Safety and Emergency procedures in LAMP

Size: px
Start display at page:

Download "Safety and Emergency procedures in LAMP"

Transcription

1 Safety and Emergency procedures in LAMP A supplement to the SOP for Laboratory for Advanced Materials Processing JM Patterson Bldg., Room 2225 Revision 1.3 March 31, 2004 Prepared by Laurent Henn-Lecordier, Institute for Systems Research Safety and Emergency procedure in LAMP Page 1 of 17

2 1 SAFETY SYSTEMS To optimize the safety in the Laboratory for Advanced Materials Processing LAMP - several safety sensors have been installed in order to generate a single alarm response in case of an emergency. Emergency situations are generally related to an accidental release of process gases. The sensors and alarm devices were initially installed in 2001 for the detection of pyrophoric silane. However the silane cylinder was removed from the lab in June 2003 and the MDA System 16 (for silane detection) is now shut down. The alarm system is still operational and activated and receives the signals from the WF6 detection sensor, smoke detectors, photohelic gauges for vent line monitoring and EPO button. This report will list the different sensors and the procedures to follow in case of an alarm within this laboratory. A section regarding the maintenance of this alarm system is also included. 1.1 Alarm Systems CSS E0883 MDA alarm unit The core of the alarm system is the MDA alarm unit located above the System 16 gas monitor. This system generates 24 V signals to the different sensor relays (MDA silane gas detector, smoke detectors, photohelic gauges, EPO button at the entrance of the lab, silane pressure switch and silane excess flow switch). If an alarm condition is detected by one of these sensors, their internal Normally Open relays will be closed and the 24 V signal will be returned to the MDA alarm interface. An appropriate response will be triggered in function of the origin (sensor) of the alarm. In all alarm cases (Level 1 or 2), the gas interlock system will be shutdown (Alarm level 1). All the pneumatic valves located on the gas cylinder outlets or downstream to the regulators will be automatically shut down. If the presence of smoke is detected in the cleanroom or pre-cleanroom, or if the EPO button at the entrance of the lab is pushed (Alarm Level 2), a visual and audible alarm (remote alarm enunciators 1 and 2) will be triggered in both rooms, in addition to the gas interlock shutdown. The MDA alarm unit is also connected to a 24 hour monitoring station that will notify UM Police Department and trigger an additional enunciator in the hallway (see next page). The building alarm will also be activated. Notice: the monitoring station and building alarm are activated via signal from MDA interface to ESL fire panel (see section ESL fire panel for additional information) If the MDA alarm unit stops functioning (power shortage, breakdown ), a Normally Open relay will open and shutdown the gas interlock system. Safety and Emergency procedure in LAMP Page 2 of 17

3 In case of a power shortage, the MDA alarm unit and the System 16 gas detector are backed up by a Stabiline SL 1500 uninterruptible power supply whose 1440 max VA will provide power to these systems for 5 to 10 minutes Remote alarm enunciators 1 and 2 Two visual (red flashlight) and aural remote alarms are connected to the MDA alarm interface and will respond to an Alarm level 2. One is located next to the MDA alarm interface above the System 16, the other on top of the ULVAC system. The 24V power supply to these units is provided by the MDA alarm unit which means that in the case of a building power shortage, these alarms will continue to function for a few minutes via the UPS. These alarms can only be shut down by deactivating the internal relays of the MDA alarm unit DSC power hour supervisory unit A 24-hour supervisory unit is connected to the Police Department. UMPD will immediately notify the fire department. This unit reports any alarm Level 2 from the MDA alarm unit. This system must be armed prior to utilization. A blue flash-light is installed in the hallway on top of the LAMP main entrance door. In addition to the visual signal, a prerecorded warning message will be annunciated. This speaker/light warning system is connected to the DSC supervisory unit. To stop the siren or rearm the system, enter on the keypad: The UMPD phone number to call or to cancel an accidental alarm is ext. # The manual related to the operation of this unit is included in this report. UMD contact: Craig Chucker, Bldg Security Systems J.M. Patterson building alarm In case of Alarm Level 2, the JM Patterson main alarm system will be also triggered. A 4 th zone was added to the ESL fire panel and connected to the MDA alarm unit. In case of alarm, the zone 4 will be energized and a relay connected to the building alarm circuitry will be activated. Only UMD work control personal should handle this circuit. UMD Contact: Jim Robinson (x 53971) 1.2 Safety Sensors and Activators MDA / Zellweger Analytics System 16 gas detector ALARM LEVEL 2 Notice: the MDA System 16 has been shut down since June The following information is therefore facultative. Safety and Emergency procedure in LAMP Page 3 of 17

4 This system is currently configured for the detection of silane and analyses sequentially the gas samples in four different locations (silane gas cabinet, ATM gas scrubber, Ulvac gas delivery system on the roof of the Ulvac cluster tool, cleanroom). If a silane concentration above 5 ppm (TLV) is detected at any of these locations, it will activate the MDA alarm interface (Alarm Level 2) The front panel of the System 16 can be read from the hallway through the glass window. Thus, in case of an alarm, the printed emergency report can be read without entering the lab. This system is backed up by a UPS station. A set of 3 green and 3 red LED s has been wired on the back panel of the System 16 gas detector. These LED s report the following fault detection: Maintenance Warning: the instrument needs attention but the instrument is still able to monitor and operate as programmed. (e.g. chemcassette needs to be changed) Loss of Coverage: the monitoring capability is reduced, but not completely absent (failure of one or more points module) Loss of Monitoring: it indicates a complete absence of monitoring. The bottom line is that these LED s are easily visible either from the precleanroom or from the cleanroom. The operators of the ULVAC should systematically make sure that all three green LED s are ON before processing with silane. (for more details concerning the System 16, cf. Section 3 of this memo) Three-zone ESL Fire Alarm Control unit ALARM LEVEL 2 3 smoke detectors are located near the silane gas cabinet, the ATM scrubber and the ULVAC CVD cluster tool. They are connected to the three-zone ESL Fire Alarm control unit, located next to the System 16 and visible from the middle door in the hallway. If the presence of smoke is detected at any of these three locations: An internal audible alarm (from the ESL fire control unit) will go off. A double throw relay will activate the MDA alarm interface (Alarm Level 2). This system is backed up by a 24 V battery. The building alarm is tied to the ESL Fire panel. In order to have the building alarm activated in case of a level 2 alarm, a fourth zone has been installed by Physical Plant. This zone is an input from the MDA alarm interface. If an alarm Level 2 is generated, the A1 N.O.-relay from the MDA alarm interface will close. A 24 V signal is consequently received to the Zone 4 terminal triggering the Fire alarm and the building alarm. Safety and Emergency procedure in LAMP Page 4 of 17

5 1.2.3 Red Mushroom push-button ALARM LEVEL 2 Located at the entrance of the lab (on the left side of the door), it can be manually activated. It is connected to a 120 V relay located in the mechanical room. The closure of this relay by pushing on the button will return a 24V signal to the MDA alarm interface (Alarm level 2). An internal flash light and an aural alarm are also activated (they are located above the entrance door in the pre-cleanroom. The 120V signal is provided by the building power supply Dwyer hi/low-setpoints photohelic gages ALARM LEVEL 1 3 photohelic gauges monitor the proper functioning of the duct exhaust system at the three following locations: silane gas cabinet exhaust, mechanical pumps and scrubber exhaust, ULVAC cluster tool exhaust. If the pressure in the exhaust is below 0.2 WG (normal pressure is between 0.5 and 1 ), it will activate the MDA alarm unit that will shutdown the gas interlock panel (the remote alarm enunciators will not be activated). 1.3 Zellweger Analytics System 16 Gas Detector Notice: the MDA System 16 has been shut down since June The following information is therefore facultative The System 16 is equipped with one 4-point module analyzer that can detect one gas at 4 different locations. This analyzer has a chemcassette formulated to react with silane: the target gas in the sample reacts with the tape, the reaction produces a stain with a density proportional to the gas concentration, the stain is measured is optically measured. This chemcassette and optical sensor must be maintained monthly (cf. Section 3-2). This system can be programmed either in Sequential Mode (System 16 monitors the sample from only one point at a time) or in Parallel Mode (the System 16 monitors simultaneously samples from the 4 points) System configuration This section is a descriptive list of the current configuration and the different steps followed by the System 16 in case of silane detection. For more details, cf. System 16 Technical Handbook. System 16 is configured in Sequential Mode. Alarm levels for the point group (i.e. 4 points): TLV1 = 5 ppm early detection of silane TLV2 = 10 ppm for critical levels which may require evacuation Safety and Emergency procedure in LAMP Page 5 of 17

6 TLV: Threshold Limit Value Based on the alarm level of the point in the point group with the lowest programmed alarm level (in our case, each point has the same alarm level) In our case, all alarms are triggered at TLV1 and will require immediate evacuation. - If the gas concentration for one point exceeds 5 ppm: 1) The appropriate alarm LEDs lights (Level 2 and/or 1) 2) The point Relay contact activates and triggers the LAMP alarm Level 2 via the MDA alarm interface. 3) The printer module prints the alarm information with time, alarm levels and concentrations MDA System 16 Scheduled Maintenance Changing the Chemcassette (i.e. every month): Clean with alcohol everywhere the tape is in contact (especially the tape guides) Use the optic verification card to check the optics system (see Section 10 of Syst. 16 Handbook). Balance the flow (see Section 9) If problem with the flow balance, check the Acid Scrubber filter. MDA Service Mode (not in the manual) Password: 0902 Select Subfunction: Real CC Transducer 140 < S < 205 (operating range) Adjusting S will probably change the flow meter value (e.g. from 42 to 45 mm). This does not matter. The transducer data S is the most accurate and is the one that will be processed by the computer (not the flow meter value) Bi-annual Maintenance Optics Calibration Enter Service Submenu, and subfunction 9 (Read Raw Optics) Raw reading should be: /- 200 Lower limit 2000 (a lower value will give an Instrument failure) Safety and Emergency procedure in LAMP Page 6 of 17

7 To clean the optics, remove the 2 screws of the optics block, pull the pipe straight back at the rear of the panel, remove the black plastic block with the LED s and clean with alcohol. When dirty, you can see a white stuff from the band. Also in case of false alarm, the optics should be always checked) Filter Maintenance (see section 12 of System 16 Technical Handbook) Acid Scrubber Filter (manufactured by Bolston, PN ) If ¾ or more of the acid scrubber turns blue, or the vacuum increases, replace it. Replacement frequency: 6 months. Primary Flow Module Filter (Replacement frequency 6-12 months) Safety and Emergency procedure in LAMP Page 7 of 17

8 Alarm System Maintenance and Check-Up Only the System 16 gas detector requires a monthly maintenance (cf. Section 3.2, p 8). If running a full alarm test, deactivate the ESL fire station. Failing to do so would result in the activation of the Building alarm and its evacuation. If you are testing the MDA gas detector, you should disable the Zone 4 of the ESL fire Control Unit. If you want to test the Fire panel itself (smoke detectors), you must contact Physical Plant to have them disconnect the relay to the building alarm MDA System 16 Gas Detector The alarm LED s and alarm relays should be tested every month: - Press the service function key on the System 16 control module. - Select Subfunction Alarm Test by pressing key 2 twice. - Press 0, to select a full-system test or 1 if you want to check each point individually. - Remote Alarm 1 and 2 should go off and gas interlock should be shutdown. - Press Service on control module to turn off the alarm ESL Fire Alarm Control Unit The three smoke detectors should be tested individually. - To test the smoke detectors, use the Smoke Detector Tester. Aim the spray for 2 seconds at detector. - The related zone relay should be activated: the internal audible alarm will go off as well as remote alarm 1 and 2. Gas interlock should be shutdown. - To deactivate the alarm, switch to Reset in the ESL system control unit Photohelic gauges The three photohelic gauges should be tested individually and every month. - For each photohelic gauges, increase the low setpoint till it reaches the current pressure in the duct system: Gas interlock should be shutdown. - To turn off alarm, reset the setpoint to its original value. Safety and Emergency procedure in LAMP Page 8 of 17

9 1.3.3 Electrical diagram of the MDA alarm interface and peripherals Safety and Emergency procedure in LAMP Page 9 of 17

10 1.3.4 MDA Interface unit: relay connections Power ON No alarm P1 and P3 normally close => V P 2 = 24V => P2 relay closed (gas interlock OK) If alarm level 2 conditions (System 16, ESL fire panel, EPO) V P 1 = 24V => P1 opens => P3 is still closed => V P 2 = 0V => P2 opens (gas interlock shutdown) V P 1 = V A 1 = V Q = 24 V => A1 closes (Alarm level 2), Q opens (monitoring station, hallway enunciator) If alarm level 1 condition (photohelic gages, overpressure switch) V P 3 = 24 V =>P3 opens => V P 2 = OV => P2 opens (gas interlock shutdown) Power OFF P2 opens => gas interlock shuts down Safety and Emergency procedure in LAMP Page 10 of 17

11 1.3.5 Floor plan of LAMP and positioning of sensor and alarm units MDA alarm unit Remote alarm 1 enunciator Gas interlock panel 8 lbnh3 gas bottle H2 gas bottle Smoke detectors WF6 sensor Silane jacketed line Silane gas detector unit UMPD unit ESL Fire HALLWAY X PRE CLEANROOM Mech. ROOM Remote alarm 2 enunciator Chemicals ESL Fire flashlight EPO flashlight EPO CLEANROOM Gas scrubber Zone 15 lbwf6 gas bottle Floor plan of LAMP and locations of sensor and alarm units Building Security Systems Unit SECURITY SYSTEM OPERATION 1). TO ARM THE SYSTEM: a). VERIFY THAT THE GREEN READY LIGHT IS ON. b). ENTER YOUR 4-DIGIT CODE 1234 OR PRESS THE AWAY BUTTON UNTIL THE SYSTEM ARMS. Safety and Emergency procedure in LAMP Page 11 of 17

12 c). EXIT THE BUILDING AND CLOSE THE DOOR WITHIN 45 SECONDS. THE SYSTEM IS NOW ON!! 2). TO DISARM THE SYSTEM: a). UPON ENTERINGTHE BUILDING YOU WILL HEAR A PRE-ALARM SOUND FROM THE KEYPAD. b). ENTER YOUR 4-DIGIT CODE INTO THE KEYPAD. c). THE SOUNDER AND THE ALARM WILL TURN OFF. d). THE SYSTEM IS NOW OFF!!! 3) WHAT DO I DO IF THE KEYPAD SOUNDER ACTIVATES? a). CHECK THE MESSAGE READ WHAT IS SAYS AND FOLLOW DIRECTION ON THE SCREEN. b). TO SILENCE THE KEYPAD OR ALARM JUST ENTER YOUR CODE. c). IF ALL ELSE FAILS CALL (301) OR (301) AND ASK FOR BUILDING SECURITY. 4) How do I set the system to arm itself at night? a). Press the * button then press the number 6. b). Enter your MASTER CODE. c). Press the right > arrow until the screen says ENTER AUTO ARM TIME d). Enter the time (Military time) that you want the system to turn on. e). Then press the right > arrow until the system says AUTO ARM CONTROL. f). Press the * button to have the system auto arm. The screen will confirm that the feature is either on or off. Please read what the system says. Often people will think they have turned on the auto arm function when in fact they have actually turned it off. Safety and Emergency procedure in LAMP Page 12 of 17

13 2 EMERGENCY RESPONSE PROCEDURE In case of an alarm in LAMP resulting from a leak of silane and/or fire detection, the following actions are automatically taken by the LAMP safety system: - Activation of 2 red flash lights in pre-cleanroom and cleanroom with audible alarms plus a blue flash light above hallway door with audible evacuation call - Enunciation to UM Police Department via supervisory unit - Activation of building alarm Anyone present in the lab must: 1. Evacuate the facility immediately 2. If main building alarm is not already sounding, pull manual station in hallway for general evacuation of the building 3. Call 911 to confirm they have been notified by automated supervisory unit (JM Patterson building (083), second floor, room 2225) 4. Call the Lab manager Laurent ). In any case, it should not be attempted by any students to evaluate the source and gravity of the alarms. Safety and Emergency procedure in LAMP Page 13 of 17

14 3 PROCESS GAS CYLINDER HANDLING PROCEDURES Full face shields and Nomex cover-all suits required for handling SiH 4 cylinders Breathing protection is required for changing WF Gas cylinder receiving Delivery of SiH 4 and WF 6 cylinders will be coordinated with the supplier (Air Products) and received at the rear entrance of the JM Patterson Building. Before entering the building the cylinder will be inspected for signs of a leak (discoloring or residue) and then will be transported to the LAMP facility and placed in a cabinet until installation. 3.2 First time installation of new cylinder This procedure is for installaion of a new cylinder on to gas system which has not been previously used with process gases Cycle purge and leak check process line Pump out the line upstream of the regulator using through the process tool. a) Set the ULVAC so that the chamber is being pumped by the turbo and monitored by the ion gauge. b) Close VS4 c) Open the process valves (PVs) and set the MFC to full open. d) Pump on the reactor chamber using the turbo pump. e) After reaching a base pressure (pump for at least 15 minutes) at the ion gauge, close the process valve immediately down stream of the MFC. f) After approximately 30 sec open this valve and monitor the effect at the ion gauge. A significant pressure jump MAY signify be a leak in the process plumbing. If this is observed, repeat the test after pumping on the chamber for another 15 minutes Installing and activation new cylinder Step 1: Connect the cylinder to the gas panel a) Secure cylinder in place in the gas cabinet b) Remove cylinder cap c) With the main cylinder valve closed (i.e., pneumatic air line NOT attached) SLOWLY remove the DISS connection cap and remove the gasket. Check for signs of a significant leak through the cylinder valve (white powder, smoke) d) Using a new DISS gasket, attach the pigtail fitting to the cylinder. Step 2: Cycle purge the gas panel Starting with all valves closed. a) Open VS5 to start N2 flow through the venturi element to generate a vacuum at VS2. b) Open VS3 and set the regulator open to near full pressure delivery. c) Open VS1 fill the panel with purge N2 and then close VS1. Safety and Emergency procedure in LAMP Page 14 of 17

15 d) Open VS2 to evacuate the panel, checking for a pressure drop to near 30" Hg at the regulator low P side. e) Close VS2. f) Repeat steps c and d AT LEAST 10 times. g) Complete this procedure with step d. h) Leave the system for approximately 1 hour noting the pressure at the regulator gauge. i) If after 1 hour the pressure does not rise, close VS3 and proceed to the next step. Step 3: Cycle purge the process line. Deliver N2 to the process line from the purge cylinder. a) Open VS1 to apply pressure to the panel. b) Open VS3 and set pressure at regulator to ~ 10 psi. c) Open VS4 to pressurize the process line with the purge N2. d) Close VS4. e) Put the SiH 4 MFCs in full open and open the upstream PV. f) Open the By-Pass PV to evacuate process line back to the cylinder valve VS4. g) Close the By-Pass PV. h) Open VS4 to fill the process line, then close VS4 i) Repeat steps f through h at least 10 times. j) End the process with step g, evacuating the process line. k) Cycle Purge the panel (see above Step 2) and leave in evacuated state Switching cylinders Removal of used cylinder Verify that the gas cylinder main valve is closed. Disconnect the air line to the pneumatically operated main cylinder valve. Step 1 Cycle purge the panel. Starting with all valves closed. a) Open VS5 to start N2 flow through the venturi element to generate a vacuum at VS2. b) Set the regulator open to near full pressure delivery. c) Open VS2 to evacuate the panel, checking for a pressure drop to near 30" Hg at the regulator low P side. d) Close VS2. e) Open VS1 to fill the panel with purge N2. f) Close VS1. g) Repeat steps c and d AT LEAST 10 times. h) Open VS1 and pressurize the panel with purge gas to provide outward purge flow when the pigtail connection is disconnected. Step 2 Disconnect and remove the used cylinder a) VERIFY that the pneumatic air line to the main cylinder valve is disconnected. b) Slowly loosen DISS cylinder fitting and disconnect. c) Place cap over cylinder valve. d) Remove cylinder and store for pick-up. Safety and Emergency procedure in LAMP Page 15 of 17

16 Installation of new cylinder Step 1: Connect the cylinder to the gas panel e) Secure cylinder in place in the gas cabinet f) Remove cylinder cap g) With the main cylinder valve closed (i.e., pneumatic air line NOT attached) SLOWLY remove the DISS connection cap and remove the gasket. Check for signs of a significant leak through the cylinder valve (white powder, smoke) h) Using a new DISS fitting gasket, attach the pigtail fitting to the cylinder. Step 2: Cycle purge the gas panel Start with VS1 open, all other valves closed. a) Open VS5 to start N2 flow through the venturi element to generate a vacuum at VS2. b) Open VS3 and set the regulator open to near full pressure delivery. c) Close VS1 d) Open VS2 to evacuate the panel, checking for a pressure drop to near 30" Hg at the regulator low P side. e) Close VS2. f) Open VS1 g) Repeat steps c and d AT LEAST 10 times. h) Complete this procedure with step e (panel evacuated). i) Leave the system for approximately 1 hour noting the pressure at the regulator gauge. [A pressure rise is associated with a leak in the panel or the DISS connection.] j) If no signs of a leak is detected, proceed to step 3. Step 3: Open the cylinder Start with all valves closed. Verify that dowstream process valves are closed. a) Verify that the electronic switch for the main cylinder valve is OFF. b) Connect the pneumatic air line for the cylinder main valve. c) Open the SiH 4 main cylinder valve using the switch at the remote panel. d) Open VS3 and check pressure at the panel regulator. e) Open VS4. f) Adjust panel regulator to 30 psi to deliver SiH 4 pressure to the ULVAC. Safety and Emergency procedure in LAMP Page 16 of 17

17 3.3 Silane Usage Policy Silane check-list This check list must be filled out before any utilization of silane. Blank copies are in a log book located at the control module. No less than two knowledgeable operators must be present at any time when using the silane gas in a deposition experiment. Is the ATM gas scrubber on? Is the scrubber line < 400 Torr? Is the System 16 Monitor on? Are the three status LEDs all green? Is SiH 4 line jacket purge flow on? YES? Operators Name 1 Name 2 Date Hour Date Hour Safety and Emergency procedure in LAMP Page 17 of 17

Standard Operating Procedures for Process Gas Handling

Standard Operating Procedures for Process Gas Handling Standard Operating Procedures for Process Gas Handling A supplement to the SOP for Laboratory for Advanced Materials Processing JM Patterson Bldg., Room 2225 University of Maryland Revision 1.1 DRAFT April

More information

Testing the System. Battery Test. Dialer Test. Fire Drill Test (Code + [#] + 69) One-Man Fire Walk-Test (Code + [#] + 68)

Testing the System. Battery Test. Dialer Test. Fire Drill Test (Code + [#] + 69) One-Man Fire Walk-Test (Code + [#] + 68) F A 1 7 0 0 c Testing the System Battery Test When AC power is present, the FA1700C runs a brief battery test every 60 seconds to determine if there is a battery connected, and runs an extended battery

More information

ZSC100 Gas Detection and Alarm System Controller

ZSC100 Gas Detection and Alarm System Controller ZSC100 Gas Detection and Alarm System Controller User Guide 1- Introduction... 3 1.1- General description... 3 1.2- Cautions and warnings... 3 2- Control panel... 4 2.1 Control panel overview... 4 2.2

More information

D8024, D9024, D10024 Analog Fire Alarm Control Panels Programming Guide

D8024, D9024, D10024 Analog Fire Alarm Control Panels Programming Guide System Reset Trou ble Silence Ala rm Silence Manual Ala rm ENTER NO YES Letters Numb ers Keyword Radionics System Reset Trouble Silence Alarm Silence Manual Alarm ENTER NO YES Le ters Numbers Keyw ord

More information

Model 17A00 Expansion Enclosure

Model 17A00 Expansion Enclosure HOME AUTOMATION, INC. Model 17A00 Expansion Enclosure Installation Manual Document Number 17I00-1 Rev A March, 2002 Home Automation, Inc. Model 17A00 Expansion Enclosure Installation Manual Document Number

More information

0 4 / 0 4 / 1 4. GE Concord 4 Quick User Guide. GE Concord 4 Quick User Guide Page 1

0 4 / 0 4 / 1 4. GE Concord 4 Quick User Guide. GE Concord 4 Quick User Guide Page 1 0 4 / 0 4 / 1 4 GE Concord 4 Quick User Guide GE Concord 4 Quick User Guide Page 1 Before Calling Is the keypad beeping? Press *. This will silence the beeping and let you know where the trouble is. Is

More information

SIMPLICITY CO CARBON MONOXIDE DETECTION & VENTILATION PANEL

SIMPLICITY CO CARBON MONOXIDE DETECTION & VENTILATION PANEL SIMPLICITY CO CARBON MONOXIDE DETECTION & VENTILATION PANEL USER MANUAL 1 Table of Contents 1 SAFETY INFORMATION...3 1.1 SAFETY PRECAUTIONS DURING NORMAL OPERATION OF PANEL...3 1.3 BATTERY INFORMATION...3

More information

Syncro AS. Analogue Addressable Fire Control Panel. User Manual

Syncro AS. Analogue Addressable Fire Control Panel. User Manual Syncro AS Analogue Addressable Fire Control Panel User Manual Man-1100 Issue 02 Nov. 2008 Index Section Page 1. Introduction...3 2. Safety...3 3. Panel Controls...4 3.1 Access Level 1...4 3.2 Access Level

More information

Pioneer-R16 Gas Monitor Operator s Manual

Pioneer-R16 Gas Monitor Operator s Manual Pioneer-R16 Gas Monitor Operator s Manual Edition 7/2/97 RKI INSTRUMENTS, INC RKI Instruments, Inc. 33248 Central Ave, Union City, CA 94587 (510) 441-5656 Chapter 1: Description About the Pioneer-R16 Gas

More information

ZM2 OPERATING AND MAINTENANCE INSTRUCTIONS

ZM2 OPERATING AND MAINTENANCE INSTRUCTIONS ZM2 OPERATING AND MAINTENANCE INSTRUCTIONS CONTENTS 1. System Overview 2. User responsibilities 3. First Line controls and indications 3.1 Logging on to the fire alarm system 3.2 Checking system status

More information

User s Information Guide R2A

User s Information Guide R2A Pi HSC505 Home Security Controller User s Information Guide R2A Page 1 of 15 of its development program. 1This document and product are copyrighted and all rights are reserved. Introduction Convention

More information

Watchguard WGAP864 User Manual

Watchguard WGAP864 User Manual Watchguard WGAP864 User Manual v1.0 Issued September 2016 1 2 Table of Contents Glossary... 5 1. Introduction to your Watchguard WGAP864... 6 2. Before Operating your Alarm System... 6 3. Understanding

More information

SILENCING AN ALARM When the alarm bell or siren is sounding, enter your user code or present your SecuraProx fob to your keypad.

SILENCING AN ALARM When the alarm bell or siren is sounding, enter your user code or present your SecuraProx fob to your keypad. SYSTEM USER GUIDE SILENCING AN ALARM When the alarm bell or siren is sounding, enter your user code or present your SecuraProx fob to your keypad. IS THIS A FALSE ALARM? YES NO displays. REAL ALARM If

More information

FMR Installation Instructions. Addressable Fire System Controller

FMR Installation Instructions. Addressable Fire System Controller FMR-7033 EN Installation Instructions Addressable Fire System Controller EN 2 FMR-7033 Installation Instructions Contents Contents Contents...2 1.0 Introduction...3 2.0 Description...3 3.0 Device Mounting...3

More information

ZX1e ZX2e ZX5e. Document No Issue 01 user manual

ZX1e ZX2e ZX5e. Document No Issue 01 user manual ZX1e ZX2e ZX5e Document No. 996-130 Issue 01 user manual MORLEY-IAS ZX2E/ZX5E Fire Alarm Control Panels Table of Contents 1 INTRODUCTION... 4 1.1 NOTICE... 4 1.2 WARNINGS AND CAUTIONS... 4 1.3 NATIONAL

More information

U ser's Guide PC6010

U ser's Guide PC6010 User's Guide PC6010 Quick Reference Guide This manual is for Basic and Advanced users. Each of these types of user can access a different set of functions. The and symbols next to the title of each procedure

More information

GasScanner 8C. Eight Channel Monitor. Operator s Manual. MINT-0281-XX Rev. A 01/29/08

GasScanner 8C. Eight Channel Monitor. Operator s Manual. MINT-0281-XX Rev. A 01/29/08 GasScanner 8C Eight Channel Monitor Operator s Manual MINT-0281-XX Rev. A 01/29/08 Product Warranty Matheson Tri-Gas, Inc., warrants gas alarm equipment sold by us to be free from defects in materials,

More information

CSP-204 CSP-208 CSP-104 CSP-108

CSP-204 CSP-208 CSP-104 CSP-108 Fire Alarm Control Panel CSP-204 CSP-208 CSP-104 CSP-108 Operation manual Firmware version 1.1 csp-x_o_en 06/15 SATEL sp. z o.o. ul. Budowlanych 66 80-298 Gdańsk POLAND tel. 58 320 94 00 www.satel.eu CONTENTS

More information

RF280THS. Installation Instructions Wireless (RF) Photoelectric Smoke Detectors

RF280THS. Installation Instructions Wireless (RF) Photoelectric Smoke Detectors RF80THS EN Installation Instructions Wireless (RF) Photoelectric Smoke Detectors RF80THS Installation Instructions Trademarks Trademarks Chamber Check is a registered trademark of Bosch Security Systems

More information

To operate the 5207 you can use either the built-in touchpad or the Model 5230 Remote Annunciator.

To operate the 5207 you can use either the built-in touchpad or the Model 5230 Remote Annunciator. Section 5 Operation To operate the 5207 you can use either the built-in touchpad or the Model 5230 Remote Annunciator. Figure 5-1 Built-in Touchpad (Seven-Segment Display) Figure 5-2 Model 5230 Remote

More information

OPERATION MANUAL RK-2006LPP AUGER FITTED SOLID FUEL BOILER TEMPERATURE CONTROLLER. Version DC19

OPERATION MANUAL RK-2006LPP AUGER FITTED SOLID FUEL BOILER TEMPERATURE CONTROLLER. Version DC19 OPERATION MANUAL RK-2006LPP AUGER FITTED SOLID FUEL BOILER TEMPERATURE CONTROLLER Version DC19 1. Application. Controller RK-2006LPP is designed for temperature control of solid fuel fired water boilers

More information

Supervised Security System Owner's Guide

Supervised Security System Owner's Guide Owner's Guide PSC06 READ THIS FIRST This equipment generates and uses radio frequency energy, and if not installed and used properly, that is, in strict accordance with the manufacturers instructions,

More information

Installation and user manual for the CF5000, MF5000 and FXP5000 range of fire panels

Installation and user manual for the CF5000, MF5000 and FXP5000 range of fire panels CF5000, MF5000 and FXP5000 Installation and user manual for the CF5000, MF5000 and FXP5000 range of fire panels 16 zone panels Contents PANEL INSTALLATION...3 Installation...3 PANEL WIRING... 3 Mains power

More information

Laptop / PC Programming Manual

Laptop / PC Programming Manual Laptop / PC Programming Manual Doc. # Fire PC Program rev B 01.07 This Document is property of Evax Systems, Inc. The Evax Fire Solutions Programmer Components 2 1.0 System Setup 4 1.1 Interface Setup

More information

VISTA-32FBPT. Commercial Fire and Burglary Partitioned Security Systems with Scheduling. User Guide /12 Rev. B

VISTA-32FBPT. Commercial Fire and Burglary Partitioned Security Systems with Scheduling. User Guide /12 Rev. B VISTA-32FBPT Commercial Fire and Burglary Partitioned Security Systems with Scheduling User Guide 800-11045 2/12 Rev. B 2 TABLE OF CONTENTS SYSTEM OVERVIEW...5 General...5 A Partitioned System...5 Zones...6

More information

Beacon 200 Gas Monitor Operator s Manual. Part Number: RK Released: 6/6/08

Beacon 200 Gas Monitor Operator s Manual. Part Number: RK Released: 6/6/08 Beacon 200 Gas Monitor Operator s Manual Part Number: 71-2102RK Released: 6/6/08 Table of Contents Chapter 1: Introduction.................................................3 Overview.............................................................3

More information

vsource GAS SOURCE SYSTEMS

vsource GAS SOURCE SYSTEMS Gas Cabinets vsource VERSA-GAS vsource Gas Cabinets address diverse high purity gas delivery requirements across a variety of environments, with safety top of mind. Every vsource model is designed to ensure

More information

AFP-300/400 OPERATORS MANUAL INTELLIGENT FIRE DETECTION AND ALARM SYSTEM SOFTWARE VERSION 2.2 REVISION AUS 3

AFP-300/400 OPERATORS MANUAL INTELLIGENT FIRE DETECTION AND ALARM SYSTEM SOFTWARE VERSION 2.2 REVISION AUS 3 OPERATORS MANUAL AFP-300/400 INTELLIGENT FIRE DETECTION AND ALARM SYSTEM SOFTWARE VERSION 2.2 REVISION AUS 3 NSW (Head Office) 7 Columbia Court Norwest Business Park Baulkham Hills NSW 2153 Ph: (02) 9899-4155

More information

Matrix2000. User Guide - CPD Advanced Analog Fire Alarm Control Panel 4 24 Zones. Version: 2.0 Revision: 1

Matrix2000. User Guide - CPD Advanced Analog Fire Alarm Control Panel 4 24 Zones. Version: 2.0 Revision: 1 Matrix2000 User Guide - CPD Advanced Analog Fire Alarm Control Panel 4 24 s Version: 2.0 Revision: 1 IMPORTANT INFORMATION Limitation of liability It is mandatory, Matrix2000 panel to be installed in accordance

More information

Interactive Fire Control Panel IFS7002 one signal loop Instruction Manual

Interactive Fire Control Panel IFS7002 one signal loop Instruction Manual Interactive Fire Control Panel IFS7002 one signal loop Instruction Manual Revision 4/01.17 Contents 1. Introduction... 6 2. Terminology... 6 3. Function... 8 4. Technical data... 8 4.1. Physical configuration...

More information

MADE IN THE U.S.A. ADVANCE FIRE CONTROL MODEL LM21-AFC MANUAL RELEASE DEVICES TEST WEEKLY TO ASSURE PROPER OPERATION OF RELEASE DEVICE/CONTROL PANEL

MADE IN THE U.S.A. ADVANCE FIRE CONTROL MODEL LM21-AFC MANUAL RELEASE DEVICES TEST WEEKLY TO ASSURE PROPER OPERATION OF RELEASE DEVICE/CONTROL PANEL RELEASE DEVICES MADE IN THE U.S.A. ADVANCE FIRE CONTROL MODEL LM21-AFC MANUAL The LiftMaster Fire Control (LM21-AFC) Release Device/Control Panel is UL/CUL listed normally energized fail-safe device incorporating

More information

abode Web app Functionality

abode Web app Functionality abode Web app Functionality System mode display controls the state of the alarm system. The user can change the mode of the system by simply clicking on the desired mode. When arming the system with the

More information

Covidien Articulating Absorba Tack Foil Pouch Leak Tester Owner s Manual IPE SN Rev 1

Covidien Articulating Absorba Tack Foil Pouch Leak Tester Owner s Manual IPE SN Rev 1 Covidien Articulating Absorba Tack Foil Pouch Leak Tester Owner s Manual IPE SN 12497 Rev 1 MANUFACTURER: INNOVATIVE PRODUCTS & EQUIPMENT, INC. 5 PROGRESS AVE TYNGSBORO, MASSACHUSETTS 01879 U.S.A Revision

More information

GLD-30 Gas Leak Detector

GLD-30 Gas Leak Detector GLD-30 Gas Leak Detector Installation, Operation & Maintenance General: The Archer Instruments GLD-30 is an ambient air monitor, used to detect the presence of a target gas (or gases) and to alert operators

More information

/10 Rev A

/10 Rev A VISTA-128BPT/ VISTA-250BPT/ VISTA-128BPTSIA Commercial Burglary Partitioned Security System With Scheduling User Guide 800-06905 6/10 Rev A Your Honeywell security system is designed for use with devices

More information

The Kryos LN2 Liquid Level Control & Cryogenic Temperature Control

The Kryos LN2 Liquid Level Control & Cryogenic Temperature Control The Kryos LN2 Liquid Level Control & Cryogenic Temperature Control Created for Taylor-Wharton Gas Equipment By Pacer Digital Systems, Inc. INTRODUCTION... 4 TEXT FORMAT NOTATION... 4 SYSTEM COMPONENTS...

More information

Across-the-Line SERIES MP300 Combined Manual and Automatic

Across-the-Line SERIES MP300 Combined Manual and Automatic Across-the-Line SERIES MP300 Combined Manual and Automatic Metron Fire Pump Controllers conform to the latest requirements of National Fire Protection Association s Standard for Centrifugal Fire Pumps

More information

Summit 3208GLD USER MANUAL. Electronics Line

Summit 3208GLD USER MANUAL. Electronics Line Summit 3208GLD USER MANUAL Electronics Line Table of Contents 1: Introduction... 2 2: Overview... 3 3: Keypad Functions... 4 3.1: Keypads... 4 3.2: 3108 LCD Keypad Layout... 4 4: Basic System Operation...

More information

**** **** A B ARMED FIRE FIRE C D READY OFF AWAY STAY 2 3 MAXIMUM TEST BYPASS PULL INSTANT CODE CHIME. First Alert READY. * 0 # Professional

**** **** A B ARMED FIRE FIRE C D READY OFF AWAY STAY 2 3 MAXIMUM TEST BYPASS PULL INSTANT CODE CHIME. First Alert READY. * 0 # Professional ARMED READY READY )$&&$&% )LUHDQG%XUJODU\ 3DUWLWLRQHG6HFXULW\6\VWHPV ZLWK6FKHGXOLQJ **** **** A B C D FIRE FIRE First Alert OFF AWAY STAY 2 3 1 4 5 6 7 8 9 MAXIMUM TEST BYPASS INSTANT CODE CHIME * 0 #

More information

&RPPHUFLDO)LUHDQG%XUJODU\ 3DUWLWLRQHG6HFXULW\6\VWHP ZLWK6FKHGXOLQJ

&RPPHUFLDO)LUHDQG%XUJODU\ 3DUWLWLRQHG6HFXULW\6\VWHP ZLWK6FKHGXOLQJ 9,67$)% &RPPHUFLDO)LUHDQG%XUJODU\ 3DUWLWLRQHG6HFXULW\6\VWHP ZLWK6FKHGXOLQJ 8VHU*XLGH FIRE FIRE * PULL K3522 3/99 TABLE OF CONTENTS SYSTEM OVERVIEW...4 General...4 A Partitioned System...4 Zones...5 Fire

More information

3500 CONVENTIONAL FIRE ALARM CONTROL PANEL

3500 CONVENTIONAL FIRE ALARM CONTROL PANEL 3500 CONVENTIONAL FIRE ALARM CONTROL PANEL USER MANUAL Protec Fire Detection PLC, Protec House, Churchill Way, Nelson, Lancashire, BB9 6RT. Telephone: +44 (0) 1282 717171 Fax: +44 (0) 1282 717273 Web:

More information

Ammonia Detection System Codes and Design Specifications

Ammonia Detection System Codes and Design Specifications Ammonia Detection System Codes and Design Specifications January 15, 2018 Revision 6 Ammonia detection system codes and design specifications Following is a discussion of ammonia detection system design

More information

CHEETAH XI INTELLIGENT SUPPRESSION CONTROL SYSTEM

CHEETAH XI INTELLIGENT SUPPRESSION CONTROL SYSTEM CHEETAH XI INTELLIGENT SUPPRESSION CONTROL SYSTEM DESCRIPTION Fike s Cheetah Xi (P/N 10-068) is a state-of-the-art true intelligent digital peer-topeer modular suppression control system. It is ideal for

More information

PF1000 CONTROLLER. for the. Heateflex Corporation 405 E. Santa Clara St. Arcadia, CA TEL: (626) ; FAX: (626)

PF1000 CONTROLLER. for the. Heateflex Corporation 405 E. Santa Clara St. Arcadia, CA TEL: (626) ; FAX: (626) PF1000 CONTROLLER for the AQUARIUS I/II DEIONIZED WATER HEATING SYSTEM Heateflex Corporation 405 E. Santa Clara St. TEL: (626)599-8566; FAX: (626)599-9567 Rev. 07 10/19/10, (PF1000) 1 2 TABLE OF CONTENTS

More information

The Dorm Room Genie. The Door Lock --To be completed by David Nedzel

The Dorm Room Genie. The Door Lock --To be completed by David Nedzel The Door Lock The Dorm Room Genie Students are issued keys by which to access their rooms, but far too often a key is forgotten inside the room leaving someone locked out. After a lockout, a student must

More information

RE6100 Series Helix Security and Automation Platform

RE6100 Series Helix Security and Automation Platform CONFIGURATION Resolution Compatibles RE6100 Series Helix Security and Automation Platform C G UI D E Configuration Settings Table 1 - Panel Settings Table 2 - Zone Settings Table 3 - Device Settings Table

More information

Digiplex LED Keypads User s Manual

Digiplex LED Keypads User s Manual KLEDEU03.fm Page -1 Friday, May 4, 2001 11:25 AM Digiplex LED Keypads User s Manual KLEDEU03.fm Page 0 Friday, May 4, 2001 11:25 AM KLEDEU03.fm Page 1 Friday, May 4, 2001 11:25 AM TABLE OF CONTENTS 1.0

More information

one Operating instructions SenTRI ONE panel-based Fire detection and alarm system

one Operating instructions SenTRI ONE panel-based Fire detection and alarm system SenTRI ONE panel-based Fire detection and alarm system one 042bc/14 4188-978 issue 2_04-10_SenTRI ONE system oper Part of Document pack 2535-235 issue 2 1 Contents User responsibility - - - - - - - - -

More information

&RPPHUFLDO%XUJODU\ 3DUWLWLRQHG6HFXULW\6\VWHP ZLWK6FKHGXOLQJ

&RPPHUFLDO%XUJODU\ 3DUWLWLRQHG6HFXULW\6\VWHP ZLWK6FKHGXOLQJ READY ARMED READY 1 OFF 7 INSTANT READY 2 AWAY 8 CODE 6BYPASS 9 CHIME 9,67$% &RPPHUFLDO%XUJODU\ 3DUWLWLRQHG6HFXULW\6\VWHP ZLWK6FKHGXOLQJ 8VHU*XLGH ARMED READY 1 OFF 2 AWAY 3 STAY 4 MAX 5 TEST 6 BYPASS

More information

Wireless Keypads LKP(E)S8M Series

Wireless Keypads LKP(E)S8M Series Wireless Keypads LKP(E)S8M Series User manual Contents Congratulations on your purchase of this Honeywell wireless keypad. To make the best out of your equipment we advise you to read this manual carefully.

More information

Flopurge TS. Operation Manual

Flopurge TS. Operation Manual Flopurge TS Operation Manual Part Number 079-0204 Spectron Gas Control Systems United Kingdom Unit 4, Herald Court, University of Warwick Science Park, Coventry, CV4 7EZ +44 (0)24 7641 6234 sales@spectron-gcs.com

More information

MODEL B2 INSTALLATION MANUAL

MODEL B2 INSTALLATION MANUAL RELEASE DEVICES GENERAL DESCRIPTION MODEL B2 INSTALLATION MANUAL S/N: The B2 Series Time Delay Release Devices are UL Listed, Canadian Listed, and CSFM Listed for use on rolling doors, single-slide and

More information

AdvisorOne Quick Operation Guide

AdvisorOne Quick Operation Guide AdvisorOne Quick Operation Guide EN CNF ES PR English: Quick Operation Guide Introduction AdvisorOne (part # AVO-1037) is a wireless, self-contained system used to detect, control and inform of security

More information

La Marche Manufacturing Company Option 46 Series. Digital Combined Accessory Package. Installation and Operation Manual

La Marche Manufacturing Company   Option 46 Series. Digital Combined Accessory Package. Installation and Operation Manual La Marche Manufacturing Company www.lamarchemfg.com Option 46 Series Digital Combined Accessory Package Installation and Operation Manual This manual is subject to change without notice. You may obtain

More information

Gas Fired Humidifier. SKG Series Combustion Field Adjustment Instructions. SKG Combustion Field Adjustment Instructions

Gas Fired Humidifier. SKG Series Combustion Field Adjustment Instructions. SKG Combustion Field Adjustment Instructions Fired Humidifier SKG Series SKG - 180301 Standard Combustion Adjustment SKG Fired Humidifier All installation work must be carried out by suitably qualified personnel and must conform to local codes and

More information

Denton Thermal Evaporator SOP. UNT Cleanroom

Denton Thermal Evaporator SOP. UNT Cleanroom Denton Thermal Evaporator SOP UNT Cleanroom Utility Checklist: The PCW water valve is open to the cryo compressor (in chase) N2 and CDA are on Cooling water valves to the tools are on 1. Mounting the substrate

More information

Section 9 System Operation

Section 9 System Operation Section 9 System Operation Operation of the control panel is simple. Menus guide you step-by-step through operations. This section of the manual is an overview of the operation menus. Please read this

More information

Soft Start Series MP700 Solid State, Reduced Voltage

Soft Start Series MP700 Solid State, Reduced Voltage Metron Fire Pump Controls and Accessories Soft Start Series MP700 Solid State, Reduced Voltage Metron Fire Pump Controllers conform to the latest requirements of National Fire Protection Association s

More information

Master Code Arming Auto-Bypass Option - Home-Away Arming Entry Delay Off Arming

Master Code Arming Auto-Bypass Option - Home-Away Arming Entry Delay Off Arming Master Code The 4 digit Master Code is used for arming and disarming the system, for programming additional access codes, and for changing other features. The Master Code will be supplied to you by your

More information

Beacon 800 Gas Monitor Operator s Manual

Beacon 800 Gas Monitor Operator s Manual Beacon 800 Gas Monitor Operator s Manual Part Number: 71-0037RK Revision: F Released: 4/18/17 www.rkiinstruments.com Product Warranty RKI Instruments, Inc. warrants gas alarm equipment sold by us to be

More information

RK/ RK Carbon Monoxide Detector Operator s Manual

RK/ RK Carbon Monoxide Detector Operator s Manual 65-2496RK/65-2499RK Carbon Monoxide Detector Operator s Manual Part Number: 71-0156RK Revision: 0 Released: 2/16/11 www.rkiinstruments.com WARNING Read and understand this instruction manual before operating

More information

FIREFLY II PLUS RELEASE DEVICES INSTALLATION MANUAL

FIREFLY II PLUS RELEASE DEVICES INSTALLATION MANUAL FIREFLY II PLUS RELEASE DEVICES INSTALLATION MANUAL MADE IN THE U.S.A. U.L. LISTED CANADIAN LISTED CSFM: 7300-1418:100 GENERAL DESCRIPTION SERIAL NUMBER The Cookson Company FIREFLY II PLUS Time Delay Release

More information

USER GUIDE HARDWIRED CONTROL PANELS

USER GUIDE HARDWIRED CONTROL PANELS USER GUIDE HARDWIRED CONTROL PANELS Scantronic Contents 1. Introduction... 3 The Alarm System... 3 The Keypads... 3 The 725r Remote Setting Device... 6 About This Guide... 6 2. Everyday Operation... 7

More information

User s Manual. System Information... 3 Using the Touchpad Buttons... 3 General Operation... 4 Using Systems with Keyswitches... 4

User s Manual. System Information... 3 Using the Touchpad Buttons... 3 General Operation... 4 Using Systems with Keyswitches... 4 ITI Part No. 60-562-03 60-562-06 $GYHQWŒ&RPPHUFLDO )LUH6\VWHP Document Number: 466-1705 Rev. B November 2001 System Information... 3 Using the Touchpad Buttons... 3 General Operation... 4 Using Systems

More information

QA16 Addressable System

QA16 Addressable System QA16 Addressable System Operating Manual HORING LIH INDUSTRIAL CO., LTD. www.horinglih.com QA16 System Characteristics Each loop can connect with 250 devices. Easy system programming through PC to panel.

More information

GSM RFID VOICE Alarm System

GSM RFID VOICE Alarm System GSM RFID VOICE Alarm System User s Manual For a better understanding of this product, please read this user manual thoroughly before using it. CONTENTS [Function Instruction] [Control Panel] Control Panel

More information

DATA SHEET CHEETAH XI 50 INTELLIGENT SUPPRESSION CONTROL SYSTEM DESCRIPTION APPROVALS: SYSTEM OPERATION. Form No. D

DATA SHEET CHEETAH XI 50 INTELLIGENT SUPPRESSION CONTROL SYSTEM DESCRIPTION APPROVALS: SYSTEM OPERATION. Form No. D DATA SHEET CHEETAH XI 50 INTELLIGENT SUPPRESSION CONTROL SYSTEM DESCRIPTION Fike s Cheetah Xi 50 (P/N 10-071) is a state-of-the-art true intelligent digital peer-to-peer modular suppression control system.

More information

WIRELESS ALARM SYSTEM WITH TELEPHONE AUTO DIALER

WIRELESS ALARM SYSTEM WITH TELEPHONE AUTO DIALER BAT.LOW AC WIRELESS ALARM SYSTEM WITH TELEPHONE AUTO DIALER THE SYSTEM THAT CALLS YOU! Our WIRELESS ALARM SYSTEM WITH TELEPHONE AUTO DIALER is designed to allow you to create your own security system.

More information

Ammonia Leak Detector Manual

Ammonia Leak Detector Manual Ammonia Leak Detector Manual Document No. 70-PHW-1023-R2.5 No part of this publication may be reproduced, stored in a retrieval system, or transmitted, in any form or by any means, electronic, mechanical,

More information

Security System. User Guide for the LED Command Center

Security System. User Guide for the LED Command Center Security System User Guide for the LED Command Center National Security Systems Inc (800)457-1999 MY SECURITY COMPANY IS: CALL BEFORE TEST: THIS SECURITY SYSTEM IS CONNECTED TO TELEPHONE NUMBER: THE SECURITY

More information

INSTRUCTION MANUAL FOR. Remote Control Car Alarm with Impact Sensor, Mini-Battery Backup Siren & Engine Immobiliser

INSTRUCTION MANUAL FOR. Remote Control Car Alarm with Impact Sensor, Mini-Battery Backup Siren & Engine Immobiliser INSTRUCTION MANUAL FOR Remote Control Car Alarm with Impact Sensor, Mini-Battery Backup Siren & Engine Immobiliser A.C.N 001 621 610 SYDNEY / AUSTRALIA Build Date: TO ARM/DISARM ALARM The alarm is activated

More information

$'(0&2 9,67$ 3DUWLWLRQHG 6HFXULW\ 6\VWHP 8VHU *XLGH N7003V3 5/04 Rev A

$'(0&2 9,67$ 3DUWLWLRQHG 6HFXULW\ 6\VWHP 8VHU *XLGH N7003V3 5/04 Rev A $'(0&29,67$ 3DUWLWLRQHG6HFXULW\6\VWHP 8VHU*XLGH N7003V3 5/04 Rev A 2 TABLE OF CONTENTS SYSTEM OVERVIEW...5 General...5 A Partitioned System...5 Zones...5 Fire Protection...6 Burglary Protection...6 Alarms...6

More information

RK-02 Multi Point Detector Operator s Manual

RK-02 Multi Point Detector Operator s Manual 65-2485RK-02 Multi Point Detector Operator s Manual Part Number: 71-0237RK Revision: A Released: 11/26/14 RKI Instruments, Inc. www.rkiinstruments.com WARNING Read and understand this instruction manual

More information

Contents. Glossary Introduction to the IDS Notes Understanding the Keypad Indicators Operation of the Keypad...

Contents. Glossary Introduction to the IDS Notes Understanding the Keypad Indicators Operation of the Keypad... 2 Contents Glossary...7 1. Introduction to the IDS805...8 1.1 Notes...8 2. Understanding the Keypad Indicators...8 3. Operation of the Keypad...9 4. System Information...10 4.1 Programmed Functions...10

More information

USER MANUAL COMBINED FIRE & EXTINGUISHING CONTROL PANEL COMPLIES WITH BS EN PART 1 & EN54 PARTS 2 & 4

USER MANUAL COMBINED FIRE & EXTINGUISHING CONTROL PANEL COMPLIES WITH BS EN PART 1 & EN54 PARTS 2 & 4 PREMIER EX8 EXTINGUISHING PANEL USER MANUAL COMBINED FIRE & EXTINGUISHING CONTROL PANEL COMPLIES WITH BS EN 12094 PART 1 & EN54 PARTS 2 & 4 USER MANUAL Approved Document No: GLT.MAN-132 INDEX Extinguishing

More information

Contents. Glossary

Contents. Glossary Contents Glossary ------------------------------------------------------------------------------------------------------ 6 1. Introduction to the IDS 1632 -------------------------------------------------------------

More information

IDS816 User Manual H Issued January 2009

IDS816 User Manual H Issued January 2009 1 Contents Glossary-------------------------------------------------------------------------------------------------------------------6 1. Introduction to the IDS 816---------------------------------------------------------------------------7

More information

Section 8 System Operation

Section 8 System Operation Section 8 System Operation Operation of the control panel is simple. Menus guide you step-by-step through operations. This section of the manual is an overview of the operation menus. Please read this

More information

Chlorine gas leak detector

Chlorine gas leak detector INSTRUCTION MANUAL FOR Chlorine gas leak detector with battery back up series M 4510 C 2007 Controlmatik ABW d.o.o. 1 CONTENT 1.0 General description:... 3 2.0 The Unit... 4 3.0 Controls and Displays...

More information

INTELLIGENT FIRE TECHNOLOGY. Twinflex and Multipoint V3. User Guide (TO BE RETAINED BY USER) Issue 3

INTELLIGENT FIRE TECHNOLOGY. Twinflex and Multipoint V3. User Guide (TO BE RETAINED BY USER) Issue 3 INTELLIGENT FIRE TECHNOLOGY Twinflex and Multipoint V3 User Guide (TO BE RETAINED BY USER) 26-0340 Issue 3 Rafiki Protection Limited Rafiki policy is one of continual improvement and the right to change

More information

Algo-Tec 6500/6600 INTERACTIVE ADDRESSABLE FIRE CONTROL SYSTEM

Algo-Tec 6500/6600 INTERACTIVE ADDRESSABLE FIRE CONTROL SYSTEM Algo-Tec 6500/6600 INTERACTIVE ADDRESSABLE FIRE CONTROL SYSTEM (1-4 LOOPS) USER MANUAL Protec Fire Detection plc, Protec House, Churchill Way, Nelson, Lancashire, BB9 6RT, ENGLAND +44 (0) 1282 717171 www.protec.co.uk

More information

VAP304 PRO INSTALLATION & OPERATION MANUAL

VAP304 PRO INSTALLATION & OPERATION MANUAL VAP304 PRO INSTALLATION & OPERATION MANUAL Panic Fire Duress Program Bypass Report Chime Test Memory B A D C Program Chime Panic Duress Exit Fire Report Bypass Memory Test Reset Password Exit Reset Password

More information

THE "SS90" SERIES RELEASE DEVICES MODEL B2 INSTALLATION MANUAL

THE SS90 SERIES RELEASE DEVICES MODEL B2 INSTALLATION MANUAL S S S S S S S S 90 OLID TATE ECURITIES, INC SOLID STATE FAIL-SAFE UNIT * PATENT PENDING LISTED U 99Y9 RESET L RELEASING DEVICE SOLID STATE SECURITIES, INC. THE "SS90" SERIES RELEASE DEVICES MADE IN THE

More information

HMS MODEL 800 Owner s Manual

HMS MODEL 800 Owner s Manual HMS MODEL 800 Owner s Manual OnQ Document Number 1307535 Rev A October, 2000 Copyright v 2000 HAI All Rights Reserved www.onqtech.com Contents INTRODUCTION... 1 Underwriter's Laboratories (UL) Listing...

More information

DATA SHEET CYBERCAT 254 INTELLIGENT FIRE ALARM CONTROL SYSTEM DESCRIPTION APPROVALS: SYSTEM OPERATION. Form No. D

DATA SHEET CYBERCAT 254 INTELLIGENT FIRE ALARM CONTROL SYSTEM DESCRIPTION APPROVALS: SYSTEM OPERATION. Form No. D DATA SHEET CYBERCAT 254 INTELLIGENT FIRE ALARM CONTROL SYSTEM DESCRIPTION Fike s CyberCat 254 (P/N 10-066) is a state-of-the-art true intelligent digital peer-to-peer modular fire alarm control system.

More information

Installation and user manual for the FX range of fire panels. 1, 2, 4 and 8 zone panels

Installation and user manual for the FX range of fire panels. 1, 2, 4 and 8 zone panels Installation and user manual for the FX range of fire panels 1, 2, 4 and 8 zone panels Contents Panel installation 3 Panel connections 3 Wiring connection drawings 4 Panel facilities 6 Installation check

More information

Installation Instructions

Installation Instructions CRSMKKIT002A00 CRSMKSEN002A00 FIELDINSTALLEDSMOKEDETECTORS FOR MEDIUM ROOFTOP UNITS 15 to 25 TONS Installation Instructions TABLE OF CONTENTS PACKAGE CONTENTS... 1 SAFETY CONSIDERATIONS... 1 GENERAL...

More information

Table of Contents. Appendix A Special Characters 31

Table of Contents. Appendix A Special Characters 31 Table of Contents Introduction 2 Section 1: General System Operation 3 1.1 Getting to Know Your System... 3 1.2 How to Arm... 4 1.3 Alternate Arming Methods... 5 1.4 Disarming... 6 1.5 Alarm Memory...

More information

Profile. For a better understanding of this product, please read this user manual thoroughly before using it.

Profile. For a better understanding of this product, please read this user manual thoroughly before using it. Intelligent GSM Auto-Dial Alarm System User s Manual Profile For a better understanding of this product, please read this user manual thoroughly before using it. Contents Function Introduction (3) Alarm

More information

NextGen Home Security. Quick Reference Guide

NextGen Home Security. Quick Reference Guide TM NextGen Home Security Quick Reference Guide Fire Auxiliary Panic (not active) Stay Away Chime Reset Bypass To view the full user manual, go to www.bellaliant.net/homesecurity and select Support. Arming

More information

DS7446KP. User Guide. Keypad

DS7446KP. User Guide. Keypad DS7446KP EN User Guide Keypad DS7446KP User Guide Command Quick Reference Command Quick Reference Command Type Basic Arming Commands Advanced Arming Commands System Disarming Commands Command Turn the

More information

ECONX FIT OFF INSTRUCTIONS

ECONX FIT OFF INSTRUCTIONS ECONX FIT OFF INSTRUCTIONS 0 1. Labelling 2. Fit Off Alarm Sensors 3. Fit Off Smoke Sensors 4. Fit of Siren and Sounder 5. Partitioning, Night Arm and Panic Buttons 6. Fit Off Temperature Sensors 7. Fit

More information

JA-63 Profi User manual

JA-63 Profi User manual JA-63 Profi User manual Contents: 1 Limited warranty... 2 2 Indicators... 3 3 Controlling the system... 4 3.1 Arming... 5 3.2 Disarming... 6 3.3 Panic Alarm... 6 3.4 To stop ALARM... 6 3.5 Home arming...

More information

D7050DH. Installation Instructions. Multiplex Photoelectric Duct Smoke Detector

D7050DH. Installation Instructions. Multiplex Photoelectric Duct Smoke Detector D700DH EN Installation Instructions Multiplex Photoelectric Duct Smoke Detector D700DH Installation Instructions.0 Description Notices These instructions cover the installation of the D700DH Multiplex

More information

Digital Precise Air Control - DPAC

Digital Precise Air Control - DPAC Digital Precise Air Control - DPAC Mode Enable Sensor Options The temperature of this sensor will determine if the unit is in heating, cooling or vent mode during Occupied operation. The following options

More information

EVO192 v3.0 Fire and Burglary What s New

EVO192 v3.0 Fire and Burglary What s New EVO192 v3.0 Fire and Burglary What s New Compatibility: EVO192 v3.0 TM50 v1.31 K641 v2.41 Overview: CP-01 Compliancy Wiring Diagram The following sections/options have been added to the EVO192 panel. They

More information

Car Park Ventilation System. Carbon Monoxide(CO)/

Car Park Ventilation System. Carbon Monoxide(CO)/ Car Park Ventilation System Based on Monitoring of Carbon Monoxide(CO)/ Temperature/ Carbon Dioxide( CO2) Page 1 CONTENTS A. FAQ on car park ventilation system with carbon monoxide (CO) monitoring and

More information

FIRE ALARM SYSTEM 1.1 GENERAL REQUIREMENTS

FIRE ALARM SYSTEM 1.1 GENERAL REQUIREMENTS VALOUR COMMUNITY CENTRE FIRE ALARM SYSTEM Section 28 00 00 GYMNASIUM ADDITION & RENOVATION Page 1 of 3 28 00 00 FIRE ALARM SYSTEM 1.1 GENERAL REQUIREMENTS 1.2 SCOPE 1. The specification covering the General

More information

Digiplex System V2.14 / V2.2ACC. Control Panel Programming Guide

Digiplex System V2.14 / V2.2ACC. Control Panel Programming Guide Digiplex System V2.14 / V2.2ACC Control Panel Programming Guide Table of Contents Getting Started...2 What Do I Do First?...2 How Do I Program the Control Panel?...2 Single Digit Entry Method...2 Multiple

More information

User Manual. with LED Keypad. Alarm Control Panel CA10 plus (software version 4.7) GDAŃSK. ca10pldu_e 07/03

User Manual. with LED Keypad. Alarm Control Panel CA10 plus (software version 4.7) GDAŃSK. ca10pldu_e 07/03 User Manual Alarm Control Panel CA10 plus (software version 4.7) with LED Keypad GDAŃSK ca10pldu_e 07/03 WARNING In order to avoid any operational problems with the control panel, it is recommended that

More information