TC100N. Fixed Gas Detector Operation Manual

Similar documents
QD6310. Fixed Gas Detector Operation Manual

CP01 Fixed gas detector

IMR EX610. Portable Gas Detector. Operation Manual

E1000 Portable gas detector

IMR IX176 Portable Gas Detector User Manual

M004 combustible gas module

User s Guide HHAQ-109. Portable Multi-gas Detector. Shop online at omega.com SM

GL-CO-RFG Channel Gas Leak Alarm System

KAE Gas Alarm Operation Manual

Operation and Maintenance Manual

3years. Detection of explosive gases, toxic gases and oxygen. Infrared XP IR version. SIL 2 high reliability IP 66.

SENSE Instruments CO., LTD. Data sheet Gas detector Page 1 / 5. Gas detector

Detection of explosive gases, toxic gases or oxygen. Infrared XP version. SIL 2 high reliability IP 66. Certifications. The Fixed Gas Detection People

CMF, CMX & CMA series

Flammable gases Detectors. Model DAX 03F DAX 03F- C DAX 03F- H. Note for installation, use and maintenance. Version 1.2

Models TX-KE and TX-KP Toxic Gas Transmitters

EST User Manual. (Please read this manual carefully before using)

COMBUSTIBLE GAS LEAK DETECTOR PEN

Warranty Registration

OLC(T) 100 FIXED GAS DETECTOR

DESIGN MANUAL MODEL GIR900 INFRARED COMBUSTIBLE GAS SENSOR

Dräger X-am 5000 Multi-Gas Detection

Detection of explosive gases, toxic gases or oxygen. Infrared XP version. SIL 2 high reliability IP 66. Aluminium or Stainless Steel version

BST-MG01 Single Gas Detecting Alarm Manual Instruction

Transmitters/Sensors. OLCT Series

Safe Operation Manual. Honeywell IQ Force Gas Detector

KAB Gas Alarm Operation Manual

P R O D U C T S P E C I F I C A T I O N MSA Ultima X Series Sensor/Transmitter Specification

3M Oldham OLCT 100. Description. Fixed Gas Detector

ULTIMA XL/XT Series Gas Monitors. [ Ultimate Features Extreme Design ]

SIMRAD GD10 IR GAS DETECTOR

REMOTE TOXIC, COMBUSTIBLE, AND OXYGEN GAS SENSORS

4-20mA CYBER Cyber Transmitter for flammable, toxic and IR gas detection Cyber Head Increased security in ATEX certified head

SAFETY CERTIFIED MODEL FP-700 COMBUSTIBLE GAS DETECTOR

BEP Marine Gas Detector 600-GDRV

MODEL: SP12C7. (Portable 4 gas detector) Operating Manual

OLCT 10N. Digital Gas Detector. User Manual. The Fixed Gas Detection Experts. Part Number: NPO10NEN Version: G.1

Installation and Wiring 1.

OLCT 100 DETECTION DE GAZ FIXE. Certifications. Better detection. Better protection. Detection of Combustible, Toxic or Oxygen gases

URD20SE. Gas detectors Carbon dioxide (CO 2 ) ATEX certified (II 2G Ex d IIC T6) Use. Operation. Ordering

SIMRAD GD10P - IR GAS DETECTOR OPERATING MANUAL

WARRANT Y AND LIMITS OF LIABILIT Y Vulcain Inc. warrants to the original purchaser that its product, and the component parts thereof, will be free fro

Dräger X-am 5600 Personal Monitor Multi-Gas Detection Device

Copying contents of this booklet or handing over to third party is strictly forbidden without written permission from Detector Oy.

testo 316-EX gas leak detector Instruction manual

MODEL: SP12C7. (Portable 4 gas detector) Operating Manual

Many different instruments Many different operating principles including: May sample many different gases & vapour From relatively simple to complex

MODEL: SP2nd (Portable Single Gas Detector) Operating Manual

Sieger Sensepoint. Easy to Use Factory set operation Simple to replace Minimal training required

BST-MG07 Multi-gas Detecting Alarm Manual Instruction

GS100M Gas Detector. Technical Specification

Carbon Monoxide (CO) Gas Detector March 15 GCO 300

5 Operating Modes GX Smallest 6 gas sample draw PID library of over 600 VOC s 2 Interchangable smart sensor slots

5 Operating Modes GX Smallest 6 gas sample draw PID library of over 600 VOC s 2 Interchangable smart sensor slots

ESD3000 ESD200 Combustible / Toxic Gas Detector

DESIGN MANUAL. The information in this doucment is applicable to sensors built in September, 2009 or later (Firmware 3.0)

TXgard-Plus. Flameproof Toxic and Oxygen Gas Detector with Non-intrusive One-man Calibration. Installation, Operation and Maintenance Instructions

Washing Machine OWNER S MANUAL TRWTL-70. Before using your washing machine, please read this manual carefully and keep it for future reference.

BEP Marine Gas Detector FD-2

Quality. Water. Water Quality. Testing. Testing. Toxic Gas Detection. Toxic Gas. Detection. Air Quality. Monitoring. Monitoring. Monitoring.

RK / RK Methane Detector Operator s Manual

Signalpoint. The Signalpoint range of flammable, toxic and Oxygen gas detectors offer users a low cost solution to their indoor gas monitoring needs

ZP04 Flammable Gas Detection Module (Model:ZP04) Manual V1.1. Zhengzhou Winsen Electronics Technology Co., Ltd

SIL 2. Ultima X Series Gas Monitors CERTIFIED

EXAMPLES OF HOW TO ORDER

Danfoss gas detection unit Type GD Heavy Duty

EXAMPLES OF HOW TO ORDER

Warranty Registration

We measure it. testo 316-Ex CH 4 testo 316-1

User Manual. GPD 3000 e. gas detector

Sieger Sensepoint. Easy to Use Factory set operation Simple to replace Minimal training required

User Manual. PolyGard CO LC-1112 V3. Electrochemical analog carbon monoxide transmitters serial no. EC-S 002. August 29, 2006

5New features & EXtreme design, now with HART Protocol and DuraSource Technology. Ultima X Series Gas Monitors

Zareba Sensepoint Pro

COSTER. Code Internal Alarm Pre-alarm Suitable sensor threshold threshold external sensor

Dräger X-am 5000 Multi-Gas Detection

Portable Oxygen Monitor OX-07. (Type B) <No Gas alarm type> Operating Manual

Detectors-Transmitters for Flammable, Toxic Gases and Oxygen. Non-intrusive operation via intrinsically safe infrared remote control.

Digital Gas Transmitters with Auxiliary Inputs & Control. PolyGard 2 DC6

Dräger X-am 7000 Multi-Gas Detection

SMART Gas Detector/Transmitter(4~20mA) with built-in LCD & explosion proof DA-500

BX444-Mc Gas Detector

GASGUARD NH 3 2% Ammonia Sensor OPERATING & INSTALLATION MANUAL

GASGUARD H 2-1% Hydrogen Sensor OPERATING & INSTALLATION MANUAL

4P-90 CiTipeL Combustible Gas Sensor Part Number: PM

Dräger X-am 2500 Multi-Gas Detection

Matters Needing Attention of Electrical Equipment Installation Engineering in Hazardous Areas. Wang Qiaoli

Sensepoint. Fit For Purpose Flammable, toxic and Oxygen versions available High performance, low cost Suitable for new and retro fit applications

Sensepoint. Fit For Purpose Flammable, toxic and oxygen versions available High performance, low cost Suitable for new and retro fit applications

ia d ATEX 100 II 2G D MT 99 ATEX G 001 x EX-METER II Series [ Versatility in Explosive Atmospheres]

Sanitary Water Tank INSTALLATION & SVC MANUAL

Xgard. Fixed detectors For flammable, toxic or oxygen gases

RK-05 Combustible Gas Transmitter Operator s Manual

Catalytic Flammable Gas Sensor. (Model: MC105B) User s Manual. Version: 1.3. Valid from: Zhengzhou Winsen Electronics Technology Co.

SIMRAD GD10PE IR GAS DETECTOR

No.THD51E To prevent accidents arising from the use of this product, please ensure the operator using it receives this manual.

GASGUARD LEL 3 OPERATING & INSTALLATION MANUAL

Gas leak detectors. For a quick overview. testo testo testo testo gas detector testo 316-Ex.

ExDetector IRCO2. ExDetector IRHC. Operating and Installation Instructions. Gas Detection and Warning Systems

Operation Manual R205BV, R209BV

Transcription:

TC100N Fixed Gas Detector Operation Manual

Safety Information Before using this device, please carefully read the following safety information: After opening the packing box, please check if there is any crack on the device or spare part missing. If the device found damaged, please don't use it. Please contact the seller. Before using, please make sure the covers are fixed tightly. In order to assure the sensitivity and accuracy, please calibrate zero before the first time of using. Working voltage of the device is 18VDC-30VDC and suggested voltage is 24VDC. Any voltage beyond 30VDC will damage the device. Please don't install the device on the place where there is fast flow passing. Otherwise, it will influence the result. Please don t expose the device always to high concentration gas (higher than the detecting range). Otherwise, it will shorten the sensor life. Please don t paint the gas sensor or detector. Only special designed gas sensor is allowed to be used on the device. Calibrate the gas sensor once at least every 180 days. We suggest the user inspect or test device regularly, so as to avoid the problems of falling dust on the sensor head or unexpected painting. Wire connection must be carried when the power is cut off. Before opening the cover, please make sure there is no mixture gas of combustible gas and air in the environment, so as to avoid any possible explosion. Please avoid water or dust to come into the cover through the wire connection holes. Otherwise, it will easily damage the device. Please avoid the device from electric shock and continuous serious mechanical striking.

calibration cap. The replacement is completed. 8.8 Electrify the detector and calibrate the detector. Installation must abide by the local electric equipment installation regulations. It s forbidden for the user to disassembly, adjust and repair the device or to replace the internal spare parts. All operation inside the device must be carried by professional personnel. Before operation, please carefully read and understand the operation manual. Precautions: To avoid personal safety injury, Instrument damage and potential dangerous accident; do not install the gas detector before reading this manual 9. Normal fault and solution Fault description Reason Solution time-lapse doesn t No response for end Wait till it ends detection gas Sensor broken Change sensor Circuit fault Factory maintain Abnormal connection with Controller Wire connection Fault Open circuit of signal output Circuit Fault Check the wire connection Inspect the open circuit Factory maintain 1. Brief Introduction TC100N fixed gas detector adopts advanced catalytic gas sensor or electrochemical sensor, which can translate the gas concentration in the air to the digital signal output. TC100N adopts three wire 4~20mA electric current signal to connect to the gas monitor panel. The detector working with gas controller can be used to detect gas in refinery, chemical plant, LPG station, boiler room, painting plant and other places with gas exists.

2. Specifications Gas Range Resolution Low alarm High alarm LEL (0~100)%LEL 1%LEL 20%LEL 50%LEL HC vapour (0~100)%LEL 1%LEL 20%LEL 50%LEL O 2 (0~30)%vol 0.1%vol 19.5%vol 23.5%vol CO (0~1000)ppm 1ppm 35ppm 200ppm H 2 S (0~100)ppm 1ppm 10ppm 15ppm CL 2 (0~20)ppm 0.1ppm 5ppm 10ppm SO2 (0~500)ppm 1ppm 10ppm 50ppm NH 3 (0~100)ppm 1ppm 25ppm 50ppm Sensor type: Catalytic or electrochemical gas sensor Sampling: Natural diffusion Working type: Continuously working Error: ±5%F.S. Response time (T90): 30s (LEL) / 60s (toxic gas) Working condition: Temperature: -40~70 (LEL) / -20~50 (toxic gas) Humidity: 95%RH (no condensation) Explosion-proof grade: ExdⅡ CT6 Protection grade: IP55 Working voltage: DC24V±25% Working pressure: 86kPa~106kPa Signal output: 4-20mA 3-line Consumption: 2W/channel Life: Catalytic sensor 3 years; electrochemical sensor: 2 years Connection screw size: G1/2 Cable required: 3-line cable, 3 mm 2 Transforming distance: 1000m Dimension and weight: l b h mm 196 186 68.5 App. 1200g Disconnect the sensor module from the detector. At the same time, hold both S2 and S3 buttons continuously for 1-2 seconds, the green power LED starts to flicker, which indicates the device enters into calibration status. Adjust the first digit of DIP switch to "ON". Red LED starts to flicker and green LED is continuously on. After several minutes, resume the first digit of DIP switch to the original position, and the detector remembers this value as zero. Red LED flickers several times and then is off. Zero calibration is completed. 7.4.2 Single Point Calibration (20.9%VOL) After zero calibration, connect the sensor module to the detector correctly and electrify the detector. After working for 10 minutes, follow the above instruction to enter into calibration status. Adjust the second digit of DIP switch to "ON" (which is 20.9%VOL. Red LED starts to flicker. After several seconds, resume the DIP switch to the original position and the detector will remember this value. Red LED flickers several times and then is off. The calibration is completed. 8. Sensor replacement In normal working environment, the catalytic sensor life is 3 years and electrochemical sensor is 2 years. When the sensor life is overdue or the sensor is damaged, please follow the following instruction to replace the sensor. Before replacing sensor, please first cut off power supply. The user doesn t need to open the detector cover, but only need to open the cover the sensor module. 8.1 Take off the calibration cover in counter-clockwise. 8.2 Pull out the decoration part downwards. 8.3 Screw off the jbckscrew from the sensor protection cover. 8.4 Take off the protection cover in counter-clockwise. 8.5 Take out the old sensor. 8.6 Put the new sensor into the sensor cover. Note: Please put the new sensor into the correct fixing holes. It's forbidden to rotate the sensor. 8.7 Then install in turns protection cover, jbckscrew, decoration part and

DIP switch Calibration Zero calibration 40%F.S calibration 50%F.S calibration DIP switch Calibration 10%F.S calibration 20%F.S calibration 60%F.S calibration Note: Black part is position of the DIP switch. 7.3.2.2 Connect gas bottle and gas flow meter correctly. Open the gas bottle and adjust the flow meter according to the above table. Connect the tube to the calibration cap of the detector. After 2 minutes of inputing gas, when the controller displaying value will be the same with the standard gas concentration, resume the DIP switch to the original position and the detector will remember this value. Red LED flickers several times and then is off. Then close the gas bottle and take off the tube. The calibration is completed. 3. Structural Drawing The device consists of the housing, sensor part and PCB. The housing consists of upper cover, lower cover, O-type circle and cable input hole part. The housing function is to protect the inside part. The sensor part consists of gas sensor, sensor protection cover, decoration part and calibration cap. The sensor and housing is connected by potting epoxy resin. The following is the structural drawing: Fig. 1 Structural drawing 7.4 Calibration of Oxygen Detector 7.4.1 Zero calibration No. Name No. Name No. Name 1 Upper cover 5 Lower cover 9 Fixing circle 2 O-type circle 6 Ground nut 10 Pad 3 Protection part 7 Sensor module 11 Seal ring 4 PCB 8 Closing plug 12 Connector 4. Installation 4.1 Position 4.1.1 For petrol gas, oil gas and alcohol gas etc. which is heavier than the air, the position is 0.3m-0.6m higher than the ground. 4.1.2 For natural gas, CH4 etc which is lighter than the air, the position is 0.5m-2m higher than the gas source. Note: Gas density more than 0.97kg/CBM, then it's heavier.

Gas density less than 0.97kg/CBM, then it's lighter. 4.1.3 The position should be far away from shocking, shattering, strong electromagnetic interference. Around the position, there should be at least 0.3m empty place. 4.1.4 The position should be within 1m around the possible gas leakage area, such as valve, pipe connection, gas outlet place. Please try to install it nearer to the above places, but avoid influencing the working of the other equipments. Please avoid the environment of high temperature and humidity. Also please keep it from water swashing, oil and mechanical damage. Please also consider the convenience of maintenance and calibration. 4.1.5 For large scale detection, we suggest install 1pc every 10-12 square meters, so as to get the best detection result. 4.2 Dimensions Fig. 2 Dimension 4.3 Installation methods Note: Fix the detector with sensor head downwards. According to the installation place, fix the detector onto the wall or gas pipes. Details as follows: Method 1 Wall mounted: According to the device dimension, choose a Normally, it includes zero calibration and single calibration. Calibration instruction is as follows: 7.1 Electrify the detector so as to ensure it can work normally. 7.2 Zero Calibration 7.2.1 Put the device into the clean air and keep it working for at least 10 minutes. Then hold both S2 and S3 buttons continuously for 1-2 seconds, the green power LED starts to flicker, which indicates the device enters into calibration status. Adjust the first digit of DIP switch to "ON". Red LED starts to flicker and green LED is continuously on. The controller displays 0. After several seconds, resume the first digit of DIP switch to the original position, and the detector remembers the present level as zero. Red LED flickers several times and then is off. Zero calibration is completed. 7.2.2 If zero calibration cannot be carried out in the clean air, then please follow the above instruction to enter into the calibration status. When the first digit of DIP switch is to "ON" and red LED starts to flicker, inhale N2 gas or synthetical air onto the sensor continuously for more than 10 minutes. Then resume the first digit of DIP switch to the original position, and the detector remembers the present level as zero. Red LED flickers several times and then is off. Zero calibration is completed. 7.3 Single Point Calibration 7.3.1 Equipment 1 bottle of standard gas: normally the concentration is 40%F.S. 1000ml/min gas flow meter 7.3.2 Procedure: 7.3.2.1 Hold both S2 and S3 buttons continuously for 1-2 seconds, the green power LED starts to flicker, which indicates the device enters into calibration status. Adjust the second digit of DIP switch to "ON" (which is 40%F.S calibration. For details of calibration s, please refer to the following table.) Red LED starts to flicker. Note: Without permission, nobody is allowed to adjust the DIP switch.

suitable wall. First fix devices on the wall by 4pcs of M6 60 bolts. 5.4 After the correct wire connection, take out the useless wire from the enclosure. Then screw on the connector, tighten the compaction circle, rubber air-proof circle and wire. Explosion-proof soft tube can also be connected with the device directly. Note: The size of the wire between the controller and detector should be not less than 6mm, and the distance should be note more than 1000m. 5.5 After checking all the connection well, install the display PCB and the front cover. Make sure O-type circle is put on and connected with the cover tightly. 6. Debugging Before adjusting, please be noted of the following information: Fig. 3 Wall Mounted 1 Wall 2 Fixing holes 3 Bolts Warning! Make sure wire connection is good and then connect it to the power. Make sure there is no combustible gas leakage on the spot. Don't use pure gas or light to test it. Otherwise, it will damage the sensor or lower the sensitivity. Method 2 Screw Buttjoint Connection: If there is pipe screw of G1/2 on the spot, the user can butt joint the device to the pipe screws and then screw tightly the device. Note: Adjusting must be carried out by professional personnels. Procedures: 6.1 Connect the wire correctly according to installation position and wire connection instruction. Then connect the device to the power. Green power LED is flickering and the detector enters into warm up status. 6.2 After 3 minutes of warm up, green power LED is on continuously and the detector enters into normal working status. 7. Calibration Calibration is to adjust the value of the detector in the specific gas environment so as to assure the detecting accuracy and reliability. Fig.4 Screw Buttjoint Connection-1 1 Pipe with screw 2 Connector

5. Wire connection 1 Pipe with screw 2 Connector Fig. 5 Screw Buttjoint Connection-2 Method 3 Pipe Installation: If there is 1" to 2" (diameter) pipe, the user can use 2 U-type bolts to fix the hanging plate on the pipe, then fix the detector onto the plate. Or, the user can first fix the detector onto the plate and then fix the plate on the pipe. 1 Installation plate 2 Detector 3 Fixing bolts 4 Pipes 5 U-type bolts Fig. 6 Pipe Installation Fig. 7 Inside the device No. Name 1 DIP switch 2 + button 3 - button 4 Calibration LED 5 Power LED 6 Terminals 7 Fixing holes/ground Warning: All the wire connection must be carried when the power is cut off. Procedure: 5.1 Take off the up cover from the device and the inside can be seen. 5.2 Screw down the pipe connector in counter-clockwise way. Take out the explosion-proof pad. In turns, put 3-line wire to the internal of the enclosure through the pipe connector, compaction circle, air-proof closing plug and the wire connection hole. According to the explosion-proof requirement, please don't take out the explosion-proof closing plug from the unused wire connection holes. Please don't throw away any part inside the enclosure or the PCB. 5.3 All the wires must be connected to the terminals as shown on the following drawing. No. Icon Name Specification 1 2 NC Low alarm --- Fig. 8 Terminals 3 4 NC High alarm --- 5 Iout 4-20mA output 4~20mA 6 - Cathode of power DC24V±25% 7 24V Anode of power DC24V±25%