1 Front page OPERATING INSTRUCTIONS EuroFID Gas Analyzers for Measurement of the Total Hydrocarbon Content Installation Operation Maintenance

2 Document Information Described Product Product name: EuroFID Components: Analyzer unit type wall housing Analyzer unit type 19" slide-in enclosure Analyzer unit type inline Analyzer unit type industrial housing Operating unit type 19" slide-in enclosure Operating unit type ½-19" slide-in enclosure Terminal box Terminal module 1 Terminal module 2 Shut-off fitting Mounting flange Heat insulation unit Hood for weather protection Software version: 5.x, mode Process or Emission (Version LEL : ) Document ID Title: Operating Instructions EuroFID Order No.: Version: 2.3 Release: Manufacturer SICK AG Erwin-Sick-Str Waldkirch Germany Tel.: Fax: info.pa@sick.de Trademarks Kalrez and Viton are trademarks of DuPont Performance Elastomers. Other product names used in this document may also be trademarks and are only used for identification purposes. Glossary AC: Alternating Current ATEX (Atmosphère explosible): Abbreviation for European directives related to safety in potentially explosive atmospheres BImSchV: FICA (Regulation for implementation of the Federal Immission Control Act of the Federal Republic of Germany) CSA: Canadian Standards Association ( DC: Direct Current FID: Flame ionization detector Firmware: Internal software of the device, normally in a deletable memory chip (EEPROM) IP XY: International Protection (also: Ingress Protection); protection type of a device in accordance with IEC/DIN EN The first digit designates protection against contact and foreign bodies, the second against moisture. LED: Light emitting diode (small indicator lamp) LEL: Lower explosion limit (Minimum concentration in a combustible gas or steam above which the gas mixture can be ignited) NAMUR: Abbreviation for Normen-Arbeitsgemeinschaft für Messund Regeltechnik in der chemischen Industrie, now Interessengemeinschaft Automatisierungstechnik der Prozessindustrie ( PC: Personal computer PTFE: Polytetrafluorethylene TA Luft: TI Air - Technical instructions for the prevention of air pollution (general administrative regulation for the Federal Immission Control Act of the Federal Republic of Germany); defines threshold values and measurement and calculation methods UV: Ultraviolet light UVRAS: Ultraviolet resonance absorption spectrometry Windows: PC operating system from Microsoft Corporation Original Documents The English edition of this document is an original document of SICK AG. SICK AG assumes no liability for the correctness of an unauthorized translation. Please contact the manufacturer or your local representative in case of doubt. Legal information Subject to change without notice. SICK AG. All rights reserved. 2 EuroFID Operating Instructions V2.3 SICK AG

3 Warning Symbols Information Symbols Hazard (general) Information about the use in potentially explosive atmospheres Hazard by voltage Important technical information for this device Hazard in potentially explosive atmospheres Important information on electrical or electronic functions Hazard by explosive substances/mixtures Supplementary information Hazard by poisonous substances Link to information at another place Warning levels / Signal words Nice to know WARNUNG Risk or hazardous situation which could result in severe personal injury or death. CAUTION Hazard or unsafe practice which could result in less severe or minor injuries and/or property damage. NOTICE Indicates a hazard or unsafe practice which could result in property damage. EuroFID Operating Instructions V2.3 SICK AG 3

4 Contents Contents 1 Important Information Product identification Main hazards Main instructions for operation Intended use Purpose of the device Installation site Restrictions of use Responsibility of the user Additional documents Product Description Measuring principle Measuring method Measuring ranges Device modules Analyzer units Operating units Terminal box Additional equipment Device options Accessories Project Planning Scope of delivery Overview of installation work Required resources (overview) Mechanical installations (overview) Mains connection (overview) Supply of operational gases Screw connections of the gas connections Safety instructions for gas supply Supply of instrument air Supply of fuel gas Exhaust gas connections Exhaust gas connection Bypass gas connection Exhaust gas lines Sample gas supply Correct sample gas supply Connection of the sample gas line Span gas supply ( Calibration gas connection) EuroFID Operating Instructions V2.3 SICK AG

5 Contents 4 Installation of the Analyzer Unit Installation of the wall housing type analyzer unit Assembly/securing Gas connections Electrical connections Installation of the 19" slide-in enclosure type analyzer unit Installation/securing Gas connections Electrical connections Installation of the inline type analyzer unit Safety notes on the lifting points Assembly Installing the hood for weather protection Gas connections Electrical connections Installation of the analyzer unit industrial housing type Supplied condition Assembly Installation of the industrial housing Installing the analyzer Gas connections Electrical connections Installation of the Terminal Box Installing the terminal box Electrical connections of the terminal box Installation of the Operating Unit Installing the operating unit Mains connection of the operating unit External mains switch Mains cable connection Signal connections Overview of signal connections Contact assignment of the plug connector Analog and digital I/O Analog outputs Analog inputs Switching outputs (threshold value alarm, status) Control inputs Printer connection EuroFID Operating Instructions V2.3 SICK AG 5

6 Contents 7 Terminal Modules (Accessory) Terminal module Functions Installation Configuration options (jumper) Terminal assignment Terminal module Functions Technical design Functional descriptions Installation Function of the relay switching contacts Configuration options (jumper) Signal connections Start-up Safety information for start-up Preparing for start-up Activation procedure Operation (General) Controls and indicators/displays Display LEDs (indicators) Keypad Menu system Main menu Navigation principle Overview of the menu system Checking the operational state (visual control) Features of a safe operational state Indications of an unsafe operational state Behavior in an emergency Main Menu Functions Measurement display Selecting the measuring range Calibration (note) Flame ignition Diagnosis Purpose of the diagnosis function Activating the diagnosis messages Malfunction messages in the diagnosis function Operational displays in the diagnosis function Maintenance (note) Language (language selection) General Help EuroFID Operating Instructions V2.3 SICK AG

7 Contents 11 Maintenance Sub-menu Functions Accessing the Maintenance sub-menu Calibration gas value (note) Conversion factors Setting the conversion factor Calculation of conversion factors Standby Error archive Function of the error archive Displaying the error archive Printing the error archive Deleting the error archive Filter change (menu) System Parameters (overview) User level (activation) Range select mode (Measuring range selection) Temperature of the sensor block I/O Parameters Integrated value (averaging for the measurement display) Display rate Setup filter (averaging for the Measured value analog output) Median value time (averaging for the Reference value analog output) Display range Output range Range select mode (note) Set measuring range (note) Configuring threshold value alarms Setting the measuring ranges Purging parameters (purging time) Device status System in Service Threshold reset Time and date settings Setting the date and time Changing the time zone Summer / winter time (automatic switchover) Supplementary text for physical unit (Unit of measurement) Name of analyzer Printer functions Paper format Log settings Printing measurement archive/average value archive Print instrument log Print diagnosis Print selftest Print all parameters (device data) Restart EuroFID Operating Instructions V2.3 SICK AG 7

8 Contents 12 Calibration Introduction to calibration of the EuroFID Types of calibration Calibration gases Zero gas Span gas Settings for calibration Calibration control Setpoint value for the span gas (calibration gas value) Calibration times for automatic calibration Manual calibration Checking calibration Zero check Sensitivity test (calibration gas test) Shutdown Preparing shutdown Securing the connected systems Protection against condensation Deactivation procedure Transport Disposal Maintenance Necessary technical knowledge for maintenance work Safety instructions for maintenance work Maintenance plan Maintenance by the user Maintenance by the manufacturer's service technicians Servicing the sample gas filter Access to the sample gas filter Replacement/cleaning of the filter element (procedure) Cleaning the filter Spare parts for the sample gas filter Clearing Malfunctions If the EuroFID does not work at all If the measured values are obviously incorrect Electrical fuses Mains fuse of the operating unit Fuses in the analyzer unit type industrial housing Fuses in the terminal box Overtemperature protection on the analyzer unit Function of the overtemperature protection Resetting the temperature monitor (after tripping) Testing the overtemperature protection (note) Malfunction messages Malfunction messages on the display (general) Possible malfunction messages EuroFID Operating Instructions V2.3 SICK AG

9 Contents 16 Technical Data Dimensions Analyzer units Operating units Terminal box Housing specifications Operating units Analyzer units Terminal box Flange specification Ambient conditions Supply voltage (mains voltage) Gas connections Auxiliary gases Instrument air Fuel gas Sample gas conditions Measuring characteristics Approvals (suitability tests) Accessories Mounting connections Heat insulation flange Hood for weather protection Shut-off fittings Plug connector for the operating unit Terminal modules Cable Older Device Versions Operating unit older version Terminal box older version EuroFID Operating Instructions V2.3 SICK AG 9

10 Contents 10 EuroFID Operating Instructions V2.3 SICK AG

11 Important Information EuroFID 1 Important Information Main hazards Main instructions for operation Intended use Own responsibility EuroFID Operating Instructions V2.3 SICK AG 11

12 Important Information 1.1 Product identification Device modules A EuroFID gas analyzer contains three main device modules: a) Analyzer unit: Contains the physical measuring system and basic electronics b) Operating unit: Contains the electronic control unit, display and controls. c) Terminal box: Connects the analyzer unit and operating unit Description of the device modules P. 24, 2.4 Product versions The EuroFID gas analyzers make up a product family featuring various modules ( P. 23, Fig. 2) which are combined to device versions for various applications. The analyzer unit version is decisive for its function: The range includes analyzer units for direct measurement in the process ( inline and industrial housing types) and for extractive gas analysis ( wall housing type, 19" slide-in enclosure type). Depending on the housing version, the analyzer unit is designed either for indoor operation or outdoor installations ( P. 15, 1.4.2). The industrial housing type analyzer unit is also available as an Ex1 industrial housing version for potentially explosive atmospheres. The Ex1 industrial housing analyzer unit is not described in these Operating Instructions. If the EuroFID is equipped with the Ex1 industrial housing analyzer unit: In addition to these Operating Instructions also observe the information in the document Operating Instructions (Annex) EuroFID Ex1 industrial housing analyzer unit. Identification of the product version The version of a device module can be recognized by its exterior ( P. 23, Fig. 2). The mains voltage is stated on the type plate of the operating unit. 12 EuroFID Operating Instructions V2.3 SICK AG

13 Important Information 1.2 Main hazards Main safety notes in abbreviated form Always observe the fully safety information (see cross-references). Health risks CAUTION: Risk of hydrogen explosion The EuroFID requires hydrogen H 2 (pure hydrogen or a hydrogen-helium mixture) for operation. Always observe the notes on explosion prevention. P. 32, CAUTION: Health risk caused by exhaust gas If there is the risk of formation of hazardous substances resulting from the combustion of the sample gas: Eliminate the exhaust gases in a safe manner. P. 32, 3.4 WARNING: Danger to life/health risk due to leaks in the gas path If the device measures toxic gases: Ensure suitable safety measures. Accuracy of the measured values CAUTION: Risks resulting from incorrect or incomplete calibration If the EuroFID is not correctly calibrated, the measured values are either incorrect or inaccurate. This can have dangerous consequences if the measured values are used directly for the control of production processes or for providing warnings for dangerous gas concentrations. Calibrate the device after each start-up ( P. 151, 12.5). Check that the setpoint value is correct ( P. 149, ). Calibrate the device regularly during operation. EuroFID Operating Instructions V2.3 SICK AG 13

14 Important Information 1.3 Main instructions for operation Start-up Ensure gas tightness; check the filters, valves etc. Ensure correct supply of operation gases (specifications P. 194, 16.6). After start-up wait until the warm-up period has elapsed (approx. 2 hours). Then calibrate the device ( P. 145, 12). Operational state Observe the LED displays: Alarm (red) = measured value is/was outside the threshold values Fault (yellow) = malfunction in the measuring system; the measured values are not valid Service (yellow) = action by user required (service requirement) OK (green) = EuroFID measuring system is ready for operation; meas. value is valid Observe the messages at the bottom of the display. If Alarm is displayed Check the current measured value. Assess the situation. Take appropriate action to deal with this situation during operation. If the alarm message is self-locking: Cancel the alarm message ( Reset P. 130, 11.13) as soon as the measured value is inside the threshold value. In hazardous situations Switch off the emergency off switch/main switch of the higher-level system. The mains switch of the EuroFID is on the rear of the operating unit ( P. 59, Fig. 18). Maintenance Calibrate the device regularly ( P. 145, 12). Clean the integrated sample gas filter regularly (every 2 months P. 162, 14.4). If the EuroFID is used as a gas detector: Check the function of the alarms regularly. NOTICE: Only open the sample gas filter when the Filter change status is activated (procedure P. 163, ). Otherwise dust-laden ambient air can enter the measuring system. Soiling which occurs as a result can render the measuring system useless. Shutdown Secure/deactivate connected systems ( P. 156, ). Prevent condensation in the measuring system. To this purpose purge the sample gas path with dry gas. The simplest solution: Use the Zero point test function ( P. 153, ). Let the analyzer unit cool before starting work on the interior of the device (risk of burns). Fragile sampling probe Only valid for inline and industrial housing type analyzer units. Observe the permissible lifting points ( P. 42, 4.3.1). Do not use the sampling probe as a lifting point. Do not subject the sampling probe to mechanical loads. This could cause breakage of the glass vacuum tube. 14 EuroFID Operating Instructions V2.3 SICK AG

15 Important Information 1.4 Intended use Purpose of the device The gas analyzer EuroFID is used to measure the total carbon content of hydrocarbon compounds in air. The possible measured concentration ranges from ppm to 100 % LEL. The EuroFID can analyze corrosive and condensing gases at gas temperatures up to 350 C (660 F). The EuroFID is tested for suitability for continuous monitoring of the emissions of organically bound carbon according to 13. FICA and TI Air 17. FICA Installation site The operating unit and analyzer unit ( P. 23, Fig. 2) are generally designed for indoor operation. These devices should not be exposed to direct atmospheric influences (wind, rain, sun). Such influences can damage the devices and negatively affect the measuring accuracy. The inline type analyzer unit can also be used outdoors if it is provided with a hood for weather protection (accessory). The industrial housing type analyzer units can be used outdoors. The terminal box can be used outdoors. The EuroFID is not suitable for operation in potentially explosive atmospheres. The maximum permissible distance between the terminal box and the analyzer unit is 500 m. The standard cable length is 6 m. EuroFID Operating Instructions V2.3 SICK AG 15

16 Important Information Restrictions of use Explosion protection The EuroFID is not suitable for operation in potentially explosive atmospheres. Operation in potentially explosive atmospheres requires additional safety equipment. Properties of the sample gas The EuroFID should never be used For measuring operationally explosive gas compounds For sample gas compounds which form an explosive mixture with air. If the EuroFID is used for measuring combustible gases or for sample gases which can form a combustible gas mixture with air, there is a potential risk of explosion in the event of a defect in the internal gas paths (leakage). The gas could ignite when it comes into contact with hot surfaces or the FID flame. For such applications: Check which regulations and laws apply at the installation site in such cases. Check whether additional safety equipment has to be installed (e.g. housing enclosure and purging). The sample gas should not contain substances which could chemically corrode the components in contact with the sample gas (materials P. 195, 16.7). Such substances should also not be formed during combustion of the sample gas (in the FID flame). The sample gas should not contain any gas components which could condense in the measuring system (temperature in the measuring system P. 117, 11.10). No solids should be formed during gas combustion.solids in the gas path could obstruct the measuring system. If the inline type analyzer unit is installed: Do not exceed the following temperature limits: Sample gas temperature at the gas inlet of the sensor tube: 350 C (660 F) max. Dewpoint temperature of the sample gas: 200 C (390 F) max. Influence of oxygen (O 2 ) and sulphur dioxide (SO 2 ) in the sample gas If the O 2 concentration in the sample gas fluctuates, the measured values can be corrupted. The EuroFID is set at the works so that the measurement error is not greater than 4 % of the measuring range end value. If the following conditions apply at the same time in the sample gas: SO 2 concentration > 50 mg/m3 O 2 concentration < 15 percent by volume then the measurement error in the measuring range tested for suitability (15 mg/m3 C) is higher than 4 % if H 2 is used as fuel gas. In such cases an H 2 -He mixture must be used as a fuel gas for measurement in the validity range of the suitability test (specification P. 195, ). The EuroFID is configured and adjusted accordingly for such applications at the manufacturer's works. For all other applications, H 2 is preferable as a fuel gas. 16 EuroFID Operating Instructions V2.3 SICK AG

17 Important Information 1.5 Responsibility of the user Designated users The gas analyzer EuroFID may only be installed and operated by skilled technicians who, based on their technical training and knowledge as well as knowledge of the relevant regulations, can assess the tasks given and recognize the hazards involved. Correct use Use the device only as described in these Operating Instructions. The manufacturer bears no responsibility for any other use. Perform the prescribed maintenance work. No components may be removed from, added to or changed on the device unless this is described and specified in the official manufacturer information. Otherwise: The device could become dangerous Any warranty by the manufacturer becomes void. The approvals according to the suitability test for the device are rendered invalid. Special local conditions In addition to these Operating Instructions, follow all local laws, technical rules and company-internal operating directives applicable at the respective installation site of the equipment. Responsibility for hazardous substances WARNING: Danger to life/health risk caused by leaks in the gas path If the device measures toxic gases: A leak in the gas path can constitute a serious risk for personnel. Take suitable safety measures. Ensure that the safety measures are complied with. Examples for safety measures: Marking of the device with warning signs Marking of the installation site with warning signs Safety-related instruction of personnel who could be in the vicinity of the installation site. Safeguarding of documents Keep these Operating Instructions ready for reference at all times. They must be passed on to new owners. 1.6 Additional documents Each EuroFID is supplied with a test log containing the most important operating parameters (e.g. measuring range, fuel gas, mains voltage). For device versions with the analyzer unit type Ex1 industrial housing (not described in these Operating Instructions) the separate Additional Operating Instructions for the EuroFID analyzer unit type Ex1 industrial housing also apply. EuroFID Operating Instructions V2.3 SICK AG 17

18 Important Information 18 EuroFID Operating Instructions V2.3 SICK AG

19 Product Description EuroFID 2 Product Description Product identification Functional principle Characteristics Versions EuroFID Operating Instructions V2.3 SICK AG 19

20 Product Description 2.1 Measuring principle Fig. 1 Measuring principle Combustion air (instrument air) U= Sample gas Fuel gas The EuroFID uses a flame ionization detector (FID) for measuring the hydrocarbons. A hydrogen flame supplied by fuel gas and combustion air burns in the FID. The sample gas is routed into this flame. The hydrocarbons contained in the sample gas are split; the produced hydrocarbon fragments are ionized. A stream of ions forms in the electric field and this electrical stream is measured. The measuring signal is proportional to the number of the listed, non-oxidized hydrocarbon atoms. Hydrocarbon atoms which are already oxidized are only partially detected.co and CO 2 are ineffective. The quantitative correlations between the measurement signal and the hydrocarbon concentration in the sample gas is determined by performing reference measurements with calibration gases which do not contain any hydrocarbons (zero gas) or of which the hydrocarbon concentration is known exactly (calibration gas e.g. 80 ppm propane in air). 2.2 Measuring method Instrument air The EuroFID requires hydrocarbon-free instrument air for operation. The instrument air is used in the EuroFID for the following purposes: As induction air for the ejector pump As combustion air for the FID As control air for regulation of the sample gas flow to the FID As zero gas for calibration Heated sensor block To prevent condensation in the sample gas path, the parts in contact with the sample gas (sample gas filter, ejector, sample gas capillaries, bypass orifice plate, sensor) are accommodated in a heated, thermostatically controlled sensor block. The constant temperature is the pre-requisite for a constant sample gas flow to the FID. The temperature in the sensor block depends on the individual measurement application ( C). 20 EuroFID Operating Instructions V2.3 SICK AG

21 Product Description Ejector pump For transporting the sample gas the EuroFID uses a patented jet pump, referred to in these instructions as an ejector pump. The ejector pump does not have any mechanical, moving parts and uses a part of the instrument air as induction air. The sample gas is diluted with the instrument air in the ejector pump. The dilution ratio is variable and depends on three pressure parameters of the ejector pump. These pressure parameters are constantly measured and used for a correction calculation which compensates for the variability of the dilution ratio. The pressure parameters are also used to automatically monitor the function of the measuring system (sample gas volume flow, condition of the sample gas filter, plausibility of the measured values). The induction air is pre-heated in the sensor block. This prevents condensation of elements of the sample gas in the ejector pump. The sample gas is diluted before it is supplied to the FID as the ejector pump is installed upstream from the FID. This reduces the effect of damaging sample gas elements. Sample gas volume flow (to the detector) The internal sample gas system comprises a sample gas capillary, a bypass gas capillary and a pressure controller. The pressure upstream from the sample gas capillary is kept constant to ensure that the volume flow through the sample gas capillary is constant. To this purpose control air is added to the bypass flow downstream from the ejector pump with the aid of a pressure regulator. To this purpose the pressure upstream from the bypass capillary is kept constant. FID The permanent FID flame is supplied with combustion air and fuel gas. Part of the instrument air serves as combustion air. H 2 or an H 2 /He mixture is used as fuel gas ( P. 32, 3.3.4). The volume flow of the fuel gas is kept constant by a system comprising a pressure regulator and a flow controller. If a malfunction occurs during operation which makes operation impossible, the EuroFID switches the fuel gas off internally. The instrument air remains switched on for several minutes to flush the sample gas out of the measuring system. The internal volume flows of each analyzer unit are optimized individually at the manufacturer's works. Supply of calibration gases An internal solenoid value is provided for the supply of both zero gas and span gas during calibration. The respective calibration gas is routed into the installed sample gas filter; pressure and volume flow are dimensioned so that the calibration gas completely displaces the sample gas. The excess calibration gas is discharged through the sample gas inlet. The ejector pump draws the calibration gas out of the sample gas filter. EuroFID Operating Instructions V2.3 SICK AG 21

22 Product Description 2.3 Measuring ranges Measuring ranges The EuroFID has 6 measuring ranges. Each measuring range starts at 0. The measuring range end values vary by the factor 10. The physical unit can be selected ( P. 127, ). The measuring ranges can be modified by changing the end value of the largest measuring range ( P. 127, ). The smaller measuring ranges are automatically derived from the largest measuring range and are therefore not variable. Table 1 Alternative scaling The EuroFID is calibrated at the works with propane, i.e. the hydrocarbon measurement is modified for propane. Measuring ranges (example) Measuring range 1 Measuring range 2 Measuring range 3 Measuring range 4 Measuring range 5 Measuring range ppm 0 20 ppm ppm Propane ppm ppm ppm It is possible to reference the measuring ranges to another physical unit (e.g. mg/m3) or another substance (e.g. butane). To this purpose corresponding conversion factors can be set ( P. 104, 11.3). 22 EuroFID Operating Instructions V2.3 SICK AG

23 Product Description Fig. 2 EuroFID device modules (overview) Operating unit type 19" slide-in enclosure with integrated terminal box Operating unit type 19" slide-in enclosure Operating unit type ½-19" slide-in enclosure Terminal box Analyzer unit type industrial housing Analyzer unit type inline Analyzer unit type wall housing Analyzer unit type 19" slide-in enclosure EuroFID Operating Instructions V2.3 SICK AG 23

24 Product Description 2.4 Device modules Analyzer units Fig. 3 Types Analyzer unit type wall housing : This analyzer unit is designed for installation in an outer housing which protects it against general environmental influences (e.g. measuring cabinet). The measuring system is mounted on a base plate under a protective hood. The sample gas must be supplied through a heated sample gas line. Analyzer unit type 19" slide-in enclosure : In this version the measuring system is installed in a 19" rack. The sample gas must be routed in the same manner as for the analyzer unit type wall housing. Analyzer unit type inline : A flange, a sampling probe and the measuring system form one unit. The sample gas is taken directly from the process. The analyzer unit is mounted with the flange directly on the process. The outer housing protects the measuring system, can be quickly removed and can also be equipped with a fan (option). A hood for weather protection is available as an accessory. Analyzer unit type industrial housing : This is a special version of the inline type in which the measuring system is accommodated in an enclosed housing. It is equipped with a flange with a sampling probe in the same way as the inline type and is directly mounted on the process. The industrial housing is either equipped with housing ventilation or an air conditioner and features a frost-protection heater. Design of the analyzer Analyzer unit: Design of the analyzer A B C D E Electronic components (with voltage converters) Sensor block heated (to prevent condensation in the measuring system) Ejector pump Sample gas filter Pneumatic components (pressure regulator, flow controller, solenoid valves) The individual operating parameters are the analyzer unit are programmed in the electronic section (e.g. characteristic curve of the ejector pump, calibration of the pressure sensors, data of basic calibration). When the system is started, the operating unit takes these operating parameters from the analyzer unit. Integrated sample gas filter The integrated sample gas filter filters particles from the sample gas before the sample gas enters the measuring system. Filter soiling is automatically monitored by the EuroFID. If the filter requires cleaning, the malfunction message Service requir. is activated. 24 EuroFID Operating Instructions V2.3 SICK AG

25 Product Description Operating units Types 19" slide-in enclosure for installation in 19" frame (3 HU) ½-19" module for installation in 19" racks (4 HU) Diagrams with dimensions P. 189, Contents Microcomputer system Operating elements Display for measurement display and user interaction Keypad LEDs (signal lamps) Signal connections Digital interface Option: Electronic card for the flow barrier option Option: Fitted terminal box (only for 19" slide-in enclosure type). An operating unit can only control and evaluate one operating unit. The operation concept complies with the recommendations of the standards committee of the German chemical industry (NAMUR) Terminal box Versions Terminal box in its own housing ( P. 190, Fig. 39) Terminal box integrated into the operating unit 19" slide-in enclosure type Terminal box integrated into the analyzer unit industrial housing type Functions Cable connections for the analyzer unit Voltage converters for the analyzer unit Voltage supply for the fan of the inline outer housing Control of the solenoid valve of the flow barrier option EuroFID Operating Instructions V2.3 SICK AG 25

26 Product Description 2.5 Additional equipment Device options Option for Flow barrier 1 Analyzer unit type inline Analyzer unit type industrial housing Housing fan Analyzer unit type inline Air conditioner Analyzer unit type industrial housing 1 prevents process gas (sample gas) flowing from the opened filter when maintenance work is being performed on the sample gas filter Accessories Accessory components Information for Hood for weather protection P. 44, Analyzer unit type inline Mounting connections P. 197, Fig. 40 Analyzer unit type inline Analyzer unit type industrial housing Heat insulation flange for mounting connections P. 197, Fig. 41 Analyzer unit type inline Shut-off fitting P. 199, Fig. 43 Analyzer units Terminal modules P. 67, 7 Counterparts of the plug connectors large connector set1 Operating units P. 57, 6 Counterparts of the plug connectors small connector set2 1 All D-SUB plug connectors + Power mains connection 2 D-SUB plug connectors Analyzer assembly, Data line Power mains connection 26 EuroFID Operating Instructions V2.3 SICK AG

27 Project Planning EuroFID 3 Project Planning Scope of delivery Overview of the installation Operating materials Gas supply Exhaust gas venting EuroFID Operating Instructions V2.3 SICK AG 27

28 Project Planning 3.1 Scope of delivery Standard scope of delivery 1 analyzer unit with 3 m connection cable to the terminal box (if not integrated into the industrial housing) 1 operating unit 1 terminal box (if not integrated into the operating unit or the industrial housing) with 6 m connection cable to the operating unit 1 mains cable, 3 m long with plug connector for connection to the terminal box 4 screws for flange + 1 flange seal (only with the analyzer unit inline type) 1 individual test log 1 Operating Instructions NOTICE: for analyzer unit types inline and industrial housing The surrounding tube of the sampling probe contains a glass vacuum tube. Incorrect use can damage the sampling probe. Observe the safety note on the lifting points ( P. 42, 4.3.1). NOTICE: Soiling in the gas paths can damage the analyzer. Only remove protective caps from the gas connections when the gas lines are connected. Possible accessories (options) Mounting connection for the inline type analyzer unit Hood for weather protection for gas analyzer type inline Shut-off fitting (for gas supply) Solenoid valve (for the flow barrier option) Terminal module 28 EuroFID Operating Instructions V2.3 SICK AG

29 Project Planning 3.2 Overview of installation work Required resources (overview) Required resource Notes, requirements Note Instrument air Constant supply of the analyzer unit with P. 31, technical compressed air CH-free; purity depending on the P. 194, application and measuring range Fuel gas Continuous supply of the analyzer unit P. 32, Normally from compressed air cylinders H 2 or H 2 /He mixture, depending on the P. 195, device version Leakage monitoring For warning of explosion risk in the event of P. 32, device1 H 2 leakage Pressure reducer For instrument air P. 31, Pressure monitor 1 Shut-off valves 1 Heated sample gas For fuel gas Or other condensation-free supply of the P. 32, P. 35, line2 sample gas to the analyzer unit Span gas Only during calibration; hydrocarbon P. 147, 12.3 concentration depending on the measuring range Exhaust gas lines Exhaust P. 32, 3.4 Bypass2 Power supply Mains connection on the operating unit P. 59, 6.2 Ventilation/air conditioning If the ambient temperature exceeds 40 C during operation 3 1 Recommendation 2 Only for the analyzer unit types wall housing and 19" slide-in enclosure 3 Take the lost heat of the analyzer unit into account (heated sensor block) EuroFID Operating Instructions V2.3 SICK AG 29

30 Project Planning Mechanical installations (overview) The following system components must be installed: Component Installation site, requirements Detailed notes Supply line or Suitable pressure reducer P. 31, compressed air Gas line to the analyzer unit cylinders for instrument air Oil-free installation Compressed gas cylinder(s) for fuel gas Mains connection (overview) At a suitable location which meets the safety requirements Suitable pressure reducer Gas line to the analyzer unit Leakage monitoring device 1 P. 32, Analyzer unit Wall housing type: In an enclosed P. 38, 4.1 interior (e.g. system cabinet) 19" slide-in enclosure type: In 19" P. 40, 4.2 frame Inline type: On a mounting connection P. 42, 4.3 Industrial housing type: On a mounting P. 47, 4.4 connection Shut-off valves Near the analyzer unit 2 Refer to Analyzer Solenoid valve 3 Near the analyzer unit unit Terminal box 4 Near the analyzer unit 5 P. 53, 5 Operating unit Near the 6 external terminal box 4 or the P. 57, 6 analyzer unit Electrical Analyzer unit ( terminal box)4 P. 37, 4 connections Analyzer unit Signal connections as required P. 60, (recommendation) 2 Pre-assembled shut-off fitting available as accessory 3 Only for devices with the flow barrier option 4 Not for device versions with an integrated terminal box 5 Standard length of the connection cable: 3 m 6 Standard length of the connection cable: 6 m The mains voltage for all device modules is supplied through the operating unit ( P. 59, 6.2). An additional, separate mains switch must normally be installed ( P. 59, 6.2.1). The terminal box receives the mains voltage from the operating unit and in turn supplies the analyzer unit with operating voltage. 30 EuroFID Operating Instructions V2.3 SICK AG

HFID Heated Total Hydrocarbon Analyzer Module

HFID Heated Total Hydrocarbon Analyzer Module Product Data Sheet PDS 103-685.A01 October, 2003 NGA 2000 HFID Heated Total Hydrocarbon Analyzer Module Heated analyzer designed to operate from 200 F to 400 F Eliminates loss of hydrocarbons due to condensation

More information

testo 316-EX gas leak detector Instruction manual

testo 316-EX gas leak detector Instruction manual testo 316-EX gas leak detector Instruction manual 2 1 Safety and the environment 1 Safety and the environment 1.1. About this document Use > Please read this documentation through carefully and familiarize

More information

IntelliDoX Operator Manual

IntelliDoX Operator Manual IntelliDoX Operator Manual OPERATOR MANUAL TABLE OF CONTENTS Table of Contents Table of Contents...1 About this Publication...3 Important Safety Information: Read First...4 Getting Started...5 About the

More information

Title: Standard Operating Procedure for Measurement of Total Hydrocarbon Using a Flame Ionization Detection

Title: Standard Operating Procedure for Measurement of Total Hydrocarbon Using a Flame Ionization Detection Procedure No: SOP-020 Revision No: 1.0 (January 21, 2011) Page No.: 1 of 7 1. INTRODUCTION AND SCOPE To obtain timely data for the purpose of air quality assessment, air quality trend reporting, and to

More information

SAFETY CERTIFIED MODEL FP-700 COMBUSTIBLE GAS DETECTOR

SAFETY CERTIFIED MODEL FP-700 COMBUSTIBLE GAS DETECTOR SAFETY MANUAL SIL 2 Certified Model FP-700 Combustible Hydrocarbon Gas Sensor Version 2.0 1 SAFETY CERTIFIED MODEL FP-700 COMBUSTIBLE GAS DETECTOR This manual addresses the specific requirements and recommendations

More information

GMS800 FIDOR TOTAL HYDROCARBON ANALYZER CONTINUOUS MONITORING OF HYDROCARBON EMISSIONS

GMS800 FIDOR TOTAL HYDROCARBON ANALYZER CONTINUOUS MONITORING OF HYDROCARBON EMISSIONS Product Information GMS800 FIDOR TOTAL HYDROCARBON ANALYZER CONTINUOUS MONITORING OF HYDROCARBON EMISSIONS Gas Analyzers SOLUTION FOR CONTINUOUS HYDROCARBON MEASUREMENTS The compact GMS800 FIDOR extractive

More information

Safe Operation Manual. Honeywell IQ Force Gas Detector

Safe Operation Manual. Honeywell IQ Force Gas Detector Safe Operation Manual Honeywell IQ Force Gas Detector Honeywell Analytics 800-663-4164 403-248-9226 Fax 403-575-3708 03JUN2012 P/N 50105843-047 Version 01 http://www.honeywell.com HONEYWELL IQ Force PERSONAL

More information

Series 9000 Total Hydrocarbon Analyzer

Series 9000 Total Hydrocarbon Analyzer Series 9000 Total Hydrocarbon Analyzer Analyzer The Baseline Series 9000 is the candidate of choice whenever accurate, reliable total hydrocarbon analysis is required. The Series 9000 analyzer provides

More information

TS400. Operating Manual. Test Station for Microtector II Series (G450/G460)

TS400. Operating Manual. Test Station for Microtector II Series (G450/G460) Operating Manual TS400 Test Station for Microtector II Series (G450/G460) GfG GESELLSCHAFT FÜR GERÄTEBAU MBH KLÖNNESTRASSE 99 44143 DORTMUND, Germany TEL. +49 / (0)2 31 / 5 64 00 0 FAX +49 / (0)2 31 /

More information

Process and Air Monitoring. Online GC systems for continuous monitoring of organic compounds in air and gaseous samples

Process and Air Monitoring. Online GC systems for continuous monitoring of organic compounds in air and gaseous samples Process and Air Monitoring Online GC systems for continuous monitoring of organic compounds in air and gaseous samples For 35 years AMA Instruments has developed and manufactured analytical devices and

More information

DESIGN MANUAL MODEL GIR900 INFRARED COMBUSTIBLE GAS SENSOR

DESIGN MANUAL MODEL GIR900 INFRARED COMBUSTIBLE GAS SENSOR DESIGN MANUAL MODEL GIR900 INFRARED COMBUSTIBLE GAS SENSOR 70086 The information in this doucment is applicable to sensors built in September, 2008 or later (Firmware 3.0) The information and technical

More information

GMA 301. Operation Manual. Worldwide Supplier of Safety Solutions. Part Number

GMA 301. Operation Manual. Worldwide Supplier of Safety Solutions. Part Number Worldwide Supplier of Safety Solutions GfG Instrumentation 1194 Oak Valley Drive #20 Phone: 734-769-0573 Fax: 734-769-1888 E-Mail: info@gfg-inc.com Internet: www.gfg-inc.com GMA 301 Operation Manual Part

More information

Flopurge TS. Operation Manual

Flopurge TS. Operation Manual Flopurge TS Operation Manual Part Number 079-0204 Spectron Gas Control Systems United Kingdom Unit 4, Herald Court, University of Warwick Science Park, Coventry, CV4 7EZ +44 (0)24 7641 6234 sales@spectron-gcs.com

More information

Advance Optima Module Magnos 17

Advance Optima Module Magnos 17 Advance Optima Module Magnos 17 Service Manual 43/24-1002-0 EN Table of Contents Page Chapter 1: Description of functions 1-1 Chapter 2: Module variants and components 2-1 Chapter 3: Analyzer variants

More information

M-Sens 2. Online-Moisture Meter for Solids. Operating Instructions. SWR engineering Messtechnik GmbH

M-Sens 2. Online-Moisture Meter for Solids. Operating Instructions. SWR engineering Messtechnik GmbH EN M-Sens 2 Operating Instructions Online-Moisture Meter for Solids SWR engineering Messtechnik GmbH CONTENTS Page 1. System Overview......................................................................

More information

User Manual. GPD 3000 e. gas detector

User Manual. GPD 3000 e. gas detector User Manual GPD 3000 e gas detector Table of contents Application...3 Unit ppm, Vol.%...4 Operation elements GDP 3000 e...5 Gas sensor...6 Important Information off sensor!...6 Measurement range...7 Turning

More information

LMV52 Control. Specifications. Document No. LV February 3, Product Description. Sample Specification

LMV52 Control. Specifications. Document No. LV February 3, Product Description. Sample Specification LMV52 Control Product Description February 3, 2017 The LMV52 is a microprocessor-based burner management system with matching system components for the control and supervision of forced draft burners.

More information

SITRANS. Temperature transmitter Functional safety for SITRANS TW. Introduction. General safety instructions 2. Device-specific safety instructions

SITRANS. Temperature transmitter Functional safety for SITRANS TW. Introduction. General safety instructions 2. Device-specific safety instructions Introduction 1 General safety instructions 2 SITRANS Temperature transmitter Device-specific safety instructions 3 Appendix List of Abbreviations/Acronyms A B Product Information Supplement to Operating

More information

M-Sens 2. Online moisture measurement for solids. Operating Instructions. SWR engineering Messtechnik GmbH PART OF THE ENVEA GROUP

M-Sens 2. Online moisture measurement for solids. Operating Instructions. SWR engineering Messtechnik GmbH PART OF THE ENVEA GROUP EN M-Sens 2 Operating Instructions Online moisture measurement for solids SWR engineering Messtechnik GmbH PART OF THE ENVEA GROUP CONTENTS Page 1. System overview.............................................

More information

Flammable gases Detectors. Model DAX 03F DAX 03F- C DAX 03F- H. Note for installation, use and maintenance. Version 1.2

Flammable gases Detectors. Model DAX 03F DAX 03F- C DAX 03F- H. Note for installation, use and maintenance. Version 1.2 Flammable gases Detectors Model DAX 03F DAX 03F- C DAX 03F- H Note for installation, use and maintenance Version 1.2 Introduction This handbook must be read carefully by any person who has or will have

More information

GAS MONITOR Model Model

GAS MONITOR Model Model Sierra Monitor Corporation 1991 Tarob Court, Milpitas, CA 95035 (408) 262-6611 (800) 727-4377 (408) 262-9042 - Fax E-mail: sierra@sierramonitor.com Web Site: www.sierramonitor.com GAS MONITOR Model 2350-00

More information

Dräger X-am 5600 Personal Monitor Multi-Gas Detection Device

Dräger X-am 5600 Personal Monitor Multi-Gas Detection Device Dräger X-am 5600 Personal Monitor Multi-Gas Detection Device Featuring an ergonomic design and innovative infrared sensor technology, the Dräger X-am 5600 is the smallest gas detection instrument for the

More information

Residential Gas Condensing Boiler Greenstar ZBR16/21/28/35/42-3A... ZWB28/35/42-3A...

Residential Gas Condensing Boiler Greenstar ZBR16/21/28/35/42-3A... ZWB28/35/42-3A... 70 80 99-00-O Residential Gas Condensing Boiler ZBR//8/35/4-3A... ZWB8/35/4-3A... 70 80 993 (03/03) CA/US Operating Instructions Contents Contents Key to symbols and safety instructions............................

More information

Operation and Maintenance Manual

Operation and Maintenance Manual Operation and Maintenance Manual GDS-48 Remote Bridge Sensor for Combustible Gases AUTHORIZED DISTRIBUTOR: GasDetectorsUSA.com - Houston, Texas USA sales@gasdetectorsusa.com - 832-615-3588 CAUTION: FOR

More information

Vacuum Controller with Control Unit VC 900

Vacuum Controller with Control Unit VC 900 Operating Instructions Read and observe these operating instructions! Vacuum Controller with Control Unit VC 900 KNF Neuberger GmbH Alter Weg 3 79112 Freiburg Germany Tel. +49 (0)7664 / 5909-0 Fax +49

More information

USERS MANUAL FOR GAS BOILERS

USERS MANUAL FOR GAS BOILERS USERS MANUAL FOR GAS BOILERS PLEASE READ THE MANUAL CAREFULLY: IT CONTAINS IMPORTANT INFORMATION REGARDING SAFETY, INSTALLATION, USE AND MAINTENANCE OF THE APPLIANCE MODELS: NOVADENS 24 NOVADENS 24C NOVADENS

More information

Operations Manual TS400. Test Station for G450/G460 Gas Detector

Operations Manual TS400. Test Station for G450/G460 Gas Detector TS400 Test Station for G450/G460 Gas Detector Operations Manual 1194 Oak Valley Dr, Ste 20, Ann Arbor MI 48108 USA (800) 959-0329 (734) 769-0573 www.goodforgas.com GfG Products for Increased Safety Congratulations

More information

PROCESS ANALYSERS. SERVOFLEX MiniMP (5200) Gas Analyser. Operator Manual. Part Number: A Revision: 8 Language: UK English

PROCESS ANALYSERS. SERVOFLEX MiniMP (5200) Gas Analyser. Operator Manual. Part Number: A Revision: 8 Language: UK English PROCESS ANALYSERS SERVOFLEX MiniMP (5200) Gas Analyser Operator Manual Part Number: 05230001A Revision: 8 Language: UK English This page intentionally blank. LIST OF CONTENTS Section Page 1 DESCRIPTION

More information

ADOS Biogas 905. Multi-channel Gas Analyser. A D O S GmbH Instrumentation and Control Trierer Str FRG

ADOS Biogas 905. Multi-channel Gas Analyser. A D O S GmbH Instrumentation and Control Trierer Str FRG ADOS Biogas 905 Multi-channel Gas Analyser Issue: 05/2006 A D O S GmbH Instrumentation and Control Trierer Str. 23-25 @ 52078 Aachen @ FRG TEL. : (0241) 9769-0 FAX : (0241) 9769-16 E-Mail : info@ados.de

More information

ALTAIR 5X Multigas Detector Product Specification

ALTAIR 5X Multigas Detector Product Specification ALTAIR 5X Multigas Detector Product Specification PHYSICAL CHARACTERISTICS Gas delivery Size, pumped unit without IR Size, pumped unit with IR Weight Handling Case material Environmental protection Display

More information

Warranty Registration

Warranty Registration WARRANT Y AND LIMITS OF LIABILIT Y Vulcain Inc. warrants to the original purchaser that its product, and the component parts thereof, will be free from defects in workmanship and materials for a period

More information

Operating instructions

Operating instructions Operating instructions Gas condensing boiler WARNING: If the information in this manual is not followed exactly, a fire or explosion may result causing property damage, personal injury or loss of life.

More information

Process Control PIP PCEA001 Fixed Gas Detection Guidelines

Process Control PIP PCEA001 Fixed Gas Detection Guidelines November 2015 Process Control PIP PCEA001 PURPOSE AND USE OF PROCESS INDUSTRY PRACTICES In an effort to minimize the cost of process industry facilities, this Practice has been prepared from the technical

More information

HIGH EFFICIENCY FIRETUBE CONDENSING GAS BOILER

HIGH EFFICIENCY FIRETUBE CONDENSING GAS BOILER This manual must be left with owner and should be hung on or adjacent to the boiler for reference. US HIGH EFFICIENCY FIRETUBE CONDENSING GAS BOILER MODELS CHS-85 through CHS-399 APPENDIX A CONTROLLER

More information

Meridian wiredhart. HART Field Device Specification Goldmine Rd. Monroe, NC USA

Meridian wiredhart. HART Field Device Specification Goldmine Rd. Monroe, NC USA HART Field Device Specification Meridian wiredhart 4320 Goldmine Rd. Monroe, NC 28110 USA HART is a registered trademark of the HART Communication Foundation TABLE OF CONTENTS 1. Introduction...5 1.1 Scope...5

More information

Operating instructions Page 14. Refrigerator Read the operating instructions before switching on for the first time

Operating instructions Page 14. Refrigerator Read the operating instructions before switching on for the first time Operating instructions Page 14 Refrigerator Read the operating instructions before switching on for the first time 7085 039-00 LKv 5710 Content Disposal notes... 14 Description of the appliance... 14 Safety

More information

Mass Sensors, Inc. Profile. Applications. Technology Baur Blvd, St. Louis, MO (314)

Mass Sensors, Inc. Profile. Applications. Technology Baur Blvd, St. Louis, MO (314) Mass Sensors, Inc. 1350 Baur Blvd, St. Louis, MO 63132 (314) 997-5779 www.mass-sensors.com Innovations in chemical microsensors Profile Mass Sensors is a development stage high-tech company operating in

More information

SERVICE AND INSTALLATION MANUAL MODELS HDO(H) OIL FOR YOUR SAFETY

SERVICE AND INSTALLATION MANUAL MODELS HDO(H) OIL FOR YOUR SAFETY Bousquet Technologies Inc. 2121, Nobel, Ste Julie, Quebec, Canada, J3E1Z9 SERVICE AND INSTALLATION MANUAL MODELS HDO(H) OIL Oil-Fired air heater for industrial and commercial use. FOR YOUR SAFETY Do not

More information

General safety precautions English

General safety precautions English English 1 1 1.1 About the documentation The original documentation is written in English. All other languages are translations. The precautions described in this document cover very important topics, follow

More information

GL-CO-RFG Channel Gas Leak Alarm System

GL-CO-RFG Channel Gas Leak Alarm System Four Elms Road Edenbridge Kent TN8 6AB UK Features & Benefits Remote sensors for natural gas, LPG and CO 1 x SPST relay outputs DIN-rail as standard, panel mounting kit available Adjustable alarm thresholds

More information

GasAlertMicroClip X3 Specifications

GasAlertMicroClip X3 Specifications GasAlertMicroClip X3 Specifications The instrument must satisfy the following: Physical Specifications Size (h x w x d) Physical size shall be no larger than 4.4 x 2.4 x 1.2 in. / 11.3 x 6.0 x 3.2 cm Weight

More information

IMR IX176 Portable Gas Detector User Manual

IMR IX176 Portable Gas Detector User Manual IMR Portable Gas Detector User Manual Read this manual carefully before using this device. (727) 328-2818 / (800) RING-IMR Fax: (727) 328-2826 www.imrusa.com Ver. 1.0A4 CONTENTS SERVICE GUIDELINES... 3

More information

SAMPLE SPECIFICATIONS CONTINUOUS MONITORING SYSTEM (Insitu)

SAMPLE SPECIFICATIONS CONTINUOUS MONITORING SYSTEM (Insitu) PART 1 - GENERAL A. RELATED DOCUMENTS SAMPLE SPECIFICATIONS CONTINUOUS MONITORING SYSTEM (Insitu) Federal, State, and local requirements as applicable. Attached Permit Attached drawings B. SUMMARY 1. Provide

More information

QAF64.2-J QAF64.6-J. Frost detector. for use on the air side

QAF64.2-J QAF64.6-J. Frost detector. for use on the air side s Frost detector for use on the air side QAF64.2-J QAF64.6-J With active capillary tube sensing element for acquiring the lowest temperature within a range of 0 15 C With startup function Operating voltage

More information

Dräger X-am 7000 Multi-Gas Detection

Dräger X-am 7000 Multi-Gas Detection Dräger X-am 7000 Multi-Gas Detection Dräger X-am 7000 is the solution for the simultaneous and continuous measurement of up to five gases. It is the ideal companion in a variety of applications where the

More information

Conversion Instructions Logano G234X. Gas boiler. Please read carefully before installing and servicing. Gas boiler

Conversion Instructions Logano G234X. Gas boiler. Please read carefully before installing and servicing. Gas boiler Gas boiler UPON COMPLETION OF THE INSTALLATION THE INSTALLER MUST INSTRUCT THE OWNER AND OPERATOR ON THE FUNCTIONALITY AND THE PROPER OPERATION OF THE BOILER AND THE HEATING SYSTEM. INSTALLER MUST REVIEW

More information

Operating Instructions

Operating Instructions Operating Instructions Low Emissions and High Efficiency Condensing Oil Boiler DANGER! If these instructions are not followed exactly, a fire or explosion may be caused with serious property damage or

More information

DESIGN MANUAL. The information in this doucment is applicable to sensors built in September, 2009 or later (Firmware 3.0)

DESIGN MANUAL. The information in this doucment is applicable to sensors built in September, 2009 or later (Firmware 3.0) DESIGN MANUAL MODEL GIR910 INFRARED CARBON DIOXIDE (CO 2 ) GAS SENSOR 70091 The information in this doucment is applicable to sensors built in September, 2009 or later (Firmware 3.0) The information and

More information

Power Flame Incorporated

Power Flame Incorporated Power Flame Incorporated SUGGESTED SPECIFICATION FOR MODEL NVC ULTRA LOW NOx GAS BURNERS SUB 9 PPM NOx THE POWER TO MANAGE ENERGY 2001 South 21st Street, Parsons, Kansas 67357 Telephone: 620-421-0480,

More information

HC Refrigerants treatment Systems

HC Refrigerants treatment Systems HC Refrigerants treatment Systems About Hydrocarbons (HC) The interest and the application for Hydrocarbons (HC) refrigerants is growing more and more, especially now that the impact of global warming

More information

Pioneer-R16 Gas Monitor Operator s Manual

Pioneer-R16 Gas Monitor Operator s Manual Pioneer-R16 Gas Monitor Operator s Manual Edition 7/2/97 RKI INSTRUMENTS, INC RKI Instruments, Inc. 33248 Central Ave, Union City, CA 94587 (510) 441-5656 Chapter 1: Description About the Pioneer-R16 Gas

More information

Monitoring Flammable Vapors and Gases in Industrial Processes

Monitoring Flammable Vapors and Gases in Industrial Processes Monitoring Flammable Vapors and Gases in Industrial Processes Flammability Hazards Industrial fires and explosions happen more frequently than most people think. They cause downtime, property damage, injury

More information

CRU-S Series Stainless Steel Condensate Recovery Unit

CRU-S Series Stainless Steel Condensate Recovery Unit IM-UK-CRU-S UK Issue 1 CRU-S Series Stainless Steel Condensate Recovery Unit 1. Safety information 2. General product information 3. Installation 4. Commissioning 5. Storage, shutdown and equipment protection

More information

MICROpeL 75 Combustible Gas Sensor Part Number: PM

MICROpeL 75 Combustible Gas Sensor Part Number: PM MICROpeL 75 Combustible Gas Sensor Part Number: PM759-000 Key Features & Benefits: ATEX, UL and CSA Approvals Withstands EN/IEC 60079-0 impact test Enhanced H 2 S and silicone poison resistance Reduced

More information

Stand Alone DYNAGARD SP Operation Manual

Stand Alone DYNAGARD SP Operation Manual Stand Alone DYNAGARD SP Operation Manual Contents Page General Description 1 Detection Principle 1 For Your Safety 1 Design 2 Mounting 2 Mounting Position of DYNAGARD SP 2 Installation of Electrical Connections

More information

Dräger X-am approved as type MQG 0010 Multi-Gas Monitor Technical Manual _01.eps

Dräger X-am approved as type MQG 0010 Multi-Gas Monitor Technical Manual _01.eps Dräger X-am 5000 approved as type MQG 0010 Multi-Gas Monitor Technical Manual 04723999_01.eps Content Content For Your Safety..................................................... 4 Safety symbols used

More information

Operations Manual TS400. Test Station for G450/G460 Gas Detector

Operations Manual TS400. Test Station for G450/G460 Gas Detector TS400 Test Station for G450/G460 Gas Detector Operations Manual 1194 Oak Valley Dr, Ste 20, Ann Arbor MI 48108 USA (800) 959-0329 (734) 769-0573 www.gfg-inc.com GfG Products for Increased Safety Congratulations

More information

Mounting and Operating Instructions EB 5610 EN. TROVIS 5600 Automation System TROVIS 5610 Heating and District Heating Controller

Mounting and Operating Instructions EB 5610 EN. TROVIS 5600 Automation System TROVIS 5610 Heating and District Heating Controller TROVIS 5600 Automation System TROVIS 5610 Heating and District Heating Controller Mounting and Operating Instructions Electronics from SAMSON EB 5610 EN Firmware version 1.40 Edition December 2014 Controller

More information

Emission Monitoring System

Emission Monitoring System Advance Cemas-FTIR Multi-component Emission Monitoring System System Description 30/23-200-1 EN System Overview Measurement principle The Advance Cemas FTIR is a multi-component emission monitoring system

More information

Operating instructions

Operating instructions EN Operating instructions SmartControl MultiControl (incl. Timer-Function) SmartControl MultiControl Table of Contents 1 Introduction...5 1.1 Purpose of the document...5 1.2 Using this document...5 1.3

More information

IMR EX610. Portable Gas Detector. Operation Manual

IMR EX610. Portable Gas Detector. Operation Manual IMR EX610 Portable Gas Detector Operation Manual IMR Environmental Equipment, Inc. 3634 Central Ave. St. Petersburg, FL 33711 Phone: 727-328-2818 email: info@imrusa.com Read this manual carefully before

More information

SEC Signature Process Gas Analyzer

SEC Signature Process Gas Analyzer SEC Signature Process Gas Analyzer Instruction and Operation Manual Sensor Electronics Corporation 5500 Lincoln Drive Minneapolis, Minnesota 55436 USA (952) 938-9486 Fax (952) 938-9617 email sensor@minn.net

More information

Products and performance

Products and performance Products and performance February 26, 2016 GfG Instrumentation overview of fixed system products Slide 1 GfG Instrumentation, Inc. 1194 Oak Valley Drive, Suite 20, Ann Arbor, Michigan 48108 Toll free (USA

More information

E1000 Portable gas detector

E1000 Portable gas detector E1000 Portable gas detector Operation manual Ver:HWWM161229CG Read this manual carefully before using the device Table of Contents SAFETY INFORMATION... - 1-1. BRIEF INTRODUCTION... - 3-2. MAIN FUNCTION

More information

Operate the Mutli-Gas Monitor PID

Operate the Mutli-Gas Monitor PID TASK: Operate the Multi-Gas Monitor Learning Objective CONDITION: In a classroom environment, given a fully charged Multi-gas Monitor with the manufacturer operator s manual, all monitor components/accessories,

More information

Residential Gas Condensing Boiler Greenstar combi 100 p / 151 p ZWB28-3A... ZWB42-3A...

Residential Gas Condensing Boiler Greenstar combi 100 p / 151 p ZWB28-3A... ZWB42-3A... Residential Gas Condensing Boiler Greenstar combi 00 p / 5 p ZWB8-3A... ZWB4-3A... Operating Instructions 708733 (07/04) US/CA Contents Contents Explanation of symbols and safety instructions.......................

More information

Model D12-IR. Digital Gas Transmitter...

Model D12-IR. Digital Gas Transmitter... Model D12-IR Digital Gas Transmitter... ATI s Series D12 gas transmitter line now includes a versatile Infrared system that can be configured for LEL or select toxic gases. The D12-IR utilizes a compact

More information

USER'S MANUAL PU SENS 01 (A11) PU SENS 01 (A19) Sensor Control Panel

USER'S MANUAL PU SENS 01 (A11) PU SENS 01 (A19) Sensor Control Panel USER'S MANUAL PU SENS 01 (A11) PU SENS 01 (A19) Sensor Control Panel PU SENS 01 CONTENTS Safety requirements... 3 Purpose... 4 Technical data... 4 Overall dimensions [mm]... 4 Mounting and set-up... 5

More information

General safety precautions English

General safety precautions English English 1 1 1.1 About the documentation The original documentation is written in English. All other languages are translations. The precautions described in this document cover very important topics, follow

More information

Operating Instructions and Maintenance Manual

Operating Instructions and Maintenance Manual Impact Series Operating Instructions and Maintenance Manual 4.9.4 Enforcer Calibration This is designed for oxygen, flammable (catalytic or IR LEL sensor), carbon monoxide and hydrogen sulfide, using a

More information

Energy management function at adsorption dryer PDAD

Energy management function at adsorption dryer PDAD Application note Energy management function at adsorption dryer PDAD Application note for use of the energy management function at adsorption dryer PDAD. Explanation, use and layout of purge air regulation

More information

ExDetector HC-100. Data Sheet. Gas Measuring and Alarm Systems

ExDetector HC-100. Data Sheet. Gas Measuring and Alarm Systems HC-100 Gas Measuring and Alarm Systems HC-100 Application / Construction The HC100 series of measuring probes, used in conjunction with evaluation systems, have the following functions: Measurement and

More information

DRYTEK 1 - User Manual ELECTRICAL PANEL FOR 1 MOTOR WITH POWER FACTOR CONTROL

DRYTEK 1 - User Manual ELECTRICAL PANEL FOR 1 MOTOR WITH POWER FACTOR CONTROL DRYTEK 1 - User Manual ELECTRICAL PANEL FOR 1 MOTOR WITH POWER FACTOR CONTROL CONTENTS 1. INTRODUCTION... 5 2. WARNINGS... 6 3. GENERAL DESCRIPTION... 7 4. INSTALLATION... 8 5. CONTROL PANEL... 9 5.1

More information

ANNEX AMENDMENTS TO THE INTERNATIONAL CODE FOR FIRE SAFETY SYSTEMS (FSS CODE) CHAPTER 1 GENERAL

ANNEX AMENDMENTS TO THE INTERNATIONAL CODE FOR FIRE SAFETY SYSTEMS (FSS CODE) CHAPTER 1 GENERAL Annex 2, page 2 ANNEX AMENDMENTS TO THE INTERNATIONAL CODE FOR FIRE SAFETY SYSTEMS (FSS CODE) CHAPTER 1 GENERAL Section 1 Application 1 The following new sentence is added to the end of paragraph 1.2:

More information

Thermo 230 / 231 / 300 / 301 / Troubleshooting

Thermo 230 / 231 / 300 / 301 / Troubleshooting Thermo 230 / 231 / 300 / 301 / 350 5 Troubleshooting 5 Troubleshooting 5.1 General This section describes troubleshooting procedures for the heaters Thermo 230, 231, 300, 301 and 350. CAUTION Troubleshooting

More information

PROCESS ANALYSERS. SERVOFLEX Micro i.s. (5100is) Gas Analyser. Operator Manual. Revision:

PROCESS ANALYSERS. SERVOFLEX Micro i.s. (5100is) Gas Analyser. Operator Manual. Revision: PROCESS ANALYSERS SERVOFLEX Micro i.s. (5100is) Gas Analyser Operator Manual Reference: Revision: Language: 05110001A 9 English 01234516ÿ8ÿ09100ÿÿ0ÿ00 ÿÿÿ $ +!ÿ ÿ"$ÿ $+(ÿÿ!ÿ+!ÿ" ÿ#$%&$'ÿ(")ÿ#$%&$'ÿÿ*$$

More information

Entwicklung + Fertigung + Service: Heerweg 15 D, Denkendorf Tel.: / Fax.: /

Entwicklung + Fertigung + Service: Heerweg 15 D, Denkendorf Tel.: / Fax.: / Planungsbüro + Service: Esteraustr. 10, 56379 Holzappel Tel.: 0 64 39 / 90 19 90 Fax.: 0 64 39 / 90 19 91 E-Mail: u.ramakers@umsitec.de Entwicklung + Fertigung + Service: Heerweg 15 D, 73770 Denkendorf

More information

GAS COOKTOP INSTALLATION INSTRUCTIONS BEFORE YOU BEGIN. IMPORTANT Save these instructions for local electrical inspector s use.

GAS COOKTOP INSTALLATION INSTRUCTIONS BEFORE YOU BEGIN. IMPORTANT Save these instructions for local electrical inspector s use. GAS COOKTOP INSTALLATION INSTRUCTIONS Please read this guide thoroughly before installation. To contact LG Electronics, 24 hours a day, 7 days a week: 1-800-243-0000 (U.S.A.) 1-888-542-2623 (Canada) Or

More information

T UNI 7000 F. Operating instructions For the user (2006/05) AU/GB

T UNI 7000 F. Operating instructions For the user (2006/05) AU/GB 6 720 648 662-00.1T UNI 7000 F Operating instructions For the user AU/G 2 Contents Contents Contents 2 1 Safety information and explanation of symbols 3 1.1 For your safety 3 1.2 Explanation of symbols

More information

Overview of Gas Range Types

Overview of Gas Range Types Residential Gas Ranges Overview of Types How Carbon Monoxide is Emitted How They Operate Overview of Gas Range Types Rick Karg R.J. Karg Associates www.karg.com November 2001 2001 R.J. Karg Associates

More information

150 & 200 RT MG-SERIES GAS FIRED DOUBLE-EFFECT EFFECT ABSORPTION CHILLER-HEATER

150 & 200 RT MG-SERIES GAS FIRED DOUBLE-EFFECT EFFECT ABSORPTION CHILLER-HEATER GUIDE SPECIFICATIONS 150 & 200 RT MG-SERIES GAS FIRED DOUBLE-EFFECT EFFECT ABSORPTION CHILLER-HEATER HEATER 1. General Description The Contractor shall furnish and install (Qty) Yazaki model CH- Gas Fired,

More information

User Manual FLOWMAX-90. for model. Condensing water heater 85,000 BTU. Installation, operating, commissioning and maintenance instructions.

User Manual FLOWMAX-90. for model. Condensing water heater 85,000 BTU. Installation, operating, commissioning and maintenance instructions. User Manual for model FLOWMAX-90 Condensing water heater 85,000 BTU WARNING If the information in these instructions is not followed exactly, a fire or explosion may result, causing property damage, personal

More information

SEC 2000 Millenium Infrared Gas Detector

SEC 2000 Millenium Infrared Gas Detector SEC 2000 Millenium Infrared Gas Detector Instruction and Operation Manual Sensor Electronics Corporation 5500 Lincoln Drive Minneapolis, Minnesota 55436 USA (952) 938-9486 Fax (952) 938-9617 Email: sales@sensorelectronic.com

More information

WMHP Series R410a Heat Pump INSTALLATION INSTRUCTIONS

WMHP Series R410a Heat Pump INSTALLATION INSTRUCTIONS WMHP Series R410a Heat Pump INSTALLATION INSTRUCTIONS **WARNING TO INSTALLER, SERVICE PERSONNEL AND OWNER** Altering the product or replacing parts with non authorized factory parts voids all warranty

More information

INSTRUCTION MANUAL. Software Revision Card 3.2 DOCUMENT SES-KS E REV. 3

INSTRUCTION MANUAL. Software Revision Card 3.2 DOCUMENT SES-KS E REV. 3 2402 COMBUSTIBLE GAS CARD 2 CHANNELS 2404 TOXIC GAS CARD 2 CHANNELS 2407 OXYGEN GAS CARD 2 CHANNELS INSTRUCTION MANUAL Software Revision Card 3.2 DOCUMENT SES-KS-2402-04-07-E REV. 3 2402 2404 2407-2 CHANNELS

More information

FP120-Z1 Power supply kit A (70 W) Technical Manual

FP120-Z1 Power supply kit A (70 W) Technical Manual FP120-Z1 Power supply kit A (70 W) Technical Manual 2015-07-08 Control Products and Systems Legal notice Legal notice Technical specifications and availability subject to change without notice. Transmittal,

More information

ISO INTERNATIONAL STANDARD. Hydrogen generators using fuel processing technologies Part 1: Safety

ISO INTERNATIONAL STANDARD. Hydrogen generators using fuel processing technologies Part 1: Safety INTERNATIONAL STANDARD ISO 16110-1 First edition 2007-03-15 Hydrogen generators using fuel processing technologies Part 1: Safety Générateurs d'hydrogène utilisant les technologies de traitement du carburant

More information

Generator Gas Analyzer (GGA)

Generator Gas Analyzer (GGA) UTILITY SYSTEMS Generator Gas Analyzer (GGA) Instruction Manual HA0112P01 Rev. B Contents Specifications... 4 Application... 5 System Description... 6 System Electronics... 6 Display Panel... 6 Sensor

More information

CERTIFICATE of Product Conformity (QAL1 )

CERTIFICATE of Product Conformity (QAL1 ) ATÜVF}heinland Precisely Right. CERTIFICATE of Product Conformity (QAL1 ) Certificate No.: 0000059868 AMs designation: Manufacturer: Test Laboratory: FIDAMAT 6 MEASURING SYSTEM fortoc SIEMENS AG Östliche

More information

BS-316. Gas detection control panel up to 16 inputs. Installation operation manual

BS-316. Gas detection control panel up to 16 inputs. Installation operation manual BS-316 Gas detection control panel up to 16 inputs Installation operation manual Page 1 from 21 Contents 1 Operation instructions--------------------------------------------------------------------------------------------------------------------3

More information

SIMPLICITY CO CARBON MONOXIDE DETECTION & VENTILATION PANEL

SIMPLICITY CO CARBON MONOXIDE DETECTION & VENTILATION PANEL SIMPLICITY CO CARBON MONOXIDE DETECTION & VENTILATION PANEL USER MANUAL 1 Table of Contents 1 SAFETY INFORMATION...3 1.1 SAFETY PRECAUTIONS DURING NORMAL OPERATION OF PANEL...3 1.3 BATTERY INFORMATION...3

More information

Dräger X-am approved as type MQG 01** Multi-Gas Monitor. Technical Handbook. CO2 Vol% CH4 %UEG. CO ppm. ppm NH _01.

Dräger X-am approved as type MQG 01** Multi-Gas Monitor. Technical Handbook. CO2 Vol% CH4 %UEG. CO ppm. ppm NH _01. Dräger X-am 5600 approved as type MQG 01** Multi-Gas Monitor Technical Handbook CO2 CH4 %UEG O2 CO H2S NH3 04733072_01.eps Content Content For Your Safety......................................................4

More information

Two-Channel Gas Controller

Two-Channel Gas Controller Two-Channel Gas Controller Specifications subject to change without notice. USA 09 Page of DESCRIPTION Highly configurable, UL 0 performance-tested and -certified, and wall-mounted gas monitor; continuously

More information

RES-210 (-211, -220, -221) / RES-420

RES-210 (-211, -220, -221) / RES-420 RESISTR GB Replacing RES-2xx/ Replacement Instructions The RESISTR temperature controller can be used as an alternative to the following controllers of the series 200 : RES-210: Analog temperature display,

More information

ULTIMA X5000. MSAX5000.com

ULTIMA X5000. MSAX5000.com ULTIMA X5000 MSAX5000.com WE KNOW YOU RE TIRED OF... NEEDING TO DISCONNECT POWER BEFORE CHANGING A SENSOR REMEMBERING HOW TO CALIBRATE THIS THING HAVING TO PULL SO MUCH WIRE AT EVERY GAS DETECTOR INSTALLATION...

More information

USER'S INFORMATION MANUAL

USER'S INFORMATION MANUAL USER'S INFORMATION MANUAL : If the information in this manual is not followed exactly, a fire or explosion may result causing property damage, personal injury or loss of life. Do not store or use gasoline

More information

PowerRouter application guideline

PowerRouter application guideline PowerRouter application guideline Software installation tool - version 3.4 Before operating the PowerRouter, you may initialize the PowerRouter by using the PowerRouter software installation tool. The

More information

INTERNATIONAL STANDARD

INTERNATIONAL STANDARD INTERNATIONAL STANDARD IEC 60079-29-2 Edition 2.0 2015-03 Explosive atmospheres Part 29-2: Gas detectors Selection, installation, use and maintenance of detectors for flammable gases and oxygen INTERNATIONAL

More information

General safety precautions English

General safety precautions English English A min (m 2 ) 550 530 540 510 520 490 500 470 480 450 460 430 440 410 420 390 400 370 380 350 360 330 340 310 320 290 300 270 280 250 260 230 240 210 220 190 200 170 180 150 160 130 140 110 120

More information

General safety precautions English

General safety precautions English English A min (m 2 ) 550 530 540 510 520 490 500 470 480 450 460 430 440 410 420 390 400 370 380 350 360 330 340 310 320 290 300 270 280 250 260 230 240 210 220 190 200 170 180 150 160 130 140 110 120

More information