1 Principles and Techniques of Instrument Processing and Sterilization Chapter 21 Copyright 2018, Elsevier Inc. All Rights Reserved. 1

2 Learning Objectives Lesson 21.1: CDC Guidelines 1. Pronounce, define, and spell the key terms. 2. Discuss the classification of patient care items, including the following: List the seven steps involved in processing dental instruments. Describe the Centers for Disease Control and Prevention (CDC) guidelines for sterilization and disinfection of patient care items. Describe the classification of instruments used to determine the type of processing. 3. Discuss how a dental assistant can be exposed to microorganisms through dental instruments or other patient care items and which CDC guidelines are in place to limit that exposure. 4. Describe the ideal instrument processing area. Copyright 2018, Elsevier Inc. All Rights Reserved. 2

3 Introduction One of the most important responsibilities of the dental assistant is to process contaminated instruments for reuse Instrument processing involves much more than sterilization Sterilization is a process intended to kill all microorganisms and is the highest level of microbial destruction Copyright 2018, Elsevier Inc. All Rights Reserved. 3

4 Rankings of Evidence Copyright 2018, Elsevier Inc. All Rights Reserved. 4

5 Classification of Patient Care Items These categories help determine which sterilization methods best ensure the safety of dental care workers and patients Categories are based on the risk of infection associated with their intended use Classifications are used to determine the minimal type of posttreatment processing Copyright 2018, Elsevier Inc. All Rights Reserved. 5

6 Classification of Instruments Critical instruments Items used to penetrate soft tissue or bone Greatest risk of transmitting infection and must be sterilized by heat Semicritical instruments Touch mucous membranes or nonintact skin Lower transmission risk Sterilized by heat or receive minimum high-level disinfection if not heat tolerant Copyright 2018, Elsevier Inc. All Rights Reserved. 6

7 Classification of Instruments (Cont.) Noncritical instruments Contact only intact skin Low risk of infection transmission Cleaned and processed with EPA-registered intermediate- or low-level disinfectant Copyright 2018, Elsevier Inc. All Rights Reserved. 7

8 Seven Steps for Instrument Processing Copyright 2018, Elsevier Inc. All Rights Reserved. 8

9 Personal Protective Equipment You must always use personal protective equipment (PPE) when processing instruments This includes utility gloves, mask, eyewear, and protective clothing Copyright 2018, Elsevier Inc. All Rights Reserved. 9

10 Transporting and Processing Contaminated Patient Care Items The dental assistant may be exposed to microorganisms through contact with contaminated instruments or other patient care items Exposure can occur through percutaneous injury (e.g., needle sticks, cuts) or contact with the mucous membranes of the eyes, nose, or mouth Copyright 2018, Elsevier Inc. All Rights Reserved. 10

11 Instrument-Processing Area Should be centrally located in the office to allow easy access from all patient care areas Dedicated only to instrument processing Physically separated from operatories and laboratory Not be part of a common walkway Copyright 2018, Elsevier Inc. All Rights Reserved. 11

12 Instrument-Processing Area (Cont.) Needs good air circulation Large enough to accommodate all equipment and supplies The area should not have a door or windows that open to the outside because dust may enter the area Multiple outlets and proper lighting, water, and an air line and a vacuum line for flushing high-speed handpieces Copyright 2018, Elsevier Inc. All Rights Reserved. 12

13 Instrument-Processing Area (Cont.) A deep sink should have hands-free control for instrument rinsing and (if space permits) a foot-operated or other hands-free trash receptacle The flooring should be an uncarpeted, seamless, hard surface The size, shape, and accessories of the instrument-processing area vary among dental offices Copyright 2018, Elsevier Inc. All Rights Reserved. 13

14 Instrument Processing Terms Copyright 2018, Elsevier Inc. All Rights Reserved. 14

15 Workflow Pattern Regardless of the size or shape of the instrument-processing area, four basic areas govern the pattern of workflow Processing of instruments should proceed in a single loop, from dirty to clean to sterile to storage, without ever doubling back If the instrument-processing area is small, you can use signs that read: Contaminated items only Precleaning area Cleaned items only Sterile items only Copyright 2018, Elsevier Inc. All Rights Reserved. 15

16 Instrument-Processing Areas Copyright 2018, Elsevier Inc. All Rights Reserved. 16

17 Contaminated Area All soiled instruments are brought into the contaminated area, the initial receiving area, where they are held for processing Any disposable items that have not already been discarded in the treatment room are removed from the instrument tray and disposed of as contaminated waste Thorough cleaning should be done before all disinfection and sterilization processes Removal of all debris and organic materials (e.g., blood and saliva) The contaminated area contains clean protective eyewear and utility gloves, counter space, a sink, a waste disposal container, holding solution, an ultrasonic cleaner, an eyewash station, and supplies for wrapping of instruments before sterilization Copyright 2018, Elsevier Inc. All Rights Reserved. 17

18 Waste Items Copyright 2018, Elsevier Inc. All Rights Reserved. 18

19 Preparation and Packaging Area In this area, cleaned instruments and other dental supplies should be inspected, assembled into sets or trays, and wrapped or placed in packages for sterilization The preparation and packaging area should consist of counter space and storage space for sterilized instruments, fresh disposable supplies, and prepared trays or instrument cassettes Clean instruments are not sterile and could harbor pathogens Instruments must be packaged and sterilized before they are used on a patient Copyright 2018, Elsevier Inc. All Rights Reserved. 19

20 Sterilization Center Courtesy Alfa Medical, H Copyright luckyraccoon/istock/getty Images.al, Hempstead, NY. Copyright 2018, Elsevier Inc. All Rights Reserved. 20

21 Learning Objectives Lesson 21.2: Precleaning and Packaging Instruments 5. Discuss precleaning and packaging of instruments, including the following: Explain the purpose of a holding solution. Describe the safety precautions that are necessary when one is operating an ultrasonic cleaner. Describe the precautions that are necessary when one is packaging materials for sterilization. Describe the CDC guidelines for preparing and packaging instruments for sterilization. Describe the CDC guidelines for sterilization of unwrapped instruments. Copyright 2018, Elsevier Inc. All Rights Reserved. 21

22 Precleaning and Packaging Instruments Instruments may be precleaned in one of three ways: Hand scrubbing Ultrasonic cleaning Instrument-washing machine Copyright 2018, Elsevier Inc. All Rights Reserved. 22

23 Holding Solution If instruments cannot be cleaned immediately after a procedure, they should be placed in a holding solution to prevent the drying of blood and debris on the instruments The holding solution may be any noncorrosive liquid A commercial enzymatic solution that partially dissolves organic debris may be used Dishwasher detergent also makes a good holding solution because it is low-cost, low-foaming, and readily available It is neither cost-effective nor desirable to use a disinfectant alone as a holding solution Copyright 2018, Elsevier Inc. All Rights Reserved. 23

24 Holding Solution (Cont.) The container must have a lid and must be labeled with: A biohazard label (because of the contaminated instruments) A chemical label (because of the cleaner/detergent) The holding solution should be changed at least twice daily, and even more frequently if it becomes clouded Remember, a holding solution is necessary only when contaminated instruments cannot be processed immediately Copyright 2018, Elsevier Inc. All Rights Reserved. 24

25 Commercial Holding Solutions Courtesy Certol, Commerce City, CO. Copyright 2018, Elsevier Inc. All Rights Reserved. 25

26 Hand Scrubbing Hand scrubbing is the least desirable method of cleaning instruments because it requires direct hand contact with the contaminated instrument Copyright 2018, Elsevier Inc. All Rights Reserved. 26

27 Hand Scrubbing Precautions Wear goggle-type eyewear and puncture-resistant gloves, as well as your protective clothing Clean only one or two instruments at a time Use only a long-handled brush, preferably one with a hand guard or wide surface Keep items above the waterline; fully immersing them in a basin of soapy water interferes with one s ability to see the sharp ends Allow instruments to air-dry or carefully pat them with thick toweling Never rub or roll instruments while they are in a towel because of the risk of accidental injury Copyright 2018, Elsevier Inc. All Rights Reserved. 27

28 Ultrasonic Cleaning Used to loosen and remove debris from instruments Also reduces the risk of cuts and punctures to the hands during the cleaning process Puncture-resistant utility gloves, a mask, protective eyewear, and a protective gown should always be worn when the ultrasonic cleaner is being used Keep a set of tongs near the ultrasonic unit; these can be used to remove instruments after the cleaning cycle has been completed Copyright 2018, Elsevier Inc. All Rights Reserved. 28

29 Ultrasonic Cleaning (Cont.) From Miller CH, Palenik CJ: Infection control and management of hazardous material, ed 5, St Louis, 2014, Mosby. Copyright 2018, Elsevier Inc. All Rights Reserved. 29

30 Ultrasonic Cleaner Copyright 2018, Elsevier Inc. All Rights Reserved. 30

31 Ultrasonic Cleaner (Cont.) Works by producing sound waves beyond the range of human hearing Sound waves, which can travel through metal and glass containers, cause cavitation (formation of bubbles in liquid) Bubbles burst by implosion Instruments should be processed in the ultrasonic cleaner until they are visibly clean Time varies from 5 to 15 minutes, depending on amount and type of material on the instruments and the efficiency of the ultrasonic unit Copyright 2018, Elsevier Inc. All Rights Reserved. 31

32 Ultrasonic Cleaning Solutions Only use ultrasonic solutions that are specially formulated for use in the ultrasonic cleaner Some ultrasonic cleaning products have enzyme activity Other ultrasonic cleaning products have antimicrobial activity, which reduces the buildup of microbes in the solutions with repeated use Antimicrobial activity does not disinfect the instruments; it merely prevents the microorganisms from multiplying Copyright 2018, Elsevier Inc. All Rights Reserved. 32

33 Commercial Cleaners Ultrasonic cleaning solutions come in a variety of sizes and types: Packets (1 oz) Tablets Concentrate solutions Ready-mixed gallon containers Courtesy Certol, Commerce City, CO Crosstex International, Inc. Copyright 2018, Elsevier Inc. All Rights Reserved. 33

34 More About Ultrasonic Cleaning Solutions Do not use other chemicals such as plain disinfectants in the ultrasonic cleaner Some disinfectants can fix the blood and debris on the instruments, making subsequent cleaning more difficult Specific ultrasonic solutions are available for the removal of difficult materials such as cement, tartar, stains, plaster, and alginate Refer to the instructions of the ultrasonic unit s manufacturer regarding the specific solution to be used The ultrasonic cleaning unit should be labeled with both a chemical label and a biohazard label because it contains a chemical and contaminated instruments Copyright 2018, Elsevier Inc. All Rights Reserved. 34

35 Special Tartar and Stain Remover Ultrasonic Solution 2017 Crosstex International, Inc. Copyright 2018, Elsevier Inc. All Rights Reserved. 35

36 Care of the Ultrasonic Cleaner The ultrasonic cleaning solution is highly contaminated and must be discarded at least once a day or sooner if it becomes visibly cloudy When the solution is being changed, the inside of the pan and lid should be rinsed with water, disinfected, rinsed again, and dried All PPE should be worn while solutions are being changed in the ultrasonic cleaner Copyright 2018, Elsevier Inc. All Rights Reserved. 36

37 Testing the Ultrasonic Cleaner If you notice that the instruments are not being cleaned completely with processing in the ultrasonic cleaner, the unit may not be functioning properly To determine whether the ultrasonic cleaner is working properly, hold a 5 5-inch sheet of lightweight aluminum foil vertically (like a curtain) half-submerged in fresh, unused solution Run the unit for 20 seconds, then hold foil up to the light Surfaces that were submerged should be evenly marked with a tiny pebbling effect over the entire surface An area without pebbling of more than ½ inch indicates a problem with the unit, and it needs to be serviced by the manufacturer Copyright 2018, Elsevier Inc. All Rights Reserved. 37

38 Automated Washers/Disinfectors Look and work similar to a household dishwasher Must be approved by the U.S. Food and Drug Administration Use a combination of very hot recirculating water and detergents to remove organic material After washing, the instruments are automatically dried These units are classified as thermal disinfectors because they have a disinfecting cycle that subjects the instruments to a level of heat that kills most vegetative microorganisms Instruments processed in an automatic washer/disinfector must be wrapped and sterilized before use on a patient Copyright 2018, Elsevier Inc. All Rights Reserved. 38

39 Drying, Lubrication, and Corrosion Control Instruments and burs made of carbon steel will rust during steam sterilization Rust inhibitors such as sodium nitrate and commercial products can be used to help reduce rust and corrosion An alternative to a rust inhibitor is to dry the instrument thoroughly with the use of dry heat or unsaturated chemical vapor sterilization (discussed later), which does not cause rusting Copyright 2018, Elsevier Inc. All Rights Reserved. 39

40 Packaging Instruments Before sterilization, the instruments should be wrapped or packaged to protect them from becoming contaminated after sterilization When instruments are sterilized without being packaged, they are exposed to the environment as soon as the sterilizer door is opened They can be contaminated by aerosols in the air, dust, improper handling, or contact with nonsterile surfaces Additional advantage to packaging instruments is they can be grouped into special setups, such as crown/bridge, amalgam, prophylactic, or composite Copyright 2018, Elsevier Inc. All Rights Reserved. 40

41 Packaging Materials Sterilization packaging materials and cassettes are medical devices and therefore must be FDA-approved It is of critical importance to use only products and materials that are labeled as sterilization packaging Never substitute products such as plastic wraps, paper, or zipper-lock freezer bags that are not registered for this purpose Specific types of packaging material are available for each method of sterilization Copyright 2018, Elsevier Inc. All Rights Reserved. 41

42 Packaging Materials and Types of Sterilization Copyright 2018, Elsevier Inc. All Rights Reserved. 42

43 Self-Seal Packages From Miller CH, Palenik CJ: Infection control and management of hazardous material, ed 6, St Louis, 2018, Elsevier. Copyright 2018, Elsevier Inc. All Rights Reserved. 43

44 Sterilization of Unwrapped Instruments An unwrapped cycle (sometimes called flash sterilization) is a method for sterilizing unwrapped patient care items for immediate use The time for unwrapped sterilization cycles depends on the type of sterilizer and the type of item (i.e., porous or nonporous) to be sterilized Unwrapped sterilization should be used only under certain conditions Copyright 2018, Elsevier Inc. All Rights Reserved. 44

45 Learning Objectives Lesson 21.3: Methods of Sterilization and Monitoring Sterilization 6. Discuss methods of sterilization, including the following: Describe the three most common methods of heat sterilization and the advantages and disadvantages of each. Explain the primary disadvantage of flash sterilization. Explain the limitations of liquid chemical sterilants. Explain how sterilization failures can occur. Copyright 2018, Elsevier Inc. All Rights Reserved. 45

46 Learning Objectives Lesson 21.3: Methods of Sterilization and Monitoring Sterilization (Cont.) 7. Discuss sterilization monitoring, including the following: Describe the three forms of sterilization monitoring. Explain the differences between process indicators and process integrators. Describe when and how biologic monitoring is done. 8. Describe the steps involved in cleaning and sterilizing the high-speed dental handpiece. Copyright 2018, Elsevier Inc. All Rights Reserved. 46

47 Methods of Sterilization Sterilization destroys all microbial forms, including bacterial spores Sterile is an absolute term; there is no partially sterile or almost sterile All reusable items (critical and semicritical instruments) that come into contact with the patient s blood, saliva, or mucous membranes must be heat sterilized The three most common forms of heat sterilization in the dental office are: Steam Chemical vapor Dry heat Copyright 2018, Elsevier Inc. All Rights Reserved. 47

48 Steam Autoclave Sterilization An autoclave is used to sterilize dental instruments and other items by means of steam under pressure Steam sterilization involves heating water to generate steam, producing a moist heat that rapidly kills microorganisms As steam fills the sterilizing chamber, the cooler air is pushed from an escape valve, which then closes and allows the pressure to increase It is actually the heat, not the pressure, that kills the microorganisms Copyright 2018, Elsevier Inc. All Rights Reserved. 48

49 Steam Autoclave Operation Cycles Dental office steam sterilizers usually operate in four cycles: Heat-up cycle Sterilizing cycle Depressurization cycle Drying cycle Different manufacturers provide different features Some have added a pre-sterilization vacuum cycle to their units to remove any air pockets from the chamber before steam enters the chamber Copyright 2018, Elsevier Inc. All Rights Reserved. 49

50 Steam Autoclave From Miller CH, Palenik CJ: Infection control and management 014, Mosby, Mosby of hazardous material, ed 6, St Louis, 2018, Elsevier. Copyright 2018, Elsevier Inc. All Rights Reserved. 50

51 Packaging Instruments for Steam Autoclave Sterilization Packaging material must be porous enough to permit steam to penetrate to the instruments inside The packaging material may be fabric but most often is sealed film or paper pouches, nylon tubing, sterilizing wrap, or paper-wrapped cassettes One disadvantage of steam sterilization is that the moisture may cause corrosion on some highcarbon steel instruments Distilled water should be used in autoclaves instead of tap water, which often contains minerals and impurities Copyright 2018, Elsevier Inc. All Rights Reserved. 51

52 Flash Sterilization Rapid, or flash, sterilization of dental instruments is accomplished by means of rapid heat transfer, steam, and unsaturated chemical vapor Flash sterilization may be used only on instruments that are placed in the chamber unwrapped Flash sterilization should also be used only for instruments that are to be promptly used on removal from the sterilizer It is always the best policy to use a method of sterilization in which the instruments can be packaged before use and remain packaged until the time of use Copyright 2018, Elsevier Inc. All Rights Reserved. 52

53 Unsaturated Chemical Vapor Sterilization Chemical vapor sterilization is similar to autoclaving, except that a combination of chemicals (alcohol, formaldehyde, ketone, acetone, and water) is used instead of water to create a vapor for sterilization OSHA requires a Safety Data Sheet, or SDS, on the chemical vapor solution because of the chemicals toxicity Copyright 2018, Elsevier Inc. All Rights Reserved. 53

54 Advantages of Unsaturated Chemical Vapor Sterilization The major advantage of the chemical vapor sterilizer is that it does not rust, dull, or corrode dry metal instruments The low water content of the vapor prevents destruction of items such as endodontic files, orthodontic pliers, wires, bands, and burs A wide range of items can be sterilized routinely without damage Other advantages include the short cycle time and the availability of a dry instrument after the cycle Copyright 2018, Elsevier Inc. All Rights Reserved. 54

55 Disadvantages of Chemical Vapor Sterilization The primary disadvantage is that adequate ventilation is essential because residual chemical vapors containing formaldehyde and methyl alcohol may be released when the chamber door is opened at the end of the cycle These vapors can temporarily leave an unpleasant odor in the area and may be irritating to the eyes Copyright 2018, Elsevier Inc. All Rights Reserved. 55

56 Filtration and Monitoring of Chemical Vapors Newer sterilizers are equipped with a special filtration device that further reduces the amount of chemical vapor remaining in the chamber at the end of the cycle Older models can usually be retrofitted Formaldehyde monitoring badges, similar to radiation monitoring devices, are available for employees Copyright 2018, Elsevier Inc. All Rights Reserved. 56

57 Packaging for Unsaturated Chemical Vapor Sterilization Standard packaging for chemical vapor sterilization includes: Film pouches or paper bags Nylon see-through tubing Sterilization wrap Wrapped cassettes Thick or tightly wrapped items require longer exposure because of the inability of the unsaturated chemical vapors to penetrate as well as saturated chemical vapors do under pressure As with autoclaving, closed containers (e.g., solid-metal trays, capped glass vials) and aluminum foil cannot be used in a chemical vapor sterilizer because they prevent sterilizing agent from reaching instruments inside Copyright 2018, Elsevier Inc. All Rights Reserved. 57

58 Pressure, Temperature, and Time The three major factors in chemical vapor sterilization are: Pressure, which should measure 20 psi Temperature, which should measure 131º C (270º F) Time, which should measure 20 to 40 minutes Copyright 2018, Elsevier Inc. All Rights Reserved. 58

59 Dry Heat Sterilization Operates by heating air and transferring that heat from the air to the instruments This form of sterilization requires higher temperatures than does steam or chemical vapor sterilization Dry heat sterilizers operate at approximately 160º C to 190º C (320º F to 375º F), depending on the type of sterilizer The advantage of dry heat is that the instruments will not rust if they are thoroughly dry before being placed in the sterilizer There are two types of dry heat sterilizers: Static air and forced air Copyright 2018, Elsevier Inc. All Rights Reserved. 59

60 Static Air Sterilizers Similar to an oven Heating coils are on the bottom of the chamber, and the hot air rises inside by way of natural convection Heat is transferred from the static (nonmoving) air to the instruments in 1 to 2 hours Disadvantages include amount of time it takes and errors due to incorrect processing time The wrapping material must be heat resistant Copyright 2018, Elsevier Inc. All Rights Reserved. 60

61 Forced Air Sterilizers Also called the rapid heat transfer sterilizers Circulate the hot air throughout the chamber at a high velocity This action permits rapid transfer of heat energy from the air to the instruments, reducing the time needed for sterilization Exposure time in a forced air sterilizer, after the sterilizing temperature has been reached, ranges from 6 minutes for unpackaged items to 12 minutes for packaged items Copyright 2018, Elsevier Inc. All Rights Reserved. 61

62 Ethylene Oxide Sterilization The use of ethylene oxide gas is a recognized method of sterilization Carried out at low temperatures, which is an advantage for plastic and rubber items that would melt in heat sterilizers Requires 4 to 12 hours, depending on the sterilizer model, and at least 16 hours of poststerilization aeration is required to remove the gas molecules bound to plastic and rubber surfaces Ineffective on wet items Toxicity is possible if the gas is not handled properly Copyright 2018, Elsevier Inc. All Rights Reserved. 62

63 Liquid Chemical Sterilants Some types of plastics, such as some rubber dam frames, shade guides, and x-ray film holding devices, are damaged by heat sterilization A liquid sterilant such as 2.0% to 3.4% glutaraldehyde must be used for sterilization of these items Glutaraldehyde requires 10 hours of contact time; anything less than 10 hours is disinfection, not sterilization Be sure you have an MSDS for these products Copyright 2018, Elsevier Inc. All Rights Reserved. 63

64 Disinfectant/Sterilizer From Bonewit-West K, Hunt SA, Applegate E: Today s medical assistant: clinical & administrative procedures, ed 3, St Louis, 2016, Elsevier. Copyright 2018, Elsevier Inc. All Rights Reserved. 64

65 Sterilization Failures Sterilization may fail when direct contact for the correct time between the sterilizing agent (chemical or steam) and all surfaces of the items being processed is insufficient Several factors can cause the sterilization process to fail, including improper instrument cleaning or packaging and sterilizer malfunction Copyright 2018, Elsevier Inc. All Rights Reserved. 65

66 Results of Sterilization Errors Copyright 2018, Elsevier Inc. All Rights Reserved. 66

67 Sterilization Monitoring It is critical that dental instruments be properly sterilized Because microorganisms cannot be seen with the naked eye, the major difficulty in sterilization is determining when an item is sterile Currently, three forms of sterilization monitoring are used: Physical, chemical, and biologic Copyright 2018, Elsevier Inc. All Rights Reserved. 67

68 Physical Monitoring Involves looking at the gauges and readings on the sterilizer and recording temperatures, pressure, and exposure time Although correct readings do not guarantee sterilization, an incorrect reading is the first signal of a problem Remember that the reading reflects the temperature in the chamber, not inside the pack Copyright 2018, Elsevier Inc. All Rights Reserved. 68

69 Chemical Monitoring Involves the use of a heat-sensitive chemical that changes color when exposed to certain conditions There are two types of chemical indicators: Process indicators Process integrators Copyright 2018, Elsevier Inc. All Rights Reserved. 69

70 Process Indicators Process indicators (external) are placed outside the instrument packages before sterilization Examples: Autoclave tape and color-change markings on packages or bags Process indicators simply identify instrument packs that have been exposed to a certain temperature; they do not indicate duration or pressure Process indicators are useful in distinguishing between packages that have been processed and those that have not Copyright 2018, Elsevier Inc. All Rights Reserved. 70

71 Chemical Indicators From Miller CH, Palenik CJ: Infection control and management of hazardous material, ed 6, St Louis, 2018, Mosby. Copyright 2018, Elsevier Inc. All Rights Reserved. 71

72 Process Integrators Placed inside instrument packages They respond to a combination of pressure, temperature, and time Process integrators are also known as multiparameter indicators All sterilization factors are integrated Examples: Strips, tabs, or tubes of colored liquid The advantage of placing integrators inside each package is that penetration of the packaging by the sterilizing agent is ensured Copyright 2018, Elsevier Inc. All Rights Reserved. 72

73 Limitations of Process Integrators Process indicators and integrators provide immediate visual control of sterilizing conditions They do not indicate sterility and are not a replacement for biologic monitoring Copyright 2018, Elsevier Inc. All Rights Reserved. 73

74 Biologic Monitoring Biologic monitoring (spore testing), is the only way to determine whether sterilization has occurred and all bacteria and endospores have been killed The CDC, American Dental Association, and Office of Safety and Asepsis Procedures Research Foundation recommend at least weekly biologic testing of sterilization equipment Several states also require routine biologic checks at weekly, monthly, or cycle-specific intervals, such as spore testing every 40 hours of use or every 30 days Copyright 2018, Elsevier Inc. All Rights Reserved. 74

75 Biologic Indicators Also known as spore tests, biologic indicators (BIs) are vials or strips of paper that contain harmless bacterial spores (which are highly resistant to heat) Three BIs are used in testing: Two BIs are placed inside instrument packs, and the sterilizer is operated under normal conditions The third strip is set aside as a control After the load has been sterilized, all BIs are cultured If the spores are killed (a negative culture), the sterilization cycle was successful The culturing of the spore test is usually handled with the use of a mail-in monitoring service Copyright 2018, Elsevier Inc. All Rights Reserved. 75

76 Mail-In Service vs. In-Office Monitoring Courtesy SPS medical Supply Corp., Rush, NY. Copyright 2018, Elsevier Inc. All Rights Reserved. 76

77 Handpiece Sterilization High-speed dental handpieces rotate at speeds up to 400,000 revolutions per minute (rpm) Blood, saliva, and tooth fragments, as well as restorative materials, may lodge in the head of the handpiece, where they may be retained and transferred to another patient Dental handpieces must be properly cleaned and heat-sterilized Copyright 2018, Elsevier Inc. All Rights Reserved. 77

78 Handpiece Flushing Techniques Flushing the handpiece is the best way to remove debris from the head To flush a dental handpiece: Attach a pressurized handpiece cleaner to the intake tube of the handpiece (where the air passes through) Flush the head of the handpiece to remove debris Blow out the handpiece using compressed air to remove debris before sterilization Running coolant water from the dental unit through the handpiece at chairside is insufficient Copyright 2018, Elsevier Inc. All Rights Reserved. 78

79 Handpiece Sterilizing Techniques Only steam sterilization and chemical vapor sterilizers are recommended because sterilization temperatures should not exceed 275º F (135º C) Handpieces should be packaged in bags, wraps, or packs to protect them from contamination before use Never run a handpiece hot out of the sterilizer, and avoid rapid cool-downs, such as running the handpiece under cold water If handpieces need to be cooled quickly after sterilization, use an air fan to blow roomtemperature air over them Copyright 2018, Elsevier Inc. All Rights Reserved. 79

80 Questions? Copyright 2018, Elsevier Inc. All Rights Reserved. 80

المملكة العربية السعودية وزارة الصحة اإلدارة العامة لمكافحة عدوى المنشئات الصحية

المملكة العربية السعودية وزارة الصحة اإلدارة العامة لمكافحة عدوى المنشئات الصحية 1. POLICY STATEMENT: Cleaning, decontamination and sterilization method should be done in CSSD service facility not in the end user department side. 2. PURPOSE: To ensure the proper way of collections,

More information

2015 Myco Industries, Inc Fax:

2015 Myco Industries, Inc Fax: Endoscope Cleaning and Handling It is highly important that you manually clean the endoscope immediately after it is removed from the patient, BEFORE it is automated or manually disinfected. Cleaning your

More information

Fiona Collins Liliana Donatelli Andrew Smith

Fiona Collins Liliana Donatelli Andrew Smith Fiona Collins Liliana Donatelli Andrew Smith DISCLOSURE STATEMENT The presenters have no conflict of interest regarding the content of this presentation Develop a database for various countries globally

More information

PERFORM Operating Document

PERFORM Operating Document PERFORM Operating Document Use and Maintenance of Tuttnauer Table-Top Autoclave PC-POD-CA-006-v03 Revision History Version Reason for Revision Date 01 New POD 13-Aug-13 02 POD Section 3.1.3 and 3.2 revised

More information

NOTE: If low flow check does not show high enough vacuum flow, repeat Steps 1-4, and adjust vacuum control knob to a higher vacuum flow setting. Effective scavenging is achieved if vacuum flow can be verified

More information

STERILIZATION PROTOCOLS

STERILIZATION PROTOCOLS Department Lead Brandy Avril, RDH Anna Allen, RDH STERILIZATION PROTOCOLS The Sterilization department is the life-blood of the AKMOM project. The goal is to maximize access to care for patients by allowing

More information

STERILIZATION PROTOCOLS

STERILIZATION PROTOCOLS Department Lead Mary Cerney, RDH Keitha Martellaro STERILIZATION PROTOCOLS The Sterilization department is the life-blood of the AKMOM project. The goal is to maximize access to care for patients by allowing

More information

Jennings Street School

Jennings Street School Infection Prevention and Control and Protocols Rationale Following standard infection control precautions can minimise the risk of norovirus outbreaks caused by person-to-person transmission in any institution

More information

Harvey Chemiclave 5000 Owner And Operator Guide

Harvey Chemiclave 5000 Owner And Operator Guide Is Happy To Present You With This Free Copy Of The Harvey Chemiclave 5000 Owner And Operator Guide Be sure to look to us when it is time to order parts and supplies for your Chemiclave Please Note: Even

More information

Cleaning and Sterilization for Instruments & Posts

Cleaning and Sterilization for Instruments & Posts Cleaning and Sterilization for Instruments & Posts EN FOR DENTAL USE ONLY CLEANING AND STERILIZATION PROCEDURE FOR ENDODONTIC FILES, HAND INSTRUMENTS, PINS AND POSTS, STAINLESS STEEL DRILLS, STAINLESS

More information

CLEANING INSTRUCTIONS

CLEANING INSTRUCTIONS Reusable Instrument Care Manual Pagoda Pedicle Screw System Manufacturer Ortho Development Corporation 12187 South Business Park Drive Draper, UT 84020 Phone 801-553-9991 Fax 801-553-9993 www.orthodevelopment.com

More information

THE BASICS OF DECONTAMINATION

THE BASICS OF DECONTAMINATION THE BASICS OF DECONTAMINATION Objectives Review the basic factors that impact cleaning and decontamination of surgical devices Present the fundamentals of manual and automatic cleaning Review various cleaning

More information

Manual Pre-disinfection/Cleaning and Sterilization Instructions for SATELEC Tips and Files

Manual Pre-disinfection/Cleaning and Sterilization Instructions for SATELEC Tips and Files Manual Pre-disinfection/Cleaning and Sterilization Instructions for SATELEC Tips and Files Cautions Do not use steel wool or abrasive cleaners. Avoid solutions containing iodine or with a high chlorine

More information

When you sterilize an instrument set, you know what the standards are.

When you sterilize an instrument set, you know what the standards are. Sterilization & Infection Control The importance of cleaning in earnest A regular column on sterilization and infection control issues. When you sterilize an instrument set, you know what the standards

More information

DRAFT. Module 4. Decontamination & Environmental Cleaning

DRAFT. Module 4. Decontamination & Environmental Cleaning Module 4. Decontamination & Environmental Cleaning Learning Objectives 1. Define the 3 steps of decontamination 2. List and describe the process for cleaning and disinfecting 3. Discuss safe management

More information

HOMES AND SENIORS SERVICES. APPROVAL DATE: April 2004 REVISION DATE: March 2016 REVISION DATE: March 2017 REVIEW DATE: November 2018

HOMES AND SENIORS SERVICES. APPROVAL DATE: April 2004 REVISION DATE: March 2016 REVISION DATE: March 2017 REVIEW DATE: November 2018 Page 1 of 6 PURPOSE: All equipment used for resident care will be appropriately cleaned, disinfected and/or sterilized to minimize the risk of transmission of infection by reducing micro-organism contamination

More information

Operating / User Manual

Operating / User Manual UniARM Surgical Support System Operating / User Manual Version 1.1 MITAKA KOHKI Co., Ltd. 1. Introduction Introduction Thank you for purchasing the UniARM for use in surgical procedures. To use the device

More information

4/8/2014 GOAL PATIENT EQUIPMENT

4/8/2014 GOAL PATIENT EQUIPMENT Module G APPLICATION OF CLEANING, DISINFECTION AND STERILIZATION PRINCIPLES TO PATIENT CARE EQUIPMENT IN OUTPATIENT HEALTHCARE SETTINGS GOAL Understand how to apply the principles of disinfection and sterilization

More information

Cleaning & Sterilization

Cleaning & Sterilization Handpieces & Attachments Instruments Cleaning & Sterilization System 7 System 6 System 5 CD4 CD3 SABO 2 SABO RemB CORE TPS Processing Instructions User/Patient Safety all contribute to the efficacy of

More information

Manual Pre-disinfection/Cleaning and Sterilization Instructions for SATELEC Tips and Files

Manual Pre-disinfection/Cleaning and Sterilization Instructions for SATELEC Tips and Files Manual Pre-disinfection/Cleaning and Sterilization Instructions for SATELEC Tips and Files Cautions Do not use steel wool or abrasive cleaners. Avoid solutions containing iodine or with a high chlorine

More information

Instructions for Use Spinal Punches IFU

Instructions for Use Spinal Punches IFU Indications for use Spinal punches are manually operated instruments indicated for cutting or biting bone during surgery involving the skull or spinal column. Contraindiciations Instruments should not

More information

Cleaning and Sterilisation for Burs, accessories and filling material

Cleaning and Sterilisation for Burs, accessories and filling material Cleaning and Sterilisation for Burs, accessories and filling material EN FOR DENTAL USE ONLY CLEANING AND STERILISATION PROCEDURE FOR BURS, ACCESSORIES AND FILLING MATERIAL 1) FOREWORD Devices that are

More information

R 7420 HANDLING AND DISPOSAL OF BODY WASTES AND FLUIDS

R 7420 HANDLING AND DISPOSAL OF BODY WASTES AND FLUIDS R 7420/page 1 of 5 M A. Definitions R 7420 HANDLING AND DISPOSAL OF BODY WASTES AND FLUIDS 1. Body wastes and fluids includes a person s blood, semen, drainage from scrapes and cuts, feces, urine, vomitus,

More information

SKLAR COATED SURGICAL INSTRUMENTS RECOMMENDED CARE & CLEANING INSTRUCTIONS

SKLAR COATED SURGICAL INSTRUMENTS RECOMMENDED CARE & CLEANING INSTRUCTIONS SKLAR COATED SURGICAL INSTRUMENTS RECOMMENDED CARE & CLEANING INSTRUCTIONS CAUTION: The following instructions are for all Sklar Coated stainless steel instruments. Read instructions prior to use. Improper

More information

LAB SAFETY FOR CUSTODIANS

LAB SAFETY FOR CUSTODIANS LAB SAFETY FOR CUSTODIANS Lab Waste Biological Waste Chemical Waste Regulated Medical Waste Non- Regulated Medical Waste Radioactive Waste Biohazardous Waste (Regulated Medical Waste) Wastes that are exposed

More information

BioSonic Ultrasonic Cleaning Systems. biosonic.coltene.com

BioSonic Ultrasonic Cleaning Systems. biosonic.coltene.com BioSonic Ultrasonic Cleaning Systems biosonic.coltene.com 2018 BioSonic An essential part of your Infection Control System Three Decades of Trusted Technology BioSonic UC150 BioSonic UC125 BioSonic UC50DB

More information

Best Practice Guidelines For Cleaning, Disinfection and Sterilization of Critical and Semi-critical Medical Devices

Best Practice Guidelines For Cleaning, Disinfection and Sterilization of Critical and Semi-critical Medical Devices Best Practice Guidelines For Cleaning, Disinfection and Sterilization of Critical and Semi-critical Medical Devices In BC Health Authorities THIS DOCUMENT IS INTENDED TO DESCRIBE BEST PRACTICES HEALTH

More information

CONVENTUS CAGE TM - DR Procedure Instruments Instrument Tray. Equipment Cleaning and Sterilization

CONVENTUS CAGE TM - DR Procedure Instruments Instrument Tray. Equipment Cleaning and Sterilization CONVENTUS CAGE TM - DR Procedure Instruments Instrument Tray Equipment Cleaning and Sterilization Table of Contents 1. WARNINGS AND PRECAUTIONS... 3 2. INSTRUMENT PREPARATION... 3 3. CAVITY PREPARATION

More information

Accessories. Maybachstraße. Symmetry Fax: Revised 10/15

Accessories. Maybachstraße. Symmetry Fax: Revised 10/15 Recommended Care, Cleaning and Sterilization Instructions for Reusable Instruments & Accessories Symmetry Surgical Inc. 3034 Owen Drive, Antioch, TN 37013 USA 1-800-251-3000 Fax: 1-615-964-5566 www.symmetrysurgical.com

More information

HYDRIM. Powerful cleaning the first step to effective sterilization. Instrument Washers A HIGHER STANDARD

HYDRIM. Powerful cleaning the first step to effective sterilization. Instrument Washers A HIGHER STANDARD HYDRIM Instrument Washers Powerful cleaning the first step to effective sterilization A HIGHER STANDARD Why Use An Automated Washer? The advantages of automated washing systems over traditional manual

More information

Use and Care Guide. Over the Patient Units and Swing Mount Units

Use and Care Guide. Over the Patient Units and Swing Mount Units Over the Patient Units and Swing Mount Units Table of Contents Page General Information 2 Operation 3-5 Dental Unit Adjustments 6-10 Cleaning, Disinfecting and Sterilization 11-12 Maintenance 13-15 Page

More information

U S E R G U I D E. Home Respiratory Equipment Cleaning Guide. Health Information

U S E R G U I D E. Home Respiratory Equipment Cleaning Guide. Health Information Health Information U S E R G U I D E Home Respiratory Equipment Cleaning Guide This document is intended for Patients and Families for use at home The goal of the cleaning guide is to properly handle,

More information

Cleaning Telemonitor & Accessories

Cleaning Telemonitor & Accessories Doc Active Date: 10/15/2012 Doc Review Date: 10/15/2013 Standard Operating Procedure (SOP) Cross Referencing: Honeywell acceptable cleaners/disinfectants; Honeywell return procedure for infected equipment;

More information

Christina Bradley Laboratory Manager Hospital Infection Research Laboratory Queen Elizabeth Hospital Birmingham, UK

Christina Bradley Laboratory Manager Hospital Infection Research Laboratory Queen Elizabeth Hospital Birmingham, UK Christina Bradley Laboratory Manager Hospital Infection Research Laboratory Queen Elizabeth Hospital Birmingham, UK Irrespective of where the decontamination is performed or who it is performed by the

More information

Rapid Heat Sterilizers with 6, 8, and 12 Minute Sterilization Cycle Times

Rapid Heat Sterilizers with 6, 8, and 12 Minute Sterilization Cycle Times Rapid Heat Sterilizers with 6, 8, and 12 Minute Sterilization Cycle Times USER MANUAL MODELS: COX 115V-N COX 220V-N COX 115V-R COX 220V-R 2364 Leicester Road, P.O. Box 175, Leicester, NY 14481 Phone (585)

More information

Aesculap Implant Systems

Aesculap Implant Systems Aesculap Implant Systems Aesculap Implant Systems Orthopedics USA Instructions for Use/Technical Description Page 1 of 14 Symbols on product packages Caution, general warning symbol Caution, see documentation

More information

www.endgermsnow.com 1-855-220-0540 Evergreen Cleaning & Sanitizing Solutions Inc. in association with Glissen Chemical Company Inc. presents: The Green Alternative Biodegradable No toxic ingredients Cosmetic

More information

Operator s Manual. Model G17-EU (International) Disinfection Soak Station for Hysteroscopes, Cystoscopes and ENT scopes

Operator s Manual. Model G17-EU (International) Disinfection Soak Station for Hysteroscopes, Cystoscopes and ENT scopes Model G17-EU (International) Disinfection Soak Station for Hysteroscopes, Cystoscopes and ENT scopes Operator s Manual CIVCO Medical Solutions 102 First Street South Kalona, IA 52247 USA Tel: 1-800-445-6741

More information

Manual Pre-disinfection/Cleaning and Sterilization Instructions for SATELEC wrenches

Manual Pre-disinfection/Cleaning and Sterilization Instructions for SATELEC wrenches Caution Manual Pre-disinfection/Cleaning and Sterilization Instructions for SATELEC wrenches Do not use steel wool or abrasive cleaners. Avoid solutions containing iodine or with a high chlorine content.

More information

Table of Contents. English

Table of Contents. English OM-E0799E 000 English Thank you for purchasing VIVA ace Motor Kit. Please read this Operation Manual and the VIVA ace Basic Set Operation Manual carefully before use for operating instructions and care

More information

The Do's and Don ts in the Instrument Sterilisation Process

The Do's and Don ts in the Instrument Sterilisation Process The Do's and Don ts in the Instrument Sterilisation Process Just like a skilfully placed filling, the dental sterilisation process requires fine attention to detail. One overlooked step during this critical

More information

CLEANING AND HANDLING OF WRIGHT INSTRUMENTS

CLEANING AND HANDLING OF WRIGHT INSTRUMENTS CLEANING AND HANDLING OF WRIGHT INSTRUMENTS 150824-0 The following languages are included in this packet: English (en) Deutsch (de) Nederlands (nl) Français (fr) Español (es) Italiano (it) Português (pt)

More information

Pharmawiki.in. Principle of HOT AIR OVEN (Dry heat sterilization)

Pharmawiki.in. Principle of HOT AIR OVEN (Dry heat sterilization) Hot Air Oven Working Principle Sterilization Labelled Diagram Temperature [ #PDF PPT ] is the main theme of this article. Sterilization and aseptic processing are essential practices for healthcare product

More information

CLEANING AND SANITATION Policy Owner: Erica Tennessen Policy Originated by (program or department): Program Applicable Programs: All Programs

CLEANING AND SANITATION Policy Owner: Erica Tennessen Policy Originated by (program or department): Program Applicable Programs: All Programs CLEANING AND SANITATION Policy Owner: Erica Tennessen Policy Originated by (program or department): Program Applicable Programs: All Programs Statutory or Regulatory Citation: Minn. Stat., Chap. 4626 Minn.

More information

Cleaning and disinfecting

Cleaning and disinfecting Cleaning and disinfecting The goal of infection control is to prevent the transmission of infection to patients/subjects and staff during pulmonary function testing. Cleaning and disinfecting instructions

More information

I.F.U Re/Processing Reusable Medical Devices

I.F.U Re/Processing Reusable Medical Devices I.F.U Re/Processing Reusable Medical Devices Distributed by: MANUFACTURER: GA 26-09-001 -EN /20130719 Contents 1 Overview of preparation methods... 3 2 Safety and responsibility... 4 3 Explanation of symbols...

More information

Commit 2 Clean TM/MC. Bloodborne Pathogens Awareness Program

Commit 2 Clean TM/MC. Bloodborne Pathogens Awareness Program Commit 2 Clean TM/MC Bloodborne Pathogens Awareness Program Commit 2 Clean TM/MC Bloodborne Pathogens Awareness Program Table of Contents Cleaning Minor Spills on Hard Surfaces... 2 Cleaning Major Spills

More information

WORK IN THE DISINFECTION ROOM WITH WASHER-DISINFECTORS

WORK IN THE DISINFECTION ROOM WITH WASHER-DISINFECTORS WORK IN THE DISINFECTION ROOM WITH WASHER-DISINFECTORS Always with you Contents Introduction 3 Levels of cleanliness 5 Cleaning and disinfection processes 6 Washer-disinfector for human-waste containers

More information

Reprocessing Reusable Medical and Dental Equipment: Policy and Protocols First Nations and Inuit Health Alberta Region

Reprocessing Reusable Medical and Dental Equipment: Policy and Protocols First Nations and Inuit Health Alberta Region Reprocessing Reusable Medical and Dental Equipment: Policy and Protocols First Nations and Inuit Health Alberta Region This document was prepared for First Nations and Inuit Health, Alberta Region in consultation

More information

FOOD SERVICES DIVISION. November Employee Safety & Sanitation Training For the School Year

FOOD SERVICES DIVISION. November Employee Safety & Sanitation Training For the School Year FOOD SERVICES DIVISION November Employee Safety & Sanitation Training For the School Year 2014 2015 The Cafeteria Manager or Sr. FSW must allocate 15 to 30 minutes each month to conduct an employee Safety

More information

Instructions for care, maintenance, cleaning and sterilization of Smith & Nephew orthopaedic devices

Instructions for care, maintenance, cleaning and sterilization of Smith & Nephew orthopaedic devices Instructions for care, maintenance, cleaning and sterilization of Smith & Nephew orthopaedic devices 71381339 REVA CLEANING AND STERILIZATION BRO.indd 1 5/21/12 3:41 PM 71381339 REVA CLEANING AND STERILIZATION

More information

Safe Use of a Class II Biological Safety Cabinet

Safe Use of a Class II Biological Safety Cabinet UBC Biosafety Committee 102-6190 Agronomy Rd. Vancouver BC, V6T 1Z3 Standard Operating Procedure General Use IBC-SOP-002 Effective date: November 20, 2013. Review date: April 30, 2014. Supersedes: N/A.

More information

SANITARY PRACTICES POLICIES AND PROCEDURES

SANITARY PRACTICES POLICIES AND PROCEDURES SANITARY PRACTICES POLICIES AND PROCEDURES December 2016 This section contains sanitary practices that must be followed for the following activities: Hand Washing for Caregivers and Children Use of Disposable

More information

The future is bright. And clean. And fully automated.

The future is bright. And clean. And fully automated. The future is bright. And clean. And fully automated. HYDRIM a remarkable collection of instrument washers intended not just to meet expectations, but exceed them. Quick, quiet and labour saving, the HYDRIM

More information

CLEANING FOR EXTENDED TOOL LIFE & IMPROVED TABLET QUALITY. natoli.com

CLEANING FOR EXTENDED TOOL LIFE & IMPROVED TABLET QUALITY. natoli.com CLEANING FOR EXTENDED TOOL LIFE & IMPROVED TABLET QUALITY natoli.com PROPER TOOL CLEANING IS CRUCIAL EXTEND TOOL LIFE DECREASE PRESS DOWNTIME REDUCE PRODUCTION-RELATED COSTS IMPROVE TABLET QUALITY Tablet

More information

Infection Control Manual Residential Care Part 3 Infection Control Standard IC9: Environmental

Infection Control Manual Residential Care Part 3 Infection Control Standard IC9: Environmental IC9: 0600 LAUNDRY 1.0 STANDARD All staff must follow Routine Practices when handling soiled Off site processing (general linen) Off site processing (general linen) will not require special separation of

More information

X99 Series. Dental Handpiece. Instruction For Use

X99 Series. Dental Handpiece. Instruction For Use X99 Series Dental Handpiece Instruction For Use 0413 Table of Contents Symbols... 2 Introduction... 3-4 Before Use... 5 Product description... 6 Technical Specifications... 7 Operation... 8-11 Hygienic

More information

CONVENTUS CAGE TM PH Procedure Instruments Instrument Tray Cleaning and Sterilization

CONVENTUS CAGE TM PH Procedure Instruments Instrument Tray Cleaning and Sterilization CONVENTUS CAGE TM PH Procedure Instruments Instrument Tray Cleaning and Sterilization 1. WARNINGS AND PRECAUTIONS 1.1. Caution should be exercised when handling instruments with sharp points, or cutting/drilling

More information

INSTRUCTIONS FOR CLEANING AND STERILIZATION OF ALL B.BRAUN AESCULAP HAND HELD SURGICAL INSTRUMENTS AND POWER PRODUCTS (UNITED KINGDOM REQUIREMENTS)

INSTRUCTIONS FOR CLEANING AND STERILIZATION OF ALL B.BRAUN AESCULAP HAND HELD SURGICAL INSTRUMENTS AND POWER PRODUCTS (UNITED KINGDOM REQUIREMENTS) INSTRUCTIONS FOR CLEANING AND STERILIZATION OF ALL B.BRAUN AESCULAP HAND HELD SURGICAL INSTRUMENTS AND POWER PRODUCTS (UNITED KINGDOM REQUIREMENTS) 1 INTRODUCTION This paper is intended to give general

More information

PICNet CONFERENCE 2013

PICNet CONFERENCE 2013 PICNet CONFERENCE 2013 SCOPING THE HORIZON Penny Brawn THE ORIGINAL STEAM STERILIZERS ORIGINAL STEAM STERILIZER A LITTLE HISTORY Back in 1450 BC in Moses time, writings were found in books of Leviticus

More information

CONVENTUS CAGE TM PH Procedure Instruments Instrument Tray Cleaning and Sterilization

CONVENTUS CAGE TM PH Procedure Instruments Instrument Tray Cleaning and Sterilization CONVENTUS CAGE TM PH Procedure Instruments Instrument Tray Cleaning and Sterilization 1. WARNINGS AND PRECAUTIONS 1.1. Caution should be exercised when handling instruments with sharp points, or cutting/drilling

More information

OXYGEN CONCENTRATORS Information For Our Customers

OXYGEN CONCENTRATORS Information For Our Customers 811 Bevins St. Lakeport, CA 95453 Phone 707-263-9888 OXYGEN CONCENTRATORS Information For Our Customers An oxygen concentrator is a machine that separates the oxygen from the nitrogen in the air, concentrates

More information

INSTRUCTIONS FOR CLEANING AND STERILISING HUMECA MEDICAL EQUIPMENT FOR BOTH ELECTRIC AND MANUAL DEVICES

INSTRUCTIONS FOR CLEANING AND STERILISING HUMECA MEDICAL EQUIPMENT FOR BOTH ELECTRIC AND MANUAL DEVICES INSTRUCTIONS FOR CLEANING AND STERILISING HUMECA MEDICAL EQUIPMENT FOR BOTH ELECTRIC AND MANUAL DEVICES Contents: Pagina 1 Introduction...2 2 Treatment instructions...2 3 Cleaning...2 3.1 Place of use...4

More information

KEY OPERATOR. Study guide TRAINING SERIES ANDERSEN PRODUCTS, LTD. Key Operator Training Kit. The power of knowledge. the future of gas sterilisation

KEY OPERATOR. Study guide TRAINING SERIES ANDERSEN PRODUCTS, LTD. Key Operator Training Kit. The power of knowledge. the future of gas sterilisation KEY OPERATOR TRAINING SERIES ANDERSEN PRODUCTS, LTD Study guide 4 Key Operator Training Kit The power of knowledge the future of gas sterilisation 4 Thank you for using the EOGas sterilisation system.

More information

INSTRUCTIONS EVIS EXERA II DUODENOVIDEOSCOPE

INSTRUCTIONS EVIS EXERA II DUODENOVIDEOSCOPE Single use soft brush (MAJ-1888) INSTRUCTIONS EVIS EXERA II DUODENOVIDEOSCOPE OLYMPUS TJF TYPE Q180V Accessories: Water resistant cap (MH-553) Biopsy valve (MB-358) Air/water valve (MH-438) Suction cleaning

More information

Corporate. Laundry SOP. Document Control Summary

Corporate. Laundry SOP. Document Control Summary Corporate Laundry SOP Document Control Summary Status: Version: Author/Title: Owner/Title: Approved by: Ratified: Related Trust Strategy and/or Strategic Aims Implementation Date: Review Date: Key Words:

More information

Effective Date: 0 8 / 1 0 / 1 7. Replaces: 08/11/16. Formulated:1997 SPECIAL (MEDICAL) WASTE MANAGEMENT

Effective Date: 0 8 / 1 0 / 1 7. Replaces: 08/11/16. Formulated:1997 SPECIAL (MEDICAL) WASTE MANAGEMENT Replaces: 08/11/16 Page 1 of 5 POLICY: Special waste generated on each Texas Department of Criminal Justice (TDCJ) unit will be handled according to the requirements of law and good infection control practices.

More information

CLEANING OF INSTRUMENTS

CLEANING OF INSTRUMENTS Disclaimer: This article is provided to Keir Surgical Ltd. by SPS Medical. The information in this document should be used in conjunction with your hospital or institution s cleaning and decontamination

More information

CLEANING AND SANITIZING

CLEANING AND SANITIZING CLEANING AND SANITIZING Cleaning and sanitizing procedures must be part of the standard operating procedures that make up your food safety program. Improperly cleaned and sanitized surfaces allow harmful

More information

Aesculap Implant Systems

Aesculap Implant Systems Aesculap Implant Systems Aesculap Implant Systems Orthopedics USA Instructions for Use/Technical Description Page 1 of 16 Symbols on product packages! Caution, general warning symbol Caution, see documentation

More information

KEY OPERATOR STUDY GUIDE

KEY OPERATOR STUDY GUIDE EOGAS 4 STERILZER 4 Thank you for using the EOGas 4 sterilization system. The active ingredient in EOGas 4 is ethylene oxide (EtO or EO). EtO is a powerful anti-microbial agent; it can also be dangerous

More information

HYGIENE AND SANITATION

HYGIENE AND SANITATION HYGIENE AND SANITATION In accordance with the policy of the board of education, the following regulation and attached procedures contains guidelines for cleaning and disposal of body fluids. For the purpose

More information

Central Sterile Services Department will receive, decontaminate, process, store and/or dispense sterile and non-sterile to departments requesting.

Central Sterile Services Department will receive, decontaminate, process, store and/or dispense sterile and non-sterile to departments requesting. Title/Description: Central Sterile Services Department: Central Sterile Services Personnel: Central Sterile Services Personnel Effective Date: 1/97 REVISED: 1/97, 11/07, 2/12 I. PURPOSE Central Sterile

More information

Instruction Manual for the Revolutionary Science Saniclave

Instruction Manual for the Revolutionary Science Saniclave Instruction Manual for the Revolutionary Science Saniclave Model RS-SC-102 and RS-SC-50 REVOLUTIONARY SCIENCE Making Science Simple Table of Contents Introduction 2 Display, Symbols, Control and Parts

More information

MARSHALLTOWN MEDICAL AND SURGICAL CENTER ENVIRONMENTAL SERVICES. SUBJECT: Room Cleaning Occupied Room (All Areas) PROCEDURE NO: 032

MARSHALLTOWN MEDICAL AND SURGICAL CENTER ENVIRONMENTAL SERVICES. SUBJECT: Room Cleaning Occupied Room (All Areas) PROCEDURE NO: 032 MARSHALLTOWN MEDICAL AND SURGICAL CENTER ENVIRONMENTAL SERVICES SUBJECT: Room Cleaning Occupied Room (All Areas) PROCEDURE NO: 032 PURPOSE: To provide a systematic approach to remove the accumulation of

More information

INSTRUCTIONS EVIS EXERA II DUODENOVIDEOSCOPE

INSTRUCTIONS EVIS EXERA II DUODENOVIDEOSCOPE Single use soft brush (MAJ-1888) INSTRUCTIONS EVIS EXERA II DUODENOVIDEOSCOPE OLYMPUS TJF TYPE Q180V Accessories: Water resistant cap (MH-553) Biopsy valve (MB-358) Air/water valve (MH-438) Suction cleaning

More information

Reusable Laryngoscope Blades and Handles

Reusable Laryngoscope Blades and Handles Manufacturer: Starkling Medizintechnik Method: Re-processing instructions Symbol: N.A. Device(s): 1 Green System Reusable Laryngoscope Blade with integrated Light Guide, Solid Stainless Steel Construction

More information

Purpose. Team Roles and Responsibilities. System Requirements. Instrument Processing Training

Purpose. Team Roles and Responsibilities. System Requirements. Instrument Processing Training Purpose Team Roles and Responsibilities System Requirements Reliably providing sterile instruments for patients is a vital component in the Trusted Care environment. Instrument processing is a critical

More information

Visao HIGH-SPEED OTOLOGIC DRILL CATALOG

Visao HIGH-SPEED OTOLOGIC DRILL CATALOG Visao HIGH-SPEED OTOLOGIC DRILL CATALOG Visao High-Speed Otologic Drill with the Integrated Power Console (IPC ) System High speed and torque 80,000 rpm max speed provides rapid, yet precise bone removal

More information

Operator s Manual. Model G17 Disinfection Soak Station for Hysteroscopes, Cystoscopes and ENT scopes

Operator s Manual. Model G17 Disinfection Soak Station for Hysteroscopes, Cystoscopes and ENT scopes Model G17 Disinfection Soak Station for Hysteroscopes, Cystoscopes and ENT scopes Operator s Manual CIVCO Medical Solutions 102 First Street South Kalona, IA 52247 USA Tel: 1-800-445-6741 Fax: 1-877-329-2482

More information

Flexible Nasopharyngoscope. Operator Manual. Manufactured by Henke-Sass, Wolf Distributed by Anthony Products, Inc.

Flexible Nasopharyngoscope. Operator Manual. Manufactured by Henke-Sass, Wolf Distributed by Anthony Products, Inc. Flexible Nasopharyngoscope Operator Manual Manufactured by Henke-Sass, Wolf Distributed by Anthony Products, Inc. 1. Introduction The flexible nasopharyngoscope is sold in unsterile condition as a reusable

More information

User Guide for Reprocessing of the SUPER VIEW Wide Angle Viewing System Sterilizable Components and Accessories

User Guide for Reprocessing of the SUPER VIEW Wide Angle Viewing System Sterilizable Components and Accessories User Guide for Reprocessing of the SUPER VIEW Wide Angle Viewing System Sterilizable Components and Accessories Table of Contents 1. Intended Use Statement... 3 2. Proper Use... 3 3. General Information...

More information

N2736/1017 HU-FRIEDY. install the new unit, follow the instructions above for installing a new product.

N2736/1017 HU-FRIEDY. install the new unit, follow the instructions above for installing a new product. You re on your way to cleaner water! Inhibits the growth of microbial contamination in dental waterlines. Meets or exceeds

More information

2364 Leicester Road, P.O. Box 175, Leicester, NY Phone (585) Fax (585)

2364 Leicester Road, P.O. Box 175, Leicester, NY Phone (585) Fax (585) Dry Heat Sterilizers with 60 Minute Sterilization Cycle Times MODEL 3100 MODEL 2100 2364 Leicester Road, P.O. Box 175, Leicester, NY 14481 Phone (585) 382-3223 Fax (585) 382-9481 www.cpac.com December

More information

Aim. Learning outcomes. 13 May Megan Reilly RN BN Cert IV Assessment & Workplace Training MHlthSc (Inf Control) Immunise Cert MACN CICP 1

Aim. Learning outcomes. 13 May Megan Reilly RN BN Cert IV Assessment & Workplace Training MHlthSc (Inf Control) Immunise Cert MACN CICP 1 Aim Apply the principles and practices of reprocessing reusable podiatry instruments and equipment to the officebased practice setting Learning outcomes Describe the chain of infection related to the office-based

More information

Series 3 + Thank you for using the EOGas 3 sterilization system.

Series 3 + Thank you for using the EOGas 3 sterilization system. Thank you for using the EOGas 3 sterilization system. The active ingredient in EOGas 3 is ethylene oxide (EtO or EO). EtO is a powerful anti-microbial agent; it can also be dangerous if not handled correctly.

More information

CHILDCARE SETTINGS. Childcare Settings

CHILDCARE SETTINGS. Childcare Settings Childcare Settings This document provides guidelines and recommendations for cleaning childcare settings such as nurseries, crèches and schools in the event of a disease outbreak in the vicinity. Cleaning

More information

LABORATORY CHEMICAL HYGIENE AND SAFETY PLAN

LABORATORY CHEMICAL HYGIENE AND SAFETY PLAN LABORATORY CHEMICAL HYGIENE AND SAFETY PLAN Laboratory Room Number (s) Building Principle Investigator or Responsible Individual (s) Date Review / Revision Date INTRODUCTION Kent State University has made

More information

STERITITE INSTRUCTIONS FOR USE

STERITITE INSTRUCTIONS FOR USE STERITITE INSTRUCTIONS FOR USE If a rigid container system is used, the manufacturers instructions regarding set preparation and assembly should be followed. -ANSI/AAMI ST79: 2010 & A1:2010 Notes from

More information

KEY OPERATOR STUDY GUIDE

KEY OPERATOR STUDY GUIDE Thank you for using the Anprolene sterilization system. The active ingredient in Anprolene is ethylene oxide (EtO or EO). EtO is a powerful anti-microbial agent; it can also be dangerous if not handled

More information

MARSHALLTOWN MEDICAL AND SURGICAL CENTER LAUNDRY/LINEN SERVICES. SUBJECT: Infection Control POLICY NO: 106

MARSHALLTOWN MEDICAL AND SURGICAL CENTER LAUNDRY/LINEN SERVICES. SUBJECT: Infection Control POLICY NO: 106 MARSHALLTOWN MEDICAL AND SURGICAL CENTER LAUNDRY/LINEN SERVICES SUBJECT: Infection Control POLICY NO: 106 PURPOSE: The activities of collecting, processing and delivering an adequate supply of clean linens

More information

When you successfully complete the test, you will receive a certificate and a registered key ring. We look forward to hearing from you.

When you successfully complete the test, you will receive a certificate and a registered key ring. We look forward to hearing from you. AN74i AN74ix Study Guide Thank you for using the Anprolene sterilization system. The active ingredient in Anprolene is ethylene oxide (EtO or EO), a chemical that can be hazardous if not handled properly.

More information

Decontamination Process Manual

Decontamination Process Manual Decontamination Process Manual (in dental practices) (First Edition) Prepared by : Superintendent, Quality Assurance & Infection Control, Dental Administration, MOH. Kuwait. Acknowledgement This is the

More information

What If No Water? Children, nursing mothers, and sick people may need more water.

What If No Water? Children, nursing mothers, and sick people may need more water. What If No Water? Water is an essential element to survival and a necessary item in an emergency supplies kit. Following a disaster, clean drinking water may not be available. Your regular water source

More information

Call our hotline at for all your instrument care and detergent questions. Medline Super Concentrates

Call our hotline at for all your instrument care and detergent questions. Medline Super Concentrates Everybody Cares About Instrument Care Surgeons, O.R. staff and C.S. workers agree that proper instrument care makes sense. Besides prolonging the life of a costly investment, regular maintenance results

More information

Appendix C IBC Safe Use of Biosafety Cabinets SOP

Appendix C IBC Safe Use of Biosafety Cabinets SOP Purpose To provide details for the safe operation of biological safety cabinets (BSCs) and ensure adequate containment of biohazards. Biological safety cabinets shall be used: a. For handling biohazardous

More information

CPAP Washbasin Manual

CPAP Washbasin Manual CPAP Washbasin Manual www.mtts-asia.com CPAP-WB-MAN-01-REV3.02EN Company information MEDICAL TECHNOLOGY TRANSFER AND SERVICES CO., LTD No. 26, Lane 41, An Duong Vuong street, Tay Ho district, Hanoi, Vietnam

More information

THE ULTRASONIC DISRUPTION OF ORGANIC SOIL ON COMPLEX INSTRUMENTATION

THE ULTRASONIC DISRUPTION OF ORGANIC SOIL ON COMPLEX INSTRUMENTATION CFSA CSSD Conference Durban ICC Wednesday 8 th March 2017 Using technology to improve patient safety THE ULTRASONIC DISRUPTION OF ORGANIC SOIL ON COMPLEX INSTRUMENTATION Helen Loudon Independent Infection

More information

PROFESSIONAL USER GUIDELINES FOR PRACTICE INFECTION CONTROL AND HYGIENE IN VETERINARY EQUINE PRACTICE

PROFESSIONAL USER GUIDELINES FOR PRACTICE INFECTION CONTROL AND HYGIENE IN VETERINARY EQUINE PRACTICE THEATRE AND STERILE PROCEDURES WALLS/TILES - Wash all surfaces from top to bottom using F10CLXD, rinse with - Wash off visible dirt (e.g. blood) using F10CLXD CHAIRS - Trigger spray using F10CL and wipe

More information

SUGGESTED POLICY FOR THE MONITORING AN ULTRASONIC CLEANER DAILY

SUGGESTED POLICY FOR THE MONITORING AN ULTRASONIC CLEANER DAILY SUGGESTED POLICY FOR THE MONITORING AN ULTRASONIC CLEANER DAILY SUBJECT: ULTRASONIC CLEANER MONITORING DAILY DEPARTMENT: CPD, CSSD, SPA, MDRD, SPD, dental offices, medical offices and department using

More information