1 WG EFA NHS WALES Estates and Facilities Alert Date: 19 December 2011 Cathays Park, Cardiff CF10 3NQ Parc Cathays, Caerdydd CF10 3NQ Electrolux (Dometic) absorption pharmacy / drug refrigerators, all models To: Chief Executives, LHBs Chief Executives, NHS Trusts NHS Wales Shared Services Partnership - Contractor Services A list of people who need to have early sight of this information is given in the Alert Estates and Facilities Alerts are agreed by NHS representatives and issued throughout the United Kingdom The Alert is endorsed by the Welsh Government as being relevant to NHS Wales and is issued by: Neil Davies, Assistant Director, NHS Wales Shared Services Partnership Facilities Services Contact point in Wales: For advice see details on the alert Nigel Porter Principal Engineer NHS Wales Shared Services Partnership Facilities Services P. O. Box 182, Bevan House Lambourne Crescent Cardiff CF14 5GS Tel: nigel.porter@wales.nhs.uk Alerts are available in full text at:
2 Estates and Facilities Alert Action Ref: EFA/2011/007 Issued: 16 December December in Wales Device Electrolux (Dometic) absorption pharmacy / drug refrigerators, all models Problem Risk of ammonia gas leakage from Electrolux (Dometic) absorption pharmacy / drug refrigerators that are: o not installed correctly o not used in line with manufacturer guidelines o not adequately maintained or inspected at regular intervals o displaying signs of impact damage, corrosion or ammonia leakage o displaying signs of failure to cool o beyond their recommended service life Ammonia gas is toxic and exposure / inhalation can cause harm. Action Absorption refrigerators should be installed and used in line with Dometic guidelines (see Appendix) and covered by a preventative maintenance programme which includes visual inspection. Any displaying signs of impact damage, corrosion or ammonia leakage should be removed from use before contacting Dometic. Any that have ceased to function correctly (i.e. maintain drugs at correct temperature) should be removed from use while awaiting repair. Any aged 8 to 10 years should be considered for replacement. Risk assessments should be carried out on the use of absorption refrigerators in all areas, particularly those occupied by vulnerable patients. Action by Managers responsible for departments which have this equipment in use. Managers and staff responsible for the procurement, supply, maintenance and asset management of pharmacy / drug fridges. Contact Dometic Technical Support technical@dometic.co.uk Telephone: Template version 5: 14 October 2011 Page 1 of 3
3 Issued: 16 December 2011 Ref: EFA/2011/007 Problem 1. An adverse incident has been reported to Health Facilities Scotland involving an Electrolux (now Dometic) pharmacy / drug absorption refrigerator which leaked ammonia gas from the cooling system into a high dependency patient care area. As a result some patients and staff required emergency medical treatment. 2. The leakage was due to a sudden de-pressuring of the cooling system. This could have been caused by a corrosion related blockage in the cooling system leading to pressure increase followed by sudden de-pressuration through a leak or pipe work rupture. 3. The refrigerator was over 10 years old. Current models are DS301H, DS601H; older models include DS300H, DS600H, RA122H, MR60 and EAxxx (where x indicates a number) 4. Dometic state that failures of this nature are nearly always indicated in the preceding days or week with a loss of cooling. Any such indications were not noticed or acted upon in this case. 5. This alert supersedes earlier advice on the risk of ammonia leak from Electrolux drug refrigerators. For specific (superseded) alerts / notices, please see Additional Information for the relevant administration. Action 6. This notice should be brought to the attention of all appropriate managers, staff and users. 7. Ensure Dometic absorption refrigerators are installed and used in line with Technical Guidelines for Dometic Pharmacy Refrigerators (see Appendix) and the manufacturer s manuals. 8. Ensure Dometic absorption refrigerators are put into a regular preventative maintenance programme which includes visual inspection of pipe work for signs of impact damage, corrosion or ammonia leakage. Dometic recommend absorption refrigerators 1 to 6 years old should be inspected annually and those over 6 years old should be inspected every 6 months. 9. Any displaying signs of impact damage, corrosion, or ammonia leakage should be disconnected from electrical mains supply and removed from use before contacting Dometic. 10. Where a Dometic absorption refrigerator has ceased to function correctly (i.e. maintain drugs at the correct temperature) it should be disconnected from electrical supply and removed from use pending repair. 11. Dometic recommend absorption refrigerators 8 to 10 years old should be considered for replacement as they will start to exhibit loss of performance. They advise that it is not economical to replace the cooling system of absorption refrigerators over 10 years old. Fridge age is identified by serial no. (e.g. 614xxxxx manufactured week 14 of 2006). Replacement plans for these refrigerators (with budget considerations) should be implemented. 12. Risk assessments should be conducted on the use of absorption refrigerators in all areas but particularly in patient occupied areas. Consideration should be given to the manufacturer s installation requirements, the risks from exposure to ammonia, the type / mobility of patients and local evacuation procedures. Page 2 of 3
4 Issued: 16 December 2011 Ref: EFA/2011/007 Suggested Onward Distribution Accident & Emergency Accommodation Services Anaesthetics Blood Transfusion Coronary Care Units Contacts Day Surgery Dental Dental Hospitals Estates/Facilities General Dental Practitioners General Medical Practitioners Health & Safety Hospices Intensive Therapy Units Neonatal Nursing Occupational Health Operating Departments Pharmacy Risk Management Special Care Baby Units Stores Wards Dometic Technical Support, Dometic UK Ltd, The Brewery, Blandford St Mary, Dorset DT11 9LS Telephone: Additional information for Wales The above sections of this Alert were compiled by Health Facilities Scotland and distributed nationally without modification. Enquiries should quote reference number WG EFA/2011/007 and be addressed to: Nigel Porter Prif Beirianydd Principal Engineer Partneriaeth Cydwasanaethau GIG Cymru Gwasanaethau Cyfleusterau N H S Wales Shared Services Partnership Facilities Services Ffon/Phone e-bost/ Nigel.Porter@Wales.nhs.uk Estates and Facilities Alerts are available in full text on the Public Health Wales website at Crown Copyright 2011 Addressees may take copies for distribution within their own organisations Page 3 of 3
5 Models: DS301HFS and DS601HFS Technical Guidelines for Dometic Pharmacy Refrigerators
6 Contents: Why Dometic are No DS 301HFS and DS 601HFS Features...2 Dometic Pharmacy Fridges...3 Absorption Refrigeration...3 Installation Requirements...3 Day to Day Use...3 Defrosting...4 Warranty...4 Maintenance...5 Older Units...5 Ammonia Leakage...5 Technical Data...6 Why Dometic are No years experience in production Silent in operation Best cooling performance among absorption products Lowest energy consumption for absorption products Intelligent thermostatic control Fuzzy Logic Compact size and design makes them adaptable to any environment Perfect solution for temperature sensitive medication Equipped with thermal cut-out device Production recognised to ISO 9001 standards Environmental policy based on ISO standards DS 301HFS DS 601HFS DS 301HFS and DS 601HFS Features Internal Fan & Door Alarm Interior Drawer Fronts Silent Operation New Thermostatic Control Digital Thermometer Lockable Manual Defrost Optional Wall Bracket DS 301HFS 28 litre capacity H 58.0 x W 42.2 x D 39.3mm DS 601HFS 53 litre capacity H 59.2 x W 48.6 x D 49.4 mm 2.
7 Dometic Pharmacy Fridges Production Dates from/to Current Models DS301H 2009 DS601H 2009 Older Models DS300H 2004 DS600H 2004 RAxxx 1996 MR EAxxx /2011 Absorption Refrigeration DS601HFS Absorption refrigerators primarily differ from normal domestic appliances in that they do not utilise a compressor. Instead they use a heat input to create a vapour cycle within a closed loop. This makes them totally silent in operation which is a unique benefit in areas where people are resting or sleeping. As with any technology, there is a trade off of benefits against limitations and in operation absorption refrigerators are slightly more sensitive than compressor fridges to their environment and manner of use. However, if the simple guidelines given below are adhered to, your Dometic fridges will give many years of satisfactory use. Installation Requirements As with any fridge, whether absorption or compression, the two key points are: Condenser Boiler (insulated) Heating Element 1) Ventilation Refrigerators have to remove heat from the interior as well as dissipating the heat input to the process itself. This is done through the unit s condenser (see diagram) and so it is vital that this area of the fridge has sufficient space (see manual) and is not adjacent to any heat sources e.g. radiators. Thought should also be give to the heat input from incident sunlight or spotlighting. 2) Levelling The cooling circuit is finely balanced and gravity assisted. Operating the fridge outside of 3 degrees of level will impair performance and reduce the life of the unit. Operating the unit within 1-3 degrees of level will provide normal operation but may still impact on the working lifespan. If at all possible, the unit should be operated within 1 degree of level at all times (this is not difficult to achieve, most hospital floors will be within this and the unit will stand perfectly flat when unobstructed). Day to Day Use Absorption fridges are relatively lightweight and easily moved as operational requirements dictate. Care should be taken when moving units that they are not subjected to sudden impacts. Also, the new location must meet the ventilation and levelling requirements described above. If the unit is laid on its side or front (never on its back!) during moving, then it must be stored in the upright position for 24 hours before being returned to service. 3.
8 Defrosting Frost will gradually form on the cooling surfaces and it must not be allowed to become too thick since it acts as an insulator and interferes with the cooling. To prevent unplanned increases in temperature, the defrost is a manual operation. We recommend that the appliance is defrosted every 3 weeks, or depending on the ice quantity on the cooling fins. While defrosting, the appliance should be switched off by disconnecting the refrigerator from the electricity supply and leave the cabinet door open. The frost will melt and the defrost water will run through the drain tube from the drip channel into an evaporating vessel located at the bottom rear of the refrigerator (see diagram). After defrosting the cabinet interior should be wiped out with a clean, damp cloth. Warranty All Dometic products sold after January 1st 2010 are offered with a 2 year warranty which begins on the date of sale to the enduser. Any faults that develop within this period will be dealt with by providing a temporary replacement while the unit is repaired at Dometic s works. The procedure for the Service Exchange Plus (S/X+) scheme is as follows, User reports Problem to Dometic UK If problem cannot be resolved over the phone, details taken for S/X+ scheme Returns form ed/faxed to user Faulty unit repaired and tested Temporary unit despatched, faulty unit collected (user informed of delivery times) User returns completed form by fax/ Original unit returned, temporary item collected (user informed of delivery times) Temporary unit tested and returned to SX+ stock Figure 1 Service Exchange Plus (S/X+) The S/X+ scheme aims to minimise the time the user is without a fridge in the unlikely event a fault should occur. In order for us to provide this high level of service, it is vital that the following points are noted and adhered to, 1) Please complete the returns form as fully and as quickly as possible. 2) Please ensure that there is a single person dealing with this issue or that details of the issue are passed from one shift to the next where appropriate. 3) Please provide a specific ward address or, if being delivered to a general receiving area, ensure that someone is aware of the imminent deliveries. 4) Please use the packaging provided (i.e. the box in which the temporary replacement arrives) to return the failed unit. The courier will not carry an unpackaged item. 5) We will provide contact details (including ) to keep you informed of the expected delivery times. Please ensure that the contact details provided are correct and relevant! 6) Please remember that the temporary unit is provided as a courtesy and should NOT be mistreated or returned broken or otherwise damaged. Because we test and if necessary repair all returned temporary units before re-use, excessive repairs are time consuming and impair the overall service. 7) If possible, hand the issue to a member of the maintenance/service team from the first point of initial failure and we will deal with them directly. 4.
9 Maintenance Dometic recommends that hospitals have some form of inspection/maintenance/service arrangement in place for when the warranty period expires. This could be hospital maintenance staff, a third party maintenance company or the local Dometic Approved Network (DAN) engineer (visit for details of local DAN engineers). There are no routine maintenance procedures necessary on a Dometic pharmacy fridge other than inspection. Attention should be paid to ventilation and levelling as well as a visual inspection of the rear of the fridge to look for impact or corrosion damage. There may be breakages or failure of components e.g. the circulation fan and these parts are readily available and simple to change. Inspection routine: Yrs 1-6: Annual Yrs 6- : 6 monthly Training can be provided to maintenance staff or Estates managers please contact Dometic UK for further details. Older Units The average lifespan of a Dometic pharmacy fridge s cooling system is around 8-10 years and after this time will start to exhibit loss of performance. This lifespan can be adversely affected by any number of operational factors and so it is not uncommon for units to fail at 6 years onwards. Parts are available for the majority of units, even those older than 10 years. However, it is never economical to change the cooling units if they have failed and we recommend a new unit in this case. We recommend that the inspection routine is increased from once to twice a year when the units have been in service for 6 years. We would also recommend that units are considered for replacement once they have been in service for 8-10 years. Older units did not benefit from the same sophisticated software that is utilised as standard in modern units and can suffer from rupture of the cooling system due to overheating. The overheating is invariably the result of a corrosion-related blockage in the cooling circuit and whereas a modern unit would sense the heat imbalance and shut down, an older unit will continue to work as it tries to refrigerate. Failures of this nature are nearly always indicated in the preceding weeks/days with a loss of cooling. Switching the unit off/on, turning the unit upside down or shaking it may restore performance but this will only be a temporary solution (if it works at all) and will leave the fridge in a potentially unsafe state. Units that are exhibiting a loss of performance should be switched off immediately, put out of service and Dometic informed. Ammonia Leakage See also NHSE HN (2003) 04 & SAN(SC)98/09 The cooling unit of a Dometic Pharmacy Fridge contains approximately 100g of ammonia along with 150g of water and some corrosion inhibitor. There are 2 methods of failure of the cooling unit, as described below, 1) Rupture Older units suffering from internal corrosion leading to blockage have the potential to overheat leading to a rupture, inevitably in the immediate vicinity of the heating element (see diagram on p.3). The mineral insulation around the boiler will absorb the major part of the rupture although it is likely to burst which can disperse the insulation around the area. Hard impact damage could also potentially cause a rupture although this is unlikely unless it is of considerable force. 2) Micro-leakage Internal corrosion may cause weakening of welds leading to microscopic cracks. Ammonia will be smelt but it might not be immediately obvious to staff that the fridge is the source. The leak may be so small that the fridge does not exhibit loss of performance for days. Impact damage may also cause a micro-leak although this is very unlikely. Should there be a leakage of ammonia at any time, follow standard NHS procedure to ensure the safety of any staff or patients in the vicinity. Switch off and unplug the affected unit and contact Dometic UK. 5.
10 Technical data DS301 H DS601 H Gross/net volume 27/23 53/51 Outer dimension 580x422x x486x494 Power input (W) Energy consumption (kw/24hr) ,05 Weight 16kg 20kg Cabinet temperature 2 C - 8 C Defrosting Manual DS301HFS DS601HFS For further information and a full list of distributors, please contact: Jon Race or Leigh Palmer: retail@dometic.co.uk Technical Support: technical@dometic.co.uk
Estates and Facilities Alert
Estates and Facilities Alert Action Ref: EFA/2010/007 Issued: 8 July 2010 Device Window blinds with looped cords or chains. All types. Problem Action Looped cords and chains on window blinds can present
More informationUser Guide and Important Warranty Information. Suprima HE. Condensing Central Heating Boiler. Please read this document and keep it safe.
User Guide and Important Warranty Information Suprima 30-80 HE Condensing Central Heating Boiler Please read this document and keep it safe. It will help you out should your boiler need attention. Introduction
More informationOPERATING INSTRUCTIONS & TROUBLESHOOTING GUIDE ECO MODELS FLUID WARMING CABINETS BLANKET WARMING CABINETS
OPERATING INSTRUCTIONS & TROUBLESHOOTING GUIDE ECO MODELS FLUID WARMING CABINETS & BLANKET WARMING CABINETS QED Scientific Ltd. Unit 21 Botany Business Park Whaley Bridge High Peak SK23 7AD Tel: 01663
More informationElectrical Safety Policy and Management System
TITLE: PAGE: 1 OF 14 Housing Solutions Limited Electrical Safety Policy and Management System Page 1 of 15 TITLE: PAGE: 2 OF 14 FOREWORD This document details the system for the management of electrical
More informationBLOOD PRODUCT STORAGE.
BLOOD PRODUCT STORAGE 082010 www.labcold.com Welcome to the Labcold Blood Product Storage brochure incorporating our new range of platelet incubators and agitators. Here at Labcold, we pride ourselves
More informationNHS Tayside. Fire Safety Policy
NHS Tayside Fire Safety Policy Policy Manager Mark Anderson, Head of Property Policy Group Fire Safety Policy Review Group Policy Established Last Updated April 2014 Policy Review Period/Expiry April 2017
More informationUser Guide and Important Warranty Information. Suprima HE Range. Condensing Central Heating Boiler. You can rely on
You can rely on User Guide and Important Warranty Information Suprima HE Range Condensing Central Heating Boiler Please keep these instructions in a safe place. If you move house, please hand them over
More informationSAFETY ACTION NOTICE By arrangement with NSS Health Facilities Scotland
By arrangement with NSS Health Scotland SUMMARY Page 1 of 5 Pages In order to control the risk of fire while maximising the considerable benefits of using alcohol-based handrubs, co-ordinated risk assessments
More informationBaxi Combi 105 HE. User s Operating Instructions. Gas Fired Wall Mounted Condensing Boiler
User s Operating Instructions Baxi Combi 05 HE Gas Fired Wall Mounted Condensing Boiler Please keep these instructions in a safe place. If you move house, please hand them over to the next occupier. Contents
More informationTechnical description. Atec.
www.thermia.com Thermia Värmepumpar is not liable or bound by warranty if these instructions are not adhered to during installation or service. The English language is used for the original instructions.
More informationUser Guide and Important Warranty Information. Promax System HE Plus A Range. Condensing System Boiler
You can rely on User Guide and Important Warranty Information Promax System HE Plus A Range Condensing System Boiler Please keep these instructions in a safe place. If you move house, please hand them
More informationCombi Eco Elite Range. User s Guide and Important Warranty Information
User s Guide and Important Warranty Information Combi Eco Elite Range Please keep these instructions in a safe place. If you move house, please hand them over to the next occupier. Baxi Heating UK Ltd
More informationUltracom cxi. Instructions for Use. High Efficiency Condensing System Boilers. To b e l e f t w i t h t h e u s e r.
Ultracom cxi High Efficiency Condensing System Boilers Instructions for Use To b e l e f t w i t h t h e u s e r WARNING GAS LEAK OR FAULT Turn off the gas emergency control valve immediately. Eliminate
More informationFire Safety Management
Fire Safety Management Procedure () FIRE SAFETY MANAGEMENT PROCEDURE Contents Section 1: Management of Fire Safety Procedure in Operation Section 2 Organisational responsibilities Section 3: Emergency
More informationvolume 11 environmental assessment section 2 environmental impact assessment Part 7 ha 218/08
DESIGN MANUAL FOR ROADS AND BRIDGES volume 11 environmental assessment section 2 environmental impact assessment Part 7 ha 218/08 glossary of terms Used in DMRB Volume 11, Sections 1 and 2 SUMMARY This
More informationReport For NHS Shetland
Report For NHS Shetland FIRE EMERGENCY MANAGEMENT PLAN GILBERT BAIN HOSPITAL C.S. TODD & ASSOCIATES LTD August 2014 FIRE EMERGENCY MANAGEMENT PLAN GILBERT BAIN HOSPITAL Report by: K. LOGIE GIFireE C.S.
More informationImportant Notes: Please read carefully before installing your refrigerator.
Model: Important Notes: Please read carefully before installing your refrigerator. This Labcold product has been made with great care, tested thoroughly before being sent to you, and is expected to give
More informationBaxi Megaflo System HE IE Range. User s Operating Instructions & Important Warranty Information. Gas Fired Wall Mounted Condensing Boiler
User s Operating Instructions & Important Warranty Information Baxi Megaflo System HE IE Range Gas Fired Wall Mounted Condensing Boiler Please keep these instructions in a safe place. If you move house,
More informationPerforma 24 Eco HE. User Guide and Important Warranty Information. Condensing Combination Boiler
0 4 You can rely on User Guide and Important Warranty Information Performa 4 Eco HE Condensing Combination Boiler Please keep these instructions in a safe place. If you move house, please hand them over
More informationProperty Health and Safety Compliance at Housing and Education Centres
Property Health and Safety Compliance at Housing and Education Centres Title: Property Health and Safety Reference number: Compliance at Housing and Education Centres Status Final Publication Internal
More informationUser s Operation Instructions & Important Warranty Information
United Kingdom en User s Operation Instructions & Important Warranty Information High-efficiency wall-hung boilers Baxi Avanta Plus 18s - 24s - 30s 24c - 28c - 35c - 39c Please keep these instructions
More informationUser s Operation Instructions & Important Warranty Information. High-efficiency wall-hung boilers Baxi Avanta Heat Only 15h - 18h - 24h - 30h
United Kingdom en User s Operation Instructions & Important Warranty Information High-efficiency wall-hung boilers Baxi Avanta Heat Only 15h - 18h - 24h - 30h Please keep these instructions in a safe place.
More informationCorporate. Laundry SOP. Document Control Summary
Corporate Laundry SOP Document Control Summary Status: Version: Author/Title: Owner/Title: Approved by: Ratified: Related Trust Strategy and/or Strategic Aims Implementation Date: Review Date: Key Words:
More informationDOMESTIC ICE MACHINE USER INSTRUCTION MANUAL MODEL TS-061
DOMESTIC ICE MACHINE USER INSTRUCTION MANUAL MODEL TS-061 Page 1 of 8 Introduction This instruction manual covers the operation of versions of the Domestic Ice Machine Model TS-061 manufactured after April
More informationBaxi System 100 HE Plus. User s Operating Instructions. Wall Mounted Powered Flue Condensing Gas Fired Central Heating Boiler
User s Operating Instructions Baxi System 100 HE Plus Wall Mounted Powered Flue Condensing Gas Fired Central Heating Boiler Please keep these instructions safe. Should you move house, please hand them
More informationOur Community Safety Information and Emergency Instructions
Our Community Safety Information and Emergency Instructions South Hook LNG Terminal, Herbrandston Please read this booklet and keep it in a safe place. Ensure the booklet is left at the property to which
More informationGrant Aerona³. Air to Water High Efficiency Heat Pump Range. User Instructions
Grant Aerona³ Air to Water High Efficiency Heat Pump Range User Instructions UK DOC 0112 Rev 1.0 April 2017 Special Tex t Formats The following special text formats are used in this manual for the purposes
More informationPanel Radiators. Installation and operating instructions. ThermaRad Classic.
ThermaRad Hurlcon Panel Radiators Installation and operating instructions ThermaRad Classic ThermaDesign ThermaStyle www.hurlconheating.com.au enquiries@hurlconheating.com.au Melbourne: 03 9554 2275 Sydney:
More informationFIRE SAFETY POLICY June 2014
FIRE SAFETY POLICY June 2014 Policy Title Fire Safety Policy Reference Number Corpo11/002 Implementation Date June 2014 1 st Review Review Date June 2017 Responsible Officer Director of Performance and
More informationBaxi Combi 133 HE Plus Baxi Combi 100 HE Plus Baxi Combi 80 HE Plus. User s Operating Instructions
User s Operating Instructions Baxi Combi 133 HE Plus Baxi Combi 100 HE Plus Baxi Combi 80 HE Plus Gas Fired Wall Mounted Condensing Combination Boiler Please keep these instructions safe. Should you move
More informationSafety Alert New Gas Regulation 2013
Safety Alert New Gas Regulation 2013 Latest Gas Safe Regulation : From 1 January 2013 onwards, any of our Gas Safe registered engineers will turn gas supply to the boiler off in any property that has flues
More informationR410a Installation and fault finding training
R410a Installation and fault finding training Contents: General introduction Disassembly and reassembly Installation guide Trouble shooting General introduction: 1. Product Introduction Development Specification
More informationFire Safety Strategy
SH HS 02 Summary: Keywords (minimum of 5): (To assist policy search engine) Target Audience: The aim of this strategy is the provision and maintenance of a safe environment for all staff, visitors and
More information3000 Series. UK Manual Accessories Warranty Registration
3000 Series UK Manual Accessories Warranty Registration Your Ebac Dehumidifier has been specifically designed to cope with Britain s unique and unsettled climate by Europe s No.1 manufacture of dehumidifier
More informationModel Carbon Monoxide (CO) Alarm Response Policy
Published on MTAS (http://www.mtas.tennessee.edu) January 26, 2019 Model Carbon Monoxide (CO) Alarm Response Policy Summary: This model policy was created to provide procedures for personnel to follow
More informationFire Safety Policy & Fire Safety Management System Version October 2013 Emergency Evacuation Procedures
Emergency Evacuation Procedures Contents 1. Raising the fire alarm 2. Schedule of patient evacuation 3. Designated fire officer roles 3.1 The Fire Team Leader 3.2 Security personnel 3.3 Shift Technicians
More informationCORRECTIONS WITHIN DESIGN MANUAL FOR ROADS AND BRIDGES AUGUST 2009
DESIGN MANUAL FOR ROADS AND BRIDGES CORRECTIONS WITHIN DESIGN MANUAL FOR ROADS AND BRIDGES AUGUST 2009 SUMMARY OF CORRECTION HA 200/08 VOLUME 11, SECTION 1, PART 1 AIMS AND OBJECTIVES OF ENVIRONMENTAL
More informationBurton Hospitals NHS Foundation Trust. On: 27 April Review Date: March Corporate / Directorate. Department Responsible for Review:
POLICY DOCUMENT Burton Hospitals NHS Foundation Trust FIRE POLICY Approved by: Trust Executive Committee On: 27 April 2016 Review Date: March 2019 Corporate / Directorate Clinical / Non Clinical Corporate
More informationUser Guide. Promax SL Condensing Central Heating Boiler
United Kingdom en User Guide Condensing Central Heating Boiler Promax SL 12-15 - 18-24 Please keep these instructions in a safe place. If you move house, please hand them over to the next occupier. Contents
More informationFire Safety Policy. Document Author: Director of Estates, Fleet & Facilities Management
Fire Safety Policy Document Author: Director of Estates, Fleet & Facilities Management Date Approved: January 2018 Document Reference Fire Safety Policy January 2020 Version V: 2.1 Responsible Committee
More informationElectrical Safety Policy
Electrical Safety Policy By acting with a business head and social heart, together we achieve our mission to enhance life chances by helping people to fulfil their potential and live better, more independent
More informationSPLIT TYPE. Instruction manual
SPLIT TYPE Instruction manual ENGLISH Instruction manual Page 1~24 To obtain the best performance and ensure years of trouble free use, please read this instruction manual completely. 1 SAFETY PRECAUTION
More informationvolume 11 environmental assessment section 2 environmental impact assessment Part 2 ha 202/08 environmental impact assessment
DESIGN MANUAL FOR ROADS AND BRIDGES volume 11 environmental section 2 environmental impact Part 2 ha 202/08 environmental impact SUMMARY This Advice Note provides guidance for undertaking statutory and
More informationS T O R E R A N G E C Y L I N D E R & G R A N T FEATURES: GRANT MONOWAVE UNVENTED DIRECT AND INDIRECT SINGLE-COIL CYLINDERS
G R A N T C Y L I N D E R & S T O R E R A N G E High Efficiency Direct and Indirect Hot Water Storage Solutions FEATURES: GRANT MONOWAVE UNVENTED DIRECT AND INDIRECT SINGLE-COIL CYLINDERS GRANT DUOWAVE
More informationA GUIDE TO THE REGULATORY REFORM ORDER (FIRE SAFETY) 2005 (RRO)
4 PEEL HOUSE,TAUNTON STREET,SHIPLEY,BN18 3NA TEL: 01274 595015 FAX: 01274 580804 EMAIL:sales@rosse-systems.co.uk A GUIDE TO THE REGULATORY REFORM ORDER (FIRE SAFETY) 2005 (RRO) THE PURPOSE OF THIS DOCUMENT
More informationUltracom hxi. Instructions for Use. High Efficiency Condensing System Boilers. To b e l e f t w i t h t h e u s e r.
Ultracom hxi High Efficiency Condensing System Boilers Instructions for Use To b e l e f t w i t h t h e u s e r WARNING GAS LEAK OR FAULT Turn off the gas emergency control valve immediately. Eliminate
More informationBANQUETLINE. Thank you for buying a Victor Unit. We hope and expect that you will get many years of satisfactory use from your purchase.
V BANQUETLINE MODEL NUMBERS BL30H1 BL50H1(S) BL70H1 BL100H1 BL100H2(M) BL100HCF2 BL200H2 BL200H4 Thank you for buying a Victor Unit. We hope and expect that you will get many years of satisfactory use
More information2245 Wyecroft Road #5, Oakville, Ontario Canada L6L 5L7 Ph: Toll Free: Fax:
2245 Wyecroft Road #5, Oakville, Ontario Canada L6L 5L7 Ph: 905-827-6157 Toll Free: 1-877-427-2266 Fax: 905-827-2027 Instructions for Installation, Operation & Service Manufacturer: Zero Appliances, Chloorkop
More informationDomestic and Residential Fire Sprinklers Design Policy
Domestic and Residential Fire Sprinklers Design Policy Content Section Page No 1. Background: Fire Sprinklers 3 2. Scope 3 3. Types of Fire Sprinklers 3 4. Types of Supply 4 5. Operating pressures and
More informationUSER GUIDE VOGUE MAX SYSTEM
USER GUIDE VOGUE MAX SYSTEM 15 18 26 32 When replacing any part on this appliance, use only spare parts that you can be assured conform to the safety and performance specification that we require. Do not
More informationFire Safety Policy and Procedures
Approved by: Date of approval: September 2015 Originator: Helen Garfield, Chief Executive Officer Policy Statement The safety of staff, patients, volunteers and visitors in the event of fire is an issue
More informationUsers Manual. Gas wall hung Boiler condensing Buderus /S. Please read thoroughly before operating the unit
7215 3200-02/2006 GB (EN) For the user Users Manual Gas wall hung Boiler condensing Buderus 500-24/S Please read thoroughly before operating the unit Introduction Please read these instructions and follow
More informationSECURITY SERVICE LEVEL AGREEMENT
Security Services Campus Services SECURITY SERVICE LEVEL AGREEMENT GENERAL STATEMENT AND GUIDANCE NOTES This document details the services provided within current resources to meet the expectations of
More informationOPERATING INSTRUCTIONS
TECHNICAL INFORMATION OPERATING INSTRUCTIONS BITZER AUSTRALIA BUFFALO TRIDENT HEAT EXCHANGE Models PS1LE016 PS2LE233 Content 1 Important Recommendations 2 Safety Recommendations 3 Application Ranges 4
More informationferroli Boiler Replacement Maxima 35C- Modena HE
ferroli Boiler Replacement Maxima 35C- Modena HE Introduction Although the new Ferroli Modena 32C HE high efficiency condensing combination boiler has an output slightly less than that of the Ferroli Maxima
More informationUnderstand and apply domestic central heating system installation and maintenance techniques
Level: 2 Credit value: 10 URN: Y/602/2888 Unit aim This combination unit provides basic learning in the installation, maintenance, decommissioning and soundness testing of a basic range of wet central
More informationLYOTRAP FREEZE-DRYING MACHINE OPERATING INSTRUCTIONS
LYOTRAP FREEZE-DRYING MACHINE OPERATING INSTRUCTIONS User Manual for Lyotrap Freeze-Drying Machine The precise answer to controlled environments 2 User Manual For Lyotrap Freeze-Drying Machine Manual Ref.
More informationSafety Information and Emergency Instructions
Safety Information and Emergency Instructions for the South Hook LNG Public Information Zone (PIZ) Our Community South Hook LNG Terminal, Herbrandston Please read this booklet and keep it in a safe place.
More informationVelocity V50 PROFESSIONAL STEAM IRON SYSTÈME DE REPASSAGE DOMESTIQUE SISTEMA DE PLANCHADO INSTRUCTION MANUAL GUIDE D UTILISATION MANUAL DE INSTRUCCIONES CONGRATULATIONS We would like to thank you and
More informationinstallation & operating instructions
installation & operating instructions RGE 410 RA/RM-2D 822706106 MO-M 0740 INTRODUCTION We are pleased that you have chosen this refrigerator and hope you will derive much satisfaction from using it.
More informationLocal Rules: Fire Safety
1. Potential Effects of a Fire A small fire is most likely to cause damage to equipment and furniture, with the potential to cause additional smoke damage to the room of origin. A large and protracted
More informationBaxi Maxflow Combi WM
Baxi Maxflow Combi WM Gas Fired Wall Mounted Combination Boiler with Unvented Hot Water Storage Please keep these instructions safe. Should you move house, please hand them over to the next occupier. User
More informationLOW TEMPERATURE FREEZER MANAGEMENT GUIDE
Introduction ULTRA-LOW LOW TEMPERATURE FREEZER MANAGEMENT GUIDE Energy-intensive laboratories present promising opportunities to help UEA reduce its carbon emissions. Ultra-low temperature freezers (ULTs)
More informationER MANUAL FOR YOUR BAUMATIC
BW18BL 50 litre thermoelectric beverage centre BW28BL 70 litre thermoelectric beverage centre ER MANUAL FOR YOUR BAUMATIC BW18BL 50 litre thermoelectric beverage centre BW28BL 70 litre thermoelectric beverage
More informationPreventing deaths and injuries in the home in Scotland
Preventing deaths and injuries in the home in Scotland www.electricalsafetyfirst.org.uk Electrical Safety First Preventing deaths and injuries in the home in Scotland 1 2 Electrical Safety First Preventing
More informationInstallation and Servicing Instructions
Installation and Servicing Instructions Zenex GasSaver GS-1 Passive Flue Gas Heat Recovery Device (PFGHRD) Recognised by the Government s Standard Assessment Procedure (SAP) IMPORTANT: Read these instructions
More informationTable of Contents. Specifications... page 2. Installation... page 3. Customizing... page 4. Reversing door swing... page 5
Introduction The Scotsman Compact Refrigerator is a unique product, capable of being built into a cabinet because of its front vented, forced-air cooling system. It s also designed to be a companion to
More informationCooled Incubators Series 1, 3 & 4. Instruction Manual
Cooled Incubators Series 1, 3 & 4 Instruction Manual Instruction Manual for LMS Cooled Incubators Series 1, 3 & 4 Series 1 Models 305-60 litre capacity 303-180 litre capacity 250-250 litre capacity Series
More informationStatesman 45/50L, 50/70L, 70/90L, 90/110L, Utility 50/70L, 70/90L, 90/110L, 110/130L, 130/150L
Users Instructions Statesman 45/50L, 50/70L, 70/90L, 90/110L, Utility 50/70L, 70/90L, 90/110L, 110/130L, 130/150L About the Boiler About Safety This is a Floor Standing, Horizontally Fired, Automatic Pressure
More informationHORNE 15 and 20 thermostatic mixing valve specification manual 2009
HORNE 15 and 20 thermostatic mixing valve specification manual 2009 contents HORNE introduction 1-4 history 1 why TMVs? 3 why HORNE? 4 HORNE HORNE 15 TMV 5 4th connection 6 HORNE HORNE 20 TMV 7 HORNE appendix
More informationBaxi Maxflow Combi FS
Baxi Maxflow Combi FS Gas Fired Floor Standing Combination Boiler with Unvented Hot Water Storage Please keep these instructions safe. Should you move house, please hand them over to the next occupier.
More informationFor Models FF2310APS, FF2310APW, FF2310APB 55cm FREESTANDING FRIDGE FREEZER. Instruction Manual
For Models FF2310APS, FF2310APW, FF2310APB 55cm FREESTANDING FRIDGE FREEZER Instruction Manual Please read these instructions carefully before use and retain for future reference. Before switching on your
More informationThese instructions are to be left with the user
ECO ADJUSTABLE HANDSET 1668.001 & 1668.002 Accessory Pack Installation & User Guide These instructions are to be left with the user 1 Introduction The Mira Eco adjustable shower handset has three different
More informationCompany Portfolio. Issue 2
Company Portfolio Issue 2 1.0 Company Introduction M&M Medical has been established since March 1991, having had in excess of 25 years experience in the complete turnkey projects of Medical & Laboratory
More informationREGULATORY REFORM (FIRE SAFETY) ORDER 2005 FIRE EMERGENCY PLAN
REGULATORY REFORM (FIRE SAFETY) ORDER 2005 FIRE EMERGENCY PLAN This Emergency Plan has been developed following fire risk assessments of the relevant building. The purpose of the Plan is to ensure the
More informationUltracom 2. sxi. Instructions for Use. 12sxi. 18sxi. 30sxi. High Efficiency Condensing System Boilers. To be left with the user. G.C. No.
Ultracom 2 sxi Instructions for Use To be left with the user 12sxi G.C. No. 41-019-12 18sxi G.C. No. 41-019-13 30sxi G.C. No. 41-019-14 High Efficiency Condensing System Boilers www.glow-worm.co.uk TABLE
More informationMitsubishi Electric Guide to Chiller Technologies. Information Guide 58
Mitsubishi Electric Guide to Chiller Technologies Information Guide 58 Information Guide 58 Mitsubishi Electric Guide to Chiller Technologies This is an independent guide produced by Mitsubishi Electric
More information2245 Wyecroft Road #5, Oakville, Ontario Canada L6L 5L7 Ph: Toll Free: Fax:
2245 Wyecroft Road #5, Oakville, Ontario Canada L6L 5L7 Ph: 905-827-6157 Toll Free: 1-877-427-2266 Fax: 905-827-2027 Instructions for Installation, Operation & Service Manufacturer: Zero Appliances, Chloorkop
More informationEnergy Performance Certificate
Energy Performance Certificate Northern Ireland 167, Killowen Grange, LISBURN, BT28 3JF Date of assessment: Date of certificate: Reference number: Type of assessment: Accreditation scheme: Assessor s name:
More informationFR 10 INSTALLATION & USER INSTRUCTIONS INTELLIGENT ROOM THERMOSTAT
INSTALLATION & USER INSTRUCTIONS INTELLIGENT ROOM THERMOSTAT FR 10 FOR USE WITH THE FOLLOWING APPLIANCES: GREENSTAR CDI COMINATION OILERS GREENSTAR CDI SYSTEM OILERS FITTED WITH OPTIONAL INTEGRAL DIVERTER
More information19 th October FAO Paul Lewis Planning Policy Branch Planning Division Welsh Assembly Government Cathays Park Cardiff CF10 3 NQ
19 th October 2009 FAO Paul Lewis Planning Policy Branch Planning Division Welsh Assembly Government Cathays Park Cardiff CF10 3 NQ Dear Sir / Madam Technical Advice Note 6 Planning for Sustainable Rural
More informationAUTOMATIC GRANULAR ICE FLAKER
AUTOMATIC GRANULAR ICE FLAKER INSTRUCTIONS AND WARNINGS 24480 rev. 01 It is strictly forbidden to reproduce this instruction manual or any part thereof. Dear Customer, Congratulations on choosing a
More informationGUARDED STATUS INFORMATION and REQUIREMENTS For OWNERS AND OPERATORS Of REFRIGERATION PLANTS
GUARDED STATUS INFORMATION and REQUIREMENTS For OWNERS AND OPERATORS Of REFRIGERATION PLANTS [Second Revised Edition] April, 2003 Issued by: Mechanical and Engineering Branch Room 500, Norquay Building
More informationPlease keep these instructions in a safe place. If you move house, please hand them over to the next occupier.
User s Operating Instructions & Important Warranty Information Baxi Combi 1 HE Plus Baxi Combi 100 HE Plus Baxi Combi 80 HE Plus Gas Fired Wall Mounted Condensing Combination Boiler Please keep these instructions
More informationFIRE SAFETY 2013 ANNUAL REPORT. Author: Senior Fire Safety Advisor Contact Details for further information: Frank Barrett
AGENDA ITEM 4.1 FIRE SAFETY 2013 ANNUAL REPORT Executive Lead: Director of Planning Author: Senior Fire Safety Advisor Contact Details for further information: Frank Barrett 02920742292 SITUATION This
More informationBaxi Bahama 100. Gas Fired Wall Mounted Combination Boiler. User s Operating Instructions
Baxi Bahama 100 Please keep these instructions safe. Should you move house, please hand them over to the next occupier. Gas Fired Wall Mounted Combination Boiler User s Operating Instructions Natural Gas
More informationOwners Manual - Incorporating Operation, Maintenance and Troubleshooting
B-Series Advanced Wastewater Treatment Units www.bluewateratu.ca Owners Manual - Incorporating Operation, Maintenance and Troubleshooting Tel: 1 877 702 4634 Certified to NSF/ANSI Standard 40, Class 1
More informationEnergy Efficiency Rating
Energy Performance Certificate Northern Ireland 37, Church Street MAGHERA BT46 5EA Date of assessment: 20 July 2011 Date of certificate: 20 July 2011 Reference number: 9639-0923-6880-4640-9926 Type of
More informationICS PACIFIC. Installation Instructions & User Manual. G135L Medical Refrigerator
ICS PACIFIC G135L Medical Refrigerator Installation Instructions & User Manual 31 Healey Road, Dandenong South, Victoria, 3175, Australia P.O. Box 4140, Dandenong South BC, Victoria, 3164, Australia E-
More informationFire risk management plan. MH/05/Revised/06/17
Fire risk management plan MH/05/Revised/06/17 1 1. Scope This policy applies to fire safety arrangements in all workplaces, common areas of residential premises and any other premises managed and/or owned
More informationPOLICY FOR THE SAFE WORKING WITH ELECTRICITY
AGENDA ITEM 2.3 7 October 2010 POLICY FOR THE SAFE WORKING WITH ELECTRICITY Report of Executive Director of Planning Paper prepared by Purpose of Paper Action/Decision required Link to Health Care Standards:
More information6100 SINGLE LOOP DIGITAL ADDRESSABLE FIRE ALARM CONTROL PANEL
6100 SINGLE LOOP DIGITAL ADDRESSABLE FIRE ALARM CONTROL PANEL USER MANUAL Protec Fire Detection plc, Protec House, Churchill Way, Nelson, Lancashire, BB9 6RT. Telephone: +44 (0) 1282 717171 Fax: +44 (0)
More informationOsprey 2 CFL
Osprey 2 CFL 125-150 - 180-220 Gas Fired Floor Standing Boiler Users Operating Instructions Please leave these instructions with the user Natural Gas Potterton Osprey 2 CFL 125 G.C.N o 41 590 54 Potterton
More informationINDOOR UNIT/OUTDOOR UNIT
SPLIT TYPE AIR CONDITIONER INDOOR UNIT/OUTDOOR UNIT ENGLISH MODEL RAS-50YHA/RAC-50YHA OUTDOOR UNIT INDOOR UNIT RAC-50YHA RAS-50YHA Instruction manual Page 1~26 To obtain the best performance and ensure
More informationvolume 11 environmental assessment section 2 environmental impact assessment Part 4 ha 204/08 scoping of environmental impact assessments
DESIGN MANUAL FOR ROADS AND BRIDGES volume 11 environmental assessment section 2 environmental impact assessment Part 4 ha 204/08 scoping of environmental impact assessments SUMMARY This Advice Note provides
More informationFACILITIES. Failure of the PSTN will not affect the internal phone system or site paging (bleeps).
.1 Telecommunications.1.1 Public Telephone Network Failure General Information The Trust uses two Telephone Service providers. Incoming traffic is carried by British Telecom (BT) and outgoing traffic is
More informationOperating and installation instructions
Operating and installation instructions Plate warming drawer EGW 3060-10 To avoid the risk of accidents or en-gb damage to the appliance it is essential to read these instructions before it is installed
More informationTechnical Bulletin 008 (Edition 2) CIP-RACL Developed by Industry
Technical Bulletin 008 (Edition 2) CIP-RACL Developed by Industry Title: Room-sealed fanned-draught chimney/flue systems concealed within voids Consumer Information Pack and Risk Assessment Checklist Date
More informationGB24 & GB30. User Manual
GB24 & GB30 User Manual BOILER OUTPUT To Domestic Hot Water:To Central Heating: GB24/30 Minimum 8.0 kw (27,296 Btu/h) GB24 Maximum 24.2 kw (82,570 Btu/h) GB30 Maximum 30.3 kw (103,384 Btu/h) GB24/30 Minimum
More informationRecommendations texts (9.82) to be used on Energy Report in Scotland 18 July 2008
s texts (9.82) to be used on Energy Report in Scotland 18 July 2008 Note. The recommendations are picked up by software using the recommendation number (see SAP 2005 version 9.82, Appendix T). Where the
More information