An Integrative Review

Size: px
Start display at page:

Download "An Integrative Review"

Transcription

1 Monitor Alarm Fatigue An Integrative Review Maria Cvach About the Author Maria Cvach, MS, RN, CCRN, is the assistant director of nursing, clinical standards, at The Johns Hopkins Hospital in Baltimore, MD. Medical devices generate enough false alarms to cause a reduction in responding known as the cry wolf effect. Frequent alarms are distracting and interfere with clinicians performing critical tasks and may lead to staff disabling alarm systems. Abstract Alarm fatigue is a national problem and the number one medical device technology hazard in The problem of alarm desensitization is multifaceted and related to a high false alarm rate, poor positive predictive value, lack of alarm standardization, and the number of alarming medical devices in hospitals today. This integrative review synthesizes research and non-research findings published between 1/1/2000 and 10/1/2011 using The Johns Hopkins Nursing Evidence-Based Practice model. Seventy-two articles were included. Research evidence was organized into five main themes: excessive alarms and effects on staff; nurse s response to alarms; alarm sounds and audibility; technology to reduce false alarms; and alarm notification systems. Non-research evidence was divided into two main themes: strategies to reduce alarm desensitization, and alarm priority and notification systems. Evidence-based practice recommendations and gaps in research are summarized. Background A cacophony of sound echoes through the modern hospital. Bells, beeps, chimes, and horns are all part of the noise-polluted environment that patients, families, and staff endure. They may be exposed to as many as 700 physiologic monitor alarms per patient per day. 1 The myriad of medical device alarms has created an environment that poses significant risk to patient safety. Device alarms are intended to alert clinicians of a hazardous condition and potential problems. However, when a caregiver is subjected to too many alarms, it disrupts his or her usual workflow and may result in errors due to omission, distraction, or inattention. The ECRI Institute, a nonprofit organization that uses applied scientific research in healthcare to establish best practices for improving patient care, publishes an annual top ten technology hazards list. Alarm hazards is the number one health technology hazard for Such hazards include inappropriate alarm modification, alarm fatigue, modifying alarms without restoring them to their original settings, and improperly relaying alarm signals to the appropriate person. 3 The problem of excessive alarms resulting in alarm fatigue has been reported in research literature for many years Studies have indicated that the presence of false and/or clinically insignificant alarms ranges from 80% 99%. 4-6 Medical devices generate enough false alarms to cause a reduction in responding known as the cry wolf effect. 11 Frequent alarms are distracting and interfere with clinicians performing critical tasks and may lead to staff disabling alarm systems. Excessive false positive alarms result in caregiver apathy and desensitization such that real events are less likely to be acted upon. 12,13 National attention to alarm hazards was spurred in 2010 by the death of a patient at Massachusetts General Hospital that was determined to be due to an alarm that had 268 Biomedical Instrumentation & Technology July/August 2012

2 inadvertently been turned off. The federal report indicated that nurses working among constantly beeping monitors contributed to the 14, 15 death of the patient. This integrative review summarizes the current research and non-research evidence available regarding alarm fatigue. Prevalence and Severity of Alarm Fatigue Alarm fatigue, the lack of response due to excessive numbers of alarms resulting in sensory overload and desensitization, is a national problem. 2 From 2005 through 2008, the U.S. Food and Drug Administration (FDA) Manufacturer and User Facility Device Experience (MAUDE) database received 566 reports of patient deaths related to monitoring device alarms. 16 A four-month review of the MAUDE database between March 1, 2010 and June 30, 2010 revealed 73 alarm related deaths with 33 attributed to physiologic monitors. 17 Physiologic monitor alarms are purposefully designed for high sensitivity to not miss a true monitor event. Cardiac monitors use single parameter thresholds that alarm when the set limit is violated. In a multisite study, Chambrin et al., determined the sensitivity and specificity of monitor alarms to be 97% and 58% respectively; positive predictive value was 27%; and negative predictive value was 99%. 6 In addition to high sensitivity, if monitor parameter thresholds are set too tight, true but clinically insignificant alarms may occur. These alarms are known as nuisance alarms. When the alarm is viewed as a nuisance, the caregiver may disable, silence, or ignore the warning that is intended to make the environment safer. Rather than creating a safer environment, a large number of nuisance alarms have an opposite effect, resulting in desensitization. Evidence-Based Practice Model And Search Strategy The Johns Hopkins Nursing Evidence-Based Practice (JHNEBP) model 18 provided an organized approach to appraise, synthesize, and translate evidence for this review. The practice question asked was, Does the amount of noise (false or nuisance alarms) as context to signal (true alarms) interfere with the nurse s response to physiologic monitor alarms? Evidence strength and quality were assessed using the standardized scoring system found Search Terms: PubMed (211), CINAHL (35), Embase (204), Hand searching and seminal articles (7), Clinical alarm OR equipment failure AND Physiologic monitor OR cardiac monitor Figure 1. Search Strategy January 1, 2000 October 1, 2011; 177 abstracts reviewed 85 articles reviewed 72 articles included on the JHNEBP appraisal tools. According to this model, research evidence has the highest strength (level I, II, and III) and non-research evidence the lowest (level IV and V). 18 When the alarm is viewed as a nuisance, the caregiver may disable, silence, or ignore the warning that is intended to make the environment safer. Rather than creating a safer environment, a large number of nuisance alarms have an opposite effect, resulting in desensitization. A comprehensive search of three databases (PubMed, Embase, and the Cumulative Index to Nursing and Allied Health Literature) was conducted. The search was limited to English with a publication date between 1/1/2000 to 10/1/2011 (Figure 1). One hundred seventyseven abstracts were reviewed and 85 articles reviewed in entirety. Seventy-two articles were included in the individual evidence table (available upon request). Biomedical Instrumentation & Technology July/August

3 Research Findings Related to Alarm Fatigue Research evidence was organized into five major themes: 1. Excessive alarms and effects on staff 2. Nurse s response to alarms 3. Alarm sounds and audibility 4. Technology to reduce false alarms 5. Alarm notification systems Excessive Alarms and Effects on Staff Excessive false alarms occur frequently and contribute to alarm desensitization, mistrust, and lack of caregiver response Many false positive alarms are induced and can be attributed to patient manipulation. Motion artifact contributes to excessive false alarms. Staff could avoid false alarms by suspending alarms for a short time Perceived alarm urgency contributes to period prior to patient the nurses alarm response, but nurses manipulation. 5,6 Statistical use additional strategies to determine methods may be suitable response including the criticality of to decrease the number of the patient, signal duration, rarity of false positive monitor alarms. alarming device, and workload. 9 The Healthcare Technology Foundation (HTF) conducted a national online survey of clinicians, engineers, technical staff, and managers in 2006 regarding the effects of alarms. The majority of respondents agreed or strongly agreed that alarms activate too frequently, disrupt patient care, and reduce trust causing caregivers to disable them. 19 Similar results were obtained when the survey was repeated in BI&T Extra To see tables that provide more detailed information about the JHNEBP Research and Non- Research Appraisal Tools, go to Nurses Response to Alarms Perceived alarm urgency contributes to the nurses alarm response, but nurses use additional strategies to determine response including the criticality of the patient, signal duration, rarity of alarming device, and workload A caregiver s probability match is the alarm response based on the perceived true alarm rate. If an alarm system is perceived to be 90% reliable, the response rate will be about 90%; if the alarm system is perceived to be 10% reliable, the response rate will be about 10%. 21 Nurses respond to alarms for different reasons, not just the fact that the alarm sounds. Nurses adjust the order of their activities by evaluating alarm urgency in relation to the patient s condition and have a greater tendency to react to alarms of longer duration and considered rare. 21,23,25 As workload or task complexity increases, alarm response and task performance deteriorates. Thus, signal duration is an important influence, but workload, patient condition, and task complexity may lead to other reaction strategies. 22,25 Adjusting alarms to patient s actual needs ensures that alarms are valid and provides an early warning to potential critical situations. Documenting alarm parameters in the medical record was found to be an effective intervention for improving alarm adjustment compliance. 25 Alarm Sounds and Audibility Humans can discriminate about five to seven different categorical sounds. 27 There is controversy in the literature on the best type of audible alarm sound. IEC is an international standard addressing alarm function and sound. 28 IEC proposes simple melodic alarm sounds to distinguish eight alarm sources and priority codes these sounds as high, medium, or low priority. Some studies have recommended redesign of IEC melodic alarm sounds indicating that the sounds are difficult to identify and cannot be discriminated when there is task overlap. These studies have shown that nurses learning of melodic sounds is poor and that nurses react quicker and more accurately to medium priority alarms despite indicating that high priority alarms sound more urgent. 29,30,31 Phonic abbreviation sounds, which are unintelligible to an untrained listener but recognizable to trained individuals, were studied as a potential type of alarm sound requiring further investigation. 32 Audibility of infusion pump alarms was researched by Sobieraj et al. 33 In this study, the authors concluded that alarms were sufficiently audible and could compete with environmental noise when patient room doors were open. However, audibility was significantly reduced when room doors were closed or during environmental producing noise events such as floor buffing. 33 Thus, it is important to have adjunct alarm notification devices to ensure alarm audibility. Alarms generate noise that may present occupational hazards or hinder patient recovery. 34, 35 Noise levels in most hospitals exceed the World Health Organization (WHO) recommen- 270 Biomedical Instrumentation & Technology July/August 2012

4 dations of 35 decibels (db) during daytime hours and 30 db for nighttime hours. 36 Noise levels 34, 35 have been consistently rising since Hirose et al., studied noise in 75 pieces of medical equipment. The db level of 54% of the equipment studied had a fixed alarm sound and most equipment exceeded 70 db. The authors concluded that alarm db level should be adjusted according to the environmental noise level, and an automatic setting of alarm db level should be set to maximum whenever the device is powered on. 37 This recommendation is in conflict with the findings from Ryherd et al. They recommended more research on the usefulness of visual and vibrating alarm systems, and concluded that noise contributes to staff stress symptoms including fatigue, concentration problems, and tension headaches. 35 Technology to Reduce False Alarms There has been much research over the past 10 years with technology aimed at decreasing false positive alarms and increasing positive predictive value. Researchers have demonstrated that alarms often self-correct. Adding short delays can significantly decrease the number of ignored or ineffective alarms, which are often caused by suctioning, washing, repositioning, and oral care. 38 Rather than using raw data, technology can base alarms on physiologic trends detected over a period of time. Signal filtering, algorithms, and/or artificial intelligence systems process alarms using filters or morphologic and timing differences to reduce the number of alarms. 39,40,64 Smart alarms, which take into account multiple parameters, rate of change, and signal quality can reduce the number of false alarms. 10,41-45,64 Manufacturers of medical devices continue to work on smart alarm technology and alarm acquisition techniques to improve alarm accuracy. 12 Alarm Notification Systems Despite few studies to support the benefit of human monitor surveillance, this alarm management approach is prevalent in many hospitals. A Health Technology Foundation Wireless technologies may be viable alternatives to human monitor surveillance. Comparative studies are needed to determine the best approach to promote positive patient outcomes. Noise levels in most hospitals exceed the World Health Organization (WHO) recommendations of 35 decibels (db) during daytime hours and 30 db for nighttime hours. Noise levels have been consistently rising since survey of 4,278 respondents indicated that central alarm management is viewed as advantageous, and many institutions (48%) use monitor watchers in their institution. 20 Monitor technicians did not show significant differences in mortality or frequency of transfer to critical care. 46 Zwieg et al., compared the use of a monitor technician versus a pager to alert nurses of arrhythmia events. Although false alarms were more frequent in the pager group, the amount of time it took to alert the nurse of an alarm event was less than one minute with both systems. Customization of alarm parameters decreased the false alarm rate thereby making a pager a viable option to arrhythmia notification. 47 Wireless technologies may be viable alternatives to human monitor surveillance. Comparative studies are needed to determine the best approach to promote positive patient outcomes. Research Strengths and Limitations There are a variety of observational research studies available on perception of alarms, alarm urgency response, and alarm fatigue. There are also a significant number of studies addressing the problem of alarm sensitivity. The quality of research studies was good to excellent. There are few randomized controlled trials (RCTs) related to monitor alarm fatigue. The RCT studies available have small sample sizes and are conducted in laboratories with volunteers who may not have the same level of experience as trained healthcare professionals. Most evidence is observational or qualitative with few studies addressing patient outcomes. These studies are limited by known biases introduced by self-report, self-selection, manual data collection, and small sample size. Many of the studies are conducted in single sites resulting in bias and limiting extrapolation of results. Biomedical Instrumentation & Technology July/August

5 Non-Research Evidence Related to Alarm Fatigue Non-research evidence supplements existing research findings as a mechanism for reducing alarm desensitization. Non-research evidence has been divided into two main themes which are summarized below. Strategies to Reduce Alarm Desensitization Clinical standards and expert opinion suggest many strategies to reduce alarm desensitization. If the alarm that is The current mechanism for alarm generation is being generated is based on setting a monitor threshold limit. When an alarm limit is breached, an audible or considered insignificant, visual signal is triggered. Currently, there are no then it should never be standards for setting default alarm parameter activated because the thresholds or graduation of alarms related to most that it can do is degree of urgency. 48 Hospitals need to develop alarm setting and response protocols. provide noise. 12,13,49,50 Technical alarms such as those on electrocardiograms (ECG) leads represent a large number of alarm occurrences. To reduce technical alarms, the ECRI Institute recommends proper skin preparation and replacing ECG leads and electrodes routinely. 13 A quality improvement project conducted on an adult medical progressive care unit and cardiology care unit demonstrated a 46% reduction in total alarms/ pt/day after performing daily electrode change. 64 Hospitals should give considerable thought to alarms that should be activated, default limit parameter settings, and customizing alarms based on the patient s needs. 13 If the alarm that is being generated is considered insignificant, then it should never be activated because the Alarm notification relies on a combination most that it can do is of technical devices and human factors. provide noise. Using a The notification system selected should quality improvement complement the monitoring equipment, approach, Graham and Cvach 49 conducted staffing model, alarm response protocol, small tests of change and unit architectural layout. by altering monitor alarm parameters and limits to actionable levels on a 15-bed medical progressive care unit. During an 18-day period, the baseline number of high priority alarms (16,953) decreased by 43% (9,647 alarms) by eliminating duplicate alarms (for example, heart rate high OR tachycardia but not both) and by setting alarm limits to actionable levels as well as individualizing patient specific parameter limits. 49 Gross et al., found that alarm loads could be controlled with alarm limits set appropriately for the population. Simple limit adjustments from heart rate 120 to 130 bpm would have resulted in a 50% reduction of alarms. 43 To reduce alarms, the Healthcare Technology Foundation recommends initial and ongoing training on alarm-based medical devices that staff are expected to operate. Training should mimic the clinical environment where the device is used. 12 Clinical competency that reflects institutional policy assures care provider skill with physiologic monitoring. 48 Standardizing alarm sounds may also be an effective way to reduce the number of alarms that staff must learn. 50 Animated steps on how to troubleshoot alarms would also be helpful. 57 Alarm Priority and Notification Systems A key aspect of alarm management is assuring that care providers are aware of alarm conditions. Audible alarms are delineated as high, medium, and low priority. High-priority alarms indicate an urgent situation requiring immediate attention; medium-priority alarms indicate a dangerous situation requiring a quick response; and low-priority alarms indicate that attention is needed. 48 An alarm risk assessment, whereby alarms are assigned an alarm priority rating, may be useful when developing alarm policies and determining proper alarm response. 52 Alarm notification relies on a combination of technical devices and human factors. The notification system selected should complement the monitoring equipment, staffing model, alarm response protocol, and unit architectural layout. 13 Basic alarm notification models include on-floor monitoring and remote monitoring. On-floor monitoring may be by direct notification to the care provider or filtered by a unit-based human monitor watch who notifies the assigned caregiver. Remote monitoring involves delivering alarm signals to a location outside of the care unit. 13 Alarm enhancement technology provides additional means to deliver alarm signals from monitors to caregivers. These technologies include auxiliary displays such as marquee signs and waveform screens. 52 The purpose of these displays is to provide additional locations to view alarms on units with long hallways or dispersed geography. 13 Issues of patient confidentiality have arisen with these devices; 272 Biomedical Instrumentation & Technology July/August 2012

6 however, they can be configured without displaying patient names to protect privacy. Integrated middleware systems link alarm systems with wireless devices. These systems route alarms to caregivers and may employ delays and alarm escalation. 54 Use of alarm notification systems that provide context to the care provider and closed-loop communication is recommended. 55,56 Organizations committed to finding solutions have formed interdisciplinary alarm management committees to conduct an alarm risk assessment and explore strategies for alarm reduction. 57 An alarm management policy is essential to define alarm accountability. Alarm data informs proper settings for unit default parameter limits, assists in determining alarm prioritization, and reduces alarm fatigue. Each unit must be analyzed to determine the proper alarm management strategy. It is difficult to apply a one-size-fits-all approach to alarm management for all types of monitored units. Initial and ongoing training on alarming devices is recommended. 58,59,60 Gaps in Knowledge, Need for Further Research There are several areas where more research is indicated. The best type of audible alarm is controversial and needs further investigation. There are no studies on the proper settings for alarm default parameter thresholds. Research is needed on the best way to set monitor limits and levels to improve alarm positive predictive value while not substantially reducing sensitivity. More research is needed on false alarm suppression algorithms. A gap in knowledge exists on the risk/benefit of alarm standardization across like medical devices. Research is needed on whether alarm standardization will improve staff s ability to distinguish device alarms, thereby improving alarm responsiveness. Finally, research is needed on alternate approaches to audible alarm notification as well as effectiveness of wireless technology for alarm notification as compared to human monitor watch. Biomedical Instrumentation & Technology July/August

7 Evidence-based Practice Recommendations To decrease monitor alarm fatigue, the following strategies are recommended: 1. Technology a. Smart alarms, which take into account multiple parameters, rate of change and signal quality, can reduce the number of false alarms. 10,41-45,64 b. Alarm technology that incorporates short delays can decrease the number of ignored or ineffective alarms caused by patient manipulation. 38 c. Standardizing alarm sounds may be an effective way to reduce the number of alarms that staff must learn. 50 d. Animated steps on the monitoring equipment for troubleshooting alarms would be helpful in assuring best practice with equipment Hospital a. Hospitals should engage an interdisciplinary alarm management committee to conduct an alarm risk assessment and explore strategies for alarm reduction. 57 b. Hospitals should develop alarm setting and response protocols. 12,13,49,50 c. Activated alarms should be set to actionable limits and levels. 43,49 d. Staffing model should consider that alarm response time is a function of primary task workload; as workload increases, time to alarm response increases, and alarm task performance gets worse. 22 e. Alarm enhancement technology provides additional means to deliver alarm signals from monitors to caregivers. 52 These technologies may include pagers, phones, and auxiliary displays such as waveform screens. 13 Use of alarm notification systems that provide context to the care provider and closed-loop communication is recommended. 55,56 f. Investment in initial and ongoing training on alarming devices. 58,59,60 Clinical competency that reflects institutional policy assures care provider skill with physiologic monitoring. 48 Training should mimic the clinical environment where the device is used. 12 g. To reduce patient and staff stress symptoms, noise reduction strategies should be employed. 35,36 3. Caregiver a. Staff could avoid false alarms by suspending alarms for a short time period prior to patient manipulation. 5,6 b. Adjustment of alarms to patients actual needs ensures that alarms are valid and provides an early warning to potential critical situations. 47,49 c. Proper skin preparation and replacing ECG leads and electrodes routinely decreases false alarms. 13,60,64 d. Documentation of alarm parameters in the medical record is an effective intervention for improving alarm adjustment compliance. 25 Summary Serious harm and death have occurred from missed alarm events. This integrative review demonstrates the research and non-research findings from the past decade related to monitor alarm fatigue. Gaps in knowledge and need for further research was discussed. Outcomes research, which generates the highest level of evidence, is needed with a focus on patient outcomes rather than just on reduction of the number of alarms. References 1. Cvach M. The Johns Hopkins Hospital Quality Improvement Alarm Data ECRI Institute. Top 10 Health Technology Hazards for Health Devices. 2011;40(11): ECRI Institute. Top 10 Health Technology Hazards for Health Devices. 2010;39 (11): Lawless ST. Crying Wolf: False Alarms in a Pediatric Intensive Care Unit. Critical Care Medicine. 1994;22: Tsien C, Fackler J. Poor Prognosis of Existing Monitors in the Intensive Care Unit. Critical Care Medicine. 1997;25(4): Chambrin MC, Ravaux P, Calvelo-Aros D, Jaborska A, Chopin C, Boniface B. Multicentric Study of Monitoring Alarms in the Adult Intensive Care Unit: a Descriptive Analysis. Intensive Care Med. 1999;25: Atzema C, Schull MJ. Alarmed: Adverse Events in Low-Risk Patients With Chest Pain Receiving Electrocardiographic Monitoring in the Emergency Department: A Pilot Study. American Journal of Emergency Medicine. 2006;24(1): Biomedical Instrumentation & Technology July/August 2012

8 8. Drews FA, Musters A, Markham B, Samore MH. Error-Producing Conditions in the Intensive Care Unit. Proceedings of the Human Factors and Ergonomics Society 51st Annual Meeting. 2007; Siebig S, Kuhls S, Imhoff M, Gather U, Scholmerich J, Wrede C. Intensive Care Unit Alarms How Many Do We Need? Critical Care Medicine. 2010;3(2): Schmid F, Goepfert M S, Kuhnt D, Eichhorn V, Diedrichs S, Reichenspurner H, et al. The Wolf Is Crying in the Operating Room: Patient Monitor and Anesthesia Workstation Alarming Patterns During Cardiac Surgery. Anesthesia and Analgesia. 2011;112(1): Breznitz S. Cry Wolf: The Psychology of False Alarms. New Jersey: Lawrence Erlbaum, Hillsdale, ACCE Health Technology Foundation Impact of Clinical Alarms on Patient Safety. Available at: White%20Paper.pdf. Accessed March 11, ECRI Institute. The Hazards of Alarm Overload: Keeping Excessive Physiological Monitoring Alarms From Impeding Care. Health Devices. Available at: Accessed July 14, Kowalczk L. MGH Death Spurs Review of Patient Monitors. The Boston Globe. Feb. 21, Available at: health/articles/2010/02/21/mgh_death_spurs_review_of_patient_ monitors. Accessed Aug. 17, Kowalczyk L. Alarm Fatigue Linked to Patient Death. The Boston Globe. April 3, Available at: massachusetts/articles/2010/04/03/alarm_fatigue_linked_to_heart_ patients_death_at_mass_general. Accessed July 18, Bustamante EA, Bliss JP, Anderson BL. Effects of Varying the Threshold of Alarm Systems and Workload on Human Performance. Ergonomics. 2007;50(7): Mondor TA, Finley GA. The Perceived Urgency of Auditory Warning Alarms Used in the Hospital Operating Room Is Inappropriate. Canadian Journal of Anesthesia. 2003;50(3): Bliss J, Fallon CK, Nica N. The Role of Alarm Signal Duration as a Cure for Alarm Validity. Applied Ergonomics. 2007;38(2): Solsona JF, Altaba C, Maull E, Rodriguez L, Bosque C, Mulero A. Are Auditory Warnings in the Intensive Care Unit Properly Adjusted? Journal of Advanced Nursing. 2001;35(3): McNeer RR, Bohorquez J, Ozdamar O, Varon AJ, Barach P. A New Paradigm for the Design of Audible Alarms that Convey Urgency Information. Journal of Clinical Monitoring and Computing. 2007;21(6): Block FE Jr. Point-Counterpoint: IEC 62D Medical Monitor Standards Will Make Alarms Worse. Horizons (BI&T suppl). Spring 2011; 60: Wee AN, Sanderson PM. Are Melodic Medical Equipment Alarms Easily Learned? Anesthesia and Analgesia. 2008;106(2): Lacherez P, Seah E, Sanderson P. Overlapping Melodic Alarms Are Almost Indiscriminable. Human Factors. 2007; 49(4): Sanderson P, Wee A, Lacherez P. Learnability and Discriminability of Melodic Medical Equipment Alarms. Anesthesia. 2006;61(2): Food and Drug Administration. Alarming Monitor Problems: Preventing Medical Errors. FDA Patient Safety News. January Available at: transcript.cfm?show=106#7. Accessed July 27, Food and Drug Administration. Manufacturer and User Device Experience (MAUDE) Database: Alarm Related Death Events 3/1/10 6/30/ Newhouse R, Dearholt S, Poe S, Pugh L, White K. Johns Hopkins Nursing Evidence-Based Practice Model and Guidelines. Sigma Theta Tau International; Korniewicz D, Clarke T, David Y. A National Online Survey on the Effectiveness of Clinical Alarms. American Journal of Critical Care. 2008;17(1): Health Technology Foundation. National Clinical Alarms Summary: Perceptions, Issues, Improvements, and Priorities of Healthcare Professionals. Presented at the AAMI Medical Device Alarm Summit. Oct. 4-5, 2011; Herndon, VA. 21. Bitan Y, Meyer J, Shinar D, Zmora E. Nurses Reaction to Alarms in a Neonatal Intensive Care Unit. Cogn Tech Work. 2004;6: Bliss JP, Dunn MC. Behavioral Implications of Alarm Mistrust as a Function of Task Workload. Ergonomics. 2000;43: NOW AVAILABLE! Disinfection and Sterilization: Best Practices in Reprocessing Surgical Instruments Gain the tools and best practices necessary to ensure sterile processing compliance. Preview today! Biomedical Instrumentation & Technology July/August

9 32. Kanagachandra G, Beatty P. Pilot Study: Designing Medical Alarm Sounds for Semantic Association and Ease of Learning. The Internet Journal of Anesthesiology. 2007;12(2). 33. Sobieraj J, Ortega C, West I, Voepel L, Battle S, Robinson D. Audibility of Patient Clinical Alarms to Hospital Nursing Personnel. Military Medicine. 2006;171: Busch-Vishniac I, West J, Barnhill C. Noise Levels in Johns Hopkins Hospital. J Acoust Soc America. 2005;118: Ryherd E, Persson WK, Ljungkvist L. Characterizing Noise and Perceived Work Environment in a Neurological Intensive Care Unit. J. Acoust. Soc. Am. 2008;123(2): West J, Busch-Vishniac I, MacLeod M, Kracht J, Orellano D, Dunn J. Characterizing Noise in Hospitals. Abstract presented at Inter-Noise. Dec. 3-6, 2006; Honolulu, Hawaii. 37. Hirose M, Sato E, Taguchi M, Kokubo K, Kobayashi H, Watanabe S. Characteristics of Auditory Alarms for Medical Equipment and Future Issues. Journal of Clinical Engineering. 2005;30(4): Gorges M, Markewitz BA, Westenskow DR. Improving Alarm Performance in the Medical Intensive Care Unit Using Delays and Clinical Context. Anesthesia and Analgesia. 2009;108(5): Borowski M, Siebig S, Wrede C, Imhoff M. Reducing False Alarms of Intensive Care Online- Monitoring Systems: An Evaluation of Two Signal Extraction Algorithms. Computational and Mathematical Methods in Medicine. doi: /2011/ Blum J M, Kruger GH, Sanders KL, Gutierrez J, Rosenberg AL. Specificity Improvement for Network Distributed Physiologic Alarms Based on a Simple Deterministic Reactive Intelligent Agent in the Critical Care Environment. Journal of Clinical Monitoring & Computing. 2009;23(1): Biot L, Holzapfel L, Becq G, Melot C, Beconnier P. Do We Need a Systematic Activation of Alarm Soundings for Blood Pressure Monitoring for the Advance Your Career With... Career Moves This one CD is packed with more than 150 career-related articles and resources. Learn how to: u Prepare for a certification exam u Improve customer service u Become a manager u Sharpen your skills and learn specialties u Keep up-to-date on IT training and much more A valuable resource for clinical and biomedical engineers and technicians. Order code: CAREER-CD List $150 AAMI member $80 SOURCE CODE: PB To order your copy today, call or visit Biomedical Instrumentation & Technology July/August 2012

10 Safety of ICU Patients? Journal of Critical Care Nurse. 2003;18(4): Burgess LP, Herdman TH, Berg BW, Feaster WW, Hebsur S. Alarm Limit Settings for Early Warning Systems to Identify At-Risk Patients. Journal of Advanced Nursing. 2009;65(9): Gross B, Dahl D, Nielsen L. Physiologic Monitoring Alarm Load on Medical/Surgical Floors of a Community Hospital. Horizons (BI&T suppl). Spring 2011; King A, Roderer A, Arney D, Chen S, Forino- Mullen M, Giannareas A, et al. GSA: A Framework for Rapid Prototyping of Smart Alarm Systems. Proceedings of the 1st ACM International Health Informatics Symposium Otero A, Felix P, Barro S, Palacios F. Addressing the Flaws of Current Critical Alarms: A Fuzzy Constraint Satisfaction Approach. Artificial Intelligence in Medicine. 2009;47(3): Funk M, Parkosewich JA, Johnson CR, Stukshis I. Effects of Dedicated Monitor Watchers on Patients Outcomes. American Journal of Critical Care. 1997;6(4): Zwieg F, Karfonta T, Jeske L, Kollauf C, White S, Drazewski R, Leske J. Arrhythmia Detection and Response in a Monitoring Technician and Pocket Paging System. Progress in Cardiovascular Nursing. 1998;13(1):16-22, Chambrin M. Alarms in the Intensive Care Unit: How Can the Number of False Alarms Be Reduced? Critical Care. 2001;5(4): Graham K, Cvach M. Alarm Fatigue: Standardizing Progressive Care Nurse s Utilization of Physiological Monitors and Decreasing Nuisance Alarms. American Journal of Critical Care Nursing. 2010;19(1): Phillips J. Barnsteiner JH. Clinical Alarms: Improving Efficiency and Effectiveness. Critical Care Nursing Quarterly. 2005;28(4): Edworthy J, Hellier E. Alarms and Human Behaviour:Iimplications for Medical Alarms. British Journal of Anesthesia. 2006;97(1): Phillips J. Clinical Alarms: Complexity and Common Sense. Critical Care Nursing Clinics of North America. 2006;18(2): Alarm Notification for Physiologic Monitoring: Could You Benefit From a New Strategy? Health Devices. 2007;36(1): Dyell D. Beyond Sound: Using Systems Integration to Advance Alarm Functionality. Horizons (BI&T suppl). Spring 2011; Gee T, Moorman B. Reducing Alarm Hazards: Selection and Implementation of Alarm Notification Systems. Patient Safety & Quality Healthcare. March/ April Available at: /799-reducing-alarm-hazards.html. Accessed Oct. 10, Moorman B, Gee T. Functional Basics of Third- Party Alarm Notification Systems. Horizons (BI&T suppl). Spring 2011; Wiklund M, Kendler J. Complementing Medical Device Alarms with Animated Guidance. Horizons (BI&T suppl). Spring 2011; Cvach M, Dang D, Foster J, Irechukwu J. Clinical Alarms and the Impact on Patient Care. Initiatives in Safe Patient Care. Sax Healthcare Communications Available at: www. initiatives-patientsafety.org/initiatives2%20.pdf. Accessed July 14, AAMI. Clinical Alarms Summit. October 4-5, 2011; Herndon, VA. Available at: alarms/index.html. 60. Hazards of Alarm Overload: Keeping Excessive Physiologic Monitoring Alarms From Impeding Care. Health Devices. 2007;36(3): Logan M. Alarm Safety: A Collaborative Effort. Horizons (BI&T suppl). Spring 2011; Keller JP, Diefes R, Graham K, Meyers M, Pelczarski K. Why Clinical Alarms Are a Top Ten Hazard: How You Can Help Reduce the Risk. Horizons (BI&T suppl). Spring 2011; Aboukhalil A, Nielsen L, Saeed M, Mark RG, Clifford GD. Reducing false alarm rates for critical arrhythmias using the arterial blood pressure waveform. Journal of Biomedical Informatics. 2008; doi: /j.jbi Cvach M. The Johns Hopkins Hospital Quality Improvement Data. Presented at the Medical Device Alarm Summit. Oct 5, Herndon, VA. Biomedical Instrumentation & Technology July/August

Running head: REDUCING THE HARM ASSOCIATED WITH CLINICAL ALARMS 1

Running head: REDUCING THE HARM ASSOCIATED WITH CLINICAL ALARMS 1 Running head: REDUCING THE HARM ASSOCIATED WITH CLINICAL ALARMS 1 Reducing the Harm Associated with Clinical Alarm Systems: Meeting the Joint Commission National Patient Safety Goal.06.01.01 Performance

More information

Alarm Fatigue in the ICU. Lynn Maguire MS, RN, PMC, NE-BC

Alarm Fatigue in the ICU. Lynn Maguire MS, RN, PMC, NE-BC Alarm Fatigue in the ICU Lynn Maguire MS, RN, PMC, NE-BC Objectives At the completion of this lecture the participant will: Define Alarm Fatigue Discuss the Joint Commission s Patient Safety Goal regarding

More information

Managing Clinical Alarms

Managing Clinical Alarms Managing Clinical Alarms Maria Cvach, DNP, RN, CCRN Assistant Director of Nursing, Clinical Standards Sharon Allan RN, MSN, ACNS BC, CCRC Clinical Nurse Specialist, CVSICU Matt Trojanowski, MSc, RRT Manager,

More information

Clinical Alarm Effectiveness: Recognizing & Mitigating Risk to Patient Care

Clinical Alarm Effectiveness: Recognizing & Mitigating Risk to Patient Care Clinical Alarm Effectiveness: Recognizing & Mitigating Risk to Patient Care Thomas Bauld, PhD, CCE, Biomedical Engineer National Center for Patient Safety (NCPS) Elena Simoncini, Clinical Engineer VA Boston

More information

You finally make it home after a long shift, shower,

You finally make it home after a long shift, shower, Clinical Alarm Safety and the 2016 HTF Clinical Alarm Survey by Ronda Bradley, MS, RRT, FAARC, and Shawna Strickland, PhD, RRT-NPS, RRT-ACCS, AE-C, FAARC You finally make it home after a long shift, shower,

More information

Why Clinical Alarms are a TOP TEN Hazard

Why Clinical Alarms are a TOP TEN Hazard 1 Why Clinical Alarms are a TOP TEN Hazard James P. Keller, Jr., M.S. Vice President, Health Technology Evaluation and Safety jkeller@ecri.org American Heart Association Conference November 6, 2012 2 An

More information

Impact Of Clinical Alarms On Patient Safety. American College of Clinical Engineering Healthcare Technology Foundation (AHTF)

Impact Of Clinical Alarms On Patient Safety. American College of Clinical Engineering Healthcare Technology Foundation (AHTF) Impact Of Clinical Alarms On Patient Safety American College of Clinical Engineering Healthcare Technology Foundation (AHTF) In 2004, the ACCE Healthcare Technology Foundation started an initiative to

More information

Focus on Respiratory Care & Sleep Medicine Irvine, CA. Alarm Management. September 14-15, 2018

Focus on Respiratory Care & Sleep Medicine Irvine, CA. Alarm Management. September 14-15, 2018 Focus on Respiratory Care & Sleep Medicine Irvine, CA Alarm Management September 14-15, 2018 ECRI INSTITUTE 1 Speaker Information Marc Schlessinger, RRT-NPS, RPFT, MBA, FACHE Senior Associate/Consultant,

More information

Alarm Management. Objectives

Alarm Management. Objectives Alarm Management Michelle Bell, RN, BSN, FISMP, CPPS Kelly Graham, BS, RN Objectives 1. Discuss the importance of appropriate alarm use 2. Describe strategies for improving alarm safety and how to begin

More information

Clinical Alarm Management Strategies Meaningful Alerts

Clinical Alarm Management Strategies Meaningful Alerts Clinical Alarm Management Strategies Meaningful Alerts Sharon H. Allan DNP, RN, ACNS-BC, CCRC Johns Hopkins School of Nursing Johns Hopkins Hospital Baltimore, MD January 12, 2018 1 Disclosures I have

More information

Alarm Management at Boston Medical Center A Roadmap to Safe Silence

Alarm Management at Boston Medical Center A Roadmap to Safe Silence Alarm Management at Boston Medical Center A Roadmap to Safe Silence Jim Piepenbrink, Director Department of Clinical Engineering Co-Chair, Clinical Alarm Task Force This presentation is a view into the

More information

2011 National Clinical Alarms Survey: Perceptions, Issues, Improvements and Priorities

2011 National Clinical Alarms Survey: Perceptions, Issues, Improvements and Priorities 2011 National Clinical Alarms Survey: Perceptions, Issues, Improvements and Priorities Healthcare Technology Foundation Clinical Alarms Task Force Presented by Tobey Clark President, Healthcare Technology

More information

Safety Innovations FOUNDATIONHTSI. Using Data to Drive Alarm System Improvement Efforts. The Johns Hopkins Hospital Experience

Safety Innovations FOUNDATIONHTSI. Using Data to Drive Alarm System Improvement Efforts. The Johns Hopkins Hospital Experience FOUNDATIONHTSI Healthcare Technology Safety Institute Safety Innovations Using Data to Drive Alarm System Improvement Efforts The Johns Hopkins Hospital Experience The key to reducing alarm signal noise

More information

Reference Case: Avera Heart Hospital Data-Driven Approach Helps Hospital Build Effective Clinical Alarm Management Program

Reference Case: Avera Heart Hospital Data-Driven Approach Helps Hospital Build Effective Clinical Alarm Management Program Reference Case: Avera Heart Hospital Data-Driven Approach Helps Hospital Build Effective Clinical Alarm Management Program Imagine that your home security system sounded an alarm at random intervals and

More information

Improving Alarm Management Utilizing Technology. NJ/DVHIMSS Annual Fall Event. Empowering the New Healthcare Era

Improving Alarm Management Utilizing Technology. NJ/DVHIMSS Annual Fall Event. Empowering the New Healthcare Era Improving Alarm Management Utilizing Technology NJ/DVHIMSS Annual Fall Event Empowering the New Healthcare Era 1 2 How big a deal is this? Up to 350 alarms per day per patient! Ranks #1 on ECRI s 2014

More information

Improving Alarm Response in ICU/CCU

Improving Alarm Response in ICU/CCU Improving Alarm Response in ICU/CCU Sasikala Thangavelu Faculty of Science and Technology Plymouth University Plymouth,England sasikala@mdb.gov.my Emmanuel Ifeachor Faculty of Science and Technology Plymouth

More information

Alarms That Deserve Attention. Mike Rayo 3/29/2016

Alarms That Deserve Attention. Mike Rayo 3/29/2016 Alarms That Deserve Attention Mike Rayo 3/29/2016 What I ll be covering Tackling Alarm Fatigue With Sound Design With Policy By Sharpening With Multimodal Design 2 Can t identify alarm Startle effect of

More information

Original Article. Intra-operative monitoring many alarms with minor impact. Summary

Original Article. Intra-operative monitoring many alarms with minor impact. Summary Original Article doi:10.1111/anae.12289 Intra-operative monitoring many alarms with minor impact F. R. de Man, 1 S. Greuters, 2 C. Boer, 3 D. P. Veerman 4 and S. A. Loer 5 1 Clinical Research Physician,

More information

Hospital Ward Alarm Fatigue Reduction Through Integrated Medical Device Instruction and Hospital System Policy

Hospital Ward Alarm Fatigue Reduction Through Integrated Medical Device Instruction and Hospital System Policy Virginia Commonwealth University VCU Scholars Compass VCU da Vinci Center Student Works VCU da Vinci Center 2014 Hospital Ward Alarm Fatigue Reduction Through Integrated Medical Device Instruction and

More information

Customizing Alarm Limits Based on Specific Needs of Patients

Customizing Alarm Limits Based on Specific Needs of Patients Customizing Alarm Limits Based on Specific Needs of Patients Maria Cvach, Melanie Kitchens, Kevin Smith, Patricia Harris, and Marilyn Neder Flack Medical device alarms are frequent, alerting for false

More information

AST Guidelines for Best Practices in Alarm Management in the Operating Room

AST Guidelines for Best Practices in Alarm Management in the Operating Room 1 Approved October 7, 2016 Revised April 14, 2017 AST Guidelines for Best Practices in Alarm Management in the Operating Room Introduction The following Guidelines for Best Practices were researched and

More information

The Joint Commission's National Patient Safety Goal on Alarm Management: How Do We Get Started?

The Joint Commission's National Patient Safety Goal on Alarm Management: How Do We Get Started? The Joint Commission's National Patient Safety Goal on Alarm Management: How Do We Get Started? September 25, 2013 9/25/2013 1 Healthcare Technology Safety Institute HTSI Brings together multidisciplinary

More information

Safety Innovations FOUNDATIONHTSI. Cardiopulmonary Monitors And Clinically Significant Events in Critically Ill Children

Safety Innovations FOUNDATIONHTSI. Cardiopulmonary Monitors And Clinically Significant Events in Critically Ill Children FOUNDATIONHTSI Healthcare Technology Safety Institute Safety Innovations Cardiopulmonary Monitors And Clinically Significant Events in Critically Ill Children Children s National Medical Center A team

More information

Capnostream 20p Bedside Capnography Monitor. The only integrated, complete picture of oxygenation and ventilation 1

Capnostream 20p Bedside Capnography Monitor. The only integrated, complete picture of oxygenation and ventilation 1 Capnostream 20p Bedside Capnography Monitor The only integrated, complete picture of oxygenation and ventilation 1 Smart Capnography and Pulse Oximetry Technology Smart Capnography is a suite of algorithms

More information

"It's Alarming Simple Steps To Reduce Nuisance Alarms

It's Alarming Simple Steps To Reduce Nuisance Alarms "It's Alarming Simple Steps To Reduce Nuisance Alarms Lisa Pahl, MSN, Principal, Practice Lead Alarm Management Healthcare Transformation Services May, 2017 1 Objective Patients and their families want

More information

Contents of ANSI/AAMI HE75:2009 Page

Contents of ANSI/AAMI HE75:2009 Page Contents of ANSI/AAMI HE75:2009 Page Glossary of equivalent standards...xiii Committee representation...xv Acknowledgments...xvi Foreword...xvii Introduction...1 1 Scope...5 1.1 General...5 1.2 Inclusions...5

More information

False Alarm Suppression in Early Prediction of Cardiac Arrhythmia

False Alarm Suppression in Early Prediction of Cardiac Arrhythmia IEEE International Conference on Bioinformatics & Bioengineering 2015 False Alarm Suppression in Early Prediction of Cardiac Arrhythmia Shoumik Roychoudhury, Mohamed F. Ghalwash, Zoran Obradovic Presented

More information

Technologic advances in hospitals have

Technologic advances in hospitals have in Safe Patient Care Enhancing patient safety through improved surveillance Clinical alarms warn caregivers of immediate or potential adverse patient conditions. Alarms and their shortcomings have been

More information

No Cause for Alarm: A Holistic Approach to Identification, Prioritization, and Reduction

No Cause for Alarm: A Holistic Approach to Identification, Prioritization, and Reduction No Cause for Alarm: A Holistic Approach to Identification, Prioritization, and Reduction Pascha E. Schafer, MD, FACC Medical Director, Cardiac Care Unit Medical College of Georgia at Georgia Regents University

More information

The Complexity of Clinical Alarm Systems

The Complexity of Clinical Alarm Systems HTM NEWS & VIEWS The Complexity of Clinical Alarm Systems Alan Lipschultz About the Author Alan Lipschultz, CCE, PE,CSP, is president of HealthCare Technology Consulting LLC in North Bethesda, MD. E-mail:

More information

2013 Compliance Report RCW 52.33

2013 Compliance Report RCW 52.33 2013 Compliance Report RCW 52.33 Revised Code of Washington (RCW) 52.33 requires substantially career fire protection districts to declare their department as established and set performance standards

More information

Regulation of Alarm Remote Notification Systems

Regulation of Alarm Remote Notification Systems Regulation of Alarm Remote Notification Systems Shawn Forrest Center for Devices and Radiological Health Food and Drug Administration Medical Device Alarms Summit October 5, 2011 Evolving Clinical Environment

More information

GE Healthcare. Dash Variable-Acuity Monitoring

GE Healthcare. Dash Variable-Acuity Monitoring GE Healthcare Dash Variable-Acuity Monitoring The RIGHT information. Right now. In your busy schedule, having accurate patient information at your fingertips is critical. And there is no time for false

More information

ABB Ability System 800xA Alarm Management

ABB Ability System 800xA Alarm Management ABB Ability System 800xA Alarm Management Effective detection, notification, analysis and corrective action of abnormal situations. When an alarm sounds in a process controlled by ABB s System 800xA, operators

More information

URGENT: MEDICAL DEVICE CORRECTION Plum 360 Infusion System List Number: 30010

URGENT: MEDICAL DEVICE CORRECTION Plum 360 Infusion System List Number: 30010 . 600 N. Field Drive Lake Forest, IL 60045 (224) 212 2000 www.icumed.com 15 June 2017 URGENT: MEDICAL DEVICE CORRECTION Plum 360 Infusion System List Number: 30010 Dear Valued Plum 360 Infusion System

More information

Clinical Alarms Summit. Convened by AAMI, FDA, TJC, ACCE, and ECRI Institute AAMI

Clinical Alarms Summit. Convened by AAMI, FDA, TJC, ACCE, and ECRI Institute AAMI Clinical Alarms 2011 Summit Convened by AAMI, FDA, TJC, ACCE, and ECRI Institute AAMI Clinical Alarms Summit Conveners AAMI The Association for the Advancement of Medical Instrumentation (AAMI), a nonprofit

More information

AUDIBLE ALARMS AND WORKPLACE NOISE IN NURSING UNITS: AN ENVIRONMENTAL FACTORS CASE STUDY

AUDIBLE ALARMS AND WORKPLACE NOISE IN NURSING UNITS: AN ENVIRONMENTAL FACTORS CASE STUDY AUDIBLE ALARMS AND WORKPLACE NOISE IN NURSING UNITS: AN ENVIRONMENTAL FACTORS CASE STUDY By Rudolph P. Santacroce, P.E., Shands HealthCare Abstract Alarms, especially on inpatient units, deserve special

More information

Alarm Safety in 2017:

Alarm Safety in 2017: Alarm Safety in 2017: Moving Beyond ECG Monitors S. Jill Ley RN, MS, CNS, FAAN Surgical Services CNS; California Pacific Medical Center Clinical Professor, UCSF; San Francisco, CA Learning Outcome Discuss

More information

2017 CIDA Standards Infused into Undergraduate Courses

2017 CIDA Standards Infused into Undergraduate Courses 2017 CIDA Standards Infused into Undergraduate Courses Department of Interior Architecture. University of North Carolina at Greensboro February 2016 101, 102 Environmental Design I, II (4:0:8), (4:0:8)

More information

The Contra Costa EMS System and Fire Station Closures: Impact and Mitigation

The Contra Costa EMS System and Fire Station Closures: Impact and Mitigation This is an EMS System response report for the communities of Clayton, Lafayette, Martinez and Walnut Creek. Each community experienced a fire station closure on January 15, 2013. The Contra Costa EMS System

More information

Unwanted Fire Alarms. Marty Ahrens National Fire Protection Association 1 Batterymarch Park Quincy, MA (617)

Unwanted Fire Alarms. Marty Ahrens National Fire Protection Association 1 Batterymarch Park Quincy, MA (617) Unwanted Fire Alarms Marty Ahrens National Fire Protection Association 1 Batterymarch Park Quincy, MA 02169-7471 (617) 984-7463 mahrens@nfpa.org Unwanted fire alarms, however they are defined, are a problem

More information

Community Development and Recreation Committee

Community Development and Recreation Committee STAFF REPORT ACTION REQUIRED Results of the Fire Insurance Grades Study Date: November 20, 2013 To: From: Wards: Reference Number: Community Development and Recreation Committee Fire Chief and General

More information

FFT Functional Family Therapy Evidence Based. Cost Effective. Sustainable. Family, Youth & Culture Sensitive.

FFT Functional Family Therapy Evidence Based. Cost Effective. Sustainable. Family, Youth & Culture Sensitive. FFT Functional Family Therapy Evidence Based. Cost Effective. Sustainable. Family, Youth & Culture Sensitive. Frequently Asked Questions Configuration of FFT 1. How many therapists can one supervisor support

More information

Kendall DL Cable and Lead Wire System. Your Disposable ECG Cable and Lead Wire Safety Solution. Safe, Simple, Secure

Kendall DL Cable and Lead Wire System. Your Disposable ECG Cable and Lead Wire Safety Solution. Safe, Simple, Secure Kendall DL Cable and Lead Wire System Your Disposable ECG Cable and Lead Wire Safety Solution Safe, Simple, Secure Covidien: Cardiac Monitoring Supply Solutions Market Leader Covidien is a market leader

More information

Bernoulli Smart Alarms & Surveillance: Applying Analytics in Real-Time

Bernoulli Smart Alarms & Surveillance: Applying Analytics in Real-Time Bernoulli Smart Alarms & Surveillance: Applying Analytics in Real-Time Introduction Patient care devices (PCDs) produce alarms base on pre-defined s that are established as defaults within the individual

More information

BeneVision Central Monitoring System. Quick Reference Guide

BeneVision Central Monitoring System. Quick Reference Guide BeneVision Central Monitoring System Quick Reference Guide NOTE: The Quick Reference Guide is not intended as a replacement to the Operating Instructions. Prior to operating equipment, the user must be

More information

Clinical Alarm Management. Compendium

Clinical Alarm Management. Compendium Clinical Alarm Management Compendium About the AAMI Foundation The AAMI Foundation is a 501(c) (3) organization, whose mission is to promote the safe adoption and safe use of healthcare technology. The

More information

LASER SAFETY TRAINING CATALOG LASER LASER INSTITUTE OF AMERICA S MAKE LASER SAFETY YOUR PRIORITY.

LASER SAFETY TRAINING CATALOG LASER LASER INSTITUTE OF AMERICA S MAKE LASER SAFETY YOUR PRIORITY. MAKE LASER SAFETY YOUR PRIORITY LASER INSTITUTE OF AMERICA S LASER SAFETY TRAINING CATALOG www.lia.org 13501 Ingenuity Dr., Suite 128 32826 Phone: +1.407.380.1553 Fax: +1.407.380.5588 INDUSTRIAL/RESEARCH

More information

A Novel Approach to Cardiac Alarm Management on Telemetry Units. Medical Device Alarm Safety in Hospitals Joint Commission Webinar May 1, 2013

A Novel Approach to Cardiac Alarm Management on Telemetry Units. Medical Device Alarm Safety in Hospitals Joint Commission Webinar May 1, 2013 A Novel Approach to Cardiac Alarm Management on Telemetry Units Patricia Covelle, RN, MMHC Director of Critical Care Nursing James Piepenbrink, BS Director, Department of Clinical Engineering Deborah Whalen

More information

EasyChair Preprint. Measuring Alarm System Quality in ICUs (WIP)

EasyChair Preprint. Measuring Alarm System Quality in ICUs (WIP) EasyChair Preprint 114 Measuring Alarm System Quality in ICUs (WIP) Dirk Hüske-Kraus, Marc Wilken and Rainer Röhrig EasyChair preprints are intended for rapid dissemination of research results and are

More information

Battery Performance Alert: A TOOL FOR IMPROVED PATIENT MANAGEMENT FOR DEVICES UNDER BATTERY ADVISORY

Battery Performance Alert: A TOOL FOR IMPROVED PATIENT MANAGEMENT FOR DEVICES UNDER BATTERY ADVISORY Battery Performance Alert: A TOOL FOR IMPROVED PATIENT MANAGEMENT FOR S UNDER BATTERY ADVISORY VERSION 1.0 AUGUST 8, 2017 Abstract: BACKGROUND: In October 2016, St. Jude Medical issued an advisory on a

More information

Australian/New Zealand Standard

Australian/New Zealand Standard AS/NZS 3200.2.4:2006 IEC 60601-2-4:2005 AS/NZS 3200.2.4:2006 Australian/New Zealand Standard Medical electrical equipment Part 2.4: Particular requirements for safety Cardiac defibrillators AS/NZS 3200.2.4:2006

More information

Track the truth. Give no false assurance. Do no harm. Nellcor SpO 2. Module for use with Philips IntelliVue * Patient Monitors

Track the truth. Give no false assurance. Do no harm. Nellcor SpO 2. Module for use with Philips IntelliVue * Patient Monitors Track the truth Give no false assurance Do no harm Nellcor Module for use with Philips IntelliVue * Patient Monitors Nellcor Module Features Designed to be used in a range of critical care environments

More information

Alarm Management for SCADA control rooms

Alarm Management for SCADA control rooms Whitepaper Alarm Management for SCADA control rooms Definition of the term alarm The concept of alarms (French for a l arme which means spring to arms ) is very old and originates from the military concept

More information

Alarm System Performance Metrics

Alarm System Performance Metrics Alarm System Performance Metrics FPID Symposium, Cork March 2016 Standards Certification Education & Training Publishing Conferences & Exhibits Presenter Kim Van Camp is the Emerson Process Management

More information

Fire Safety Policy SH HS 06. Version: 4. Summary:

Fire Safety Policy SH HS 06. Version: 4. Summary: SH HS 06 Fire Safety Policy Summary: Keywords: Target Audience: The Fire Safety Policy sets out the Trusts approach to a proactive fire safety culture to protect its staff, patients, service users, all

More information

Hospital Preparedness: A Pilot Program for Radiation Monitoring in Emergency Departments

Hospital Preparedness: A Pilot Program for Radiation Monitoring in Emergency Departments Hospital Preparedness: A Pilot Program for Radiation Monitoring in Emergency Departments Susan E. Eckert, RN, MSN Washington Hospital Center/ ER One Institute CAPT Michael A. Noska, MS, USPHS Dept. of

More information

Ensuring the Effectiveness of Clinical Alarm Systems

Ensuring the Effectiveness of Clinical Alarm Systems Ensuring the Effectiveness of Clinical Alarm Systems Requirements: Ensure that critical alarms used in the clinical environment are tested on an ongoing, scheduled basis and they are sufficiently audible

More information

80BPM 36.5 C 35/ / 80. Panorama CENTRAL MONITORING SYSTEM. efficient flexible versatile (18) (94) Adult A SMITH, ERICA (92) 5RPM.

80BPM 36.5 C 35/ / 80. Panorama CENTRAL MONITORING SYSTEM. efficient flexible versatile (18) (94) Adult A SMITH, ERICA (92) 5RPM. Panorama CENTRAL MONITORING SYSTEM 5RPM 78 (92) 80BPM SMITH, ERICA mmhg 35/ 20 (18) 36.5 C mmhg Adult 1427 - A 120/ 80 (94) 1 mmhg efficient flexible versatile Focused on clinical efficiency, the Panorama

More information

Safety Innovations FOUNDATIONHTSI. Plan, Do, Check, Act: Using Action Research to Manage Alarm Systems, Signals, and Responses

Safety Innovations FOUNDATIONHTSI. Plan, Do, Check, Act: Using Action Research to Manage Alarm Systems, Signals, and Responses FOUNDATIONHTSI Healthcare Technology Safety Institute Safety Innovations Plan, Do, Check, Act: Using Action Research to Manage Alarm Systems, Signals, and Responses The Beth Israel Deaconess Medical Center

More information

ENHANCED ODOR-DETECTION SYSTEMS. A Patented System Developed by K-9 Search on Site (K-9 SOS)

ENHANCED ODOR-DETECTION SYSTEMS. A Patented System Developed by K-9 Search on Site (K-9 SOS) ENHANCED ODOR-DETECTION SYSTEMS A Patented System Developed by K-9 Search on Site (K-9 SOS) The canine is better able to detect target odors than any other method known and EODS takes it to a whole new

More information

Battery Performance Alert

Battery Performance Alert Battery Performance Alert A TOOL FOR IMPROVED PATIENT MANAGEMENT FOR S UNDER BATTERY ADVISORY VERSION 2.0 Abstract BACKGROUND: In October 2016, St. Jude Medical (now Abbott) issued an advisory on a family

More information

Remote Monitoring of Offshore Structures using Acoustic Emission.

Remote Monitoring of Offshore Structures using Acoustic Emission. 11th European Conference on Non-Destructive Testing (ECNDT 2014), October 6-10, 2014, Prague, Czech Republic Remote Monitoring of Offshore Structures using Acoustic Emission. More Info at Open Access Database

More information

1. EXPLORE A CAREER PLAN IN DESIGN AND MERCHANDISING

1. EXPLORE A CAREER PLAN IN DESIGN AND MERCHANDISING 21 st Century Learning Interior Design and Merchandising High School Interior Design and Merchandising is an elective that focuses on creating beauty, comfort and convenience in rooms and homes. The principles

More information

Framework for Alarm Management Process Maturity

Framework for Alarm Management Process Maturity Framework for Alarm Management Process Maturity James Welch, Michael Rayo, Benjamin Kanter, Tandi Bagian, Katrina Jacobs, Hassan Shanawani, Rory Jaffe, Marlyn Conti, and Lynn Razzano Editor s note: The

More information

Why Building Occupants Ignore Fire Alarms

Why Building Occupants Ignore Fire Alarms Page 1 of 6 Français Contact Us Help Search Canada Site Publications FAQ Business Research Activities Product Evaluations Codes and Guides NRC Site Home Subject Index Home > Publications > Construction

More information

Scope of Work Urban Design Review Framework Monitoring Program Item #7.1

Scope of Work Urban Design Review Framework Monitoring Program Item #7.1 1.0 Background The Urban Design Review Framework (Framework) project (PUD2017-0601) explored design review protocols and processes, and proposed revisions to achieve better built outcomes. In 2017 July,

More information

Ovation Alarm Management System

Ovation Alarm Management System Ovation Alarm Management System Features Multiple types of fully configurable, easy-tounderstand alarm displays Convenient access to past alarms, returns and state changes Simplified ability to acknowledge,

More information

THE APPLICATION OF THE COMMON DESIGN RECOMMENDATIONS (CDR) IN ASSESSING RESTORATIVE GREEN OUTDOOR ENVIRONMENTS

THE APPLICATION OF THE COMMON DESIGN RECOMMENDATIONS (CDR) IN ASSESSING RESTORATIVE GREEN OUTDOOR ENVIRONMENTS THE APPLICATION OF THE COMMON DESIGN RECOMMENDATIONS (CDR) IN ASSESSING RESTORATIVE GREEN OUTDOOR ENVIRONMENTS Siti Nur Fathirah Adnan 1 and Shureen Faris Abd Shukor 2 Faculty of Design and Architecture,

More information

Preventing ECMO Patient Harm While Using Remote Monitoring: A Case Report. Emily Thunstrom

Preventing ECMO Patient Harm While Using Remote Monitoring: A Case Report. Emily Thunstrom Preventing ECMO Patient Harm While Using Remote Monitoring: A Case Report Emily Thunstrom Disclosure I have no disclaimers Outline Alarms Perfusion alarm management ECMO Alarms and Rounding Case Report

More information

Acoustics in open-plan offices A laboratory study

Acoustics in open-plan offices A laboratory study Acoustics in open-plan offices A laboratory study Petra Larm, Jukka Keränen, Riikka Helenius, Jarkko Hakala, Valtteri Hongisto Finnish Institute of Occupational Health, Lemminkäisenkatu 14-18 B, 20720

More information

FORT MYERS BEACH FIRE DEPARTMENT JOB DESCRIPTION

FORT MYERS BEACH FIRE DEPARTMENT JOB DESCRIPTION This Position Description is established by the Fort Myers Beach Fire Department ( Department ) to outline the basic requirements, duties, and general responsibilities of the Deputy Fire Marshal, Fire

More information

ELECTRICAL SAFETY IN HEALTHCARE FACITILIES

ELECTRICAL SAFETY IN HEALTHCARE FACITILIES ELECTRICAL SAFETY IN HEALTHCARE FACITILIES BY ENGR. S.M. NASEERUDDIN Dar Al-Majd Consulting Engineers Riyadh, Saudi Arabia ABSTRACT The increasing use of electrical diagnostic and treatment equipment in

More information

June 17, Skanray Technologies Pvt Ltd. c/o Ms. Yolanda Smith Regulatory Consultant Smith Associates 1468 Harwell Avenue Crofton, Maryland 21114

June 17, Skanray Technologies Pvt Ltd. c/o Ms. Yolanda Smith Regulatory Consultant Smith Associates 1468 Harwell Avenue Crofton, Maryland 21114 DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 June 17, 2015 Skanray

More information

LASER SAFETY TRAINING CATALOG LASER INSTITUTE OF AMERICA S MAKE LASER SAFETY YOUR PRIORITY LASER

LASER SAFETY TRAINING CATALOG LASER INSTITUTE OF AMERICA S MAKE LASER SAFETY YOUR PRIORITY LASER MAKE LASER SAFETY YOUR PRIORITY LASER INSTITUTE OF AMERICA S LASER SAFETY TRAINING CATALOG 1.800.34.LASER www.lia.org 13501 Ingenuity Dr., Suite 128 Orlando, FL 32826 Phone: +1.407.380.1553 Fax: +1.407.380.5588

More information

INTEGRATED LIFE SAFETY COMPLIANCE TESTING AND ITS PROCEDURES

INTEGRATED LIFE SAFETY COMPLIANCE TESTING AND ITS PROCEDURES INTEGRATED LIFE SAFETY COMPLIANCE TESTING AND ITS PROCEDURES By: Mehrdad Ghods Integrated Engineering Student Sept. 06, 2002 University of British Columbia Vancouver, BC i PREFACE: The purpose of this

More information

Effective Alarm Management for Dynamic and Vessel Control Systems

Effective Alarm Management for Dynamic and Vessel Control Systems DYNAMIC POSITIONING CONFERENCE October 12-13, 2010 OPERATIONS SESSION Effective Alarm Management for Dynamic and Vessel Control Systems By Steve Savoy Ensco Offshore Company 1. Introduction Marine control

More information

Healthcare PROTEZIONE TOTALE PER I VOSTRI INTERESSI AZIENDALI PRESENTI E FUTURI. When every second counts, Honeywell delivers.

Healthcare PROTEZIONE TOTALE PER I VOSTRI INTERESSI AZIENDALI PRESENTI E FUTURI. When every second counts, Honeywell delivers. Healthcare PROTEZIONE TOTALE PER I VOSTRI INTERESSI AZIENDALI PRESENTI E FUTURI When every second counts, Honeywell delivers. Your Challenges are Complex. Honeywell Can Help. Evolving regulations, greater

More information

Optimum Room Acoustic ComfortTM (RACTM) can be achieved by using a selection of appropriate acoustic descriptors

Optimum Room Acoustic ComfortTM (RACTM) can be achieved by using a selection of appropriate acoustic descriptors Optimum Room Acoustic ComfortTM (RACTM) can be achieved by using a selection of appropriate acoustic descriptors C. Svensson and E. Nilsson Saint-Gobain Ecophon AB, Box 500, SE-260 61 Hyllinge, Sweden

More information

Diagnostics with fieldbus

Diagnostics with fieldbus 2002 Emerson Process Management. All rights reserved. View this and other courses online at www.plantwebuniversity.com. Fieldbus 105 Diagnostics with fieldbus Overview More than device maintenance Equipment

More information

The Mode of Urban Renewal Base on the Smart City Theory under the Background of New Urbanization

The Mode of Urban Renewal Base on the Smart City Theory under the Background of New Urbanization Frontiers of Engineering Management DOI 10.15302/J-FEM-2015035 ENGINEERING MANAGEMENT THEORIES AND METHODOLOGIES Yi-hua Mao, Hong-yu Li, Qin-rui Xu The Mode of Urban Renewal Base on the Smart City Theory

More information

MSU Landscape Architecture Program

MSU Landscape Architecture Program The Importance of Nature to Health & Well-being Dr. Joanne Westphal, PhD,, M.D. Background Medicine Landscape Architecture Background Medicine effects of nature on human health» Conscious sensory sensory

More information

OMNI TOUCH SCREEN PATIENT MONITOR

OMNI TOUCH SCREEN PATIENT MONITOR OMNI TOUCH SCREEN PATIENT MONITOR OMNI Intuitive Designed for a fast paced work environment, the Infinium Omni patient monitor offers an extremely simple and adaptable user interface. Patient information

More information

AS/NZS IEC :2016

AS/NZS IEC :2016 Australian/New Zealand Standard AS/NZS IEC 60601.2.25:2016 (IEC 60601-2-25, Ed. 2.0:2011, IDT) Medical electrical equipment Part 2.25: Particular requirements for the basic safety and essential performance

More information

LASER SAFETY TRAINING CATALOG LASER INSTITUTE OF AMERICA S MAKE LASER SAFETY YOUR PRIORITY LASER

LASER SAFETY TRAINING CATALOG LASER INSTITUTE OF AMERICA S MAKE LASER SAFETY YOUR PRIORITY LASER MAKE LASER SAFETY YOUR PRIORITY LASER INSTITUTE OF AMERICA S LASER SAFETY TRAINING CATALOG 1.800.34.LASER www.lia.org 13501 Ingenuity Dr., Suite 128 Orlando, FL 32826 Phone: +1.407.380.1553 Fax: +1.407.380.5588

More information

Document Title: Management of Research and Development Freezers

Document Title: Management of Research and Development Freezers Document Title: Management of Research and Development Document Number: SOP029 Staff involved in development: Job titles only Document author/owner: Directorate: Department: For use by: RM&G Manager, R&D

More information

MATRIX TO DETERMINE YOUR PRESENT POSITION IN MEETING N.F.P.A. 1720

MATRIX TO DETERMINE YOUR PRESENT POSITION IN MEETING N.F.P.A. 1720 MATRIX TO DETERMINE YOUR PRESENT POSITION IN MEETING N.F.P.A. 1720 Standard for the Organization and Deployment of Fire Suppression Operations, Emergency Medical Operations, and Special Operations to the

More information

DESIGN FOR ELDERLY EGRESS IN FIRE SITUATIONS

DESIGN FOR ELDERLY EGRESS IN FIRE SITUATIONS DESIGN FOR ELDERLY EGRESS IN FIRE SITUATIONS Lauren Folk i, John Gales ii, Steven Gwynne iii and Michael Kinsey iv i Research Student, Architectural Conservation and Sustainability Engineering, Carleton

More information

6 th Pipeline Technology Conference 2011

6 th Pipeline Technology Conference 2011 6 th Pipeline Technology Conference 2011 Pipeline Leak Detection and Theft Detection Using Rarefaction Waves Authors: Dr Alex Souza de Joode, VP International Operations; ATMOS International, UK. Andrew

More information

BUSINESS PLAN CEN/TC 215 RESPIRATORY AND ANAESTHETIC EQUIPMENT EXECUTIVE SUMMARY

BUSINESS PLAN CEN/TC 215 RESPIRATORY AND ANAESTHETIC EQUIPMENT EXECUTIVE SUMMARY Page: 1 BUSINESS PLAN CEN/TC 215 RESPIRATORY AND ANAESTHETIC EQUIPMENT Business environment EXECUTIVE SUMMARY The following European Council New Approach directive and regulation are relevant to the work

More information

Civic Nature A Competitive Advantage

Civic Nature A Competitive Advantage Civic Nature A Competitive Advantage Dr. Kathleen Wolf Research Social Scientist University of Washington College of the Environment January 2010 Trees & Change hope, peace & inspiration 1 Trees & Ethics

More information

White Paper. Critical Considerations for Monitoring Low Levels of Hydrogen Sulfide. by Jacob Spector and Kirk Johnson

White Paper. Critical Considerations for Monitoring Low Levels of Hydrogen Sulfide. by Jacob Spector and Kirk Johnson White Paper Critical Considerations for Monitoring Low Levels by Jacob Spector and Kirk Johnson 2015 Honeywell International, Inc. A New Threshold of Safety Critical Considerations When it comes to gas

More information

Every day is filled with decisions let us simplify your day with our technical solutions

Every day is filled with decisions let us simplify your day with our technical solutions Every day is filled with decisions let us simplify your day with our technical solutions 11033 English Profilbrochure 2016_TRYK.indd 1 21/11/2016 16.20 Lindpro, in numbers 21 Branches Nationwide and local

More information

Fire Sprinklers Working Group Final Report

Fire Sprinklers Working Group Final Report Introduction The Building Act received Royal Assent on March 25, 2015. The Act aims to establish more consistent building requirements across British Columbia and create a more robust and modern building

More information

One call protects it all.

One call protects it all. One call protects it all. 2 // Tyco Integrated Fire & Security Two industry leaders. One great resource. Driven by our shared passion for protecting the employees and operations of businesses and institutions

More information

Current SEI SAT Initiative Technology Investigations. Architecture-Related Technology. Current SEI SAT Initiative Technology Investigations 5/1/2008

Current SEI SAT Initiative Technology Investigations. Architecture-Related Technology. Current SEI SAT Initiative Technology Investigations 5/1/2008 Current SEI SAT Initiative Technology Investigations Software Engineering Institute Carnegie Mellon University Pittsburgh, PA 15213 Phil Bianco, Andres Diaz-Pace May 1 st, 2008 2008 Carnegie Mellon University

More information

Scholars Research Library. The Role of Plant Clinic in Protecting Vertical Urban Green Spaces in Tehran

Scholars Research Library. The Role of Plant Clinic in Protecting Vertical Urban Green Spaces in Tehran Available online at www.scholarsresearchlibrary.com Annals of Biological Research, 2011, 2 (3) :255-261 (http://scholarsresearchlibrary.com/archive.html) ISSN 0976-1233 CODEN (USA): ABRNBW The Role of

More information

Omni (K) PATIENT MONITOR

Omni (K) PATIENT MONITOR Omni (K) PATIENT MONITOR Omni (K) Intuitive Designed for a fast paced work environment, the Infinium Omni (K) patient monitor offers an extremely simple and adaptable user interface. Patient information

More information

Where Technology Shapes Solutions. Alarm management : Wasn t that problem already solved years ago?

Where Technology Shapes Solutions. Alarm management : Wasn t that problem already solved years ago? Honeywell Users Group 2011 Where Technology Shapes Solutions. Luc De Wilde, Total Petrochemicals Dal Vernon Reising, Human Centered Solutions Alarm management : Wasn t that problem already solved years

More information

Benefits of Enhanced Event Analysis in. Mark Miller

Benefits of Enhanced Event Analysis in. Mark Miller Benefits of Enhanced Event Analysis in Data Center OTDR Testing Mark Miller Dr. Fang Xu AFL/Noyes Test & Inspection Overview Challenges Topics Techniques and Improvements Benefits of enhanced event analysis

More information