CER-1 & CER-2 Site Requirements

Similar documents
CER OPTIMA. Endoscope Reprocessing System CER-1 & CER-2 Site Requirements. Introduction. Space Requirements

CER-1 & CER-2 Installation & Testing Instructions

Customer Requirements Pre-Installation Instructions

Endoscope Reprocessor Installation Instructions Endoscope Reprocessing System

Water Pre-Filtration System Installation Instructions Endoscope Reprocessing Systems

SITE REQUIREMENTS & INSTALLATION DSD-201 & SSD-102

Site Requirements. MODEL 2.0 Australia

SITE REQUIREMENTS 1 USER MANUAL INSTALLATION PACKAGE SITE REQUIREMENTS

Plumbing fittings should be hand tightened followed by one complete turn with the appropriate wrench.

Site Requirements ADVANTAGE PLUS

CER OPTIMA. Service Manual. Endoscope Reprocessing System. CER OPTIMA Service Manual

CER-1 / CER-2 (MV-1 / MV-2) (CLM-5001X / CLM-5002VX)

Chemical Transfer Pump Tube Set and Adapter Instructions

Cascade Comfort Bathing Systems with Aqua-Aire Safe Operation & Daily Maintenance Instructions

AACH50 1/2 HP GALLONS (200 TO 1200L) INSTRUCTION MANUAL

FOR THE RECORD. You should record the model and serial numbers, which can be found on the back of the unit, for future reference.

SAVE THESE INSTRUCTIONS

Operator s Manual. Model G32-S Model G32-E Disinfection Soak Stations

VANTAGE. Installation Manual. Release Date: April 16, 2003 Publication Number: INS Revision Date: January 22, 2015 Revision: E

advanced : technology

Advantage Series Bathing System Specifications

Dansereau Dental. P r e - C o n s t r u c t i o n G u i d e

Operator s Manual. Model G17-EU (International) Disinfection Soak Station for Hysteroscopes, Cystoscopes and ENT scopes

PART 1 - GENERAL 1.1 WORK INCLUDED

Installing the Turbo Floor Model System

USER MANUAL SAVE THESE INSTRUCTIONS

How To Size A Big Dipper Unit

PRE-INSTALLATION PROCEDURE

WM-450-PT WM-900-PT Commercial Wall Mounted Reverse Osmosis

MICRO CLEANER. OPERATOR MANUAL Advance MODEL /10 Rev E Document Number

Top Control Dishwasher

OX100, OX102. Limiting-current type Oxygen Analyzer OX100, OX102. Bulletin 11M10A01-03E

BUILT-IN DISHWASHER INSTALLATION INSTRUCTIONS

INSTRUCTIONS FOR OPERATION AND CARE OF

Installation, Operation, and Maintenance Manual RTE14S & RTE14S-2. For the Taco Bell "Rethermalizer" Model Numbers

1 ESMA, Inc. P. O. BOX 734 * SOUTH HOLLAND, IL * (800) * FAX (708)

Glass and Dishwashers 402/452/502. (original instructions) (incl. Australian /502) Starting from Serial No.:

LABORATORY ULTRA ZERO AIR GENERATOR. ZAC-ULT Series USER MANUAL

PART 1 - GENERAL 1.1 WORK INCLUDED

DI/RO GAS TANKLESS WATER HEATER

Superior Spa with Aqua-Aire

D E S I G N G U I D E

Signature Features. Stainless Steel Interior. Quiet Dishwasher. Delay Start. Multiple-Cycle Options. More Easy-To-Use Features. Product Dimensions

Multi-Function Cooktop

INSTALLATION INSTRUCTIONS

APPENDIX 4. Thank you for choosing the Biomega BioClave steam sterilizer. HYDRAULIC DRAWING

Schuco byallied MEDICAL ASPIRATOR. USER'S MANUAL A Caution

Installation Plan. Professional Dishwasher PG Mat. no.: en - US

OG5000 Ozone Generator Model OG5000B-A. User Manual

1-866-PENNERS

Cascade Contour Bathing Systems with Aqua-Aire Safe Operation & Daily Maintenance Instructions

UF424 and UN324. Technical Service

INSTRUCTION MANUAL AMS WATER MANAGEMENT SYSTEM (AMS-WMS 230V) INSTRUCTION MANUAL AMS WMS 230V Manual Issue 1 15/10/09 Page 1 of 15

Installation Manual CARBONATOR With Plain-Water Booster

BEFORE OPERATING THE MACHINE: WARNING

Installation plan PG 8581 PG 8582 PG en - US

Viking Installation Guide

INSTALLATION INSTRUCTIONS

10. Water Dispensers

LIBBE BASIC INSTRUCTIONS & MEASUREMENTS. 8 (2.44 meter) (73.66 cm) 5 2 (1.6 meter) Approximately 6 (1.83 meter)

SOLUSCOPE SERIES 3 DESIGNED WITH THE FUTURE IN MIND FEATURES FLEXIBLE ENDOSCOPE DISINFECTOR / STERILISER

Signature Features. 30" 24" Electric Built-In. Quiet Dishwasher. Tall-Tub Design. SpaceWise Silverware Basket. Energy Saver Dry Option

Jabsco Marine Toilets SIMPLY RELIABLE SOLUTIONS FOR EVERY BOAT AND BUDGET

Celebrating over 30 years of excellence in manufacturing ultrasonic cleaning equipment, electropolishing equipment and associated chemistries

Technical Manual and Replacement Parts List. Cream Dispenser

Installation Plan. Professional Dishwasher PG Mat. no.: en - US

Operating Instructions

Installation plan: PG Mat. no Version 00 US

SAFETY ALERT SYMBOLS SAFETY PRECAUTIONS

Hoshizaki America, Inc.

Hoshizaki America, Inc.

Warmette Blanket Cabinet Warmette Dual Cabinet. User Manual

Tall-Tub Design. Our large capacity, tall-tub dishwasher fits up to 12 place settings so you can wash more at once. SpaceWise Silverware Basket

W-750-IS Specifications

DISHWASHER INSTALLATION GUIDE SPECIFICATIONS, INSTALLATION, AND MORE

DESIGN CLASS FUNCTION

DOUBLE DISHDRAWER TM DISHWASHER

MODEL 7000 SUCTION UNIT

Installation/Service Manual VANGUARD 245 POST-MIX OVERCOUNTER DISPENSER

Penner Stationary Patient Lift Safe Operation & Daily Maintenance Instructions

OPERATOR S MANUAL FOR YOUR SAFETY. 4 Gallon / Liter Part No. PB408SV

A m e r i c a s D e n t a l E q u i p m e n t C o m p a n y. Rev 2/15. D a n s e r e a u

Installation Instructions

GROWMAX WATER Perfect Water for Plants and Gardens

REMOTE COOLER AIR COOLED & WATER COOLED

P/N CC-100. Coolant Circulator. Operation manual

TBJ Model SPCS B Scope Pre-Cleaning Sink. Product Specification. Product Description

J120 STEAM BOOSTER INSTALLATION, OPERATION, AND SERVICE MANUAL J120 STEAM BOOSTER. J120 Steam Booster Manual D

Installation. 324 Series Built-In Dishwashers U L. Viking Range, LLC 111 Front Street Greenwood, Mississippi USA (662)

π H-6621 INDUSTRIAL DEHUMIDIFIER WARNINGS SPECIFICATIONS uline.com WATER REMOVAL ELECTRICAL REQUIREMENTS BUILT-IN ELECTRICAL SAFETY

Operator s Manual. Model G17 Disinfection Soak Station for Hysteroscopes, Cystoscopes and ENT scopes

Installation Plan. Laboratory Glassware Washer. en US PG 8583 CD. M. Nr

STT700 Temperature Transmitter Lightning Protection Device Specifications 34-TT-03-20, June 2017

WL380 QUICK START GUIDE

Installation. 15 W. Undercounter/Freestanding Ice Machine FGIM515 / CFGIM515 FPIM515 / CFPIM515

SPEAKMAN GRAVITYFLO PORTABLE EYEWASH SE-4000

Quick Start Installation Guide

Installation and User Manual for Model UF424 and UN324

COMMERCIAL WASHERS MODELS T-300/350/400/450/600/650/750/900/950/1200/1450 VENDED C-SERIES CONTROL

Rev B, 9/2/2009. Kodiak Chiller Overview

Transcription:

CER-1 & CER-2 Site Requirements 1 2 INTRODUCTION The following document outlines the space requirements, water supply, drain and electrical specifications required for the installation of the MEDIVATORS CER OPTIMA endoscope reprocessor models CER-1 and CER-2 models. The CER-1 OPTIMA model is capable of reprocessing a single flexible endoscope, while the CER-2 OPTIMA model can accommodate up to two endoscopes, allowing single or dual endoscope reprocessing. This document is a guideline for installation specifications and requirements, in addition to any facility/hospital requirements. SPACE REQUIREMENTS Chassis Dimensions Height = 16 inches (41cm), 33 inches (84cm) with the lid fully open Width = 22 inches (56cm), 26 inches (66cm) with barcode scanner mounted to CER Depth = 20 inches (51cm), Allow a minimum of 4 additional inches (10.2cm) for water/drain/disinfectant hose connections located in the back of the CER machine). Empty Weight (approximated) CER-1 OPTIMA model: 71 pounds (32kg) CER-2 OPTIMA model: 77 pounds (35kg) Cart Dimensions Height = 33.0 inches (84cm) Width = 21.5 inches (55cm) Depth = 24.5 inches (62cm) Active Vapor Management System Dimensions Height = 32 inches (81cm) Width = 23 inches (69 cm) Depth = 24 inches (61cm) Counter Top Installation To install the endoscope reprocessor on a counter top, the counter top must be a minimum of 24 inches (61cm) deep, (from front to back) and at a comfortable working height. Also ensure there is adequate clearance between the open CER lid and any cabinets. For the counter top, it is necessary to cut two 2 inch (5.1cm) holes in the back/rear of the counter top to allow passage of the drain hose and the disinfectant chemical reservoir tubing. Two additional holes are also required if the water inlet hose and electrical connection are located under the counter. Cart Mounted Installation The endoscope reprocessor can be placed on the cart, which is designed to provide easy access to the disinfectant chemical reservoir and water filtration system.

3 WATER SUPPLY SPECIFICATIONS Inlet water flow Minimum of 2 to 3 gallons (7.6L to 11.3L) per minute Water pressure Minimum of 40 PSI (2.8 bar) Maximum of 60 PSI (4.1 bar) NOTE: PRESSURES OUTSIDE OF THIS RANGE WILL AFFECT THE OPERATION OF THE MACHINE. Consult a qualified plumber for options to bring the water pressure within the 40-60 PSI range. Water temperature Cold water supply: Not to exceed 104 F (40 C) NOTE: Endoscope manufacturers recommend that endoscopes not be exposed to temperatures above 130 F (55 C). Water Hardness Ensure water hardness is less than 12 gpg (200 ppm) for optimal performance. De-ionized water (DI water), due to its corrosive nature, is not recommended for use in MEDIVATORS reprocessors. Reverse Osmosis (RO water) can be used however, not necessary since potable water can be used with the MEDIVATORS reprocessors. If RO water is utilized, the pre-filter system must still be used. Water Shut-off Valve A shut-off valve must be installed before the water filtration system. It is important for the valve to be mounted within the operator's reach to allow for easy and convenient daily use. The shut-off valve is used for turning the water on at the beginning of the day, and shutting the water off at the end of the day. Water Connection All fittings on the reprocessor and the water filtration system are 3/4" male GHT (Garden Hose Thread). All tubing necessary to connect the water filtration system to the reprocessor is included with the CER OPTIMA system. The water filtration system is installed in-line. Local plumbing regulations vary regarding the installation of vacuum breakers for water line and back-flow protection for the drain. Consult local plumbing regulations to ensure compliance. NOTE: MEDIVATORS recommends checking local plumbing regulations regarding anti-siphon valves, vacuum breakers, and plumbing connections, before attempting to install the endoscope reprocessor. 2

4 DRAIN SPECIFICATIONS The drain fitting on the CER OPTIMA reprocessor is 1.25 inch (32 mm) OD. It is recommended that the reprocessor be installed with the 4-foot (1.2 m) flexible drain hose supplied with the CER OPTIMA system. A drainpipe of no less than 1-1/2 inch (38 mm) diameter must be placed at least 1 foot (30.5 cm) below the drain fitting of the reprocessor (having no loops), to ensure complete draining of fluids from the system. The drain hose requires an air gap at the point it enters the drain. Consult local plumbing codes for size of air gap. The facility drain must be capable of handling a discharge rate of 5 gallons per minute (19L per minute). A facility drain having a dishwasher type "Y" drain is not an acceptable drain method and cannot be used. The drain "Y" does not provide adequate flow rates nor air gap and must not be used. Refer to figure 1. Figure 1 3

5 ELECTRICAL SPECIFICATIONS CER OPTIMA Reprocessor Electrical 120 VAC at 60 Hz, 3.0 amps, 1Ø 230 VAC at 50/60 Hz, 3.0 amps, 1Ø (230V models not available in North America) CER OPTIMA Reprocessor Fuses (located in power entry module) 120VAC, Two 4.0 amp slow-blow fuses 230VAC, Two 2.0 amp slow-blow fuses Power Cord The CER OPTIMA reprocessor is supplied with one 10-foot (3.05 m) hospital grade removable power cord. A minimum of one duplex (two outlets) outlet is required for the reprocessor and heated disinfectant reservoir. An additional duplex outlet is required for the Active Vapor Management System and the Disinfectant Transfer Pump. Safety Classification Ordinary protection The reprocessor is not suitable for use in the presence of a flammable anesthetic mixture with air, oxygen or nitrous oxide. NOTE: For maximum safety, MEDIVATORS recommends the CER OPTIMA system only be used when connected to an electrical outlet containing a Ground Fault Interrupter (GFI), or Residual Current Device (RCD). NOTE: MEDIVATORS recommends surge protectors be used for protection against electrical power spikes and surges. Consult local electrical codes to ensure proper compliance. Heated Disinfectant Reservoir (option) Electrical 120 VAC at 50/60 Hz, 3.5 amps, 1Ø 230 VAC at 50/60 Hz, 1.75 amps, 1Ø Heated Disinfectant Reservoir Fuses (located in power entry module) 120VAC, 5.0 amp slow-blow fuses 230VAC, 2.0 amp slow-blow fuses Disinfectant Transfer Pump (option) Electrical 120 VAC at 50/60 Hz, 0.5 amps, 1Ø 230 VAC at 50/60 Hz, 0.25 amps, 1Ø Disinfectant Transfer Pump Fuses (located in power entry module) 120VAC, 1.5 amp slow-blow fuses 230VAC, 1.0 amp slow-blow fuses 6 VENTILATION REQUIREMENTS During cycle operation and chemical heating process, chemical vapors are produced. It is a health and safety requirement to have sufficient air changes, at least 10 to 12 times per hour, in the room to dissipate the chemical vapor. The flow of air should be considered by the customer as well. Air should flow away from the front of the unit to the back, thus drawing fumes away from the operator's face. PPE should be used as required. The unit may be supplied with an Active Vapor Management System to aid in chemical vapor management.

7 HIGH LEVEL DISINFECTANTS The CER OPTIMA reprocessor will monitor the minimum and maximum temperature of the high level disinfectant to assure proper disinfections of the endoscope. The temperature limits are automatically programmed when the disinfectant type is selected. If a disinfectant requires heating, this will require the use of the heated disinfectant reservoir option. If a non- heated reservoir is utilized and the minimum required disinfectant temperatures are not achieved, the CER OPTIMA reprocessor will halt and error. The cycle cannot be completed without the proper disinfectant temperature. For further details, consult your local MEDIVATORS Representative. NORTH AMERICAN MINIMUM AND MAXIMUM ALARM PARAMETERS DISINFECTANT TYPE Minimum Temp alarm HLD Contact Time Number of Rinses Maximum Temp alarm RAPICIDE Disinfectant 35 C 5 min 2 40 RAP OPA28H 25 C 5 min 2 40 RAP OPA28U 20 C 10 min 2 40 OPA 20 C 12 min 3 40 GLUT 20 C 20 Min 2 40 GLUT 25 C 20 Min 2 40 GLUT 20 C 45 min 2 40 INTERNATIONAL MINIMUM AND MAXIMUM ALARM PARAMETERS DISINFECTANT TYPE Minimum Temp alarm HLD Contact Time Number of Rinses Maximum Temp alarm RAP OPA INT 20 C 5 min 2 40 MEDIVATORS, RAPICIDE and CER OPTIMA are registered trademarks of Medivators Inc. Medivators Website Resource Center Go to: www.medivators.com, Select Resource Center and User Library for detailed user guides and hookup information, report forms and logs, and product bulletins MEDIVATORS Customer and Technical Support Toll Free: Phone: 1-800-444-4729 FAX: 1-866-553-9444

Manufactured in the USA by: Medivators Inc. 14605 28th Avenue North Minneapolis, MN 55447 USA Toll Free: +1.800.444.4729 www.medivators.com Medivators BV Sourethweg 11 6422PC Heerlen The Netherlands Tel: +31.45.5.471.471 2016 Medivators Inc. All rights reserved. 50097-820/G Cantel Medical Asia/Pacific Pte. Ltd. 1A International Business Park #05-01 Singapore 609933 Tel: +65.6227.9698 Cantel Medical Devices (China) Co. Ltd. Unit 804-805, Innov Tower Block A, Hongmei Road, Xuhui 200233 Shanghai Tel: +86 21 60161380 Fax: +86 21 61210913