1 MANUAL No E2.6 Published in Jul.2002 Revised in Oct.2012 M-15 SERIES MONITOR Model M-15B1 ACCELERATION MONITOR INSTRUCTION MANUAL

2 FOREWORD M-15B1 FOR THE SAFE OPERATION OF THIS UNIT Thank you for using the MODEL M-15 Series Monitor. SHINKAWA Sensor Technology, Inc. applies strict inspection and quality control standards to ensure the high reliability of its products. Please study the contents of this and related manuals carefully before installing or operating the equipment. Keep them handy for future reference. REGARDING PRECAUTIONS Please read and understand completely the Precautions on pages ii-v of this manual before using this or any of our products. Indicators and other symbols that are used in the Precautions section appear throughout this manual on the left side of the page. This product is designed to be used by specialists or persons with a thorough knowledge of our products. Please make certain that the end user receives the instruction manual delivered with this product. BEFORE USING After this system is delivered, inspect it for possible damage that occurred during shipment, and be sure it is the item that was ordered. In the unlikely event it was damaged or does not function according to specification, please contact the SHINKAWA office or dealer nearest you. Store this system under the ambient conditions given in the specification. Avoid places where it may be exposed to high humidity or corrosive gases, but rather in the environmental conditions specified in the manual. i

3 PRECAUTIONS M-15B1 Follow closely all instructions written in the manual. Important information is contained in this manual that will prevent serious injuries to individuals and damage to this product and other equipment. Read and understand the contents of this section, including the Indicators and Symbols, before reading other sections of the manual. Indicators Indicator GER WARNING Description If this unit is not handled properly, dangerous situations may occur that could result in death or serious injury. If this unit is not handled properly, dangerous situations may occur that could result in slight or moderate injury, or damage to the unit or other equipment. Serious injury means loss of sight, cuts, burns by high or low temperatures, electrocution, broken bones, poisoning, etc. which would require hospitalization or long-term outpatient care, and unpleasant after-effects. Slight or moderate injury refers to cuts or electrical shocks that would not require any hospitalization or long-term outpatient care. Damage to equipment refers to breakage of a single unit or to destruction of several pieces of equipment. Other Symbols Symbol Description Prohibited. Details may be found inside or near the symbol, and further explained in words or pictures. Mandatory. Details may be found inside or near the symbol, and further explained in words or pictures. ii

4 PRECAUTIONS M-15B1 In order to use this unit safely, strictly follow the installation and operation instructions that are required by law, and observe all warnings as described in this manual. SHINKAWA Sensor Technology does not accept responsibility for the misuse of this or any of its products. To Use This Unit Safely GER Grounding Prohibited A ground connection is mandatory. Failure to do so may cause electrocution or improper operation of the unit. Do not perform wiring work while the power supply is on. This may result in electrocution. To Use the M-15B1 Monitoring System Safely WARNING Mandatory Mandatory Mandatory Connect the input power supply terminal within the permissible power supply voltage and frequency, as written in the specifications. Instrument may be damaged and cause incorrect operation. To avoid incorrect operation in case this monitoring system experiences a problem, the Interlock function should be setup in the system. Before the Signal Transmission Cable Mega-test is performed, it is mandatory to disconnect all cables from this respective monitor. When reconnecting the cables to the monitor after the test is completed, the cables must be forced to short circuit in order to discharge the electric charge. Reconnection of the cables while the monitor is still charged may cause damage. iii

5 PRECAUTIONS M-15B1 To Use the M-15B1 Monitoring System Safely WARNING Prohibited Prohibited Prohibited Prohibited Prohibited Since this monitoring system is a highly precise instrument, care should be taken while handling so that there are no heavy shocks or excessive vibration. Do not set up any specifications while the unit is in operation. Alarm contact might cause output. Do not touch the main board of this unit. Instruments may be damaged and cause incorrect operation of the system. Do not measure the insulation resistance or withstand voltage in any places other than the specifications of insulation resistance and withstand voltage are prescribed. Instrument may be damaged. Do not remodel without permission. The product guarantee will be voided. iv

6 PRECAUTIONS M-15B1 EMC (Electro-Magnetic Compatibility) The M-15 series monitoring system has an excellent reputation for EMC features. However, these characteristics may be affected depending upon the surrounding conditions, such as the amount of generated noise, temperature, humidity or machine arrangement. Therefore, please study the Precautions below before using this unit. WARNING Mandatory Mandatory Mandatory Mandatory Mandatory Prohibited This unit should be used by specialists or persons with a thorough knowledge of our products. In case shielded cables are used, the length of unshielded wire should be 100 mm or shorter. Install the surge killer on the relay that is used around this unit, related units and connected signal cables. Tighten securely the screws. The faceplate on the monitor unit should be anchored with screws. To avoid incorrect operation in case this unit experiences a problem, the Interlock function should be setup in the system. Input signal cables should not be arranged in a row with other cables associated with the Control System or Power Supply System. It is recommended to use separate wiring ducts. Do not use radio instruments near this unit. Prohibited Noise Filters Inserting the FERRITE CORE (see note 1) is effective in preventing noise mediated through cables. In case the surrounding conditions are poor, the use of any parts, including this part, in order to prevent noise is highly recommended. (note 1) ZCAT : TDK v

7 CONTENTS M-15B1 1. GENERAL SPECIFICATONS AND OUTLINE DRAWINGS 2-1 SPECIFICATIONS OUTLINE DRAWINGS WIRING METHODS 3-1 SIGNAL TRANSMISSION CABLE WIRING INTRINSICAL SAFETY CONSTRUCTION WIRING NON-INTRINSICAL SAFETY CONSTRUCTION WIRING FUNCTION DESCRIPTION 4-1 INDICATION FUNCTION ALARM FUNCTION ALARM DELAY FUNCTION UENCE FUNCTION SUPPRESSION FUNCTION POWER-ON TIMER MODE ALARM RELAY MODE ALARM RE METHOD MONITOR OUTPUT RECORDER OUTPUT SELF DIAGNOSTIC FUNCTION LAMP TEST PART NAMES AND DESCRIPTIONS 5-1 PART NAMES ON THE FRONT PANEL PART NAMES ON THE REAR PANEL PART NAMES AND DESCRIPTIONS vi

8 CONTENTS M-15B1 6. SCREEN DESCRIPTION AND KEY OPERATION METHOD 6-1 DIGITAL INDICATOR DESCRIPTION SWIT INDICATION ANNELS OF MEASURED IBRATION ALUE SWIT INDICATION TO ( ALUE OR ALARM UP ALUE INDICATION, UP MEASURED ALUE INDICATION) ALARM FUNCTION UP BASIC FUNCTON UP BLOCK DIAGRAMS 7-1 BLOCK DIAGRAMS BLOCK DIAGRAM DESCRIPTION MAINTENANCE AND INSPECTION METHODS 8-1 CONTENTS OF PERIODIC INSPECTION AND MAINTENANCE PERIODIC INSPECTION AND MAINTENANCE ECK LIST MAINES TO USE FOR INSPECTION AND MAINTENANCE CONFIRMATION OF OK (INPUT ABNORMALITY ALARM) OPERATION CONFIRMATION OF GER / ALERT ALARM OPERATION CONFIRMATION OF THE MEASURED ALUE AND RECORDER 0UTPUT TROUBLESHOOTING vii

9 1. GENERAL M-15B1 The M-15B1 monitoring system monitors the operating condition of rotating machinery, using the signals from the transducer that is installed in the rotating machinery. The monitored measured value and alarm setup value are digitally displayed on the 7-segment LED display. With the built-in CPU, switching the indication mode and alarm setup with the key operation is easily conducted. Warning Regarding the Cable Mega-test (Insulation Test) Before the Cable Mega-test is performed, it is mandatory to disconnect all cables from their respective instruments. When reconnecting the cables to the instruments after the test is completed, the cables must be forced to short circuit in order to discharge the electric charge. Reconnection of the cables while the unit is still charged may cause damage to the instruments. -1-

10 2. SPECIFICATIONS AND OUTLINE DRAWINGS M-15B1 2-1 SPECIFICATIONS When add l specifications not Model / Spec. Code upon delivery / Add l Spec. Code ( requested, leave blank. ) M-15B /NB1 /EX /MC /CSU Monitor Range Input Transducer Frequency Response Recorder Output Non-incendive*1 Sensitivity correction Memory Card 1 25mm/s pk 1 CA Series 0 10Hz to 10kHz (-3dB) 0 4 to 20mADC Class I 1 TIIS(IEC) 1 Compact Flash 2 1in/s pk (100m/9.8 pk) 1 1 to 5DC 1 Division pk CSA 4 2G pk Custom Setup Specifications No. of Channels No. of Channels Monitor Range Monitor Range elocity Acceleration : 4ch Field changeable from 1 to 4ch. : 25mm/s pk or 1in/s pk Field changeable from 25mm/s pk to 50mm/s pk, or from 1in/s pk to 2in/s pk. : 20 pk or 2G pk Field changeable from 20 pk to 50 pk, or from 2G pk to 5G pk! WARNING Monitor range should stay within the setup range. Otherwise, precise measurement can not be expected. Input Transducer CA Series Acceleration Transducer Input Signal : 2 to 22DC Input Impedance : Approx. 50kΩ Frequency Response 2nd-Order Band-pass Filter : 10Hz to 10kHz(-3dB) Sequence Function Used to prevent alarm output that is caused by excessive vibration during machine startup. Block off the GER/ALERT alarm, or switch the alarm setup value to another number magnified by setup number. Sequence Setup : Field changeable to alarm blockade or magnification number from 1 to 5.! WARNING This function may leave the machine unprotected. While the. magnification value is set to -, machine is unprotected as alarms are disabled. While the. magnification set to a value between 2 to 5, and. function is in effect, alarm setpoints are multiplied by the set value. When alarm setpoints exceed 150% of the monitor range, alarm may not output. Factory Preset alue : 1 magnification Suppression Function If the vibration value is less than the setup value, this function is forced to suppress the measured vibration value and recorder output. 100(%) Suppression Setup alue : Field changeable from 0 to 5%. (in 0.1% Increments) Factory Preset alue : 1.0% Conversion Accuracy Output & display 0 0 *1 10 Input 100(%) Output & Display vs. Input I/O Conversion Accuracy : ±1.5% of F.S at 100Hz at 25 C ±3.0% of F.S at 100Hz at 0 to 50 C -2- Output Recorder Output : 1 to 5DC (Output Impedance: 250Ω) 4 to 20mADC (Max. Load Resistance: 600Ω) (Select from Model Code) Recorder Output Gain : Proportional to measured value, can be setup at magnification of 0.5, 1 or 2, field changeable. Factory Preset alue : 1 magnification Monitor output : Input signal of CA Series is output via buffer amplifier. Output Signal : 0.8 to 20DC Output Impedance : Approx. 100Ω(Max.5mA) Transducer Power Supply : Output oltage : 24DC (4mA constant current) Alarm Alarm Setup : GER Alarm: 1 point, from 0 to 100% of monitor range, field changeable ALERT Alarm: 1 point, from 0 to 100% of monitor range, field changeable Factory Preset alue : GER Alarm 50% of range. ALERT Alarm 50% of range. Alarm Delay Time : 0.5, 1 or 3 sec. field changeable. Factory Preset alue : GER Alarm 3 sec. ALERT Alarm 3 sec. Alarm Reset : AUTO-RE or SELF-HOLD field changeable. Factory Preset Mode : GER Alarm AUTO-RE ALERT Alarm AUTO-RE OK Alarm AUTO-REEST Relay Mode : NORMALLY DE-ENERGIZED or NORMALLY ENERGIZED field changeable. Factory Preset Mode : GER Alarm normally de-energized ALERT Alarm normally de-energized OK Alarm normally de-energized Alarm Contact Type : C Contact, Dry contact Alarm Contact Output : GER 1 point (OR of all channels) ALERT 1 point (OR of all channels) OK 1 point (OR of all channels) Max. Contact oltage : 250AC, 220DC Max. Contact Current : 2A Max. Contact Power : 125A, 60W (Resistance load) Note)*1 When the specification is CSA:30DC 2A only Display Display : 3 digits 7-segment LED display. Display Accuracy : ±(1.5% of F.S.+1 digit) at 25 C Measured alue : SCAN, FIX or AUTO Mode, field changeable. Factory Preset Mode : AUTO(Max. value indicated) Sensor Gap : Sensor gap voltage value Setup alue : GER Alarm setup value ALERT Alarm setup value Alarm Indication : GER Alarm () (Red LED lighting) ALERT Alarm () (Red LED flashing) Input Abnormality Alarm (OK) (Green LED flashing) Sequence Lamp : Yellow LED (.) 30112E2.3

11 2. SPECIFICATIONS AND OUTLINE DRAWINGS M-15B1 Specifications Operation Contact Input Normally open. Alarm Reset (RES) : Reset the SELF-HOLD Alarm. Sequence () : Prevent the alarm output during machine startup. When sequence circuit is in progress,. lamp on front panel is lit. Contact Type : Dry contact Temperature Range Operating Temperature : 0 to +50 C Storage Temperature : -30 to +85 C -20 to +60 C (with Memory card slot) Relative Humidity : 20 to 90%RH (non-condensing) Insulation Resistance Input-Output-Power-Contact-GND : 100M at 500DC Withstand oltage Power-Contact-GND : 1500AC for 1 minute. Input-Output-Power, Contact, GND : 100AC for 1 minute. Power Consumption When all channels are used : Less than including 80A Power Supply oltage Rated Power Supply oltage : 100 to 240AC (50/60Hz) Power Supply oltage Range : 85 to 264AC Materials and Finish Main Body : SPCC Unichromated plating finish (white) Face Plate : Sheet on aluminum Sheet : Reinforced polyester top Weight Main Body : Approx. 3.5kg Notes Memory contents ibration alue Input oltage (GAP voltage) Alarm Status Sampling Time External Operation Contact Status (Reset, Sequence Contact) Alarm Set alue Alarm Relay Mode Alarm Reset Mode Alarm Delay Time Sequence Set alue Suppression Set alue Input Abnormal Alarm Set oltage alue Sampling Time 10 sec. ~ 24 hrs. (Can be set up in 10 sec. increments) File File Type File Name : Text : log1.txt, log2.txt Memory Card Slot (/MC1) Recording Media Recommended Memory Card : CompactFlash TM Manufactured by SanDisk Recommended Memory Capacity : 32MB, 64MB Others : Compact Flash Card Adapter is needed Recording Time : Approx. 30 days when 10 sec. sampling is set up. (Memory Capacity 32MB is used) Display Busy Status Stand-by Status Communication Error Status Card Not Inserted Status Others ALARM CONTACT OPERATION : Yellow LED : GreenLED : Green LED flashes : Red LED Power On Monitor Alarm Relay Mode Power Off Normal Alarm NO Contact NORMALLY DE-ENERGIZED Open Open Close NO Contact NORMALLY ENERGIZED Open Close Open NC Contact NORMALLY DE-ENERGIZED Close Close Open NC Contact NORMALLY ENERGIZED Close Open Close E2.3

12 2. SPECIFICATIONS AND OUTLINE DRAWINGS M-15B1 2-2 OUTLINE DRAWINGS 4 - ø 5.5 (0.217) 1 M - 15 MONITORING SYSTEM 2 4 IR (10.00) 238(9.37) STATUS DISPLAY SCAN FIX AUTO ECK SHIFT MONITOR Ref. (0.47 Ref.) 6 114(4.49) 190(7.48) 3 (0.118) Ref (7.70 Ref.) 26 Ref. (1.02 Ref.) 8 9 NO. NAME DESCRIPTION 1 7 Segment Digital Indicator Ref. (8.60 Ref.) A1 IN - 1 B 1 A 2 IN - 3 B2 A 5 IN - 2 B5 A 6 IN - 4 B6 A 9 REC1 B9 A 10 REC3 B10 A 11 REC2 B11 A 12 REC4 B12 A 13 B13 A 14 RES B14 NC NO NC NO NC OK NO L1 L2 GND AC 50/60Hz 2 Unit Indicator Red LED 3. Lamp Yellow LED 4 Status Indicator 2 Color LED (Red/ Green) 5 Display Indicator Yellow LED 6 Key Operation Switch 7 Monitor Output Terminal BNC 8 Input/Output Terminal Block Screw: M3 Pitch between terminals: 7.62 mm 9 Power Supply Input, Alarm Output Terminal Block Screw: M3 Pitch between terminals: 7.62 mm 10 Serial No. Plate Ref. (6.59 Ref.) 10 Dimension : mm(inches) Note 1) The chart below refers to the Input/Output Terminal Block ( 8 in the picture above) when each channel is used. in use Input Terminal Block IN-1(A1),(A2) IN-2(A5),(A6) IN-3(B1),(B2) IN-4(B5),(B6) Output Terminal Block REC1(A9),(A10) REC2(A11),(A12) REC3(B9),(B10) REC4(B11),(A12) Note 2) Do not use any open terminals on the Power Supply input or Alarm Output Terminal Block ( 9 in the picture above) that are not specified. -4-

13 3. WIRING METHODS M-15B1 Prohibited Mandatory Do not use the blank terminals. 3-1 SIGNAL TRANSMISSION CABLE WIRING Use the exclusive cable for connection with the CA Series Acceleration Transducer. Use a twisted pair shielded cable with AWG No. 16 to 20 gage ( mm 2 ) stranded conductors for wiring from the junction terminal and monitoring device. Select suitable cable for operating environments. The cable length between CA Series Acceleration Transducer and Monitor should not exceed 300 meters (intrinsic safety expolosion-proof construction : 200m). Excessive length may limit the system s frequency response. Keep cabling away from power and control lines, to prevent noise interference. Cable should be securely fixed. Prevent the cable from bending sharply, twisting, kinking, knotting, and straining. The recommended minimum bend radius is 80mm. SHINKAWA recommends that the cable should be routed, through conduit to prevent physical damage and noise interference. -5-

14 3. WIRING METHODS M-15B1 3-2 INTRINSICAL SAFETY CONSTRUCTION WIRING Terminal No. Terminal Name Signal Name Terminal No. Terminal Name Signal Name A1 IN-1 INPUT SIGNAL B1 IN-3 INPUT SIGNAL A2 MON, SHIELD B2 MON, SHIELD 1 3 A3 - - B3 - - A4 - - B4 - - A5 IN-2 INPUT SIGNAL B5 IN-4 INPUT SIGNAL A6 MON, SHIELD B6 MON, SHIELD 2 4 A7 - - B7 - - A8 - - B8 - - A9 REC1 RECORDER OUTPUT (+) B9 REC3 RECORDER OUTPUT (+) 1 3 A10 RECORDER OUTPUT (-) B10 RECORDER OUTPUT (-) A11 REC2 RECORDER OUTPUT (+) B11 REC4 RECORDER OUTPUT (+) 2 4 A12 RECORDER OUTPUT (-) B12 RECORDER OUTPUT (-) A13.CONTACT INPUT B13 RES RE CONTACT INPUT A14.CONTACT INPUT B14 RE CONTACT INPUT Terminal block wiring screw: M3 A 6 6 Shield Earth busbar or Earth terminal C B A 3 MTL 728+ Safety barrier 1 C A B Shield 5 6 Shield 5 6 B A C 4 C B Shield 5 Shield 5 B A C Shield C B A Safe area Hazardous area MT Barrier enclosure M-15 Monitor Terminal Block A Shield B C Junction box T-20E Junction Box Shield terminals must no be connected to ground at junction box. Signal, Power supply (White) 0(Black) Shield CA-301 Acceleration Transducer Machine case 2 Monitor connecting terminal Signal name Wire no. Terminal name Wire no. Signal A IN- A Common B,C B,C Structure connected to supply source ground. -6-

15 3. WIRING METHODS M-15B1 Mandatory Regarding wiring work, follow the warnings below. Warning No. 1 2 Description Use a twisted pair shielded cable with AWG No. 16 to 20 gage ( mm 2 ) stranded conductors for wiring from the junction terminal. Ex ia IIB T3 : Cw 0.1µF, Lw 1.5mH Ex ia IIB T4 : Cw 0.1µF, Lw 1.5mH Ex ia IIC T3 : Cw 0.012µF, Lw 2.0mH Keep wiring away from power and control lines to prevent noise interference. Transducer mounted directly on a conductive structure which is connected to supply source ground. 3 The safety barrier should be installed in the Barrier Enclosure Perform First Class Grounding Construction (The resistance to Ground 10Ω or less, Grounding should be independent.) Connecting part should be as short as possible. (Shield effect) Connect the shield wire to the common. If this wiring is used, according to the resistance from the Safety Barrier, the monitor input (voltage) decreases approx. 1%. Danger To prevent electrocution, only perform wiring work after turning off the Power Supply. -7-

16 3. WIRING METHODS M-15B1 3-3 NON-INTRINSICAL SAFETY CONSTRUCTION WIRING Terminal No. Terminal Name Signal Name Terminal No. Terminal Name Signal Name A1 IN-1 INPUT SIGNAL B1 IN-3 INPUT SIGNAL A2 MON, SHIELD B2 MON, SHIELD 1 3 A3 - - B3 - - A4 - - B4 - - A5 IN-2 INPUT SIGNAL B5 IN-4 INPUT SIGNAL A6 MON, SHIELD B6 MON, SHIELD 2 4 A7 - - B7 - - A8 - - B8 - - A9 REC1 RECORDER OUTPUT (+) B9 REC3 RECORDER OUTPUT (+) 1 3 A10 RECORDER OUTPUT (-) B10 RECORDER OUTPUT (-) A11 REC2 RECORDER OUTPUT (+) B11 REC4 RECORDER OUTPUT (+) 2 4 A12 RECORDER OUTPUT (-) B12 RECORDER OUTPUT (-) A13.CONTACT INPUT B13 RES RE CONTACT INPUT A14.CONTACT INPUT B14 RE CONTACT INPUT Terminal block wiring screw: M3 A 1 C B Shield 3 Shield 3 B A C A C B Shield 3 Shield 3 B A C M-15 Monitor Terminal Block A Shield B C Junction box T-20E Junction Box Shield terminals must no be connected to ground at junction box. Signal, Power supply (White) 0(Black) Shield CA-301 Acceleration Transducer Machine case 2 Monitor connecting terminal Signal name Wire no. Terminal name Wire no. Signal A IN- A Common B,C B,C Structure connected to supply source ground. -8-

17 3. WIRING METHODS M-15B1 Mandatory Regarding wiring work, follow the warnings below. Warning No Description Use a twisted pair shielded cable with AWG No. 16 to 20 gage ( mm 2 ) stranded conductors for wiring from the junction terminal. (Max. length: 300m) Keep wiring away from power and control lines to prevent noise interference. Transducer mounted directly on a conductive structure which is connected to supply source ground. Connecting part should be as short as possible. (Shield effect) Connect the shield wire to the common. Danger To prevent electrocution, only perform wiring work after turning off the Power Supply. -9-

18 4. FUNCTION DESCRIPTIONS M-15B1 4-1 INDICATION FUNCTION This function selects the value from the Measured Displacement, Alarm Setup and Sensor Set Gap oltage, etc., and shows the value on the LED display. It is also possible to select from different kinds of adjustment contents and indicate on the LED display by switch operation. 4-2 ALARM FUNCTION This function forces the ALERT Alarm, GER Alarm and OK (input abnormality) Alarm to contact-output and display on the LED display. Alarm Setup and Indication are performed individually on each channel. All channels of each contact-output simultaneously (OR) output. The limit value specified for GER alarm must be higher than that for ALERT alarm GER ALARM When the measured value exceeds the GER Alarm setup value, this function outputs the GER Alarm contact and the red STATUS lamp is lit. Regarding the GER Alarm setup, refer to Section 6 Screen Description and Key Operation Method ALERT ALARM When the measured value exceeds the ALERT Alarm setup value, this function outputs the ALERT Alarm contact and the red STATUS lamp flashes. Regarding the ALERT Alarm setup, refer to Section 6 Screen Description and Key Operation Method OK ALARM When the input signal level from the Transducer is abnormal, the OK Alarm contact outputs and the green STATUS lamp flashes. If the input signal level abnormality from the Transducer is detected, the OK Alarm, according to the OFF DELAY TIMER, operates for about 15 seconds, even after the input signal level is back to the normal level. Detectable Transducer Abnormal Input Type CA Transducer Detectable Abnormality Condition Cables between the Transducer and Monitor short-circuit or disconnect While OK Alarm Output is in progress, GER/ALERT Alarm and Recorder Warning Output are disconnected. -10-

19 4. FUNCTION DESCRIPTIONS M-15B1 4-3 ALARM DELAY FUNCTION Avoid incorrect operation of GER Alarm and ALERT Alarm because of noise, etc. It is possible to provide a delay time for each channel. Regarding the Alarm Delay Time setup, refer to Section 6 Screen Description and Key Operation Method. 4-4 UENCE FUNCTION Prevent alarm actuation that is caused by excessive vibration during machine startups and stops. Disconnect the GER/ALERT ALARM, or switch the alarm setup to another number magnified by the specified number, according to the. contact input (from the power supply input and alarm output terminal block) during the specified time. The. lamp is lit. Warning This function may leave the machine unprotected. While the. magnification value is set to -, machine is unprotected as alarms are disabled. While the. magnification is set to a value between 2 to 5, and. function is in effect., alarm setpoints are multiplied by the set value. When alarm setpoints exeed 150% of the monitor range, alarm may not output. It is necessary to keep short circuit of the terminals for more than 0.5 seconds to operate these functions. 4-5 SUPPRESSION FUNCTION When the condition ibration value < Measured range Suppression setup value is met, this function disconnects the Measured ibration alue (M) Display and Recorder Output. Regarding the Suppression function setup, refer to Section 6 Screen Description and Key Operation Method. 4-6 POWER-ON TIMER MODE This function is performed for about 5 seconds after the power is turned on. While the function is in progress, 7-segment LED display flashes and all alarm functions are disconnected. Warning While the Power-On TIMER is in progress, measured value and recorder output are disconnected at the same time. 4-7 ALARM RELAY MODE On each alarm contact, it is possible to select from: Normally De-energized or Normally Energized. Regarding the Alarm Relay Mode setup, refer to Section 6 Screen Description and Key Operation Method. Power Off Power On Normal Alarm NO Contact NORMALLY DE-ENERGIZED Open Open Close NO Contact NORMALLY ENERGIZED Open Close Open NC Contact NORMALLY DE-ENERGIZED Close Close Open NC Contact NORMALLY ENERGIZED Close Open Close -11-

20 4. FUNCTION DESCRIPTIONS M-15B1 4-8 ALARM RE METHOD If the input value is released from the alarm level, each alarm contact output and indication can be canceled automatically by using AUTO-RE. The alarm according to the key operations key + SHIFT key, or the alarm situation, can be held until it is short-circuited between the RES terminal and the terminal on the terminal block on rear panel, by using SELF-HOLD. Regarding the Alarm Reset Method, refer to Section 6 Screen Description and Key Operation Method. Warning If input reaches the alarm level, the alarm cannot be reset. It is necessary to keep short circuit of the terminals for more than 0.5 second to operate these functions. 4-9 MONITOR OUTPUT For signal analysis, this function outputs the input signal, via a buffer amplifier, to the BNC connector on the front. Warning The monitor output signal is not insulated from the input RECORDER OUTPUT The Analog signal that is proportional to the measured value is output from the terminal block on the rear panel of the Instrument Rack SELF-DIAGNOSTIC FUNCTION This function detects Sensor disconnect, or a poor connection of a connector, which is caused by the Input System. When the abnormality is detected, the OK Alarm contact outputs and the OK indicator flashes LAMP TEST Press ECK for 5 seconds or more. The 7-segment digital indicator and other indication lamps will blink for 3 seconds for check. Do not operate this function while machine is running to avoid wrong operation. Prohibited -12-

21 5. PART NAMES AND DESCRIPTIONS M-15B1 5-1 PART NAMES ON THE FRONT PANEL Unit of oscillating measured value is indicated as an example of representation. 1 M-15 MONITORING SYSTEM 2 IR STATUS MONITOR ECK 2 DISPLAY 3 SCAN FIX AUTO SHIFT No. Name 1 7 Segment Digital Indicator 2 Unit Indicator 3. Lamp 4 Status Indicator 5 Display Indicator 6 Key Operation Switch 7 Monitor Output Terminal -13-

22 5. PART NAMES AND DESCRIPTIONS M-15B1 5-2 PART NAMES ON THE REAR PANEL 9 8 A1 IN-1 B1 A2 IN-3 B2 A5 IN-2 B5 A6 IN-4 B6 A9 REC1 B9 A10 REC3 B10 A11 REC2 B11 A12 REC4 B12 A13 B13 A14 RES B14 NC NO NC NO NC OK NO L1 L2 GND AC 50/60Hz 10 No. Name 8 Input/Output Terminal Block Power Supply Input, Alarm 9 Output Terminal Block 10 Ser. No. Plate -14-

23 5. PART NAMES AND DESCRIPTIONS M-15B1 5-3 PART NAMES AND DESCRIPTIONS 1 7-SEGMENT DIGITAL DISPLAY Indicates the channels, measured value, alarm setup value, sensor set gap voltage value and other specification setups. 2 UNIT INDICATOR Indicates the unit for the digitally displayed values, such as measured values, alarm setup values, etc. Warning 3 UENCE LAMP While the. circuit is in progress, this lamp is lit. (Regarding the Alarm Operation during the Sequence circuit progression, refer to Section 6 Screen Description and Key Operation Method. ) 4 STATUS INDICATOR Displays the condition of the monitor by a constantly lit or flashing green or red LED. When the condition is normal, the green LED is lit. Status Indicator Lamp Red (lit) Red (flashing) Green (lit) Green (flashing) Off Condition GER Alarm ALERT Alarm Normal Input Abnormality (Not-OK) Power Off or Channel Off 5 DISPLAY INDICATOR The selected indication mode LED is on. (Regarding how to change the Display Indicator, refer to Section 6 Screen Description and Key Operation Method. ) 6 KEY OPERATION SWIT The keys to switch, select or setup the channel, measured value and alarm setup value screens. 7 MONITOR OUTPUT CONNECTOR The BNC Connector used to signal-output to the oscilloscope waveform observation and analytic instruments. The input signal of each channel is output via the buffer amplifier. Mandatory Monitor output should be insulated from the ground. If this does not occur, noise may be generated in the signal. Warning Do not remove this unit while the power is on. Serious damage may result. -15-

24 5. PART NAMES AND DESCRIPTIONS M-15B1 8 INPUT/OUTPUT TERMINAL BLOCK Contact signal connecting the terminal block for the Transducer, Recorder output and Operation. Transducer Input Connects the Transducer designated in the specifications. On the terminal block, the power supply for the Transducer drive outputs. Recorder Output Outputs the current output (4 to 20mADC) or voltage output (4 to 20mADC) response to the measured range. Contact for Operation RES. (Reset Input) : Resets the SELF-HOLD Alarm according to a short-circuit between the terminals. If the situation is in the alarm, the alarm cannot be reset. (For details, refer to Section 4 Function Description and Section 6 Screen Description and Key Operation Method. ).(Sequence Input) : The Sequence circuit operates according to a short-circuit between the terminals. Depending upon the Sequence setup, the machine may not be protected. (For details, refer to Section 4 Function Description and Section 6 Screen Description and Key Operation Method. ) 9 Power Supply, Alarm Output Terminal Block The terminal block of the power supply input for the unit drive. Power Supply Input Provides power to the unit. Alarm Contact Output (C Contact Output) : : GER Alarms on all channels simultaneously (OR) output. : ALERT Alarms on all channels simultaneously (OR) output. OK : OK Alarms on all channels simultaneously (OR) output. 10 Ser. No. Plate Plate on which the Serial Number is written. -16-

25 6. SCREEN DESCRIPTIONS AND KEY OPERATION METHOD M-15B1 Unit of oscillating measured value is indicated as an example of representation. 6-1 DIGITAL INDICATOR DESCRIPTION 1 3 IB 2 1 ANNEL DISPLAY ( Dis.) Displays the number of the channel. While the value is being collected, the display is constantly on. setup can be changed, the display flashes. When the specification 2 MEASURED ALUE/SPECIFICATION UP DISPLAY (MEAS Dis.) Displays the measured value and specification setup condition. When the specification setup change is available, the display flashes. 3 UNIT INDICATOR The unit lamp for the digitally displayed values, such as measured values, alarm setup values, etc, is on. -17-

26 6. SCREEN DESCRIPTIONS AND KEY OPERATION METHOD M-15B1 6-2 SWIT INDICATION ANNELS OF MEASURED IBRATON ALUE (M) It is possible to change the display channel of the M by using the key or key, if desired. It is also possible to change other displays, such as voltage value and GER/ALERT Alarms by using the same process. The chart below is the procedure to switch the indication channel of the M. Step Key Operation Items Display 1 Press once, 2 appears. Press once, 4 appears. Default display E.g., Channel 1 is isplayed. Dis. is 1. Dis. Is 2. IB 2 Press once, 3 appears. Press once, 1 appears. 2 M is displayed. IB Dis. Is 3. 3 Press once, 4 appears. Press once, 2 appears. 3 M is displayed. IB Dis. Is 4. 4 Press once, 1 appears. Press once, 3 appears. 4 M is displayed. IB -18-

27 6. SCREEN DESCRIPTIONS AND KEY OPERATION METHOD M-15B1 6-3 SWIT INDICATION (TO ALUE, ALARM UP ALUE INDICATION, OR MEASURED ALUE INDICATION UP) Explains how to switch to the voltage value, GER Alarm setup value or ALERT Alarm setup value from the M screen, and how to select the M indication setup SWIT INDICATION PROCEDURE In order to switch the indication, use the key or SHIFT + keys. On each screen, it is possible to change the indicated channels by using the and keys. On the M indication setup screen, by using the and keys, the indication style can be changed. For details about the measured vibration indication setup screen, refer to Section Changes in Dis. (Sec. 6-2) M SCREEN OLTAGE ALUE SCREEN Dis. Change GER ALARM UP ALUE SCREEN Dis. Change ALERT ALARM UP ALUE SCREEN Dis. Change M INDICATION UP SCREEN ANGE IND. UP : KEY KEYS : SHIFT + KEY -19-

28 6. SCREEN DESCRIPTIONS AND KEY OPERATION METHOD M-15B SWIT INDICATION OPERATION Use the or SHIFT + keys to change the contents of the MEAS Dis. The chart below is an example of the procedure to change the MEAS Dis. on one channel. Regarding how to change the Channel, refer to Section 6-2. Step Key Operation Items Display 1 2 Press ; GAP voltage value in step 2 appears. Press SHIFT and simultaneously; M indication setup in step 5 appears. Press ; GER alarm setup value in step 3 appears. Press SHIFT and simultaneously; M in step 1 appears. Default M is on MEAS Dis. voltage value is on MEAS Dis. IB unit lamp is on. M is on MEAS Dis. Picture below shows 1 M is 11. IB unit lamp is on. voltage value is on MEAS Dis. Picture below shows 1 voltage value is IB 3 Press ; ALERT alarm setup value in step 4 appears. Press SHIFT and simultaneously; GAP voltage value in step 2 appears. GER alarm setup value is on MEAS Dis. unit lamp is on. GER alarm setup value is on MEAS Dis. Picture below shows 1 setup value is 16. IB 4 Press ; M indication setup screen in step 5 appears. Press SHIFT and simultaneously; GER alarm setup value in step 3 appears. ALERT alarm setup value is on MEAS Dis. Proceed to next step unit lamp is on. ALERT alarm setup value is on MEAS Dis. screen. Picture below indicates 1 setup value is 12. IB -20-

29 6. SCREEN DESCRIPTIONS AND KEY OPERATION METHOD M-15B1 Continued from previous step Step Key Operation Items Display 5 Press ; M screen in step 1 appears. Press SHIFT + ; ALERT alarm setup value in step 4 appears. M indication setup is on MEAS Dis. Refer to Section IB unit lamp is on and M is on MEAS Dis. Selected indication setup lamp flashes. IB If the situation remains unchanged for one minute, the M screen will automatically return. -21-

30 6. SCREEN DESCRIPTIONS AND KEY OPERATION METHOD M-15B HOW TO ANGE THE MEASURED IBRATION ALUE (M) INDICATION In order to indicate the M, it is possible to select from the following: SCAN : Approx. every 2 seconds, the displayed channel changes. The channels change in the following order: FIX : The specified channel is displayed continuously. AUTO : The maximum value of the channel being collected is displayed. The chart below shows how to setup the M indication. Step Key Operation Items Display IB unit lamp is on. M is on MEAS Dis. 1-A Press or and SHIFT simultaneously, and M indication setup screen appears. Default display IB 1-B SCAN setting When setup screen appears, currently selected setting on DISPLAY Indicator (DISPLAY) flashes. Picture below shows SCAN is selected. DISPLAY SCAN FIX AUTO 2 Press once in step 1, and FIX lamp flashes. Press again, and lamp moves in this order: FIX AUTO SCAN. Press, and lamp moves backward FIX setting Currently selected setting on DISPLAY flashes. Picture below shows FIX is selected. DISPLAY SCAN FIX AUTO Flashing lamp stays on when setting confirmation is complete. Picture below shows FIX is confirmed. 3 Press or to confirm setting. FIX confirmed DISPLAY SCAN FIX AUTO -22-

31 6. SCREEN DESCRIPTIONS AND KEY OPERATION METHOD M-15B1 6-4 ALARM FUNCTION UP Prohibited Do not setup any specifications while the unit is in operation. Alarm contact might cause output. The limit value specified for GER alarm must be higher than that for ALERT alarm ALARM FUNCTION UP PROCEDURE After the protection is cancelled, it is possible to select one alarm function by using the key or SHIFT + key. In order to change the channels, use the or key. The chart below is the Alarm function setup procedure. Step Key Operation Items Display 1 Press continuously for more than 5 sec. Protection cancellation Picture below is shown for 3 sec. as Alarm function setup mode. This is GER alarm (GER) value setup screen. IB 2 1 GER value setup screen After picture in step 1 is shown for 3 sec., 1 GER setup value appears. IB : KEY KEYS : SHIFT + KEY GER ALUE UP SCREEN Dis. Change ALERT ALUE UP SCREEN Dis. Change Proceed to next step -23-

32 6. SCREEN DESCRIPTIONS AND KEY OPERATION METHOD M-15B1 Continued from previous step GER DELAY TIME UP SCREEN Dis. Change ALERT DELAY TIME UP SCREEN Dis. Change GER RE UP SCREEN Dis. Change ALERT RE UP SCREEN Dis. Change OK RE UP SCREEN Dis. Change GER RELAY MODE UP SCREEN Dis. Change ALERT RELAY MODE UP SCREEN Dis. Change OK RELAY MODE UP SCREEN Dis. Change GER UENCE MAGNIFICATION NUMBER UP SCREEN Dis. Change ALERT UENCE MAGNIFICATION NUMBER UP SCREEN Dis. Change On each setup screen press the key continuously for more than 5 sec., and the M screen will return. If the situation remains unchanged for approx. 1 minute, the M screen will automatically return. -24-

33 6. SCREEN DESCRIPTIONS AND KEY OPERATION METHOD M-15B GER/ALERT ALARM UP After the procedure in Section is completed, in order to cancel the protection, each alarm setup value setting screen appears. To change the setup value, follow the procedure below. Step Key Operation Items Display Refer to Section and go to the alarm function setup mode. Picture below shows ALERT value setup screen is selected. 1 Use to select GER or ALERT value setup screen. Use or to select the channel. Select alarm value setup screen Select. IB After picture above is shown for 3 sec., 1 ALERT setup value appears. IB 2 Press, and far right number of alarm setup value flashes. Setup value can be changed. 3 Use and to change number. Use SHIFT + or SHIFT + to change position. 4 ECK Press ECK and simultaneously to confirm setup. 5 Press continuously for 5 sec., and M appears. Setup value change available Change setup value Confirm setup value Return to default mode Picture below shows 2 alarm setup value is 15. When setup is confirmed, number stops flashing. IB Return to M display. Picture below shows 1 12 is being measured. IB -25-

34 6. SCREEN DESCRIPTIONS AND KEY OPERATION METHOD M-15B ALARM DELAY TIME UP After the procedure in Section is completed, in order to cancel the protection, each alarm delay time setup screen appears. To change the setup value, follow the procedure below. Step Key Operation Items Display 1 Refer to Section and go to the alarm function setup mode. Use to select GER or ALERT delay time setup screen. Use or to select channel. Select alarm delay time setup screen Select. Picture below shows ALERT delay time setup screen is selected. If GER delay time setup, d T is on MEAS dis. IB After picture above is shown for 3 sec., 2 ALERT alarm delay time appears. IB 2 Press key, and setup value flashes. Setup value can be changed. 3 Use and keys to change setup value. Select from 0.5, 1, 2 or 3 sec. 4 ECK Press ECK and keys simultaneously to confirm setup. 5 Press continuously for 5 sec., and M appears. Setup value change available Change setup value Confirm setup value Return to default mode Picture below shows 2 alarm delay time is set at 3 sec. When setup is confirmed, number stops flashing. IB Return to M display. Picture below shows 1 12 is being measured. IB -26-

35 6. SCREEN DESCRIPTIONS AND KEY OPERATION METHOD M-15B ALARM RE MODE UP After the procedure in Section is completed, in order to cancel the protection, each alarm reset setup screen appears. To change the setup, follow the procedure below. Step Key Operation Items Display Refer to Section and go to the alarm function setup mode. 1 Use to select GER, ALERT or OK reset mode setup mode screen. There is no select. (All together) Select alarm reset mode setup screen Picture below shows GER reset mode setup screen is selected. If ALERT or OK alarm setup, AL or ol is on display. IB After picture above is shown for 3 sec., GER setup appears. IB 2 Press, and setup flashes. Setup can be changed. Alarm reset mode change available 3 Use and to change the setup. Change setup Picture below shows AUTO-RE is selected. If SELF-HOLD is selected, L is on display. IB 4 ECK Press ECK and simultaneously to confirm setup. 5 Press continuously for 5 sec., and M appears. Confirm setup Return to default mode When setup is confirmed, indication stops flashing. Return to M display. Picture below shows 1 12 is being measured. IB -27-

36 6. SCREEN DESCRIPTIONS AND KEY OPERATION METHOD M-15B ALARM RE If the alarm reset mode is setup as SELF-HOLD, the alarm will output until the reset signal inputs from outside. There are two ways to reset the alarm, by the front panel key operation, and by short-circuiting between the RES terminal and the terminal on the terminal block on the rear panel. Warning Warning If the input reaches the alarm level, the alarm cannot be reset. The reset operation affects all alarms. All alarms are reset. 1 FRONT PANEL KEY OPERATION Step Key Operation Items Display 1 SHIFT Press SHIFT and simultaneously for more than 5 sec. Alarm reset operation Press keys for more than 5 sec., and alarm will reset and STATUS indication will be changed. 2 INPUT CONTACT SIGNAL FROM OUTSIDE According to the short-circuit between RES and on the Input/Output terminal block on the rear panel, the alarm reset will operate. External Reset Contact Input A13 A14 B13 RES B14 It is necessary to keep short circuit of the terminals for more than 0.5 seconds to operate these functions. -28-

37 6. SCREEN DESCRIPTIONS AND KEY OPERATION METHOD M-15B RELAY MODE (NORMALLY DE-ENERGIZED, NORMALLY ENERGIZED) After the procedure in Section is completed, in order to cancel the protection, alarm relay mode screen appears. To change the setup, follow the procedure below. Step Key Operation Items Display Refer to Section and go to the alarm function setup mode. Picture below shows ALERT alarm relay mode setup screen is selected. If GER or OK setup, de or oe is on display. 1 Use to select GER, ALERT or OK relay mode setup screen. There is no select. (All together) Select alarm relay mode setup screen IB After picture above is shown for 3 sec., ALERT setup appears. IB 2 Press, and setup flashes. Setup can be changed. Alarm relay mode change available Picture below shows NORMALLY DE-ENERGIZIED (ENERGIZED ALARM) is selected. 3 Use and to change setup. Change setup Picture below shows NORMALLY ENERGIZED (DE-ENERGIZED ALARM) is selected. IB IB 4 ECK Press ECK and simultaneously to confirm setup. 5 Press continuously for 5 sec., and M appears. Confirm setup Return to default mode When setup is confirmed, setup stops flashing. Return to M display. Picture below shows 1 12 is being measured. IB -29-

38 6. SCREEN DESCRIPTIONS AND KEY OPERATION METHOD M-15B UENCE SIGNAL INPUT 1 UENCE FUNCTION While the Sequence signal inputs, according to the Sequence magnification number in Section 6-4-8, the alarm setup values of GER and ALERT increase. The Sequence magnification number can be setup from 1 to 5 in 1 magnification increments. It is possible to force to cut off the alarm when the Sequence signal inputs. Warning This function may leave the machine unprotected. While the. magnification value is set to -, machine is unprotected as alarms are disabled. While the. magnification is set to a value between 2 to 5, and. function is in effect., alarm setpoints are multiplied by the set value. When alarm setpoints exeed 150% of the monitor range, alarm may not output. 2 UENCE SIGNAL INPUT When there is a short-circuit between and on the INPUT/OUTPUT terminal block on the rear panel, the Sequence function operates. UENCE CIRCUIT CONTACT INPUT A13 A14 B13 RES B14 It is necessary to keep short circuit of the terminals for more than 0.5 seconds to operate these functions. -30-

39 6. SCREEN DESCRIPTIONS AND KEY OPERATION METHOD M-15B UENCE MAGNIFICATION NUMBER UP The Sequence magnification for GER alarm must be higher than that ALERT alarm. After the procedure in Section is completed, in order to cancel the protection, each alarm Sequence magnification number setup screen appears. To change the setup, follow the procedure below. Step Key Operation Items Display 1 Refer to Section and go to the alarm function setup mode. Use to select GER or ALERT Sequence magnification number setup screen. Use or to select channel. Select Sequence magnification number setup screen Select. Picture below shows GER sequence magnification number setup screen is selected. If ALERT setup, AS is on display. IB After picture above is shown for 3 sec., 1 GER sequence magnification number appears. Picture below shows setup is at 4. IB 2 Press key, and setup value flashes. Setup can be changed. 3 Use and keys to change setup value. Magnification number from 1 to 5 or Alarm cut off can be selected. ECK 4 Press ECK and simultaneously to confirm setup. 5 Press continuously for 5 sec., and M appears. Setup value change available Change setup value Confirm setup value Return to default mode Picture below shows Alarm cut off is selected. If a number from 1 to 5 is selected, the number is on display. When setup is confirmed, number stops flashing. IB Return to M display. Picture below shows 1 12 is being measured. IB -31-

40 6. SCREEN DESCRIPTIONS AND KEY OPERATION METHOD M-15B1 6-5 BASIC FUNCTION UP Prohibited Do not setup any specifications while the unit is in operation. Alarm contact might cause output BASIC FUNCTION UP PROCEDURE According to the protection cancellation operation, the basic function setup appears. Use the or SHIFT + keys to select the setup items. To change the setup channel, use the and keys. For the basic function setup, follow the procedure below. Step Key Operation Items Display 1 Press and simultaneously for more than 5 sec. Protection cancellation Picture below is shown for 3 sec., after basic function setup mode appears. IB 2 Measured range decimal point setup screen After picture above is shown for 3 sec., picture below appears. IB : KEY KEYS : SHIFT + KEY MEASURED RANGE DECIMAL POINT UP SCREEN MEASURED RANGE 100% UP SCREEN MEASURED RANGE 20% UP SCREEN RECORDER OUTPUT GAIN UP SCREEN Dis. Change Dis. Change Dis. Change Dis. Change Proceed to next step -32-

41 6. SCREEN DESCRIPTIONS AND KEY OPERATION METHOD M-15B1 Continued from previous step RECORDER OUTPUT 0% UP SCREEN RECORDER OUTPUT 100% ADJUSTMENT SCREEN SUPRESSION UP SCREEN. BYPASS UP SCREEN Dis. Change Dis. Change Dis. Change Dis. Change On each setup screen, press the key continuously for more than 5 sec., and the M screen will return. If the situation remains unchanged for approx. 1 minute, the M screen will return. -33-

42 6. SCREEN DESCRIPTIONS AND KEY OPERATION METHOD M-15B MEASURED RANGE UP BASIC OPERATION DESCRIPTION Mandatory Danger Mandatory Warning The alarm operates. During adjustment, various alarms may output. If necessary, consider alternate protection plans, such as disconnecting the circuit between the alarm contact output and the connecting system, etc. During adjustment, the machine is not protected. After the measured range setup change, it is necessary to adjust the recorder output written in Section From the acceleration range to the velocity range, it cannot modify to the acceleration range from the velocity range. 1 WIRING METHOD According to the wiring method in Section 8-6-1, connect the Transducer input terminal block of the channel, which is to be adjusted to the D.C. voltage generator and the function generator. Connect the recorder output terminal block of the channel to DMM. 2 STANDARD SINE WAE INPUT OLTAGE CALCULATION Use the formula below and the measured range, which is to be setup to figure out the input voltage value. Measured range: Acceleration ( ) INPUT OLTAGE E(mrms) = (MEASURED RANGE TRANSDUCER SENSITIITY) / 2 E.g., measured range is 50 INPUT OLTAGE E(mrms) = ( m/ ) / 2 = 360.8mrms Measured range: elocity (mm/s) INPUT OLTAGE E(mrms) = (6.28 MEASURED RANGE CALIBRATION FREQUENCY m/ ) / 2 E.g., measured range is 50mm/s, calibration frequency is 100Hz. INPUT OLTAGE E(mrms) = ( mm/s / Hz m/ ) / 2 = 226.6mrms The standard sine wave input voltage setup at the time of shipment is written on the serial no. plate on the back of the unit. -34-

VM-15 SERIES MONITOR. Model VM-15U1 VELOCITY MONITOR INSTRUCTION MANUAL

VM-15 SERIES MONITOR. Model VM-15U1 VELOCITY MONITOR INSTRUCTION MANUAL MANUAL No. 20203E2.4 Published in Jul.2002 Revised in Mar.2012 M-15 SERIES MONITOR Model M-15U1 ELOCITY MONITOR INSTRUCTION MANUAL FOREWORD M-15U1 FOR THE SAFE OPERATION OF THIS UNIT Thank you for using

More information

VM-702B ABSOLUTE VIBRATION MONITOR MODULE Page 1 of 5

VM-702B ABSOLUTE VIBRATION MONITOR MODULE Page 1 of 5 Page 1 of 5 No entry if additional Model Code / Additional Spec. Code ( spec. code is not specified. ) VM-702B /ALY /NB1 /CS1 /CS2 /TRP /TB Analysis Function Non-incendive * 1 Monitor Function Analysis

More information

VM-5 SERIES MONITOR INSTRUCTION MANUAL

VM-5 SERIES MONITOR INSTRUCTION MANUAL MANUAL. 29507E4.0 Published in Jun.1996 Revised in Aug.2009 VM-5 SERIES MONITOR MODEL VM-5T DUAL THRUST MONITOR INSTRUCTION MANUAL SAFETY CONVENTIONS VM-5T For safe operation... Thank you for using our

More information

VM-5 SERIES MONITOR MODEL VM-55 VIBRATION MONITOR INSTRUCTION MANUAL

VM-5 SERIES MONITOR MODEL VM-55 VIBRATION MONITOR INSTRUCTION MANUAL MANUAL.20501E2.1 Published in Mar.2005 Revised in Jul.2010 VM-5 SERIES MONITOR MODEL VM-55 VIBRATION MONITOR INSTRUCTION MANUAL SAFETY CONVENTIONS VM-55 For safe operation... Thank you for using our VM-5

More information

VM-5 SERIES MONITOR. MODEL VM-5P3 Phase Marker Unit INSTRUCTION MANUAL

VM-5 SERIES MONITOR. MODEL VM-5P3 Phase Marker Unit INSTRUCTION MANUAL MANUAL NO. 20002E2.0 Published in Dec.2000 Revised in Nov. 2012 VM-5 SERIES MONITOR MODEL VM-5P3 Phase Marker Unit INSTRUCTION MANUAL SAFETY CONVENTIONS VM-5P3 For safe operation... Thank you for using

More information

VM-704B & VM-704 Temperature Monitor Module Manual

VM-704B & VM-704 Temperature Monitor Module Manual MANUAL No. 21105E1.0 Published in Nov. 2011 VM-7 Monitoring System VM-704B & VM-704 Temperature Monitor Module Manual This manual is solely designed for technicians to set, maintain, check for the VM-7

More information

MODEL MP-1S OVERSPEED UNIT

MODEL MP-1S OVERSPEED UNIT MODEL MP-S OVERSPEED UNIT SPEIFIATIONS Model Spec. ode upon delivery/add l Spec. ode( not requested, leave blank ) MP-S- - - - Hysteresis Set up Monitor Range Input Signal No. of Input Pulse Trigger Mode

More information

Pioneer-R16 Gas Monitor Operator s Manual

Pioneer-R16 Gas Monitor Operator s Manual Pioneer-R16 Gas Monitor Operator s Manual Edition 7/2/97 RKI INSTRUMENTS, INC RKI Instruments, Inc. 33248 Central Ave, Union City, CA 94587 (510) 441-5656 Chapter 1: Description About the Pioneer-R16 Gas

More information

Analog Room Pressure Monitor RPC Series

Analog Room Pressure Monitor RPC Series Description The Room Pressure Monitor is used to measure differential pressure in the range of 0.125 to 1"wc or 30 to 250 Pa. It combines precision high sensitivity silicon sensing capabilities and the

More information

ET10 Series Stand-alone LCD Fan Coil Thermostat

ET10 Series Stand-alone LCD Fan Coil Thermostat ET10 Series Stand-alone LCD Fan Coil Thermostat Features Modern Appearance Stylish Rotary dial and buttons Large LCD with backlight Separate power supply unit with high performance relay outputs Retention

More information

DIGITAL METERS Large Display Temperature Meters Instruction Manual

DIGITAL METERS Large Display Temperature Meters Instruction Manual DIGITAL METERS Large Display Temperature Meters PD755 PD756 PD757 Handles Thermocouple & RTD Inputs with Simplicity J, K, T, E, R, & S Thermocouples 100 Ω Platinum RTD (0.00385 or 0.00392 curve) Large

More information

Advanced Ultraviolet Burner Controller

Advanced Ultraviolet Burner Controller No. CP-SS-7E Advanced Ultraviolet Burner Controller Model AUR300C/AUR30C Overview The AUR300C Advanced Ultraviolet Burner Controller and AUR30C Advanced Ultraviolet Burner Controller with Communication

More information

Beacon 200 Gas Monitor Operator s Manual. Part Number: RK Released: 6/6/08

Beacon 200 Gas Monitor Operator s Manual. Part Number: RK Released: 6/6/08 Beacon 200 Gas Monitor Operator s Manual Part Number: 71-2102RK Released: 6/6/08 Table of Contents Chapter 1: Introduction.................................................3 Overview.............................................................3

More information

Buzzer Unit. TAN-5000 Series. Operating Manual

Buzzer Unit. TAN-5000 Series. Operating Manual PT1E-1040 Buzzer Unit TAN-5000 Series Operating Manual Request for the Customers Read and understand this operating manual before using the buzzer unit. Use the buzzer unit in accordance with the operating

More information

AUR300C, AUR350C. Advanced Ultraviolet Burner Controller

AUR300C, AUR350C. Advanced Ultraviolet Burner Controller No. CP-SS-7E Overview The AUR300C is a burner controller with a dynamic selfchecking function and is used in combination with the AUD300C or AUD00C Advanced Ultraviolet Flame Detector. This AUR300C controls

More information

SC-9 Controller INSTALLATION AND OPERATING INSTRUCTIONS. SC-9 Applications CONTENTS

SC-9 Controller INSTALLATION AND OPERATING INSTRUCTIONS. SC-9 Applications CONTENTS SC-9 Controller MODULAR BOILER CONTROLLER SERIES INSTALLATION AND OPERATING INSTRUCTIONS SC-9 Applications Space heat systems with outdoor reset Constant temperature setpoint control Any of these in combination

More information

S304 and S305 Dust Emission Monitors. User Manual. Distributor

S304 and S305 Dust Emission Monitors. User Manual. Distributor S304 and S305 Dust Emission Monitors User Manual Distributor Version 6.4 30/5/2012 Table of Contents 1. INTRODUCTION... 3 1.1 Safety... 3 1.2 Product overview... 4 1.3 Principle of operation... 4 2. INSTALLATION...

More information

INSTRUCTION MANUAL FOR MICROCOMPUTER BASED TEMPERATURE INDICATING CONTROLLER JCD-13A, JCR-13A

INSTRUCTION MANUAL FOR MICROCOMPUTER BASED TEMPERATURE INDICATING CONTROLLER JCD-13A, JCR-13A INSTRUCTION MANUAL FOR MICROCOMPUTER BASED TEMPERATURE INDICATING CONTROLLER JCD-13A, JCR-13A Preface Thank you for the purchase of our Microcomputer based Temperature Indicating Controllers JCD-13A or

More information

HIn3550 Magnetic Field Monitor User's Manual

HIn3550 Magnetic Field Monitor User's Manual HIn3550 Magnetic Field Monitor User's Manual Copyright 1993 by Holaday Industries, Inc. Manual #600053 10/97 $12.50 Revision Record Manual #600053 HIn3550 Magnetic Field Monitor Revision Description Date

More information

Tri-Stack Smart System

Tri-Stack Smart System Tri-Stack Smart System TM Notes & Warnings - The protection provided by this equipment may be impaired if it is not used in the manner specified herein. - Ensure all wiring meets applicable national and

More information

Micro-controller X C C1 C2 AL1 AL2 AL3. Model: PXR4/5/9. Operation Manual SEL PXR. ECNO:406e

Micro-controller X C C1 C2 AL1 AL2 AL3. Model: PXR4/5/9. Operation Manual SEL PXR. ECNO:406e C C1 C2 AL1 AL2 AL3 Micro-controller X Model: PXR4/5/9 PXR SEL Operation Manual ECNO:46e Table of Contents Part Names and Functions... 6 Operations... 7 2-1 Parameter list... 7 2-2 Basic operations...

More information

Software Version 2.01 LEVEL MONITOR MODEL 220

Software Version 2.01 LEVEL MONITOR MODEL 220 Software Version 2.01 LEVEL MONITOR MODEL 220 19 April 2000 CONTENTS 1. Introduction 1 1.1 Model Number Designation 2 1.2 Intrinsic Safety Considerations 3 2. Specification 4 3. Operation 6 3.1 Display

More information

INTRINSICALLY SAFE S314X AND S315X

INTRINSICALLY SAFE S314X AND S315X INTRINSICALLY SAFE S314X AND S315X Ex SERIES DUST MONITORS Atex models S314X and S315X are not TUV approved versions User Manual Distributor Version 1.3 9.9.2009 Table of Contents 1. INTRODUCTION... 3

More information

Product Manual SZ1009

Product Manual SZ1009 Product Manual SZ1009 Conventional Heating & Cooling Thermostats with Heat Pump Mode Communicating Thermostats Description The SZ1009 is a microprocessor-based mable thermostats designed for conventional

More information

MAINTENANCE & TROUBLESHOOTING GUIDE LEAK ALARM CHANNEL DRY OIL WATER AUX ALARM HIGH LOW CRITICAL WATER TANK LEVEL ALARM MODEL LDE-740 ADVANCE PAPER

MAINTENANCE & TROUBLESHOOTING GUIDE LEAK ALARM CHANNEL DRY OIL WATER AUX ALARM HIGH LOW CRITICAL WATER TANK LEVEL ALARM MODEL LDE-740 ADVANCE PAPER PNEUMERCATOR Liquid Level Control Systems Electronic Systems Excluding LC2000 And TMS Series MAINTENANCE & TROUBLESHOOTING GUIDE 400 300 500 FUEL LEVEL LEAK MONITOR 1 200 600 OIL/GAS NORMAL WATER PNEUMERCATOR

More information

OilSET Installation and Operating Instructions. Oil Separator Alarm Device with SET/DM3AL sensor

OilSET Installation and Operating Instructions. Oil Separator Alarm Device with SET/DM3AL sensor Labkotec UK Ltd Adminicle House 1 Lumb Lane Audenshaw Manchester M34 5WH GREAT BRITAIN Tel: 0844 3350 477 Fax: 0161 4281 179 E-mail: info@labkotec.co.uk 10.8.2012 Internet: www.labkotec.co.uk 1/13 OilSET-1000

More information

SUPERINTEND VRE-08/16/32/64

SUPERINTEND VRE-08/16/32/64 SUPERINTEND VRE-08/16/32/64 Residual Current Monitoring System (RCM) for Grounded Electrical Systems 8/16/32/64 supervision points Operator Manual v2.00 Superintend VRE-08/16/32/64 operator manual v2.00

More information

ETAF Series VDC Fan output LCD Fan Coil Thermostat

ETAF Series VDC Fan output LCD Fan Coil Thermostat ETAF Series 0...10 VDC Fan output LCD Fan Coil Thermostat Features Modern Appearance Stylish rotary dial and buttons Large LCD with backlight 0-10 VDC (2-10 VDC) Fan speed output Retention of temperature

More information

OilSET Installation and Operating Instructions. Oil Separator Alarm Device

OilSET Installation and Operating Instructions. Oil Separator Alarm Device Labkotec Oy Myllyhaantie 6 FI-33960 PIRKKALA FINLAND Tel: +358 29 006 260 Fax: +358 29 006 1260 18.11.2010 Internet: www.labkotec.fi 1/10 OilSET-1000 Oil Separator Alarm Device Copyright 2010 Labkotec

More information

SET Installation and Operating Instructions. Level switch for one sensor

SET Installation and Operating Instructions. Level switch for one sensor Labkotec Oy Myllyhaantie 6 FI-33960 PIRKKALA FINLAND Tel.: +358 29 006 260 Fax: +358 29 006 1260 7.11.2013 Internet: www.labkotec.fi 1/14 SET-1000 Level switch for one sensor Copyright 2013 Labkotec Oy

More information

Series Temperature Controller Instruction Sheet

Series Temperature Controller Instruction Sheet Series Temperature Controller Instruction Sheet Thank you very much for purchasing DELTA A Series. Please read this instruction sheet before using your A series to ensure proper operation and please keep

More information

TEMPERATURE CONTROLLER

TEMPERATURE CONTROLLER TEMPERATURE CONTROLLER E5CX(-H) DIN-sized super-compact (48 x 48-mm) Temperature Controller Featuring Advanced PID Control Advanced PID control with two degrees of freedom improves stability and response

More information

Operating & Maintenance Manual. Alert-4 Ethernet LCD Master Alarm

Operating & Maintenance Manual. Alert-4 Ethernet LCD Master Alarm Operating & Maintenance Manual Alert-4 Ethernet LCD Master Alarm w w w. a m i c o. c o m Contents User Responsibility 4 Introduction 4 Features 5 Description of the Alarm 5 Shipment Details 5 The Alarm

More information

MICRO CONTROLLER S Z SERIES

MICRO CONTROLLER S Z SERIES Instruction Manual MICRO CONTROLLER S Z SERIES TYPE: PYW 4 5 7 9 Fuji Electric Co.,Ltd. INP-TN1PYWf-E INTRODUCTION You are now the owner of Fuji's digital Temperature Controller. Before using, be sure

More information

Table of Contents SECTION PAGE SECTION 1 INTRODUCTION Description Features Models... SECTION 2 RS-232 COMMUNICATIONS...

Table of Contents SECTION PAGE SECTION 1 INTRODUCTION Description Features Models... SECTION 2 RS-232 COMMUNICATIONS... SECTION Table of Contents SECTION 1 INTRODUCTION...................... 1.1 Description............................... 1.2 Features................................. 1.3 Models..................................

More information

Dual Column Dehydrating Breather Manual

Dual Column Dehydrating Breather Manual Dual Column Dehydrating Breather Manual DCB-MANUAL 1.6 Read and understand this manual prior to operating or servicing the products. SPX Transformer Solutions, Inc., Service & Components Division U.S.

More information

Carbon Monoxide Transmitter

Carbon Monoxide Transmitter Introduction The CO Transmitter uses an electrochemical sensor to monitor the carbon monoxide level and outputs a field-selectable 4-20 ma or voltage signal. The voltage signal may also be set to 0-5 or

More information

Beacon 800 Gas Monitor Operator s Manual

Beacon 800 Gas Monitor Operator s Manual Beacon 800 Gas Monitor Operator s Manual Part Number: 71-0037RK Revision: F Released: 4/18/17 www.rkiinstruments.com Product Warranty RKI Instruments, Inc. warrants gas alarm equipment sold by us to be

More information

EasyTronic III MANUAL SERVICE

EasyTronic III MANUAL SERVICE rev.6 EasyTronic III MANUAL SERVICE General characteristics: Power supply 24 Vac ±15% Max consumption at 24Vac 300mA Relay outputs 6 Maximum relay current 8 A res. Serial standard RS232 2 Serial standard

More information

SET-2000 Hi Level/Oil

SET-2000 Hi Level/Oil Labkotec Oy Labkotie 1 FI-36240 KANGASALA FINLAND Tel: + 358 29 006 260 Fax: + 358 29 006 1260 13.03.2008 Internet: www.labkotec.fi Alarm Device for Oil Separators Copyright 2008 Labkotec Oy We reserve

More information

AUR300C. Advanced Ultraviolet Burner Controller

AUR300C. Advanced Ultraviolet Burner Controller No. CP-SS-1807E AUR300C Advanced Ultraviolet Burner Controller Overview The AUR300C is a burner controller with a dynamic selfchecking function and is used in combination with the AUD300C or AUD500C Advanced

More information

OilSET-1000 (12 VDC)

OilSET-1000 (12 VDC) Labkotec Oy Myllyhaantie 6 FI-33960 PIRKKALA FINLAND Tel: +358 29 006 260 Fax: +358 29 006 1260 Internet: www.labkotec.fi 20.03.2009 1/11 OilSET-1000 (12 VDC) Oil Separator Alarm Device Copyright 2009

More information

Communications None RS-232C RS-422* RS-485* BCD Transmission output*/** (4 to 20 ma) E5AX- E5AX- L(M)A02 L(M)A03

Communications None RS-232C RS-422* RS-485* BCD Transmission output*/** (4 to 20 ma) E5AX- E5AX- L(M)A02 L(M)A03 Digital Controller A 96 x 96-mm (DIN) Digital Process Controller Optimum PID control with feed-forward control circuitry. High accuracy (+0.3% FS +1 digit max.). Replaceable Output Units. Models with communications

More information

Mark 25 Ultrapure Water Conductivity Analyzer

Mark 25 Ultrapure Water Conductivity Analyzer Martek Instruments, Inc. Mark 25 Ultrapure Water Conductivity Analyzer Instruction Manual WARRANTY POLICY Unless otherwise stated, MARTEK INSTRUMENTS, INC. warrants this equipment to be free from defects

More information

TABLE OF CONTENTS. NOTE: Read the entire instruction manual before starting the installation. TROUBLESHOOTING... 13

TABLE OF CONTENTS. NOTE: Read the entire instruction manual before starting the installation. TROUBLESHOOTING... 13 R 410A Duct Free Split System Air Conditioner and Heat Pump Product Family: DFS4(A/H) System, DFC4(A/H)3 Outdoor, DFF4(A/H)H Indoor NOTE: Read the entire instruction manual before starting the installation.

More information

Snifter ATEX22 VERSION. User Manual. Distributor

Snifter ATEX22 VERSION. User Manual. Distributor Snifter ATEX22 VERSION User Manual Distributor Version 1.4 09/09/2009 Table of Contents 1. INTRODUCTION... 3 1.1. Safety... 3 1.2. Product overview... 4 1.3. How does it work?... 4 2. INSTALLATION... 5

More information

Table of Contents SECTION PAGE

Table of Contents SECTION PAGE Table of Contents SECTION PAGE SECTION 1 INTRODUCTION................... 1.1 Description.............................. 1.2 Features................................ 1.3 Models.................................

More information

Refrigeration Controller Operator s Manual (HRC) PO Box 6183 Kennewick, WA

Refrigeration Controller Operator s Manual (HRC) PO Box 6183 Kennewick, WA Refrigeration Controller Operator s Manual (HRC) PO Box 6183 Kennewick, WA 99336 www.jmcvr.com 1-509-586-9893 Table of Contents TABLE OF FIGURES...1 OVERVIEW OF THE HRC CAPABILITIES...2 INSTALLATION AND

More information

Duct and Rough Service Carbon Monoxide Sensor

Duct and Rough Service Carbon Monoxide Sensor Product Identification and Overview Duct and Rough Service Carbon Monoxide Sensor BAPI s Carbon Monoxide Sensor offers enhanced electrochemical sensing with outstanding accuracy at low concentrations.

More information

OWNER S MANUAL Venstar Inc. 08/07

OWNER S MANUAL Venstar Inc. 08/07 Digital Thermostat residential THERMOSTAT T1 800 7-DAY PROGRAMMABLE up to 3-heat & 2-cool HEAT COOL HEAT PUMP Control up to 3 Heat & 2 Cool Stages 3 Configurable Outputs Adjustable 2nd & 3rd Stage Timers

More information

HALOGUARD TM IR MONITOR (OPTIONAL) FOUR or EIGHT POINT SCANNER INSTRUCTION MANUAL

HALOGUARD TM IR MONITOR (OPTIONAL) FOUR or EIGHT POINT SCANNER INSTRUCTION MANUAL HALOGUARD TM IR MONITOR (OPTIONAL) FOUR or EIGHT POINT SCANNER INSTRUCTION MANUAL SE R IA L N O. M O D E L N O. Tem p. Range 1 - = > 60 o F 2 - = 40-60 o F 3 - = < 40 o F G as Type 1 - R-11 2 - R-12 3

More information

INSTRUCTION MANUAL FOR MICROCOMPUTER BASED TEMPERATURE INDICATING CONTROLLER FCS-23A

INSTRUCTION MANUAL FOR MICROCOMPUTER BASED TEMPERATURE INDICATING CONTROLLER FCS-23A INSTRUCTION MANUAL FOR MICROCOMPUTER BASED TEMPERATURE INDICATING CONTROLLER Thank you for your purchase of our Microcomputer based Temperature Indicating Controller. This manual contains instructions

More information

F PC and AO OUTPUT BOARDS INSTRUCTION MANUAL. Blue-White. Industries, Ltd.

F PC and AO OUTPUT BOARDS INSTRUCTION MANUAL. Blue-White. Industries, Ltd. F-2000 PC and AO OUTPUT BOARDS INSTRUCTION MANUAL Blue-White R Industries, Ltd. 500 Business Drive Huntington Beach, CA 92649 USA Phone: 714-89-8529 FAX: 714-894-9492 E mail: sales@blue-white.com or techsupport@blue-white.com

More information

Fisher LCP100 Local Control Panel

Fisher LCP100 Local Control Panel Instruction Manual LCP100 Local Control Panel Fisher LCP100 Local Control Panel Contents Introduction... 1 Scope of Manual... 1 Description... 2 Specifications... 2 Installation... 3 Hazardous Area Classifications

More information

SET-2000 Oil/Sludge. Alarm Device for Oil Separators. Installation and Operating Instructions

SET-2000 Oil/Sludge. Alarm Device for Oil Separators. Installation and Operating Instructions SET-000 Oil/Sludge Alarm Device for Oil Separators Copyright 007 Labkotec Oy We reserve the right for changes without notice TABLE OF CONTENTS GENERAL... INSTALLATION... 4. SET-000 Oil/Sludge Control Unit...

More information

GG-EM ENTRANCE MONITOR. Installation and Operation Manual

GG-EM ENTRANCE MONITOR. Installation and Operation Manual GG-EM ENTRANCE MONITOR Installation and Operation Manual 2 GG-EM Warning Use this product only in the manner described in this manual. If the equipment is used in a manner not specified by Calibration

More information

Smart Transmitter/ Gas Detector Head

Smart Transmitter/ Gas Detector Head PT2E-1760 Smart Transmitter/ Gas Detector Head SD-1EC Operating Manual (PT2-176) Safety information -04 Specific conditions of use - The plastic window is submitted to tests corresponding

More information

Table of Contents. T538 Instruction Manual Rev # 6 June R6T538 June rtf E L Switchgear and Control Inc

Table of Contents. T538 Instruction Manual Rev # 6 June R6T538 June rtf E L Switchgear and Control Inc Table of Contents Introduction...2 Mounting...2 Dimensions...2 Controls...3 Screen Displays...3 Screen Messages...4 Light Test...5 Control Power...5 Wiring...5 Inputs...7 Outputs...7 Programming...8 Technical

More information

ADC150 Automatic Draft Control

ADC150 Automatic Draft Control Installation & Operating Manual ADC150 Automatic Draft Control USA CAN Product Information... Chapters 1 + 2 Mechanical Installation... Chapter 3 Electrical Installation... Chapter 4 Start Up and Configuration...

More information

Tempco Part Number PCT30006 Temperature Control Enclosure with Relay Output for Tote Tank Heating Applications

Tempco Part Number PCT30006 Temperature Control Enclosure with Relay Output for Tote Tank Heating Applications Instruction Manual Tempco Part Number PCT30006 Temperature Control Enclosure with Relay Output for Tote Tank Heating Applications Manual PCT30006, Revision 9/20/2016 The PCT30006 control enclosure incorporates

More information

Instruction VIBROCONTROL 850. Single channel vibration monitor. Keep it accessible for future reference.

Instruction VIBROCONTROL 850. Single channel vibration monitor. Keep it accessible for future reference. Instruction VIBROCONTROL 850 Single channel vibration monitor Keep it accessible for future reference. Copyright 2017 Brüel & Kjaer Vibro GmbH All rights to this technical documentation remain reserved.

More information

ATA11U03. Non-Programmable Heat Pump Thermostat. Installation Manual. 97B0082N06 Revised: 16 September, 2013 FOR INSTALLER AND OWNER ONLY

ATA11U03. Non-Programmable Heat Pump Thermostat. Installation Manual. 97B0082N06 Revised: 16 September, 2013 FOR INSTALLER AND OWNER ONLY Table of Contents Non-Programmable Heat Pump Thermostat Installation Manual 97B0082N06 Revised: 16 September, 2013 Safety Considerations 3 General 3 Package Contents 3 Installation Considerations 3 Power

More information

M2500 Engine Controller Installation Manual

M2500 Engine Controller Installation Manual M2500 Engine Controller Installation Manual Revision: 23-04-2012 Page 1 Contents 1 Preface... 4 2 Installation... 5 3 Terminal Connections... 6 4 Inputs... 7 4.1 Power Supply... 7 4.2 Mode/ Control Inputs...

More information

GasScanner 8C. Eight Channel Monitor. Operator s Manual. MINT-0281-XX Rev. A 01/29/08

GasScanner 8C. Eight Channel Monitor. Operator s Manual. MINT-0281-XX Rev. A 01/29/08 GasScanner 8C Eight Channel Monitor Operator s Manual MINT-0281-XX Rev. A 01/29/08 Product Warranty Matheson Tri-Gas, Inc., warrants gas alarm equipment sold by us to be free from defects in materials,

More information

GETTING STARTED GUIDE NI AI, ±29.4 mv/v, 24 Bit, 10 ks/s/ch Simultaneous, 350 Ω Quarter-Bridge Completion

GETTING STARTED GUIDE NI AI, ±29.4 mv/v, 24 Bit, 10 ks/s/ch Simultaneous, 350 Ω Quarter-Bridge Completion GETTING STARTED GUIDE NI 9236 8 AI, ±29.4 mv/v, 24 Bit, 10 ks/s/ch Simultaneous, 350 Ω Quarter-Bridge Completion This document explains how to connect to the NI 9236. Note Before you begin, complete the

More information

Beacon 410A Gas Monitor Operator s Manual

Beacon 410A Gas Monitor Operator s Manual Beacon 410A Gas Monitor Operator s Manual Part Number: 71-0397 Revision: F Released: 12/5/17 www.rkiinstruments.com Product Warranty RKI Instruments, Inc., warrants gas alarm equipment sold by us to be

More information

Isolated Thermocouple Module

Isolated Thermocouple Module User Manual for the HE693THM665, HE693THM666, HE693THM667, HE693THM668 Isolated Thermocouple Module Twelfth Edition 19 April 2004 PREFACE 19 APR 2004 PAGE 3 PREFACE This manual explains how to use the

More information

Refer to Bulletin E-1101 for detailed information on the FLAME-MONITOR System.

Refer to Bulletin E-1101 for detailed information on the FLAME-MONITOR System. The Fireye EP260, EP270 (early spark termination), or EP265 (pilot stabilization) programmer modules are used with the FLAME-MONITOR Burner Management Control System (P/N E110). Several operational characteristics

More information

Interchangeable thermistor (THE types) Output Contact E5CS-R1KJX-F E5CS-R1PX-F E5CS-R1GX-F Voltage E5CS-Q1KJX-F E5CS-Q1PX-F E5CS-Q1GX-F

Interchangeable thermistor (THE types) Output Contact E5CS-R1KJX-F E5CS-R1PX-F E5CS-R1GX-F Voltage E5CS-Q1KJX-F E5CS-Q1PX-F E5CS-Q1GX-F Temperature Controller 1/16 DIN Sized Controller Offers Selectable Control, Alarm Modes Accurate to ±0.5% of full scale Multiple temperature scale ranges allow flexibility to match application Field-selectable

More information

Explorer F210 Fuel Instrument Installation and Operation Manual

Explorer F210 Fuel Instrument Installation and Operation Manual Explorer F210 Fuel Instrument Installation and Operation Manual www.northstarnav.com 2 Northstar Explorer F210 Installation and Operation Manual IMPORTANT SAFETY INFORMATION Please read carefully before

More information

DIGITAL INDICATOR DI25

DIGITAL INDICATOR DI25 DIGITAL INDICATOR DI25 Operating instructions Preface Thank you for purchasing our digital indicator DI25. This manual contains instructions for the mounting, functions, operations and notes when operating

More information

Instructions for the fan motor control system with integrated wiring terminals SILVER C

Instructions for the fan motor control system with integrated wiring terminals SILVER C Instructions for the fan motor control system with integrated wiring terminals SILVER C 1. General The motor control system is used for controlling the type EC, 0.41-10 kw fan motors in the SILVER C units.

More information

Table of Contents 1. OVERVIEW SYSTEM LAYOUT SPECIFICATIONS FUNCTION... 11

Table of Contents 1. OVERVIEW SYSTEM LAYOUT SPECIFICATIONS FUNCTION... 11 Table of Contents 1. OVERVIEW... 3 2. SYSTEM LAYOUT... 4 3. SPECIFICATIONS... 8 3.1 SYSTEM COMPONENTS...9 3.2 PLC INPUTS AND OUTPUTS...9 3.3 FUNCTION KEYS...10 3.4 DEFAULT SET POINTS AND TIMERS...10 4.

More information

SC-3 Controller INSTALLATION AND OPERATING INSTRUCTIONS FOR GAS-FIRED, OIL FIRED AND DUAL FUEL HOT WATER HEATING PLANTS. SC-3 Applications CONTENTS

SC-3 Controller INSTALLATION AND OPERATING INSTRUCTIONS FOR GAS-FIRED, OIL FIRED AND DUAL FUEL HOT WATER HEATING PLANTS. SC-3 Applications CONTENTS FOR GAS-FIRED, OIL FIRED AND DUAL FUEL HOT WATER HEATING PLANTS SC-3 Controller MODULAR BOILER CONTROLLER SERIES INSTALLATION AND OPERATING INSTRUCTIONS SC-3 Applications Space Heat Systems With Outdoor

More information

SAFETY CERTIFIED MODEL FP-700 COMBUSTIBLE GAS DETECTOR

SAFETY CERTIFIED MODEL FP-700 COMBUSTIBLE GAS DETECTOR SAFETY MANUAL SIL 2 Certified Model FP-700 Combustible Hydrocarbon Gas Sensor Version 2.0 1 SAFETY CERTIFIED MODEL FP-700 COMBUSTIBLE GAS DETECTOR This manual addresses the specific requirements and recommendations

More information

DTM Distributed Transmitter-Monitor

DTM Distributed Transmitter-Monitor DTM Distributed Transmitter-Monitor DTM10 Proximity Distributed Transmitter-Monitor (Shaft Vibration, Thrust Position and Speed) The DTM10 distributed vibration transmitter-monitor is ideal for monitoring

More information

CORONA BURNER CONTROLLER INSTRUCTION MANUAL RCF15 EXCLUSIVE FOR FLAME ROD. Corona Corporation

CORONA BURNER CONTROLLER INSTRUCTION MANUAL RCF15 EXCLUSIVE FOR FLAME ROD. Corona Corporation CORONA BURNER CONTROLLER INSTRUCTION MANUAL RCF15 EXCLUSIVE FOR FLAME ROD Corona Corporation This instruction manual contains important information for safe use of this device. Keep this instruction manual

More information

Rooftop Thermostat Controller Specification and Installation Instructions. Model TRT2422

Rooftop Thermostat Controller Specification and Installation Instructions. Model TRT2422 ºF / º C Rooftop Thermostat Controller Model TRT2422 Description The TRT2422 is a combination controller and thermostat with a built-in scheduler, which is designed for simple and accurate control of single

More information

DSF-2010-SB KEEP THIS MANUAL IN A SAFE PLACE FOR FUTURE REFERENCE!

DSF-2010-SB KEEP THIS MANUAL IN A SAFE PLACE FOR FUTURE REFERENCE! DSF-2010-SB KEEP THIS MANUAL IN A SAFE PLACE FOR FUTURE REFERENCE! Read this manual before using this product. Failure to follow the instructions and safety precautions in this manual can result in serious

More information

INSTRUCTION MANUAL. Handy Controller Exclusive for SF4B / SF4B<V2> Series. SFB-HC Ver.2.1 FUNCTION CANCEL ENTER

INSTRUCTION MANUAL. Handy Controller Exclusive for SF4B / SF4B<V2> Series. SFB-HC Ver.2.1 FUNCTION CANCEL ENTER INSTRUCTION MANUAL Handy Controller Exclusive for SF4B / SF4B Series SFB-HC Ver.2.1 FUNCTION - + CANCEL ENTER Panasonic Industrial Devices SUNX Co., Ltd. 2013 MJE-SFBHC No.0037-78V (MEMO) 1 Panasonic

More information

CP CP

CP CP CP-8161-333 CP-8161-433 Electronic Programmable Controller Six Stage, Dual Setpoint General Instructions APPLICATION Electronic six stage programmable controller with proportional output for heating, cooling

More information

D A R A - 4. Data Aire Relay Auto-Changeover

D A R A - 4. Data Aire Relay Auto-Changeover D A R A - 4 Data Aire Relay Auto-Changeover WARNING: If adding the DARA-4 panel to existing units that did not include a DARA-4 panel when originally purchased it will be necessary to add a relay or relays.

More information

HALOGUARD IR MULTI-POINT, COMPOUND SPECIFIC MONITOR INSTRUCTION MANUAL

HALOGUARD IR MULTI-POINT, COMPOUND SPECIFIC MONITOR INSTRUCTION MANUAL HALOGUARD IR MULTI-POINT, COMPOUND SPECIFIC MONITOR INSTRUCTION MANUAL S E R I A L N O. M O D E L N O. T e m p. R a n g e 1 - = > 6 0 o F 2 - = 4 0-6 0 o F 3 - = < 4 0 o F - G a s T y p e 1 - R -1 1 2

More information

Owners Manual. HTC621W 7-Day Programmable Thermostat 3600W. HTC625W 7-Day Programmable Thermostat 4000W

Owners Manual. HTC621W 7-Day Programmable Thermostat 3600W. HTC625W 7-Day Programmable Thermostat 4000W Owners Manual 7213110100R06 HTC621W 7-Day Programmable Thermostat 3600W HTC625W 7-Day Programmable Thermostat 4000W HTC627-347W 7-Day Programmable Thermostat 5200W Congratulations on your purchase of this

More information

Digital Panel Meter K3TL. Ordering Information. Low-cost, High-quality Digital Thermometer with Built-in Microcomputer. Accessories (Order Separately)

Digital Panel Meter K3TL. Ordering Information. Low-cost, High-quality Digital Thermometer with Built-in Microcomputer. Accessories (Order Separately) Digital Panel Meter Low-cost, High-quality Digital Thermometer with Built-in Microcomputer Compact DIN-size (96 x 48 x 66 (W x H x D)) body. Mounting thickness of only 3.5 mm required. Highly visible display

More information

Sintrol Snifter A1+ Manual. Version 1.2.2

Sintrol Snifter A1+ Manual. Version 1.2.2 Sintrol Snifter A1+ Manual Version 1.2.2 Table of Contents 1 INTRODUCTION... 3 1.1 Safety... 3 1.2 Product overview... 3 1.3 How does it work?... 3 2 INSTALLATION... 4 2.1 Selecting the installation location...

More information

SEC 2000 Millenium Infrared Gas Detector

SEC 2000 Millenium Infrared Gas Detector SEC 2000 Millenium Infrared Gas Detector Instruction and Operation Manual Sensor Electronics Corporation 5500 Lincoln Drive Minneapolis, Minnesota 55436 USA (952) 938-9486 Fax (952) 938-9617 Email: sales@sensorelectronic.com

More information

INSTRUCTION MANUAL FOR

INSTRUCTION MANUAL FOR INSTRUCTION MANUAL FOR Cool Block The 1st Edition Yamato Scientific Co.,, Ltd. 2-1-6, Nihonbashi Honcho, Chuo-ku Tokyo 103-8432 JAPAN Congratulations on your selection of Yamato Scientific s Cool Block

More information

Rosemount 4500 Hygienic Pressure Transmitter. Start

Rosemount 4500 Hygienic Pressure Transmitter. Start Quick Installation Guide June 2007 Rosemount 4500 Rosemount 4500 Hygienic Pressure Transmitter ProductDiscontinued Start Step 1: Mount the Transmitter Step 2: Set the Switches Step 3: Connect Wiring and

More information

SV-8 Universal Process Indicator User Manual

SV-8 Universal Process Indicator User Manual Uniersal Process Indicator User Manual ABUS TECHNOLOGIES INC. WARNING This manual should be passed on to the end user. The contents of this manual are subject to change without prior notice. All rights

More information

TCB-IFTH1GUL. 24V Thermostat Interface. Installation Manual. Model name: English FOR INSTALLER

TCB-IFTH1GUL. 24V Thermostat Interface. Installation Manual. Model name: English FOR INSTALLER 24V Thermostat Interface Model name: TCB-IFTH1GUL FOR INSTALLER Read this manual before using the TCB- IFTH1GUL 24V Thermostat Interface. Refer to installation manuals supplied with Indoor Unit, Remote

More information

INSTALLATION INSTRUCTIONS

INSTALLATION INSTRUCTIONS TT-1343 5/06b INSTALLATION INSTRUCTIONS Original Issue Date: 8/03 Model: Automatic Transfer Switches Equipped with Series 1000 Programmable Controller Market: ATS Subject: Remote Annunciator Kits GM28938-KP1,

More information

Non-contact voltage (for SSR drive): V DC DC current: 4 to 20mA DC Thermocouple: K, J, E Input. RTD: Pt100, JPt100

Non-contact voltage (for SSR drive): V DC DC current: 4 to 20mA DC Thermocouple: K, J, E Input. RTD: Pt100, JPt100 INSTRUCTI MANUAL Micro-computer based Temperature Indicating Controller 1 GCS-00 No.GCS1E 200.11 To prevent accidents arising from the use of this controller, please ensure the operator receives this manual.

More information

AMC-RAM-4. Refrigerant Alarm Module USER MANUAL. Please read the installation and operating instructions completely and carefully before starting.

AMC-RAM-4. Refrigerant Alarm Module USER MANUAL. Please read the installation and operating instructions completely and carefully before starting. AMC-RAM-4 Refrigerant Alarm Module USER MANUAL IMPORTANT: Please read the installation and operating instructions completely and carefully before starting. Filename: 3310405B, DOC, AMC_RAM4_Manual.doc

More information

MODEL AC-SCC-5 INSTALLATION, OPERATION & MAINTENANCE MANUAL

MODEL AC-SCC-5 INSTALLATION, OPERATION & MAINTENANCE MANUAL 5CIM2-0413 W30-AC0056 MODEL AC-SCC-5 INSTALLATION, OPERATION & MAINTENANCE MANUAL Multi-zone Hydronic Chiller Interface Module SECTION 1: INTRODUCTION Unit description... 5 Location... 5 Handling... 5

More information

- Data Brochure Steam Control 279

- Data Brochure Steam Control 279 - Data Brochure Steam Control 279 D 279 12/07 The tekmar Steam Control 279 can operate a single on-off steam boiler or an on-off steam valve using outdoor reset. The control determines the on time of the

More information

AutomationDirect PM24. Microprocessor - Based Process/Temperature Limit Controller. Operator s Manual. Technical Support.

AutomationDirect PM24. Microprocessor - Based Process/Temperature Limit Controller. Operator s Manual. Technical Support. AutomationDirect 1/16 DIN Series Technical Support We strive to make our manuals the best in the industry. We rely on your feedback to let us know if we are reaching our goal. If you cannot find the solution

More information

VERSAFLOW VORTEX Supplementary instructions

VERSAFLOW VORTEX Supplementary instructions VERSAFLOW VORTEX Supplementary instructions Vortex flowmeter Equipment category II 2G CONTENTS VERSAFLOW VORTEX 1 Safety instructions 3 1.1 General notes... 3 1.2 EC conformity... 3 1.3 Approval according

More information

EC3-X33 Stand-alone Superheat Controller Technical Data

EC3-X33 Stand-alone Superheat Controller Technical Data Technical Data EC3-X33 is a stand-alone universal superheat controller for air conditioning, refrigeration and industrial applications such as chillers, industrial process cooling, rooftops, heat pumps,

More information

Instruction manual Temperature Controller KT7

Instruction manual Temperature Controller KT7 Instruction manual Temperature Controller KT7 No. KT71E7 201.05 To prevent accidents arising from the misuse of this controller, please ensure the operator receives this manual. SAFETY PRECAUTIS Be sure

More information