Wire chambers E B Beam pipe E E g.hall@ic.ac.uk www.hep.ph.ic.ac.uk/~hallg/ 10 B

Pulse formation in gas detectors Signal is due to induction - positive ions dominate - why? charge Ne migrates toward anode, crosses voltage drop φ Work done transporting charge => charge movement in external circuit Ne = V 0 q V 0 = bias voltage - constant since avalanches form close to anode, electrons do not experience large φ ions drift across whole chamber so feel full bias voltage applied to chamber Ion velocity is much less than electrons fast rise to signal pulse drift time to cathode to induce full signal charge: pulse duration ~ µs - ms This is a drawback for some applications => designs to shorten ion drift paths while operating at the same voltage amplifiers with pulse shaping so slow part of signal can be ignored or cancelled g.hall@ic.ac.uk www.hep.ph.ic.ac.uk/~hallg/ 11

MicroStrip Gas Chambers (MSGC) Etch anode and cathode strips on same glass substrate avalanche region near glass surface very accurate postions of strips control gain short distance for e - & ions to travel fast signal, with short tail duration (for charged particle signals) due to drift time in gas x-ray Note uniform E field in drift region charged particle g.hall@ic.ac.uk www.hep.ph.ic.ac.uk/~hallg/ 12

MSGC properties Excellent performance in high intensity (rate) conditions gain reduction over time due to substrate charging coatings and low resistivity to avoid also good energy resolution gain reduces with intensity in MWPC because of space charge build-up g.hall@ic.ac.uk www.hep.ph.ic.ac.uk/~hallg/ 13

Gas discharges bane of all gas detectors and troublesome for MSGC charge is stored in capacitances high E fields present discharge typically initiated by heavily ionising nuclear events manufacturing defects deposits on electrodes discharge is very fast ~ns difficult to predict or prevent g.hall@ic.ac.uk www.hep.ph.ic.ac.uk/~hallg/ 14

Gas Electron Multiplier GEM Principle Thin polymer (kapton) foil ~50µm thick, metal clad on both sides perforated by large number of holes, ~10 4.cm -2, using photolithographic process & etching Apply voltage ~500V across foil V g.hall@ic.ac.uk www.hep.ph.ic.ac.uk/~hallg/ 15

GEM electric field amplification field is mainly confined to holes reduces risk of complete discharge between electrodes g.hall@ic.ac.uk www.hep.ph.ic.ac.uk/~hallg/ 16

GEM operation Operated in proportional region amplification in holes in foil but gain is property of foil, so little dependence on external fields good mechanical tolerance signal on readout electrode due only to electron collection no slow ion tail or ion feedback can cascade several GEM foils to increase gain cathode separation of gas amplification and readout = flexible design readout electrodes at OV anodes g.hall@ic.ac.uk www.hep.ph.ic.ac.uk/~hallg/ 17

Problems with gaseous detectors Excellent and widely used detectors, but some drawbacks Detectors with built-in amplification need careful attention small voltage changes can produce large variations in gain gas quality must be carefully monitored impurities can affect operation or, worse, cause long term damage sparking, if chamber enters discharge region - noise and damage gas amplification has inherent instability - for long term use organic molecules break into new polymers radiation damage deposits in gas can build up on wires wires strung in chambers must be maintained under tension gravitational displacement can distort field in multi-wire systems long thin wires subject to stress, particularly at connection points Nevertheless, large chambers with 10 4 wires or more have been operated successfully for many years g.hall@ic.ac.uk www.hep.ph.ic.ac.uk/~hallg/ 18