Instruction Sheet. Important Note: Information that requires special emphasis. Note: Information that supplements points in the main text.

Similar documents
Instruction Sheet. DANGER Indicates a potentially or imminently hazardous situation which, if not avoided, will result in death or serious injury.

INSTALLATION INSTRUCTIONS

INSTALLATION INSTRUCTIONS

INSTALLATION INSTRUCTIONS

Installation Instructions PANOS 5.25

PANOS. Installation Instructions Square, 5.25 Round Flangeless, Downlight. The following installation instructions relate to the options below

OPERATING INSTRUCTIONS

Installation Instructions

Tank, Heater, and Control Operation Manual REV. E DATE: 04May2015

Installation Instruction. Puritan Bennett. Wall Air Water Trap For 800 Series Ventilator Basic Cart. Pla

HEARTH PRODUCTS KITS AND ACCESSORIES. KIT CONTENTS (See Figure 1) TOOLS NEEDED

INSTALLATION INSTRUCTIONS

Stopper Station Series SS-2008 Push Button Pneumatic Timer Button

200i User s information manual

Instant Hot Water Dispenser Owner s Manual

installation and start-up instructions HUMIDIFIERS AND HUMIDISTAT

QAA 142 Termination instructions Self-regulating trace heaters type ELSR-NA and -HA in Div1 with heater entry kit EL-HAZELECT Kit part No: 09CA051

P-K MACH GAS-FIRED BOILER C1500H C2000H OUTDOOR BOILER KIT GUIDE

Installation Instructions. For the 18 Built-In Dishwasher and Front Color Panels

Alpine CONDENSING HIGH EFFICIENCY GAS-FIRED HOT WATER BOILER

Installation Instruction

APEX CONDENSING HIGH EFFICIENCY GAS-FIRED HOT WATER BOILER

INSTALLATION INSTRUCTIONS

INSTALLATION INSTRUCTIONS Cased N Coil, Horizontal ENH4X

DISHWASHER. Models DW2432 and DW2432SS. Installation Manual. Write Serial Number (on inner door of unit) here:

LED Seamless VaporProof Fixture

Installation & Operation Manual. AromaFlo. mr.steam. Feel Good Inc.

INSTALLATION INSTRUCTIONS

INSTALLATION INSTRUCTIONS

INSTALLATION INSTRUCTIONS

OWNER S MANUAL. Models: Herkules Equipment Corporation 2760 Ridgeway Court Walled Lake, MI USA Fax

SAVE THESE INSTRUCTIONS

CR330 Owner s Manual

BUILT-IN DISHWASHER INSTALLATION INSTRUCTIONS

Installation Instructions

Operator Manual Autofill System

GBX GAS BOILERS USER S INFORMATION MANUAL

User Manual. Glass Countertop Mechandiser Refrigerator. Single Door Model: 360GSM3HCB, 360GSM3HCW Pass-Through Model: 360GSM32HCB, 360GSM32HCW

Mortex INSTALLATION INSTRUCTIONS AIR CONDITIONING & HEAT PUMP INDOOR COILS

installation manual DM3862 RM2554 DM2663 RM2551 Refrigerator for LP-gas & electric operation DM2852 ! AVERTISSEMENT ! WARNING FOR YOUR SAFETY

Installation Guide. 15 W. Undercounter/Freestanding Nugget Ice Machine U L. Viking Range, LLC. 111 Front Street

INSTALLATION INSTRUCTIONS

Installation Instructions

Service Procedures for the Model SL31-31 Dryer

TECHNICAL DATA. Wet 26a. February 22, 2009

INSTALLATION INSTRUCTIONS

NATURAL GAS CONVERSION KIT

Heat Exchanger Block Replacement Instructions

TECHNICAL INSTRUCTIONS

DRAIN WATER TEMPERING INSTALL

INSTALLATION INSTRUCTIONS

SEL-3422 Motor Relay HMI Quick-Start Guide

Installation Instructions

HELIATOS Bottom Feed Connector Installation Instructions Introduction. Step 1

HX Field Replacement Kit

Installation Instructions

PUREFIRE GAS BOILERS USER S INFORMATION MANUAL PB HEAT, LLC 131 S. CHURCH ST BALLY, PA WARNING MISE EN GARDE TO BE COMPLETED BY THE INSTALLER

2100 Series Laboratory Turbidimeters. Pioneering technology in turbidity monitoring.

USER S INFORMATION MANUAL PB HEAT, LLC 131 S. CHURCH ST BALLY, PA WARNING MISE EN GARDE TO BE COMPLETED BY THE INSTALLER.

DOC JAN06 LT 200

Installation Instructions

Built-In Dishwasher. Installation Instructions. BEFORE YOU BEGIN Read these instructions completely and carefully. IMPORTANT The dishwasher MUST be

Installation Instructions for P/N: Algas-SDI thermostat retrofit kit

Installation Instructions

INSTALLATION INSTRUCTIONS TXV Horizontal Slab Coils WLSH

Installation Instructions

Installation Manual PS-225 & PS-275

24 and 30 Gas Cooktops Tables de cuisson à gaz de 24 et 30 Superficies de cocción a gas de 24 y 30

Installation. S19-320A (6' cable) S19-320D (12' cable) Flow Switch Alarm System. Table of Contents

DISCONTINUED. Installation. S19-320A (6' cable) S19-320D (12' cable) Flow Switch Alarm System. Table of Contents

Air Conditioner

SS1095 Ice Maker.

TECHNICAL INSTRUCTIONS

Installation Instructions

Owner's Manual. WS Series. Water Softener

SIDEBURNER WARRANTY REPAIR KIT FOR OUTDOOR USE ONLY Repair kit For use with Char-Broil RED Series Grills Only.

LGB Gas fired boiler

SAFE DRINKING WATER AND TOXIC ENFORCEMENT ACT

INSTALLATION INSTRUCTIONS

Installation Instructions

Installation Instructions

INSTALLATION INSTRUCTIONS TXV Coils for Manufactured Housing EMA

INSTALLATION INSTRUCTIONS TXV Horizontal Duct Coils EHD

An LP-cylinder not connected for use shall not be stored in the vicinity of this or any other appliance. WARNING

USER'S INFORMATION MANUAL ES2 GAS - FIRED

Installation/Care/Use Manual EZH20 In Wall Bottle Filling Station

FMP-ULS UNIVERSAL LIQUID SENSOR INSTALLATION GUIDE

Installation Instruction

Installation, Operation, and Maintenance Manual RTE14S & RTE14S-2. For the Taco Bell "Rethermalizer" Model Numbers

Dust Collector. Sizes DC1 and DC2 U S E R G U I D E UGC

Kit Instructions COMMON VENT KIT (PVC AND DURAVENT)

Installation Instructions

TABLE OF CONTENTS. NOTE: Read the entire instruction manual before starting the installation. TROUBLESHOOTING... 13

Installation Instructions

Installation GUIDE VDWU524SS VDWU524WSSS FDWU524WS FDWU524 VDWU324SS FDWU324

OWNER S MANUAL INSTALLATION OPERATION MAINTENANCE CONVECTION OVEN MODEL # RDCO-32

INSTALLATION INSTRUCTIONS Cased N Coil, Horizontal ENH4X

USER'S INFORMATION MANUAL SERIES 2 GAS - FIRED

TEMPERATURE CONTROLLERS

PERFECT FIT IN-DASH HEAT/ COOL/ DEFROST PLYMOUTH BELVEDERE

Transcription:

Instruction Sheet 4500089 Converting the Surface Scatter 7 sc Turbidimeter to SS7 sc-hst Safety information Please read this entire document before unpacking, setting up, or operating this equipment. Pay attention to all danger and caution statements. Failure to do so could result in serious injury to the operator or damage to the equipment. To ensure that the protection provided by this equipment is not impaired, do not use or install this equipment in any manner other than that specified in this document. Use of hazard information DANGER Indicates a potentially or imminently hazardous situation which, if not avoided, could result in death or serious injury. CAUTION Indicates a potentially hazardous situation that may result in minor or moderate injury. Important Note: Information that requires special emphasis. Note: Information that supplements points in the main text. Precautionary labels Read all labels and tags attached to the instrument. Personal injury or damage to the instrument could occur if not observed. A symbol, if noted on the instrument, will be included with a danger or caution statement in the manual. Electrical equipment marked with this symbol may not be disposed of in European public disposal systems after 12 August of 2005. In conformity with European local and national regulations (EU Directive 2002/96/EC), European electrical equipment users must now return old or end-of-life equipment to the Producer for disposal at no charge to the user. Note: For return for recycling, please contact the equipment producer or supplier for instructions on how to return end-of-life equipment, producer-supplied electrical accessories, and all auxiliary items for proper disposal. CAUTION Read and follow these instructions carefully to avoid personal injury and damage to the instrument. Only personnel qualified to do so should conduct the installation/maintenance tasks described in this instruction sheet. ATTENTION Lire et suivre ces instructions attentivement pour éviter les risques de blessures de personnes et de détérioration de l appareil. Les tâches d installation et d entretien décrites dans cette partie de la fuelle d instructions doivent être seulement effectuées par le personnel qualifié pour le faire. ADVERTENCIA Lea y siga cuidadosamente estas instrucciones a fin de evitar accidentes personales y daños al instrumento. Las tareas de instalación y mantenimiento descritas en esta hoja de instrucciones deberán ser efectuadas únicamente por personas debidamente cualificadas. 1

WARNHINWEIS Lesen und befolgen Sie diese Instruktionen aufmerksam, um Verletzungen von Personen oder Schäden am Gerät zu vermeiden. In diesem Abschnitt beschriebene Installations- und Wartungsaufgaben dürfen nur von qualifiziertem Personal ausgeführt werden. AVISO Leia e siga atentamente as presentes instruções para evitar ferimentos pessoais e não danificar o instrumento. As tarefas de instalação/manutenção descritas nestas instruções só poderão ser executadas por pessoal qualificado para o fazer. Benefits Unpacking Installation The SS7 sc-hst was developed to measure turbidity in samples at high temperatures. The SS7 sc-hst uses fittings molded from black polypropylene to minimize leakage. These fittings can withstand higher working temperatures than standard PVC. A flow multiplier pulls moisture-laden air out of the enclosure and reduces condensation. The flow multiplier is powered by compressed air and captures the vapors before they can condense inside the instrument enclosure. The vents in the enclosure act as air inlets rather than air outlets. Remove the kit from the shipping container and inspect each item for damage. Verify that all items are included. If any items are missing or damaged, contact Hach Customer Service in Loveland, Colorado. The Hach Company toll free phone number for customers within the United States is 800-227-4224. Customers outside the United States should contact their regional Hach office or distributor. Note: This kit will only work on an SS7 sc Turbidimeter that has a molded slant tube (one continuous piece). If the turbidimeter has a fabricated slant tube (made from PVC components), order a molded one (Cat. No. 4500200) and install it before proceeding. DANGER All electrical connections should be made by a qualified technician to assure compliance to all applicable electrical codes. DANGER Tous les branchements électriques doivent être effectués par un technicien qualifié, afin d'assurer le respect de toutes les normes électriques applicables. PELIGRO Todas las conexiones eléctricas deben ser realizadas por un técnico capacitado para asegurar el cuplimiento con todos los códigos eléctricos. GEFAHR Um zu gewährleisten, daß alle elektrischen Anschlüsse den VDE-Vorschriften und gegebenenfalls den Zusatzvorschriften der zuständigen Elektrizitätswerke sowie anderen zutreffenden Sicherheitsnormen entsprechen, dürfen diese Anschlüsse nur von geschultem Fachpersonal hergestellt werden. PERIGO devem ser realizadas por técnico qualificado para garantir o cumprimento de todas as normas elétricas aplicáveis. 2

1. If upgrading a standard SS7 sc (or older) turbidimeter, remove the hose fittings from the slant tube and the bulkhead adapters along with the hoses inside the unit. Then remove the bulkhead adapters at the bottom of the unit and proceed to step 2. If upgrading an older SE (severe environment) turbidimeter, complete the steps above and then disconnect power to the heater (if existing). a. Remove the heater and the control box and disconnect the wiring to the power source. b. Remove the mounting screws from the electrical control box, the heater, and the adapter to the enclosure. Discard the screws. c. Plug the hole used for the electrical connections with a small, oil-tight hole seal. d. Thoroughly clean the unit and make any required repairs to it. 2. Using a 15 /8-inch hole saw, drill a 1.625 inch diameter hole on the upper-left side of the unit as shown in Figure 1 and Figure 2 (this hole may already exist). 3. Install the ¾-in. gray PVC bulkhead fittings in the new hole as shown in Figure 2. 4. Install the white polypropylene bulkhead fittings in the bottom of the unit as shown in Figure 2. 5. Install the new (black) hose adapter fittings using Teflon tape and tighten them properly with a wrench. Orient the ¾-in. elbows as shown in Figure 2. 6. Install the hoses between the slant tube and the bulkhead fitting on the bottom of the unit. Attach the hoses with the stainless steel clamps. Note: Hose is furnished in one piece. Cut each piece to desired length. 7. Install the ¾-in. hose barb elbow in the vent cover and the straight hose fitting in the PVC bulkhead fitting leading to the flow multiplier. Teflon tape is not required. 8. Loosen the screws that hold the light source in place before installing the vent cover. Figure 1 Location of 1.625-inch diameter hole (side view of instrument case) 3

9. Install the vent cover on the top of the slant tube using the threaded disk and ¼-inch screw. Tighten until snug. 10. Tighten the screws that hold the light source. 11. Check the alignment of the light source after installing the cover and readjust if necessary, using the new alignment plate included in the kit. See the Surface Scatter 7 sc Turbidimeter User Manual (Cat. No. DOC026.53.00769) for alignment instructions. 12. Connect the elbow in the vent cover and the straight hose fitting in the PVC bulkhead adapter with a section of ¾-in. hose. No clamps are required. 13. Screw the flow multiplier into the outside of the PVC bulkhead adapter as shown in Figure 2. 14. Connect a filter/pressure regulator to the flow multiplier. (The filter/pressure regulator, tubing, and tubing fittings are not provided in this kit.) The flow multiplier has a 1 /8-in. NPT male fitting. A ¾-in. hose should be connected to the output of the flow multiplier and routed to a suitable place to vent the fumes and/or condensate coming out of the hose. 15. Set the pressure regulator to 20 psi initially. Adjust the flow up or down as needed. 16. The enclosed drain trap should be used to prevent air from being drawn in through the drain. See Figure 2. 17. When reconnecting the hydraulics to the bottom of the unit, hold the ¾-in. bulkhead fittings on the inside of the enclosure with the door open. 18. Replace the label displaying the part number and serial number of the sensor unit (not the control unit) with the label enclosed in the upgrade kit. Thoroughly clean the enclosure surface where the old label was attached and place the new label in the same location. This identifies the product as a reconditioned product (for service and compliance purposes). 19. Calibrate the instrument using the calibration cup as described in the user manual. Operation If condensation forms in the enclosure, increase the air pressure (and flow) by increasing the air pressure setting of the pressure regulator for the flow multiplier. Make sure the bubble trap is working. Bubbles on the surface of the liquid will cause incorrect readings. If deposits accumulate inside the unit, wash the inside with warm water spray. The vent cover at the top of the slant tube can be removed for cleaning if necessary. Loosen the light source before removing the cover. Check the alignment of the light after reinstalling the cover using the new alignment plate included in the kit. Make sure the cover sits flat on top of the slant tube when reinstalled. Calibrate the instrument using the calibration cup as described in the user manual. Note: Do not operate the instrument without the cover. Note: Do not operate the instrument unless the flow multiplier is also operating. 4

Figure 2 Surface Scatter 7 sc and HST upgrade components 5

Replacement parts Description Quantity Catalog number SS7 sc Turbidimeter HST Upgrade Kit 1 4500043 Includes: Clamp, hose, worm drive 316 SST for ¾ in. ID tubing 4 4666700 Clamp, hose, worm drive 316 SST for 1 in. ID tubing 2 4666800 Cover, vent SS7 sc 1 4029400 Fitting, bulkhead, ¾ in. NPT, polypropylene 2 4031100 Fitting, bulkhead, 1 in. NPT, polypropylene 1 4035500 Fitting, ¾ in. NPT, modified street elbow (flow multiplier) 1 4030100 Fitting, hose barb, 1.00 OD x 1 NPT 3 4037200 Fitting, elbow, ¾ OD tube x ¾ NPT 3 4049600 Fitting, hose barb, ¾ OD x ¾ NPT, gray 6 4043900 Fitting, bulkhead ¾ FPT x ¾ FPT PVC 1 4501100 Instructions, SS7 sc HST upgrade 1 4500089 Label, serial number 1 4507200 Orifice, flow multiplier 1 4030200 Screw ¼-20 x 1 PAN SST PHIL 1 785811 Seal, oil-tight 0.88 in. diameter hole, street 1 4221000 Tubing Norprene, ¾ in. ID 5 ft. 4666500 Tubing Norprene 1 in. ID 1 ft. 4666600 Template, light source alignment 1 4507600 Threaded Disk 1 4029900 Trap Assembly, 1 in. CPVC 1 4052600 Accessories Description Quantity Catalog number Auto Flush Kit (120 VAC) 1 4669212 Auto Flush Kit (220 VAC) 1 4669222 Bubble Trap/Head Regulator 1 4668000 Flow Meter, 100 1600 ml/min. 1 4028400 Heat Exchanger Unit (sample cooler) 1 4855100 FOR TECHNICAL ASSISTANCE, PRICE INFORMATION AND ORDERING: In the U.S.A. Call toll-free 800-227-4224 Outside the U.S.A. Contact the HACH office or distributor serving you. On the Worldwide Web www.hach.com; E-mail techhelp@hach.com HACH COMPANY WORLD HEADQUARTERS Telephone: (970) 669-3050 FAX: (970) 669-2932 Hach Company, 2006 Edition 2 August 2006 jk/dk