Measured and Modeled Light Scattering Values for Dry and Hydrated Laboratory Aerosols*

Size: px
Start display at page:

Download "Measured and Modeled Light Scattering Values for Dry and Hydrated Laboratory Aerosols*"

Transcription

1 VOLUME 21 JOURL OF ATMOSPHERIC AND OCEANIC TECHNOLOGY JULY 2004 Measured and Modeled Light Scattering Values for Dry and Hydrated Laboratory Aerosols* PIR KUS University of Illinois at Urbana Champaign, Urbana, Illinois CHRISTIAN M. CARRICO Colorado State University, Fort Collins, Colorado MARK J. ROOD University of Illinois at Urbana Champaign, Urbana, Illinois ALLEN WILLIAMS Illinois State Water Survey, Champaign, Illinois (Manuscript received 31 December 2003, in final form 5 February 2004) ABSTRACT Closure experiments were completed to compare measured and modeled aerosol optical properties and their dependence on controlled relative humidity (RH) and wavelength of light. NaCl, (NH 4 ) 2 SO 4, and NH 4 NO 3 aerosol particles with approximate geometric mass mean diameters of 0.2 m and geometric standard deviations of 1.7 were tested as part of this study. High evaporative losses (up to 40%) were observed for NH 4 NO 3 aerosol at this particle size range due to heating, and the results from these tests have been excluded from the closure analysis. Aerosol optical properties were measured with a RH-scanning nephelometry system (humidograph) and modeled with a Mie Lorentz light scattering model. Particle size distributions were measured with a scanning differential mobility analyzer. Closure between the measured and modeled values of the total light scattering coefficient ( sp ), backscatter ratio (b), and Ångström exponent (å) for dry (low RH) aerosols was achieved within 0.0% 5%, 4% 15%, and 3% 17%, respectively. The values of f (RH), hemispheric b, and å at 80% RH agreed within 2% 27%, 1% 27%, and 1% 28%, respectively. Correcting for nephelometer nonidealities, including a heating artifact, improved the agreement between the measured and predicted sp values at RH 80% from 35% to 13% for the TSI nephelometer at the maximum heating condition, and from 18% to 11% for the Radiance Research, Inc. (RR), nephelometer. Accurate quantification of the closure for these optical properties is important when establishing visibility standards, and assessing the progress toward meeting those standards, as well as reducing the uncertainties in estimating radiative forcing due to those aerosols. 1. Introduction Investigation of the effects of aerosols on atmospheric radiative transfer and hence visibility and climate relies on a range of techniques for measuring and modeling aerosol optical properties. The integrating nephelometer, an instrument that can measure total and back light scattering coefficients by aerosol particles ( sp and bsp, respectively), has been developed, commercialized, and * Illinois State Water Survey Manuscript Number J-482. Corresponding author address: Prof. Mark. J. Rood, Civil and Environmental Engineering, University of Illinois at Urbana Champaign, 205 N. Mathews St., Urbana, IL mrood@uiuc.edu used for visual air quality studies over many decades (Charlson et al. 1969; Heintzenberg and Charlson 1996). Nephelometers have been characterized during laboratory and ambient aerosol experiments at low relative humidity (RH) conditions [e.g., RH 40%; Bodhaine et al. (1991); Anderson et al. (1996); Anderson and Ogren (1998)]. The intent of this study is to investigate the performance of nephelometers in measuring an aerosol s optical properties as a function of RH. Accurate quantification of the optical properties of aerosols is important when establishing visibility standards and assessing the progress toward meeting those standards (Watson 2002), as well as reducing the uncertainties in estimating radiative forcing due to those aerosols (Quinn et al. 1996). Measurements by integrating nephelometers provide 2004 American Meteorological Society 981

2 982 JOURL OF ATMOSPHERIC AND OCEANIC TECHNOLOGY VOLUME 21 extensive and intensive aerosol optical properties that are important in determining the impacts of aerosols on visibility and climate (Charlson et al. 1992). The optical extensive property investigated here is sp and the optical intensive parameters of interest are the hygroscopic growth factor [ f (RH), the ratio of sp at a given RH to sp for a low RH condition], the hemispheric backscatter fraction (b, the ratio of bsp to sp ), and the Ångström exponent (å; Ångström 1964). The Ångström exponent characterizes the dependence of sp on wavelength ( ) of scattered light and is defined as log[ sp ( 1 )/ sp ( 2 )]/log( 1 / 2 ). Relative humidity is one of the most important meteorological influences on aerosol particle size and light scattering as determined from modeling studies (Orr et al. 1958; Hegg et al. 1993; Nemesure et al. 1995), experimental laboratory investigations (Tang 1980, 1996, 1997), and experimental ambient aerosol studies (Covert et al. 1979; Charlson et al. 1984). As a result, investigations of aerosol optical properties using controlled RH nephelometry (humidography) have been pursued for at least three decades (Covert et al. 1972). Experiments have been implemented in many forms and with many RH control strategies ranging from passive systems with one dry RH measurement and one ambient RH measurement (Li-Jones et al. 1998; Gassó et al. 2000; Bergin et al. 2001), dilution/mixing systems for RH control (Covert et al. 1972; Rood et al. 1985), and using a combination of water vapor addition and/or thermal control (Carrico et al. 1998, 2003; Kotchenruther et al. 1999; ten Brink et al. 2000; Day et al. 2000). Also, these experiments have investigated diverse aerosol populations including background marine aerosols (Carrico et al. 1998), biomass smoke (Kotchenruther and Hobbs 1998), class 1 National Park sites (Malm and Day 2001), mineral dust (Li-Jones et al. 1998; Carrico et al. 2003), and polluted aerosols in urban (Rood et al. 1987) and rural continental regions (Koloutsou-Vakakis et al. 2001). Despite the numerous field applications of humidography, limited closure studies are available regarding these measurements, particularly for polydisperse and hydrated aerosols that typically exist in the lower troposphere. An important concern is sample losses and the extent to which the measurement itself can alter the aerosol s properties [e.g., through sample heating and volatility losses; Willeke and Baron (1990)]. For example, previous studies have shown that sample heating within a nephelometer with a continuous light source is of several degrees Celsius and as large as 4 7 C depending on experimental conditions (e.g., aerosol sample flow rate, inlet temperature, lamp power) (Anderson et al. 1996; Molenar 1997; Heintzenberg and Erfurt 2000). Studies by Dougle et al. (1998), Bergin et al. (1997), and ten Brink et al. (2000) examined the losses of NH 4 NO 3 in a nephelometer for a range of temperatures, residence times, and aerosol size distributions while heating the aerosol. Bergin et al. observed a 10% decrease in sp due to NH 4 NO 3 particle loss because of the aerosol particles volatility at 27 C and a 40% decrease at 47 C for the typical mean residence time of 4.8 s. The particles tested had a geometric number median diameter of 0.2 m and geometric standard deviation of 1.6. Closure studies compare agreement between different methods of measurements and modeling of aerosol properties (Quinn et al. 1996). Measured and modeled values of sp and bsp for monodisperse aerosol at low RH provided closure within 10% (Anderson et al. 1996). Studies of dry polydisperse ambient marine aerosol have demonstrated closure within 3% 19% for sp and from 8% to 40% for bsp (Quinn et al. 1996). This study evaluates integrating nephelometer measurements of sp, bsp, f (RH), b, and å for dry and hydrated polydisperse laboratory aerosols with increasing and decreasing RH conditions and examines the closure between these measured and modeled optical properties for dry and hydrated polydisperse laboratory aerosols (Fig. 1). The aerosol particles consist of single component dry or aqueous solutions that contain sodium chloride (NaCl), ammonium sulfate [(NH 4 ) 2 SO 4 ], or ammonium nitrate (NH 4 NO 3 ) solute, which are three common components of ambient aerosols. These results are also important in that they provide a quantitative analysis of the effects of heating within the nephelometer and recommendations for correcting for this heating, especially when studying aerosols at ambient conditions. 2. Experimental approach a. Measured physical and optical aerosol properties The experimental setup used to measure light scattering coefficients and particle size consisted of an aerosol generator, measurement of dry particle size distributions, and measurement of the temperature, RH, pressure, and optical properties of the aerosol (Fig. 2). Solutions were prepared by mixing reagent-grade solute with deionized, distilled, and filtered water (NOpure; Barnstead International, Dubuque, Iowa, model D4741). A polydisperse aerosol was then generated with an atomizer, dried with filtered dry air in a mixing chamber, and then passed through an impactor that allows particles with an aerodynamic diameter (D p,a ) 1 m to remain in the gas stream (Fig. 2). The aerosol flow was then split with 1 alpm (actual liters per minute) flowing to the aerosol particle sizing system and 22 alpm flowing through an RH-scanning nephelometry system (humidograph), as described below. The aerosol size distribution was measured with a scanning differential mobility analyzer (SDMA) that included an electrostatic classifier (TSI Inc., Shoreview, Minnesota, model 3936 differential mobility analyzer) interfaced to a condensation particle counter (CPC; TSI, model 3010-S). Appropriate operating conditions for the

3 JULY 2004 KUS ET AL. 983 FIG. 1. Schematic describing the method used to evaluate closure between the measured and modeled light scattering. SDMA were tested by comparing measured and vendor reported particle sizes for polystyrene latex (PSL) spheres (Duke Scientific Corporation, Palo Alto, California). These PSL particles were generated in the same manner as the inorganic salt particles. The aerosol flow through the humidograph first passed through a low-rh reference nephelometer (TSI, model 3563) that measured sp and bsp at dry conditions (RH 10%). Subsequently, RH was then controlled by adding water vapor with a humidifier using a water permeable membrane and by cooling with a custom thermoelectric cooler (Melcor Inc., Trenton, New Jersey). The humidifier consisted of a water-jacketed encasing of a 1.59 cm (5/8 in.) inner-diameter water vapor permeable membrane composed of Goretex that was custom manufactured by W. L. Gore and Associates, Inc., Newark, Delaware (Koloutsou-Vakakis et al. 2001). The aerosol sample stream then split with FIG. 2. Experimental apparatus used in this study.

4 984 JOURL OF ATMOSPHERIC AND OCEANIC TECHNOLOGY VOLUME 21 approximately 11 alpm of aerosol flowing through two RH-scanning nephelometers in parallel. One nephelometer had a continuous light source (TSI, model 3563) and the other nephelometer had a flash lamp light source [Radiance Research Inc. (RR), Seattle, Washington, model M903]. The TSI nephelometers measured sp and bsp at 450, 550, and 700 nm and used a quartz halogen continuous light source, while the RR nephelometer measured sp at 530 nm with a quartz halogen flash lamp as its light source. A complete experiment consisted of increasing and then decreasing RH scans between 35% RH 85% in the RH-scanning nephelometers during a 45-min period. Increasing RH scans began from a sample condition with RH 35% in all of the nephelometers. The (NH 4 ) 2 SO 4 and NaCl aerosols are dry at this condition since negligible growth was observed until the deliquescent relative humidity (DRH) of these salts (Tang 1996). Decreasing RH scans, on the other hand, began from a hydrated aerosol condition at RH 85% (Carrico et al. 2003). Dry filtered air and CO 2 were used separately to calibrate the light scattering measurements of both nephelometers, and those measured scattering values agreed within 5% with published values (Anderson et al. 1996). All light scattering measurements were corrected to most accurately describe the true optical properties with corrections for known angular nonidealities, namely, truncation and non-lambertian error examined for the TSI nephelometer (Anderson et al. 1996; Anderson and Ogren 1998). The same angular nonidealities were assumed for the RR nephelometer. To facilitate the comparison of light scattering values for the two different nephelometers, the results from the RR nephelometer were adjusted to 550 nm using å values calculated for wavelengths of 450 and 550 nm from the TSI nephelometer. Corrections for nephelometer nonidealities were applied as a function of RH (Carrico et al. 2000). An adjustment of the optical properties to 25 C and 1013 mb was performed using the nephelometers sample temperature (T) and pressure values to compare the modeled results to the measured results. The pressure sensor in the TSI nephelometer is located just after the inlet of the instrument and before the location of the continuous light source. The sample temperature sensor is located just after the continuous light source. The RH of the sample volume for the RR nephelometer was calculated and based on the dry-bulb temperature as measured with a type K thermocouple located 2.5 cm inside of the nephelometer s outlet and the dewpoint temperature determined for the RH-scanning TSI nephelometer. The total pressure of the RR nephelometer s sample volume was assumed equal to the total pressure of the TSI nephelometer. Experimental f (RH) measurements were then fit to functions of the forms given below for the increasing branch of a deliquescent curve [Eq. (1)] and for the decreasing branch of a deliquescent curve [Eq. (2); Kotchenruther et al. (1999)] to facilitate comparisons of the experimental and modeled results: [ ] [ ] g RH 1 RH d [ ] b RH 1 RH d f (RH) 1 a 1 arctan 1 10 c 1 arctan 1 10 and (1) g RH f (RH) c 1. (2) 100 The RH was measured with four capacitive-type RH sensors [two Vaisala, Helsinki, Finland, model HMP- 233 RH sensors and two TSI sample RH sensors (collocated with the TSI outlet temperature sensor) located within the TSI nephelometers]. Dewpoint temperatures were measured with three chilled mirror type sensors (General Eastern Instruments, Wilmington, Massachusetts, models Hygro M1 and M4). Dry-bulb temperatures were measured with thermocouples (OMEGA Engineering, Inc., Stamford, Connecticut, type K), the Vaisala temperature sensors (Vaisala, model HMP-233), and with thermistors (defined as an outlet sample temperature sensor in the TSI-3563 manual) internal to the TSI nephelometers. The detailed locations of the TSI temperature, pressure, and RH sensors can be found in the online manual of the instrument ( com/particle/products/nephelometers/neph.htm). The capacitive-type temperature sensors were calibrated in an isothermal gas flow (T 22 C) at RH 25% and RH 85% using a third Vaisala RH sensor (Vaisala, model HMP-233) that was recently factory calibrated and used as a transfer standard. Agreement between the two dewpoint sensors and the capacitive-type detectors at 22 C was within 2%. The RH sensor in the RR nephelometer was not altered from its factory calibration and was not used in this study as its reading was 10% lower than the other sensors. Intercomparison of the calibrated temperature sensors was in agreement

5 JULY 2004 KUS ET AL. 985 TABLE 1. Manufacturer-specified uncertainties ( ) for temperature and RH measurements for sensors used in this experiment. Sensor T ( C) RH (%) Vaisala HMP-233 Omega type K thermocouple General Eastern dewpoint hygrometer M1 TSI nephelometer RR nephelometer b 0.5 d a 2 c 5 2 a For RH 90%. Given as 1% when calibrated against a high quality certified humidity standard. b Uncertainty in measuring dewpoint temperature. c RH uncertainty at RH 80% found from dewpoint uncertainty and thermocouple dry-bulb temperature uncertainty. d The uncertainty in the temperature measurements in the TSI nephelometer is specified as 0.3 C in Heintzenberg and Erfurt (2000), 0.5 C in Anderson et al. (1996), and as 2 C from the manufacturer (M. Havlicek, TSI Inc. 2003, personal communication). We adopt the middle value here of 0.5 C. to within 0.2 C. The manufacturers specified uncertainties ( ) in these RH and temperature measurements are given in Table 1. The RH was measured and derived with two different instruments [Vaisala RH sensor (RH2, Fig. 2) and two collocated dewpoint hygrometers (DP2, Fig. 2)] at the inlet position of the RH-scanning TSI nephelometer. The RH can then be described most accurately by taking the average of all of the RH measurements representing the same physical position during periods of constant low and high RH conditions. Therefore the RH from the Vaisala RH sensor was calibrated based on the regression line between Vaisala RH versus this averaged RH. The RH of the sample volume for the RH-scanning TSI nephelometer was determined by its internal sample RH sensor (RH3, Fig. 2). The sample RH for the RH-scanning RR nephelometer, however, was determined from the calculated dewpoint temperature [using the inlet Vaisala dry-bulb temperature and RH sensors (T2 and RH2, respectively, Fig. 2)] and the dry-bulb temperature for its sample volume (T4, Fig. 2). b. Description of experiments Experiments were performed under conditions to produce light scattering values that are comparable to those at ambient conditions, and with select TSI nephelometer lamp power settings and ventilation conditions to investigate the effects of instrumental heating on measured optical properties. Tests included initial conditions with a cold start, where the nephelometers were off for 1 h, and initial conditions with a steady state condition, where instruments were powered on for 1 h. Experiments with maximum instrumental heating occurred at full lamp power (75 W) with no additional ventilation other than that provided by the instrument and removal of the TSI nephelometer s enclosures. The maximum heating case is nearly a worst case scenario with about a 4.5 C warming for the TSI nephelometer, though greater heating may occur if the instrument has its cover in place and is operated at a lower sample flow rate. Intermediate instrumental-heating experiments occurred similarly with lowered lamp power (50 W) provided to the RH-scanning TSI nephelometer. The minimum instrumental heating occurred with further reduced lamp power (25 W) and with additional ventilation using external forced and induced draft fans ventilating the nephelometers. The fans were located external to but surrounding the light scattering chamber of the nephelometer. c. Modeled aerosol optical parameters and evaluation of closure A computer program based on the Mie Lorentz light scattering (BHMIE) code of Bohren and Huffman (1983) was used to calculate aerosol optical properties for the conditions of the experiment. Particles were assumed to be homogeneous spheres of uniform density for dry and hydrated aerosol particles. Aerosol chemical properties including the refractive index and density as a function of RH for the Mie light scattering model were taken from Tang (1996) and Tang and Munkelwitz (1994). Diameter growth factors (D p,wet /D p,dry ) for each particle size bin across each size distribution were determined from the mean diameter for each channel of the SDMA. Scattering coefficients as a function of RH were calculated by assuming that all of the particles in each size bin grow equally and aerosol particle number concentration is conserved (Fig. 1). 3. Results and discussion a. Aerosol particle size distributions Measured geometric mean particle diameters (D p,g ) of the PSL spheres were 54 and 304 nm, and were measured with the SDMA to within 1% of the diameters reported by the vendor of the spheres. The measured dry particle size distributions from the laboratory-generated inorganic aerosols had approximately lognormal distributions with D p,g and g based on masses of approximately 0.2 m and 1.7, respectively (Fig. 3). Measurement of one dry particle size distribution took 1 2 min while an increasing or decreasing RH scan took 15 min. Therefore, generation of a stable test aerosol was essential. The stability of the output from the aerosol generator was checked by monitoring the change in dry sp during a complete RH scan. The changes in sp were within 1% during an RH scan, indicative of stable aerosol generation. b. Effects of sample heating on measurements of aerosol optical properties Dry-bulb temperature values at startup for all of the tests were within 0.2 C of one another. As previously mentioned, the locations of these temperature sensors

6 986 JOURL OF ATMOSPHERIC AND OCEANIC TECHNOLOGY VOLUME 21 FIG. 3. Typical dry particle size distributions for (NH 4 ) 2 SO 4, NH 4 NO 3, and NaCl. Geometric mass means (D p,g ) and geometric standard deviations ( g ) are provided above. are described schematically in Fig. 2. Steady-state temperature profiles were then recorded for an aerosol passing through the humidograph at three lamp power settings as described in Fig. 4. The maximum rise in the sample temperature of 4.5 C is for the 75-W lamp power setting and occurs between the inlet of the RH-scanning TSI nephelometer (T2, Fig. 2) and its sample temperature within the nephelometer (T3, Fig. 2). This increase in temperature was achieved within 35 min of a cold start. A more modest heating of 2.0 C was observed in these nephelometers when the lamp power is 25 W. The RR nephelometer also shows a temperature increase of 2 C from a cold start though the exact profile is uncertain due to the precision of the temperature measurement ( 0.5 C as it reports only integer values). Heating in the RR nephelometer is due to a combination of factors including its light source and possibly heating due to heat transfer from the surroundings as a result of the close proximity of having all three instruments located in the same sampling container. The agreement between three nephelometers for low RH conditions (RH 35% for all three nephelometers) is observed to be within 5% for the nonvolatile NaC1 and (NH 4 ) 2 SO 4 for the minimum, intermediate, and maximum heatings. The minimum, intermediate, and maximum heating cases represent temperature rises of 2, 3, and 4 C and measured sample temperatures of 23,25, and 29 C, respectively. However, substantial decreases in sp values for volatile NH 4 NO 3 aerosols were observed when comparing measured sp by the RH-scanning and low RH reference TSI nephelometers. A 10% 15% reduction in sp values for the minimum heating case and a 35% 45% reduction in sp values for the maximum heating case were observed with NH 4 NO 3 aerosols. The larger reduction in sp for NH 4 NO 3 when compared to NaCl and (NH 4 ) 2 SO 4 are attributed to the evaporation of the NH 4 NO 3 particles. Tests with NH 4 NO 3 therefore have been excluded from the closure analysis. Although these losses were likely FIG. 4. Temperature profile in the humidograph system for minimum, moderate, and maximum heating cases during the steady-state tests. Locations of temperature sensors are described in Fig. 2. exacerbated due to the small sizes of the generated aerosol, the need to minimize the heating with regard to the volatilization losses of such semivolatile components of the aerosol is noted. Ratios of dry sp values from the low RH reference nephelometer to the RH-scanning nephelometers range from 0.90 to 1.04 for NaCl and (NH 4 ) 2 SO 4 aerosol. Ratios of these sp values were used in the following analyses to normalize f (RH) to account for small calibration and particle loss differences between the reference and scanning RH nephelometer so that it is unity at the low RH condition. Results for bsp values were also normalized in the same manner as their corresponding bsp values, except for the NaCl aerosol tests. The NaCl tests had low signal-to-noise ratios for the bsp values due to the low backscattering values caused by the low NaCl mass concentrations used during the tests. The difference between bsp values for the low RH reference and RH-scanning nephelometers was used as an offset to allow f (RH) for bsp values to begin at one. All the correction factors [i.e., for standard temperature and pressure (STP), angular nonidealities, normalization, and wavelength adjustment for the RR nephelometer] applied to the measured scattering coefficients at 550 nm were tabulated in Table 2. c. Closure between modeled and measured total scattering coefficients Measured and corresponding predicted sp values at 550 nm are compared in Fig. 5 for dry and hydrated NaCl aerosols during the minimum and maximum heating cases. The minimum heating cases described in Fig. 5a report sp values, obtained directly from the nephelometers without any corrections, versus RH measured immediately upstream of the nephelometers. Figure 5b

7 JULY 2004 KUS ET AL. 987 TABLE 2. Correction factors that are used to correct the total scattering coefficients at 550 nm for the TSI and RR nephelometers for selected tests. Corrections for angular nonidealities for the RR nephelometer are assumed equal to the same factors for the TSI nephelometer. Test conditions (NH 4 ) 2 SO 4 TSI- 25W-CS (NH 4 ) 2 SO 4 TSI 74W-SS NaCl TSI 25W-CS NaCl TSI 75W-SS Correction factors for TSI nephelometer Angular nonidealities STP Normalization* Low RH High RH Correction factors for RR nephelometer Wavelength adjustment STP Low RH High RH Normalization TSI, RH-scanning continuous light source nephelometer; RR, RH-scanning flash lamp light source nephelometer; CS, cold start tests; SS, steady-state tests; 25 and 75 W are the power settings for the TSI nephelometer; and STP, standard temperature and pressure. * Normalization is the multiplication of the sp values by the ratio of sp from the reference nephelometer to sp from the RH-scanning nephelometer during the low RH sample period. reports corrected sp values (i.e., nephelometer nonidealities, standard temperature and pressure, and the wavelength correction for RR nephelometer measurements) versus internal sample RH (i.e., accounting for instrumental heating). Figures 5c and 5d are the same as Figs. 5a and 5b, respectively, except they apply to the maximum heating condition. Predicted and measured scattering coefficients reveal very good agreement (i.e., within 2%) for initially dry polydisperse NaCl particles at RH DRH for both the TSI and RR nephelometers. However, discrepancies between measured and modeled sp values are apparent for hydrated aero- FIG. 5. Total light scattering coefficients at 550 nm as a function of RH for NaCl before and after corrections: (a) before any corrections to sp vs RH inlet (Vaisala measured RH just before the nephelometer) for minimum heating, (b) corrected sp vs actual sample RH (RH as experienced by the aerosols in the scattering volume of the instrument) for minimum heating, and (c), and (d) as is (a), (b) except for maximum heating cases, respectively.

8 988 JOURL OF ATMOSPHERIC AND OCEANIC TECHNOLOGY VOLUME 21 FIG. 6. For minimum heating, controlled RH measurements of hygroscopic growth in light scattering factor [ f (RH) sp (RH)/ sp (dry)] at 550 nm for (a) NaCl and (b) (NH 4 ) 2 SO 4. sols when the sample heating is not considered and the inlet RH (or ambient RH for field testing conditions) is used to represent the RH within the sample volume (Figs. 5a and 5c). Discrepancies become even more apparent for the hydrated aerosol at the maximum heating case. For example, the modeled sp value at a sample RH of 80% is 130 Mm 1 (inverse megameters), while sp values measured at an inlet RH of 80% by the RR and TSI nephelometers are 107 and 86 Mm 1, respectively. The discrepancies resulting in f (RH 80%) 9.5 (predicted), 7.8 (measured by RR nephelometer), and 6.2 (measured by TSI nephelometer), causing the differences between modeled and experimental results to range from 18% to 35%. Implementation of corrections due to nephelometer nonidealities (Anderson and Ogren 1998) and heating within the nephelometers reduced the differences between the modeled and experimental results to 10% over the full range of increasing and decreasing RH conditions tested here. Light scattering values obtained during increasing RH conditions as described in Figs. 5b and 5d (open dark and gray triangles for the TSI and RR nephelometers, respectively) also display the effect of heating at RH values just below the DRH of the aerosol. Aerosols can deliquesce immediately upstream of the increasing RH nephelometer as the RH of the nephelometer s sample volume approaches the DRH of the aerosol. Such a response occurs as a result of the reduction in RH due to heating within the nephelometer. Since sample heating lowers the RH in the nephelometer s sample volume, particles that deliquesce immediately upstream of the sample volume move down the metastable branch of the aerosol s hysteresis loop to an extent determined by the magnitude of the heating in the nephelometer. Therefore, these increasing RH points fall on the upper branch of the hysteresis loop as seen in the Figs. 5b and 5d. Heating within the nephelometers also causes an artifact when measuring DRH values. The aerosol DRH value is identified as the large step change in sp values for a small change in RH during increasing RH conditions. This rapid change in sp occurs at a sample volume RH well below the previously reported DRH value of 75% for NaCl. For example, the measured DRH is 62% for NaCl during the maximum heating case when using the RH-scanning TSI nephelometer and its sample RH. Since sample heating lowers the RH in the sample volume, particles that deliquesce immediately upstream of the sample volume move down the metastable branch of the aerosol s hysteresis loop to an extent determined by the magnitude of the heating in the nephelometer. Thus, the step change in sp caused by deliquescence as reported by the nephelometer represents the transition from measuring sp for the complete population of dry particles to measuring sp of metastable droplets on the upper branch of their hysteresis loop at a sample RH the aerosol s actual DRH. Intermediate sp values that occur between the beginning and end of the step change (i.e., reported DRH value) result from the existence of a mixed population of dry particles and hydrated metastable droplets. Therefore, the lower branch of the hysteresis loops in Figs. 5b and 5d are plotted versus the inlet RH until the end of the step change in the light scattering values. The sample RH is then used for the light scattering values above this step change in addition to light scattering values along the upper branch of the hysteresis loop. The RH corresponding to the top of this step change in the light scattering values during increasing RH conditions is 75% (Fig. 6) for the NaCl particles. In practice as measured with the humidograph, the deliquescence step change occurs over an RH range of 3% 4% and is a result of the RH gradients within the instrumentation (e.g., particles near the walls of the humidifier may experience higher RH conditions than along the centerline of the humidifier). Measured DRHs and crystallization RHs (CRHs) for the two salts tested here are compared to previously reported values in Table 3. The DRH and CRH values for (NH 4 ) 2 SO 4 and NaCl are in

9 JULY 2004 KUS ET AL. 989 TABLE 3. Experimental deliquescence and crystallization relative humidities for the (NH 4 ) 2 SO 4 and NaCl. Inorganic salt (NH 4 ) 2 SO 4 NaCl Measured by this study DRH CRH N Previously reported DRH CRH Reported by Tang (1996) ten Brink et al. (2000) Tang (1996) ten Brink et al. (2000) not available. N number of tests. agreement with previously reported values. These results underscore the need to characterize both the inlet RH (for the correct measurement of DRH) and the sample RH internal to the nephelometer (for the correct correspondence between the measured scattering values and the RH of the nephelometer s sample volume) when heating occurs within the nephelometer. Another reason to limit the sample heating is to minimize the volatility of the aerosol. The magnitude of the sample heating also limits the maximum sample RH for which measurements are possible without causing condensation upstream of the instrument, which is apparent in Fig. 5d where the maximum sample RH was 74% when the inlet RH was 91%. The results for measured and modeled sp values for (NH 4 ) 2 SO 4, which were obtained before the tests with NaCl, show poorer agreement when compared to the NaCl tests. Measured results for absolute sp values for (NH 4 ) 2 SO 4 were consistently 40% 3% high compared to the modeled values due to the use of an aerosol dilution system. Therefore, the absolute sp values for those tests are excluded from this analysis. This is a result of undercounting of particles by the CPC due to a problem with the aerosol dilution system that was located upstream of the CPC, resulting in an inaccurately measured dilution factor for the (NH 4 ) 2 SO 4 aerosol. This is seen in a constant percentage difference between the measured and modeled sp for low and high RH and for all tests with (NH 4 ) 2 SO 4, which is not subject to the volatility losses that were previously discussed for NH 4 NO 3. Improved agreement between the measured and predicted sp values was observed with the subsequent tests using NaCl after rearrangement of the dilution system s plumbing (Table 4). Also, as will be discussed below, measured and modeled intensive properties showed similar agreement for both salts. d. Measured and modeled intensive optical properties The final values for sp, as described above, were then used to calculate f (RH), b, and å as a function of RH for NaCl and (NH 4 ) 2 SO 4 as illustrated in Figs. 6 and 7 for steady-state conditions with minimum heating. The measured and modeled f (RH) values demonstrate agreement within 7% 12% over the entire hysteresis loop for NaCl and (NH 4 ) 2 SO 4. Also of note are the much greater f (RH) values for NaCl when compared to the corresponding values for (NH 4 ) 2 SO 4, even though both salts have similar dry size distributions. These results are consistent with previous results comparing the hygroscopic properties of NaCl and (NH 4 ) 2 SO 4 (Tang 1996; Tang and Munkelwitz 1994). The largest dis- TABLE 4. Comparison of measured and modeled optical parameters for the tests discussed here. Percent difference of measured value from modeled value for TSI Percent difference for RR Test number Test conditions Low RH sp b å RH 80% f(rh) b å Low RH RH 80% sp f(rh) A1 A2 A3 A4 A5 C1 C2 C3 C4 C5 C6 (NH 4 ) 2 SO 4 TSI 25W CS (NH 4 ) 2 SO 4 TSI 25W SS (NH 4 ) 2 SO 4 TSI 50W SS* (NH 4 ) 2 SO 4 TSI 75W SS* (NH 4 ) 2 SO 4 TSI 75W SS* NaCl TS1 10 W CS NaCl TS1 25 W CS NaCl TS1 25 W SS NaCl TS1 50 W SS* NaCl TS1 75 W CS* NaCl TS1 75 W SS* TS1, RH-scanning continuous light source nephelometer; RR, RH-scanning flash lamp light source nephelometer; CS, cold start tests; SS, steady-state tests; and 25W, 50W, and 75W are the power settings for the TSI nephelometer. * In these tests, 80% RH along the metastable branch was not achieved due to high heating. The value from the extrapolated fit at 80% was used instead.

10 990 JOURL OF ATMOSPHERIC AND OCEANIC TECHNOLOGY VOLUME 21 FIG. 7. For minimum heating, controlled RH measurements of hemispheric backscatter ratio (b) at 550 nm and Ångström exponent (å) for a nm wavelength pair for (a) NaCl and (b) (NH 4 ) 2 SO 4. Vertical error bars represent the standard deviation of the averaged parameter. crepancies with the theory for f (RH) values exist at high RH values (RH 80%). This may be due to a number of factors ranging from increased particle loss for larger more hygroscopic particles (i.e., NaCl), increased particle loss due to aerosol flowing through the smaller RR nephelometer, and larger uncertainties in measurement of RH at high RH conditions. Fitted equations used for increasing and decreasing RH portions of the deliquescent type [i.e., NaCl and (NH 4 ) 2 SO 4 ] aerosols agreed well with the experimental data. The goodness of the fit was evaluated based on the root-mean square of the standard error (rmse), which is the square root of the sum of the square of the differences between the measured and fitted points divided by the degrees of freedom. A value closer to 0 means a better fit. The rmse values ranged from 0.05 to 0.76, having higher error in runs with lower lamp power where the noise is higher. Another parameter that was considered for the quality of the fit is the confidence width on the calculated f (RH 90%) value. The maximum achievable RH in the nephelometer is reduced in the TSI nephelometer when operating with a high lamp power setting and necessitates the use of the fitted equation to obtain f (RH) values beyond the measured RH values. The confidence width on the f (RH) value at 90% (i.e., RH that is out of the measured RH range) was calculated at the 95% confidence level. The predicted value is actually within the lower (extrapolated value confidence width) and upper (extrapolated value confidence width) limits of the confidence interval. For example, an f (RH) value of 2 with a confidence width of 0.3 indicates that the f (RH) value can be between 1.7 and 2.3 with a 95% confidence level. A smaller confidence width suggests higher confidence in the extrapolated portion of the fit. Confidence intervals for f (RH 90%) ranged from 0.2 for runs with lower lamp power to 3.1 for the runs with higher lamp power and the most hygroscopic reagent NaCl. It should be noted that the extrapolation of the f (RH) values to RH of 90% was done to assess the quality of the fit. The confidence interval at RH 90% represents simply the statistical uncertainty for the fit, not the ability of the nephelometer to make measurements at RH 90%. Optical parameters b and å demonstrate similarly good agreement between measured and modeled values for the entire hysteresis loop for NaCl and (NH 4 ) 2 SO 4 for the select minimum heating runs (Fig. 7). The agreement between the measured and predicted b and å values ranged from 4% to 15% at RH 35% and 2% to 11% at RH 80% (Fig. 7). At low RH conditions, the agreement for b is better when the lamp power is 75 W compared to the reduced lamp power since the signalto-noise ratios for bsp values with reduced lamp power were affected significantly with lamp power settings between 25 and 75 W. As RH increases and particles grow, the hemispheric backscatter fraction decreases as larger particles preferentially scatter light in the forward direction and the wavelength dependence of sp decreases (Bohren and Huffman 1983). Both parameters are observed to have a nearly linear dependence on the controlled RH conditions when existing as an hydrated aerosol along the upper branch of the hysteresis loop as observed for ambient aerosol as well (Carrico et al. 2000, 2003). e. Sensitivity analysis of measured and derived optical aerosol properties to temperature and RH The use of a sample temperature or RH internal to the nephelometer is dependent on how representative the measurement is of the sample volume as well as the accuracy of the measurement. For example, the sample temperature measurement in the RR nephelometer is recorded as an integer resulting in an uncertainty of 0.5 C while the uncertainty of the RH measurement internal to the TSI nephelometer is reported by the manufacturer as 5% (Table 1). The effects on the optical properties of the hydrated aerosol from an error in the measured sample temperature (0.2, 0.5, and 1.0 C) and the resulting error in RH (1%, 2%, and 5% RH) are tabulated in Table 5 assuming a base case sample tem-

11 JULY 2004 KUS ET AL. 991 TABLE 5. Resulting uncertainties in aerosol optical parameters at RH 80% for given temperature/relative humidity uncertainties using modeled optical properties for salts in this study. Temperature uncertainty RH uncertainty (at RH 80%) T 0.2 C RH 1% T 0.5 C RH 2% T 1.0 C RH 5% f(rh 80%) b å NaCl (NH 4 ) 2 SO 4 NaCl (NH 4 ) 2 SO 4 NaCl (NH 4 ) 2 SO 4 5%; 3% 3% 1% 1% 1% 1% 13%; 9% 10%; 8% 2% 3% 2% 2% 26%; 15% 19%; 15% 5% 8%; 7% 4%; 5% 3%; 4% perature of 25 C and RH 80%. Resulting relative uncertainties for NaCl aerosol and T 1.0 C ( RH 5%) are 26% and 15% for f (RH 80%), 5% and 5% for b, and 4% and 5% for å.similar results in terms of the percentage uncertainties in the optical parameters are observed for (NH 4 ) 2 SO 4 (Table 5). The effects on b and å are less than 10% for a temperature uncertainty of 1.0 C. Another effect on f (RH) is the 2% uncertainty in measuring RH, which results in an error of 13% to 9% for f (RH 80%). In addition to temperature and RH uncertainties, heating in the sample volume of the instrument results in artifacts such as volatile particle loss and inaccurate estimation of DRH when attempting to measure aerosols at ambient conditions. Furthermore, use of a high-accuracy RH measurement device external to the nephelometer is dependent on simultaneously measuring temperature at both locations where RH and light scattering measurements are made in order to correct the RH measurements as discussed above. If the sample RH was not characterized accurately, especially for ambient samples at high RH (e.g., 85%), the resulting measured and derived parameters will diverge significantly from the actual values. Values of f (RH), b, and å as a function of inlet RH (RH inlet ) and temperature differentials ( Ts) between the inlet and sample temperature of 2.0, 3.0, and 4.0 C are shown in Figs. 8a d. These correspond to RH differences of approximately 11%, 16%, and 21% RH for a base case of T 25 C and RH 85%. These calculations used the modeled optical parameters for (NH 4 ) 2 SO 4, assuming a dry size distribution with a geometric mean volume diameter of 0.3 m and geometric standard deviation of 1.7, which is representative of the ambient aerosol. Heating of the aerosol by 4 C by the instrument s light source leads to a 40% underestimation in f (RH 85%) and an increase in b and å by 18% and 14% at the same RH, respectively. Due to the exponential nature of the growth curve, the difference in the optical properties as a function of RH undoubtedly increases for RH 85%. Heating of the ambient aerosol sample, without adequate corrections, will result in underestimation of f (RH) but an overestimation of b, which are important parameters in quantifying the radiative effects of ambient aerosols. f. Discussion and recommendations It should be emphasized that the results of this experiment apply strictly to submicron single chemical component aerosols that have no imaginary component of the refractive index in the visible spectrum. These results demonstrate the importance of isothermal sampling when measuring the optical properties of hydrated aerosols. For example, reducing lamp power for the TSI nephelometer and increasing ventilation decreased the sample heating from 4.5 to 1.3 C, which reduced the error in measuring f (RH 80%) from 21% to 7%. Alternately, the use of a lower-wattage light source that maintains the same spectral characteristics may assist in reducing the instrumental noise, particularly in the shorter wavelengths, which are most affected by lowering the lamp s power. More extensive modifications of the TSI nephelometer have been described to further reduce the sample heating to 0.5 C without lowering the lamp power and thus producing a negligible reduction in the instrument s sensitivity while completing measurements for haze and fog studies (Heintzenberg and Erfurt 2000). Though the internal instrumental heating of the RR nephelometer is likely small, a more extensive characterization of its nonideal properties (e.g., truncation error, non-lambertian light source, spectral output of light source, particle losses) is clearly needed to better analyze the uncertainties when measuring the optical properties of aerosol. This closure study is constrained by the accuracy of the sample RH and temperature sensors and how well the sensors characterize the nephelometer s sample volume. Use of higher accuracy temperature and RH measurement devices internal to the sample volume of the nephelometer will assist in reducing the uncertainties when measuring the optical properties of aerosol particles. 4. Summary and conclusions In situ optical properties of dry and hydrated NaCl and (NH 4 ) 2 SO 4 aerosols were characterized in the laboratory by comparing measured and modeled (Mie Lorentz theory) total and backscattering coefficients ( sp and bsp, respectively) while using two different commercially available nephelometers. Closure was then

12 992 JOURL OF ATMOSPHERIC AND OCEANIC TECHNOLOGY VOLUME 21 FIG. 8. Influence of increasing dry-bulb temperature within the nephelometer s scattering volume compared to ambient conditions (RH inlet ) when considering the sample s (a) RH, (b) hygroscopic growth factor [ f (RH)], (c) hemispheric backscatter fraction (b), and (d) Ångström exponent (å) for the (NH 4 ) 2 SO 4 aerosol. The aerosol has a dry size distribution with a geometric volume mean diameter 0.3 m and a geometric standard deviation 1.7. evaluated for sp, the hygroscopic growth response [ f (RH)], hemispheric backscatter fraction (b), and the Ångström exponent (å). Measured sp values obtained by the TSI and RR nephelometers were in agreement to within 5% for low RH (RH 35%) conditions with NaCl and (NH 4 ) 2 SO 4 aerosols. Closure for measured and modeled sp values at the low RH condition for NaCl was achieved within 2% and 5% for the TSI and RR nephelometers, respectively. The scattering coefficients at higher RH values showed larger discrepancies from the modeled values when the inlet RH (or ambient RH in the case of the ambient measurements) was used as the sample RH within the nephelometer. Correcting for the nephelometer nonidealities and the heating improved the agreement between the measured and predicted sp values at RH 80% from 35% to 13% for the TSI nephelometer at the maximum heating condition, and from 18% to 11% for the RR nephelometer. Each correction step has its own uncertainty, but comprehensive uncertainty analyses for the measured and predicted optical properties were not carried out in this study. Obtaining closure for the absolute value of sp when using (NH 4 ) 2 SO 4 aerosols was hindered due to uncertainties in the total particle concentration resulting from a dilution system upstream of the particle sizing system. However, comparable agreement among all tests between measured and modeled results was observed with the intensive properties f (RH 80%), b, and å. The hygroscopic growth properties of the aerosol were fitted to three equations according to the growth type (monotonic or deliquescent) and the direction of the RH scan to be able to evaluate closure. The equations described the hygroscopic properties very well with a low root-meansquare error, of 0.2 on average. The capability of the fit equation to predict f (RH) values outside the measured RH range was assessed by a 95% confidence width on the predicted f (RH 90%). The confidence intervals were within 10% of the f (RH 90%) value for the minimum heating cases. At RH 80%, the percent differences between the measured and modeled f (RH 80%), b, and å values for all three salts after correction for nephelometer nonidealities and heating ranged from 2% to 27%, 1% to 27%, and 1% to 28%, respectively. Deliquescent and crystallization relative humidity

13 JULY 2004 KUS ET AL. 993 (DRH and CRH, respectively) values were also measured and compared to previously reported values for NaCl and (NH 4 ) 2 SO 4 aerosols. The DRH and CRH values for (NH 4 ) 2 SO 4 and NaCl are in agreement with previously reported values. Determining DRH values accurately for ambient or test aerosols when heating occurs will require RH measurements before the sample enters the nephelometers and within the sample volume of the nephelometer since particles can form hydrated solution droplets that do not crystallize as the aerosol approaches its DRH value upstream of the nephelometer. The experiment demonstrated the substantial influences of instrumental heating both with regard to loss of volatile species as well as ambiguities in the sample RH, thus underscoring the desirability of eliminating sample heating. Heating in the sampling volume of the nephelometer also reduces the achievable maximum RH (i.e., inside the nephelometer). Heating of 2.0 and 4.0 C would cause 85% RH of the ambient sample to decrease to 74% and 64%, respectively. These results underscore the limitations of using the TSI nephelometer for high RH or near ambient conditions without taking additional measures to limit its heating. The sensitivity analysis with respect to the magnitude of the heating that a sample has experienced was performed. It revealed that heating of the aerosol by 4 C by the instrument s light source leads to a 40% underestimation in f (RH 85%). Care must be taken when interpreting controlled RH measurements, as heating causes an artifact in measuring the optical properties of aerosol particles. Ideally, measurements of inlet RH and sample RH should be made when considering controlled RH measurements from a nonisothermal nephelometer. The closure experiment revealed the importance of keeping the nephelometry system isothermal for measuring particle scattering coefficients as a function of RH or as particles exist in the ambient environment. In the case of a nonisothermal measurement, the RH should be adequately measured in the nephelometer s sample volume and correction to the ambient measurement should be performed if at all possible. In addition, accurate changes in an aerosol s light scattering properties can be obtained during decreasing and increasing RH conditions. The f (RH) values can differ significantly depending on the composition and RH history of the ambient particles, and the aerosol s type of hygroscopic growth. Acknowledgments. The authors acknowledge Prof. Susan Larson for use of aerosol sizing instrumentation and Drs. Tad Anderson, David Covert, John Ogren, Mike Bergin, Jost Heintzenberg, and Doug Orsini for helpful discussions related to these experiments. This work was supported by the NSF under Grant ATM and by NOAA under Grant COM NH06GP0412. REFERENCES Anderson, T. L., and J. A. Ogren, 1998: Determining aerosol radiative properties using the TSI 3563 integrating nephelometer. Aerosol Sci. Technol., 29, , and Coauthors, 1996: Performance characteristics of a high sensitivity, three wavelength, total scatter/backscatter nephelometer. J. Atmos. Oceanic Technol., 13, Ångström, A., 1964: Parameters of atmospheric turbidity. Tellus, 16, Bergin, M. H., J. A. Ogren, S. E. Schwartz, and L. M. McInnes, 1997: Evaporation of ammonium nitrate aerosol in a heated nephelometer: Implications for field measurements. Environ. Sci. Technol., 31, , and Coauthors, 2001: Aerosol radiative, physical, and chemical properties in Beijing during June, J. Geophys. Res., 106, Bodhaine, B. A., N. C. Ahlquist, and R. C. Schnell, 1991: Three wavelength nephelometer suitable for aircraft measurement of background aerosol scattering coefficient. Atmos. Environ., 25A, Bohren, C. F., and D. R. Huffman, 1983: Absorption and Scattering of Light by Small Particles. Wiley and Sons, 530 pp. Carrico, C. M., M. J. Rood, and J. A. Ogren, 1998: Aerosol light scattering properties at Cape Grim, Tasmania, during the First Aerosol Characterization Experiment (ACE 1). J. Geophys. Res., 103, ,,, C. Neusüß, A. Wiedensohler, and J. Heintzenberg, 2000: Aerosol optical properties at Sagres, Portugal, during ACE 2. Tellus, 52B, , P. Kus, M. J. Rood, P. Quinn, and T. Bates, 2003: Mixtures of pollution, dust, and seasalt aerosol during ACE-Asia: Light scattering properties as a function of relative humidity. J. Geophys. Res., 108, 8650, doi: /2003jd Charlson, R. J., N. C. Ahlquist, H. Selvidge, and P. B. MacCready, 1969: Monitoring of atmospheric parameters with the integrating nephelometer. J. Air Waste Manage. Assoc., 19, , D. S. Covert, and T. V. Larson, 1984: Observations of the effect of humidity on light scattering by aerosols. Hygroscopic Aerosols, L. Ruhnke and A. Deepak, Eds., A. Deepak Publishers, , S. E. Schwartz, J. M. Hales, R. D. Cess, J. A. Coakley Jr., J. E. Hansen, and D. J. Hofmann, 1992: Climate forcing by anthropogenic aerosols. Science, 255, Covert, D. S., R. J. Charlson, and N. C. Ahlquist, 1972: A study of the relationship of chemical composition and humidity to light scattering by aerosols. J. Appl. Meteor., 11, , A. P. Waggoner, R. E. Weiss, N. C. Ahlquist, and R. J. Charlson, 1979: Atmospheric aerosols, humidity and visibility. Character and Origin of Smog Aerosols, G. M. Hidy, Ed., Wiley and Sons, Day, D. E., W. C. Malm, and S. M. Kreidenweis, 2000: Aerosol light scattering measurements as a function of relative humidity. J. Air Waste Manage. Assoc., 50, Dougle, P. G., J. P. Veefkind, and H. M. ten Brink, 1998: Crystallization of mixtures of ammonium nitrate, ammonium sulfate, and soot. J. Aerosol Sci., 29, Gassó, S., and Coauthors, 2000: Influence of humidity on the aerosol scattering coefficient and its effect on the upwelling radiance during ACE 2. Tellus, 52B, Hegg, D., T. Larson, and P.-F. Yuen, 1993: A theoretical study of the effect of relative humidity on light scattering by tropospheric aerosols. J. Geophys. Res., 98 (D10), Heintzenberg, J., and R. J. Charlson, 1996: Design and applications of the integrating nephelometer: A review. J. Atmos. Oceanic Technol., 13, , and G. Erfurt, 2000: Modification of a commercial integrating nephelometer for outdoor measurements. J. Atmos. Oceanic Technol., 17, Koloutsou-Vakakis, S., and Coauthors, 2001: Aerosol properties at a

A Method for Fire Detecting by Volume and Surface Area Concentration Based on Dual Wavelengths

A Method for Fire Detecting by Volume and Surface Area Concentration Based on Dual Wavelengths A Method for Fire Detecting by Volume and Surface Area Concentration Based on Dual Wavelengths Xiao Xiao, Tian Deng, Ming Zhu, Shu Wang School of Electronic Information and Communications, Huazhong University

More information

Optec NGN-2 LED Ambient Integrating Nephelometer

Optec NGN-2 LED Ambient Integrating Nephelometer Optec NGN-2 LED Ambient Integrating Nephelometer Extended Abstract # 77 John V. Molenar 1 Gerald Persha 2 1 Air Resource Specialists, Inc., 1901 Sharp Point Drive, Fort Collins, CO, 80525 2 Optec Inc.,

More information

GAW - WCCAP recommendation for aerosol drying

GAW - WCCAP recommendation for aerosol drying GAW - WCCAP recommendation for aerosol drying Alfred Wiedensohler and Wolfram Birmili Leibniz Institute for Tropospheric Research Humidity control One particular problem during summertime or in warm and

More information

Polarized Light Scattering of Smoke Sources and Cooking Aerosols

Polarized Light Scattering of Smoke Sources and Cooking Aerosols Polarized Light Scattering of Smoke Sources and Cooking Aerosols Thomas Cleary, Amy Mensch National Institute of Standards and Technology, Gaithersburg, MD, USA Abstract Light scattering data was gathered

More information

Hygrometry. The art or science of humidity observation

Hygrometry. The art or science of humidity observation Hygrometry The art or science of humidity observation Humidity Under normal atmospheric conditions, the amount of water vapor any sample of air can hold depends primarily on its temperature Various terms

More information

GE General Eastern Instruments

GE General Eastern Instruments GE General Eastern Instruments The Humilab Bench Top NIST Traceable Humidity Calibration System Introduction GE General Eastern s Humilab provides the user with the ability to verify, calibrate and document

More information

Investigation of the Potential Use of Blue Light in Forward Scattering Optical Smoke Chambers to Detect all UL217 Fires in the New Standard

Investigation of the Potential Use of Blue Light in Forward Scattering Optical Smoke Chambers to Detect all UL217 Fires in the New Standard Investigation of the Potential Use of Blue Light in Forward Scattering Optical Smoke Chambers to Detect all UL217 Fires in the New Standard David Richardson, Daniel O Shea, Stephen Daniels, Michael Byrne

More information

2. HEAT EXCHANGERS MESA

2. HEAT EXCHANGERS MESA 1. INTRODUCTION Multiport minichannel and microchannel aluminium tubes are becoming more popular as components in heat exchangers. These heat exchangers are used in various industrial applications and

More information

THE NEXT GENERATION IN VISIBILITY SENSORS OUTPERFORM BOTH TRADITIONAL TRANSMISSOMETERS AND FORWARD SCATTER SENSORS

THE NEXT GENERATION IN VISIBILITY SENSORS OUTPERFORM BOTH TRADITIONAL TRANSMISSOMETERS AND FORWARD SCATTER SENSORS THE NEXT GENERATION IN VISIBILITY SENSORS OUTPERFORM BOTH TRADITIONAL TRANSMISSOMETERS AND FORWARD SCATTER SENSORS Steve Glander: Senior Sales Engineer All Weather, Inc. 1165 National Dr. Sacramento, CA

More information

Compression of Fins pipe and simple Heat pipe Using CFD

Compression of Fins pipe and simple Heat pipe Using CFD Compression of Fins pipe and simple Heat pipe Using CFD 1. Prof.Bhoodev Mudgal 2. Prof. Gaurav Bhadoriya (e-mail-devmudgal.mudgal@gmail.com) ABSTRACT The aim of this paper is to identify the advantages

More information

ZONE MODEL VERIFICATION BY ELECTRIC HEATER

ZONE MODEL VERIFICATION BY ELECTRIC HEATER , Volume 6, Number 4, p.284-290, 2004 ZONE MODEL VERIFICATION BY ELECTRIC HEATER Y.T. Chan Department of Building Services Engineering, The Hong Kong Polytechnic University, Hong Kong, China ABSTRACT Selecting

More information

Scientific Principals and Analytical Model. Charcoal Cooler. Lisa Crofoot MECH 425, Queens University

Scientific Principals and Analytical Model. Charcoal Cooler. Lisa Crofoot MECH 425, Queens University Scientific Principals and Analytical Model Charcoal Cooler Lisa Crofoot MECH 425, Queens University 1.0 Scientific Principles Evaporative cooling is based on the principle that water requires heat energy

More information

4th International Conference on Sensors, Measurement and Intelligent Materials (ICSMIM 2015)

4th International Conference on Sensors, Measurement and Intelligent Materials (ICSMIM 2015) 4th International Conference on Sensors, Measurement and Intelligent Materials (ICSMIM 2015) Multi-stage Series Heat Pump Drying System with Dehumidification Simulation and Experiment Verification Chao

More information

Audit Report. Global GAW station Pallas, Finland

Audit Report. Global GAW station Pallas, Finland Audit Report Global GAW station Pallas, Finland February 25, 2015 Prof. Dr. Alfred Wiedensohler, Dr. Thomas Tuch General: This station audit was a re-audit of the audit performed in August 2004. As in

More information

3.4 Humidity. Types of instruments

3.4 Humidity. Types of instruments Practical Handbook of Tower Flux Observation (Ver..0) Chapter 3 3.4 Humidity Humidity is the amount of water vapor in the air, which is expressed by various indexes depending on the study objectives. These

More information

CALIBRATION HOW IT WORKS

CALIBRATION HOW IT WORKS HIGH POWER SOLUTIONS POWER DETECTORS ENERGY DETECTORS At Gentec-EO, we understand that the essence of our business since over 45 years has been delivering accuracy. There are no half measures: it either

More information

Understanding total measurement uncertainty in power meters and detectors

Understanding total measurement uncertainty in power meters and detectors Understanding total measurement uncertainty in power meters and detectors Jay Jeong, MKS Instruments. Inc. INTRODUCTION It is important that users of calibrated power meters and detectors understand and

More information

Open and Closed Door Moisture Transport and Corresponding Energy Consumption in Household Refrigerator

Open and Closed Door Moisture Transport and Corresponding Energy Consumption in Household Refrigerator 18 R. Saidur et al./journal of Energy & Environment, Vol. 6, May 2007 Open and Closed Door Moisture Transport and Corresponding Energy Consumption in Household Refrigerator R. Saidur, M. A. Sattar, M.

More information

The Shortwave Reflectance of the Yankee Environmental Systems Hotplate

The Shortwave Reflectance of the Yankee Environmental Systems Hotplate 1 - Introduction The Shortwave Reflectance of the Yankee Environmental Systems Hotplate David Siuta, Jefferson R. Snider and Bujidmaa Borkhuu University of Wyoming September 2010 (Updated September 2017)

More information

CALIBRATION HOW IT WORKS

CALIBRATION HOW IT WORKS HIGH POWER SOLUTIONS POWER DETECTORS ENERGY DETECTORS At Gentec-EO, we understand that the essence of our business since 40 years has been delivering accuracy. There are no half measures: it either measures

More information

LABORATORY STUDY ON THE CONSOLIDATION SETTLEMENT OF CLAY-FILLED GEOTEXTILE TUBE AND BAGS

LABORATORY STUDY ON THE CONSOLIDATION SETTLEMENT OF CLAY-FILLED GEOTEXTILE TUBE AND BAGS Journal of GeoEngineering, Vol. 6, No. 1, pp. Chew 41-45, et al.: April Laboratory 2011 Study on the Consolidation Settlement of Clay-Filled Geotextile Tube and Bags 41 LABORATORY STUDY ON THE CONSOLIDATION

More information

ISO AIR FILTERS FOR GENERAL VENTILATION: DETERMINING FRACTIONAL EFFICIENCY

ISO AIR FILTERS FOR GENERAL VENTILATION: DETERMINING FRACTIONAL EFFICIENCY ISO 16890-2 AIR FILTERS FOR GENERAL VENTILATION: DETERMINING FRACTIONAL EFFICIENCY APPLICATION NOTE AFT-005 (US) Contents Introduction to ISO 16890... 1 How ISO 16890 Compares to EN 779 and ASHRAE 52.2...

More information

A Statistical Analysis of a Liquid Desiccant Dehumidifier/Regenerator in an Air Conditioning System

A Statistical Analysis of a Liquid Desiccant Dehumidifier/Regenerator in an Air Conditioning System Int. J. of Thermal & Environmental Engineering Volume 5, No. 1 (013) 41-50 A Statistical Analysis of a Liquid Desiccant Dehumidifier/Regenerator in an Air Conditioning System Abdulrahman T. Mohammad *,

More information

EVALUATING LOOP HEAT PIPE PERFORMANCE IMPROVEMENT USING CIRCUMFERENTIAL GROOVES. Nadjara dos Santos, Roger R. Riehl

EVALUATING LOOP HEAT PIPE PERFORMANCE IMPROVEMENT USING CIRCUMFERENTIAL GROOVES. Nadjara dos Santos, Roger R. Riehl Abstract EVALUATING LOOP HEAT PIPE PERFORMANCE IMPROVEMENT USING CIRCUMFERENTIAL GROOVES Nadjara dos Santos, Roger R. Riehl National Institute for Space Research INPE - Space Mechanics and Control Division-DMC

More information

Design and performance of an automatic regenerating adsorption aerosol dryer for continuous operation at monitoring sites

Design and performance of an automatic regenerating adsorption aerosol dryer for continuous operation at monitoring sites Atmos. Meas. Tech., 2, 417 422, 2009 Author(s) 2009. This work is distributed under the Creative Commons Attribution 3.0 License. Atmospheric Measurement Techniques Design and performance of an automatic

More information

Prediction of Soil Infiltration Rate Based on Sand Content of Soil

Prediction of Soil Infiltration Rate Based on Sand Content of Soil Agricultural Engineering Research Journal 3 (): 0-4, 013 ISSN 18-3906 IDOSI Publications, 013 DOI: 10.589/idosi.aerj.013.3..1109 Prediction of Soil Infiltration Rate Based on Sand Content of Soil 1 1 Majid

More information

EXPERIMENTAL STUDY OF CENTRIFUGAL HUMIDIFIER FITTED IN AN INDUSTRIAL SHED LOCATED IN TROPICAL CLIMATES

EXPERIMENTAL STUDY OF CENTRIFUGAL HUMIDIFIER FITTED IN AN INDUSTRIAL SHED LOCATED IN TROPICAL CLIMATES THERMAL SCIENCE, Year 2011, Vol. 15, No. 2, pp. 467-475 467 EXPERIMENTAL STUDY OF CENTRIFUGAL HUMIDIFIER FITTED IN AN INDUSTRIAL SHED LOCATED IN TROPICAL CLIMATES by Krishnasamy SENTHILKUMAR a* and Pss

More information

The Effect of Quantity of Salt on the Drying Characteristics of Fresh Noodles

The Effect of Quantity of Salt on the Drying Characteristics of Fresh Noodles Available online at www.sciencedirect.com ScienceDirect Agriculture and Agricultural Science Procedia 2 ( 2014 ) 207 211 ST26943, 2nd International Conference on Agricultural and Food Engineering, CAFEi2014

More information

Optical and Scene Networks Frostburg, MD October 25, John V. Molenar Air Resource Specialists, Inc.

Optical and Scene Networks Frostburg, MD October 25, John V. Molenar Air Resource Specialists, Inc. Optical and Scene Networks Frostburg, MD October 25, 2011 John V. Molenar Air Resource Specialists, Inc. Number of Optical Monitors 70 60 50 40 30 20 10 0 64 67 65 52 50 48 46 29 25 2003 2004 2005 2006

More information

IMPROVE Aerosol Monitoring Network

IMPROVE Aerosol Monitoring Network IMPROVE Aerosol Monitoring Network Bret A. Schichtel and William C. Malm National Park Service and CIRA, Colorado State University Rodger B. Ames CIRA, Colorado State University Robert A. Eldred and Lowell

More information

Experimental Study to Evaluate Smoke Stratification and Layer Height in Highly Ventilated Compartments

Experimental Study to Evaluate Smoke Stratification and Layer Height in Highly Ventilated Compartments Experimental Study to Evaluate Smoke Stratification and Layer Height in Highly Ventilated Compartments Jason Huczek a, Marc Janssens a, Kentaro Onaka b, Stephen Turner c a SwRI, 6220 Culebra Road, San

More information

CFD Analysis of temperature dissipation from a hollow metallic pipe through circular fins using Ansys 14.5

CFD Analysis of temperature dissipation from a hollow metallic pipe through circular fins using Ansys 14.5 IJAET International Journal of Application of Engineering and Technology ISSN: 2395-3594 Vol-1 No.-2 CFD Analysis of temperature dissipation from a hollow metallic pipe through circular fins using Ansys

More information

Available online Journal of Scientific and Engineering Research, 2017, 4(10): Research Article

Available online   Journal of Scientific and Engineering Research, 2017, 4(10): Research Article Available online www.jsaer.com, 2017, 4(10):305-311 Research Article ISSN: 2394-2630 CODEN(USA): JSERBR Low Cost Humidity/Temperature Calibration System Yasser A. Abdelaziz 1 Central Laboratories Department,

More information

Conceptual Design of a Better Heat Pump Compressor for Northern Climates

Conceptual Design of a Better Heat Pump Compressor for Northern Climates Purdue University Purdue e-pubs International Compressor Engineering Conference School of Mechanical Engineering 1976 Conceptual Design of a Better Heat Pump Compressor for Northern Climates D. Squarer

More information

LEAKAGE MONITORING OF SEAL PLUG FOR INDIAN PRESSURIZED HEAVY WATER REACTOR BY PSYCHOMETRICS ANALYSIS METHOD

LEAKAGE MONITORING OF SEAL PLUG FOR INDIAN PRESSURIZED HEAVY WATER REACTOR BY PSYCHOMETRICS ANALYSIS METHOD Proceedings of the National Seminar & Exhibition on Non-Destructive Evaluation NDE 2011, December 8-10, 2011 LEAKAGE MONITORING OF SEAL PLUG FOR INDIAN PRESSURIZED HEAVY WATER REACTOR BY PSYCHOMETRICS

More information

CEE:3371 Principles of Hydraulics and Hydrology Project #2 Flow Measurement with a Weir

CEE:3371 Principles of Hydraulics and Hydrology Project #2 Flow Measurement with a Weir CEE:3371 Principles of Hydraulics and Hydrology Project #2 Flow Measurement with a Weir Problem Statement The Iowa DNR plans to monitor a proposed prairie restoration project in eastern Iowa as an experiment.

More information

Experimental Investigation of a Multi Effect Membrane Based Regenerator for High Concentration Aqueous LiCL Solution as Desiccant

Experimental Investigation of a Multi Effect Membrane Based Regenerator for High Concentration Aqueous LiCL Solution as Desiccant Proceedings of the World Congress on Momentum, Heat and Mass Transfer (MHMT 16) Prague, Czech Republic April 4 5, 216 Paper No. ENFHT 111 DOI: 1.11159/enfht16.111 Experimental Investigation of a Multi

More information

Helicoil and Shell-Tube Heat Exchanger. Abdullah AlThomairi, Ahmed Sajjad, and Faris AlShuaibi. October 25, 2014

Helicoil and Shell-Tube Heat Exchanger. Abdullah AlThomairi, Ahmed Sajjad, and Faris AlShuaibi. October 25, 2014 Helicoil and Shell-Tube Heat Exchanger Abdullah AlThomairi, Ahmed Sajjad, and Faris AlShuaibi October 25, 2014 Submitted to: NORTHEASTERN UNIVERSITY Department of Chemical Engineering CHME 3313 Transport

More information

Jülich Solar Power Tower System Behavior During Downtime

Jülich Solar Power Tower System Behavior During Downtime Jülich Solar Power Tower System Behavior During Downtime Torsten Baumann 1, a), Felix Göhring 1, b), Hannes Stadler 1,c) and Till Doerbeck 2,d) 1 German Aerospace Center (DLR), Institute of Solar Research,

More information

TESTS OF ADSIL COATING

TESTS OF ADSIL COATING TESTS OF ADSIL COATING Test B - Long Term Test FSEC-CR-1259-01 July 11, 2001 Prepared for: Bob Suggs Florida Power & Light Company 9250 W. Flagler Street Miami, Florida 33174 Principal Investigator Dr.

More information

Evaluation the Effect of Washing on the Heat Transfer Capacity and Air-Side Flow Resistance of Air Cooled Condensers

Evaluation the Effect of Washing on the Heat Transfer Capacity and Air-Side Flow Resistance of Air Cooled Condensers University of Nebraska - Lincoln From the SelectedWorks of David Yuill Summer July 9, 2018 Evaluation the Effect of Washing on the Heat Transfer Capacity and Air-Side Flow Resistance of Air Cooled Condensers

More information

RADIATION BLOCKAGE EFFECTS BY WATER CURTAIN

RADIATION BLOCKAGE EFFECTS BY WATER CURTAIN , Volume 6, Number 4, p.248-254, 04 RADIATION BLOCKAGE EFFECTS BY WATER CURTAIN C.L. Choi Department of Building Services Engineering, The Hong Kong Polytechnic University, Hong Kong, China ABSTRACT Water

More information

PREDICTION OF THE PRESSURE DROP OF CO 2 IN AN EVAPORATOR USED FOR AIR COOLING ABSTRACT 1. INTRODUCTION 2. EXPERIMENTAL SET-UP AND PROCEDURE

PREDICTION OF THE PRESSURE DROP OF CO 2 IN AN EVAPORATOR USED FOR AIR COOLING ABSTRACT 1. INTRODUCTION 2. EXPERIMENTAL SET-UP AND PROCEDURE PREDICTION OF THE PRESSURE DROP OF CO 2 IN AN EVAPORATOR USED FOR AIR COOLING M. POIRIER, D. GIGUÈRE, Z. AIDOUN, M. OUZZANE Natural Resources Canada, CANMET Energy Technology Centre-Varennes 1615, Lionel

More information

Evaluation of a dynamic model for a cold climate counter flow air to air heat exchanger

Evaluation of a dynamic model for a cold climate counter flow air to air heat exchanger Evaluation of a dynamic model for a cold climate counter flow air to air heat exchanger Toke Rammer Nielsen, Associate Professor, Department of Civil Engineering, Technical University of Denmark; trn@byg.dtu.dk

More information

Some Modeling Improvements for Unitary Air Conditioners and Heat Pumps at Off-Design Conditions

Some Modeling Improvements for Unitary Air Conditioners and Heat Pumps at Off-Design Conditions Purdue University Purdue e-pubs International Refrigeration and Air Conditioning Conference School of Mechanical Engineering 2006 Some Modeling Improvements for Unitary Air Conditioners and Heat Pumps

More information

Investigation of a Novel Ceiling Panel for Heat and Moisture Control in Buildings

Investigation of a Novel Ceiling Panel for Heat and Moisture Control in Buildings Investigation of a Novel Ceiling Panel for Heat and Moisture Control in Buildings Melanie T. Fauchoux, PhD Student University of Saskatchewan http://www.engr.usask.ca/dept/mee melanie.fauchoux@usask.ca

More information

Humidity Generation and Humidity Measurement The Complete Guide

Humidity Generation and Humidity Measurement The Complete Guide OWLSTONE NANOTECH INC Humidity Generation and Humidity Measurement The Complete Guide Billy Boyle (billy.boyle@owlstone.co.uk) www.owlstonenanotech.com OW-002349-MC Keywords: humidity, gas sensors, calibration,

More information

Brine Generation Study

Brine Generation Study DOE/WIPP 00-2000 Brine Generation Study April 2000 Waste Isolation Pilot Plant Carlsbad, New Mexico Processing and final preparation of this report was performed by the Waste Isolation Pilot Plant Management

More information

STACK EFFECT IN LIGHT WELL OF HIGH RISE APARTMENT BUILDING

STACK EFFECT IN LIGHT WELL OF HIGH RISE APARTMENT BUILDING STACK EFFECT IN LIGHT WELL OF HIGH RISE APARTMENT BUILDING H. Kotani, R. Satoh, T. Yamanaka Dept. of Architectural Engineering, School of Engineering, Osaka University, Osaka, Japan ABSTRACT The purpose

More information

Refrigerant Mass and Oil Migration During Start-up Transient

Refrigerant Mass and Oil Migration During Start-up Transient Refrigerant Mass and Oil Migration During Start-up Transient Steffen Peuker, Predrag Hrnjak* Department of Mechanical and Industrial Engineering University of Illinois at Urbana-Champaign 1206 West Green

More information

Mathematical Simulation of Longan Fruit Drying

Mathematical Simulation of Longan Fruit Drying Mathematical Simulation of Longan Fruit Drying Aree Achariyaviriya, Somchart Soponronnarit and Jirawan Tiansuwan ABSTRACT The objectives of this study were to develop a mathematical model and to simulate

More information

Performance Study of Triple Concentric Pipe Heat Exchanger

Performance Study of Triple Concentric Pipe Heat Exchanger Performance Study of Triple Concentric Pipe Heat Exchanger #1 Pravin M. Linge, #2 A. B. Korane, #3 V. N. Kapatkar #123 Mechanical Engineering Department, Sinhgad College of Engineering, SPPU, Pune, India

More information

Publication American Institute of Physics. Reprinted with permission.

Publication American Institute of Physics. Reprinted with permission. P2 Publication 2 M. Noorma, P. Kärhä, A. Lamminpää, S. Nevas and E. Ikonen, Characterization of GaAsP trap detector for radiometric measurements in ultraviolet wavelength region, Rev. Sci. Instrum. 76,

More information

Optimizing Chemistry and Coater Settings for Faster Drying

Optimizing Chemistry and Coater Settings for Faster Drying Optimizing Chemistry and Coater Settings for Faster Drying David Keely, Field Tech Service Manager, The Dow Chemical Company, Louisville, KY Dr. Bill Griffith, Principal Scientist, The Dow Chemical Company,

More information

Meteorology 432. Hygrometry Spring 2013

Meteorology 432. Hygrometry Spring 2013 Meteorology 432 Hygrometry Spring 2013 Hygrometry Objective: Determine the amount of water vapor present in the atmosphere. Pure water vapor in equilibrium (saturated conditions) with a plane surface of

More information

ω = (1) Kirby S. Chapman Ph.D.

ω = (1) Kirby S. Chapman Ph.D. White Paper: Relative Humidity Impacts of the AirFloor System in the Built Environment Kirby S. Chapman Ph.D. kirby.chapman@scavengetech.com Introduction Relative humidity is specified as one of the seven

More information

Measure Lifetime Derived from a Field Study of Age at Replacement

Measure Lifetime Derived from a Field Study of Age at Replacement Measure Lifetime Derived from a Field Study of Age at Replacement David Robison, MicroGrid David Cohan Bruce True, Portland General Electric The traditional engineering technique for estimating the expected

More information

A STUDY ON THE BEHAVIOR OF STEAM CONDENSATION IN U-SHAPED HEAT TUBE

A STUDY ON THE BEHAVIOR OF STEAM CONDENSATION IN U-SHAPED HEAT TUBE A STUDY ON THE BEHAVIOR OF STEAM CONDENSATION IN U-SHAPED HEAT TUBE Satoru Ito 1, Keisuke Tsukada 1, Nobuyoshi Tsuzuki 2, Takao Ishizuka 3 and Hiroshige Kikura 3 1 Department of nuclear Engineering, Graduate

More information

An FT-NIR Primer. NR800-A-006

An FT-NIR Primer. NR800-A-006 Industry: Product: Oils and Gas NR800 Near Infra-red Spectroscopy Like other forms of spectroscopy, light from a lamp (usually called the source ) is passed through a sample and measured by a detector.

More information

A Fresnel Reflection-Based Optical Fiber Sensor System for Remote Refractive Index Measurement Using an OTDR

A Fresnel Reflection-Based Optical Fiber Sensor System for Remote Refractive Index Measurement Using an OTDR PHOTONIC SENSORS / Vol. 4, No. 1, 2014: 48 52 A Fresnel Reflection-Based Optical Fiber Sensor System for Remote Refractive Index Measurement Using an OTDR Jianying YUAN, Chunliu ZHAO *, Manping YE, Juan

More information

Product data sheet Palas Fidas 200 S

Product data sheet Palas Fidas 200 S Product data sheet Palas Fidas 200 S Applications Regulatory environmental monitoring in measuring networks Ambient air measurement campaigns Oriented indicative measurements (spot checks) Long-term studies

More information

CALIBRATION HOW IT WORKS

CALIBRATION HOW IT WORKS THZ DETECTORS PHOTO DETECTORS HIGH POWER DETECTORS POWER DETECTORS ENERGY DETECTORS MONITORS At Gentec-EO, we understand that the essence of our business since 40 years has been delivering accuracy. There

More information

MODELLING AND OPTIMIZATION OF DIRECT EXPANSION AIR CONDITIONING SYSTEM FOR COMMERCIAL BUILDING ENERGY SAVING

MODELLING AND OPTIMIZATION OF DIRECT EXPANSION AIR CONDITIONING SYSTEM FOR COMMERCIAL BUILDING ENERGY SAVING MODELLING AND OPTIMIZATION OF DIRECT EXPANSION AIR CONDITIONING SYSTEM FOR COMMERCIAL BUILDING ENERGY SAVING V. Vakiloroaya*, J.G. Zhu, and Q.P. Ha School of Electrical, Mechanical and Mechatronic Systems,

More information

Pneumatic Atomization of a Viscous Solution for Spray Drying: Investigation of Droplet and Particle Formation.

Pneumatic Atomization of a Viscous Solution for Spray Drying: Investigation of Droplet and Particle Formation. ILASS Americas, 19 th Annual Conference on Liquid Atomization and Spray Systems, Toronto, Canada, May 2006 Pneumatic Atomization of a Viscous Solution for Spray Drying: Investigation of Droplet and Particle

More information

by Richard W. Bukowski Center for Fire Research Program for Fire Detection and Control Systems National Bureau of Standards Washington,

by Richard W. Bukowski Center for Fire Research Program for Fire Detection and Control Systems National Bureau of Standards Washington, LETTER TO THE EDITOR by Richard W. Bukowski Center for Fire Research Program for Fire Detection and Control Systems National Bureau of Standards Washington, DC Reprinted from The Journal of Fire and Flammability,

More information

Reducing energy consumption of airconditioning systems in moderate climates by applying indirect evaporative cooling

Reducing energy consumption of airconditioning systems in moderate climates by applying indirect evaporative cooling Reducing energy consumption of airconditioning systems in moderate climates by applying indirect evaporative cooling Aleksandra Cichoń,*, Anna Pacak, Demis Pandelidis, and Sergey Anisimov Department of

More information

Temperature dependent Components within the Pyroelectric Detector

Temperature dependent Components within the Pyroelectric Detector 1 Introduction This application note is offered to provide some insight into how temperature affects the various components of a typical pyroelectric detector and how the effects of ambient temperature

More information

The SAG Aerosol Urs Baltensperger

The SAG Aerosol Urs Baltensperger The SAG Aerosol Urs Baltensperger Laboratory of Atmospheric Chemistry Paul Scherrer Institut, CH-5232 Villigen PSI Web site: http://gaw.tropos.de:8087/ (Host: IfT Leipzig) GAW2005 Workshop WMO, 14-16 March

More information

Optimization of Sensitivity Characteristics of Photoelectric Smoke Detector to Various Smokes

Optimization of Sensitivity Characteristics of Photoelectric Smoke Detector to Various Smokes Optimization of Sensitivity Characteristics of Photoelectric Smoke Detector to Various Smokes T NAGASHIMA, A MAMMOTO, Y YAMAUCHI, K WATANABE and M MIZOBUCHI R&D Laboratories Hochiki Corporation 246 Tsuruma,

More information

TRANSMISSIVITY BEHAVIOR OF SHREDDED SCRAP TIRE DRAINAGE LAYER IN LANDFILL COVER SYSTEM *

TRANSMISSIVITY BEHAVIOR OF SHREDDED SCRAP TIRE DRAINAGE LAYER IN LANDFILL COVER SYSTEM * TRANSMISSIVITY BEHAVIOR OF SHREDDED SCRAP TIRE DRAINAGE LAYER IN LANDFILL COVER SYSTEM * Krishna R. Reddy, Aravind Marella and Prasanth Ala University of Illinois at Chicago, Department of Civil and Materials

More information

THEORETICAL EQUILIBRIUM MOISTURE CONTENT OF WOOD UNDER VACUUM Zhangjing Chen? Fred M. Lamb1

THEORETICAL EQUILIBRIUM MOISTURE CONTENT OF WOOD UNDER VACUUM Zhangjing Chen? Fred M. Lamb1 THEORETICAL EQUILIBRIUM MOISTURE CONTENT OF WOOD UNDER VACUUM Zhangjing Chen? Research Associate Fred M. Lamb1 Professor Department of Wood Science and Forest Products Virginia Tech Blacksburg, VA 2406

More information

AN ANALYSIS OF THE PERFORMANCE OF RESIDENTIAL SMOKE DETECTION TECHNOLOGIES UTILIZING THE CONCEPT OF RELATIVE TIME

AN ANALYSIS OF THE PERFORMANCE OF RESIDENTIAL SMOKE DETECTION TECHNOLOGIES UTILIZING THE CONCEPT OF RELATIVE TIME AN ANALYSIS OF THE PERFORMANCE OF RESIDENTIAL SMOKE DETECTION TECHNOLOGIES UTILIZING THE CONCEPT OF RELATIVE TIME Elizabeth L. Milarcik, Stephen M. Olenick*, and Richard J. Roby Combustion Science & Engineering,

More information

AN ANALYSIS OF THE ACTUAL THERMAL PLUMES OF KITCHEN APPLIANCES DURING COOKING MODE

AN ANALYSIS OF THE ACTUAL THERMAL PLUMES OF KITCHEN APPLIANCES DURING COOKING MODE AN ANALYSIS OF THE ACTUAL THERMAL PLUMES OF KITCHEN APPLIANCES DURING COOKING MODE R Kosonen 1,*, H Koskela 2 and P Saarinen 2 1 Halton Oy, Haltonintie 1-3, 474 Kausala, Finland 2 Finnish Institute of

More information

AHRI Standard 1250P (I-P) 2009 Standard for Performance Rating of Walk-In Coolers and Freezers

AHRI Standard 1250P (I-P) 2009 Standard for Performance Rating of Walk-In Coolers and Freezers AHRI Standard 1250P (I-P) 2009 Standard for Performance Rating of Walk-In Coolers and Freezers - ERRATA SHEET FOR AHRI STANDARD 1250 (I-P)-2009, PERFORMANCE RATING OF WALK-IN COOLERS AND FREEZERS December

More information

Analysis of Pullout Resistance of Soil-Nailing in Lateritic Soil

Analysis of Pullout Resistance of Soil-Nailing in Lateritic Soil Analysis of Pullout Resistance of Soil-Nailing in Lateritic Soil B,L.A. Isaka 1, B.C. Madushanka 1 and N.H. Priyankara 1 1 Department of Civil and Environmental Engineering Faculty of Engineering University

More information

NEBRASKA MODIFIED ROOF POND: 1985 SUMMER PERFORMANCE RESULTS

NEBRASKA MODIFIED ROOF POND: 1985 SUMMER PERFORMANCE RESULTS NEBRASKA MODIFIED ROOF POND: 1985 SUMMER PERFORMANCE RESULTS by Drs. Bing Chen, Raymond Guenther, John Maloney, and John Kasher and Chuck Sloup, Jay Kratochvil, Tom Goetzinger Passive Solar Research Group

More information

Lab 1b Cooling Tower Performance and Sensors for Thermal-Fluid Systems

Lab 1b Cooling Tower Performance and Sensors for Thermal-Fluid Systems Lab 1b Cooling Tower Performance and Sensors for Thermal-Fluid Systems OBJECTIVES Warning: though the experiment has educational objectives (to learn about boiling heat transfer, etc.), these should not

More information

MODEL-BASED OPTIMIZATION OF AN INFRARED GAS SENSOR

MODEL-BASED OPTIMIZATION OF AN INFRARED GAS SENSOR MODEL-BASED OPTIMIZATION OF AN INFRARED GAS SENSOR ABSTRACT Ingo Sieber Forschungszentrum Karlsruhe P.O. Box 3640 76021 Karlsruhe Germany sieber@iai.fzk.de Manufacturing test structures of microsensors

More information

INDOOR CLIMATE IN HEATING CONDITION OF A LARGE GYMNASIUM WITH UNDER-FLOOR SUPPLY/RETURN SYSTEM

INDOOR CLIMATE IN HEATING CONDITION OF A LARGE GYMNASIUM WITH UNDER-FLOOR SUPPLY/RETURN SYSTEM INDOOR CLIMATE IN HEATING CONDITION OF A LARGE GYMNASIUM WITH UNDER-FLOOR SUPPLY/RETURN SYSTEM Mingjie Zheng Research Laboratory, SANKO AIR CONDITIONING CO., LTD. Nagoya, 450-0003, JAPAN ABSTRACT In large

More information

An experimental study of the impact of tunnel suppression on tunnel ventilation

An experimental study of the impact of tunnel suppression on tunnel ventilation An experimental study of the impact of tunnel suppression on tunnel ventilation Yoon J. Ko and George Hadjisophocleous Civil and Environmental Engineering, Carleton University 1125 Colonel By Drive, Ottawa,

More information

EFFECT OF COMPACTION ON THE UNSATURATED SHEAR STRENGTH OF A COMPACTED TILL

EFFECT OF COMPACTION ON THE UNSATURATED SHEAR STRENGTH OF A COMPACTED TILL EFFECT OF COMPACTION ON THE UNSATURATED SHEAR STRENGTH OF A COMPACTED TILL Vanapalli, S.K., Pufahl, D.E., and Fredlund, D.G. (University of Saskatchewan, Saskatoon, SK., Canada, S7N 5A9) Abstract An experimental

More information

The New DUSTTRAK II and DRX Aerosol Monitors

The New DUSTTRAK II and DRX Aerosol Monitors The New DUSTTRAK II and DRX Aerosol Monitors HEALTH AND SAFETY Exposure Monitoring TRUST. SCIENCE. INNOVATION. Real-Time Dust Monitoring Takes A Giant Leap Forward. The Revolution in Real-Time Dust Monitoring

More information

Calculated Energy Savings for Disabling Anti- Sweat Door Heat on Glass Display Case Doors

Calculated Energy Savings for Disabling Anti- Sweat Door Heat on Glass Display Case Doors Calculated Energy Savings for Disabling Anti- Sweat Door Heat on Glass Display Case Doors ET Project Number: Product Manager: Sachin Andhare Pacific Gas and Electric Company` Prepared By: Danielle Geers

More information

Humidity sensors. Behavior in different environments. Supervisor: Vaclav Vacek Collaborators: Martin Doubek, Martin Janda and Michal Vitek 28.9.

Humidity sensors. Behavior in different environments. Supervisor: Vaclav Vacek Collaborators: Martin Doubek, Martin Janda and Michal Vitek 28.9. Humidity sensors Behavior in different environments Supervisor: Vaclav Vacek Collaborators: Martin Doubek, Martin Janda and Michal Vitek 28.9.29 CZECH TECHNICAL UNIVERSITY IN PRAGUE CERN Content Content...

More information

Analysis of Evaporative Cooler and Tube in Tube Heat Exchanger in Intercooling of Gas Turbine

Analysis of Evaporative Cooler and Tube in Tube Heat Exchanger in Intercooling of Gas Turbine IJIRST International Journal for Innovative Research in Science & Technology Volume 1 Issue 11 April 2015 ISSN (online): 2349-6010 Analysis of Evaporative Cooler and Tube in Tube Heat Exchanger in Intercooling

More information

Construction of a Calibration System for Power Responsivity of Optical detectors Using Different Laser Sources

Construction of a Calibration System for Power Responsivity of Optical detectors Using Different Laser Sources J. Basic. Appl. Sci. Res., ()971-976, 01 01, TextRoad Publication ISSN 090-4304 Journal of Basic and Applied Scientific Research www.textroad.com Construction of a Calibration System for Power Responsivity

More information

A Method for Predicting the Matric Suction of Unsaturated Soils with Soil Color Recognition

A Method for Predicting the Matric Suction of Unsaturated Soils with Soil Color Recognition 2017 2 nd International Conference on Architectural Engineering and New Materials (ICAENM 2017) ISBN: 978-1-60595-436-3 A Method for Predicting the Matric Suction of Unsaturated Soils with Soil Color Recognition

More information

A Cost Effective Multi-Spectral Scanner for Natural Gas Detection

A Cost Effective Multi-Spectral Scanner for Natural Gas Detection A Cost Effective Multi-Spectral Scanner for Natural Gas Detection Semi-Annual Progress Report No. 1 Reporting Period Start Date: October 1, 2003 Reporting Period End Date: March 31, 2004 Principal Authors:

More information

14 Drying. I Basic relations and definitions. Oldřich Holeček, Martin Kohout

14 Drying. I Basic relations and definitions. Oldřich Holeček, Martin Kohout 14 Drying Oldřich Holeček, Martin Kohout I Basic relations and definitions Drying is a process which is used for removing the liquid from the solid material. In standard chemical engineering practice drying

More information

Modeling of Ceiling Fan Based on Velocity Measurement for CFD Simulation of Airflow in Large Room

Modeling of Ceiling Fan Based on Velocity Measurement for CFD Simulation of Airflow in Large Room Modeling of Ceiling Fan Based on Velocity Measurement for CFD Simulation of Airflow in Large Room Y. Momoi 1, K. Sagara 1, T. Yamanaka 1 and H. Kotani 1 1 Osaka University, Graduate School of Eng., Dept.

More information

The New DUSTTRAK II and DRX Aerosol Monitors

The New DUSTTRAK II and DRX Aerosol Monitors The New DUSTTRAK II and DRX Aerosol Monitors HEALTH AND SAFETY Exposure Monitoring TRUST. SCIENCE. INNOVATION. Real-Time Dust Monitoring Takes A Giant Leap Forward. The Revolution in Real-Time Dust Monitoring

More information

Development of a Transient Simulation Model of a Freezer Part II: Comparison of Experimental Data with Model

Development of a Transient Simulation Model of a Freezer Part II: Comparison of Experimental Data with Model Purdue University Purdue e-pubs International Refrigeration and Air Conditioning Conference School of Mechanical Engineering 199 Development of a Transient Simulation Model of a Freezer Part II: Comparison

More information

Summary of Comments (Washington Revisions November 7, 2000) Update November 27, 2000

Summary of Comments (Washington Revisions November 7, 2000) Update November 27, 2000 SAE Alternate Refrigerant Cooperative Research Program Summary of Comments (Washington Revisions November 7, 2000) Update November 27, 2000 To: Alternate Refrigerant Task Force Members From: Ward Atkinson

More information

Link loss measurement uncertainties: OTDR vs. light source power meter By EXFO s Systems Engineering and Research Team

Link loss measurement uncertainties: OTDR vs. light source power meter By EXFO s Systems Engineering and Research Team Link loss measurement uncertainties: OTDR vs. light source power meter By EXFO s Systems Engineering and Research Team INTRODUCTION The OTDR is a very efficient tool for characterizing the elements on

More information

The effect of write current on thermal flying height control sliders with dual heater/insulator elements

The effect of write current on thermal flying height control sliders with dual heater/insulator elements Microsyst Technol (2011) 17:959 964 DOI 10.1007/s00542-010-1216-4 TECHNICAL PAPER The effect of write current on thermal flying height control sliders with dual heater/insulator elements Hao Zheng Hui

More information

Theoretical Performance Analysis Of Direct Evaporative Cooler In Hot And Dry Climates

Theoretical Performance Analysis Of Direct Evaporative Cooler In Hot And Dry Climates INTERNATIONAL JOURNAL OF SCIENTIFIC & TECHNOLOGY RESEARCH VOLUME 3, ISSUE 4, APRIL 214 ISSN 2277-8616 Theoretical Performance Analysis Of Direct Evaporative Cooler In Hot And Dry Climates Ibrahim U. Haruna,

More information

CROWD-SOURCED REMOTE ASSESSMENTS OF REGIONAL-SCALE POST-DISASTER DAMAGE

CROWD-SOURCED REMOTE ASSESSMENTS OF REGIONAL-SCALE POST-DISASTER DAMAGE Eleventh U.S. National Conference on Earthquake Engineering Integrating Science, Engineering & Policy June 25-29, 2018 Los Angeles, California CROWD-SOURCED REMOTE ASSESSMENTS OF REGIONAL-SCALE POST-DISASTER

More information

Assessment of aluminium stress tolerance of triticale breeding lines in hydroponics

Assessment of aluminium stress tolerance of triticale breeding lines in hydroponics Assessment of aluminium stress tolerance of triticale breeding lines in hydroponics X.G. Zhang 1, 2, R.S. Jessop 2, F. Ellison 3, and P. Evans 1 1 Agriculture Victoria, Pastoral and Veterinary Institute,

More information

Thick-Film Heater achieves Superior Performance in Thermal Response, Uniformity and Efficiency.

Thick-Film Heater achieves Superior Performance in Thermal Response, Uniformity and Efficiency. Thick-Film Heater achieves Superior Performance in Thermal Response, Uniformity and Efficiency. by Mary Ruggiero, P.Eng., PhD and John Stockton, P.Eng Abstract IntegrAL thick-film heaters by Datec Corporation

More information

Performance Investigation of Refrigerant Vapor- Injection Technique for Residential Heat Pump Systems

Performance Investigation of Refrigerant Vapor- Injection Technique for Residential Heat Pump Systems Purdue University Purdue e-pubs International Refrigeration and Air Conditioning Conference School of Mechanical Engineering 2008 Performance Investigation of Refrigerant Vapor- Injection Technique for

More information