Installation and Operating Instructions DÜRR Regeneration Unit for X-ray developers XR 24, XR24 II, XR 24 Nova, XR 24 Pro

Similar documents
INSTALLATION AND OPERATING INSTRUCTIONS DÜRR DAYLIGHT LOADER DL 24, DL 26, DA 24

Periomat Plus. Installation and Operating Instructions / /08

Dürr Dental X-ray Film Developer XR 24 Pro

XR 24 NDT. Manual /30

Automatic Mixer Operating Instructions

Operator Guide. Publication No. OI /97-Eng. KODAK Automixer II Plus

INSTALLATION AND OPERATING INSTRUCTIONS DÜRR XR 24 PRO X-RAY FILM DEVELOPING APPLIANCE

HOT WASHER MODEL NO: KING150

Specifications. Vacuum motor power consumption(w/hp) 1200 / 1.6. Exhaust water pump power consumption(w/hp 800 / 1.1

SHORT WAVE INFRARED PANEL DRYER

Operational Instructions. Glue Applicator MS Translation from German to English

Trotec PAC FT 2600 Portable air conditioner

Antari X-310 ORDERCODE 60663

PRESSURE WASHER MODEL NO: JETSTAR 1750 OPERATION & MAINTENANCE INSTRUCTIONS. WARNING Read the instructions before using the machine PART NO:

INSTRUCTIONS FOR OPERATION AND CARE OF

Waste water ejection unit

Energy management function at adsorption dryer PDAD

TECHNICAL INFORMATION Touchtronic Clothes Dryers

B 353 COFFEE BREWING MACHINES MACHINES A CAFE KAFFEEBRUEHMASCHINEN KOFFIEZETAPPARATEN

Millo / Millo pro. Nr x000 / 1805-x000. Ideen für die Dentaltechnik A

Glass and Dishwashers AMX / AUX Series

Instructions for the fan motor control system, SILVER C

BOILING UNIT REDITAP. Installation and User Guide. IMPORTANT: This booklet should be left with the user after installation and demonstration

1. Take the hoses and elbow out of the holders, depending on model. max. 100 cm. min. * cm

541D19 SERIES. Technical Manual. A Division of Aquion Partners L.P.

MW001 Integrated Dishwasher. Manual for Installation, Use and Maintenance

Product Manual Glue Applicators MS 80 and MS 200

USER MANUAL. Version: 1.0. ROBE Show Lighting s.r.o. Czech Republic

POLYMIX PX-IG 2000 Operating Instructions

SCHMIDT Flow Sensor SS Ex - Supplementary instructions for use in explosive atmospheres ATEX

Installation and Operating Manual

EHA Hoffmann International GmbH

Piccolo Cyclonic Vacuum Cleaner

AirStar Clinical Air Compressor AS60, AS90 and AS120. Installation & Operating Instructions

Instructions for the fan motor control system with integrated wiring terminals SILVER C

Operating Guide. Termix Compact 28 VX-FI / HWP / HWS. 1.0 Table of Contents. 1.0 Table of Contents

TQ Series. Electronic Control Pump Instruction Manual. 50Hz. ISO 9001 Certified Walrus Pump Co., Ltd.

Service package PDAD- -SP (en) Assembly instructions d [ ]

UL U TR LTRASONIC S ONIC SCALE ALER PIEZO MINI

NEPTUN I Sole cleaning unit Art.-No.:

MODEL 7000 SUCTION UNIT

USER GUIDE. DRENA 2 - User Manual ELECTRICAL PANEL FOR 2 MOTORS - WASTE WATER -

Safety. DANGER Indicates potentially fatal situations. WARNING Indicates possible danger to life and limb.

IH 025 VOLCANO. Portable Induction Heater. smart mounting

DEUTSCH Bio-Master Bio-Max

User and Installation Instructions. Wall Mounted Water Boiler WMB3F/B and WMB3F/W IS435 ECN3461

U88-T2-B. User Guide. Bagless Upright. Get Cleaning... vax.co.uk. U88-T2 series. What s your Vax model number? (Located on the flap of the packaging)

FINO JETSTEAM PRO 52072

Glass and Dishwashers AMX / AUX Series

Vacu-Mixer. Operating Instructions

WF140 Freestanding Dishwasher

Owner s Manual READ AND SAVE THESE INSTRUCTIONS

CP-200 Push Button. CP Warewash Pump Instruction & Operation Manual Push Button , Rev1.0,

Z-800II / Z-1000II / Z-1020 Fog Machine User Manual

Portable Air Conditioner with Heat Pump Technology Operating Instructions. Model No.: HCB-P13HP-D. Reference No.: BPD13HP V140217

Page 1 of 18. Part# /5/2013

Bagless upright vacuum

Installation instructions Air drying unit for Bubble Sensor OTT CBS. English

5KW ELECTRIC FAN HEATER

Servicing manual. 600 Series - 11S / 19S / 24S / 24C. Wall-mounted condensing gas boiler. For trade use

OPERATOR S MANUAL SF-12 AND SF12A HI VOLUME WATER HEATER

BarAid 400. Installation and. Starting from Serial No.:

WC141 Integrated Dishwasher Installation, Use and Maintenance

Bagless upright. Vax Careline (UK) (ROI) vax.co.uk VS-19EC. Vax model number:

User Manual GV25 GV35 GV702. Company information: Original instructions GV12066 (1)

Servicing manual. Wall-mounted condensing gas boiler 600 Series - 11S / 19S / 24S / 24C /2002 GB(EN) For trade use

15KW ELECTRIC FAN HEATER

Alpha CombiMax 350 and 600

Operating manual UNIQUE UNIQUE [HAZER]

MW401 Integrated Dishwasher

V-ZUG Ltd. Range hood DW-SE/DI-SE. Operating instructions

Operating Instructions for Reka Hot-Melt Glue Applicators

POWERFORCE HELIX. Vacuum USER GUIDE 2110 & 2111 SERIES. (Model 2110 Shown)

Owner's Manual. WS Series. Water Softener

14,000 BTU Portable Air Conditioner 14,200 BTU Heat Pump Model Number: LX-140 Operating Instructions. Model No. LX Soleus Air International

Antari Z-300 ORDERCODE 60661

Part No KEMPER Filter-Master XL

PV3000S Professional Pond Vacuum

Operator s Manual. Model G14TC-3 Disinfection Soak Station for Transesophageal Ultrasound Probes

SIME FORMAT WALL HUNG BOILERS MODEL 34i AND MODEL 34e. cod A

OPERATING MANUAL. for the modular heat presses. Secabo TC5 SMART and TC7 SMART

Installation Guide. Termix Distribution unit. 1.0 Table of Contents. 1.0 Table of Contents

OPERATING MANUAL Gfp 255C Please read this manual carefully before operating!

EP8020 STEAM GENERATOR IRON PROFESSIONAL TYPE IRONING SYSTEM WITH DETACHABLE WATER TANK FOR HOUSEHOLD USE ONLY

Bench-Top Vacuum Oven with 28 segments programmable controller

Reacti-Therm I and III Heating Modules

Z-1500II and Z-3000II USER MANUAL Z-1500II und Z-3000II BEDIENUNGSANLEITUNG Mode D Emploi des Z-1500II et Z-3000II Machines à Fumée

Model 5060/5070. Letter Opener Revision 3 Brush Transport

PowerVent. Installation manual (GB) English. Store this document in a safe place UK

GB Operating instructions

INDUSTRIAL VACUUM CLEANER WITH AUTO START/STOP

TECHNICAL PROPERTIES OF THE APPLIANCE

Bagless upright. vax.co.uk. Vax Careline: (UK) (ROI) VS-190 Series. Vax model number: Version 1.1

CORREX MP Anode with external voltage source. Installation & Operation Manual

Heating, Air Conditioning, Ventilation. Отопление-Кондиционеры-Вентиляция. MTM 8-30 kw UNIVERSAL OIL HEATER OPERATING MANUAL

LABOPORT. Chemically-resistant Laboratory Pumps with Modular Accessories. Operating Instructions for Laboratory Pumps. and for LABOPORT Systems

5) Do not start or stop the unit by inserting or pulling out the power plug.

Steam Mop User Manual

Moisture inside the drum is due to final testing.

Mechanical seals external, single or double, to DIN EN 12756

Transcription:

Installation and Operating Instructions DÜRR Regeneration Unit for X-ray developers XR 24, XR24 II, XR 24 Nova, XR 24 Pro 2006/01

Content Important Information 1. Notes... 3 1.1 CE - Labeling... 3 1.2 Electromagnetic Compatibility... 3 1.3 General Notes... 3 1.4 General Safety Information... 3 1.5 Combining Appliances... 4 1.6 Electrical Safety Notes... 4 1.7 Warnings and Symbols... 4 2. Product information... 4 2.1 Correct Usage... 4 2.2 Incorrect usage... 4 3. Contents... 5 3.1 Accessories... 5 3.2 Special Accessories... 5 4. Technical Data... 5 5. Functional Diagram... 6 5.1 Regeneration Unit connected to XR 24, XR 24 II, XR 24 NOVA... 6 5.2 Regeneration Unit connected to XR 24 PRO... 6 6. Functional Description... 7 6.1 Regeneration Unit RU 24 connected to XR 24, XR 24 II, XR 24 NOVA... 7 6.2 Regeneration Unit RU 24 connected to XR 24 PRO... 7 Installation 7. Set-up alternatives for Regeneration Unit... 8 8. Preliminary measures... 9 8.1 Remove front cover... 9 8.2 Remove rear cover... 9 8.3 Remove side covers... 9 9. Hose connections... 10 10. Electrical connections... 11 10.1 XR 24, XR 24 II, XR 24 NOVA... 11 10.2 XR 24 PRO... 11 11. Setting the Regeneration Electronics12 11.1 XR 24... 12 11.2 XR 24 II and XR 24 NOVA... 13 11.3 XR 24 PRO... 14 12. Check Settings... 15 12.1 XR 24, XR 24 II, XR 24 NOVA... 15 12.2 XR 24 PRO... 15 13. Setting Regeneration Pump Piston Stroke... 16 14. Commissioning... 16 Operation 15. Cleaning the Regeneration Unit... 17 15.1 XR 24, XR 24 II, XR 24 NOVA... 17 15.2 XR 24 PRO... 17 16. Cleaning the X-ray Developer... 17 17. Disposal of Chemicals... 17 Troubleshooting 18. Troubleshooting Tips for XR 24, XR 24 II, XR 24 NOVA... 18 19. Troubleshooting Tips XR 24 PRO... 19 2

IMPORTANT INFORMATION 1. Notes 1.1 CE - Labeling This product bears the CE- Labeling in accordance with the guidelines of the Medical Products Committee 93/42/EWG and satisfies all criteria of the Appendix I to said guidelines. 1.2 Electromagnetic Compatibility This product conforms to all requirements laid down for Electromagnetic Compatibility of Medical Appliances. 1.3 General Notes These Installation and Operating Instructions form an integral part of the unit. They must be kept close to the unit and in readiness whenever required. Precise observance of these instructions is a precondition for use of the unit for the intended purpose and for its correct installation and operation. These Installation and Operating Instructions should be passed on to any future purchaser or explained to any new operator. Safety for the operator as well as troublefree operation of the unit are only ensured if use is made of original equipment parts. Moreover, use may only be made of those accessories that are specified in the technical documentation or that have been expressly approved and released by Dürr Dental for the intended purpose. Any warranty or guarantee becomes invalid in the case where parts or accessories are used which are not supplied by Dürr Dental. Dürr Dental only regard themselves as being responsible for the equipment with regard to safety, reliability and proper functioning if assembly, resettings, changes or modifications, extensions and repairs have been carried out by Dürr Dental or an agency authorized by Dürr Dental and if the equipment is used in conformity with the Installation and Operating Instructions. The developer unit herein described conforms to the relevant safety standards valid at this time. All switches, processes, trade marks, software programs and appliances named in this document are registered names. Any reprinting of the technical documentation, in whole or in part, is subject to prior written approval of Dürr Dental. 1.4 General Safety Information This appliance has been so designed and constructed by Dürr Dental that any danger arising from the use of this product is virtually excluded provided it is fitted according to the instructions. However, in order to avoid any possible damage or injury we are obliged to point out the following safety measures. When operating this appliance all local rules and regulations must be observed! Any modification whatsoever is not permitted: any guarantees covered by Dürr Dental thus become invalid. In the interests of ensuring problem-free running the owner and operator are responsible for observing all rules and regulations. Retain all packing material for possible return of the product to the manufacturer. Take care that it does not fall into the hands of children. Only the original packing guarantees optimal safety of the appliance during transport. Should return of the product to the manufacturers be necessary during the guarantee period, Dürr Dental accepts no responsibility for damage occurring during transport where the original packaging was not used! This product is a technical appliance and may only be installed and operated or used by someone who has been trained or who has experience of how to use the appliance correctly. Before every use the operator must check the functional safety and the condition of the appliance. 3

The operator must be knowledgeable in the operation of the appliance. The product is not designed to be used in medical treatment areas where there exists the danger of explosion. Areas where explosions could occur are those where flammable anesthetic material, skin cleansers, oxygen and skin disinfectants are present. This appliance is not to be used in any areas where the atmosphere could cause fire. 1.5 Combining Appliances Units may only be connected to the system or to other units when it has been established that there is no reduction of safety for the patient, the operator or the environment through such connection. Where it is not absolutely clear from the documentation whether the safety is reduced by such connection, then the operator must establish, e.g. by contacting either the manufacturer or an expert, that there is no danger to the patient, the operator or the environment through such connection. 1.6 Electrical Safety Notes This regeneration unit may only be used with a correctly fitted standard earthed plug and socket. Before connecting the appliance to the power supply check that the current required and the frequency is the same as that provided at the power outlet. No extension cables should be used to connect the machine to the power outlet. Warning! Check the appliance and the power supply cables for correct installation and possible damage before switching on. Damaged cables, plugs and sockets must be replaced before use. The appliance should not be used in the patient s vicinity. A minimum distance to patient of 1.5 meters is recommended. 1.7 Warnings and Symbols In the Installation and Operating Instructions use is made of the following terms or symbols to denote information of special importance: Information, requirements and prohibitions on preventing personal injury and extensive damage to propert. Special information concerning the economical use of the appliance and other instructions. Observe Installation and Operating Instructions CE-Labeling Power on/off / Select program Confirm selection 2. Product information 2.1 Correct Usage The DÜRR Regeneration Unit is designed solely for use with DÜRR DENTAL X-ray developer units of the model series XR... Precise observance of the installation and operating instructions is a pre-condition for use of the unit for the intended purpose and for its correct installation, operation and maintenance. 2.2 Incorrect usage Any other use or use beyond what is specified is deemed to be not for the intended purpose. The manufacturer accepts no liability for damage resulting therefrom. All risk is borne solely by the user. 4

3. Contents Regeneration Unit RU 24 for X-ray developers XR 24... 230V - 240V, 50Hz - 60Hz... 1700-820-00 100V - 110V, 50Hz - 60Hz... 1700-830-00 Before connecting the regeneration unit to the developer please make sure to check the following: Is the regeneration unit to be connected to a developer of the model XR 24, XR 24 II, XR 24 NOVA or to an XR 24 PRO? For connection to an XR 24 PRO: Use the assembly set 1700-284-00 (included) For connection to an XR 24, XR 24 II, XR 24 NOVA: Use the conversion set PCB 1700-212-00! This must be ordered separately. See section 3.2 and installation as described in section 11.2. The assembly set 1700-284-00 is then no longer required. 3.1 Accessories 1 Assembly set for XR 24 Pro only... 1700-284-00 4. Technical Data Regeneration Unit RU 24 for X-ray developers XR 24... Model 1700-820-00 (230V) Model 1700-830-00 (110V) Model 1700-820-00 Voltage (V) 230-240 230 Frequency (Hz) 50 60 Power consumption (A)0.65 / 0.9 0.5 Output (W) 50-80 50-80 Model 1700-830-00 Voltage (V) 100-110 100-110 Frequency (Hz) 50 60 Power consumption (A) 0.8 / 1.0 0.6 / 0.75 Output (W) 40-80 40-80 Model 1700-820-00 1700-830-00 Measurements (cm) L x B x H 32 x 36 x 37 Weight (empty) (kg) 4 3.2 Special Accessories The following parts are not part of the standard contents. Order separately if required! 1 PCB conversion set, only for XR 24 II and XR 24 NOVA... 1700-212-00 5

5. FUNCTIONAL DIAGRAM 5.1 Regeneration Unit connected to XR 24, XR 24 II, XR 24 NOVA A1 H A2 20 30 40 D G 10 0 REG. 50% OFF 50 S1 60 100% ON C B 5.2 Regeneration Unit connected to XR 24 PRO J H A I B 6

6. FUNCTIONAL DESCRIPTION 6.1 Regeneration Unit RU 24 connected to XR 24, XR 24 II, XR 24 NOVA The regeneration pumps of the X-ray developer are activated every time that the appliance is initially switched on and, furthermore, are reactivated after every length of 1.3m of developed film material when the regeneration switch is correctly set (switch (C) to ON). For the duration of time that the film intake is activated (via light barrier H) on a film being fed into the appliance, an impulse generator (A) produces rotational impulses which are counted, the number of impulses being proportional to the length of the film. The potentiometer (D) allows the running time of the regeneration pumps to be set at any interval between 0 s and 60 seconds. The running time of the regeneration pump is factory set at 30 s. In this time the pumps feed c. 190 ml fresh chemicals into the developer and fixer baths of the X-ray developer from the 5l regeneration canisters. Switching off the pumps, and / or switching off the X-ray developer transmits a delete impulse to the counter which then resets at zero. A) Impulse generator A1) XR24: Drive shaft light barrier A2) XR24 II: Drive motor impulse generator A2) XR24 Nova: Drive motor impulse generator B) Regeneration Unit C) PCB regeneration switch On/Off D) Potentiometer for time setting G) Reduction of regeneration time to 50% / 30s H) Film intake light barrier 6.2 Regeneration Unit RU 24 connected to XR 24 PRO The regeneration pumps of the XR 24 PRO are reactivated after every length of 1.3m of developed film material. For the duration of time that the film intake is activated (via light barrier H) on a film being fed into the appliance, a generator produces impulses which are counted, the number of impulses being proportional to the length of the film. The software in the service program of the XR 24 PRO allows the running time of the regeneration pumps to be set at any interval between 0 s and 60 seconds, see section 11.3. The running time of the regeneration pump is factory set at 30 s. In this time the pumps feed c. 190 ml fresh chemicals into the developer and fixer baths of the X-ray developer from the 5l regeneration canisters. The switching off of the pumps after a regeneration phase automatically transmits an impulse to the counter which then resets at zero. The current counter reading is stored whenever the X-ray developer is swiotched off or in case of a power down. A) Drive motor impulse generator B) Regeneration Unit H) Film intake light barrier I) PCB J) Display 7

INSTALLATION 7. Set-up alternatives for Regeneration Unit 1 Set-up of Regeneration Unit RU 24 next to the X-ray developer: Where sufficient space is available, this set-up is preferred. Set-up of Regeneration Unit RU 24 in a cupboard: The operator must be instructed to check the state of the chemicals regularly. Set-up of Regeneration Unit RU 24 above the level of the X-ray developer is not permissible. This set-up position makes it possible that chemicals leak into the X-ray developer baths. Please ensure that all connecting hoses from the Regeneration Unit RU 24 to the X-ray developers are laid without any twists or bends. The distance between Regeneration Unit and X-ray developer should not exceed 2m. When setting-up the X-ray developer, ensure good access to it and to the collecting tanks. 8

2 2 1 3 4 5 6 8. Preliminary measures The following measures need only be taken when the Regeneration Unit is subsequently being connected to a X-ray developer. Remove mains plug (1) from socket Lift cover (2) Drain any chemicals from the X-ray developer Open the green catch (3) on the film intake roller set (6) Remove roller set (6) Open the green catch (3) and the clips (4) of the dryer roller set (5) Swing the dryer roller set (5) towards the rear and lift out Rinse the baths with warm water Close the water tap Remove the cover necessary for working 8.1 Remove front cover Unscrew the two quick-release screws (11) from the front cover panel (12) Unscrew the three Philip s screws (7) and remove the front cover panel (12) 3 7 8 7 9 8.2 Remove rear cover Unscrew the three Philip s screws (7) and remove the rear cover panel (9) 8.3 Remove side covers Remove front (12) and rear panels (9), lift the side panels (10) upwards to remove 10 11 12 4 9

5 13 14 16E 9. Hose connections 9.1 Hose connection to XR24, XR 24 II, XR 24 NOVA or to XR 24 PRO with serial number up to T005999 The regeneration connection is effected using the nozzles to the tank at 16E and 16F, see figs. 6 and 7. The connecting hose from the Regeneration Unit for developer (black) (13) and fixer (red) (14) must be slipped through the prepared holes on the rear of the machine and fed to the grey connecting sleeves (16E) and (16F) on the baths. Uncsrew the locking nuts (18) on the connectors (16E) and (16F). Remove the black plugs (20). Slide the locking nuts (18) and the grey hose locking ring (19) onto the connecting hose (13), and (14). Slide the connecting hose for developer (black) (13) and fixer (red) (14) all the way onto the appropriate connector sleeve, (16E) and (16F), of the bath and secure in place with the locking nut (18). E Developer bath F Fixer bath 6 16F 9.2 Hose connection to XR 24 PRO with serial number from T006000 The tank design has been optimised and the regeneration connection is effected at the inlet pipe, see fig. 7a. Use the assembly set 1700-284-00 inlet pipe for regeneration unit with enclosed assembly instructions. 16E 16F 18 F E 1 13 14 19 20 2 5 7 10 7a 6

10. Electrical connections 10.1 XR 24, XR 24 II, XR 24 NOVA Place the appliance plug into the socket (15a) to the rear of the X-ray developer and secure in place with the locking screw. 8 15a 10.2 XR 24 PRO Remove socket cap (15b) to the rear of the X-ray developer. Use the connection port (15a) from the 15a assembly set provided 1700-284-00. Put connection port into the prepared opening using the 4 screws. Connect the cable set as illustrated (Fig. 9). Place the Regeneration Unit plug in the socket and secure in place with the locking screw. 8a 15b Circuit diagram XR 24 PRO 1700-216-50 X1/X2 X3 Plug Regeneration Unit 9 11

C20 C19 C18 V 74 R 59 R 67 R 61 C23 11. Setting the Regeneration Electronics Before setting, fill the X-ray developer baths with water. The pump running time should be set according to the individual requirements of the surgery. We recommend, for example, that where a large number of films are developed the regeneration time should be lowered in order to avoid unnecessary consumption of chemicals. 11.1 XR 24 Switch S1 ON / OFF: If switch (22) is set to OFF, regeneration will take place after 1.3m of developed film. If switch (22) is set to ON, then additionally regeneration will take place every time the appliance is switched on. Switch S1 50% / 100%: If switch (21) is set to 100%, then the maximum pump running time of the Regeneration Unit is 60s. By turning the switch (21) to 50% then the maximum pump running time of the Regeneration Unit is halved to 30s. Potentiometer REG regeneration time 10 11 20 10 0 30 REG. R60 40 60 50% OFF 50 Start- Reg. S1 100% ON 21 22 23 The amount of chemical transported to the X-ray developer depends on the pump running time of the Regeneration Unit. The potentiometer (23) can be used to set the pump time from >0s to <60s (<30s). The pump operating time is factory-set at 30s. In this time approx. 190ml fresh developer and fresh fixer are pumped to the baths of the X-ray developer. If further adjustment is necessary the potentiometer knob (23) can be used to change the pump operating time. 12

24 11.2 XR 24 II and XR 24 NOVA The controller for the Regeneration Unit of XR 24 II and XR 24 Nova is not integrated into the main board. For these models the controller can be found on the PCB (24), Order No. 1700-212-00, which is slotted onto the main board. The slots for this printed circuit board can be found in the upper left corner of the main board, and are marked X13 and X14. Before slottting PCB into position (24) remove power supply to the X-ray developer. 12 Setting for regeneration and pump operating time are carried out on the PCB (24). 25 Switch X3 Regeneration on appliance operation ON/ OFF If switch (27) is set to ON, regeneration on appliance operation is additionally activated and when the X-ray developer is switched on regeneration occurs. If switch (27) is set to OFF, regeneration occurs only after 1.3m of film has been developed. 13 26 27 28 Switch X4 50% / 100% If switch (26) is set to 60s then the maximum pump running time of the Regeneration Unit is 60s. If switch (26) is set to 30s then the maximum pump running time of the Regeneration Unit is halved to 30s. Potentiometer P1 Reg.-time The amount of chemical transported to the X-ray developer depends on the pump running time of the Regeneration Unit. The potentiometer (28) can be used to set the pump time from >0s to <60s (<30s). The pump operating time is factory-set at 30s. In this time approx. 190ml fresh developer and fresh fixer are pumped to the X-ray developer baths. If further adjustment is necessary the potentiometer knob (28) can be used to change the pump operating time. 13

14 11.3 XR 24 PRO The regeneration function ON / OFF and the regeneration time can be read on the display of the X-ray developer (Fig. 14). Please refer to the Installation and Operating Instructions of the X-ray developer XR 24 PRO (order no. 9000-600-74/30) See section 18.6 Service Routine. Start Service Routine Switch off appliance at main switch (press min. 2 s) Simultaneously press + (min. 4s) Appliance is in service set-up routine Service function selection Using keys and display select REGENERATE, then select REGEN.ON. Alter setting (Factory setting 30s) or changes the parameter value confirms new value Finish service set-up routine Switch off appliance at main switch (press min. 2 s) 14

12. Check Settings 12.1 XR 24, XR 24 II, XR 24 NOVA Check regeneration at start of operation Main switch ON When the appliance is switched on the Regeneration Unit must be activated and developer and fixer will be pumped to the baths of the X-ray developer. Check the appliance for leaks. Check regeneration impulse for film length Main switch ON Set fast forward to ON (film transport set to 3 min.). Lay a film in the film intake such that the film intake flap is activated but so that the film is not transported. After approx. 8 min. the regeneration pumps should switch on and developer and fixer should be pumped into the baths of the X-ray developer. Check the appliance for leaks. Check regeneration pump performance The developer and fixer pumps should transport the same amount of liquid into the baths of the X-ray developer. Fill two measuring beakers with the same amount of water (approx. 0.5 liter). Place these in the Regeneration Unit and place the two hoses from the regeneration containers into the measuring beakers. Check that the pumps feed the identical amount of water. If necessary alter the pump transport settings. See section 13. Setting regeneration pump piston stroke. 12.2 XR 24 PRO A processor in the electronics of the XR 24 PRO stores the complete developing time, therefore setting option selection is not required. Check regeneration impulse for film length Switch on the X-ray developer and select program Endo (film transport time 2:45 min.). Lay a film in the film intake such that the film intake flap is activated but so that the film is not transported. After approx. 7 1/2 min. the regeneration pumps should switch on and developer and fixer should be pumped into the baths of the X-ray developer. Check the appliance for leaks. Check regeneration pump performance The developer and fixer pumps should transport the same amount of liquid into the baths of the X-ray developer. Fill two measuring beakers with the same amount of water (approx. 0.5 liter). Place these in the Regeneration Unit and place the two hoses from the regeneration containers into the measuring beakers. The pumps can be activated via the Service Routine of the XR 24 PRO, in order to check the amount of liquids transported. (Please refer to the Installation and Operating Instructions of the X-ray developer XR 24 PRO, Order No. 9000-600-74/30. Section 18.6 Service Routine) 15

% OF STROKE 100 75 50 29 13. Setting Regeneration Pump Piston Stroke The amount of chemicals actually transported by the Regeneration Unit can be adjusted by altering the setting of the piston stroke. If the amounts pumped by the developer pump and the fixer pump are not the same then this can be regulated by altering the piston stroke of each pump. 15 25 0 30 Loosen the socket screw (30). Use a screwdriver or spanner to turn the hexagonal screw (29) to change the piston stroke setting: Turning in the direction 100% = the amount pumped increases Turning in the direction 0% = the amount pumped decreases. The optimal pump volume is factory set (Fig. 15). Finally, tighten the hex. socket screw (30). 14. Commissioning After finishing all re-setting that may be necessary drain all water from the baths in the regeneration unit of the X-ray developers and place the chemicals into position as appropriate for the model of X-ray developer and according to the Installation and Operating Instructions of that model. The X-ray developer should never be operated without the roller set! If a roller set is missing then it is possible for developer and/or fixer to be forced under pressure vertically and spray out of the X-ray developer. Chemicals irritate eyes and skin and there is danger of severe damage to the eyes. Without the developer and fixer roller sets transport of chemicals to the baths is no longer possible. There is danger of chemicals getting in the eyes and of chemicals mixing together. If chemicals get into eyes, wash thoroughly with water and consult doctor. 16

OPERATION 15. Cleaning the Regeneration Unit Every 3 months Check whether the collection container of the X-ray developer is empty, if necessary empty into the main canisters. It is important that these collection containers are empty, otherwise overflowing may result when the chemicals are drained from the baths. Drain the chemicals from the baths of the X-ray developers and empty the Regeneration Unit containers and rinse throroughly with warm water. Fill the appliance and the regeneration unit containers with warm water and close the top cover. Now refer to the operating instructions of the appropriate model of X-ray developer for next steps: See section 15.1 or 15.2. 15.1 XR 24, XR 24 II, XR 24 NOVA Regeneration on operation activated Switch the X-ray developer on and wait until the pumps of the Regeneration Unit have begun operation. Finally switch off the X-ray developer. Repeat this operation at least 4 times, so that the Regeneration Unit (pumps and hoses) are thoroughly rinsed through. Once the appliance is again in Stand-By Phase (lifting bath in upper position) turn off at main switch. Regeneration on operation deactivated Switch the X-ray developer on and select fast forward operation. Lay a film in the film intake such that the film intake flap is activated but so that the film is not transported. Wait approx. 30 min., until the Regeneration Unit has operated 4 times. Remove the film from the film intake and let the cycle continue til the end (lifting bath in upper position). Switch off X-ray developer. 15.2 XR 24 PRO Switch on the X-ray developer and select program Endo (film transport time 3 min.). Lay a film in the film intake such that the film intake flap is activated but so that the film is not transported. Wait approx. 30 min., until the Regeneration Unit has operated 4 times. Remove the film from the film intake and let the cycle continue til the end (lifting bath in upper position). Switch off X-ray developer. 16. Cleaning the X-ray Developer Every 3 months Change chemicals (See Installation and Operating Instructions 9000-600-74/30) Chemical cleaning of roller sets in cleaning bath, see enclosed notes, A3- Poster 9000-600-14/01 or Instructions for Cleaning Roller Sets 9000-600-22 (included with XR Clean-Set). 17. Disposal of Chemicals When disposing of developer and fixer observe all local and regional regulations! In Germany and many other countries X-ray chemicals must be disposed of as special waste. 17

TROUBLESHOOTING The following descriptions of troubleshooting procedures are designed for technical staff only. Repairs may only be carried out by Dürr Dental or by a person or agency authorised by Dürr Dental. Before carrying out any repair or maintenance procedures remove power supply! 18. Troubleshooting Tips for XR 24, XR 24 II, XR 24 NOVA Fault Possible causes Remedy 1.Regeneration containers are empty after only a few hours 2.Films not fixed completely and veil on film Frequent use of the appliance main switch, surgery mains switch or constant interruption of power supply. (Only when Regeneration on Operation activated) Electronic relay K3 (of XR24) sticks or electronics defect. Regeneration containers empty. Pump operating time too short. Regeneration containers or their hoses are interchanged. Non-return valve of regeneration pump stuck or wrongly positioned. Regeneration hoses (from the Regeneration Unit to the X-ray developer) twisted. Film shelf-life exceeded. Chemicals shelf-life exceeded. Switch off the machine only during the midday break and keep on Stand-By during working hours. Clean contacts or replace electronics. (Order No. 1700-210-52) Refill the regeneration containers and repeatedly restart at main appliance switch using the regeneration on operation function. If there is still no improvement replace all chemicals in the machine. Pump operating time needs adjusting via potentiometer. Drain chemicals from the machine and clean thoroughly. Connect the regeneration containers in the correct positions and refill with fresh chemicals. Clean non-return valve, replace if required. Check for correct positioning Lay the hoses with no twists or bends. Observe film shelf-life date. Observe chemicals shelf-life date. If shelf-life of chemicals is exceeded do not use and dispose of appropriately. 18

3.Regeneration pump not working Light barrier on the drive shaft (XR24), tachometer generator (XR24-II) or electronics defect. Cable connections to Regeneration Unit loose or disconnected. Regeneration pump motor blocked. Regeneration time at potentiometer set at 0s. The impulse of the light barrier can be checked via Service PCB XR24 (Order No. 1700-031-00); check impulse of the LED drive motor (XR24-II) or replace electronics. Check all cable connections. Check the motor for any mechanical faults (e.g. turn the fan wheel). Set potentiometer to ca. 30s and check amount of regeneration material pumped. 19. Troubleshooting Tips XR 24 PRO Fault 1.Regeneration containers are empty after only a few hours 2.Films not fixed completely and veil on film Possible causes Regeneration time set too long. Electronics do not switch off. Regeneration containers empty. Pump operating time too short. Regeneration containers or their hoses are interchanged. Non-return valve of regeneration pump stuck or wrongly positioned. Regeneration hoses (from the Regeneration Unit to the X-ray developer) twisted. Film shelf-life exceeded. Chemicals shelf-life exceeded. Remedy Check regeneration timer setting. Replace electronic parts. Refill the regeneration containers. If there is still no improvement replace all chemicals in the machine. Re-set pump operating time at higher value. Drain chemicals from the machine and clean thoroughly. Connect the regeneration containers in the correct positions and refill with fresh chemicals. Clean non-return valve, replace if required. Check for correct positioning Lay the hoses with no twists or bends. Observe film shelf-life date. Observe chemicals shelf-life date. If shelf-life of chemicals is exceeded do not use and dispose of appropriately. 3.Regeneration pump not working Motor tachometer generator or electronics defect. Cable connections to Regeneration Unit loose or disconnected. Regeneration pump motor blocked. Check the impulses of the light barrier and the impulses of the roller drive or replace electronics. Check all cable connections. Check the motor for any mechanical faults (e.g. turn the fan wheel). 19