PM1703MO-1A PM1703MO-1B PERSONAL COMBINED RADIATION DETECTOR/DOSIMETER. Operation Manual

Size: px
Start display at page:

Download "PM1703MO-1A PM1703MO-1B PERSONAL COMBINED RADIATION DETECTOR/DOSIMETER. Operation Manual"

Transcription

1 PM1703MO-1 PM1703MO-1A PM1703MO-1B PERSONAL COMBINED RADIATION DETECTOR/DOSIMETER Operation Manual

2 TABLE OF CONTENTS 1 DESCRIPTION AND OPERATION Application Delivery kit Specifications Description Design Battery Battery charger OPERATION Pre-operation General Safety precautions Battery installation Functional check Operation Turning the instrument ON/OFF Low battery indication Operation modes Search mode. Detection and location of gamma sources DER measurement mode DE measurement mode Setup Communication with computer Alarm thresholds Dose rate alarm thresholds Dose alarm threshold Permanent memory of the instrument MAINTENANCE TROUBLESHOOTING CALIBRATION TEST PROCEDURE (REFERENCE) STORAGE AND SHIPPING WARRANTY ANNEX A - Typical energy response of the instrument in search mode ANNEX B - Typical instrument anisotropy

3 Thank you for purchasing a Polimaster Personal Combined Radiation Detector/Dosimeter. Before operating this unit, please review this manual thoroughly and use the instrument for its intended application. When in use, don't keep the instrument in immediate proximity to radio emitting sources, such as cell phones, to avoid false alarms. Avoid mechanical shocks and damages, corrosive media, organic solvents, open fire, submerging the instrument into water and other unfavorable effects. Observe the radioactive materials handling and radiation safety standards and regulations in case of detection of radiation sources/emissions. 3

4 1 DESCRIPTION AND OPERATION 1.1 Application 1) Personal Combined Radiation Detector/Dosimeter PM1703MO-1 PM1703MO-1A PM1703MO-1B (hereinafter referred to as the instrument ) is designed to measure ambient dose equivalent rate (DER) * (10) of gamma and X-ray (photon) radiation, ambient dose equivalent (DE) Н * (10), to search for (detect and locate) radioactive materials and alerting the user with audible, visual and vibration alarms. The instrument is available in three versions: - PM1703MO-1 - PM1703MO-1A differs from the PM1703MO-1 by additional USB interface for communication with a personal computer (PC) - PM1703MO-1B differs from the PM1703MO-1 by additional Bluetooth radio channel for communication with a Pocket PC (PPC) or PC. The instrument is intended for use by customs and border guards, personnel of nuclear power plants, isotope laboratories, first responders, firefighters, military, police, etc. All operations history is stored in the instrument s permanent memory, protecting the data even when the battery is removed. The stored data can be transferred to a personal computer via infrared interface (all models), Bluetooth (РМ1703МО-1B) and USB interface (РМ1703МО-1A). 1 Some changes may be introduced in the instrument electrical circuit, design and software that do not influence the specifications and, therefore, may be not specified in this manual. 4

5 1.2 Delivery Kit The instrument s delivery kit is shown in Table 1. Table 1 Description Quantity, pcs. PM1703MO-1 PM1703MO-1A PM1703MO-1B PM1703MO-1 Personal Combined Radiation Detector/Dosimeter PM1703MO-1A Personal Combined Radiation Detector/Dosimeter PM1703MO-1B Personal Combined Radiation Detector/Dosimeter Battery (Alkaline) 1.5 V, min ma/h, AA (LR6) 1) Cable USB Operation manual Package Accessories: - software 1 CD 2 CD 2 CD - flash Card with software for Pocket PC charger 2) holster 2) infra red adapter 2) ) or rechargeable battery Ni-MH (Nickel Metal Hydride) 1.2 V, not less than 2000 ma/h, AA (R6) on customer request. Other batteries with similar parameters are allowed 2) optional items 5

6 1.3 Specifications Detector DER indication range DER measurement range Maximum permissible intrinsic relative error of DER measurement in measurement range CsI (Tl) scintillator Geiger-Muller tube 0.01 µsv/h Sv/h 0.1 µsv/h Sv/h ( 20 K 1 H K2 H ) %, where - DER value, msv/h K 1 coefficient msv/h K 2 coefficient (msv/h) µsv Sv 0.1 µsv Sv DE indication range DE measurement range Maximum permissible intrinsic relative error of DE measurement in measurement range 20 % Sensitivity in search mode, not less: - PM1703MO cps/(µsv/h) for 241 Am 85 cps/(µsv/h) for 137 Сs - PM1703MO-1A, PM1703MO-1B 200 cps/(µsv/h) for 241 Am 100 cps/(µsv/h) for 137 Сs Energy range in DER measurement mode Energy response in DER measurement mode relative to MeV ( 137 Cs) Energy range in Search mode (Typical energy response in search mode is shown in Annex A) Typical instrument anisotropy MeV 30 % MeV see Annex B Response (time to alarm) when the radiation intensity at the instrument s reference point is increased by 0.5 μsv/h in a period of not more than 0.5 seconds using 241 Am, 137 Сs, 60 Сo gamma sources (at n-coefficient set to a value corresponding to mean time to false alarm > 10 hours and ambient background level of not more than 0.25 µsv/h), not more than 2 s Response (time to alarm) when the radiation intensity at the instrument s reference point is increased gradually by 0.5 μsv/h using a 137 Сs gamma source at a speed of not more than 0.5 m/s (at n-coefficient set to a value corresponding to mean time to false alarm > 10 hours and ambient background level of not more than 0.25 µsv/h), not more than 2 s Response time (display indication of new dose rate with a relative error of ±20 %) following an increase in radiation field by 1 µsv/h in a period of not more than 0.5 seconds produced by a 137 Cs source, not more than 4 s Detection of gamma radiation sources with 0.95 probability at the n-coefficient set to a value corresponding to mean time to false alarm > 10 hour and ambient background level of not more than 0.1 µsv/h according to Table 2 6

7 PM1703MO-1A, PM1703MO-1B: Detection of gamma radiation sources with 0.95 probability at the n-coefficient set to a value corresponding to mean time to false alarm > 1 hour and ambient background level of not more than 0.25 µsv/h according to Table 3 Rate of false alarms in search mode: - at n-coefficient = at n-coefficient = 5.3 Background calibration Type of alarm Communication with PC (reading data from storage, setting parameters) - all models - additional reading of data from storage, identification of isotopes with PC or PPC: PM1703MO-1A PM1703MO-1B Power requirements Battery life (at average ambient background 0.3 µsv/h, LCD backlight, audible and vibration alarms 5 min/24 hours) Battery life at continuous communication with PC via mean time to false alarm > 1 h mean time to false alarm > 10 h - automatic upon power on - auto calibration after background variations and change of n-coefficient - user initiated as needed by pressing a button - audible - vibration - light infra red USB interface Bluetooth radio channel AA (LR6) battery (Alkaline) 1.5 V, min ma/h or rechargeable battery AA (R6) (Ni-MH (Nickel Metal Hydride)) 1,5 V, min ma/h 1000 h Bluetooth (PM1703MO-1B) Data collection Environmental: - temperature range 28 h 2000 data points -30 to +50 С (LCD indication from -20 to +50 С) up to 98 % at 35 C 84 to kpa IP65 - relative humidity - pressure Ingress protection Drop test 0.7 m onto concrete Dimensions: PM1703MO-1 72 x 32 x 87 mm (2 13/16 " x 1 ¼" x 3 7/16 ") PM1703MO-1A PM1703MO-1B 75 x 35 x 98 mm (2 15/16 " x 1 3/8 " x 3 7/8 ") Weight, including battery, max.: 250 g 7

8 Table 2 Parameter 241 Am Source 137 Cs Activity of gamma source, MBq 30 % Average velocity (source/instrument), m/s Source-to-detector surface distance, m Table 3 Parameter 133 Ba Source 137 Cs Activity of gamma source, kbq (µci) 55.0 (1.5) (2.7) 50.0 (1.35) Velocity (source/instrument), m/s Source-to-detector surface distance, m Co 60 Co 8

9 1.4 Description Design The body of the instrument is made of a shockproof plastic protected against water jets and dust. The instrument, its functional parts, detectors, their effective centers and direction of calibration are shown in Figure 1. The LCD screen may display any of the following information: - Dose Equivalent Rate (DER); - Dose Equivalent (DЕ); - n-coefficient, which determines the alarm threshold in the Search mode (also called alarm multiplier ); - messages: "tst" testing, "P Х,Х"- firmware version, "CAL" - calibration, "OL" overload, "off" the instrument is off, "bat" battery status, "to" temperature overload, etc. - audible alarm sign (if audible alarm is enabled) - vibration alarm sign (if vibration alarm is enabled) - Bluetooth sign (if Bluetooth is enabled) - blinking h (history) symbol (in the Search and DER measurement modes) - indicates that there has been DER alarm activation and information about it hasn t been uploaded to a computer - blinking R (respond) symbol indicates that the Search mode is currently on - - indicates critical low battery. An analogue bar in the lower area of the LCD display indicates the time left until self-tests or background calibration are completed. In the Search mode, DER measurement mode and DE measurement mode the analogue scale segments represent the dynamics of radiation environment change. The instrument is provided with a detachable clip (item 9, Figure 1) allowing the user to affix the instrument to a waist belt for hand-free operation. The clip can be removed as shown in Figure 2 a). Draw aside the spring from the latch using a screwdriver and carefully move the clip towards the face panel. Install the clip as shown in Figure 2 b). Put the clip on the bracket and move until it latches. The instrument can come complete with an optional protective holster made of synthetic material which also can be worn on a belt. It s recommended to remove the clip while using the holster. 9

10 1 MODE button - used for switching the instrument ON; switching between operation modes; background update (calibration); changing parameters in Setup mode. 2 - LIGHT button used for switching on the backlight; switching on the indication of the battery power level; enabling communication with a PC; changing parameters in Setup mode, switching the instrument OFF. 3 - infra red (IR) window 4 - red light-emitting-diode (LED) alarm 5 liquid crystal display (LCD) 6 audible alarm 7 battery 8 battery compartment lid 9 clip 10 USB connector closed with a plastic plug and marked. Figure 1 10

11 a) b) Figure 2 11

12 1.4.2 Battery The instrument is powered by an Alkaline battery, 1.5 V min 2000 ma/h, AA (LR6) size supplied with the instrument. It is allowed to use a rechargeable battery NiMH (Nickel Metal Hydride), 1.2 V min 2000 ma/h, AA (R6) size Battery Charger The battery charger is available as an option. WARNING! USE THE BATTERY CHARGER ONLY WHEN RECHARGEABLE BATTERY RECOMMENDED BY THE MANUFACTURER IS INSTALLED. ATTEMPTING TO CHARGE NON-RECHARGEABLE BATTERIES MAY CAUSE DAMAGE TO THE INSTRUMENT! The battery charger is combined with the instrument holder and is designed for charging the battery installed in the instrument. Install the charger on the dash panel of an automobile or any other horizontal surface. There are four 3.5-mm dia. holes to fix the charger. To protect the instrument from sun thermal effect, the charger can be equipped with an optional protective screen. Place the instrument in the holder and fix with its clip as described below (see Figure 3). Depending on the battery charger version, it may be connected to a standard automobile cigarette lighter socket 12 V or domestic supply network: Figure 3 - install the holder on the automobile s dash panel (or other surface) and fasten it with four screws (not included in the delivery package) - Move the instrument clip down by pressing on it - Install the instrument to the holder and release the clip, ensure that the clip claw is firmly fixed in the holder - If the power is available and the charger is properly connected to the vehicle power supply or domestic supply network, the green LED will be activated on the charger s back side - If the instrument is properly installed in the holder, the battery sign will be activated on the LCD until the battery completely charges. 12

13 2 OPERATION CAUTION! WHEN THE INSTRUMENT OPERATES AT TEMPERATURES BELOW minus 20 o C, THE LCD IS NOT FAIL-PROOF. RELY ONLY ON THE AUDIBLE OR VIBRATION ALARM WHEN DETECTING RADIATION SOURCES. THE LCD RESUMES ITS NORMAL OPERATION WHEN TEMPERATURE CONDITIONS ARE ABOVE -20 o C. When the temperature inside the instrument exceeds the upper limit, the message to will be displayed. The instrument functionality will be restored after it cools down. In this case it is recommended to re-calibrate instrument. 2.1 Pre-Operation General Check the instrument s delivery kit according to paragraph 1.2 and its functionality according to paragraph Remove the instrument from its package Before operating the instrument, review this operation manual thoroughly and retain it for the future references Safety Precautions The instrument is a safe product provided it is used for its intended application and according to the operating manual. The instrument is powered by a 1.5 V battery. There are no live parts with dangerous voltage and current accessible under normal operation. RF electromagnetic fields, electrostatic fields, equivalent and maximum level of sound and level of local vibration comply with safety requirements. Observe the radioactive materials handling and radiation safety standards and regulations during adjustment, repair, maintenance and calibration of the instrument, if radioactive sources are used or detected Battery Installation The instrument is packaged without a battery installed. To install the battery, unscrew cover 8 of the battery compartment (as shown in Figure 1) unit with a coin or screwdriver. Once opened, insert a single AA battery loading the positive (+) end first. Reinstall the cover and tighten it. Once the battery has been installed, the instrument will turn on automatically Functional Check The instrument must correctly go through the self-tests, background calibration and enter the operation mode (2.1.1) and other modes (2.2.3). 2.2 Operation Turning the Instrument ON/OFF Once the battery has been installed, the instrument will turn ON automatically. To turn the instrument ON in the process of operation (if it was turned off by control buttons) it is necessary to press the Mode button. All display segments and alarms (audible, LED and vibration) will be activated for approximately 1 second and the instrument will 13

14 than perform its self-test procedures. The display will indicate for few seconds the firmware version (for example, Р-1.7):, then bat message, flashing icon and the analogue scale, which represents remaining battery power:, then tst message and the analogue with decreasing segments which countdown the test time:. After self-tests are completed the instrument goes into the background calibration mode, in which LCD indicates CAL and the analog scale is being filled up:. During calibration the instrument measures radiation background and calculates the detection alarm threshold using the n-coefficient values (see ). When the background calibration is finished, the instrument goes into the Dose rate measurement mode and is ready for operation. When the instrument is turned ON, it continuously monitors radiation environment, measures DER and DE. To turn the instrument OFF, press and hold the LIGHT button for 6-7 seconds until the LCD screen displays the flashing OFF message. Confirm switching instrument off by pressing the MODE button. The instrument is switched off if the LCD displays OFF message Low Battery Indication The instrument automatically monitors the battery voltage. When the battery voltage drops below critical value (approx V), the instrument LCD will indicate a blinking battery sign and signals will be produced (audible, light and/or vibration). This means that non-rechargeable battery must be replaced (or rechargeable battery must be charged). The instrument will work up to 8 hours after the appearance of the critical low battery warning (at a normal background radiation level). By pressing the MODE button the user can switch off the critical low battery warning for about 30 minutes. At the same time the alarm thresholds will remain turned on. To check the battery power level in any operation mode, press and hold the LIGHT button, the bat message and the remaining battery power in relative units on the analogue scale will be displayed. The instrument automatically detects battery type. If non-rechargeable battery is detected, the bat message will blink; if rechargeable battery is detected, the bat message will be still. 14

15 N O T E - USED NON-RECHARGEABLE BATTERY WITH VOLTAGE LOWER THAN NOMINAL MAY NOT BE RECOGNIZED CORRECTLY. THE MANUFACTURER IS NOT RESPONSIBLE FOR ANY MALFUNCTIONS CAUSED BY INCORRECT BATTERY USE Operation Modes - Self-tests - Background calibration - Search mode - Dose rate (DER) measurement mode - Dose (DE) measurement mode - Setup - Communication with computer Use the MODE button to switch between the main modes: Search DER DE Bluetooth (on/off) Search mode. Detection and Location of Gamma Radiation Sources In the Search mode the instrument displays the current DER value in Sv/h, msv/h, Sv/h and an R (Respond) symbol is blinking which indicates that the Search mode is on. The scale and the units of measurement are changed automatically. In the Search mode the instrument detects and locates sources of photon radiation in the DER range from 0.01 Sv/h to 0.1 msv/h (for 137 Cs). If the value of the dose rate exceeds the upper limit of this range, the LCD screen will display the overload message -OL-. In this case the user should switch to the DER mode to retain control over the radiation environment. ATTENTION! AFTER THE INSTRUMENT IS EXPOSED TO A DOSE RATE HIGHER THAN 0.1 SV/H, IT MAY PRODUCE ALARMS ( CHIRPS ) FOR SOME TIME EVEN AFTER THE RADIATION BACKGROUND RETURNS TO THE NORMAL CONDITION. THE DURATION DEPENDS ON THE RATE AND LENGTH OF THE EXPOSURE AND ATTRIBUTED TO PHYSICAL PROCESSES IN THE CSI DETECTOR AFTER THE EXPOSURE. The instrument automatically compares the average count rate from the detector with the alarm threshold value (the value calculated based on the background radiation measured during calibration and the n-coefficient set by the user). If the threshold value is exceeded, the instrument triggers an alarm. The n-coefficient changes the search alarm threshold (minimum detection level): the smaller is the value of n-coefficient, the smaller is the threshold value and the more sensitive is the instrument to changes in background radiation. However, this also increases the probability of false alarms (false positives). The n-coefficient is set by the user in the Setup mode if this function is enabled in the PC communication mode (the manufacturer preset n-coefficient value is 5.3). The value of n-coefficient can be set in the range of 1.0 to 9.9 through every

16 To update the background radiation levels, press and hold down the Mode button for more than 2 seconds until the message CAL appears in the display. The instrument is able to perform background update (calibration) automatically. This feature provides more convenient instrument usage and decreases the probability of false alarms when the radiation background is changed slowly and there are no alarm signals. The auto-calibration function can be enabled/disabled in the PC communication mode. The instrument is highly sensitive to radiation level variations. It can give alarm signals if, for example, it is moved from the outside (street where the radiation background is normally lower than in a room) to a building where there are materials including natural radioactive isotopes (K, Th, Ra, U), which produce a higher natural radiation background. These materials mostly include: concrete and similar materials containing sand, natural stone (especially granite), ceramic tile, glass, etc. In case the alarm is actuated, automatic calibration is not performed, therefore it is recommended that the user re-calibrate the instrument manually to accommodate the new background. You can also change the n-coefficient to alter the sensitivity threshold. When the background radiation decreases (when the instrument is moved from a higher level of background radiation to a lower one) the instrument will be re-calibrated automatically (if auto-calibration is enabled). This increases the instrument s sensitivity and is very useful. Detection of gamma radiation sources When searching for radiation sources hold the instrument with its side without clip directed towards the scanned object. For greater efficiency, scan the object (baggage, people, container, vehicle, etc.) closely and slowly. Use the light or vibration alarm when audible alarms are hard to hear or may not be advisable due to security considerations. Bear in mind that sensitivity of the instrument and false alarm rate are dependent on: - preset n-coefficient, and - background level during calibration used by the instrument to calculate the search alarm threshold. Auto-calibration function of the instrument automatically adjusts/calibrates to slow changes in the ambient radiation level and update background approximately every ten minutes with a decreasing ambient background level or in slightly longer periods of time with an increasing ambient background level. However, the instrument will automatically recalibrate itself only if there are no alarms or abrupt background changes over the algorithm-defined periods of time. False alarms Occasionally, the instrument may produce random signals, called false alarms. The number of false alarms depends on many factors including changes in the ambient background radiation level as well as changes to the system setup. Mean time to false alarm under normal conditions, if there are no mechanical impacts, is specified in technical specifications. Such alarms (light, audible and/or vibration) are irregular and can, therefore, be easily distinguished from true alarm signals produced by the gamma radiation sources which tend to have continuous repetition rate or increase as the instrument moves closer to a radiation source. Location of gamma radiation sources As the instrument moves closer to a radiation source, the frequency alarm signals increases, thus enabling the user to determine the direction toward the source location. 16

17 When the light, audible and/or vibration alarm signal frequency reaches the maximum value, pinpointing the source more closely is impossible without updating the background. During the background calibration it is important to keep the instrument in the same place if possible. Once the background calibration is completed, the user may then continue to search for the radiation source. This procedure may be repeated as many times as necessary until the gamma radiation source is located DER Measurement Mode In this mode the instrument LCD displays the current dose rate value in Sv/h, msv/h, Sv/h. The scale and units of measurement are changed automatically. The dose rate indication range is from 0.01 Sv/h to 9.99 Sv/h. If the value of the dose rate exceeds the instrument s upper limit (9.99 Sv/h), the LCD screen will display the overload message OL DE Measurement Mode In this mode the instrument LCD displays the value of the accumulated dose in Sv, msv, Sv. The scale and units of measurement are changed automatically. The dose indication range is from 0 Sv to 9.99 Sv. At the same time the analogue scale is filled. The number of displayed scale segments corresponds to the accumulated dose relative to the preset dose threshold level. If the value of the accumulated dose exceeds the instrument s upper limit (9.99 Sv), the LCD screen will display the overload message OL. To reset the accumulated dose value and start a new accumulation it is necessary to connect the instrument to a PC and use the Software or the control buttons, if this function is enabled in PC communication mode (see ). To reset the accumulated dose, press and hold down the Mode button until the message appears in the display. The dose indication mode is displayed on the LCD screen for 6-8 seconds, after that the instrument automatically switches to the Search mode Setup The instrument is supplied with the settings as shown in Table 4. Table 4 Factory (default) settings * ON /OFF Search mode (display is µsv/h) DER measurement mode (display in µsv/h, msv/h, Sv/h) DE measurement mode (display in µsv, msv, Sv) Auto calibration Possibility to change the n-coefficient by control buttons 17

18 Possibility to enable/disable alarm types by control buttons Audible alarm Vibration alarm LED (light) alarm DER data logging interval 60 minutes n-coefficient 5.3 * - The factory settings can be changed before shipment on customer request Using the instrument front panel buttons it is possible to adjust (change) several parameters of the instrument: - n-coefficient (setup range 1.0 to 9.9 with increment of 0.1) - enable or disable the Audible alarm on reaching the alarm threshold - enable or disable the Vibration alarm on reaching the alarm threshold. To enter the Setup Menu from the DER measurement mode or the Search mode, press and hold down the Mode button for more than 5 seconds. The CAL message will be displayed first. Continue to hold the Mode button down until the screen displays the value of n-coefficient:. Press the Mode button to go to parameters in the following succession: n-coefficient audible on/off vibration on/off temperature inside ( о C). Setting the n-coefficient To set (or change) the n-coefficient value press and hold down the Mode button for more than 5 seconds. The CAL message will be displayed first, then the n-coefficient value. To enter the n- coefficient adjustment menu press the Light button, the figures will start blinking. Then press the Mode button to increase the coefficient value, or the Light button to decrease it: 18

19 Six seconds after the last adjustment of the n-coefficient value the instrument automatically performs calibration to calculate the new alarm threshold based on the new n-coefficient. Enabling/disabling the audible and vibration alarms To adjust it, make sure the instrument is in any of the operation modes, then press the Mode button and hold it for more than 5 seconds. CAL message will be displayed first, then the value of the n- coefficient. Press the Mode button once again to enter the audible alarm adjustment menu: the On (or Off) message and a blinking loudspeaker sign : To change the setting press the Light button, i.e. switch on or off the audible alarm. To enter the vibration alarm adjustment menu press the Mode button once again. The LCD will display the On (or Off) message and a blinking vibration sign. Use the Light button to switch on or off the vibration alarm. Then press the Mode button and LCD will display the temperature inside the instrument. ATTENTION! DON T USE THE BUILT-IN TEMPERATURE FOR ENVIRONMENTAL TEMPERATURE MEASUREMENTS. In the Setup mode if you don t press the Mode button within six seconds the instrument automatically goes to the CAL calibration mode Communication with Computer ATTENTION! THIS IS ADVANCED MODE FOR ADMINISTRATOR USERS WITH PASSWORD-PROTECTED ACCESS. The manufacturer preset password: 1. The instrument has an internal clock which is responsible for keeping track of date and time. If the battery is removed or allowed to go dead, the clock will no longer function. To verify the clock, the instrument can be synchronized with PC using the PM1703MO1 Data Processing Software. The following instructions will walk through setting of the clock: To synchronize the time between the instrument and PC, simply connect the instrument to the PC via IrDA. Start the Polimaster software for the instrument. Read the history of the instrument and the time will automatically be synchronized after reading the history or reading the settings. After that the event history will be saved with time and data stamps referred to local PC time. If it is impossible to sync time after a battery replacement, for example there is no access to the PC, change the battery and resume 19

20 operation without issue. However the internal clock will be lagging behind approximately the duration of time there is no battery power to the instrument. The instrument performs data exchange with a PC with OS WINDOWS via infra red communication at a distance up to 0.2 meters. The PM1703МO-1A has additional communication with PC via USB interface, and the PM1703МO-1B via Bluetooth radio channel. The following functions can be performed in the mode of communication with PC: - downloading information from the instrument s memory: instrument ID, date and time the instrument is switched on and off; - date and time of each self-testing and self-test report; - dose rate values in preset data logging intervals; - date, time and value of dose rate alarms; - date, time and value of dose; - date, time and dose rate value of each background calibration; - enabling/disabling audible and/or vibration alarm; - checking or changing the n-coefficient value (in the range from 1 to 9.9 in 0.1 steps); - checking and automatic correction of current time and date; - setting the logging intervals in which dose rate values are stored in instrument s memory; - enabling/disabling the control buttons to switch on/off audible and vibration alarm; - setting dose rate alarm threshold levels; - setting dose alarm threshold levels; - resetting accumulated dose; - enabling/disabling the control buttons to resetting accumulated dose; - switching the instrument off Communication with computer via infra red and PM1703MO-1A via USB interface For infra red communication use a personal computer with IrDA or infra red adapter, for communication via USB interface use a PC with USB port and USB cable supplied with the instrument and the user software supplied on CD. Minimum requirements to a computer and its software: - Intel Pentium or similar processor Mb RAM - Windows 2000, NT, XP, Vista (small font in Windows settings) - 20 Mb free on HDD and an adequate space for the database - 800x600 resolution of a monitor - IrDA - database compatible with Microsoft Office Access. Install the Software supplied on CD on your personal computer in order to be able to communicate with the instrument. Run a program. For more information about the user software please see the text document installed together with the user software. In communication via USB interface the instruments PM1703MO-1A allow to transfer the acquired scintillation spectra for radionuclide identification to PC. For this purpose it is included in a delivery kit of PM1703MO-1A a special PoliIdentify Software Laptop/Desktop Edition on a separate CD. The detailed information on the software installing and its operation is described in the text document that is installed with the programm. 20

21 To initialize the infra red communication, place the instrument cm away from the infrared adapter (IrDA) and press the Light button. The instrument will enter the data exchange mode the message Ir will appear on the LCD screen: For communication with computer via USB interface connect the instrument with the computer using a USB cable. N o t e - As soon as the instrument s operation history is transferred to computer, the instrument is turned off automatically (the LCD displays OFF ) Communication via Bluetooth (PM1703MO-1B) The instrument can perform Bluetooth wireless communication with a PC with built-in or external compatible Bluetooth and transfer history of operation and parameters stored in its permanent memory, as well as the acquired scintillation spectra for radionuclide identification. The user software for operation of the instrument via Bluetooth and its description are included in the delivery kit: - for PC on CD; - for PPC on memory card. Press the Mode button in the Dose measurement mode to enter the Bluetooth communication on/off menu. The LCD will display a blinking sign and message OFF, i.e. Bluetooth is off. Press the Light button to switch on (off) the Bluetooth. When Bluetooth is on, the LCD displays message on. The instrument starts searching (blinking sign Bluetooth is on) and automatically establishes communication with an available pocket PC. On request PIN N o t e - The PM1703MO-1B with activated Bluetooth may occasionally automatically restart if low battery is used. This is caused by increased power consumption by Bluetooth. The battery should be replaced. It is also recommended to replace the battery if the instrument is expected to be used for a long time with activated Bluetooth, audible and vibration alarms in places where the dose rate exceeds the natural radiation background Alarm Thresholds The instrument can produce warning signals when the following thresholds are exceeded: - Dose rate (two threshold levels); - Dose (one threshold level). When a preset threshold level is reached, the instrument immediately produces a warning signal, notifying the user about a dangerous radiation level. The audible, vibration and visual signals are different for each of the three thresholds. Exceeding the threshold values is also indicated by an analogue scale on the LCD. Use the PM1703MO-1 Data Processing Software to change the instrument threshold values. When several thresholds are exceeded simultaneously, the Dose threshold has the highest priority, then goes the Dose rate threshold. The search threshold has the lowest priority Dose Rate Alarm Thresholds The value range of Dose rate thresholds I and II correspond to the instrument dose rate measurement range (0.1 Sv/h 9.99 Sv/h). 21

22 The factory preset threshold values: First dose rate threshold 50 Sv/h. Alarms - the instrument produces a light (red LED), audible and/or vibration signals and LCD backlight is activated. The signals are single, of the same tone, constant and periodic with the same interval and duration of 1 s. Analogue scale filling the first five segments up to Δ segment represents the preset threshold value. For example: Second dose rate threshold 9.99 Sv/h. Alarms - the instrument produces a light (red LED), audio and/or vibration signals and LCD backlight is activated. The signals are of the same tone, alternating single and double ones with a constant duration and interval of 1 second. Analogue scale - filling the scale up to segment represents the preset threshold value. If the radiation background rises, the scale can be filled to the end. For example: Whenever information about exceeding a threshold value is stored in the instrument memory, the LCD screen indicates it by a blinking h (HISTORY) sign Dose Alarm Threshold The value range of the Dose threshold corresponds to the instrument dose indication range (0,1 Sv 9,99 Sv). The factory preset Dose alarm threshold 9.99 Sv. Alarms the instrument produces a light (red LED), audible and/or vibration signals and LCD backlight is activated. Triple signals of the same tone, constant with the same interval and duration of 1 c. Analogue scale filling the scale to the end. The preset threshold value corresponds to the last segment of the scale. For example: Permanent Memory of the Instrument The following data is automatically stored (and can be changed using the software) in the instrument s permanent memory, protecting the data even when the battery is removed: - instrument ID (serial number), date and time stamp each time the instrument is switched on/off; 22

23 - radiation background history in preset data logging intervals; - dose accumulation history; - date, time and values of each dose rate alarm and dose alarm; - date, time and dose rate value of each background calibration; - date, time each time audible and/or vibration alarm is enabled or disabled - date, time each time new n-coefficient is set - current time and date - dose rate data logging intervals - dose rate value each time alarm thresholds are exceeded - information if audible and vibration alarm can be enabled/disabled by control buttons - dose rate alarm threshold values - dose alarm threshold value. ATTENTION! WHENEVER INFORMATION ABOUT ALARM EVENTS IS STORED IN THE INSTRUMENT MEMORY, THE LCD SCREEN INDICATES IT BY A BLINKING H (HISTORY) SIGN. THIS SIGN DISAPPEARS AFTER THE INFORMATION IS READ FROM THE INSTRUMENT MEMORY USING THE PM1703MO-1 DATA PROCESSING SOFTWARE. 23

24 3 MAINTENANCE Maintenance of the instrument involves preventive treatment, battery replacement (see 2.1.3) and regular check up of the instrument. Routinely examine the instrument body, remove any dust, and decontaminate the instrument in situation of direct contact with radioactivity. Wipe the body with and an ethyl alcohol-saturated tissue to decontaminate the instrument. Do not use the instrument if the body of the instrument or LCD protective glass is damaged. 24

25 4 TROUBLESHOOTING 4.1 Troubleshooting is listed in Table 4. Table 4 Possible faults Possible reasons Troubleshooting The dosimeter won t turn on Symbol LCD is indicated on the For PM1703MO-1A and PM1703MO-1B: - indication of zero values in search and DER measurement modes; - OL indication at natural radiation background; - other faults No battery installed or it is placed incorrectly Battery discharge Microprocessor failure Replace the battery or place it observing the polarity Replace the battery Restart the microprocessor: remove the battery and then place it back, pressing continuously the backlight button. Release the button once the LCD indication appears All the other faults are eliminated by the manufacturer. 25

26 5 CALIBRATION TEST PROCEDURE (REFERENCE) 5.1 General This procedure describes the calibration procedure for the PM1701MO-1, PM1701MO-1A, PM1701MO-1B, PM1701MO-2 instruments. The calibration tests shall be performed by qualified calibration facilities. The instruments shall be calibrated during manufacture, after repair and periodically during operation and storage as required by national regulations. 5.2 For operations and measuring instruments refer to the Table below. Operation Paragraph Reference and auxiliary measuring instruments External examination Check-up Testing of , Dosimeter calibration assembly with a reference metrological source 137 Cs in compliance with ISO characteristics Certified uncertainty of the assembly not more than 5 % at 0.95 confidence probability Barometer. Least division 1 kpa. Measurement range from 60 to 120 kpa Thermometer. Least division 0.1 C. Measurement range from 10 to 30 C Hygrometer. Measurement range from 30 to 90 % , Stopwatch. Measurement range from 1 to 600 s Dosimeter. Intrinsic error 15 %. 5.3 Safety precautions During the calibration procedure observe the existing national standards for handling radioactive materials and other sources of ionizing radiation and radiation safety regulations, as well as on-site accident prevention instructions. 5.4 Test conditions: - ambient temperature (20 5) C; - relative air humidity 60 (+20; -30) %; - atmospheric pressure 101,3 (+5,4; -15,3) kpa; - background gamma radiation: not more than 0.20 Sv/h. 5.5 Preparation for calibration tests Prior to calibration, study this Operating Manual and prepare the instrument according to the Operating Manual. 5.6 Calibration procedure During the external examination check the instrument against the following requirements: - the instrument delivery kit shall comply with the operating manual; - the instrument must be marked with clear inscriptions; - the instrument must be free from pollution and mechanical damages that may affect operation of the instrument. 26

27 5.6.2 During check-up it is necessary to check operation of the instrument as described in the operating manual Testing of metrological performance Check the relative intrinsic error of DER *(10) measurement as follows: 1) switch on the instrument; 2) after completion of self-tests, enter the DER measurement mode; 3) place the instrument on the dosimetric calibration assembly so that the gamma detector faces a gamma radiation source and the normal through the detector geometric center coincides with the radiation flux axis (see Figure 1); 4) calculate the average DER of ambient gamma background with no radiation sources. For this purpose, within 600 seconds after entering the DER measurement mode take five readings of the ambient background at intervals of not less than 150 seconds and calculate the average background dose rate, by formula where 1 5 b н bi 5 Н, (1) i 1 bi the i-th reading of the instrument of the background, µsv/h; 5) create reference DER = 3.0 µsv/h in the point coinciding with geometrical center of the detector and irradiate the instrument; 6) in not less than 300 seconds after the beginning of irradiation take 5 readings of DER measurement at intervals of not less than 60 seconds and calculate the average DER value j by formula H j H ji (2) i 1 where ji the i-th reading of the instrument when measuring DER in check point; 7) repeat measurements for the points where reference DER is equal to 8.0; 80.0; 800 µsv/h; 8) create reference DER = 8.0 msv/h in the point coinciding with the detector geometric center; 9) irradiate the instrument; 10) in not less than 100 seconds after the beginning of irradiation take 5 readings of DER measurement at intervals of not less than 30 seconds and calculate the average DER value j by formula (2); 11) models PM1703MO-1, PM1703MO-1A, PM1703MO-1B: perform additional tests in points where reference DER is equal to 80; 800 msv/h and 8.00 Sv/h; 12) calculate relative error of measurement Q j, %, by formula Q j H j H H b oj Н oj 100 where оj reference DER at check point; j average DER value at check point; b average background DER at check point; 13) calculate confidence limits of relative intrinsic error of DER measurement, %, with 0.95 confidence probability, by formula 2 2 (4) 1,1 (Q o ) (Q jmax ) (3) 27

28 where Q o error (uncertainty) of the reference dosimetric assembly, %; Q jmax maximum relative measurement error Q j, %. 14) compare the confidence limit of relative intrinsic error calculated by formula (4) with the limits of relative intrinsic error acc : - for models PM1703MO-1, PM1703MO-1A, PM1703MO-1B calculated by formula acc. = (20+K 1 / +K 2 ) %, (5) where - DER value, msv/h; K 1 coefficient = msv/h; K 2 coefficient = (msv/h) -1 ; - for model PM1703MO-2: acc. = 30 %. The instrument passes the test if acc Test the relative intrinsic error of DE measurements for models PM1703MO-1, PM1703MO-1A, PM1703MO-1B as follows: 1) set the maximum values of DER threshold and DE threshold on the instrument and enter the DE measurement mode; 2) perform item , item 3); 3) read the initial DE value from the instrument; 4) using a reference gamma source 137 Cs, create DER equal to 0.08 msv/h at the point of geometric center of the detector and irradiate the instrument during the time period T equal to 60 minutes; 5) read the final DE value after completion of irradiation; 6) calculate the relative intrinsic relative error of DE measurement G j by formula where Gj H H oj T Н fj final value of DE, msv; Н ij initial value of DE, msv; oj reference DER value at check point, msv/h; Т irradiation time in hours; fj H ij H oj T, (6) 100 7) repeat measurements 1) - 6) for the points at reference DER equal to 8.0 msv/h and msv/h, at T = 120 minutes; 8) calculate the error confidence limit of error of the instrument under test for each measured point with 0.95 confidence probability, by formula 1.1 (G o ) 2 (G j ) 2, (7) where G o error of the reference dosimetric assembly, %; G j relative error of DE measurement, calculated by formula (6), %. Compare the confidence limit of relative intrinsic error calculated by formula (7) with the limits of relative intrinsic error acc. = 20 %. The instrument passes the test if acc.. 28

29 6 STORAGE AND SHIPPING The instruments must be stored in the manufacturer s package without batteries installed at the air temperature between -15 C to +50 C and humidity less than 95 % at +35 C. The shelf life does not exceed the average service life of the instrument 8 years. Instruments without its original packaging must be stored at 10 C to 35 C and humidity less than 80 % at 25 C. The storage place must be free of dust and/or vapors from strong chemicals and other substances that may cause corrosion. Instruments (switched off, in a package) may be shipped by any closed transport at air temperature from -50 C to +50 C. The packaged instruments must be placed and fastened in a vehicle to exclude shocks against each other and against walls of the vehicle. When carried by sea, instruments in package must be packaged in hermetic plastic bags with silica gel. When carried by air, instruments in package must be placed in hermetic compartments. 29

30 7 WARRANTY The manufacturer warrants this instrument to be free from defects in material and workmanship under normal use and service for a period of 18 months provided that the equipment has been used, transported and stored in a proper manner and not subjected to abuse. The warranty is void if the instrument is repaired or altered by others than those authorized by Polimaster. The warranty period is prolonged for a period of warranty repair. The warranty does not cover the batteries. The battery replacement is not considered as the warranty repair. 30

31 ANNEX A (reference) Typical Energy Response of the Instrument in Search Mode % kev

32 ANNEX B (reference) Typical instrument anisotropy Figure B.1 Rotation in horizontal plane Figure B.2 Rotation in vertical plane 32

PERSONAL COMBINED RADIATION DETECTOR/DOSIMETER РМ1703МО-1 OPERATION MANUAL

PERSONAL COMBINED RADIATION DETECTOR/DOSIMETER РМ1703МО-1 OPERATION MANUAL PERSONAL COMBINED RADIATION DETECTOR/DOSIMETER РМ1703МО-1 OPERATION MANUAL 1 TABLE OF CONTENTS 1 DELIVERY OPTIONS...3 2 BEFORE YOU BEGIN...4 3 PRECAUTIONS...4 4 GENERAL DESCRIPTION...4 4.1 DISPLAY AND

More information

Using the GammaRAE II Responder Personal Radiation Detector and Dosimeter in One. Firmware Version 3.00 D August 2006

Using the GammaRAE II Responder Personal Radiation Detector and Dosimeter in One. Firmware Version 3.00 D August 2006 Using the GammaRAE II Responder Personal Radiation Detector and Dosimeter in One Firmware Version 3.00 D August 2006 GammaRAE II Responder Features Getting Started Replacing the batteries Turn the unit

More information

User s Manual. Electronic Personal Dosimeter (For Gamma(X)-ray) NRF30. Fuji Electric Co., Ltd. TN514629i 1/31

User s Manual. Electronic Personal Dosimeter (For Gamma(X)-ray) NRF30. Fuji Electric Co., Ltd. TN514629i 1/31 User s Manual Electronic Personal Dosimeter (For Gamma(X)-ray) NRF30 Fuji Electric Co., Ltd. TN514629i 1/31 Introduction Thank you for purchasing the Electronic Personal Dosimeter NRF30 (Hereinafter described

More information

MODEL 23 AND 23-1 ELECTRONIC PERSONAL DOSIMETER. August 2017

MODEL 23 AND 23-1 ELECTRONIC PERSONAL DOSIMETER. August 2017 MODEL 23 AND 23-1 ELECTRONIC PERSONAL DOSIMETER August 2017 MODEL 23 AND 23-1 ELECTRONIC PERSONAL DOSIMETER August 2017 ii STATEMENT OF WARRANTY Ludlum Measurements, Inc. warrants the products covered

More information

LUDLUM MODEL 25 AND 25-1 PERSONAL RADIATION MONITOR USER S MANUAL. October 2014 Serial Number and Succeeding Serial Numbers

LUDLUM MODEL 25 AND 25-1 PERSONAL RADIATION MONITOR USER S MANUAL. October 2014 Serial Number and Succeeding Serial Numbers LUDLUM MODEL 25 AND 25-1 PERSONAL RADIATION MONITOR USER S MANUAL October 2014 Serial Number 25000100 and Succeeding Serial Numbers LUDLUM MODEL 25 AND 25-1 PERSONAL RADIATION MONITOR USER S MANUAL October

More information

MetRad1 Handheld Metal/Gamma Detection Unit. Operation Manual. Berkeley Nucleonics Test, Measurement and Nuclear Instrumentation since 1963 BNC

MetRad1 Handheld Metal/Gamma Detection Unit. Operation Manual. Berkeley Nucleonics Test, Measurement and Nuclear Instrumentation since 1963 BNC MetRad1 Handheld Metal/Gamma Detection Unit Operation Manual BNC Berkeley Nucleonics Test, Measurement and Nuclear Instrumentation since 1963 GENERAL Thank you for purchasing the Model MetRad 1 - Handheld

More information

Contamination Meter Users Manual

Contamination Meter Users Manual RDS-80 Contamination Meter U Users Manual Version 1.01 LAURUS Systems, Inc. - Ph: 410-465-5558 - Fax: 410-465-5257 - www.laurussystems.com NOTE This document and the information herewith are copyrighted

More information

PORTABLE ISOTOPE IDENTIFIER Search Tool / Sample Counting System

PORTABLE ISOTOPE IDENTIFIER Search Tool / Sample Counting System FEATURES: HIGH GAMMA SENSITIVITY MULTI-CHANNEL ANALYZER (MCA) WORKS WITH VARIETY OF PROBES RADIONUCLIDE LIBRARY TAILORED TO CUSTOMER REQUEST SIMULTANEOUSLY MEASURES ; DISPLAYS; ALARMS 1-3 SEPARATE DETECTORS

More information

Using the GammaRAE II Personal Radiation Detector

Using the GammaRAE II Personal Radiation Detector Using the GammaRAE II Personal Radiation Detector GammaRAE II Features Normal Usage Turning Unit On Operating Mode User Screens Advanced Features Programming Mode Training Agenda GammaRAE II Features Detector:

More information

EcotestVIP PERSONAL RADIOACTIVITY DETECTOR. Operating manual ВIСТ РЭ

EcotestVIP PERSONAL RADIOACTIVITY DETECTOR. Operating manual ВIСТ РЭ EcotestVIP PERSONAL RADIOACTIVITY DETECTOR Operating manual ВIСТ.412129.020-02 РЭ The enterprise works in accordance with the introduced and certificated Quality Management System in compliance with the

More information

RAD-60 USER GUIDE (ALSO FOR THE RAD-50 MODEL AND ISOTRAK DOSIMETERS)

RAD-60 USER GUIDE (ALSO FOR THE RAD-50 MODEL AND ISOTRAK DOSIMETERS) RAD-60 USER GUIDE (ALSO FOR THE RAD-50 MODEL AND ISOTRAK DOSIMETERS) 2 RAD-60 DOSIMETER USER S GUIDE The RAD-60 is a Digital Pocket Alarm Stand Alone Dosimeter designed for use by people who may be exposed

More information

Portable Oxygen Monitor OX-07. (Type B) <No Gas alarm type> Operating Manual

Portable Oxygen Monitor OX-07. (Type B) <No Gas alarm type> Operating Manual PT0E-0922 Portable Oxygen Monitor OX-07 (Type B) Operating Manual Request for the Customers Read and understand this operating manual before using this gas monitor. Use it in accordance

More information

IMR IX176 Portable Gas Detector User Manual

IMR IX176 Portable Gas Detector User Manual IMR Portable Gas Detector User Manual Read this manual carefully before using this device. (727) 328-2818 / (800) RING-IMR Fax: (727) 328-2826 www.imrusa.com Ver. 1.0A4 CONTENTS SERVICE GUIDELINES... 3

More information

Advanced Radiation Detection Equipment

Advanced Radiation Detection Equipment Advanced Radiation Detection Equipment Electronic Dosimeters Personal Radiation Detectors Spectroscopic Personal Radiation Detectors Portable Monitors & Radioisotope Identifiers Radiation Fixed Monitors

More information

Our Extended Personal Electronic Dosimeter (PED) Family

Our Extended Personal Electronic Dosimeter (PED) Family Our Extended Personal Electronic Dosimeter (PED) Family Tracerco s Award Winning Radiation Monitors and Servicing business unit provides a range of products and services accredited to ISO9001:2008 and

More information

Controls Surface mounted pushbuttons Suitable for gloved use. Buttons Side - On/Off Front Menu Scroll Pin Hole Setup access

Controls Surface mounted pushbuttons Suitable for gloved use. Buttons Side - On/Off Front Menu Scroll Pin Hole Setup access 8.0 Safe-EPD Dosimeter 8.1 Technical data STS Safe-Series Series Field Survey Instruments Instrument Name STS Safe-EPD Dosimeter Description The STS Safe-EPD simulator is a simulated generic Electronic

More information

MODEL 951 RADIATION DETECTOR

MODEL 951 RADIATION DETECTOR P R O D U C T G U I D E MODEL 951 RADIATION DETECTOR I N T R O D U C T I O N About Berkeley Nucleonics Founded in 1963, Berkeley Nucleonics Corporation (BNC), is a leading manufacturer of precision electronic

More information

DMC 3000 User s Guide

DMC 3000 User s Guide Health Physics Division DMC 3000 User s Guide 151153EN-D Table of Contents -2-1- Presentation...3 2- General...4 2.1 Use and Functionality... 4 2.2 Compatibility with Mirion Technologies Products... 5

More information

II 2 G. SENSIT HXG-2d. Ex ib IIB T3 IP20. Suitable for combustible gases such as Methane, Butane, Propane and Natural Gas MADE IN USA

II 2 G. SENSIT HXG-2d. Ex ib IIB T3 IP20. Suitable for combustible gases such as Methane, Butane, Propane and Natural Gas MADE IN USA SENSIT HXG-2d Combustible Gas Detector INSTRUCTION MANUAL Read and understand Instructions before use. PENDING II 2 G Ex ib IIB T3 IP20 PENDING (USA) Class 1, Division 1, Groups C and D, T3 Suitable for

More information

MODEL OX-07 Type A INSTRUCTION MANUAL FOR RIKEN PORTABLE OXYGEN MONITOR PT0E-0860

MODEL OX-07 Type A INSTRUCTION MANUAL FOR RIKEN PORTABLE OXYGEN MONITOR PT0E-0860 INSTRUCTION MANUAL FOR RIKEN PORTABLE OXYGEN MONITOR MODEL OX-07 Type A Safety Precautions Read and understand the instructions in this manual before operating this detector. Keep manual accessible all

More information

IT-09 DATA PANEL. Operating manual ВІСТ НЕ

IT-09 DATA PANEL. Operating manual ВІСТ НЕ IT-09 DATA PANEL Operating manual ВІСТ.468382.019 НЕ CONTENTS 1 DESCRIPTION AND OPERATION...2 1.1 APPLICATION OF THE DATA PANEL...2 1.2 SPECIFICATIONS...3 1.3 DELIVERY KIT OF THE PANEL...5 1.4 DESIGN AND

More information

Ranger EXP RADIATION A L E R T

Ranger EXP RADIATION A L E R T RADIATION A L E R T TM Ranger EXP INTERNATIONAL S.E. International, Inc. P.O. Box 39, 436 Farm Rd. Summertown, TN 38483 USA 1.800.293.5759 1.931.964.3561 Fax: 1.931.964.3564 www.seintl.com radiationinfo@seintl.com

More information

Portable Oxygen Monitor OX-07. (Type A) Operating Manual

Portable Oxygen Monitor OX-07. (Type A) Operating Manual PT0E-0863 Portable Oxygen Monitor OX-07 (Type A) Operating Manual Request for the Customers Read and understand this operating manual before using this gas monitor. Use it in accordance with this operating

More information

Radioactivity dosimeter RD01. User's Manual

Radioactivity dosimeter RD01. User's Manual Radioactivity dosimeter RD01 User's Manual 2011 Contents Introduction 2 Technical specifications 4 How does RD01 work? 5 How to use RD01? 7 Power 10 How to carry out measures correctly? 11 Warranty 12

More information

SENSIT HXG-2d Combustible Gas Detector INSTRUCTION MANUAL Read and understand Instructions before use. PENDING Ex. II3G Ex ic llb T3 IP20

SENSIT HXG-2d Combustible Gas Detector INSTRUCTION MANUAL Read and understand Instructions before use. PENDING Ex. II3G Ex ic llb T3 IP20 SENSIT HXG-2d Combustible Gas Detector INSTRUCTION MANUAL Read and understand Instructions before use. PENDING Ex II3G Ex ic llb T3 IP20 PENDING (USA) Class 1, Zone 2, AEx nl llb T3 PENDING (Canada) Class

More information

HIn3550 Magnetic Field Monitor User's Manual

HIn3550 Magnetic Field Monitor User's Manual HIn3550 Magnetic Field Monitor User's Manual Copyright 1993 by Holaday Industries, Inc. Manual #600053 10/97 $12.50 Revision Record Manual #600053 HIn3550 Magnetic Field Monitor Revision Description Date

More information

DMC 3000 Quick User s Guide

DMC 3000 Quick User s Guide Health Physics Division Radiation. Safety. DMC 3000 Quick User s Guide Welcome to your DMC 3000! In approaching the opportunity to create a more refined Personal Electronic Dosimeter, we wanted to address

More information

GasAlertMicroClip X3 Specifications

GasAlertMicroClip X3 Specifications GasAlertMicroClip X3 Specifications The instrument must satisfy the following: Physical Specifications Size (h x w x d) Physical size shall be no larger than 4.4 x 2.4 x 1.2 in. / 11.3 x 6.0 x 3.2 cm Weight

More information

GEOPHYSICAL ENVIRONMENTAL ANALYSIS EQUIPMENT AND SERVICES

GEOPHYSICAL ENVIRONMENTAL ANALYSIS EQUIPMENT AND SERVICES GENERAL SERVICES ADMINISTRATION Federal Supply Service AUTHORIZED FEDERAL SUPPLY SERVICE SCHEDULE PRICELIST GEOPHYSICAL ENVIRONMENTAL ANALYSIS EQUIPMENT AND SERVICES FSC GROUP: 66, Part II, Section Q FSC

More information

CO2 RESPONDER Portable Carbon Dioxide Detector QUICK REFERENCE GUIDE

CO2 RESPONDER Portable Carbon Dioxide Detector QUICK REFERENCE GUIDE CO2 RESPONDER Portable Carbon Dioxide Detector QUICK REFERENCE GUIDE GasAlertMicro 5 IR from Refer to for more details. Manual provided on CD with unit at time of purchase 2 Table of Contents Getting Started.

More information

IT-09T DATA PANEL. Operating manual ВІСТ НЕ

IT-09T DATA PANEL. Operating manual ВІСТ НЕ IT-09T DATA PANEL Operating manual ВІСТ.468382.019-04 НЕ CONTENTS 1 DESCRIPTION AND OPERATION...2 1.1 APPLICATION OF THE DATA PANEL...2 1.2 SPECIFICATIONS...3 1.3 DELIVERY KIT OF THE PANEL...5 1.4 DESIGN

More information

MINI-RADIAC INTRODUCTION

MINI-RADIAC INTRODUCTION THE MINI-RADIAC IS A PERSONAL DOSIMETER/RADIATION MONITOR. IT IS USED TO DETECT ACCUMULATED DOSE AND/OR RATE OF RADIATION EXPOSURE. IT IS DESIGNED TO ALERT AND INFORM THE WEARER OF THE PRESENCE OF GAMMA

More information

Bacharach Bodyguard 4 User's Guide

Bacharach Bodyguard 4 User's Guide Bacharach Bodyguard 4 User's Guide This manual and the software described in it are copyrighted, with all rights reserved. Under the copyright laws, this manual or the software may not be copied, in whole

More information

Personal Radiation Alarm Dosimeter LK Operating Instruction Manual. Contents

Personal Radiation Alarm Dosimeter LK Operating Instruction Manual. Contents Personal Radiation Alarm Dosimeter LK-3600 Operating Instruction Manual Contents 1. Overview.. 1 2. Features 1 3. Equipment Appearance Instructions. 1 4. Working Principle. 1 5. Main Technical Performances

More information

SENSIT HXG-2d. SENSIT TECHNOLOGIES 851 Transport Drive Valparaiso, IN Phone: (219) Fax: (219)

SENSIT HXG-2d. SENSIT TECHNOLOGIES 851 Transport Drive Valparaiso, IN Phone: (219) Fax: (219) WARRANTY AND REPAIR POLICY Your SENSIT HXG-2d instrument is warranted to be free from defects in materials and workmanship for a period of two years after purchase (excluding sensor, calibration and batteries).

More information

LUDLUM MODEL 26-1 FRISKER USER S MANUAL. April 2017 Serial Number PF and Succeeding Serial Numbers

LUDLUM MODEL 26-1 FRISKER USER S MANUAL. April 2017 Serial Number PF and Succeeding Serial Numbers LUDLUM MODEL 26-1 FRISKER USER S MANUAL April 2017 Serial Number PF002366 and Succeeding Serial Numbers LUDLUM MODEL 26-1 FRISKER USER S MANUAL April 2017 Serial Number PF002366 and Succeeding Serial Numbers

More information

LUDLUM MODEL 3276 SERIES FRISKER/AREA MONITOR USER S MANUAL. November 2017 Serial Number: and Succeeding Firmware: 49830n30 and Higher

LUDLUM MODEL 3276 SERIES FRISKER/AREA MONITOR USER S MANUAL. November 2017 Serial Number: and Succeeding Firmware: 49830n30 and Higher LUDLUM MODEL 3276 SERIES FRISKER/AREA MONITOR USER S MANUAL November 2017 Serial Number: 25009185 and Succeeding Firmware: 49830n30 and Higher LUDLUM MODEL 3276 SERIES FRISKER/AREA MONITOR USER S MANUAL

More information

ALTAIR 5X Multigas Detector Product Specification

ALTAIR 5X Multigas Detector Product Specification ALTAIR 5X Multigas Detector Product Specification PHYSICAL CHARACTERISTICS Gas delivery Size, pumped unit without IR Size, pumped unit with IR Weight Handling Case material Environmental protection Display

More information

PURPOSE AND SPECIFICATIONS

PURPOSE AND SPECIFICATIONS PURPOSE AND SPECIFICATIONS РМ1406 Food Contamination Monitor (hereafter referred to as Instrument ) is intended to measure the radionuclide activity in various samples food stuffs, drinking water, convenience

More information

RIID capability in the palm of your hand

RIID capability in the palm of your hand RIID capability in the palm of your hand Thermo Scientific RadEye SPRD-GN Personal Radiation Detector Changing the game The Thermo Scientific RadEye SPRD-GN delivers unprecedented detection. The first

More information

CA870 INFRARED THERMOMETER E N G L I S H. User Manual. Auto OFF

CA870 INFRARED THERMOMETER E N G L I S H. User Manual. Auto OFF INFRARED THERMOMETER CA870 Auto OFF E N G L I S H User Manual Owner s Record The serial number for the Model CA870 is located inside the battery compartment of the instrument. Please record this number

More information

RAM GENE-1 ERK. Contamination & Radiation Meter. Operating Manual. Document #PRIR90N2.DOC Version 2.2 October 2009

RAM GENE-1 ERK. Contamination & Radiation Meter. Operating Manual. Document #PRIR90N2.DOC Version 2.2 October 2009 RAM GENE-1 ERK MK-Ι & MK-ΙΙ Contamination & Radiation Meter Operating Manual Document #PRIR90N2.DOC Version 2.2 October 2009 Version / Revision Log: RAM GENE-1 ERK/ Operating Manual Ver./Rev.# Date Revised

More information

SENSIT HXG-2d Combustible Gas Detector

SENSIT HXG-2d Combustible Gas Detector INSTRUCTION MANUAL SENSIT HXG-2d Combustible Gas Detector Read and understand instructions before use. 710183 UL 913 Intrinsically Safe for Use in Class I, Groups C and D, T3 Hazardous Locations For use

More information

ADVANCED RADIATION DETECTION EQUIPMENT

ADVANCED RADIATION DETECTION EQUIPMENT ADVANCED RADIATION DETECTION EQUIPMENT Electronic Dosimeters Personal Radiation Detectors Spectroscopic Personal Radiation Detectors Portable Monitors and Radioisotope Identifiers Radiation Fixed and Deployable

More information

Refrigerated Incubator Model and Operating Instructions

Refrigerated Incubator Model and Operating Instructions Refrigerated Incubator Model 165000 and 165000-2 Operating Instructions N2400379 - Rev. 1 08May2018 1 Contents 1. SAFETY...3 1.1. EMF INTERFERENCE...4 1. PRODUCT INFORMATION...5 1.1 INTRODUCTION...5 2.

More information

E N G L I S H FIRE ALARM ASPIRATION SENSING TECHNOLOGY QUICK INSTALLATION GUIDE STAND-ALONE FAAST LT MODELS FL0111E FL0112E FL0122E. 367 mm.

E N G L I S H FIRE ALARM ASPIRATION SENSING TECHNOLOGY QUICK INSTALLATION GUIDE STAND-ALONE FAAST LT MODELS FL0111E FL0112E FL0122E. 367 mm. E N G L I S H FIRE ALARM ASPIRATION SENSING TECHNOLOGY QUICK INSTALLATION GUIDE STAND-ALONE FAAST LT MODELS FL0E FL0E FL0E mm mm 0 mm DESCRIPTION The LT FL0 Series is part of the Fire Alarm Aspiration

More information

64 Max IR Thermometer

64 Max IR Thermometer 64 Max IR Thermometer Users Manual January 2017 2017 Fluke Corporation. All rights reserved. Specifications are subject to change without notice. All product names are trademarks of their respective companies.

More information

Sirius Technologies, Inc.

Sirius Technologies, Inc. Model#: ST501 CO2/Temp/RH Indoor Air Quality Meter Users Manual Page 1 TABLE OF CONTENTS FEATURES & SPECIFICATIONS... 3 WHAT S IN THE BOX... 7 START UP AND GENERAL OPERATION... 8 Recording and Datalogging...

More information

TRACERCO T402 and T402HR USER MANUAL

TRACERCO T402 and T402HR USER MANUAL TRACERCO T402 and T402HR USER MANUAL Quality Assurance Tracerco UK has accreditation for ISO9001 and ISO14001. Confidentiality Notice Information contained in this publication or as otherwise supplied

More information

Dosimeters/Survey meters

Dosimeters/Survey meters Dosimeters/Survey meters 40 Gracechurch Str., EC3V OBT, London, United Kingdom (24 Hour Customer Line) 0044 208 720 6926 RAD-ION (RI-02) Especially designed for the Detection of Pulsed (Golden Engineering)

More information

RadEye B20 RadEye B20-ER Multi-Purpose Survey Meter

RadEye B20 RadEye B20-ER Multi-Purpose Survey Meter Operating Instructions DB-065 E RadEye B20 RadEye B20-ER Multi-Purpose Survey Meter REVISIONS SHEET: Rev. Rev. Dept. Name Rev. Cat. Explanation state resp. page *) 27.07.07 RM&P-E Pij > V 1.26 A 22.07.08

More information

A bright alpha numeric display and the simplified keypad (three buttons) allow the operation even under harsh conditions.

A bright alpha numeric display and the simplified keypad (three buttons) allow the operation even under harsh conditions. Personal Alpha/Beta Continuous Air Monitor (CAM) The pocamon (personal online continuous air monitor) monitors continuously the breathing air to detect airborne radioactive aerosols (LLRD). Typical application

More information

RadEye G20 RadEye G20-ER RadEye G20-10 RadEye G20-ER10 X-Ray Gamma Survey Meter

RadEye G20 RadEye G20-ER RadEye G20-10 RadEye G20-ER10 X-Ray Gamma Survey Meter Operating Instructions DB-067 E RadEye G20 RadEye G20-ER RadEye G20-10 RadEye G20-ER10 X-Ray Gamma Survey Meter REVISIONS SHEET: Rev. Rev. Dept. Name Rev. Cat. Explanation state resp. page *) 10.12.07

More information

TS400. Operating Manual. Test Station for Microtector II Series (G450/G460)

TS400. Operating Manual. Test Station for Microtector II Series (G450/G460) Operating Manual TS400 Test Station for Microtector II Series (G450/G460) GfG GESELLSCHAFT FÜR GERÄTEBAU MBH KLÖNNESTRASSE 99 44143 DORTMUND, Germany TEL. +49 / (0)2 31 / 5 64 00 0 FAX +49 / (0)2 31 /

More information

Design Manual Installation Operation Maintenance

Design Manual Installation Operation Maintenance Design Manual Installation Operation Maintenance Model FT194 UV/IR Portable Flame Detector Test Lamp 23282 Mill Creek Drive, Suite 215 Laguna Hills, CA 92653 USA +1.949.583.1857 Phone +1.949.340.3343 Fax

More information

PRM470 Portable Monitor Operation and Service Manual D00092 June 02, 2016 Rev. 2

PRM470 Portable Monitor Operation and Service Manual D00092 June 02, 2016 Rev. 2 PRM470 P/N D00092 PRM470 Portable Monitor D00092 June 02, 2016 PRM470 P/N D00092 Foreword Thank You for choosing a Rapiscan Detection product. This manual is intended for the installation and normal daily

More information

Electronic Self-Leveling Horizontal Rotary Laser Level Model No Instruction Manual

Electronic Self-Leveling Horizontal Rotary Laser Level Model No Instruction Manual 1467H-English 8/6/08 10:07 AM Page 1 Electronic Self-Leveling Horizontal Rotary Laser Level Model No. 40-6537 Instruction Manual Congratulations on your choice of this Electronic Self-Leveling Horizontal

More information

Operations Manual TS400. Test Station for G450/G460 Gas Detector

Operations Manual TS400. Test Station for G450/G460 Gas Detector TS400 Test Station for G450/G460 Gas Detector Operations Manual 1194 Oak Valley Dr, Ste 20, Ann Arbor MI 48108 USA (800) 959-0329 (734) 769-0573 www.goodforgas.com GfG Products for Increased Safety Congratulations

More information

LD-5000 REFRIGERANT LEAK DETECTOR

LD-5000 REFRIGERANT LEAK DETECTOR LD-5000 REFRIGERANT LEAK DETECTOR www.jbind.com TM PRODUCT OVERVIEW JB Industries is proud to unveil the new, revolutionary LD-5000 PROWLER refrigerant gas leak detector. Using our new proprietary JB Industries

More information

TRACERCO Radiation Monitor

TRACERCO Radiation Monitor TRACERCO Radiation Monitor Instruction and Maintenance Manual for Model No. T402 Customer Notice: Please note that the T402 is NOT an Intrinsically Safe Monitor and should not be confused with the TRACERCO

More information

Interscan Corporation. Instruction Manual Series Compact Portable Analyzer

Interscan Corporation. Instruction Manual Series Compact Portable Analyzer Interscan Corporation Instruction Manual 4000 Series Compact Portable Analyzer 1 2 INTERSCAN CORPORATION Simplified 4000 SERIES MANUAL TABLE OF CONTENTS SECTION TITLE PAGE 1.0 Equipment Description 4 2.0

More information

1, 2, 3, and 4-Gas Detector. Technical Reference Guide

1, 2, 3, and 4-Gas Detector. Technical Reference Guide 1, 2, 3, and 4-Gas Detector Technical Reference Guide Limited Warranty and Limitation Liability BW Technologies LP (BW) warrants the product to be free from defects in material and workmanship under normal

More information

THL2. Temperature/Humidity USB Datalogger INSTRUCTION MANUAL

THL2. Temperature/Humidity USB Datalogger INSTRUCTION MANUAL The THL2 is compatible with computers using Windows 2000, XP, Vista, Windows 7 and Windows 8. INSTRUCTION MANUAL 2 THL2 1-800-547-5740 Fax: (503) 643-6322 www.ueitest.com email: info@ueitest.com Temperature/Humidity

More information

RAM GENE / RAM GENE-1

RAM GENE / RAM GENE-1 RAM GENE / RAM GENE-1 Radiation Meters Operating Manual Document #PRIR69N0.DOC Version 6.1 - December 2000 ROTEM INDUSTRIES LTD. ROTEM Industrial Park P.O.Box 9046, Beer-Sheva 84190, ISRAEL Tel: 972-8-6571312/6568652

More information

Model R9400. Instruction Manual. CO Meter with Temperature. wwwreedinstruments. com

Model R9400. Instruction Manual. CO Meter with Temperature. wwwreedinstruments. com Model R9400 CO Meter with Temperature Instruction Manual com Table of Contents Safety... 3 Features... 3 Specifications... 4 Instrument Description... 5 Operating Instructions...6-7 Selecting Temperature

More information

EL-USB-2-LCD+ High Accuracy Humidity, Temperature and Dew Point Data Logger with LCD

EL-USB-2-LCD+ High Accuracy Humidity, Temperature and Dew Point Data Logger with LCD High Accuracy Humidity, Temperature and Dew Point Data Logger with LCD ORDERING INFORMATION Standard Data Logger (Data Logger, Software on CD, Battery) Replacement Battery 99 Washington Street Melrose,

More information

RAE Systems Systems. User Presentation. October Protection Through Detection

RAE Systems Systems. User Presentation. October Protection Through Detection RAE Systems Systems AreaRAE Inert User Presentation October 2009 www.raesystems.com AGENDA Overview & Features Basic Operation Turning the AreaRAE Inert monitor ON Warnings NORMAL & INERT Mode Displays

More information

MGC Dock User s Manual

MGC Dock User s Manual User s Manual Contents Warnings Statements/Avertisseement... 3 READ FIRST BEFORE OPERATION... 3 Description... 4 Basic Operation... 5 Clip Dock Components... 5 LEDs... 5 User Operation... 6 Turning the

More information

i.c³ User Guide For Helmer i.series Ultra-Low Freezers A/A

i.c³ User Guide For Helmer i.series Ultra-Low Freezers A/A i.c³ User Guide For Helmer i.series Ultra-Low Freezers 360175-A/A Document History Revision Date CO Supersession Revision Description A 18 APR 2014* 9275 n/a Initial release. * Date submitted or change

More information

MGC Dock User s Manual

MGC Dock User s Manual Operator s Manual Contents Warnings Statements/Avertisseement... 3 READ FIRST BEFORE OPERATION... 3 Basic Operation... 4 Clip Dock Components... 4 LEDs... 4 Operation... 5 Turning the Clip Dock On and

More information

Model: LD5 Dual Mode Refrigerant Gas Leak Detector. Operating Instructions

Model: LD5 Dual Mode Refrigerant Gas Leak Detector. Operating Instructions 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.testequipmentdepot.com Operating Instructions Model: LD5 Dual Mode Refrigerant Gas Leak Detector PAGE

More information

Operations Manual TS400. Test Station for G450/G460 Gas Detector

Operations Manual TS400. Test Station for G450/G460 Gas Detector TS400 Test Station for G450/G460 Gas Detector Operations Manual 1194 Oak Valley Dr, Ste 20, Ann Arbor MI 48108 USA (800) 959-0329 (734) 769-0573 www.gfg-inc.com GfG Products for Increased Safety Congratulations

More information

Using the QRAE Plus personal multigas monitor

Using the QRAE Plus personal multigas monitor Using the QRAE Plus personal multigas monitor Firmware v 1.10 QRAE Plus The Hazardous Environment Detection Company QRAE Plus features Turning on the QRAE Plus Recommended Daily Start-up Procedure User

More information

RC4000 Radiation Detection System

RC4000 Radiation Detection System RC4000 Radiation Detection System LAURUS Systems Inc. 3460 Ellicott Center Drive Suite 101 Ellicott City, MD 21043 Tel. +1 (410) 465-5558 Fax. +1 (410) 465-5257 Web Site: www.laurussystems.com Email: radcomm.sales@laurussystems.com

More information

User Manual. Non-contact Infrared Thermometer.

User Manual. Non-contact Infrared Thermometer. User Manual Non-contact Infrared Thermometer www.paryvara.com info@paryvara.com Introduction Thank you for purchasing this Non-contact Infrared Thermometer. Please read the User Manual carefully to ensure

More information

A FIRST RESPONDERS GUIDE TO PURCHASING RADIATION PAGERS

A FIRST RESPONDERS GUIDE TO PURCHASING RADIATION PAGERS EML-624 A FIRST RESPONDERS GUIDE TO PURCHASING RADIATION PAGERS FOR HOMELAND SECURITY PURPOSES Paul Bailey Environmental Measurements Laboratory U.S. Department of Homeland Security 201 Varick Street,

More information

minidose A compact and colorful personal electronic dosimeter, your family can use everyday.

minidose A compact and colorful personal electronic dosimeter, your family can use everyday. minidose A compact and colorful personal electronic dosimeter, your family can use everyday. The minidose is an easy-to-use electronic personal dosimeter that records X-ray and/or Gamm radiation dose accumulated

More information

Operate the Mutli-Gas Monitor PID

Operate the Mutli-Gas Monitor PID TASK: Operate the Multi-Gas Monitor Learning Objective CONDITION: In a classroom environment, given a fully charged Multi-gas Monitor with the manufacturer operator s manual, all monitor components/accessories,

More information

WS-9117U-IT Wireless 915 MHz Temperature Station. Instruction Manual

WS-9117U-IT Wireless 915 MHz Temperature Station. Instruction Manual WS-9117U-IT Wireless 915 MHz Temperature Station Instruction Manual 1 TABLE OF CONTENTS Topic Inventory of Contents Quick Setup Detailed Setup Guide Battery Installation Setting the Time Features Minimum

More information

Instruction Sheet THERMOCOUPLE DATA LOGGER WITH LCD DISPLAY AND USB INTERFACE

Instruction Sheet THERMOCOUPLE DATA LOGGER WITH LCD DISPLAY AND USB INTERFACE Instruction Sheet OM-EL-USB-TC-LCD THERMOCOUPLE DATA LOGGER WITH LCD DISPLAY AND USB INTERFACE Thermocouple Data Logger with LCD ORDERING INFORMATION Standard Data Logger OM-EL-USB-TC-LCD (Data Logger,

More information

SENSIT HXG-2. Instruction Manual For use with combustible gases Gas Leak Detector

SENSIT HXG-2. Instruction Manual For use with combustible gases Gas Leak Detector SENSIT HXG-2 Instruction Manual For use with combustible gases Gas Leak Detector CLASSIFIED 7WA6 Approved UL913, For Class 1, Division 1, Groups C & D hazardous locations when used with alkaline batteries.

More information

Model 850 Investigator s Aid Accelerant and Gas Detector Operation and Maintenance

Model 850 Investigator s Aid Accelerant and Gas Detector Operation and Maintenance Model 850 Investigator s Aid Accelerant and Gas Detector Operation and Maintenance www.graceindustries.com M850TM1215 December 2015 1 1. Cover 2. Table of Contents 3. Top View 4. Features & Benefits 5.

More information

LUDLUM MODEL 26 FRISKER USER S MANUAL. December 2016 Serial Number PF and Succeeding Serial Numbers

LUDLUM MODEL 26 FRISKER USER S MANUAL. December 2016 Serial Number PF and Succeeding Serial Numbers LUDLUM MODEL 26 FRISKER USER S MANUAL December 2016 Serial Number PF001075 and Succeeding Serial Numbers LUDLUM MODEL 26 FRISKER USER S MANUAL December 2016 Serial Number PF001075 and Succeeding Serial

More information

Patriot Systems Limited

Patriot Systems Limited COPYRIGHT 1997 - The Patriot Systems Ltd. Patriot Alarm Monitoring Automation Package is licensed for use on one computer, by the original person, or company, or organisation whose name is registered with

More information

Sintrol Snifter A2 EX. Manual. Version 1.2.3

Sintrol Snifter A2 EX. Manual. Version 1.2.3 Sintrol Snifter A2 EX Manual Version 1.2.3 Table of Contents 1 INTRODUCTION... 3 1.1 Safety... 3 1.2 Product overview... 4 1.3 How does it work?... 4 2 INSTALLATION... 4 2.1 Selecting the installation

More information

Microprocessor Digital Thermometer

Microprocessor Digital Thermometer Microprocessor Digital Thermometer Instruction Manual Models 819A, 820A and 821A This owner s manual was as current as possible when this product was manufactured. However, products are constantly being

More information

Snifter ATEX22 VERSION. User Manual. Distributor

Snifter ATEX22 VERSION. User Manual. Distributor Snifter ATEX22 VERSION User Manual Distributor Version 1.4 09/09/2009 Table of Contents 1. INTRODUCTION... 3 1.1. Safety... 3 1.2. Product overview... 4 1.3. How does it work?... 4 2. INSTALLATION... 5

More information

Portable Alpha/Beta Continuous Online Air Monitor (CAM)

Portable Alpha/Beta Continuous Online Air Monitor (CAM) Aer 5000 Desktop Portable Alpha/Beta Continuous Online Air Monitor (CAM) The Aer5200 monitors the ambient air continuously to detect airborne radioactive aerosols (LLRD). Typical application fields are

More information

RGR150 USER S MANUAL. Wireless Rain Gauge with Thermometer and Clock

RGR150 USER S MANUAL. Wireless Rain Gauge with Thermometer and Clock RGR150 manual-final-091908:layout 1 9/19/08 8:59 AM Page 1 RGR150 USER S MANUAL Wireless Rain Gauge with Thermometer and Clock INTRODUCTION Thank you for selecting this Wireless Rain Gauge. This device

More information

E1000 Portable gas detector

E1000 Portable gas detector E1000 Portable gas detector Operation manual Ver:HWWM161229CG Read this manual carefully before using the device Table of Contents SAFETY INFORMATION... - 1-1. BRIEF INTRODUCTION... - 3-2. MAIN FUNCTION

More information

Dryer Controller M720

Dryer Controller M720 User Manual Dryer Controller M720 Hardware version 2.00 Software version 2.00 Manual M720 Dryer controller Page 1 of 60 Document history Preliminary version: - Created in April, 2009 Hardware Version 2.00,

More information

LUDLUM MODEL 35 SERIES DIGITAL SURVEY METER USER S MANUAL

LUDLUM MODEL 35 SERIES DIGITAL SURVEY METER USER S MANUAL LUDLUM MODEL 35 SERIES DIGITAL SURVEY METER USER S MANUAL September 2016 LUDLUM MODEL 35 SERIES DIGITAL SURVEY METER USER S MANUAL September 2016 STATEMENT OF WARRANTY Ludlum Measurements, Inc. warrants

More information

6-1 Installing Batteries 6-2 Automatic Circuit/Reset Feature 6-3 Feature Sensitivity Adjustment

6-1 Installing Batteries 6-2 Automatic Circuit/Reset Feature 6-3 Feature Sensitivity Adjustment CONTENT Page 1. GENERAL INFORMATION ------------------ 2 2. FEATURES ---------------------------------------- 2 3. SPECIFICATION --------------------------------- 3 4. OPERATION GUIDE ---------------------------

More information

TEK-Mate. Refrigerant Leak Detector

TEK-Mate. Refrigerant Leak Detector O P E R A T I N G M A N U A L TEK-Mate Refrigerant Leak Detector Declaration Of Conformity This is to certify that this equipment, designed and manufactured by INFICON Inc., 2 Technology Place, East Syracuse,

More information

WIRELESS 868 MHz TEMPERATURE STATION Instruction Manual

WIRELESS 868 MHz TEMPERATURE STATION Instruction Manual WIRELESS 868 MHz TEMPERATURE STATION Instruction Manual INTRODUCTION: Congratulations on purchasing this temperature station with wireless 868 MHz transmission of outdoor temperature and display of indoor

More information

Interactive Fire Control Panel IFS7002 one signal loop Instruction Manual

Interactive Fire Control Panel IFS7002 one signal loop Instruction Manual Interactive Fire Control Panel IFS7002 one signal loop Instruction Manual Revision 4/01.17 Contents 1. Introduction... 6 2. Terminology... 6 3. Function... 8 4. Technical data... 8 4.1. Physical configuration...

More information

Sensor Control Panel

Sensor Control Panel USER S MANUAL PU SENS 01 Sensor Control Panel V55-6EN-03(SENS).indd 1 18.08.2015 10:37:16 2 PU SENS 01 CONTENTS Safety Requirements 2 Main Technical Data 3 Control Panel Mounting 3 Control Panel Operation

More information

requirements. Serial: USB or RSXXX Bluetooth GPS/GSM Ethernet Standard or rugged desktop or laptop PC or tablet; Android smartphone or tablet

requirements. Serial: USB or RSXXX Bluetooth GPS/GSM Ethernet Standard or rugged desktop or laptop PC or tablet; Android smartphone or tablet Photo credit goodluz - contrastwerkstatt - piai - liravega - FOTOLIA.COM - IMS 2012 IMS and its Products IMS Innovation & Measurement Systems is a French designer and manufacturer of electronic instruments

More information

INSTRUCTION MANUAL MODEL GX-82A

INSTRUCTION MANUAL MODEL GX-82A INSTRUCTION MANUAL MODEL GX-82A PORTABLE THREE GAS DETECTOR RKI Instruments, Inc. 1855 Whipple Road Hayward, CA 94544 Tel: 510-441-5656 Table of Contents INTRODUCTION 1 DESCRIPTION 2 COMPONENTS AND CONTROLS

More information

INSTRUCTIONS FOR. Wireless Refrigerator Freezer Thermometer (#10378)

INSTRUCTIONS FOR. Wireless Refrigerator Freezer Thermometer (#10378) CONTENTS Unpacking Instructions... 2 Package Contents... 2 Product Registration... 2 Features & Benefits: Sensors... 2 Features & Benefits: Display... 3 Setup... 4 Install or Replace Batteries... 4 Temperature

More information