Quick Start Guide LIQ-QSG-Hx338, Rev M July Rosemount Hx338. Steam Sterilizable and Autoclavable ph Sensor

Similar documents
Rosemount 372. HF Resistant ph sensor. Quick Start Guide LIQ-QSG-372 Rev. C June 2017

Quick Start Guide LIQ-QSG-Hx438, Rev A June Rosemount Hx438. Amperometric Steam Sterilizable Dissolved Oxygen Sensors

Retractable ph/orp Sensor

Rosemount 3900/3900VP

Net Safety ST3 XChem Toxic Gas Sensor

Duct Mount. Installation Instructions

Dissolved Oxygen Sensors

Rosemount High Density Temperature Measurement. Simplify Your Project

Rosemount 0085 Pipe Clamp Sensor Assembly. Quick Start Guide , Rev CA June 2016

Rosemount 3200HP. ph Sensors. Instruction Manual LIQ-MAN-3200HP Rev. G April 2017

Rosemount 3490 Series 4 20 ma + HART Compatible Controller

PURSense Sanitary Flange Conductivity Sensors

STEAM TRAP HEALTH MONITORING SERVICE EXECUTIVE SUMMARY

ENDURANCE Conductivity Sensors

Rosemount 3051S Electronic Remote Sensors

SmartPower Solutions. Quick Start Guide , Rev CA March 2015

Emerson Plantweb Insight

Rosemount 396/396VP/397/398/398VP

Quick Start Guide , Rev. BB February Rosemount 1495 Orifice Plate Rosemount 1496 Orifice Flange Union

Advanced Diagnostics for HART Protocol Rosemount 3051S Scalable Pressure, Flow, and Level Solutions

Autocalibration Systems for In Situ Oxygen Analyzers

Rosemount 2110 Level Switch

Emerson Wireless SmartPower Solutions

Rosemount 214C Sensor. Quick Start Guide , Rev AF August 2017

Rosemount 0085 Pipe Clamp Sensor Assembly. Quick Start Guide , Rev DC August 2017

Rosemount 752 FOUNDATION Fieldbus Remote Indicator

Rosemount 1595 Conditioning Orifice Plate

Xi Advanced Electronics for Zirconium Oxide Oxygen Probes

Quick Start Guide , Rev DA December Rosemount Volume 1 Sensor Assembly

Rosemount 2100 Series

Rosemount 148 Temperature Transmitter. Quick Start Guide , Rev GA September 2016

Rosemount 2090P Pulp and Paper Pressure Transmitter

Flame Simulator: FS-UVIR-975. User Manual

Rosemount Functional Safety Manual. Manual Supplement , Rev AF March 2015

Flame Simulator: FS-IR-975. User Manual

SmartPower Solutions. SmartPower Solutions. Product Data Sheet , Rev AC January Contents

Rosemount 2090F Hygienic Pressure Transmitter

Rosemount 3051S MultiVariable Transmitter

Quick Start Guide , Rev EA December Rosemount 0065/0185 Sensor Assembly

Durafet Non-Glass ph Electrodes Specifications

Rosemount 4600 Oil & Gas Panel Pressure Transmitter

Sensor Wiring Instructions and Approval Drawings

Rosemount 2090F Hygienic Pressure Transmitter

Rosemount 3100 Series Ultrasonic Level Transmitters

RK Carbon Monoxide Transmitter Operator s Manual

Rosemount 3100 Series Ultrasonic Level Transmitters

Rosemount 3490 Series

Rosemount 3490 Series 4 20 ma + HART Compatible Controller

Rosemount 4500 Hygienic Pressure Transmitter Demonstrated best-in-class performance during SIP/CIP for process temperatures up to 400 F (204 C)

Multipurpose test lamp kit

RK-02 Multi Point Detector Operator s Manual

Free Chlorine Measuring System

Digital Salt / TDS Meter. 390a. The Value Leader TM

ph Measuring System Start-up and maintenance manual e Contents

19 mm Series Low Cost, Stainless Steel, Isolated Pressure Sensors

User Guide. ROSS Process Combination ph Electrode

CT-7 Series Toxic Gas Transmitter Operator s Manual

245 Sanitary Flow-Through Toroidal Conductivity Sensor. Instruction Manual LIQ_MAN_245 Rev. D January 2014

Xi Advanced Electronics for Zirconium Oxide Oxygen Probes

Durafet II and Meredian II Combination ph/orp Electrodes for Honeywell 7777 Mountings Series 7777 Specifications

MB-T Moisture analyzers

Catalytic Bead DrägerSensor Sensors for Fixed Gas Detectors

Product waste disposal - Protection of the environment:

82 Constant Voltage SPECIFICATIONS. 316L SS Pressure Sensor 19mm Diameter Package 0-100mV Output Absolute and Gage Temperature Compensated

Configuration Data Sheet

PLATE HEAT EXCHANGERS. Operation and Maintenance Manual

154BC SPECIFICATIONS FEATURES APPLICATIONS STANDARD RANGES

Instruction Manual. CAT 100 & CAT 200 Continuous Analyzer Transmitter Addendum for Hazardous Area Applications. 3 rd Edition 05/2007

Rosemount Pipe Clamp Sensor. Reference Manual , Rev BA February 2014

Note To The User: This product has updated features not yet mentioned in the product manual.

RK Hydrogen Sulfide Transmitter Operator s Manual

MiniCO, MiniH 2 S, MiniOX, MiniOX Remote, MiniCl 2. , and MiniClO 2. Responder Detectors Instruction Manual "! WARNING

ASPIRE Laboratory Aspirator

Instruction Manual LIQ-MAN-FCL-1056, Rev F July FCL with 1056 Transmitter

YARWAY PROCESS THERMOSTATIC STEAM TRAPS SERIES 151 AND AV5

154BC SPECIFICATIONS. Please refer to the 154N uncompensated and constant voltage datasheets for more information on different features of the 154.

RK Multi Point Detector Operator s Manual

Rosemount 2090F Hygienic Pressure Transmitter

86 Compensated SPECIFICATIONS. 316L SS Pressure Sensor Small Profile 0-100mV Output Absolute and Gage Temperature Compensated

Rosemount 4600 Oil & Gas Panel Pressure Transmitter. Quick Start Guide , Rev GA September 2017

PME. Series Enclosures. Battery Enclosure (BE) Field Upgrade Installation Instructions. Effective: January, Alpha Technologies

Hydrazine Analyzer Series 2171 Specifications

JENCO INSTRUMENTS, INC. JENCO ELECTRONICS, LTD ARJONS DR., SUITE C P.O. BOX LINKOU-117

7758 Pipeline Electrode Mountings for Industrial ph and ORP Measurements. Specification. Overview. Description

INSTALLATION INSTRUCTIONS

FAILURE TO ADHERE TO THESE WARNINGS CAN RESULT IN FIRE, SERIOUS INJURY, ELECTRICAL SHOCK AND/OR DEATH. Technical Service

Rosemount 2140:SIS Level Detector

RK-05 Carbon Monoxide Transmitter Operator s Manual

LAKEWOOD INSTRUMENTS MODEL 353 ORP/MILLIVOLT CONTROLLER

Model OEM-2 INSTRUCTIONS FOR USE

55-Gallon Dispenser Package

MSP340 Pressure Transducer

EVAPORATIVE COOL MIST HUMIDIFIER Instruction Manual

85 Compensated SPECIFICATIONS. 316L SS Pressure Sensor Small Profile 0-100mV Output Absolute and Gage Temperature Compensated

RK/ RK Carbon Monoxide Detector Operator s Manual

Rosemount 2090F Sanitary Pressure Transmitter with HART Protocol

RK-05 Carbon Monoxide Detector Operator s Manual

Rosemount 4500 Hygienic Pressure Transmitter

Karbon Kitchen Sink Tap

Dräger PIR 3000 Detection of flammable gases and vapours

Transcription:

Quick Start Guide LIQ-QSG-Hx338, Rev M July 2017 Rosemount Hx338 Steam Sterilizable and Autoclavable ph Sensor

Safety Information WARNING! Do not exceed temperature and pressure limitations of 266 F (130 C) and 43 psig (600 kpa, 4 bar). WARNING! HIGH PRESSURE AND TEMPERATURE HAZARD Before removing the senosr, reduce the process pressure to 0 psig and cool down the process temperature. Failure to reduce the pressure and temperature may cause serious injury to personnel. WARNING! CORROSIVE SUBSTANCE Internal electrolyte solution may cause skin or eye irritation. WARNING! TOXIC AND CORROSIVE SUBSTANCE Hydrofluoric acid and the mixture of hydrochloric acid and thiourea is toxic and highly corrosive. Avoid skin contact and wear protective gloves. Use only in a well-ventilated area. Do not inhale fumes. In case of an accident, consult a doctor immediately. CAUTION! EQUIPMENT DAMAGE The wetted sensor materials may not be compatible with process composition and operating conditions. Application compatibility is entirely your responsibility.

Contents Contents Chapter 1 Plan...1 1.1 Unpacking...1 1.2 Storage and handling... 1 1.3 Specifications...1 Chapter 2 Install...3 2.1 Installing the sensor... 3 Chapter 3 Calibration and maintenance...9 3.1 Two point buffer calibration... 9 3.2 Recommended ph sensor standardization... 9 Chapter 4 Troubleshooting... 11 4.1 Cleaning or replacing the electrodes... 11 4.2 Accessories... 11 Rosemount Hx338 ph Sensor i

Contents ii Quick Start Guide

Plan 1 Plan 1.1 Unpacking Electrode preparation 1. Remove electrode from shipping container. 2. Remove the protective boot covering the electrode bulb. 3. Rinse away salt film with clean water and then shake the electrode so that the internal solution fills the bulb, thus removing any air trapped there. Note Do not allow lubricant to coat electrode bulb or reference junction. If it does, wipe it clean before installation. 1.2 Storage and handling Proper storage and handling instructions of the electrodes must be followed to ensure correct measurements. 1. Store the electrodes in their original shipping containers until needed. 2. Do not store at temperatures below -5 C (23 F). 3. Store the electrodes with a protective cap containing KCI solution (PN 9210342). 4. For overnight storage, immerse the sensor in tap water or 4 ph buffer solution. 5. Use the startup and calibration procedure described in Section 3.1 to properly calibrate the sensors. 1.3 Specifications Table 1-1: Specifications Hx338 ph Sensor Specifications Operating Temperature Range Maximum Pressure ph Range Wetted Materials Process Connections Cable Connector Minimum Conductivity of Sample Materials and Units 32 to 266 F (0 to 130 C) 43 psig (600 kpa abs, 4 bar) 2 to 12 (0 to 14 short periods) Glass, EPDM (FDA approved), Ceramic (Diaphragm) PG 13.5 Threads Single Pole S7 Connector Plug or VP6 Cable Connection 50 µs/cm Rosemount Hx338 ph Sensor 1

Plan 2 Quick Start Guide

Install 2 Install 2.1 Installing the sensor Figure 2-1: Rosemount Hx338 dimensional drawing Rosemount Hx338 ph Sensor 3

Install A. VP 6.0 plug B. SW17 free spinning PG 13.5 thread C. Washer D. O-ring Procedure 1. Mount the sensor in a vertical pipe run with flow from top to bottom. If the sensor must be mounted in a horizontal pipe run, orient the sensor in the or 9 o'clock position. 2. Orient the sensor within 80 of vertical. See Figure 2-2. Figure 2-2: Sensor Orientation 3. Wire the sensor to the transmitter. See wiring diagrams below. 4 Quick Start Guide

Install Figure 2-3: Rosemount Hx338-01-72 wiring to Rosemount 1066 Transmitter Figure 2-4: Rosemount Hx338-01 wiring to Rosemount 1066 Transmitter Rosemount Hx338 ph Sensor 5

Install Figure 2-5: Rosemount Hx338-01-72 wiring to Rosemount 5081 Transmitter Figure 2-6: Rosemount Hx338-01 wiring to Rosemount 5081 Transmitter 6 Quick Start Guide

Install Figure 2-7: Rosemount Hx338-01-72 wiring to Rosemount 1057 Transmitter Figure 2-8: Rosemount Hx338-01 wiring to Rosemount 1057 Transmitter Rosemount Hx338 ph Sensor 7

Install Figure 2-9: Rosemount Hx338-01-72 wiring to Rosemount 1056 and Rosemount 56 Transmitters Figure 2-10: Rosemount Hx338-01 wiring to Rosemount 1056 and Rosemount 56 Transmitters 8 Quick Start Guide

Calibration and maintenance 3 Calibration and maintenance 3.1 Two point buffer calibration Prerequisites Select two stable buffer solutions, preferably ph 4.0 and 7.0 (ph buffers other than ph 4.0 and ph 7.0 can be used as long as the ph values are at least two ph units apart). Note A ph 7.0 buffer solution reads an mv value of approximately zero, and ph buffers read approximately ±59.1 mv for each ph unit above or below ph 7.0. Check the ph buffer manufacturer specifications for millivolt values at various temperatures, as it may affect the actual value of the buffer solution mv/ph value. Procedure 1. Immerse sensor in the first buffer solution. Allow sensor to equilibrate to the buffer temperature (to avoid errors due to temperature differences between the buffer solution and sensor temperature) and wait for reading to stabilize. The value of the buffer can now be acknowledged by the transmitter. 2. Once the first buffer has been acknowledged by the transmitter, rinse the buffer solution off of the sensor with distilled or deionized water. 3. Repeat steps 1 and 2 using the second buffer solution. The theoretical slope value, according to the Nernst equation for calculating ph, is approximately 59.17 mv/ph. Over time the sensor will age, both in the process and in storage, and this will result in reduced slope values. To ensure accurate readings, it is recommended that the electrode be replaced when the slope value falls below 47 to 49 mv/ph. 3.2 Recommended ph sensor standardization For maximum accuracy, the sensor can be standardized on-line or with a process grab sample after a buffer calibration has been performed and the sensor has been conditioned to the process. Standardization accounts for the sensor junction potential and other interferences. Standardization will not change the sensor s slope, but will simply adjust the transmitter s reading to match that of the known process ph. Rosemount Hx338 ph Sensor 9

Calibration and maintenance 10 Quick Start Guide

Troubleshooting 4 Troubleshooting 4.1 Cleaning or replacing the electrodes Electrodes should respond rapidly. Sluggishness, offsets, and erratic readings are indicators that the electrodes may need cleaning or replacement. 1. Use a mild non-abrasive detergent to clean and remove oil deposits from the electrode. 2. Soak electrodes for 30 to 60 minutes in a 5% hydrochloric acid solution to remove scale deposits. Temperature effect on life expectancy: If glass electrode life expectancy is 100% at 25 C (77 F), then it will be approximately 25% at 80 C (176 F), and approximately 5% at 120 C (248 F). 4.2 Accessories Part number Description 24281-00 15 ft (4.6 m) cable with mating VP8 connector 24281-01 25 ft (7.6 m) cable with mating VP8 connector 24281-02 2.5 ft (0.8 m) cable with mating VP8 connector 24281-05 4 ft (1.3 m) cable with mating VP8 connector 24281-06 10 ft (3 m) cable with mating VP8 connector 24281-07 20 ft (6.1 m) cable with mating VP8 connector 24281-08 30 ft (9.1 m) cable with mating VP8 connector 23645-06 15 ft (4.6 m) cable with mating VP8 connector with BNC on transmitter end 23645-07 15 ft (4.6 m) cable with mating VP8 connector with bare wires on transmitter end 9160494 3.3 ft (1 m) cable with AS7 connector, bare wire on analyzer end 9160495 16.4 ft (5 m) cable with AS7 connector, bare wire on analyzer end 9160496 32.8 ft (10 m) cable with AS7 connector, bare wire on analyzer end 9160498 16.4 ft (5 m) cable with AS7 connector, BNC on analyzer end 9160478 Insertion assembly, 70 mm insertion, for 120 mm sensor, G 1-1/4 in. 9160483 15 degree weld-in socket, G 1-1/4 in. thread, 44 mm 9260484 Service kit for insertion mounting assembly 9210012 Buffer solution, ph 4.01, 16 oz 9210013 Buffer solution, ph 6.86, 16 oz 9210014 Buffer solution, ph 9.18, 16 oz Rosemount Hx338 ph Sensor 11

Troubleshooting 12 Quick Start Guide

Troubleshooting Rosemount Hx338 ph Sensor 13

Troubleshooting 14 Quick Start Guide

Troubleshooting Rosemount Hx338 ph Sensor 15

LIQ-QSG-Hx338 Rev M 2017 www.emerson.com/rosemountliquidanalysis Emerson Automation Solutions 8200 Market Blvd Chanhassen, MN 55317 Toll Free +1 800 999 9307 F +1 952 949 7001 liquid.csc@emerson.com www.emerson.com/rosemountliquidanalysis EUROPE Emerson Automation Solutions Neuhofstrasse 19a P.O. Box 1046 CH-6340 Baar Switzerland T + 41 (0) 41 768 6111 F + 41 (0) 41 768 6300 liquid.csc@emerson.com www.emerson.com/rosemountliquidanalysis MIDDLE EAST AND AFRICA Emerson Automation Solutions Emerson FZE Jebel Ali Free Zone Dubai, United Arab Emirates, P.O. Box 17033 T +971 4 811 8100 F +971 4 886 5465 liquid.csc@emerson.com www.emerson.com/rosemountliquidanalysis ASIA-PACIFIC Emerson Automation Solutions 1 Pandan Crescent Singapore 128461 Singapore T +65 777 8211 F +65 777 0947 liquid.csc@emerson.com www.emerson.com/rosemountliquidanalysis 2017 Rosemount. All rights reserved. The Emerson logo is a trademark and service mark of Emerson Electric Co. Rosemount is a mark of one of the Emerson Process Management family of companies. All other marks are the property of their respective owners. The contents of this publication are presented for information purposes only, and, while effort has been made to ensure their accuracy, they are not to be construed as warranties or guarantees, express or implied, regarding the products or services described herein or their use or applicability. All sales are governed by our terms and conditions, which are available on request. We reserve the right to modify or improve the designs or specifications of our products at any time without notice.