1 Gas Liquid & Slurry Solid & Powder Steam & Water Installation, Operation & Maintenance Manual Original Instructions Liquid & Slurry Sampling Sampler Controllers S-LS-IOM Sentry COS 966 Blue Ribbon Circle North, Oconomowoc, WI

2 2 Sentry COS

3 Sentry COS 3 Table of Contents Safety Information... 4 General Description... 5 Features... 5 Touchscreen Navigation... 6 Specifications... 7 Special Conditions for Safe Use... 8 Installation... 9 Grounding Requirements... 9 Power/IS Wiring... 9 Set-Up Checklist...10 Setting Parameters...11 IO Setup...12 Sampler Setup...13 Cans...14 System Settings...15 Operation...16 Reading the Main Screen...16 Starting the Sampler...17 Stopping the Sampler...17 Resetting the Counter...18 Replacing Sample Containers...18 Tare Scales...19 Changing Number of Cans...19 Acknowledging/Clearing Alarms...20 List of Alarms...21 Performance Factors...21 Maintenance...23 Troubleshooting...23 Parts & Accessories...24 Standard Warranty...25 Customer Assistance...27

4 4 Sentry COS Safety Information Please read the entire manual before attempting to unpack, set up or operate this product. Pay careful attention to all Warnings, Cautions, and Notes. Failure to do so could result in serious personal injury and/or equipment damage. Use of Hazard Information If multiple hazards exist, the signal word corresponding to the greatest hazard shall be used. Definitions DANGER: Indicates a hazardous situation which, if not avoided, will result in death or serious injury. WARNING: Indicates a hazardous situation which, if not avoided, could result in death or serious injury. CAUTION: Used with the safety alert symbol, indicates a hazardous situation which, if not avoided, could result in minor or moderate injury. NOTICE: Used to address practices not related to personal injury. NOTE: Information that requires special emphasis. SHALL: This word is understood to be mandatory. SHOULD: This word is understood to be advisory. Product Selection, Installation & Use Improper selection, installation, or use can cause personal injury or property damage. It is solely the responsibility of users, through their own analysis and testing, to select products suitable for their specific application requirements, ensure they are properly maintained, and limit their use to their intended purpose. Follow proper local, state and federal regulations for proper installation and operational requirements. Always use caution and common sense when working with any chemical. Read the product label and MSDS carefully and follow the instructions exactly. Potential Equipment Hazards Hot surfaces! This equipment may have very hot surfaces. If an operator contacts a hot surface, injury may occur. Use protective clothing to prevent injury. If other equipment comes in contact with a hot surface, damage to the equipment may occur. Ensure the area around this equipment is kept clear to prevent damage from occurring. High pressures! This equipment may contain fluids at very high pressures. Prior to installing, removing or maintaining this equipment, ensure that the equipment is isolated from all connecting piping, the equipment is depressurized, the contents have been drained, and the equipment is cool. Moving parts! This equipment may contain moving parts. All drive guards and doors must be secured in place when this machine is being operated.

5 Sentry COS 5 General Description The Sentry COS sampler controller is designed to operate the API-P and API-C samplers to capture accurate and repeatable samples from liquid & slurry process lines. The COS sampler controller is user-friendly with a simple menu system that provides quick access to frequently used parameters. The COS sampler controller is available in an explosion proof/safe area enclosure with a touchscreen interface. Features Simple to use Quick access to commonly used parameters Easy control via a DCS Choice of operating modes: time proportional or flow proportional Up to 30 samples per minute Flow meter inputs can be analog or pulse On-display alarm indicators Alarms indicated until acknowledged Automatic or manual can selection Analog scale(s) for can weighing Automatic stop to prevent overfilling cans Performance factor calculations to verify system integrity Live density monitoring option

6 6 Sentry COS General Description, continued Touchscreen Navigation To navigate through the various screens on the controller, use the arrow keys on the controller touchscreen display or the touch keys on the window of the enclosure. Touchscreen Touch Keys Description Up Arrow: Move up through a list of selections or increase the numerical value of a parameter Down Arrow: Move down through a list of selections or decrease the numerical value of a parameter Enter: Move to the next screen or select a parameter in order to change its value Back/Escape: Move to the previous screen or deselect a parameter after changing its value u Note: The touch keys on the window of the enclosure are very sensitive. Be aware that a key may detect your finger or hand hovering close to the window and activate without actual physical contact.

7 Sentry COS 7 Specifications Model No: COS Controller Power Input Voltage: VAC / 1 ph / Hz / 5.5 A Frequency Range: Hz 1 phase Input Current: 5.5 A Motor rating: V, 1.2 A max IS Inputs Three user configurable digital inputs CPC 1 prox Can 1 prox CPC 2 prox Can 2 prox General alarm Switching valve FB DPT alarm Skid stop Two 4 20 ma analog inputs for flow rate and density One pulse input for flow rate Motor VFD ABB #ACS150-01U-02A4-2 IS Outputs Two digital outputs for can weight alarms Two 4 20 ma analog outputs for can weight values Communications Modbus TCP/IP Keypad Four-button capacitive keypad Display 320 x 240 pixel touchscreen display graphics and text Enclosure Flameproof, aluminum NEMA 4X, IP 66 Mounting Surface Ambient Temperature -4 F to 122 F (-20 C to 50 C) Shipping Weight ~150 lb (68 kg) Agency Approvals Available CE0891 II 2(1) G Exd [ia Ga] IIA T6 Gb ATEX certificate # EMT16ATEX0014X

8 8 Sentry COS Special Conditions for Safe Use All cable entry devices shall be suitably ATEX approved (type d ) and rated at least IP66. All unused openings shall be closed with suitably ATEX approved (type d ) and rated at least IP66 blanking elements. For ambient temperatures below -10 C, use field wiring that is suitable for the minimum ambient temperature. Zinc plated steel fasteners with a minimum yield strength of 240 N/mm2 or greater bolts must be used. All bolts must be securely fastened with a tightening torque of 129 Nm. 5. All connections and components used with the intrinsic safety barriers must be used and connected in accordance with the separate ATEX approvals for each barrier. Intrinsic safety connections shall be done in accordance with Sentry drawings D and E. 6. The motor used with the controller must be suitably ATEX approved for use with VFD Drive ABB, PN ACS150-01U-02A4-2. The motor shall not draw more than 1.2 Amps (per phase) under full load conditions. 7. All field wiring connections must be made using wire that is rated at least 80 C. 8. No holes or modification to the enclosure shall be made by the end user. 9. Enclosures must be bonded to ground both internally and externally with suitably sized conductors (EN/IEC , cl. 15.3). 10. Enclosure shall not be opened when energized or when an explosive atmosphere is present.

9 Sentry COS 9 Installation Do not install or use this equipment in any manner other than that which is specified in this manual or the protection offered by this equipment could be impaired. Installation should be performed by qualified personnel in accordance with applicable state/ provincial and local codes and procedures. 1. Mount the controller in a location that is easily accessible and vibration-free. 2. Have a qualified electrician wire the controller to the electrical supply. Surge suppression and filtering is recommended but not required. 3. Mount the enclosure with proper structural support to accommodate the 150 lb controller. 4. The controller should be hard wired to instrument-quality power using appropriate ATEX type d certified conduit, fittings, and wiring or cable. 5. Add ferrite to the AC mains input (double loop). 6. A shielded cable (Belden 29550C or equivalent) must be used for the connection to the motor, and it must be tied to ground on the VFD and motor. 7. A suitable external over-current protection device, such as a fuse or circuit breaker (15 A), and disconnect device is required. Reference the electrical wiring diagram supplied with your equipment for actual circuit power requirements. 8. The overcurrent protection and disconnect should be near the equipment and easily reached. 9. The overcurrent protection should be marked as the disconnect for the controller. 10. The overcurrent protection and disconnect must meet IEC and IEC The over-current protection and disconnect devices should be installed on both the hot (L) and neutral (N) leads. 12. The disconnect device shall be located near the equipment and marked with appropriate On(1)/Off(O) markings as specified by local codes. 13. To ensure a safe circuit is maintained, follow intrinsic safety connections, wiring, parameters, etc. Reference drawing D and E. Grounding Requirements Connect a ring terminal ground lug to the ground wire. Secure the ground lug to the enclosure ground stud, which is marked with a ground symbol and is located next to the customer power connection. For an insulated ground conductor, use a green with yellow strip conductor that is the same size as the mains conductor. Reference drawing H. For external ground, connect ground lug to ground wire. The location is marked with a ground symbol and is located on the lower right corner behind the door flange. Power/IS Wiring Cable length is determined by capacitance/inductance of the IS barriers. Maintain 50mm distance between IS/non IS wiring.

10 10 Sentry COS Installation, continued Set-Up Checklist Before operating the controller, set the parameters for your application. Use the following checklist as a guide for items that you should consider and check when installing the unit for the first time. IO Setup Flow input I/O select (4 20 ma or pulse) Flow input ma settings Flow input pulse settings Flow units Sampler Setup Cans Sample mode (time proportional or flow proportional) Total samples Density Pipe diameter Transfer period (used in time proportional mode) Barrels per grab setpoint (used in flow proportional mode) Sample size Enable or disable cans Tare cans Select can Tare hysteresis Switching volume Can volume Enable or disable can proximity sensor Number of cans (1 or 2)

11 Sentry COS 11 Installation, continued Setting Parameters To set or change parameters on the controller, the sampler must be turned off. From the Main Screen, press enter to access the Menu screen.

12 12 Sentry COS Installation, continued IO Setup From the Menu screen, select IO Setup and press enter. From the IO Setup screens, set the appropriate parameters for your application: a. b. c. d. e. Flow Input I/O Select: Choose either 4-20mA or pulse input. Flow Input ma Settings: Set the scale for the ma input. Flow Input Pulse Settings: Set the scale for the pulse input. Flow Units: Select the units for the flow rate. To return to the Menu, press the back/escape button.

13 Sentry COS 13 Installation, continued Sampler Setup From the Menu screen, select Sampler Setup and press enter. On the Sampler Control Settings screen, set the appropriate parameters for your application: a. b. c. d. e. f. g. h. Sample Mode: Choose either Time Proportional or Flow Proportional. Total Samples: Set the total number of samples in the batch. Sampling will stop once this number is reached. Density: Set the density of the fluid you are sampling. Value is in grams per cubic centimeter (g/cc). Pipe Diameter: Set the diameter of the pipe that the sampler is attached to. Transfer Period: Set the duration of time in which the batch or total samples should be completed (used for time proportional sampling). Barrels per Grab Setpoint: Set the number of barrels that flow through the pipeline between samples (used for flow proportional sampling). Sample Size: Set the sample volume in cc. To return to the Menu, press the back/escape button.

14 14 Sentry COS Installation, continued Cans From the Menu screen, select Cans and press enter. On the Cans screen, set the appropriate parameters for your application: a. b. c. d. e. f. g. h. i. Cans: Enable or disable cans, or select CPC (constant pressure cylinder). Tare: Determine the tare of the cans (if weigh scales are used; not an option for CPC). Can Select: Choose which can is active; number changes automatically when controller switches to next can. Tare Hysteresis: Set the amount of variation allowed after a tare for a can to still be considered at 0. Switching Volume: Set the volume (in %) at which the controller automatically switches to the next can or stops sampling. Can Volume: Set the total volume of each can. Can Prox: Enable or disable a proximity sensor which indicates the presence of a can. Number of Cans: Indicate number of cans used (1 or 2). To return to the Menu, press the back/escape button.

15 Sentry COS 15 Installation, continued System Settings System settings must be made directly on the controller. There is no Settings button on the window of the enclosure. Do not change any system settings other than those detailed below. Changing system settings may affect the accuracy and operation of the controller. Set Time/Date a. b. c. d. Press the blue Settings button on the controller to access the system menu. Use the arrow keys to scroll down to Set Time/Date, then press enter. Follow the prompts to set the current time and date. When finished, press escape to return to the system menu. Set Networks a. b. c. d. Press the blue Settings button on the controller to access the system menu. Use the arrow keys to scroll down to Set Networks, then press enter. Enter the appropriate IP address, Netmask, Gateway information. When finished, press escape to return to the system menu.

16 16 Sentry COS Operation During sampling, the controller turns on the motor based on the parameters entered during installation, and stops the motor when a stop signal is received from the motor prox sensor. Sampling continues until the batch count is reached or until the sample container(s) can no longer be filled. Reading the Main Screen When power is applied to the controller, the unit cycles through several start-up screens until it reaches the Main Screen. The Main Screen displays the current status of the sampler and can(s) as well as the progress of the current batch: On/Off Indicator Green = sampler is on Red = sampler is off Sampling Indicator Green = sampler is actively sampling Red = sampler is not actively sampling (between grabs) Can 1/Can 2 Indicators Green = can is available Yellow = can needs to be tared Red = can is full Can Weight Current weight of the can in kilograms (kg) Count Number of grabs completed in the current batch Flowrate Current flowrate of the process fluid given in the selected unit of measure Performance Factor Displayed after the first 100 samples are taken

17 Sentry COS 17 Operation, continued Starting the Sampler The sampler must be turned on and off from the Main Screen. To turn the sampler on, hold the up arrow for three seconds. A confirmation screen appears. Press Enter to turn the sampler on, or press Back/Escape to leave the sampler off. If there is an issue that prevents the sampler from turning on, a notification screen appears. When the issue is resolved, try starting the sampler again. Stopping the Sampler The sampler stops automatically at the end of a batch, and also can be stopped before the end of a batch by turning it off. The sampler must be turned on and off from the Main Screen. To turn the sampler off, hold the down arrow for three seconds. A confirmation screen appears. Press Enter to turn the sampler off, or press Back/Escape to leave the sampler on.

18 18 Sentry COS Operation, continued Resetting the Counter The sampler stops automatically at the end of a batch. The sampler will not start sampling again until the batch counter is reset. The counter must be reset from the Main Screen. To reset the counter, hold the Back/Escape button for three seconds. A confirmation screen appears. Press Enter to reset the counter, or press Back/Escape to exit the confirmation screen without resetting the counter. Replacing Sample Containers Do not overfill constant pressure cylinders (CPCs). Use of proximity sensors with CPCs is required. Equipment failure due to overfilling a CPC could cause death or serious injury. Do not overfill cans. Process fluid leakage could create a slip hazard and/or cause equipment damage. Replace sample containers at the end of each batch or whenever they become full.

19 Sentry COS 19 Operation, continued Tare Scales If a weigh scale is used, tare the scale after replacing the sample container and before resuming operation. a. From the Menu screen, select Cans. b. From the Cans screen, select Tare. c. d. e. On the Tare screen, use the up/down arrows to select the scale to Tare. Press enter to Tare. Select another scale to Tare, or press Back/Escape to return to the Cans screen. Changing Number of Cans If a can is not replaced, or if an additional can is added, change the parameter for number of cans. a. From the Menu screen, select Cans. b. c. d. On the Cans screen, arrow down to Number of Cans and press Enter to select the parameter. Use the arrow keys to change the value, then press Back/Escape to deselect the parameter. Press Back/Escape again to return to the Menu screen.

20 20 Sentry COS Operation, continued Acknowledging/Clearing Alarms If an alarm condition occurs, the Alarms screen appears with a description of the alarm(s). Press the up arrow to acknowledge all alarms. Active alarms must be acknowledged before you can access the Main Screen. Press the down arrow to clear all alarms. A confirmation screen appears; press Enter to clear all alarms, or Back/Escape to return to the previous screen. Any alarm condition that is not resolved remains on the Alarm screen. Critical alarms stop the sampler immediately and must be resolved before operation can resume. Non-critical alarms do not stop the sampler and operation may continue.

21 Sentry COS 21 Operation, continued List of Alarms Type Alarm Description Critical Can 1 Over Limit Can 1 is over its switching weight Critical Can 2 Over Limit Can 2 is over its switching weight Critical Can 1 Removed Unexpectedly Can 1 was removed while unit was operating Critical Can 2 Removed Unexpectedly Can 2 was removed while unit was operating Non-critical Can 1 PF Failed Performance factor calculation has failed for current batch on can 1 Non-critical Can 2 PF Failed Performance factor calculation has failed for current batch on can 2 Non-critical Differential Pressure Sensor Differential pressure sensor has given an alarm Non-critical Low Flow Low flow sensor has given an alarm Non-critical Switching Valve FB [feedback] Position indication from valve does not match desired valve position Non-critical General Customer defined Active Alarms Active alarms include alarms that have been acknowledged, but not cleared. To view active alarms, select Alarms from the Menu screen. Alarm History All alarm conditions that occur are recorded and maintained in an alarm history list. To view this list, select Alarm History from the Menu screen. Performance Factors Viewing Performance Factor Values The PF (performance factor) value is displayed on the Main Screen after the batch count reaches 100 samples. Before the batch count reaches 100, or to view previous PF values, select PF Results from the Menu screen. The Performance factor values screen displays recent performance factors with a timestamp and can number. Use the up/down arrows to scroll through the previous 99 values.

22 22 Sentry COS Operation, continued Performance Factor Calculations There are two performance factors calculated by the COS controller: Can and Batch. Can performance factor is based on the expected sample volume for the current can versus the actual sample volume of the can. The can PF value calculation is based on total can weight since the last tare. CanPF = AccumulatedCanSampleVolume CalculatedCanSampleVolume The accumulated can sample volume is the actual volume collected for the can so far, where AccumulatedCanSampleVolume = AccumulatedCanSampleWeight MeanCanDensityinput The calculated can sample volume is the theoretical sample volume collected so far, where CalculatedCanSampleVolume = BatchVolumeThisCan x RequiredSampleVolume RequiredBatchVolume - BatchVolumeUptoPreviousCan Batch performance factor is based on the expected sample volume for the total batch processes so far versus the actual sample volume for the batch. This performance factor corresponds to the ANSI/ASTM D4177 standard. To get a batch PF calculation, you must tare the can between batches. BatchPF = AccumulatedBatchSampleVolume CalculatedBatchSampleVolume The accumulated batch sample volume is the actual volume collected so far, where AccumulatedBatchSampleVolume = AccumulatedBatchSampleWeight MeanDensityinput The calculated batch sample volume is the theoretical sample volume collected so far, where CalculatedBatchSampleVolume = MeasuredBatchVolume x RequiredSampleVolume RequiredBatchVolume

23 Sentry COS 23 Maintenance The COS controller does not require any scheduled maintenance. When necessary, wipe any dust off the outside window with a damp cloth. Electrical shock hazard. Disconnect all power, including voltages connected to relays, before servicing or replacing fuses. Should it be necessary to service the controller for any reason: Turn power off before opening the enclosure. Secure all bolts on the enclosure before applying power to the controller. Troubleshooting symptom possible problem(s) remedy Sampler will not start or stops unexpectedly - The unit is in alarm - Cans are not in place - Check controller settings: Is the proximity sensor engaged or - Correct items listed disengaged? Is the can volume correct? Is the switching volume correct? - Perform a span calibration. You need a 10 kg calibrated weight to complete the calibration procedure: 1. From the Menu screen, select Cans. Weight scales are incorrect - Span calibration needed 2. From the Cans screen, select Tare. 3. Tare both scales (see the Tare Scales section) 4. From the Tare Scale screen, press the up arrow for five (5) seconds. 5. A screen appears asking for confirmation that you wish to perform a span calibration. Press Yes. 6. The Cal Scale screen appears. Use the up/down arrows to select a scale to calibrate. 7. Place a calibrated 10 kg weight on the selected scale and press enter. 8. A screen appears, asking for confirmation that you want to save a span calibration value. Press Yes. 9. Repeat steps 6-8 for the next scale, as needed. 10. When span calibration is complete for both scales, press Back/Escape several times until you return to the Main Screen.

24 24 Sentry COS Parts & Accessories Description SEC P/N Quantity IS Barrier - mv Measuring Amplifier For Load Cells C 1 IS Barrier - Switching Amplifier Flow Pulse Input, DI F 1 IS Barrier - 2 ch Analogue Data Transmitter, 4-20mA AI G 1 IS Barrier - 2 ch Switching Amplifier, DI H 1 IS Barrier - 2 ch Power Supply For Load Cells, DO D 1 IS Barrier - 2 ch Analogue Data Transmitter, 4-20mA AO J 1 PLC/Display A 1 4 Channel AI, IO Card B 1 SmartMod Cable For IO Card P 1 Capacitive Touch Keypad E 1 24VDC Power Supply E 1 Inverter, Single Phase, VAC U 1 Holder, 1-Pole Fuse Class CC C 1 Fuse, 10A Time Delay Class CC G 1 Fuse, 6A Time Delay Class CC D 1 19 Pin Double Ended Cable, 5 Meters M 1 19 Pin Cabinet Receptacle N 1 4-Pin Minifast Conn, 6-8 mm Cable Diameter P 1 4-Pin Minifast Conn, mm Cable Diameter R 1 Shielded Cable, VFD (Belden 29550C or equivalent) X 1 Ferrite (Fair-Rite, 1.5 OD) Y 1

25 Sentry COS 25 Updated June 2015 Sentry Equipment Corp Standard Warranty Standard Warranty Sentry Equipment Corp ( Seller ) warrants products manufactured by it and supplied hereunder ( Products ) to be free from defects in workmanship and, to the extent materials are selected by Seller, to be free from defects in materials, in each case for a period as defined in the table below: Brand Product Line Warranty Period Sentry Steam & Water Sampling Products and Systems Solid & Powder Sampling Products and Systems Gas Sampling Products and Systems Liquid & Slurry Sampling Products and Systems Pipeline Integrity Products Eighteen months from date of shipment or twelve months from startup (whichever occurs first) Waters Equipment Steam & Water Sampling Products and Systems Twelve months from date of shipment If within such period the original purchaser ( Customer ) proves, to Seller s satisfaction, that any such Products are defective, such Products shall be repaired or replaced at Seller s option. Seller s sole obligation and Customer s exclusive remedy hereunder shall be such repair and replacement and shall be conditioned upon Seller s receiving written notice of any alleged defect within 10 days after its discovery and, at Seller s option, return of such Product to Seller, FOB Sentry Equipment Corp or provision of evidence (e.g., photographs) of such defect satisfactory to Seller. Warranty Conditions & Limitations This Warranty shall not apply to any Product which, in the opinion of Sentry Equipment Corp, has been (a) altered or repaired in a manner affecting the efficiency or performance of the unit, or (b) incorrectly installed or operated, or (c) damaged in shipment, or (d) damaged by flood or fire, or (e) if the serial number is missing, altered or defaced. Any materials required to be used by Seller as provided in Customer specifications or instructions are excluded from the foregoing warranty and Customer assumes sole responsibility for the selection of such materials. Customer further acknowledges and agrees that any Seller recommendations with respect to materials to be used in connection with Products will be based on published reference literature, that any third-party studies to which Seller refers may not correlate directly with the Customer s intended usage or process (i.e., chemical composition, concentrations, temperatures, etc.), and that Customer is solely responsible for the final determination with respect to which materials are to be used in connection with the Products. EXCEPT FOR THE LIMITED WARRANTIES SET FORTH HEREIN, SELLER HEREBY DISCLAIMS ANY AND ALL WARRANTIES AND REPRESENTATIONS (EXPRESS OR IMPLIED, ORAL OR WRITTEN), INCLUDING ANY AND ALL IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR ANY PURPOSE WHETHER OR NOT SELLER KNOWS, OR HAS REASON TO KNOW, HAS BEEN ADVISED, OR IS OTHERWISE IN FACT AWARE OF ANY SUCH PURPOSE, WHETHER ALLEGED TO ARISE BY LAW, BY REASON OF CUSTOM OR USAGE IN THE TRADE, OR BY COURSE OF DEALING OR PERFORMANCE. Without limiting the generality of the foregoing, Seller makes no warranty regarding the ability of Products to withstand erosion or corrosion, or regarding material compatibility of elastomers in specific services, and no warranty made hereunder shall apply to Products which have been subjected to adverse storage. Customer shall be responsible for maintenance of Products following delivery. Wear or damage caused by lack of normal maintenance or by misuse of Products shall not be considered as defective workmanship or material. Sentry Equipment Corp, and its subsidiaries, reserves the right to make product design changes or improvements without notice and without imposing any obligation upon itself to install these changes or improvements on its products previously manufactured. This warranty is for the sole benefit of the original purchaser and is not transferable unless agreed to in writing by Sentry Equipment Corp. 966 Blue Ribbon Circle North, Oconomowoc, WI fax sales@sentry-equip.com

26 26 Sentry COS Updated June 2015 Sentry Equipment Corp Standard Warranty Receiving Shipments Including Loss or Damage by Transportation It is the Customer s responsibility to check for missing cartons and sign of damage to cartons. If found, Customer should note missing and/or damaged cartons on the delivery receipt and have delivery receipt signed by the representative of the transportation company. If unpacking discloses concealed damage from rough handling, the Customer should request a concealed damage inspection from the transportation company. The Sentry Customer Service Department will aid your organization in any claim proceeding for shortages or damages in shipment, but it is the Customer s responsibility to file claim with the carrier for damage or loss. Liability Limitation IN NO EVENT, WHETHER FOR BREACH OF WARRANTY OR OTHER CONTRACT BREACH, NEGLIGENCE OR OTHER TORT, OR ON ANY STRICT LIABILITY THEORY, SHALL SENTRY EQUIPMENT CORP, ITS SUBSIDIARIES OR ITS SUPPLIERS BE LIABLE FOR ANY INCIDENTAL OR CONSEQUENTIAL DAMAGES WHATSOEVER (INCLUDING BUT NOT LIMITED TO DAMAGES FOR LOSS OF PROFITS, BUSINESS INTERRUPTION, LOSS OF INFORMATION, OR OTHER PECUNIARY LOSS) ARISING OUT OF THE USE OF, OR INABILITY TO USE, THE PRODUCTS, EVEN IF SENTRY EQUIPMENT CORP OR ANY OF ITS SUBSIDIARIES HAS BEEN ADVISED OF THE POSSIBILITY OF SUCH DAMAGES. Customer Actions for Claims on Products During the Warranty Period 1. Contact the Customer Service Department, Sentry Equipment Corp, Oconomowoc, WI, at , to obtain a Return Material Authorization (RMA) number. 2. You will be sent an RMA and a Decontamination Statement that is required to be filled out and returned with the equipment. 3. The following information must appear on the outside of the package: a. RMA number marked on outside of box. b. Decontamination Statement filled out and attached to outside of box. 4. Return defective equipment FREIGHT PREPAID. Collect shipments will be refused. 5. Seller will not process warranty claims until the Customer has properly accomplished the above items. 6. Sentry Equipment Corp may accept the entire claim, a part of the claim or none of the claim if Seller s inspection of returned parts proves that the failure was for reasons other than defective material or workmanship. Important Notes Sentry Equipment Corp will not be responsible for damage incurred during the return shipment. 2. All returns subject to inspection and a minimum $ evaluation fee for any products found not to be covered by this warranty. 3. An RMA is not authorization for credit. Repair and/or replacements will be issued upon evaluation of returned goods. 4. RMA is valid for thirty (30) days from issue date A Rev Blue Ribbon Circle North, Oconomowoc, WI fax sales@sentry-equip.com

27 Sentry COS 27 Customer Assistance Sentry Equipment Corp engineers, manufactures and services equipment for sampling, monitoring, injection and analysis for worldwide customers in utility and process industries. Our equipment enables analytical and operational professionals to obtain safe, accurate and repeatable results. Recognized as a technology leader, Sentry Equipment Corp has been relied on by operational and analytical professionals since 1924, and we pride ourselves in offering exceptional technical support. Sentry Equipment Corp offers experienced Customer Service associates to assist our valued customers. Please call us at for assistance with your application. Please do not return any equipment before discussing your application issue with a Sentry representative and obtaining a Return Material Authorization (RMA) number. We consider every customer a commitment, and we thank you for your business.

28 Enclosure Openings

29 μ μ μ μ n SENTRY EQUIPMENT CORP OCONOMOWOC, WI USA

30 μ μ μ μ SENTRY EQUIPMENT CORP OCONOMOWOC, WI USA

31 SENTRY EQUIPMENT CORP OCONOMOWOC, WI USA

32 Sentry Equipment Corp is a global, employee-owned company that designs, engineers, manufactures, markets, distributes and services high-quality products for steam & water sampling, liquid & slurry sampling, solid & powder sampling, gas sampling, natural gas odorization systems, and pipeline corrosion monitoring & chemical injection equipment. Our equipment enables analytical and operational professionals in the power and process industries to obtain safe, accurate and repeatable results. It is solely the responsibility of end users, through their own analysis and testing, to select products and materials suitable for their specific application requirements, ensure they are properly installed, safely applied and properly maintained, and to limit their use to their intended purpose. Improper selection, installation, or use may result in personal injury or property damage. 966 Blue Ribbon Circle North, Oconomowoc, WI fax sales@sentry-equip.com European Authorized Rep: OBELIS Bd. General Wahis Brussels, Belgium

GASGUARDIAN Channel Controller OPERATING & INSTALLATION MANUAL

GASGUARDIAN Channel Controller OPERATING & INSTALLATION MANUAL GASGUARDIAN 2 3 2-Channel Controller OPERATING & INSTALLATION MANUAL GasGuardian 2 3 Operating and Installation Manual Table of Contents General description.... 3 Installation. 3 Locating the GasGuardian-2..

More information

GG-2 2-CHANNEL GAS DETECTION CONTROL PANEL. Installation and Operation Manual

GG-2 2-CHANNEL GAS DETECTION CONTROL PANEL. Installation and Operation Manual GG-2 2-CHANNEL GAS DETECTION CONTROL PANEL Installation and Operation Manual 2 GG-2 Warning Use this product only in the manner described in this manual. If the equipment is used in a manner not specified

More information

Installation and Operations Manual

Installation and Operations Manual Installation and Operations Manual H-IM-LLC February 2018 Part No. 25092501 Replaces H-IM-LLC (01/2014) Lead Lag Control System Table of Contents General Safety Information 2 Inspection 2 Warranty Statement

More information

Instruction Manual WARNING

Instruction Manual WARNING Controllers Instruction Manual WARNING THIS MANAUL MUST BE CAREFULLY READ BY ALL INDIVIDUALS WHO HAVE OR WILL HAVE THE RESPONSIBILITY FOR INSTALLING, USING OR SERVICING THIS PRODUCT. Like any piece of

More information

Controllers. Instruction Manual WARNING

Controllers. Instruction Manual WARNING Controllers Instruction Manual WARNING THIS MANUAL MUST BE CAREFULLY READ BY ALL INDIVIDUALS WHO HAVE OR WILL HAVE THE RESPONSIBILITY FOR INSTALLING, USING OR SERVICING THIS PRODUCT. Like any piece of

More information

COOLING WATER MIXING SKID

COOLING WATER MIXING SKID COOLING WATER MIXING SKID 35, 80, and 140 GPM Installation, Operation and Maintenance Manual SPD 8.1.13 Rev. 5 8/04 Web: www.sentry-equip.com e-mail: sales@senry-equip.com REVISION HISTORY Rev. ID Date

More information

GG-EM ENTRANCE MONITOR. Installation and Operation Manual

GG-EM ENTRANCE MONITOR. Installation and Operation Manual GG-EM ENTRANCE MONITOR Installation and Operation Manual 2 GG-EM Warning Use this product only in the manner described in this manual. If the equipment is used in a manner not specified by Calibration

More information

CLIMAGUARD Air-to-Air Outdoor Heat Exchangers INSTRUCTION MANUAL. Rev. H 2015 Pentair Equipment Protection P/N

CLIMAGUARD Air-to-Air Outdoor Heat Exchangers INSTRUCTION MANUAL. Rev. H 2015 Pentair Equipment Protection P/N CLIMAGUARD Air-to-Air Outdoor Heat Exchangers TX23, TX33, TX38, TX52 Model INSTRUCTION MANUAL Rev. H 2015 Pentair Equipment Protection P/N 10-1008-221 87976519 TABLE OF CONTENTS RECEIVING THE HEAT EXCHANGER...3

More information

Remote Relay Module (RRM)

Remote Relay Module (RRM) Remote Relay Module (RRM) Instruction Manual WARNING THIS MANUAL MUST BE CAREFULLY READ BY ALL INDIVIDUALS WHO HAVE OR WILL HAVE THE RESPONSIBILITY FOR INSTALLING, USING OR SERVICING THIS PRODUCT. Like

More information

Beacon 800 Gas Monitor Operator s Manual

Beacon 800 Gas Monitor Operator s Manual Beacon 800 Gas Monitor Operator s Manual Part Number: 71-0037RK Revision: F Released: 4/18/17 www.rkiinstruments.com Product Warranty RKI Instruments, Inc. warrants gas alarm equipment sold by us to be

More information

Series: MBC1-TC Mini Benchtop Temperature Controller

Series: MBC1-TC Mini Benchtop Temperature Controller User s Guide Series: MBC1-TC Mini Benchtop Temperature Controller Imagine Instruments LLC:: 4500 Williams Drive, Ste 212-318 :: Georgetown, TX 78633 :: p. 855.574.6243 e-mail: info@imagineinstruments.com

More information

Sentry LIQUID LEVEL CONTROLLER MODEL 120 OPERATING MANUAL.

Sentry LIQUID LEVEL CONTROLLER MODEL 120 OPERATING MANUAL. Sentry LIQUID LEVEL CONTROLLER MODEL 120 OPERATING MANUAL www.aquaticsentry.com TABLE OF CONTENTS 1. SAFETY PRECAUTIONS... 3 2. APPLICATION... 3 2.1 HIGH AND LOW LEVEL ALARM 2.2 PUMP DOWN CONTROLLER 2.3

More information

Blade Series Heat Exchanger Operation and Installation

Blade Series Heat Exchanger Operation and Installation Blade Series Heat Exchanger Operation and Installation *IMPORTANT* For safe and satisfactory operation, please read the following instructions. Keep these instructions for future reference. Some information

More information

GG-CO-EXP CARBON MONOXIDE SENSOR. Installation and Operation Manual

GG-CO-EXP CARBON MONOXIDE SENSOR. Installation and Operation Manual GG-CO-EXP CARBON MONOXIDE SENSOR Installation and Operation Manual 2 GG-CO-EXP Warning Use this product only in the manner described in this manual. If the equipment is used in a manner not specified by

More information

Installation Manual. THM-0100 Setpoint Thermostat Version THM HBX Control Systems Inc.

Installation Manual. THM-0100 Setpoint Thermostat Version THM HBX Control Systems Inc. Installation Manual M-000 Setpoint Thermostat Version.04 M-000 HBX Control Systems Inc. TABLE OF CONTENTS Introduction Index Safety symbols and Warnings Index Receipt and Inspection Description Technical

More information

Installation, Operation & Maintenance Manual

Installation, Operation & Maintenance Manual Original Instructions Installation, Operation & Maintenance Manual Sentry AutoVREL Flow Controller Sample Conditioning Automation S-SW-IOM-00457-2 02-18 Table of Contents Safety Information... 3 General

More information

Silicone Rubber Heating Tape with Adjustable Thermostat Control (HSTAT Series) Instruction Manual

Silicone Rubber Heating Tape with Adjustable Thermostat Control (HSTAT Series) Instruction Manual Silicone Rubber Heating Tape with Adjustable Thermostat Control (HSTAT Series) Instruction Manual Read and understand this material before operating or servicing these heating tapes. Failure to understand

More information

GASGUARD Cl 2. Chlorine Sensor OPERATING & INSTALLATION MANUAL

GASGUARD Cl 2. Chlorine Sensor OPERATING & INSTALLATION MANUAL GASGUARD Cl 2 Chlorine Sensor OPERATING & INSTALLATION MANUAL Operating and Installation Manual 2 Table of Contents GasGuard Cl 2 Operating and Instruction Manual General description. 4 Installation..

More information

Installation and Operation Manual. 445 Procedures Cart. Page 3. Page 5. Page 8. Page 10. Important Information Page 2. Installation.

Installation and Operation Manual. 445 Procedures Cart. Page 3. Page 5. Page 8. Page 10. Important Information Page 2. Installation. Important Information Page 2 Installation and Operation Manual 445 Procedures Cart Installation Page 3 Description Page 5 Components Overview Page 6 Controls & Indicators Page 7 Operation Page 8 HOT Operator

More information

GASGUARD H2S EXP. Hydrogen Sulfide Sensor OPERATING & INSTALLATION MANUAL

GASGUARD H2S EXP. Hydrogen Sulfide Sensor OPERATING & INSTALLATION MANUAL GASGUARD H2S EXP Hydrogen Sulfide Sensor OPERATING & INSTALLATION MANUAL Operating and Installation Manual Warning Use this product only in the manner described in this manual. If the equipment is used

More information

TMC. Installation and Operation Manual TMC. Temperature and Pressure Monitoring for Heating and Cooling Applications. Temperature Monitoring Control

TMC. Installation and Operation Manual TMC. Temperature and Pressure Monitoring for Heating and Cooling Applications. Temperature Monitoring Control Installation and Operation Manual Temperature and Pressure Monitoring for Heating and Cooling Applications Temperature Monitoring Control VALVE OPEN ALARM System= 128 o F Alarm At= 130 o F RESET /BACK

More information

Installation and Operation Manual. 445 Procedures Cart. Page 3. Page 5. Page 8. Page 10. Important Information Page 2. Installation.

Installation and Operation Manual. 445 Procedures Cart. Page 3. Page 5. Page 8. Page 10. Important Information Page 2. Installation. Important Information Page 2 Installation and Operation Manual 445 Procedures Cart Installation Page 3 Description Page 5 Components Overview Page 6 Controls & Indicators Page 7 Operation Page 8 HOT Operator

More information

Module de relais Relay Module. Manuel de l utilisateur 301RW. User Manual ERP /07

Module de relais Relay Module. Manuel de l utilisateur 301RW. User Manual ERP /07 Module de relais Relay Module Manuel de l utilisateur 301RW User Manual ERP 511397 02/07 Relay Module 301R User Manual ERP 511397 2/07 Notices and Trademarks Copyright by Honeywell International Inc.

More information

GG-NO2 NITROGEN DIOXIDE GAS SENSOR. Installation and Operation Manual

GG-NO2 NITROGEN DIOXIDE GAS SENSOR. Installation and Operation Manual GG-NO2 NITROGEN DIOXIDE GAS SENSOR Installation and Operation Manual 2 GG-NO2 Warning Use this product only in the manner described in this manual. If the equipment is used in a manner not specified by

More information

Model Gas Alarm Panel APPLICABILITY & EFFECTIVITY. This manual provides instructions for the following Sierra Monitor products:

Model Gas Alarm Panel APPLICABILITY & EFFECTIVITY. This manual provides instructions for the following Sierra Monitor products: Model 2102 Gas Alarm Panel APPLICABILITY & EFFECTIVITY This manual provides instructions for the following Sierra Monitor products: Model Description 2102-00 Alarm Panel 2 Channel 2102-01 Alarm Panel 2

More information

BX Series. Installation, Operation, Maintenance Instructions & Spare Parts

BX Series. Installation, Operation, Maintenance Instructions & Spare Parts ISO 9001 BX Series Installation, Operation, Maintenance Instructions & Spare Parts WARNING WARNING. The BX convection heater is a heavy duty baseboard designed for use in industrial environments and is

More information

Sentry LIQUID LEVEL GAUGE MODEL 200 or 200C OWNER MANUAL REV 1.7 SEPT08 PAGE 1 OF 12

Sentry LIQUID LEVEL GAUGE MODEL 200 or 200C OWNER MANUAL REV 1.7 SEPT08 PAGE 1 OF 12 PAGE 1 OF 12 TABLE OF CONTENTS PAGE 1. SAFETY PRECAUTIONS 1.1. Electrical shock 3 2. APPLICATION 3 3. INSTALLATION 3.1. Mount indoor alarm display 3.2. Mount the outdoor junction box 3.3. Install interconnecting

More information

GASGUARD SO2 Sulfur Dioxide Sensor OPERATING & INSTALLATION MANUAL

GASGUARD SO2 Sulfur Dioxide Sensor OPERATING & INSTALLATION MANUAL GASGUARD SO2 Sulfur Dioxide Sensor OPERATING & INSTALLATION MANUAL Operating and Installation Manual Warning Use this product only in the manner described in this manual. If the equipment is used in a

More information

GG-NH3 AMMONIA GAS SENSOR. Installation and Operation Manual

GG-NH3 AMMONIA GAS SENSOR. Installation and Operation Manual GG-NH3 AMMONIA GAS SENSOR Installation and Operation Manual 2 GG-NH3 Warning Use this product only in the manner described in this manual. If the equipment is used in a manner not specified by Calibration

More information

GG-LEL2 COMBUSTIBLE GAS SENSOR. Installation and Operation Manual

GG-LEL2 COMBUSTIBLE GAS SENSOR. Installation and Operation Manual GG-LEL2 COMBUSTIBLE GAS SENSOR Installation and Operation Manual 2 GG-LEL2 Warning Use this product only in the manner described in this manual. If the equipment is used in a manner not specified by Calibration

More information

LV-5 Direct Contact Low Voltage Detector and LV-5/K01 Kit including LV-PT Tester, Holster and Available Accessories

LV-5 Direct Contact Low Voltage Detector and LV-5/K01 Kit including LV-PT Tester, Holster and Available Accessories LV-5 Direct Contact Low Voltage Detector and LV-5/K01 Kit including LV-PT Tester, Holster and Available Accessories Operating & Instruction Manual 1475 Lakeside Drive Waukegan, Illinois 60085 U.S.A. 847.473.4980

More information

GASGUARD NH 3. Ammonia Sensor OPERATING & INSTALLATION MANUAL

GASGUARD NH 3. Ammonia Sensor OPERATING & INSTALLATION MANUAL GASGUARD NH 3 Ammonia Sensor OPERATING & INSTALLATION MANUAL Operating and Installation Manual 2 Table of Contents GasGuard NH 3 Operating and Instruction Manual General description. 4 Installation.. 4

More information

GasScanner 8C. Eight Channel Monitor. Operator s Manual. MINT-0281-XX Rev. A 01/29/08

GasScanner 8C. Eight Channel Monitor. Operator s Manual. MINT-0281-XX Rev. A 01/29/08 GasScanner 8C Eight Channel Monitor Operator s Manual MINT-0281-XX Rev. A 01/29/08 Product Warranty Matheson Tri-Gas, Inc., warrants gas alarm equipment sold by us to be free from defects in materials,

More information

GASGUARD O 2. Oxygen Sensor OPERATING & INSTALLATION MANUAL

GASGUARD O 2. Oxygen Sensor OPERATING & INSTALLATION MANUAL GASGUARD O 2 Oxygen Sensor OPERATING & INSTALLATION MANUAL 2 Table of Contents General description. 4 Installation.. 4 Locating the sensor. 4 Installation guidelines 5 Wiring.. 6 Operation.. 7 Start-up...

More information

Compact Equipment Warranty Policy

Compact Equipment Warranty Policy 006 07.15 Compact Equipment Warranty Policy 07/2015 Table of Contents Preface Page 1 Distributor s Warranty Responsibilities Page 1 Standard Warranties Page 2 Warranty Coverage Page 2 Warranty Limitations

More information

WATLOW IND. FIREBAR Flange Heater Installation & Maintenance Manual I&M NUMBER: Page: 1 Date: 6/11/2008 Rev: 2.00

WATLOW IND. FIREBAR Flange Heater Installation & Maintenance Manual I&M NUMBER: Page: 1 Date: 6/11/2008 Rev: 2.00 I&M NUMBER: 316-42-2-1 Page: 1 Pre Installation Check to make sure that heater received is the same as that ordered. Elements may come in contact with each other during shipment. Minor adjustments to elements

More information

(HSTAT Series) Instruction Manual

(HSTAT Series) Instruction Manual Silicone Rubber Heating Tapes with Adjustable Thermostat Control (HSTAT Series) Instruction Manual Read and understand this material before operating or servicing these heating tapes. Failure to understand

More information

WATLOW IND. WATROD Flange Heater Installation & Maintenance Manual I&M NUMBER: Page: 1 Date:6/11/2008 Rev: 2.00

WATLOW IND. WATROD Flange Heater Installation & Maintenance Manual I&M NUMBER: Page: 1 Date:6/11/2008 Rev: 2.00 I&M NUMBER: 316-42-8-1 Page: 1 _ Pre Installation Check to make sure that heater received is the same as that ordered. Elements may come in contact with each other during shipment. Minor adjustments to

More information

Spa Control System OWNER S MANUAL

Spa Control System OWNER S MANUAL LIMITED WARRANTY ONE YEAR LIMITED WARRANTY: UNITED SPAS, INC. warrants, to the original purchaser, the Spa Equipment against defects in materials or workmanship for a period of one year from date of purchase.

More information

A10. Installation & Operation Manual. Flow Switch. For the latest version of this manual, visit kteksolidslevel.com.

A10. Installation & Operation Manual. Flow Switch. For the latest version of this manual, visit kteksolidslevel.com. Installation & Operation Manual A10 For the latest version of this manual, visit kteksolidslevel.com. K-TEK Solids Level l 6100 West by Northwest #140 l Tel: +(1) 713.462.7665 l Toll Free 800.245.7056

More information

GG-H2S HYDROGEN SULFIDE GAS SENSOR. Installation and Operation Manual

GG-H2S HYDROGEN SULFIDE GAS SENSOR. Installation and Operation Manual GG-H2S HYDROGEN SULFIDE GAS SENSOR Installation and Operation Manual 2 GG-H2S Warning Use this product only in the manner described in this manual. If the equipment is used in a manner not specified by

More information

! The Caution Symbol (exclamation point) alerts you to a "CAUTION", a safety or

! The Caution Symbol (exclamation point) alerts you to a CAUTION, a safety or I&M NUMBER: 316-42-10-1 Page: 1 Pre Installation Check to make sure that heater received is the same as that ordered. Elements may come in contact with each other during shipment. Minor adjustments to

More information

TL-105 COLUMN HEATER

TL-105 COLUMN HEATER TL-105 COLUMN HEATER INSTRUCTION MANUAL Timberline Instruments Inc. 1880 S. Flatiron Ct., Unit I Boulder, Colorado 80301 Ph: (303) 440-8779 Fx: (303) 440-8786 info@timberlineinstruments.com www.timberlineinstruments.com

More information

ModSync Sequencing System Installation & Operation Manual. For use with Fulton Steam Boilers.

ModSync Sequencing System Installation & Operation Manual. For use with Fulton Steam Boilers. ModSync Sequencing System Installation & Operation Manual For use with Fulton Steam Boilers. Revision 3.0 8/21/2008 - 2 - Table of Contents Introduction Page 4 Features Page 4 Sequence of Operation Page

More information

MaxLite LED Round Pendant Highbay Series

MaxLite LED Round Pendant Highbay Series General Safety Information To reduce the risk of death, personal injury or property damage from fire, electric shock, falling parts, cuts/abrasions, and other hazards read all warnings and instructions

More information

Beacon 200 Gas Monitor Operator s Manual. Part Number: RK Released: 6/6/08

Beacon 200 Gas Monitor Operator s Manual. Part Number: RK Released: 6/6/08 Beacon 200 Gas Monitor Operator s Manual Part Number: 71-2102RK Released: 6/6/08 Table of Contents Chapter 1: Introduction.................................................3 Overview.............................................................3

More information

GASGUARD H 2-1% Hydrogen Sensor OPERATING & INSTALLATION MANUAL

GASGUARD H 2-1% Hydrogen Sensor OPERATING & INSTALLATION MANUAL GASGUARD H 2-1% Hydrogen Sensor OPERATING & INSTALLATION MANUAL 2 Table of Contents GasGuard H 2-1% General description. 4 Installation.. 4 Locating the sensor. 4 Installation guidelines 5 Wiring.. 6 Operation..

More information

Henny Penny Island Warmer Model HMI-103 Model HMI-105 TECHNICAL MANUAL

Henny Penny Island Warmer Model HMI-103 Model HMI-105 TECHNICAL MANUAL Henny Penny Island Warmer Model HMI-103 Model HMI-105 TECHNICAL MANUAL THIS PAGE INTENTIONALLY LEFT BLANK. Section TABLE OF CONTENTS Page Section 1. TROUBLESHOOTING... 1-1 1-1. Introduction... 1-1 1-2.

More information

Ion Genesis II Pump Controller Digital Level Control with Pump Alternation and High Water Alarm

Ion Genesis II Pump Controller Digital Level Control with Pump Alternation and High Water Alarm Page 1 of 8 General Overview Thank you for purchasing an Ion Genesis controller. Take the time to read the instructions carefully before using this appliance. We strongly recommend that you keep this instruction

More information

HTLS. Heat Tracing Line Sensing Control Panel

HTLS. Heat Tracing Line Sensing Control Panel PDS Heat Tracing Line Sensing Control Panel Series 8000/12000/24000 Features: Color Touchscreen Operator Panel Eight/Twelve/Twenty Four Points of Independent Control NEMA 4/12 Enclosure High/Low Temperature

More information

HotPoly Tote Tank / IBC Heaters (TTH Series)

HotPoly Tote Tank / IBC Heaters (TTH Series) HotPoly Tote Tank / IBC Heaters (TTH Series) Instruction Manual Read and understand this material before operating or servicing these heaters. Failure to understand how to safely operate these heaters

More information

GG-NH3-2% AMMONIA GAS SENSOR. Installation and Operation Manual

GG-NH3-2% AMMONIA GAS SENSOR. Installation and Operation Manual GG-NH3-2% AMMONIA GAS SENSOR Installation and Operation Manual 2 GG-NH3-2% Warning Use this product only in the manner described in this manual. If the equipment is used in a manner not specified by Calibration

More information

AUTOMATION. Operator s Manual RST Series Web Enabled Input Module. Rev. A2, 1/12

AUTOMATION. Operator s Manual RST Series Web Enabled Input Module. Rev. A2, 1/12 AUTOMATION P R O D U C T S GROUP, INC. Operator s Manual RST-5000 Series Web Enabled Input Module Rev. A2, 1/12 Tel: 1/888/525-7300 Fax: 1/435/753-7490 www.apgsensors.com E-mail: sales@apgsensors.com RST-5000

More information

GASGUARD VENT LINE3. Ammonia Sensor OPERATING & INSTALLATION MANUAL

GASGUARD VENT LINE3. Ammonia Sensor OPERATING & INSTALLATION MANUAL GASGUARD VENT LINE3 Ammonia Sensor OPERATING & INSTALLATION MANUAL Operating and Installation Manual Warning Use this product only in the manner described in this manual. If the equipment is used in a

More information

6502B RF Distribution

6502B RF Distribution 6502B RF Distribution User Guide Revision A - March 2004 Part Number 15247-201 Table of Contents Table of Contents About this User Guide Purpose...................................................... 1

More information

User s Manual and Warranty Information for Counterweighted Chain Drive ThyssenKrupp Access

User s Manual and Warranty Information for Counterweighted Chain Drive ThyssenKrupp Access II User s Manual and Warranty Information for Counterweighted Chain Drive ThyssenKrupp Access Part #2139703 Rev. G II Table of Contents Introduction...3 Elevator Overview...4 Description of Features...5-7

More information

Intrinsically Safe Transmitters, Option Code 2n

Intrinsically Safe Transmitters, Option Code 2n Installation Manual for Intrinsically Safe Transmitters, Option Code 2n ATEX PERSONAL INJURY WARNING DO NOT USE these products as safety or emergency stop devices or in any other application where failure

More information

WATLOW ELECTRIC MFG CO. FIREBAR Screw Plug Installation & Maintenance Manual I&M NUMBER: Page: 1 Date: 11/25/2013 Rev: 4.

WATLOW ELECTRIC MFG CO. FIREBAR Screw Plug Installation & Maintenance Manual I&M NUMBER: Page: 1 Date: 11/25/2013 Rev: 4. I&M NUMBER: 316-42-3-1 Page: 1 Pre Installation Check to make sure that heater received is the same as that ordered. Elements may come in contact with each other during shipment. Minor adjustments to elements

More information

WATLOW IND. WATROD Modular Duct Heater Installation & Maintenance Manual I&M NUMBER: Page: 1 Date:6/11/2008 Rev: 2

WATLOW IND. WATROD Modular Duct Heater Installation & Maintenance Manual I&M NUMBER: Page: 1 Date:6/11/2008 Rev: 2 I&M NUMBER: 316-42-15-1 Page: 1 _ Pre Installation Check to make sure that heater received is the same as that ordered. Elements may come in contact with each other during shipment. Minor adjustments to

More information

Wrap-Around TOTE Tank / IBC Heaters (TOTE and TOT Series)

Wrap-Around TOTE Tank / IBC Heaters (TOTE and TOT Series) Wrap-Around TOTE Tank / IBC Heaters (TOTE and TOT Series) Instruction Manual Read and understand this material before operating or servicing these heating tapes. Failure to understand how to safely operate

More information

GASGUARD NH 3 2% Ammonia Sensor OPERATING & INSTALLATION MANUAL

GASGUARD NH 3 2% Ammonia Sensor OPERATING & INSTALLATION MANUAL GASGUARD NH 3 2% Ammonia Sensor OPERATING & INSTALLATION MANUAL 2 Table of Contents GasGuard NH 3 2% General description. 4 Installation.. 4 Locating the sensor. 4 Installation guidelines 5 Wiring.. 6

More information

Installation, Operation & Service Manual

Installation, Operation & Service Manual Installation, Operation & Service Manual WARNING Improper installation, adjustment, alteration, service or maintenance can result in death, injury or property damage. Read the Installation, Operation and

More information

Mark 25 Ultrapure Water Conductivity Analyzer

Mark 25 Ultrapure Water Conductivity Analyzer Martek Instruments, Inc. Mark 25 Ultrapure Water Conductivity Analyzer Instruction Manual WARRANTY POLICY Unless otherwise stated, MARTEK INSTRUMENTS, INC. warrants this equipment to be free from defects

More information

GG-VL-R REFRIGERANT VENT LINE SENSOR. Installation and Operation Manual

GG-VL-R REFRIGERANT VENT LINE SENSOR. Installation and Operation Manual GG-VL-R REFRIGERANT VENT LINE SENSOR Installation and Operation Manual 2 GG-VL-R Warning Use this product only in the manner described in this manual. If the equipment is used in a manner not specified

More information

GASGUARD LEL 3 OPERATING & INSTALLATION MANUAL

GASGUARD LEL 3 OPERATING & INSTALLATION MANUAL GASGUARD LEL 3 Ammonia Sensor OPERATING & INSTALLATION MANUAL GasGuard LEL 3 Operating and Installation Manual 2 Table of Contents 3GasGuard LEL Operating and Instruction Manual General description. 4

More information

OPERATION & INSTALLATION MANUAL

OPERATION & INSTALLATION MANUAL OPERATION & INSTALLATION MANUAL Model: SIO 14 & SIO 18 Electric Tankless Hot Water Generators Table of Contents SAFETY INFORMATION... 1 INTRODUCTION... 2 Unit Operation:... 2 Unit Freezing:... 3 Maintenance:...

More information

Sentry LIQUID LEVEL ALARM MODEL 100 OPERATING MANUAL.

Sentry LIQUID LEVEL ALARM MODEL 100 OPERATING MANUAL. Sentry LIQUID LEVEL ALARM MODEL 100 OPERATING MANUAL www.aquaticsentry.com TABLE OF CONTENTS 1. SAFETY PRECAUTIONS... 3 2. APPLICATION... 3 2.1 HIGH Liquid Level Alarm 2.2 LOW Liquid Level Alarm 3. INSTALLATION...

More information

PM-200 Power Module Operating Instructions

PM-200 Power Module Operating Instructions PM-200 Power Module Operating Instructions Terahertz Technologies Inc. 169 Clear Road, Oriskany, NY 13424 (315) 736-3642 FAX (315) 736-4078 E-mail tti@borg.com Contents Introduction... 1 Specifications...

More information

ISOLOK series SAA Sampler

ISOLOK series SAA Sampler Features Two different models for sampling fixed volumes of 3 cc or 8 cc per sample cycle Can handle line pressures from vacuum up to 200 psi (13.8 bar) in most applications Rugged 316 stainless steel

More information

ion Genesis Pump Controller

ion Genesis Pump Controller High Water Alarm Document No.: IONG_OM Page 1 of 7 Table of Contents Safety Precautions.......................... 1 General Overview.......................... 1 Installation.................................2

More information

Henny Penny Island Warmer Model HMI-103 Model HMI-105 TECHNICAL MANUAL

Henny Penny Island Warmer Model HMI-103 Model HMI-105 TECHNICAL MANUAL Henny Penny Island Warmer Model HMI-103 Model HMI-105 TECHNICAL MANUAL Section TABLE OF CONTENTS Page Section 1. TROUBLESHOOTING... 1-1 1-1. Introduction... 1-1 1-2. Safety... 1-1 1-3. Troubleshooting...

More information

Air Pump Up to 800 gallons

Air Pump Up to 800 gallons Air Pump Up to 800 gallons REMINDER CALL 1-888-755-6750 BEFORE RETURNING TO STORE. PACKAGE CONTENTS ITEM #PBPAPK40W Questions, problems, missing parts? Before returning to your retailer, call our customer

More information

WATLOW IND. WATROD Circulation Heater Installation & Maintenance Manual I&M NUMBER: Page: 1 Date: 6/11/2008 Rev: 2.00

WATLOW IND. WATROD Circulation Heater Installation & Maintenance Manual I&M NUMBER: Page: 1 Date: 6/11/2008 Rev: 2.00 I&M NUMBER: 316-42-5-1 Page: 1 Pre Installation Check to make sure that heater received is the same as that ordered. Watlow heaters are built to comply with UL and CSA dielectric requirements, it may be

More information

RATIO:GUARD Model E-1S EC Monitor

RATIO:GUARD Model E-1S EC Monitor EC probe Probe tees Probe retention clips Temperature probe E-1S monitor box UNPACKING Please open and inspect your package upon receipt. Your package was packed with great care and all the necessary packing

More information

SPA BLOWER OWNER'S MANUAL XXXX, XXXX, XXXX, XXXX, XXXX, XXXX fax

SPA BLOWER OWNER'S MANUAL XXXX, XXXX, XXXX, XXXX, XXXX, XXXX fax SPA BLOWER OWNER'S MANUAL 80015-XXXX, 80016-XXXX, 80017-XXXX, 80018-XXXX, 80019-XXXX, 80020-XXXX fax 888.610.3839 2015 323300-015 6/15 THIS PAGE INTENTIONALLY LEFT BLANK. 2 Operating Instructions and Parts

More information

10810 W. LITTLE YORK RD., STE. 130 HOUSTON TX VOICE (713) FAX (713) web: IMPORTANT!!!

10810 W. LITTLE YORK RD., STE. 130 HOUSTON TX VOICE (713) FAX (713) web:   IMPORTANT!!! 10810 W. LITTLE YORK RD., STE. 130 HOUSTON TX 77041-4051 VOICE (713) 973-6905 FAX (713) 973-9352 web: www.twrlighting.com IMPORTANT!!! PLEASE TAKE THE TIME TO FILL OUT THIS FORM COMPLETELY. FILE IT IN

More information

ELECTRONIC FAUCETS INSTRUCTION MANUAL. Toll Free: (800) Haul Road Wayne, NJ krowne.

ELECTRONIC FAUCETS INSTRUCTION MANUAL. Toll Free: (800) Haul Road Wayne, NJ krowne. ELECTRONIC FAUCETS INSTRUCTION MANUAL 100 Haul Road Wayne, NJ 07470 Toll Free: (800) 631-0442 customerservice@krowne.com krowne.com Model 16-196 1 3 2 4 6 5 7 1. 2. 3. 4. 5. 6. 7. Aerator Sensor with Cable

More information

MaxLite LED MICRO-T PANEL

MaxLite LED MICRO-T PANEL ` Installation Instructions General Safety Information To reduce the risk of death, personal injury or property damage from fire, electric shock, falling parts, cuts/abrasions, and other hazards read all

More information

MN908 No. 282C Replaces 282B LB7013

MN908 No. 282C Replaces 282B LB7013 No. 282C Replaces 282B LB7013 Instruction Manual For Baldor Dust Control Units Models DC7, DC7 3, DC8, DC8 3, DC10, DC10 3, DC12, DC12 3 and DC14 3. For use on Grinders mounted to GA16 and GA20 pedestals

More information

MODEL FPT-130 SINGLE POINT FREEZE PROTECTION HEAT TRACE CONTROL

MODEL FPT-130 SINGLE POINT FREEZE PROTECTION HEAT TRACE CONTROL TRACON MODEL FPT-130 SINGLE POINT FREEZE PROTECTION HEAT TRACE CONTROL TABLE OF CONTENTS FPT 130 Overview... 2 Installation... 3 Power Source and Load Connections... 4 Temperature Sensor... 5 External

More information

Ion Endeavor Pump Controller Digital Level Control with Pump Alternation and High Water Alarm

Ion Endeavor Pump Controller Digital Level Control with Pump Alternation and High Water Alarm Ion Endeavor Controller Digital Level Control with Alternation Page 1 of 8 General Overview The Ion Endeavor is a pump controller that senses a water level of up to 72", has a configurable water level/pump

More information

PodView Level Indicator LI40 Series Manual

PodView Level Indicator LI40 Series Manual PodView Level Indicator LI40 Series Manual Flowline Inc. 10500 Humbolt Street Los Alamitos, CA 90720 Tel: (562) 598 3015 Fax: (562) 431 8507 www.flowline.com Rev A MN204260 1 of 16 INTRODUCTION / TABLE

More information

OCH-SS series Direct Wired Units Indoor * and Outdoor Comfort Heaters

OCH-SS series Direct Wired Units Indoor * and Outdoor Comfort Heaters 1200 North Main Street Fostoria, OH 44830 Phone: 800-495-4525 Fax: 419-435-0842 A DIVISION OF www.fostoriaindustries.com OCH-SS series Direct Wired Units Indoor * and Outdoor Comfort Heaters *EXCLUDING

More information

Installation & Operation Manual

Installation & Operation Manual Installation & Operation Manual Radiant Thermostat 519 519_D 06/16 Zoning Replaces: 03/13 Introduction The Radiant Thermostat 519 accurately controls the room and/or floor temperature for a hydronic heating

More information

RAM3 Remote Alarm Module

RAM3 Remote Alarm Module RAM3 Remote Alarm Module INSTRUCTIONS Installation and Operation of the AMC-RAM3 with AMC Monitors IMPORTANT: Please read this installation and operating instructions completely and carefully before starting.

More information

EQUIPMENT Operation Manual

EQUIPMENT Operation Manual EQUIPMENT Operation Manual Loctite SD15 Heat Controller Part Number 1667508 1 TABLE OF CONTENTS 1 Please Observe The Following... 3 1.1 Emphasized Sections... 3 1.2 For Your Safety... 3 1.3 Unpacking and

More information

XT Series Thermostats

XT Series Thermostats ISO 9001 Explosion-Proof & Moisture Resistant XT Series Thermostats Installation, Operation, & Maintenance Instructions XTWA Thermostat XTWL Thermostat XTB Thermostat Part No.MI133.Rev.2.00 Date of Issue:

More information

LV-S-5. Instruction Manual DIRECT CONTACT STRAY VOLTAGE DETECTOR GENERAL DESCRIPTION HOW IT WORKS

LV-S-5. Instruction Manual DIRECT CONTACT STRAY VOLTAGE DETECTOR GENERAL DESCRIPTION HOW IT WORKS LV-S-5 DIRECT CONTACT STRAY VOLTAGE DETECTOR Instruction Manual GENERAL DESCRIPTION The LV-S-5 is a hand held stray voltage detector for testing exposed metallic surfaces and conductors for the presence

More information

Installation & Operation Manual

Installation & Operation Manual Humidity & Temperature Sensor 086 Accessories 086_D 08/12 Replaces: New Installation & Operation Manual Introduction The Humidity & Temperature Sensor 086 can be connected to select tekmar thermostats

More information

MODEL GPT-130 SINGLE POINT HEAT TRACE CONTROL THERMOSTAT

MODEL GPT-130 SINGLE POINT HEAT TRACE CONTROL THERMOSTAT TRACON MODEL GPT-130 SINGLE POINT HEAT TRACE CONTROL THERMOSTAT TABLE OF CONTENTS GPT 130 Overview... 2 Installation... 3 Power Source and Load Connection... 4 Temperature Sensor Installation... 5 Panel

More information

USER S OPERATING AND INSTRUCTION MANUAL

USER S OPERATING AND INSTRUCTION MANUAL Grand Rapids, Michigan, U.S.A. 49504-5298 USER S OPERATING AND INSTRUCTION MANUAL MODEL 1508-NLG SELF-ACTUATING TRAY LIDDER 1508S20000CV2 INDEX SAFETY INSTRUCTIONS... 1508S20002 SET UP... 1508S20003 OPERATING

More information

Field Top Mounted Condensing Unit

Field Top Mounted Condensing Unit Field Top Mounted Condensing Unit INSTALLATION & OPERATIONS MANUAL RF/RF-N, RM/RM-N IMPORTANT Keep in store for future reference! REVISION HISTORY VERSION 1.0 New manual This manual is specific to the

More information

TrueClean ToteCleaner

TrueClean ToteCleaner TrueClean ToteCleaner Installation, Operation, and Maintenance Manual www.trueclean.us TrueClean Check Valve Table of Contents Introduction.... 1 About... 1 Warranty... 2 Safety... 3 Important Safety Information....

More information

AG Series Electric Heaters

AG Series Electric Heaters AG Series Electric Heaters Auxiliary Electric Heat Installation, Operation & Maintenance Instructions 97B0005N04 Table of Contents Overview 3 Vertical Upflow or Downflow Installation - Internal 3 Vertical

More information

Retrofit Instructions for Tower Dryer Moisture Sampler Blower Kit

Retrofit Instructions for Tower Dryer Moisture Sampler Blower Kit Retrofit Instructions for Tower Dryer Moisture Sampler Blower Kit Installation Manual PNEG-1982 Date: 03-26-15 PNEG-1982 2 PNEG-1982 Retrofit Instructions for Tower Dryer Moisture Sampler Blower Kit Table

More information

GG-R REFRIGERANT SENSOR. Installation and Operation Manual

GG-R REFRIGERANT SENSOR. Installation and Operation Manual GG-R REFRIGERANT SENSOR Installation and Operation Manual 2 GG-R Warning Use this product only in the manner described in this manual. If the equipment is used in a manner not specified by Calibration

More information

Power Conditioners User s Manual. For use with Standard Power Conditioners, Medical Power Conditioners and Ground Guard Power Conditioners

Power Conditioners User s Manual. For use with Standard Power Conditioners, Medical Power Conditioners and Ground Guard Power Conditioners Power Conditioners User s Manual For use with Standard Power Conditioners, Medical Power Conditioners and Ground Guard Power Conditioners TABLE OF CONTENTS 1.0 - Introduction... 4 2.0 - Safety Instructions...

More information

WATLOW ELECTRIC MFG CO. WATROD Duct Heater (module only) I& M Manual I&M NUMBER: Page: 1 Date: 11/25/2013 Rev: 4.00

WATLOW ELECTRIC MFG CO. WATROD Duct Heater (module only) I& M Manual I&M NUMBER: Page: 1 Date: 11/25/2013 Rev: 4.00 I&M NUMBER: 316-42-16-1 Page: 1 Pre Installation Check to make sure that heater received is the same as that ordered. Elements may come in contact with each other during shipment. Minor adjustments to

More information

Owner s Manual PCS PRO

Owner s Manual PCS PRO Owner s Manual PCS PRO Designed and Manufactured Exclusively by Ozotech, Inc. M43-31017 REV A Table of Contents 1.0 Limited Warranty 3 2.0 Service Returns 4 3.0 Caution 5 4.0 Theory of Operation 5 4.1

More information

Miniature Thermoelectric Cooler Controller MTCC-01 user's guide

Miniature Thermoelectric Cooler Controller MTCC-01 user's guide Miniature Thermoelectric Cooler Controller MTCC-01 user's guide 1 Table of Contents 1. Warranty... 3 2. MTCC-01-xxx overview... 4 2.1. MTCC-01-xxx specification... 4 2.2. TE cooled detector temperature

More information