1 Operating Manual ALTAIR QuickCheck Station ALTAIR 4 QuickCheck Station Order No /01
2 MSA AUER GmbH Thiemannstrasse 1 D Berlin Germany MSA AUER GmbH. All rights reserved
3 MSA Declaration of Conformity Declaration of Conformity Manufactured by: Mine Safety Appliances Company 1000 Cranberry Woods Drive Cranberry Township, PA USA European authorized representative: MSA AUER GmbH Thiemannstrasse 1 D Berlin We declare that the ALTAIR QuickCheck Station and ALTAIR 4 QuickCheck Station is in conformance with the European EMC Directives 2004/108/EC and in accordance with European standards EN :2007 and EN :2007 MSA AUER GmbH Dr. Axel Schubert R&D Instruments Berlin, June 2009 GB ALTAIR QuickCheck Station/ALTAIR 4 QuickCheck Station 3
4 Contents MSA Contents 1 Safety Regulations Correct Use Liability Information Safety and Precautionary Measures to be Adopted Description System overview Instrument Setup Operation Automatic Version Manual Version for ALTAIR QuickCheck Station Test Failure Gas Tolerance Levels Maintenance Troubleshooting Instructions for Troubleshooting Technical Specifications Ordering Information ALTAIR QuickCheck Station Versions ALTAIR 4 QuickCheck Station Versions Replacement parts ALTAIR QuickCheck Test Gas Cylinders ALTAIR 4 Quick Check Test Gas Cylinders ALTAIR QuickCheck Station/ALTAIR 4 QuickCheck Station GB
5 MSA Safety Regulations 1 Safety Regulations 1.1 Correct Use The QuickCheck Stations, hereafter referred to as test units, verify instrument responses to gas [bump check] and alarm functionality, with the ALTAIR QuickCheck Station testing ALTAIR and ALTAIR PRO gas detectors and the ALTAIR 4 QuickCheck Station testing ALTAIR 4 gas detectors. It is imperative that this operating manual be read and observed when using the product. In particular, the safety instructions, as well as the information for the use and operation of the product, must be carefully read and observed. Furthermore, the national regulations applicable in the user's country must be taken into account for a safe use. Danger! This product is supporting life and health. Inappropriate use, maintenance or servicing may affect the function of the device and thereby seriously compromise the user's life. Before use the product operability must be verified. The product must not be used if the function test is unsuccessful, it is damaged, a competent servicing/maintenance has not been made, genuine MSA spare parts have not been used. Alternative use, or use outside this specification will be considered as non-compliance. This also applies especially to unauthorised alterations to the product and to commissioning work that has not been carried out by MSA or authorised persons. 1.2 Liability Information MSA accepts no liability in cases where the product has been used inappropriately or not as intended. The selection and use of the product are the exclusive responsibility of the individual operator. Product liability claims, warranties also as guarantees made by MSA with respect to the product are voided, if it is not used, serviced or maintained in accordance with the instructions in this manual. GB ALTAIR QuickCheck Station/ALTAIR 4 QuickCheck Station 5
6 Safety Regulations MSA 1.3 Safety and Precautionary Measures to be Adopted Attention! The following safety instructions must be observed implicitly. Only in this way can the safety and health of the individual operators, and the correct functioning of the instrument, be guaranteed. - Please ensure the test unit in use corresponds to the gas type of instrument to be tested. All test units are only to be used with the specific gas for which they are labelled. - Use the test unit only in non-hazardous environments. - Repair or alteration of the test unit, beyond the procedures described in this manual or by anyone other than a person authorized by MSA, could cause the instrument to fail to perform properly. - Use only genuine MSA replacement parts when performing any maintenance procedures described in this manual. Substitution of components can seriously impair instrument performance, alter intrinsic safety characteristics or void agency approvals. 6 ALTAIR QuickCheck Station/ALTAIR 4 QuickCheck Station GB
7 MSA Description 2 Description 2.1 System overview a Fig. 1 Instrument view 2 3 4b 4c 1 Test unit housing 4c Fail LED 2 Alarm test LED 5 ALTAIR instrument *) 3 Gas test LED 6 Gas cylinder *) 4a Pass/Fail and confidence pulse LED 7 Regulator *) 4b Pass and confidence Pulse LED *) These parts may be sold seperately. GB ALTAIR QuickCheck Station/ALTAIR 4 QuickCheck Station 7
8 Description MSA The Test Unit - uses a microphone to detect the audible alarm. - electronically checks the instrument s visual LED alarm. - electronically checks the instrument's standard vibrating alarm for ALTAIR and ALTAIR PRO. The ALTAIR 4 QuickCheck does not detect vibrating alarm funcionality. - tests the instrument s response to a known concentration of gas. The audible alarm is not checked on ALTAIR instruments with software version 1.0. The ALTAIR, ALTAIR PRO and ALTAIR 4 instruments record the date, time, and result of a bump test internally in their event recorder. 2.2 Instrument Setup (1) Place the test unit on a level surface in an atmosphere with an ambient noise level of 60 db or less. Test unit power requirement is 6 VDC from a VAC Hz supply to the power cord. (2) Connect 6 VDC power cord to the receptor at the back of the unit. (3) To connect tubing, attach one end of tubing to the regulator barb fitting and the other end to the test unit barb fitting. (4) Connect regulator to the necessary gas cylinder [ chapter 6.4 and 6.5]. For ALTAIR 4 QuickCheck Station, place the plastic cap over the unused barb fitting on the opposite side of test unit. Attention! Only the specified gas concentrations may be used with the test unit. Attention! The vehicle power supply for the test unit must only be used when the vehicle is stationary. If it is intended to be used when the vehicle is moving both the cylinder and test unit must be fixed to the interior of the vehicle. 8 ALTAIR QuickCheck Station/ALTAIR 4 QuickCheck Station GB
9 MSA Description The ALTAIR 4 QuickCheck can be: - used with an ALTAIR 4 QuickCheck Cylinder Holder with the supplied adapter and tubing - mounted on either side of the ALTAIR 4 QuickCheck Cylinder Holder. When using the cylinder holder: - mount the cylinder holder adapter to the desired side of the cylinder holder with the provided screws. - attach one end of of provided tubing to the cylinder holder gas fitting and the other to the gas fitting on the test unit. Ensure that the tubing is fully seated over the barbs to prevent gas from leaking. - connect the test unit to the remaining side of the cylinder holder with the provided screws. - up to two ALTAIR 4 QuickCheck test units can be used with an ALTAIR 4 QuickCheck Cylinder Holder. Warning! Use of more than two test units per regulator may cause erroneous results. Attention! Before operation, be sure to cover unused gas fittings located on each side of the test unit and cylinder holder with the supplied fitting cover. Fig. 2 ALTAIR 4 QuickCheck Station Setup GB ALTAIR QuickCheck Station/ALTAIR 4 QuickCheck Station 9
10 Operation MSA 3 Operation 3.1 Automatic Version (1) Verify tubing is connected and power is supplied. The Pass/Fail LED or Pass LED flashes green in 30 s intervals, serving as a confidence pulse to indicate that unit is powered. (2) Verify that automatic regulator is ON. (3) Press ALTAIR or ALTAIR PRO TEST button for one second. On the ALTAIR 4 instrument no button needs to be pressed. (4) Insert instrument into the test unit. The following test sequence occurs: Alarm Test - checks for functionality of audible and visual alarm - checks for functionality of vibrating alarm for ALTAIR and ALTAIR PRO - corresponds with "Alarm Test" LED - LED flashes amber during test - green LED indicates passing - red LED indicates failure - Alarms disabled by the user will not be tested by the test unit; the test unit will view these as passing, but ALTAIR 4 QuickCheck Station displays a solid amber LED Gas Test - verifies that sensor responds to gas - corresponds with "Gas Test" LED - LED flashes amber during test - green LED indicates passing - red LED indicates failure. 10 ALTAIR QuickCheck Station/ALTAIR 4 QuickCheck Station GB
11 MSA Operation Overall Pass/Fail Test - corresponds with "Pass/Fail" LED or "Pass" or "Fail" LED - green LED indicates passing - red LED indicates failure - checkmark symbol appears on instrument display when it is successfully bump checked - Failure of either of the two previous tests results in an overall failure. If alarms are turned OFF, but all other functions are working correctly, the overall test passes. (5) After a Pass/Fail indication is given, remove instrument from the test unit. The ALTAIR QuickCheck LEDs automatically turn OFF if the instrument is left in the cradle for more than 30 seconds after a passed test. The ALTAIR 4 QuickCheck LEDs do not turn OFF until the instrument is removed from the QuickCheck Station. 3.2 Manual Version for ALTAIR QuickCheck Station (1) Verify tubing is connected and power is supplied. The Pass/Fail LED or Pass LED flashes green in 30 s intervals, serving as a confidence pulse to indicate that unit is powered. (2) Press the instrument TEST button for one second. (3) Insert the instrument into the test unit. (4) Turn ON regulator. The following test sequence occurs: Alarm Test - checks for functionality of the three alarms [audible, visual and vibrating] - corresponds with "Alarm Test" LED - LED flashes amber during test - green LED indicates passing - red LED indicates failure - Alarms disabled by the user will not be tested by the test unit; the test unit will view these as passing. GB ALTAIR QuickCheck Station/ALTAIR 4 QuickCheck Station 11
12 Operation MSA Gas Test - verifies that sensor responds to gas - corresponds with "Gas Test" LED - LED flashes amber during test - green LED indicates passing - red LED indicates failure. Overall Pass/Fail Test - corresponds with "Pass/Fail" LED - green LED indicates passing - red LED indicates failure - checkmark symbol appears on instrument display when it is successfully bump checked - Failure of either of the two previous tests results in an overall failure. (5) After a Pass/Fail indication is given, remove instrument from the test unit. (6) Turn OFF the regulator. Attention! Failure to turn OFF the regulator will empty the gas cylinder. The ALTAIR QuickCheck LEDs automatically turn OFF if the instrument is left in the cradle for more than 30 seconds after a passed test. 3.3 Test Failure If the detector fails the gas test, recalibrate instrument and perform another bump test. If the detector fails the Alarm Test or continues to fail the bump test see chapter 4.1. If an instrument is removed from the test unit before test completion, the: - instrument is not properly bump tested - test fails - checkmark symbol does not appear on instrument display. A failed or aborted test removes the checkmark symbol from the instrument display. 12 ALTAIR QuickCheck Station/ALTAIR 4 QuickCheck Station GB
13 MSA Operation 3.4 Gas Tolerance Levels Testing time may vary based on: - instrument software version - sensor condition - gas tolerance level The gas tolerance level can be found on the label located on the bottom of the ALTAIR 4 QuickCheck Station. During a test, the gas tolerance level indicates where the read gas levels must be to pass the gas test. For example If a 30% gas tolerance level QuickCheck is used and the calibration cylinder contains 60 ppm CO, the instrument sensor must read above 42 ppm CO for the gas test to pass. [60 - (60*0.30) = 42]. Oxygen is the only exception to this calculation. Oxygen passes a test if the reading: - for the 20% gas tolerance unit is less than or equal to 17% O 2 - for the 30% gas tolerance unit is less than or equal to 18% O 2 - for the 40% gas tolerance unit is less than or equal to 19% O 2 - for the 50% gas tolerance unit is less than or equal to 19.5% O 2. Attention! For ALTAIR 4, a value of 15% vol must be used as a calibration concentration for O 2. The gas tolerance level indicates the accuracy of the instrument reading, verified by the bump check [based on a +0% accurate gas cylinder] as follows: LEL, CO, and H 2 S Units Gas tolerance level Accuracy 20% 80% 30% 70% 40% 60% 50% 50% GB ALTAIR QuickCheck Station/ALTAIR 4 QuickCheck Station 13
14 Maintenance MSA O 2 Unit Gas tolerance level Accuracy 20% 87% 30% 80% 40% 73% 50% 70% As a reference, a full calibration sets the instrument response to the accuracy of the calibration cylinder. 4 Maintenance Warning! Repair or alteration of the test unit, beyond the procedures described in this manual or by anyone other than a person authorized by MSA, could cause the unit to fail to perform properly. Use only genuine MSA replacement parts when performing any maintenance procedures described in this manual. Substitution of components can seriously impair performance of the unit, alter intrinsic safety characteristics or void agency approvals. Failure to follow this warning can result in serious personal injury or death. Clean with a damp cloth only when unit is unplugged. Allow to dry before providing power to the unit. 4.1 Troubleshooting The test unit operates reliably when handled properly. If the unit becomes inoperative, refer to the Troubleshooting Guidelines, which indicate the most likely causes of a problem. Inoperative units that are under warranty may be returned to MSA. 14 ALTAIR QuickCheck Station/ALTAIR 4 QuickCheck Station GB
15 MSA Maintenance 4.2 Instructions for Troubleshooting Problem Description Reaction Alarm Test Failure Alarm chamber plugged with dirt or debris. Check for contamination in alarm chamber and on instrument; clean as needed. Alarm non-functional Press instrument TEST button; if no sound occurs during alarm check, contact MSA. Instrument LEDs activated, but LEDs do not light during alarm test LEDs non-functional Vibrating alarm is activated on ALTAIR or tional Vibrator Alarm non-func- ALTAIR PRO instrument, but it does not vibrate during Alarm test Gas Test Failure No gas in calibration cylinder Instrument requires calibration Calibration cylinder tubing improperly installed Station gas delivery outlet plugged with dirt. Instrument inlet is blocked Wrong calibration cylinder Wrong calibration gas concentration Regulator closed Press instrument TEST button; if LEDs are non-functional, contact MSA. Press instrument TEST button; if vibrator alarm is non-functional, contact MSA. Check pressure gauge on calibration cylinder. Calibrate instrument. Verify tubing is fully connected to station fitting. Clean gas delivery outlet. Check for blockage and remove if necessary. Verify that calibration cylinder is the same gas type as the instrument. Check manual and calibration cylinder for proper concentration. Ensure regulator is opened or turned ON before placing the instrument into the station. GB ALTAIR QuickCheck Station/ALTAIR 4 QuickCheck Station 15
16 Maintenance MSA Problem Description Reaction Station not responding to instrument Expired Cylinder Incorrect regulator The unit is not powered ALTAIR or ALTAIR PRO TEST button pressed for less than one second prior to instrument placement in station Instrument not fully seated No checkmark appears on instrument af- to test completion Instrument removed prior ter a bump test Failed test Check expiration date of calibration cylinder; replace if necessary. Verify that station is used with the recommended regulator. Check power connections to unit Remove instrument; press and hold TEST button for one full second before placing instrument into the unit Fully seat instrument in station Re-test instrument Check instrument for operation 16 ALTAIR QuickCheck Station/ALTAIR 4 QuickCheck Station GB
17 MSA Technical Specifications 5 Technical Specifications For use with ALTAIR Single Gas Detector All versions Tests performed ALTAIR PRO Single Gas Detector ALTAIR 4 Gas Detector O 2, CO, CO Steel, CO Fire, H 2 S, HCN, Cl 2 and ClO 2, SO 2, NO 2 and NH 3 All versions Visual alarm activation Vibrating alarm activation with ALTAIR QuickCheck Station only Audible alarm activation and verification Sensor response to gas General pass fail indication Visual indicators LED indicators For alarm test, gas test and general pass/fail or pass and fail Flashing amber Test in progress Solid red Failed test Solid green Passed test Solid amber Test not perfomed Checkmark Displays on instrument for 24 hours following a passed test ALTAIR Quick- Check Station ALTAIR Quick- Check Station ALTAIR 4 Quick- Check Station Manual Automatic Features a 0.25 l/min manual regulator that user turns ON during testing Features a 0.25 l/min automatic regulator that pulls gas as needed Testing time Average Less than 30 seconds for ALTAIR QuickCheck Less than 1 minute for ALTAIR 4 QuickCheck GB ALTAIR QuickCheck Station/ALTAIR 4 QuickCheck Station 17
18 Technical Specifications MSA Dimensions [L x B x H] Weight Power supply Temperature range Ingress Protection Warranty ALTAIR QuickCheck ALTAIR 4 QuickCheck ALTAIR QuickCheck ALTAIR 4 QuickCheck ALTAIR QuickCheck IP 30 ALTAIR QuickCheck ALTAIR 4 QuickCheck 16.5 cm x 12.1 cm x 5.1 cm 22.2 cm x 16.6 cm x 9.2 cm 245 g Less than 700 g 115/230 VAC, 50/60 Hz 6 VDC [Only use MSA power supply chapter 6] 0 C to +40 C 2 years 18 ALTAIR QuickCheck Station/ALTAIR 4 QuickCheck Station GB
19 MSA Ordering Information 6 Ordering Information 6.1 ALTAIR QuickCheck Station Versions Gas type Regulator Regulator included Part No. O 2, CO, H 2 S, SO 2, NO 2 Manual *) No O 2, CO, H 2 S, SO 2, NO 2 Automatic Yes Cl 2, ClO 2 Manual *) No Cl 2, ClO 2 Automatic Yes NH 3 Manual *) No NH 3 Automatic Yes HCN Manual *) No HCN Automatic Yes *) For manual versions of the ALTAIR QuickCheck Station, a manual regulator and regulator tubing must be purchased separately. 6.2 ALTAIR 4 QuickCheck Station Versions Gas type 6.3 Replacement parts Regulator Gas SPAN Regulator included Part No. Ex, O 2, CO, H 2 S Automatic 50% No Ex, O 2, CO, H 2 S Automatic 40% No Ex, O 2, CO, H 2 S Automatic 30% No Ex, O 2, CO, H 2 S Automatic 20% No Description Part No. European Power Supply Vehicle Power Supply Regulator Tubing Automatic-Gas Regulator Manual Regulator Single Cylinder Holder ALTAIR 4 QuickCheck Cylinder Holder GB ALTAIR QuickCheck Station/ALTAIR 4 QuickCheck Station 19
20 Ordering Information MSA 6.4 ALTAIR QuickCheck Test Gas Cylinders Sensors Gas Concentration for QuickCheck Cylinder type Part No. H 2 S 40 ppm RP CO 60 ppm RP O 2 5% RP NO 2 10 ppm RP SO 2 10 ppm RP NH 3 25 ppm RP Cl 2 10 ppm RP ClO 2 2 ppm Cl 2 RP HCN 10 ppm RP ALTAIR 4 Quick Check Test Gas Cylinders Description Part No. Gas Can RP (58 l) 1.45% CH 4, 15% O 2, 60 ppm CO, 20 ppm H 2 S ALTAIR QuickCheck Station/ALTAIR 4 QuickCheck Station GB
21 MSA in Europe [ & ] Northern Europe Southern Europe Eastern Europe Central Europe Netherlands MSA Nederland Kernweg 20, 1627 LH Hoorn Phone +31 [229] Telefax +31 [229] info@msaned.nl Belgium MSA Belgium Duwijckstraat 17, 2500 Lier Phone +32 [3] Telefax +32 [3] msabelgium@msa.be Great Britain MSA Britain East Shawhead Coatbridge ML5 4TD Scotland Phone +44 [12 36] Telefax +44 [12 36] info@msabritain.co.uk Sweden MSA NORDIC Kopparbergsgatan Malmö Phone +46 [40] Telefax +46 [40] info@msanordic.se MSA SORDIN Rörläggarvägen Värnamo Phone +46 [370] Telefax +46 [370] info@sordin.se Italy MSA Italiana Via Po 13/ Rozzano [MI] Phone +39 [02] Telefax +39 [02] info-italy@msa-europe.com Spain MSA Española Narcís Monturiol, 7 Pol. Ind. del Sudoeste Sant-Just Desvern [Barcelona] Phone +34 [93] Telefax +34 [93] info@msa.es France MSA GALLET Zone Industrielle Sud Châtillon sur Chalaronne Phone +33 [474] Telefax +33 [474] message@msa-gallet.fr Poland MSA Poland Safety ul. Wschodnia 5A Raszyn k/warszawy Phone +48 [22] Telefax +48 [22] mee@msa-europe.com Czech Republic MSA Safety Czech Pikartská 1337/ Ostrava-Radvanice Phone +420 [59] Telefax +420 [59] info@msa-auer.cz Hungary MSA Safety Hungaria Francia út Budapest Phone +36 [1] Telefax +36 [1] info@msa.hu Romania MSA Safety Romania Str. Virgil Madgearu, Nr. 5 Ap. 2, Sector Bucuresti Phone +40 [21] Telefax +40 [21] office@msanet.ro Russia MSA Russia Leninsky Prospekt 2 9th Floor, office Moscow Phone +7 [495] Telefax +7 [495] msa-russia@ msa-europe.com Germany MSA AUER Thiemannstrasse Berlin Phone +49 [30] Telefax+49 [30] info@auer.de Austria MSA AUER Austria Kaplanstrasse Tulln Phone +43 [22 72] Telefax +43 [22 72] info@msa-auer.at Switzerland MSA Schweiz Eichweg Oberglatt Phone + 41 [43] Telefax+41 [43] info@msa.ch European International Sales [Africa, Asia, Australia, Latin America, Middle East] MSA EUROPE Thiemannstrasse Berlin Phone +49 [30] Telefax+49 [30] contact@msa-europe.com
SAFETY MANUAL. UltraSonic IS-5 Ultrasonic Gas Leak Detector
SAFETY MANUAL UltraSonic IS-5 Ultrasonic Gas Leak Detector The information and technical data disclosed in this document may be used and disseminated only for the purposes and to the extent specifically
More informationSAFETY MANUAL. FlameGard 5 MSIR Multi-Spectral Infrared Flame Detectors
SAFETY MANUAL FlameGard 5 MSIR Multi-Spectral Infrared Flame Detectors The information and technical data disclosed in this document may be used and disseminated only for the purposes and to the extent
More informationOperating Manual ALTAIR PRO Single Gas Detector
Operating Manual ALTAIR PRO Single Gas Detector Order No.: 10079352/02 MSA AUER GmbH D-12059 Berlin Thiemannstrasse 1 Germany MSA AUER GmbH. All rights reserved MSA DECLARATION OF CONFORMITY Declaration
More informationOperation Manual Series 47K Gas Detector [SIL 2]
Operation Manual Series 47K Gas Detector [SIL 2] Order No.: 10052472/08 MSA AUER GmbH D-12059 Berlin Thiemannstrasse 1 Germany MSA AUER GmbH. All rights reserved AUER EC DECLARATION OF CONFORMITY EC Declaration
More informationFlameGard 5 UV/IR-E Ultraviolet/Infrared Flame Detector
Installation & Maintenance Instructions FlameGard 5 UV/IR-E Ultraviolet/Infrared Flame Detector Registered Address ABLE Instruments & Controls Ltd Cutbush Park, Danehill, Lower Earley, Reading, Berkshire,
More informationULTIMA XIR ULTIMA X 3 ULTIMA XE. ULTIMA X Series. [ State-of-the-Art Gas Monitoring]
ULTIMA XIR ULTIMA X 3 ULTIMA XE ULTIMA X Series [ State-of-the-Art Gas Monitoring] ULTIMA X Gas Monitors [ Providing a unique Range of Capabilities] ULTIMA X are state-of-the-art gas monitors for continuous
More informationOperating Manual. ALTAIR Pump Probe. Order No /00. MSAsafety.com
Operating Manual ALTAIR Pump Probe Order No. 10153106/00 MSAsafety.com Jiangsu Province No. 8 Rui En Lane, Xingpu Road, Suzhou Industrial Park MSA Suzhou China MSA The Safety Company. All rights reserved
More informationALTAIR 2X Gas Detectors It s what s inside that really counts
ALTAIR 2X Gas Detectors It s what s inside that really counts Because every life has a purpose The ALTAIR 2X Gas Detector The ALTAIR 2X Gas Detector is the first one or two gas detector that incorporates
More informationia d ATEX 100 II 2G D MT 99 ATEX G 001 x EX-METER II Series [ Versatility in Explosive Atmospheres]
II 2G EEx ia d ATEX 100 D MT IIC T6 99 ATEX G 001 x Series [ Versatility in Explosive Atmospheres] [ The flexible solution for all your needs] After accidents with hazardous material the will provide reliable
More informationCHILLGARD M-100. Refrigerant Sensor. Operating Manual. Order No.: /02
CHILLGARD M-100 Refrigerant Sensor Operating Manual GB Order No.: 10086391/02 For the Declaration of Conformity, please visit the product page on MSAsafety.com. MSA Europe GmbH Schlüsselstrasse 12 8645
More informationALTAIR 5X Multigas Detector Product Specification
ALTAIR 5X Multigas Detector Product Specification PHYSICAL CHARACTERISTICS Gas delivery Size, pumped unit without IR Size, pumped unit with IR Weight Handling Case material Environmental protection Display
More informationSafe Operation Manual. Honeywell IQ Force Gas Detector
Safe Operation Manual Honeywell IQ Force Gas Detector Honeywell Analytics 800-663-4164 403-248-9226 Fax 403-575-3708 03JUN2012 P/N 50105843-047 Version 01 http://www.honeywell.com HONEYWELL IQ Force PERSONAL
More informationMODEL 9010/9020 LCD CONTROL UNIT
Data Sheet MODEL 9010/9020 LCD CONTROL UNIT The Sophisticated Hazardous Gas Warning System Supplied by.com Call us on +44 (0)118 916 9420 Email info@247able.com Model 9010/9020 LCD Control Unit [ The Sophisticated
More informationaltair 2X Gas Detectors
altair 2X Gas Detectors i t S W H at S i n S i d E t H at REALLY co u n t S + = XCell Pulse Technology With our patented* Xcell pulse technology, we have created a simple and inexpensive approach to bump
More informationaltair 2X Gas Detectors
altair 2X Gas Detectors i t S W H at S i n S i d E t H at R E A L LY co u n t S Because every life has a purpose... + = XCell Pulse Technology With our patented* Xcell pulse technology, we have created
More informationMiniCO, MiniH 2 S, MiniOX, MiniOX Remote, MiniCl 2. , and MiniClO 2. Responder Detectors Instruction Manual "! WARNING
MiniCO, MiniH 2 S, MiniOX, MiniOX Remote, MiniCl 2 TM, and MiniClO 2 TM Responder Detectors Instruction Manual "! WARNING THIS MANUAL MUST BE CAREFULLY READ BY ALL INDIVIDUALS WHO HAVE OR WILL HAVE THE
More informationOperating Manual ALTAIR 2X. Single or Two Gas Detector. Order No /00. MSAsafety.com
Operating Manual ALTAIR 2X Single or Two Gas Detector Order o. 10148950/00 MSAsafety.com MIE SAFET APPLIACES COMPA 2014 - All Rights Reserved This manual is available on the Internet at: www.msasafety.com
More informationIT S WHAT S INSIDE THAT COUNTS
IT S WHAT S INSIDE THAT COUNTS ALTAIR 4X Multigas Detector With MSA XCell Sensor Technology IT S WHAT S INSIDE THAT COUNTS OUR VISION OF SAFETY Workers who face potentially dangerous situations deserve
More informationMSA XCell Sensors. Because every life has a purpose
MSA XCell Sensors Because every life has a purpose XCell Sensors MSA XCell Sensors are a breakthrough in sensor design, enabling faster response and shorter span calibrations, saving you time and money.
More informationTGM 8 GAS Monitor. instruction manual. INSTRUMENT DIVISION MSA-INST or FAX (412) MSA INTERNATIONAL (412) or FAX (412)
TGM 8 GAS Monitor instruction manual THIS MANUAL MUST BE CAREFULLY READ BY ALL INDIVIDUALS WHO HAVE OR WILL HAVE THE RESPONSIBILITY FOR INSTALLING, USING OR SERVICING THIS PRODUCT. Like any piece of complex
More informationIT S WHAT S INSIDE THAT
ALTAIR 4X Multi-Gas Detector with MSA XCell Sensor Technology IT S WHAT S INSIDE THAT COUNTS Because every life has a purpose Performance beyond what you see Our Vision of Safety. Workers who face potentially
More informationTS400. Operating Manual. Test Station for Microtector II Series (G450/G460)
Operating Manual TS400 Test Station for Microtector II Series (G450/G460) GfG GESELLSCHAFT FÜR GERÄTEBAU MBH KLÖNNESTRASSE 99 44143 DORTMUND, Germany TEL. +49 / (0)2 31 / 5 64 00 0 FAX +49 / (0)2 31 /
More informationIT S WHAT S INSIDE THAT COUNTS
IT S WHAT S INSIDE THAT COUNTS AltAir 4X Multigas Detector With MSA XCell Sensor technology Rugged stainless steel clip MSA Link Software-ready MotionAlert sensor signals man down Ultra-bright LEDs on
More informationALTAIR 5X Multigas Detector
ALTAIR 5X Multigas Detector With MSA XCell Sensor Technology IT S WHAT S INSIDE THAT COUNTS Now with advanced PID for VOC detection Because every life has a purpose... Save Time, Save Money, Save Lives
More informationTGM 1 GAS Monitor. instruction manual. INSTRUMENT DIVISION MSA-INST or FAX (412) MSA INTERNATIONAL (412) or FAX (412)
TGM 1 GAS Monitor instruction manual THIS MANUAL MUST BE CAREFULLY READ BY ALL INDIVIDUALS WHO HAVE OR WILL HAVE THE RESPONSIBILITY FOR INSTALLING, USING OR SERVICING THIS PRODUCT. Like any piece of complex
More informationAutomated Test System SIMPLICITY COUNTS MSA s New GALAXY GX2 Automated Test System / Advanced Safety Management Effortless Operation
Automated Test System SIMPLICITY COUNTS MSA s New GALAXY GX2 Automated Test System / Advanced Safety Management Effortless Operation Advanced MSA Link Pro Software Analyze All Data, Any Time MSA Link Pro
More informationPac Ex 2. Elegant and smart Explosive hazards and lack or surplus of oxygen under control
Pac Ex 2 Elegant and smart Explosive hazards and lack or surplus of oxygen under control Portable instruments protecting people against explosive hazards and lack or surplus of oxygen are used in a large
More informationALTAIR 5X Multigas Detector
ALTAIR 5X Multigas Detector With MSA XCell Sensor Technology IT S WHAT S INSIDE THAT COUNTS Now with advanced PID for VOC Detection Because every life has a purpose IT S WHAT S INSIDE THAT COUNTS WORKERS
More informationOperating Manual ALTAIR 5X PID. Multigas Detector. Order No.: /00. MSAsafety.com
Operating Manual ALTAIR 5X PID Multigas Detector Order No.: 10165710/00 MSAsafety.com 1000 Cranberry Woods Drive Cranberry Township, PA 16066 A Phone 1-800-MSA-2222 Fax 1-800-967-0398 For your local MSA
More informationIT S WHAT S INSIDE THAT COUNTS
IT S WHAT S INSIDE THAT COUNTS ALTAIR 4X Multigas Detector With MSA XCell Sensor Technology IT S WHAT S INSIDE THAT COUNTS OUR VISION OF SAFETY Workers who face potentially hazardous situations deserve
More informationIMR IX176 Portable Gas Detector User Manual
IMR Portable Gas Detector User Manual Read this manual carefully before using this device. (727) 328-2818 / (800) RING-IMR Fax: (727) 328-2826 www.imrusa.com Ver. 1.0A4 CONTENTS SERVICE GUIDELINES... 3
More informationMacurco Single-Gas XL Series Monitor, CM-1XL Carbon Monoxide (CO), HS-1XL Hydrogen Sulfide (H2S) User Instructions
Macurco Single-Gas XL Series Monitor, CM-1XL Carbon Monoxide (CO), HS-1XL Hydrogen Sulfide (H2S) User Instructions Important: Keep these User Instructions for reference 2 TABLE OF CONTENTS GENERAL SAFETY
More informationmicropac. Personal Gas Detection Instruments
micropac. Personal Gas Detection Instruments ST-53-99 Chemical Industry... Metal Industry... ST-51-99 In many fields of industry and at numerous workplaces people have to be protected against gas hazards.
More informationSIRIUS. [ MultiGas Detector ] Instruction Manual. Manufactured by: MSA INSTRUMENT DIVISION P.O.Box 427, Pittsburgh, Pennsylvania 15230
SIRIUS [ MultiGas Detector ] Instruction Manual Manufactured by: MSA INSTRUMENT DIVISION P.O.Box 427, Pittsburgh, Pennsylvania 15230 [Europe] Rev.03 / 11 2010 10092677 Y-HLS / GB 1 Warranty and Liability
More informationVICOTEC450 EXTRACTIVE VISIBILITY MEASURING DEVICES EXTRACTIVE VISIBILITY MEASUREMENT FOR INCREASED SAFETY IN TUNNELS
PRODUCT INFORMATION VICOTEC450 EXTRACTIVE VISIBILITY MEASURING DEVICES EXTRACTIVE VISIBILITY MEASUREMENT FOR INCREASED SAFETY IN TUNNELS Tunnel sensors EXTRACTIVE VISIBILITY MEASUREMENT FOR INCREASED SAFETY
More informationSERIES 47K GAS SENSOR
Data Sheet SERIES 47K GAS SENSOR Remote Combustible Gas Detection Supplied by.com Call us on +44 (0)118 916 9420 Email info@247able.com Series 47K Gas Sensor Remote Combustible Gas Detection MSA remote
More informationInstruction Manual. Alarm Unit For Low Gas Level # Read manual before use! Observe all safety information! Keep manual for future use!
Mess-, Regel- und Überwachungsgeräte für Haustechnik, Industrie und Umweltschutz Lindenstraße 20 74363 Güglingen Telefon +49 7135-102-0 Service +49 7135-102-211 Telefax +49 7135-102-147 info@afriso.de
More informationOperation Manual Series 47K Gas Detector [SIL 2]
Operation Manual Series 47K Gas Detector [SIL 2] Order No.: 10052472/11 MSA AUER GmbH D-12059 Berlin Thiemannstrasse 1 Germany MSA AUER GmbH. All rights reserved MSA EC DECLARATION OF CONFORMITY EC Declaration
More informationToxiRAE II. PGM-1100 Series Personal Toxic Gas Monitor. User Manual
ToxiRAE II PGM-1100 Series Personal Toxic Gas Monitor User Manual 045-4003-000, Rev C May 2007 - READ BEFORE OPERATING - This manual must be carefully read by all individuals who have or will have the
More informationHPLC Column Heater Instruction Manual
HPLC Column Heater Instruction Manual For Technical Service Call 1-800-4BIORAD (1-800-424-6723) Warranty Unless otherwise specified, instruments sold by Bio-Rad Laboratories are warranted for 1 year against
More informationToxiRAE II. PGM-1100 Series Personal Toxic Gas Monitor. User Manual
ToxiRAE II PGM-1100 Series Personal Toxic Gas Monitor User Manual 045-4003-000, Revision B, March 2005 - READ BEFORE OPERATING - This manual must be carefully read by all individuals who have or will have
More informationActivation Operation Troubleshooting Part Number: Edition: Release Date:
Personal Single Gas Monitor Activation Operation Troubleshooting Part Number: Edition: Release Date: 17123100-1 8 July 21, 2017 Table of Contents Warnings and Cautionary Statements... 4 Recommended Practices...
More informationDräger X-am 5000 Multi-Gas Detection
Dräger X-am 5000 Multi-Gas Detection The Dräger X-am 5000 belongs to a generation of gas detectors, developed especially for personal monitoring applications. This 1- to 5-gas detector reliably measures
More informationPersonal Single Gas Monitor. Activation Operation Troubleshooting
Personal Single Gas Monitor Activation Operation Troubleshooting Part Number: 17123100 Revision: 3.0 Release Date: January 1, 2009 Table of Contents Warnings and Cautionary Statements... 3 Hardware Overview...
More information10yr O2NE & Safe-Ox. User Manual
Analox Ltd. 15 Ellerbeck Court, Stokesley Business Park North Yorkshire, TS9 5PT, UK T: +44 (0)1642 711400 F: +44 (0)1642 713900 W: www.analox.net E: info@analox.net This support line is closed on UK public
More informationALTAIR 4 Four Gas Multigas Detector
Operating Manual ALTAIR 4 Four Gas Multigas Detector Manufactured by MSA INSTRUMENT DIVISION P.O. Box 427, Pittsburgh, Pennsylvania 15230, USA P/N 10088975 WARNING Read this manual carefully before using
More informationReturn loss measurement of fiber optic components
Return loss measurement of fiber optic components Solutions Brief 815-1 How the HP 8153A/HP 81534A measure return loss of fiber optic components? The significance of return loss The introduction of new
More informationClip SGD. Portable single Gas Detector. User manual
Portable single Gas Detector User manual M07716 Issue 1 October 2017 WARNING Clip SGD is a hazardous area certified gas detector and as such must be operated and maintained in strict accordance with the
More informationG450. Multi-gas Detector. Field Operation Manual Oak Valley Dr, Ste 20, Ann Arbor MI USA (800) (734)
G450 Multi-gas Detector Field Operation Manual 1194 Oak Valley Dr, Ste 20, Ann Arbor MI 48108 USA (800) 959-0329 (734) 769-0573 www.gfg-inc.com Warranty GfG Instrumentation warrants our products to be
More informationStand Alone DYNAGARD SP Operation Manual
Stand Alone DYNAGARD SP Operation Manual Contents Page General Description 1 Detection Principle 1 For Your Safety 1 Design 2 Mounting 2 Mounting Position of DYNAGARD SP 2 Installation of Electrical Connections
More informationGasAlert. Introduction. WWarning. Table 1. GasAlert Models. Gas Monitored
Introduction Table 1. GasAlert Models WWarning To ensure your personal safety, read Safety Information before you use the detector. The GasAlert gas detector ( the detector ) warns of hazardous gas at
More informationDräger X-am 5000 Multi-Gas Detection
Dräger X-am 5000 Multi-Gas Detection The Dräger X-am 5000 belongs to a generation of gas detectors, developed especially for personal monitoring applications. This 1- to 5-gas detector reliably measures
More informationMSA Portable Gas Detection
Because every life has a purpose MSA Portable Gas Detection Marine Chemist Association Seminar 2015 Paul Jung Product Line Manager Portable Instruments Our Mission & Vision Mission to see to it that men
More informationPRODUCT INFORMATION. GM901 Carbon Monoxide Gas Analyzers CO MEASUREMENT FOR EMISSION MONITORING AND PROCESS CONTROL. In-situ Gas Analyzers
PRODUCT INFORMATION GM901 Carbon Monoxide Gas Analyzers CO MEASUREMENT FOR EMISSION MONITORING AND PROCESS CONTROL In-situ Gas Analyzers MONITORING CARBON MONOXIDE EMISSIONS INNOVATIVELY AND EFFICIENTLY
More informationTechnical Reference Guide. 1, 2, 3, and 4-Gas Detector
Technical Reference Guide 1, 2, 3, and 4-Gas Detector Introduction 1 Zeroing the Sensors 1 Contacting BW Technologies by Honeywell 2 Safety Information - Read First 2 Sensor Poisons and Contaminants 5
More informationSAFETY CERTIFIED MODEL FP-700 COMBUSTIBLE GAS DETECTOR
SAFETY MANUAL SIL 2 Certified Model FP-700 Combustible Hydrocarbon Gas Sensor Version 2.0 1 SAFETY CERTIFIED MODEL FP-700 COMBUSTIBLE GAS DETECTOR This manual addresses the specific requirements and recommendations
More informationDräger X-am 7000 Multi-Gas Detection
Dräger X-am 7000 Multi-Gas Detection Dräger X-am 7000 is the solution for the simultaneous and continuous measurement of up to five gases. It is the ideal companion in a variety of applications where the
More informationInstruction Manual. CAT 100 & CAT 200 Continuous Analyzer Transmitter Addendum for Hazardous Area Applications. 3 rd Edition 05/2007
CAT 100 & CAT 200 Continuous Analyzer Transmitter Addendum for Hazardous Area Applications 3 rd Edition www.emersonprocess.com CAT 100 & CAT 200 Addendum for Hazardous Area Applications ESSENTIAL INSTRUCTIONS
More informationBeacon 200 Gas Monitor Operator s Manual. Part Number: RK Released: 6/6/08
Beacon 200 Gas Monitor Operator s Manual Part Number: 71-2102RK Released: 6/6/08 Table of Contents Chapter 1: Introduction.................................................3 Overview.............................................................3
More informationGasAlertMicro 5. 1, 2, 3, 4, and 5 Gas Detectors. Quick Reference Guide
O 2, CO, H 2 S, PH 3, SO 2, Cl 2, NH 3, NO 2, HCN, ClO 2, O 3, VOC, and Combustibles 1, 2, 3, 4, and 5 Gas Detectors Quick Reference Guide Limited Warranty & Limitation of Liability BW Technologies Ltd.
More informationCT-7 Series Toxic Gas Transmitter Operator s Manual
65-2341 CT-7 Series Toxic Gas Transmitter Operator s Manual Part Number: 71-0424 Revision: P1 Released: 6/8/17 RKI Instruments, Inc. www.rkiinstruments.com WARNING Read and understand this instruction
More informationToxi Vision IR CH 4. Single Gas Detector. Reference Manual
Toxi Vision IR CH 4 Single Gas Detector Reference Manual Sperian Instrumentation 651 South Main Street Middletown, CT 06457 (800) 711-6776 (860) 344-1079 Fax (860) 344-1068 Part number 13-316 Version 1.10
More informationUser Manual. GPD 3000 e. gas detector
User Manual GPD 3000 e gas detector Table of contents Application...3 Unit ppm, Vol.%...4 Operation elements GDP 3000 e...5 Gas sensor...6 Important Information off sensor!...6 Measurement range...7 Turning
More informationOperating Instructions and Maintenance Manual
Impact Series Operating Instructions and Maintenance Manual 4.9.4 Enforcer Calibration This is designed for oxygen, flammable (catalytic or IR LEL sensor), carbon monoxide and hydrogen sulfide, using a
More informationDräger X-am 2500 Multi-Gas Detection
Dräger X-am 2500 Multi-Gas Detection The Dräger X-am 2500 was especially developed for use as personal protection. The 1 to 4 gas detector reliably detects combustible gases and vapours, as well as O2,
More informationFF-SR Series. FF-SR25933 Two-hand Safety Control Module FF-SR25933
Two-hand Safety Control Module FF-SR Series FEATURES Complies with the Machinery Directive 98/37/EC and UL 508 Type III C per EN 574, Category 4 per EN 954-1 Inputs for two dual-contact pushbuttons 0,5
More informationSTT 3000 Series STT250 /150 Surge Protection Device Specifications
STT 3000 Series STT250 /150 Surge Protection Device Specifications Part Number: 46188660-501 EN2I-6061 June 2011 Introduction This device is designed to give the Smart Temperature Transmitter maximum protection
More informationOperations Manual TS400. Test Station for G450/G460 Gas Detector
TS400 Test Station for G450/G460 Gas Detector Operations Manual 1194 Oak Valley Dr, Ste 20, Ann Arbor MI 48108 USA (800) 959-0329 (734) 769-0573 www.gfg-inc.com GfG Products for Increased Safety Congratulations
More informationSENSIT HXG-2d Combustible Gas Detector
INSTRUCTION MANUAL SENSIT HXG-2d Combustible Gas Detector Read and understand instructions before use. 710183 UL 913 Intrinsically Safe for Use in Class I, Groups C and D, T3 Hazardous Locations For use
More informationPRODUCT INFORMATION. SureSense THE SURE WAY TO DETECT ANY OBJECT. Photoelectric Sensors
PRODUCT INFORMATION SureSense THE SURE WAY TO DETECT ANY OBJECT Photoelectric Sensors SureSense PHOTOELECTRIC SENSORS A FAMILY BUILT TO SOLVE ALL YOUR SENSING NEEDS Solving a wide range of sensing applications
More informationCDL 210. Operating manual. CO2 Logger. Best.-Nr
PCE Instruments France EURL 76, Rue de la Plaine des Bouchers 67100 Strasbourg France Téléphone: +33 (0) 972 3537 17 Numéro de fax: +33 (0) 972 3537 18 info@pce-france.fr www.pce-instruments.com/french
More informationSingle Gas Monitor 777
Single Gas Monitor 777 Carbon Monoxide The Value Leader TM www.tpi-thevalueleader.com Contents Agency Listings... 1 Safety Information... 2 General Description & Physical Specifications... 3 Product Features...4
More informationPRODUCT INFORMATION HISIC450 OVERHEIGHT DETECTORS RELIABLE OVERHEIGHT DETECTION OF VEHICLES. Traffic sensors
PRODUCT INFORMATION HISIC450 OVERHEIGHT DETECTORS RELIABLE OVERHEIGHT DETECTION OF VEHICLES Traffic sensors RELIABLE OVERHEIGHT DETECTION OF VEHICLES Reliable detection of vehicles in flowing, slow-moving
More informationGasAlertMicroClip X3 Specifications
GasAlertMicroClip X3 Specifications The instrument must satisfy the following: Physical Specifications Size (h x w x d) Physical size shall be no larger than 4.4 x 2.4 x 1.2 in. / 11.3 x 6.0 x 3.2 cm Weight
More informationSirius MultiGas Detector
Sirius MultiGas Detector Operating Manual In North America, to contact your nearest stocking location, dial toll-free 1-800-MSA-2222 To contact MSA International, dial 1-412-967-3354 or 1-800-MSA-7777
More informationCertificate of Compliance
FF0 001 Certificate of Compliance Certificate: 70116284 (LR 064969_0_00) Issued to: Mine Safety Appliances Company North American Division 1000 Cranberry Woods Dr Cranberry Township, Pennsylvania 16066-5296
More informationDräger Pac 7000 Single-Gas Detection Device
Single-Gas Detection Device Safety at the workplace always takes priority: depending on the sensor selection, the single gas detector,, provides a reliable warning against dangerous concentrations of 14
More informationInstruction Manual. Electronic gas and smoke detector GRM
Mess-, Regel- und Überwachungsgeräte für Haustechnik, Industrie und Umweltschutz Lindenstraße 20 DE-74363 Güglingen Telefon +49(0)7135-102-0 Service +49(0)7135-102-211 Telefax +49(0)7135-102-147 E-Mail
More informationII 2 G. SENSIT HXG-2d. Ex ib IIB T3 IP20. Suitable for combustible gases such as Methane, Butane, Propane and Natural Gas MADE IN USA
SENSIT HXG-2d Combustible Gas Detector INSTRUCTION MANUAL Read and understand Instructions before use. PENDING II 2 G Ex ib IIB T3 IP20 PENDING (USA) Class 1, Division 1, Groups C and D, T3 Suitable for
More informationGMS800 FIDOR TOTAL HYDROCARBON ANALYZER CONTINUOUS MONITORING OF HYDROCARBON EMISSIONS
Product Information GMS800 FIDOR TOTAL HYDROCARBON ANALYZER CONTINUOUS MONITORING OF HYDROCARBON EMISSIONS Gas Analyzers SOLUTION FOR CONTINUOUS HYDROCARBON MEASUREMENTS The compact GMS800 FIDOR extractive
More informationDräger X-am approved as type MQG 0010 Multi-Gas Monitor Technical Manual _01.eps
Dräger X-am 5000 approved as type MQG 0010 Multi-Gas Monitor Technical Manual 04723999_01.eps Content Content For Your Safety..................................................... 4 Safety symbols used
More informationOperator s Manual. 1, 2, 3, and 4-Gas Detector
Operator s Manual 1, 2, 3, and 4-Gas Detector Limited Warranty and Limitation Liability BW Technologies LP (BW) warrants the product to be free from defects in material and workmanship under normal use
More informationHazLoc Essential Guides:
HazLoc Essential Guides: Intertek Deeside Lane, Chester, CH1 6DD UK hazloc@intertek.com +44 1244 882590 www.intertek-hazloc.com Contents Contents... 1 Introduction... 1 Introduction to Ex... 1 ATEX Overview...
More informationAM54. Armored Variable Area Flowmeter Series AM54331/32 Series AM54371/72/73/74 Series AM54431/32 Series AM54471/72/73/74
AM54 Instruction Bulletin North American Installation Addendum Document Number PN25016 Armored Variable Area Flowmeter Series AM54331/32 Series AM54371/72/73/74 Series AM54431/32 Series AM54471/72/73/74
More informationInternational Product Compliance and Certifications. Product Safety Consulting, Inc.
International Product Compliance and Certifications - USA - Canada - Europe - Other International - CB Scheme Agenda This was the only mark. Now, these are the marks A bit of history.. UL founded in 1894
More informationPROTÉGÉ ZM SINGLE GAS MONITOR
PROTÉGÉ ZM SINGLE GAS MONITOR KEEPING YOUR WORKFORCE SAFE JUST GOT EASIER. PROTÉGÉ ZM SINGLE GAS MONITOR. Scott Safety announces the introduction of the Protégé ZM Single Gas Monitor an easy-to-use, zero-maintenance
More informationPERSONAL COMBUSTIBLE GAS DETECTOR
PERSONAL COMBUSTIBLE GAS DETECTOR FEATURES Continuous LCD Reading with Back Light Electronic Audio Alarm Visual Alarms include Icons and Red LED Easy Handling Certified Intrinsically Safe Auto Test When
More informationUser manual CTX 300 CO 2
User manual CTX 300 CO 2 Analogic CO 2 Gas Detector Part Number: NP300CO2GB Revision: B.0 07/2015 The Fixed Gas Detection Experts Copyright 2016 by Oldham S.A.S. April 2016 All rights reserved. No reproduction
More informationAERO 20 AERO ENGLISH - Operating Instructions
www.nilfisk-alto.com AERO 20 AERO 25 302002328-2008-05-21 ENGLISH - Operating Instructions Important safety instructions Before using the vacuum cleaner, always read the operating instructions and keep
More informationBST-MG01 Single Gas Detecting Alarm Manual Instruction
BST-MG01 Single Gas Detecting Alarm Manual Instruction Notice: BST-MG01 single gas detecting alarm is applied to operation of portable gas detectors like Oxygen(O2), Hydrogen Sulphid(H2S), Carbon Monoxide(CO),
More informationAccuracy counts! Sensors with analog output
Accuracy counts! Sensors with analog output Company Contrinex is a leading international high-tech sensor company with headquarters in Switzerland. Contrinex offers quality and highly innovative sensing
More informationSAFETY MANUAL. Electrochemical Gas Detector GT3000 Series Includes Transmitter (GTX) with H 2 S or O 2 Sensor Module (GTS)
SAFETY MANUAL Electrochemical Gas Detector GT3000 Series Includes Transmitter (GTX) with H 2 S or O 2 Sensor Module (GTS) Sensor Module (GTS) Transmitter (GTX) Detector (GT3000) SAFETY CERTIFIED GT3000
More informationInstruction Manual. Seco. Dry-running Rotary Vane Vacuum Pumps SV 1003 D, SV 1005 D
Instruction Manual Seco Dry-running Rotary Vane Vacuum Pumps SV 1003 D, SV 1005 D Busch Vyroba CZ s.r.o. Svárovská 620, CZ 460 01, Liberec 11 Czech Republic 0870140843/A0002_en / Original instructions
More information3. To protect against electric shock, do not immerse cord, plugs, or machine in water or other liquid.
1 IMPORTANT SAFEGUARDS When using electrical appliances, basic safety precautions should always be followed to reduce the risk of fire, electric shock, and/or injury to persons, including the following:
More informationRLD-H10G Refrigerant Leak Detectors
FANs 125, 121 Product/Technical Bulletin RLD-H10G Issue Date 0396 RLD-H10G Refrigerant Leak Detectors The RLD-H10G is a professional grade leak detector for use by refrigeration and air conditioning technicians.
More informationControllers. Instruction Manual WARNING
Controllers Instruction Manual WARNING THIS MANUAL MUST BE CAREFULLY READ BY ALL INDIVIDUALS WHO HAVE OR WILL HAVE THE RESPONSIBILITY FOR INSTALLING, USING OR SERVICING THIS PRODUCT. Like any piece of
More informationSCHMIDT LED Measured Value Display MD Instructions for Use
SCHMIDT LED Measured Value Display MD 10.010 Instructions for Use Table of Contents 1 Important Information... 3 2 Application range... 4 3 Mounting instructions... 4 4 Electrical connection... 6 5 Signalizations...
More informationOperations Manual TS400. Test Station for G450/G460 Gas Detector
TS400 Test Station for G450/G460 Gas Detector Operations Manual 1194 Oak Valley Dr, Ste 20, Ann Arbor MI 48108 USA (800) 959-0329 (734) 769-0573 www.goodforgas.com GfG Products for Increased Safety Congratulations
More information1.1 Definitions of alert icons. For your safety
For your safety 1 For your safety Before using this product, carefully read these Instructions for Use and those of the associated products. Strictly follow the Instructions for Use. The user must fully
More informationMulti-gas made simple
Multi-Gas (1 4) Detector H 2 S, CO, O 2, LEL Multi-gas made simple The easy to wear, slim and compact GasAlertMicroClip provides affordable protection from atmospheric gas hazards and extended battery
More information