Semiannual Maintenance and Calibration Report Prepared by Air Resource Specialists, Inc.

Size: px
Start display at page:

Download "Semiannual Maintenance and Calibration Report Prepared by Air Resource Specialists, Inc."

Transcription

1 Semiannual Maintenance and Calibration Report Prepared by Air Resource Specialists, Inc. Client: National Park Service Field Personnel: Mike Farinacci Site: Island in the Sky (CANY-IS) Service Date(s): /29/26 Site Operator: Cody Marsh Subject: Semiannual Maintenance All site visit and calibration forms are attached, detailing the pre- and post-maintenance calibrations and test results. SUMMARY OF FINDINGS GASEOUS POLLUTANT SUPPORT EQUIPMENT Zero-Air System (Werther / Panther Compressor PC7/4E): Maintenance - No maintenance for the pump or compressor was required at this time. The media for the zero-air system was replaced. Station Temperature: Pre-Maintenance Testing - The sensor was compared against a certified temperature transfer standard. The sensor was found to be responding outside of calibration acceptance criteria. Maintenance - There was a -3.5 offset on the STP channel, which was removed. Post-Maintenance Testing - The sensor was compared against a certified temperature transfer standard. The sensor was found to be responding within calibration acceptance criteria. GASEOUS POLLUTANT ANALYZERS Ozone Analyzer (Thermo 49i): Pre-Maintenance Testing - The ozone analyzer was compared against a Level 2 ozone transfer standard. The ozone analyzer was found to be responding within calibration acceptance criteria. Maintenance - Cooling fan filter was cleaned and the instrument was calibrated at 2 ppb. Post-Maintenance Testing - The ozone analyzer was compared against a Level 2 ozone transfer standard. The ozone analyzer was found to be responding within calibration acceptance criteria. Ozone Station Reference (Thermo 49i): Pre-Maintenance Testing - The Level 3 ozone station reference was recertified using a Level 2 ozone transfer standard. The ozone station reference was found to be responding within calibration acceptance criteria. Maintenance - New coefficients were entered in the datalogger to reflect the updated certification. Post-Maintenance Testing - The ozone station reference was compared against a Level 2 ozone transfer standard. The ozone station reference was found to be responding within calibration acceptance criteria.

2 METEOROLOGICAL SENSORS Ambient Temperature and Vertical Temperature Difference (RM Young 4342VC): Pre-Maintenance Testing - The sensor was compared against a certified temperature transfer standard in three water baths controlled at temperatures between and 5 degrees Celsius. The sensor was confirmed to be responding within calibration acceptance criteria. The aspirator fan was functioning correctly. Maintenance - No maintenance was performed. Post-Maintenance Testing - No post-maintenance checks were performed. Relative Humidity (Rotronic MP6): Pre-Maintenance Testing - The sensor was collocated with a certified relative humidity transfer standard. The sensor was confirmed to be responding within calibration acceptance criteria. The aspirator fan was functioning correctly. Maintenance - The sensor was swapped with a laboratory certified sensor. Post-Maintenance Testing - The sensor was collocated with a certified relative humidity transfer standard. The sensor was confirmed to be responding within calibration acceptance criteria. Wind Speed (Climatronics 75): Pre-Maintenance Testing - The sensor was challenged with a certified anemometer drive. The sensor was confirmed to be responding within calibration acceptance criteria. The starting threshold test for the sensor was within acceptance criteria. The heater for the sensor was found to be functioning correctly. Maintenance - The sensor was swapped with a laboratory rebuilt sensor. Post-Maintenance Testing - The sensor was challenged with a certified anemometer drive. The sensor was confirmed to be responding within calibration acceptance criteria. The starting threshold test for the sensor was within acceptance criteria. The heater for the sensor was functioning correctly. Wind Direction (Climatronics 76): Pre-Maintenance Testing - The reference alignment for the sensor was checked using a compass. The reference alignment for the sensor was confirmed to be within acceptable limits. The accuracy of the sensor was tested by comparison to a reference. The sensor was confirmed to be responding within calibration acceptance criteria. The linearity of the sensor was within acceptable limits. The starting threshold test results were within acceptance criteria. The heater for the sensor was confirmed to be functioning correctly. Maintenance - The sensor was swapped with a laboratory rebuilt sensor. Post-Maintenance Testing - The reference alignment for the sensor was checked using a compass. The reference alignment for the sensor was confirmed to be within acceptable limits. The accuracy of the sensor was tested by comparison to a reference. The sensor was confirmed to be responding within calibration acceptance criteria. The linearity of the sensor was within acceptable limits. The starting threshold test for the sensor was within acceptance criteria. The heater for the sensor was functioning correctly.

3 Solar Radiation (Li-Cor Pyranometer): Pre-Maintenance Testing - The sensor was collocated with a certified solar radiation transfer standard. The sensor was confirmed to be responding outside of calibration acceptance criteria. (SELECT) Maintenance - The sensor was re-leveled and the zero was calibrated on the Climatronics translator. The solar levels were very low during this visit. A shunted sensor will be installed during the next SAM visit. Post-Maintenance Testing - The sensor was collocated with a certified solar radiation transfer standard. The sensor was confirmed to be responding outside of calibration acceptance criteria. Precipitation (Climatronics 58): Pre-Maintenance Testing - The sensor was challenged using a known volume of water. The sensor was found responding within calibration acceptance criteria. The sensor was found to be level and clean. The heater was found to be functional. Maintenance - No maintenance was performed. Post-Maintenance Testing - No post-maintenance checks were performed. PARTICULATE MONITORS AND SAMPLERS CASTNET Filter Pack Flow: Pre-Maintenance Testing - A leak check on the system was performed and results were within acceptable limits. The flow was checked using a certified flow standard measuring flow in standard conditions. The measured flow was found to be within calibration acceptance criteria. Maintenance - The balston filter was replaced, the brass elbow was cleaned, and the pump diaphragm was replaced. No calibration was performed. Post-Maintenance Testing - The flow was checked using a certified flow standard measuring flow in standard conditions. The measured flow was confirmed to be within calibration acceptance criteria. ADDITIONAL COMMENTS The site is in good condition.

4 SITE INFORMATION ABBR. CANY-IS SITE NAME NETWORK TYPE CLIENT NPS FIELD SPECIALIST M. Farinacci DATE /29/26 Island in the Sky DATE OF LAST VISIT 5/4/26 NPS Deg Min Sec LATITUDE North LONGITUDE West CALCULATE-> Decimal Decimal Deg Min Sec --CALCULATE-> ELEVATION Meters --CALCULATE-> Feet Feet CALCULATE-> Meters 789 Photo Documentation Completed? Yes No N/A Cellular Phone Coverage Cellular Phone Coverage Protocol? Carrier? # of Bars? Signal Strength? -X dbm -X dbm DAY TIME IN TIME OUT /29/ SITE STANDARDS PM Flow Reference Please verify site standards used by the site operator MANUFACTURER MODEL SERIAL # Calibration Expiration Date NOTES:

5 TEMPERATURE / VERTICAL TEMPERATURE DIFFERENCE SYSTEM VERIFICATION & CALIBRATION ABBR. CANY-IS CLIENT NPS FIELD SPECIALIST M. Farinacci DATE /29/26 SITE NAME Island in the Sky DATE OF LAST VISIT 5/4/26 Network type NPS MANUFACTURER MODEL SERIAL NUMBER EXPIRATION DATE Temperature Reference Eutechnics /8/27 2m Temperature Sensor Climatronics List sensors according to height on tower, from highest to lowest. Climatronics Temp. Deltas CALIBRATION ACCEPTANCE CRITERIA (<=) Ambient Temperature Difference ( o C).5 Vertical Temperature Difference ( o C) N/A 2m Temperature Bath Temp ( o C) DAS Difference MAX ABS Difference.29 MAX ABS Difference Aspirator fan functional 2m? Yes No N/A Yes No N/A Yes No N/A Yes No N/A Each sensor was verified against its data channel? Yes No N/A Each Temperature Difference = Upper - Lower? Yes No N/A Bath Temp ( o C) MAX ABS Difference MAX ABS Difference NOTES:

6 STATION TEMPERATURE SENSOR VERIFICATION & CALIBRATION ABBR. CANY-IS CLIENT NPS FIELD SPECIALIST M. Farinacci DATE SITE NAME Island in the Sky DATE OF LAST VISIT Network type NPS /29/26 5/4/26 Temperature Reference MANUFACTURER Eutechnics MODEL SERIAL NUMBER EXPIRATION DATE 9/8/27 CALIBRATION ACCEPTANCE CRITERIA (<=) Temperature Difference ( o C). Temperature Temperature Reference ( o C) DAS ( o C) Difference Reference ( o C) DAS ( o C) Difference FAIL NOTES: -3.5 offset was removed from STP channel.

7 RELATIVE HUMIDITY SENSOR VERIFICATION & CALIBRATION ABBR. CANY-IS CLIENT NPS FIELD SPECIALIST M. Farinacci DATE SITE NAME Island in the Sky DATE OF LAST VISIT Network type NPS /29/26 5/4/26 RH SENSOR REFERENCE MANUFACTURER Vaisala MODEL HMP55 SERIAL NUMBER G2837 EXPIRATION DATE 4/5/27 Rotronic MP Rotronic MP6A 5264 FIELD CALIBRATION ACCEPT. CRITERIA (<=) Relative Humidity Difference (%) 5% Hour 2 3 Relative Humidity (%) Relative Humidity (%) STD DAS Difference V Hour STD DAS Difference 49.% 52.% 2.9% 4 4.% 42.7%.7% 42.2% 44.9% 2.7% % 49.% 2.5% 42.2% 44.6% 2.4% % 56.8% 2.5% V Average 2.7% Average 2.2% Aspirator fan functional? Yes No N/A NOTES: Sensor was swapped.

8 WIND SPEED SENSOR VERIFICATION & CALIBRATION ABBR. CANY-IS CLIENT NPS FIELD SPECIALIST M. Farinacci DATE /29/26 SITE NAME Island in the Sky DATE OF LAST VISIT 5/4/26 Network type NPS MANUFACTURER MODEL SERIAL NUMBER EXPIRATION DATE Wind Speed Reference RM Young 882 CA /8/27 Wind Speed Torque Gauge RM Young 83 n/a and Climatronics - 75 / and Climatronics - 75 / Heavy Duty Aluminum Heavy Duty Aluminum Sensor Serial # 478 Sensor Serial # 284 Cups Serial # 44 Cups Serial # 44 CALIBRATION ACCEPTANCE CRITERIA (<=) Wind Speed Difference (m/s) Wind Speed Difference (%).25 5.% if wind speed <= 5 m/s if wind speed > 2 m/s Wind Speed Threshold (m/s vs % Diff.) Select UNITS m/s Motor Speed (rpm) Target Speed Wind Speed DAS Difference.93 N/A N/A N/A % % % % % Starting Threshold Torque <=.3 g-cm TORQUE.3 NO ACTION REQUIRED Heater sleeve functional? Yes No N/A Wind Speed Motor Speed (rpm) Target Speed DAS Difference..99 N/A N/A N/A % % % % % Starting Threshold Torque <=.3 g-cm TORQUE.3 NO ACTION REQUIRED NOTES: Sensor was swapped.

9 WIND DIRECTION SENSOR VERIFICATION & CALIBRATION ABBR. CANY-IS CLIENT NPS FIELD SPECIALIST M. Farinacci DATE /29/26 SITE NAME Island in the Sky DATE OF LAST VISIT 5/4/26 Network type NPS Direction Alignment Reference Direction Linearity Reference Direction Torque Gauge MANUFACTURER MODEL SERIAL NUMBER Brunton Transit Wilson Machining 8 pt jig 85 RM Young 83 n/a EXPIRATION DATE & Sensor Serial # Vane Serial # Climatronics - 76 & Climatronics Sensor Serial # #2 35 Vane Serial # 35 Local Magnetic Declination (degrees). Method Solar Azimuth CALIBRATION ACCEPTANCE CRITERIA (<=) Cross-arm Alignment Error (degrees) 2 Total Align. Diff (degrees) 5 Sensor Linearity (degrees) 5 Mag. Dec. from NOAA (deg/min/sec) Landmarks From the North From the South From the East From the West Degrees Is the Reference Alignment intended to be N-S? YES Is the Reference Alignment intended to be N-S? YES Reference Alignment (degrees) 358. Reference Alignment (degrees) 358. SENSOR ALIGNMENT N-S Reference Degrees DAS From the North 36.5 From the South From the East From the West Total Alignment MAX ABS Diff OR SENSOR ALIGNMENT N-S Reference Degrees DAS From the North From the South From the East From the West Total Alignment MAX ABS Diff SENSOR ALIGNMENT SENSOR ALIGNMENT Landmark Degrees DAS Difference Landmark Degrees DAS Difference From the North From the North From the South 8 From the South 8 From the East 9 From the East 9 From the West 27 From the West 27 Total Alignment MAX ABS Diff Total Alignment MAX ABS Diff OR Difference SENSOR ALIGNMENT SENSOR ALIGNMENT X Reference Degrees DAS Difference X Reference Degrees DAS 358 Align with Ref (N) Align with Ref (S) Perp with Ref (E) Perp with Ref (W) 628 Total Alignment MAX ABS Diff Total Alignment MAX ABS Diff OR OR Difference Difference SENSOR LINEARITY Point DAS Difference N/A MAX Difference 3 SENSOR LINEARITY Point DAS Difference 33.3 N/A MAX Difference 3 NO ACTION REQUIRED NO ACTION REQUIRED Starting Threshold TORQUE Starting Threshold Torque <= 6. g-cm 5.3 Torque <= 6. g-cm NO ACTION REQUIRED TORQUE 5.3 NO ACTION REQUIRED Heater sleeve functional? Yes No N/A NOTES: Sensor was swapped.

10 SOLAR RADIATION SENSOR VERIFICATION & CALIBRATION ABBR. CANY-IS CLIENT NPS FIELD SPECIALIST M. Farinacci DATE /29/26 SITE NAME Island in the Sky DATE OF LAST VISIT Network type NPS 5/4/26 Solar Radiation Reference MANUFACTURER MODEL SERIAL NUMBER EXPIRATION DATE Licor Pyranometer PY /5/27 MULTIPLIER 2. Translator Logger Type ESC Licor Pyranometer PY Climatronics High Input (V) 5. Low Input (V). High Output 396. Low Output. Licor Pyranometer PY Translator Climatronics Logger Type High Input (V) 5. ESC Low Input (V). High Output 396. Low Output. CALIBRATION ACCEPTANCE CRITERIA (<=) Difference from CTS (%) % Solar Radiation Hour CTS (W/m 2 ) DAS (W/m 2 ) Difference % % % V DARK RESPONSE DAS (W/m 2 ) MEAN ABS % DIFF 5.4% ACTION Sensor found clean? Yes No Sensor found level? Yes No Solar Radiation DAS (W/m 2 ) Hour CTS (W/m 2 ) Difference 7.2% 5.% 6.5% V DAS (W/m 2 ) DARK RESPONSE MEAN ABS % DIFF 6.2% ACTION NOTES: Sensor level was adjusted, zero was calibrated. Low solar levels and clouds prevented a valid check after sensor was leveled and zero calibrated.'

11 PRECIPITATION SENSOR VERIFICATION & CALIBRATION ABBR. CANY-IS CLIENT NPS FIELD SPECIALIST M. Farinacci DATE /29/26 SITE NAME Island in the Sky DATE OF LAST VISIT 5/4/26 Network type NPS MANUFACTURER MODEL SERIAL NUMBER Precipitation Reference Texas Electronics FC-525 n/a EXPIRATION DATE Climatonics 58 NPS 987 Climatonics 58 CALIBRATION ACCEPTANCE CRITERIA (<=) Difference from Input Volume (%) 5% x Reference Chart Input Volume (ml) Climatronics Climatronics Met One 97--G-H Diameter (in.) mm/tip ml/tip Met One RM Young Texas Electronics TR-525I-HT DAS target Conversions Value Units Value. inch mm. Units mm inch Reference (ml) Target (mm) Precipitation DAS (mm) Difference % Heater functional? Yes No N/A Sensor found level? Yes No Sensor found clean? Yes No Precipitation Reference (ml) Target (mm) DAS (mm) Difference 94 NOTES:

12 CASTNET FLOW SYSTEM VERIFICATION & CALIBRATION ABBR. CANY-IS CLIENT NPS FIELD SPECIALIST M. Farinacci DATE /29/26 SITE NAME Island in the Sky DATE OF LAST VISIT 5/4/26 Network type NPS MFC High Flow Reference MANUFACTURER MODEL SERIAL NUMBER BIOS Definer 22-H 2694 EXPIRATION DATE 2/8/27 MFC Tylan FC-28 AW94322 MFC Tylan FC-28 AW94322 Pump Flow Target 3. Thomas 7CAB8A 682A ENTER FLOWS IN SLPM Pump Thomas 7CAB8A 682A MFC Display Linear MFC VDC Linear Regression Regression Date 5/3/26 Setpoint Slope Intercept Setpoint Slope Intercept ESC Logger Scaling High Input (V) Low Input (V) High Output (LPM) Low Output (LPM) Operational Checks Vacuum DAS Flow MFC Flow Flow Standard Target Flow Standard Rotometer MFC Display MFC Volts DC DAS Display Difference Value % Pump Off. SLPM Leak Test. SLPM SLPM % 2 (low) 2.5 SLPM % 3 (target) 3. SLPM % 4 (high) 3.5 SLPM % 5 4. SLPM % MAX > 4. SLPM % Leak Test Results Flow (SLPM).3 *Note: A leak is present if the difference between the zero and leak test value is greater than. lpm Recalibration Not Required *Note: A calibration is required if the difference between the transfer standard and the DAS value is greater than +/- 2.% at the target setting or greater than +/- 2.5% at either the low or high point NEW CALIBRATION FACTORS MFC Display Linear MFC VDC Linear Regression Regression ESC Logger Scaling Setpoint 2.73 Setpoint High Input (V) 5. Slope. Slope. Low Input (V). Intercept -.5 Intercept -.4 High Output (LPM) Correlation. Correlation. Low Output (LPM) -.4 Target Flow Standard Rotometer MFC Display MFC Volts DC DAS Display Value % Pump Off. SLPM Leak Test. SLPM SLPM % 2 (low) 2.5 SLPM % 3 (target) 3. SLPM % 4 (high) 3.5 SLPM % 5 4. SLPM % MAX > 4. SLPM % Leak Test Results Flow (SLPM).2 *Note: A leak is present if the difference between the zero and leak test value is greater than. lpm Operational Checks Vacuum 9. DAS Flow 3. MFC Flow 2.8 Flow Standard 3. NOTES: Pump diaphragm replaced, filter replaced, elbow cleaned. No calibration was necessary. Difference

13 OZONE ANALYZER VERIFICATION & CALIBRATION () ABBR. CANY-IS CLIENT NPS FIELD SPECIALIST M. Farinacci DATE /29/26 SITE NAME Island in the Sky DATE OF LAST VISIT 5/4/26 Network type NPS Logger Slope Correction.9966 Logger Intercept Correction.848 CALIBRATION ACCEPTANCE CRITERIA (<=) TRANSFER STANDARD AMBIENT ANALYZER STATION REFERENCE Mean Absolute Difference (%) 3% Thermo Thermo Thermo Maximum Absolute Difference (%) 3% Linearity % of Fullscale % 49i 49i 49i DATA ACCEPTANCE CRITERIA (<=) Mean Absolute Difference (%) % Coefficient Coefficient.6 Coefficient. Maximum Absolute Difference (%) % Background Background -.2 Background Linearity % of Fullscale 2% Cell A Intensity (Hz) 53 Cell A Intensity (Hz) 7389 Cell A Intensity (Hz) 8638 SLOPE INT Cell B Intensity (Hz) 8769 Cell B Intensity (Hz) 866 Cell B Intensity (Hz) Transfer Standard Correction Factors Flow A (lpm).744 Flow A (lpm).78 Flow A (lpm).67 Flow B (lpm).762 Flow B (lpm).697 Flow B (lpm).67 Full Scale (ppb) 25 Pressure (mmhg) Pressure (mmhg) Pressure (mmhg) Linear Regression Plot Bench Temp ( C) 26.8 Bench Temp ( C) 36.4 Bench Temp ( C) 33.2 Bench Lamp Temp ( C) 54.3 Bench Lamp Temp ( C) 53.9 Bench Lamp Temp ( C) 53.9 Ozone Lamp Temp ( C) 68.3 Ozone Lamp Temp ( C) 67. Ozone Lamp Temp ( C) Ozone Source X Ozone Source Transfer Standard Level LEVEL 2 Transfer Standard Level LEVEL 3 O3 LAMP TRANSFER STANDARD AMBIENT ANALYZER STATION REFERENCE POINT TARGET % or mv Display Corrected DAS Diff %Diff Linearity % DAS Diff %Diff Linearity % ZERO N/A.% N/A.% % -.2% %.% %.% % -.% %.% % -.2% %.% 7.2..%.% %.% %.2% ZERO N/A -.% N/A -.% Mean ABS % Diff 2.4% Mean ABS % Diff.8% Max ABS % Difference 2.9% Max ABS % Difference.2% Analog Test Analyzer DAS Reference DAS Slope.28 Slope.5 Line Loss Test Analyzer Display Span (site tubing) 25 Span (bypass tubing) 23 NOTES: Zero Full Scale % P/F.% Y-Intercept -.7 Y-Intercept -.99 Correlation. Correlation. Analyzer DAS (ppb) Residuals Plot Residual (ppb) y =.289x Transfer Standard (ppb) R² = Regression Concentration Value (ppb)

14 OZONE ANALYZER VERIFICATION & CALIBRATION () ABBR. CANY-IS CLIENT NPS FIELD SPECIALIST M. Farinacci DATE /29/26 SITE NAME Island in the Sky DATE OF LAST VISIT 5/4/26 Network type NPS Logger Slope Correction.9964 Logger Intercept Correction.36 CALIBRATION ACCEPTANCE CRITERIA (<=) TRANSFER STANDARD AMBIENT ANALYZER STATION REFERENCE Mean Absolute Difference (%) 3% Thermo Thermo Thermo Maximum Absolute Difference (%) 3% Linearity % of Fullscale % 49i 49i 49i DATA ACCEPTANCE CRITERIA (<=) Mean Absolute Difference (%) % Coefficient Coefficient.8 Coefficient. Maximum Absolute Difference (%) % Background Background -.2 Background Linearity % of Fullscale 2% Cell A Intensity (Hz) Cell A Intensity (Hz) 7375 Cell A Intensity (Hz) 8622 Cell B Intensity (Hz) 858 Cell B Intensity (Hz) 7988 Cell B Intensity (Hz) Transfer Standard Correction Factors.987 Flow A (lpm).733 Flow A (lpm).77 Flow A (lpm).674 Flow B (lpm).763 Flow B (lpm).697 Flow B (lpm).669 Full Scale (ppb) 25 Pressure (mmhg) Pressure (mmhg) Pressure (mmhg) Linear Regression Plot Bench Temp ( C) 33.4 Bench Temp ( C) 34.7 Bench Temp ( C) 35 Bench Lamp Temp ( C) 54.3 Bench Lamp Temp ( C) 53.7 Bench Lamp Temp ( C) 53.9 Ozone Lamp Temp ( C) 68.5 Ozone Lamp Temp ( C) 67.3 Ozone Lamp Temp ( C) Ozone Source X Ozone Source Transfer Standard Level LEVEL 2 Transfer Standard Level LEVEL 3 POINT TARGET O3 LAMP TRANSFER STANDARD AMBIENT ANALYZER STATION REFERENCE % or mv Display Corrected DAS Diff %Diff Linearity % DAS Diff %Diff Linearity % ZERO N/A.% N/A.% %.% %.% % -.% %.% %.% %.% %.% 7.7..%.% %.% % -.% ZERO N/A -.2% N/A.% Mean ABS % Diff.7% Mean ABS % Diff.4% Max ABS % Difference.9% Max ABS % Difference.6% Analog Test Analyzer DAS Reference DAS Slope.4 Slope.7 Zero.7 Full Scale Y-Intercept.3 Y-Intercept -.3 Correlation. Correlation Residuals Plot Residual (ppb) Analyzer DAS (ppb) SLOPE INT y =.26x Transfer Standard (ppb) R² = Regression Concentration Value (ppb) NOTES:

15 CALIBRATION AND VERIFICATION STANDARDS ABBR. CANY-IS SITE NAME Network type CLIENT NPS Island in the Sky NPS FIELD SPECIALIST M. Farinacci DATE DATE OF LAST VISIT /29/26 5/4/26 MANUFACTURER MODEL SERIAL # Calibration Expiration Date Ozone Transfer Standard Thermo 49i /28/27 MFC High Flow Reference BIOS Definer 22-H /8/27 MFC Low Flow Reference Temperature Reference Eutechnics /8/27 AT/RH Sensor Reference Vaisala HMP55 G2837 4/5/27 Barometric Pressure Reference Wind Speed Reference (high rpm) RM Young 882 CA /8/27 Wind Speed Reference (low rpm) Wind Speed Torque Gauge RM Young 83 n/a Wind Direction Alignment Reference Brunton Transit Wind Direction Linearity Reference Wilson Machining 8 pt jig 85 Wind Direction Torque Gauge RM Young 83 n/a Solar Radiation Reference Multiplier 2. Licor Pyranometer PY /5/27 UV Radiation Reference Multiplier Precipitation Reference Volume 94 ml Texas Electronics FC-525 n/a Voltage Measurement Reference Voltage Source PM Flow Standard # PM Flow Standard #2 PM Flow Standard #3 PM Flow Standard #4 PM Temperature Standard # PM Temperature Standard #2 PM Temperature Standard #3 PM Temperature Standard #4 PM Barometric Pressure Standard # PM Barometric Pressure Standard #2 PM Barometric Pressure Standard #3 PM Barometric Pressure Standard #4 TEOM MTV Standard HiVol Direct Flow Standard HiVol Orifice Plate Orifice Manometer Stagnation Manometer

16

17 Calibration Certificate CertificateNo Sold To: Air Resource Specialists, Inc. Product 2-22H Definer 22 High Flow 9 Sharp Point Drive Serial No Fort Collins, CO 8525 Cal. Date 8-Feb-26 US All calibrations are performed at Mesa Laboratories, Inc., Park Place, Butler, NJ, 745, an ISO 725:25 accredited laboratory through NVLAP of NIST. This report shall not be reproduced except in full without the written approval of the laboratory. Results only relate to the items calibrated. This report must not be used to claim product certification, approval, or endorsement by NVLAP, NIST, or any agency of the Federal Government. As Received Calibration Data Lab. Pressure 757 mmhg Technician Lilianna Malinowska Lab. Temperature 22.2 C Instrument Reading Lab Standard Reading Deviation Allowable Deviation As Received sccm sccm -.%.% Out of Tolerance sccm sccm -.%.% Out of Tolerance sccm sccm -.%.% Out of Tolerance 2.4 C 2.4 C - ±.8 C In Tolerance 757 mmhg 757 mmhg - ± 3.5 mmhg In Tolerance Mesa Laboratories Standards Used Description Standard Calibration Date Calibration Due Date ML Nov Nov-26 Percision Thermometer Sep-25 2-Sep-26 Precision Barometer Jun Jun-26 Mesa Laboratories Inc. Park Place Butler, NJ 745 USA (973) FAX (973) Symbol MLAB on the NAS of 2 CAL2-48 Rev G5

18 As Shipped Calibration Data Certificate No Lab. Pressure 766 mmhg Technician Lilianna Malinowska Lab. Temperature 22.2 C Instrument Reading Lab Standard Reading Deviation Allowable Deviation As Shipped sccm sccm.79%.% In Tolerance sccm 5.8 sccm.54%.% In Tolerance 5.3 sccm 5.65 sccm.58%.% In Tolerance 22.2 C 22.2 C - ±.8 C In Tolerance 766 mmhg 766 mmhg - ± 3.5 mmhg In Tolerance Mesa Laboratories Standards Used Description Standard Calibration Date Calibration Due Date ML Nov Nov-26 Percision Thermometer Sep-25 2-Sep-26 Precision Barometer Jun Jun-26 Calibration Notes The expanded uncertainty of flow, temperature, and pressure measurements all have a coverage factor of k = 2 for a confidence interval of approximately 95%. Flow testing is in accordance with our test number PR8-3 with an expanded uncertainty of.8% using high-purity nitrogen or filtered laboratory air. Flow readings in sccm are performed at STP of 2. C and 76 mmhg. Pressure testing is in accordance with our test number PR8- with an expanded uncertainty of.6 mmhg. Temperature testing is in accordance with our test number PR8-2 with an expanded uncertainty of.4 ºC. Traceability to the International System of Units (SI) is verified by accreditation to ISO/IEC 725 by NVLAP under NVLAP Code Technician Notes: Louis Guido, Chief Metrologist Mesa Laboratories Inc. Park Place Butler, NJ 745 USA (973) FAX (973) Symbol MLAB on the NAS 2 of 2 CAL2-48 Rev G5

19

20

21

22

Semiannual Maintenance and Calibration Report Prepared by Air Resource Specialists, Inc.

Semiannual Maintenance and Calibration Report Prepared by Air Resource Specialists, Inc. Semiannual Maintenance and Calibration Report Prepared by Air Resource Specialists, Inc. Client: National Park Service Field Personnel: Mike Slate Site: Canyonlands National Park (CANY-IS) Service Date(s):

More information

Semiannual Maintenance and Calibration Report Prepared by Air Resource Specialists, Inc.

Semiannual Maintenance and Calibration Report Prepared by Air Resource Specialists, Inc. Semiannual Maintenance and Calibration Report Prepared by Air Resource Specialists, Inc. Client: National Park Service Field Personnel: Jake Zaragoza Site: Big Bend NP (BIBE-KB) Service Date(s): 2/21/217

More information

Semiannual Maintenance and Calibration Report Prepared by Air Resource Specialists, Inc.

Semiannual Maintenance and Calibration Report Prepared by Air Resource Specialists, Inc. Semiannual Maintenance and Calibration Report Prepared by Air Resource Specialists, Inc. Client: National Park Service Field Personnel: Will Yahr Site: Big Bend NP - (BIBE-KB) Service Date(s): 7/26/217

More information

Title: Standard Operating Procedure for Measurement of Total Hydrocarbon Using a Flame Ionization Detection

Title: Standard Operating Procedure for Measurement of Total Hydrocarbon Using a Flame Ionization Detection Procedure No: SOP-020 Revision No: 1.0 (January 21, 2011) Page No.: 1 of 7 1. INTRODUCTION AND SCOPE To obtain timely data for the purpose of air quality assessment, air quality trend reporting, and to

More information

Air Resource Specialists, Inc.

Air Resource Specialists, Inc. Air Resource Specialists, Inc. Optec NGN-2A Nephelometer Support Systems Price List July 2006 FULLY INTEGRATED OPTEC NGN-2A MONITORING AND DATA LOGGING SUPPORT SYSTEMS (costs do not include the Optec NGN-2A

More information

Model OEM-2 INSTRUCTIONS FOR USE

Model OEM-2 INSTRUCTIONS FOR USE ECO SENSORS, INC. 3-03.2 OEM OZONE CONTROLLER Model OEM-2 INSTRUCTIONS FOR USE GENERAL The model OEM-2 is a system to control ozone generators and alarms based on an adjustable ozone concentration set

More information

SIEMENS ULTRAMAT 6 IR CARBON MONOXIDE ANALYZER METHOD VALIDATION FOR TESTING CARBON MONOXIDE IN NITROGEN, NF

SIEMENS ULTRAMAT 6 IR CARBON MONOXIDE ANALYZER METHOD VALIDATION FOR TESTING CARBON MONOXIDE IN NITROGEN, NF 2015 SIEMENS ULTRAMAT 6 IR CARBON MONOXIDE ANALYZER METHOD VALIDATION FOR TESTING CARBON MONOXIDE IN NITROGEN, NF FINAL APPROVAL REVISION #: 0 ISSUE DATE: December 4, 2015 PAGE: 1 of 13 This Methods

More information

FORM ACCL-PC1 ACCL WITNESS TEST PROCEDURE AND CHECKLIST

FORM ACCL-PC1 ACCL WITNESS TEST PROCEDURE AND CHECKLIST FORM ACCL-PC1 ACCL WITNESS TEST PROCEDURE AND CHECKLIST AHRI CERTIFICATION PROGRAM FOR AIR-COOLED WATER-CHILLING PACKAGES USING THE VAPOR COMPRESSION CYCLE ACCL Participant: Date of Test: Witness Test

More information

MP 601A Humidity Probe INSTRUCTION MANUAL

MP 601A Humidity Probe INSTRUCTION MANUAL MP 601A Humidity Probe INSTRUCTION MANUAL CONTENTS DESCRIPTION... 3 OPERATION... 3 WIRING... 5 INSTALLATION... 7 MAINTENANCE...7 CALIBRATION BASICS... 8 CALIBRATION PROCEDURE... 9 SPECIFICATIONS... 11

More information

Test Report. Appliance EMF Testing

Test Report. Appliance EMF Testing Project: 11014506A Date: November 6, 2015 Model: 120V_WWF13NW Test Report On Appliance EMF Testing Copyright 2005 UL LLC UL LLC authorizes the above-named company to reproduce this Report provided it is

More information

Model 1062 Environment Monitor

Model 1062 Environment Monitor Model 1062 Environment Monitor Continuously Monitors Environment Temperature Humidity Barometric Pressure Line Voltage Line Frequency Applications Manufacturing The 1062A Environment Monitor will ensure

More information

Understanding total measurement uncertainty in power meters and detectors

Understanding total measurement uncertainty in power meters and detectors Understanding total measurement uncertainty in power meters and detectors Jay Jeong, MKS Instruments. Inc. INTRODUCTION It is important that users of calibrated power meters and detectors understand and

More information

QUALITY ASSURANCE/QUALITY CONTROL DOCUMENTATION SERIES OPTICAL DATA REDUCTION AND VALIDATION (IMPROVE PROTOCOL) STANDARD OPERATING PROCEDURE

QUALITY ASSURANCE/QUALITY CONTROL DOCUMENTATION SERIES OPTICAL DATA REDUCTION AND VALIDATION (IMPROVE PROTOCOL) STANDARD OPERATING PROCEDURE QUALITY ASSURANCE/QUALITY CONTROL DOCUMENTATION SERIES TITLE OPTICAL DATA REDUCTION AND VALIDATION (IMPROVE PROTOCOL) TYPE STANDARD OPERATING PROCEDURE NUMBER 4400 DATE NOVEMBER 2004 AUTHORIZATIONS TITLE

More information

AutomationDirect PM24. Microprocessor - Based Process/Temperature Limit Controller. Operator s Manual. Technical Support.

AutomationDirect PM24. Microprocessor - Based Process/Temperature Limit Controller. Operator s Manual. Technical Support. AutomationDirect 1/16 DIN Series Technical Support We strive to make our manuals the best in the industry. We rely on your feedback to let us know if we are reaching our goal. If you cannot find the solution

More information

C-NRPP Quality Assurance Guidance for Radon Test Devices

C-NRPP Quality Assurance Guidance for Radon Test Devices C-NRPP Quality Assurance Guidance for Radon Test Devices Table of Contents: 1. Purpose 2. Definitions 3. Quality Assurance Plan 4. Calibration 5. Spikes/Performance Tests 6. Duplicates 7. Blanks 8. Cross-Checks

More information

SimpleVIS Viscometer Reference Guide P

SimpleVIS Viscometer Reference Guide P SimpleVIS Viscometer Reference Guide P82.0090 Run Status Indicator Lamps Sample/Solvent Port Waste Bottle Dust Cover Temperature Calibration Port Touch Screen Display RS-232 Serial Port Exhaust Muffler

More information

Integrating Nephelometer - Visibility and Particulate Monitoring. Summary

Integrating Nephelometer - Visibility and Particulate Monitoring. Summary Summary This paper investigate the correlation between the Aurora 1000 integrating Nephelometer single wavelength green source 525 nm and the Thermo Fisher 1400A/B TEOM, Met-One 1020 BAM, and FRM sampler.

More information

CMT MANUAL 6. NMC SENSORS

CMT MANUAL 6. NMC SENSORS CMT MANUAL 6. NMC SENSORS TABLE OF CONTENTS INDEMNITY... 5 DOCUMENT INFORMATION... 5 ACKNOWLEDGEMENTS... 5 6.1 INTRODUCTION... 6 6.2 TEMPERATURE AND HUMIDITY SENSORS... 7 6.2.1. TEMPERATURE SENSOR... 7

More information

QUALITY ASSURANCE/QUALITY CONTROL DOCUMENTATION SERIES OPTEC NGN-2 NEPHELOMETER AUDIT PROCEDURES (IMPROVE PROTOCOL) STANDARD OPERATING PROCEDURE

QUALITY ASSURANCE/QUALITY CONTROL DOCUMENTATION SERIES OPTEC NGN-2 NEPHELOMETER AUDIT PROCEDURES (IMPROVE PROTOCOL) STANDARD OPERATING PROCEDURE QUALITY ASSURANCE/QUALITY CONTROL DOCUMENTATION SERIES TITLE OPTEC NGN-2 NEPHELOMETER AUDIT PROCEDURES (IMPROVE PROTOCOL) TYPE STANDARD OPERATING PROCEDURE NUMBER 4700 DATE JANUARY 1995 AUTHORIZATIONS

More information

HOBO U30 Remote Monitoring Systems

HOBO U30 Remote Monitoring Systems onset HOBO U30 Remote Monitoring Systems Web-based environmental monitoring All electronics housed within a research-grade, tamperproof enclosure Setup is quick and easy with plug-and-play sensors GSM

More information

Refrigerator/Freezer Thermometer

Refrigerator/Freezer Thermometer The McKesson triple display digital thermometer simultaneously shows the current, minimum and maximum temperatures and updates continuously. An audible and visual alarm will trigger when the temperatures

More information

Using the Model 1500 Portable Chilled Mirror Hygrometer

Using the Model 1500 Portable Chilled Mirror Hygrometer APPLICATION NOTE 2016_0120 MODEL 1500 Rev C Using the Model 1500 Portable Chilled Mirror Hygrometer The Model 1500 is a portable, multi-function, optical chilled mirror hygrometer designed to accurately

More information

PRODUCT CONFORMITY CERTIFICATE

PRODUCT CONFORMITY CERTIFICATE PRODUCT CONFORMITY CERTIFICATE This is to certify that the GC 866 PID airtoxic (Model A73022) manufactured by: Chromatotec / airmotec 15, Rue d Artiguelongue Saint -Antoine 33240 Val de Virvée France has

More information

DE20 INSTRUCTION MANUAL

DE20 INSTRUCTION MANUAL DE20 INSTRUCTION MANUAL Now UL Listed! SET POINT DE20 SET YOUR DISTRIBUTOR: Chemical Distributors, Inc. 80 Metcalfe Street - Buffalo, NY 14206 Phone: 800-777-2436 716-856-2300 Fax: 716-856-7115 sales@cdibuffalo.com

More information

Emissioni convogliate in atmosfera: Sistemi di monitoraggio in continuo delle emissioni SME

Emissioni convogliate in atmosfera: Sistemi di monitoraggio in continuo delle emissioni SME Emissioni convogliate in atmosfera: Sistemi di monitoraggio in continuo delle emissioni SME Aspetti gestionali e verifiche periodiche (norma UNI EN 14181:2015) Walter Battaglia - Alessandro Migatta - Sandro

More information

Equipment specifications (D6)

Equipment specifications (D6) Equipment specifications (D6) This report has been produced for the Cyprus Research Promotion Foundation in fulfilment of deliverable number D6 of the project Advanced Photovoltaics Research and Testing

More information

Refrigerator/Freezer Thermometer

Refrigerator/Freezer Thermometer The McKesson triple display digital thermometer simultaneously shows the current, minimum and maximum temperatures and updates continuously. An audible and visual alarm will trigger when the temperatures

More information

573 Dew Point Mirror. Industrial Chilled Mirror Hygrometer. Typical applications:

573 Dew Point Mirror. Industrial Chilled Mirror Hygrometer. Typical applications: Industrial Chilled Mirror Hygrometer Integral frost/dew point measuring head Internal sample pump Optimal Response Injection System ForceFrost TM function Ice-Test user calibration verification User configurable

More information

SAMPLE SPECIFICATIONS CONTINUOUS MONITORING SYSTEM (Insitu)

SAMPLE SPECIFICATIONS CONTINUOUS MONITORING SYSTEM (Insitu) PART 1 - GENERAL A. RELATED DOCUMENTS SAMPLE SPECIFICATIONS CONTINUOUS MONITORING SYSTEM (Insitu) Federal, State, and local requirements as applicable. Attached Permit Attached drawings B. SUMMARY 1. Provide

More information

IQ/OQ Protocol Installation Qualification/ Operation Qualification

IQ/OQ Protocol Installation Qualification/ Operation Qualification IQ/OQ Protocol Installation Qualification/ Operation Qualification ZZ FreeZone Freeze Dry Systems (To be used with some FreeZone Systems manufactured prior to August, 2004- see model number chart inside)

More information

LI-830 and LI-850 Gas Analyzers. For Continuous Monitoring Applications

LI-830 and LI-850 Gas Analyzers. For Continuous Monitoring Applications LI-830 and LI-850 Gas Analyzers For Continuous Monitoring Applications Easy Operation The LI-830 CO2 and LI-850 CO2/H2O Analyzers are high-performance monitoring solutions that give accurate, stable readings.

More information

s ter k Hea y Bloc Dr

s ter k Hea y Bloc Dr Advanced Dry Block Heaters Exceptional temperature uniformity and stability Optional external temperature probe Holds interchangeable modular blocks Designed for applications that require repeatable results

More information

Summary of Comments (Washington Revisions November 7, 2000) Update November 27, 2000

Summary of Comments (Washington Revisions November 7, 2000) Update November 27, 2000 SAE Alternate Refrigerant Cooperative Research Program Summary of Comments (Washington Revisions November 7, 2000) Update November 27, 2000 To: Alternate Refrigerant Task Force Members From: Ward Atkinson

More information

Contact person Date Reference Page Björn Folkeson P (15) Energy and bioeconomy

Contact person Date Reference Page Björn Folkeson P (15) Energy and bioeconomy Contact person Björn Folkeson 2016-06-27 6P04381 1 (15) Energy and bioeconomy +46 10 516 54 46 bjorn.folkeson@sp.se Thermal Integration Ltd 8 Curzon Road Sudbury Suffolk CO10 2XW Great Britain Testing

More information

Southern California Gas Company Customer Programs Department

Southern California Gas Company Customer Programs Department Appliance Test Report SCGAT140220A Revision 0 Issued Date: March 2014 Duke 613Q-G1V Gas Convection Oven Performance Test Per ASTM F1496-13 Southern California Gas Company Customer Programs Department Food

More information

QUALITY ASSURANCE AND SAFETY AT A TID RADIATION TEST LABORATORY

QUALITY ASSURANCE AND SAFETY AT A TID RADIATION TEST LABORATORY QUALITY ASSURANCE AND SAFETY AT A TID RADIATION TEST LABORATORY Seville, 31.03.2016 Pedro Martin, RadLab Radiation Facility Supervisor TÜV NORD GROUP QUALITY ASSURANCE AND SAFETY «Anything that can go

More information

MICHELL GROUP. Advanced Industrial Hygrometry and Process Instrumentation. Company Profile, 2011

MICHELL GROUP. Advanced Industrial Hygrometry and Process Instrumentation. Company Profile, 2011 MICHELL GROUP Advanced Industrial Hygrometry and Process Instrumentation Company Profile, 2011 Who is the Michell Group? Leading in the world of solutions for humidity and gas analysers More than 30 years

More information

LAB ID and/or LABORATORY NAME: ASSESSOR NAME: ALPHA TRACK. Method Number: ELAP method number SOP Number: Revision Number: SOP Date:

LAB ID and/or LABORATORY NAME: ASSESSOR NAME: ALPHA TRACK. Method Number: ELAP method number SOP Number: Revision Number: SOP Date: LAB ID and/or LABORATORY NAME: ASSESSOR NAME: DATE: ALPHA TRACK Method Number: ELAP method number 7035 SOP Number: Revision Number: SOP Date: Personnel / Data Records observed: Page 1 of 6 RADON IN AIR

More information

Inductive Conductivity

Inductive Conductivity Inductive Conductivity THORNTON Leading Pure Water Analytics 7100E Non-Fouling Sensors PEEK or PFA Wetted Parts Excellent Chemical Resistance 0/4 to 20mA Outputs for Conductivity & Tempreature 4 Relays

More information

ST710-KHJV.03. Wiring diagram. Product description. Temperature controller. Order number

ST710-KHJV.03. Wiring diagram. Product description. Temperature controller. Order number ST71-KHJV.3 Temperature controller Order number 921.8 Wiring diagram Product description The switching outputs of the thermal controller can be programmed as -two-point controller with alarm -three-point

More information

SITE REPORT FOR CARE (CANADA) (Submitted by Rodica Nitu) Summary and purpose of document

SITE REPORT FOR CARE (CANADA) (Submitted by Rodica Nitu) Summary and purpose of document WORLD METEOROLOGICAL ORGANIZATION COMMISSION FOR INSTRUMENTS AND METHODS OF OBSERVATION INTERNATIONAL ORGANIZING COMMITTEE (IOC) FOR THE WMO SOLID PRECIPITATION INTERCOMPARISON EXPERIMENT (SPICE) Fourth

More information

Technical-Report for SL Lamp 5 W

Technical-Report for SL Lamp 5 W Technical-Report for SL Lamp 5 W Technical Report No. 68.184.17.478.01 Dated 2017-06-27 Client: Manufacturing place: Name: PHOCOS China Ltd. Address: 3F 1st Bldg. 4A Chang Yi Road, 266021, Qingdao, PEOPLE'S

More information

RF TEST REPORT. ZTE Corporation SRQ-VFD510 ZTE LTE/WCDMA/GSM GPRS RXC RF07

RF TEST REPORT. ZTE Corporation SRQ-VFD510 ZTE LTE/WCDMA/GSM GPRS RXC RF07 RF TEST REPORT Applicant FCC ID Brand Product ZTE Corporation SRQ-VFD510 ZTE LTE/WCDMA/GSM GPRS Model VFD 511 Multi-Mode Digital Mobile Phone Report No. RXC1703-0087RF07 Issue Date May 24,2017 TA Technology

More information

LI-830 and LI-850 Gas Analyzers. For Continuous Monitoring Applications

LI-830 and LI-850 Gas Analyzers. For Continuous Monitoring Applications LI-830 and LI-850 Gas Analyzers For Continuous Monitoring Applications Easy Operation The LI-830 CO2 and LI-850 CO2/H2O Analyzers are high-performance monitoring solutions that give accurate, stable readings.

More information

LAB ID and/or LABORATORY NAME: ASSESSOR NAME: ELECTRET by Voltametry. Method Number: ELAP method number SOP Number: Revision Number: SOP Date:

LAB ID and/or LABORATORY NAME: ASSESSOR NAME: ELECTRET by Voltametry. Method Number: ELAP method number SOP Number: Revision Number: SOP Date: LAB ID and/or LABORATORY NAME: ASSESSOR NAME: DATE: ELECTRET by Voltametry Method Number: ELAP method number 7034 SOP Number: Revision Number: SOP Date: Personnel / Data Records observed: General LLD:

More information

SENTRY-MS Mono-static Multi-gas Open Path Analyzer

SENTRY-MS Mono-static Multi-gas Open Path Analyzer SENTRY-MS Mono-static Multi-gas Open Path Analyzer Air Monitoring Where You Need it... The Cerex Sentry-MS Multi-gas Open Path Analyzer is specifically designed for rapid deployment for a variety of indoor

More information

C-NRPP Quality Control and Quality Assurance Manual for Radon Sampling and Analysis conducted by Radon Measurement Professionals and Laboratories

C-NRPP Quality Control and Quality Assurance Manual for Radon Sampling and Analysis conducted by Radon Measurement Professionals and Laboratories C-NRPP Quality Control and Quality Assurance Manual for Radon Sampling and Analysis conducted by Radon Measurement Professionals and Laboratories October 2018 C-NRPP/PNCR-C 1 (855) 722-6777 c-nrpp.ca Canadian

More information

INSTRUCTION MANUAL. Pyrheliometer MS-57

INSTRUCTION MANUAL. Pyrheliometer MS-57 INSTRUCTION MANUAL Pyrheliometer MS-57 1. Index 1. Index 1 2. Important User Information 2 2-1. Contact Information 2 2-2. Warranty and Liability 2 2-3. About Operating Manual 3 2-4. Environment 3 2-5.

More information

Honeywell Zephyr TM Analog Airflow Sensors. HAF Series High Accuracy ±50 SCCM to ±750 SCCM. Datasheet

Honeywell Zephyr TM Analog Airflow Sensors. HAF Series High Accuracy ±50 SCCM to ±750 SCCM. Datasheet Honeywell Zephyr TM Analog Airflow Sensors HAF Series High Accuracy ±50 SCCM to ±750 SCCM Datasheet Honeywell Zephyr TM Analog Airflow Sensors HAF Series - High Accuracy Honeywell Zephyr HAF Series sensors

More information

HSD Dry Cabinet

HSD Dry Cabinet HSD 1106-52 Dry Cabinet Outstanding performance fast recovery time The HSD cabinet series has an outstanding performance for drying moisture sensitive components and pcb s as a result of a high performance

More information

REPORT 3933 US ROUTE 11 CORTLAND, NEW YORK 13045

REPORT 3933 US ROUTE 11 CORTLAND, NEW YORK 13045 REPORT 3933 US ROUTE 11 CORTLAND, NEW YORK 13045 FOR THE SCOPE OF ACCREDITATION UNDER NVLAP LAB CODE 100402-0. Project No. G101030158 Date: February 26, 2013 REPORT NO. 101030158CRT-001 TEST OF ONE LED

More information

STT 3000 Series STT250 Smart Temperature Transmitters Specifications Models STT25T EN0l-6091 February 2009

STT 3000 Series STT250 Smart Temperature Transmitters Specifications Models STT25T EN0l-6091 February 2009 STT 3000 Series STT250 Smart Temperature Transmitters Specifications Models STT25T EN0l-6091 February 2009 Introduction Honeywell s STT 3000 family of microprocessor based transmitters covers the full

More information

Our photo detectors are offered for both power or energy measurements. Measure as low as a few femtojoules in energy or a few picowatts in power.

Our photo detectors are offered for both power or energy measurements. Measure as low as a few femtojoules in energy or a few picowatts in power. PHOTO DETECTORS BEAM DIAGNOSTICS SPECIAL PRODUCTS OEM DETECTORS THZ DETECTORS PHOTO DETECTORS HIGH POWER SOLUTIONS POWER DETECTORS ENERGY DETECTORS MONITORS PRESENTATION OVERVIEW OF THE DIFFERENT MODELS

More information

IQ/OQ Protocol Installation Qualification/ Operation Qualification zz FreeZone Stoppering Tray Dryer & Bulk Tray Dryer Freeze Dry Systems

IQ/OQ Protocol Installation Qualification/ Operation Qualification zz FreeZone Stoppering Tray Dryer & Bulk Tray Dryer Freeze Dry Systems IQ/OQ Protocol Installation Qualification/ Operation Qualification zz FreeZone Tray Dryer & Bulk Freeze Dry Systems Labconco o: 1059501 Rev. - Purpose and Scope IQ and OQ This Qualification Protocol is

More information

I/A Series 716C 1/16 DIN Temperature Controller

I/A Series 716C 1/16 DIN Temperature Controller Product Specifications I/A Series 716C 1/16 DIN Temperature Controller PSS 2C-1B5 A The Foxboro 716C is a powerful compact, 1/16 DIN, microprocessor-based temperature controller that offers a variety of

More information

Datasheet: K-22 LO Sensor

Datasheet: K-22 LO Sensor Datasheet: K-22 LO Sensor The K- 22 LO is a CO2 sensor module designed to be built- in into stationary ventilation equipment, such as window vent or duct exhaust actuators, serving as a linear transmitter

More information

ST Wiring diagram. Product description. Four-stage controller. Order number

ST Wiring diagram. Product description. Four-stage controller. Order number ST96-35.16 Four-stage controller Order number 99.2 Wiring diagram Product description The four-stage controller with 4-digit setpoint and actual value display, 3 keys and 4 relays was developed for the

More information

Honeywell Zephyr Analog Airflow Sensors: HAF Series High Accuracy ±50 SCCM to ±750 SCCM

Honeywell Zephyr Analog Airflow Sensors: HAF Series High Accuracy ±50 SCCM to ±750 SCCM Short Port Fastener Mount Honeywell Zephyr Analog Airflow Sensors: HAF Series High Accuracy ±50 SCCM to ±750 SCCM Long Port Fastener Mount Short Port Snap Mount DESCRIPTION Honeywell Zephyr HAF Series

More information

Application of a Reagentless Direct Conductivity Online TOC Analyzer for Clean-in-Place Verification

Application of a Reagentless Direct Conductivity Online TOC Analyzer for Clean-in-Place Verification Application of a Reagentless Direct Conductivity Online TOC Analyzer for Clean-in-Place Verification By Paul Farrell, Emek Blair (Hach Company) and Torben Andersen (McFlusion) ABSTRACT This paper describes

More information

Temperature Controllers

Temperature Controllers Model TEC-4100 1/4 DIN Model TEC-4100 1/4 DIN Temperature Controller Ordering Code: Power Input BOX 1 4 = 90-250 VAC 5 = 11-26 VAC / VDC TEC-4100- Configurable for 4 Programmable Outputs and NEMA 4X/IP65

More information

SELECTION GUIDE SELECTION GUIDE. Pressure, Flow & Liquid Level Products for Aerospace & Military Uses

SELECTION GUIDE SELECTION GUIDE. Pressure, Flow & Liquid Level Products for Aerospace & Military Uses SELECTION GUIDE SELECTION GUIDE Pressure, Flow & Liquid Level Products for Aerospace & Military Uses P R E C I S I O N S E N S O R S Altitude Switches PRECISION SENSORS 50 Seemans Lane Milford, CT 06460

More information

ABB Instrumentation. Specifications. COMMANDER 150 Universal Process Indicator. the 1 /8 DIN indicator to match all your display requirements

ABB Instrumentation. Specifications. COMMANDER 150 Universal Process Indicator. the 1 /8 DIN indicator to match all your display requirements Specifications COMMANDER 150 Universal Process Indicator High visibility 6-digit LED display the clearest view of your process status 0.1% measurement accuracy precise indication of process measurement

More information

SDR Dry Room. Outstanding performance

SDR Dry Room. Outstanding performance SDR Dry Room Outstanding performance The SDR-51 dry room has an outstanding performance for drying moisture sensitive components, pcb s and other moisture sensitive products. The dynamic drying unit of

More information

Contact person Date Reference Page Daniel Månsson P07886E-rev2 1 (14) Energy and bioeconomy

Contact person Date Reference Page Daniel Månsson P07886E-rev2 1 (14) Energy and bioeconomy Contact person Daniel Månsson 2016-04-15 5P07886E-rev2 1 (14) Energy and bioeconomy +46 10 516 53 28 Daniel.Mansson@sp.se FairHeat Ltd, 10 Stoney Street, London SE1 9AD16 United Kingdom Testing of heat

More information

Determining Real Time Performance of Residential AC Systems. Presented at the RESNET Conference San Antonio, TX February 28, 2006

Determining Real Time Performance of Residential AC Systems. Presented at the RESNET Conference San Antonio, TX February 28, 2006 Determining Real Time Performance of Residential AC Systems Presented at the RESNET Conference San Antonio, TX February 28, 2006 Bill Spohn testo, Inc. testo Worldwide 100 90 80 70 60 50 40 30 20 10 0

More information

ENVIRONMENTAL MONITORING IN PREPARATION FOR THE INSTALLATION OF A GREEN ROOF

ENVIRONMENTAL MONITORING IN PREPARATION FOR THE INSTALLATION OF A GREEN ROOF ENVIRONMENTAL MONITORING IN PREPARATION FOR THE INSTALLATION OF A GREEN ROOF Andrew Martin School of Civil Engineering, Purdue University, West Lafayette, Indiana ABSTRACT Green roofs are becoming an increasingly

More information

Our photo detectors are offered for both power or energy measurements. Measure as low as a few femtojoules in energy or a few picowatts in power.

Our photo detectors are offered for both power or energy measurements. Measure as low as a few femtojoules in energy or a few picowatts in power. BEAM DIAGNOSTICS SPECIAL PRODUCTS OEM DETECTORS THZ DETECTORS PHOTO DETECTORS HIGH POWER SOLUTIONS POWER DETECTORS ENERGY DETECTORS MONITORS PRESENTATION OVERVIEW OF THE DIFFERENT MODELS Our photo detectors

More information

QUALITY ASSURANCE/QUALITY CONTROL DOCUMENTATION SERIES AUTHORIZATIONS TITLE NAME SIGNATURE

QUALITY ASSURANCE/QUALITY CONTROL DOCUMENTATION SERIES AUTHORIZATIONS TITLE NAME SIGNATURE QUALITY ASSURANCE/QUALITY CONTROL DOCUMENTATION SERIES TITLE SITE SELECTION FOR OPTICAL MONITORING EQUIPMENT (IMPROVE PROTOCOL) TYPE STANDARD OPERATING PROCEDURE NUMBER 4050 DATE FEBRUARY 1994 AUTHORIZATIONS

More information

Electro Optical Components, Inc Skylane Boulevard, Suite D, Santa Rosa, CA Toll Free: 855-EOC

Electro Optical Components, Inc Skylane Boulevard, Suite D, Santa Rosa, CA Toll Free: 855-EOC Electro Optical Components, Inc. 5464 Skylane Boulevard, Suite D, Santa Rosa, CA 95403 Toll Free: 855-EOC-6300 www.eoc-inc.com info@eoc-inc.com Genicom Co., Ltd. is one of leading manufacturers in the

More information

2017 Company Overview

2017 Company Overview 2017 Company Overview WHO IS SERVOFLO SERVOFLO IS A STOCKING DISTRIBUTOR PROVIDING TECHNICAL AND LOGISTICAL SERVICES FOR: Pressure sensors Mass Flow sensors Humidity sensors Temperature Sensors Micropumps

More information

Supertron Sensing Pte Ltd Certificate No. : LA C Blk 16 Kallang Place #07-04 Issue No. : 4 Singapore Date : 04 February 2015

Supertron Sensing Pte Ltd Certificate No. : LA C Blk 16 Kallang Place #07-04 Issue No. : 4 Singapore Date : 04 February 2015 Supertron Sensing Pte Ltd Certificate No. : LA-2011-0485-C Blk 16 Kallang Place #07-04 Issue No. : 4 Singapore 339156 Date : 04 February 2015 Page : 1 of 5 FIELD OF TESTING : Calibration and Measurement

More information

Certificate: / 29 April 2014

Certificate: / 29 April 2014 Test report: 936/21217455/A of 10 September 2013 Initial certification: 01 April 2014 Expiry date: 31 March 2019 Publication: BAnz AT 01 April 2014 B12, chapter I, No. 1.1 Approved application The tested

More information

I NTEGRATED S ENSOR S UITE WITH F AN-ASPIRATED R ADIATION SHIELD ADDENDUM, REV B. For Vantage Pro and Vantage Pro Plus

I NTEGRATED S ENSOR S UITE WITH F AN-ASPIRATED R ADIATION SHIELD ADDENDUM, REV B. For Vantage Pro and Vantage Pro Plus I NTEGRATED S ENSOR S UITE WITH F AN-ASPIRATED R ADIATION SHIELD ADDENDUM, REV B For Vantage Pro and Vantage Pro Plus The Vantage Pro Integrated Sensor Suite (ISS) with the Fan-Aspirated Radiation Shield

More information

Ultrasonic Clamp-On Flow Meter Testing

Ultrasonic Clamp-On Flow Meter Testing Ultrasonic Clamp-On Meter Testing Carried out for Micronics Ltd By Colin Judd Ultrasonic Clamp-On Meter Testing Carried out for: Micronics Ltd Unit B4 Business Centre Davies Way Loudwater High Wycombe

More information

CALIBRATION HOW IT WORKS

CALIBRATION HOW IT WORKS HIGH POWER SOLUTIONS POWER DETECTORS ENERGY DETECTORS At Gentec-EO, we understand that the essence of our business since over 45 years has been delivering accuracy. There are no half measures: it either

More information

UV Sentry. Cost Effective Perimeter & LDAR Monitoring. ...Precise, Species Specific Fugitive Emissions Quantification

UV Sentry. Cost Effective Perimeter & LDAR Monitoring. ...Precise, Species Specific Fugitive Emissions Quantification UV Sentry Open Path Multi-gas UVDOAS Analyzer Ammonia Benzene Carbon Disulfide Chlorine Cost Effective Perimeter & LDAR Monitoring The UV Sentry offers perimeter monitoring with among the lowest cost per

More information

Vibration tests on a TNAML Luminaire

Vibration tests on a TNAML Luminaire Sebert Trillingstechniek B.V. Report M13.001-P13.001 rev.01 Date 26 august 2013 Vibration tests on a TNAML Luminaire Client Technor Benelux B.V. Veersteeg 15 4212 LR Spijk The Netherlands Contact Person

More information

Model 5190-F Temperature/Humidity Probe

Model 5190-F Temperature/Humidity Probe Model 5190-F Temperature/Humidity Probe FAA APPROVED ECP234 2017 Mar 13 NOT FAA APPROVED User s Manual Rev. C All Weather Inc. 1165 National Drive Sacramento, CA 95834 USA 800.824.5873 www.allweatherinc.com

More information

TEST REPORT #65. Compressor Calorimeter Test of Refrigerant L-41-2 (R-447A) in a R-410A Scroll Compressor

TEST REPORT #65. Compressor Calorimeter Test of Refrigerant L-41-2 (R-447A) in a R-410A Scroll Compressor Air-Conditioning, Heating, and Refrigeration Institute (AHRI) Low-GWP Alternative Refrigerants Evaluation Program (Low-GWP AREP) TEST REPORT #65 Compressor Calorimeter Test of Refrigerant L-41-2 (R-447A)

More information

Simple, digital wireless reporting.

Simple, digital wireless reporting. Get Connected Get Notified Get TraceableLIVE Simple, digital wireless reporting. Monitor a wide range of conditions from high to ultra-low temperatures and humidity to barometric pressure and CO 2 anywhere

More information

INSTRUCTION MANUAL T119 1MN0110 REV. 0. operates with ISO9001 certified quality system

INSTRUCTION MANUAL T119 1MN0110 REV. 0. operates with ISO9001 certified quality system INSTRUCTION MANUAL 1MN0110 REV. 0 operates with ISO9001 certified quality system TECSYSTEM S.r.l. 20094 Corsico (MI) Tel.: +39-024581861 Fax: +39-0248600783 http://www.tecsystem.it R. 1.1 25/08/16 ENGLISH

More information

Contact person Date Reference Page Daniel Månsson P07886C-rev2 1 (14) Energy and bioeconomy

Contact person Date Reference Page Daniel Månsson P07886C-rev2 1 (14) Energy and bioeconomy Contact person Daniel Månsson 2016-04-15 5P07886C-rev2 1 (14) Energy and bioeconomy +46 10 516 53 28 Daniel.Mansson@sp.se FairHeat Ltd, 10 Stoney Street, London SE1 9AD16 United Kingdom Testing of heat

More information

ENVIRONMENTAL TEST REPORT

ENVIRONMENTAL TEST REPORT MECHANICAL TESTING 0839.04 Hermon Laboratories Ltd. Harakevet Industrial Zone, Binyamina 30500, Israel Tel. +972-4-6288001 Fax. +972-4-6288277 E-mail: mail@hermonlabs.com ENVIRONMENTAL TEST REPORT ACCORDING

More information

Calculated Energy Savings for Disabling Anti- Sweat Door Heat on Glass Display Case Doors

Calculated Energy Savings for Disabling Anti- Sweat Door Heat on Glass Display Case Doors Calculated Energy Savings for Disabling Anti- Sweat Door Heat on Glass Display Case Doors ET Project Number: Product Manager: Sachin Andhare Pacific Gas and Electric Company` Prepared By: Danielle Geers

More information

Page 2 of 5. SOP SF SOP for Water Bath Maintenance

Page 2 of 5. SOP SF SOP for Water Bath Maintenance Page 2 of 5 4.1. The SOP applies to all SSF staff and is intended to outline maintenance, calibration, and documentation procedures used within the laboratory to ensure that water baths perform with reliable

More information

BlueSpin Magnetic Hotplate Stirrer

BlueSpin Magnetic Hotplate Stirrer BlueSpin Magnetic Hotplate Stirrer BlueSpin LCD Digital 7 Inch Square Magnetic Hotplate Stirrer MS7-H550-Pro 7 inch square hotplate magnetic stirrers are widely used in chemical synthesis, physical and

More information

User Guide IM/AX4CO4 Rev. K. AX413, AX430, AX433, AX436 & AX438 Single and dual input analyzers for high level conductivity

User Guide IM/AX4CO4 Rev. K. AX413, AX430, AX433, AX436 & AX438 Single and dual input analyzers for high level conductivity User Guide IM/AX4CO4 Rev. K AX413, AX430, AX433, AX436 & AX438 Single and dual input analyzers for high level conductivity The Company We are an established world force in the design and manufacture of

More information

PQS Independent type-testing protocol

PQS Independent type-testing protocol PQS Independent type-testing protocol WHO/PQS/E03/RF05-VP.1 Original: English Distribution: General TITLE: Refrigerator or combined refrigerator-icepack freezer: compression-cycle. Solar direct drive without

More information

SAFETY CERTIFIED MODEL FP-700 COMBUSTIBLE GAS DETECTOR

SAFETY CERTIFIED MODEL FP-700 COMBUSTIBLE GAS DETECTOR SAFETY MANUAL SIL 2 Certified Model FP-700 Combustible Hydrocarbon Gas Sensor Version 2.0 1 SAFETY CERTIFIED MODEL FP-700 COMBUSTIBLE GAS DETECTOR This manual addresses the specific requirements and recommendations

More information

S304 and S305 Dust Emission Monitors. User Manual. Distributor

S304 and S305 Dust Emission Monitors. User Manual. Distributor S304 and S305 Dust Emission Monitors User Manual Distributor Version 6.4 30/5/2012 Table of Contents 1. INTRODUCTION... 3 1.1 Safety... 3 1.2 Product overview... 4 1.3 Principle of operation... 4 2. INSTALLATION...

More information

CALIBRATION HOW IT WORKS

CALIBRATION HOW IT WORKS HIGH POWER SOLUTIONS POWER DETECTORS ENERGY DETECTORS At Gentec-EO, we understand that the essence of our business since 40 years has been delivering accuracy. There are no half measures: it either measures

More information

02/11/2015

02/11/2015 MIC48 With RS 485 link Part number 89422418 Heating and / or cooling function 2 independent alarms Load break detection 2 setpoint which can be selected remotely Manual / automatic power adjustment RS

More information

MD-MINI Dry Bath. Instruction Manual

MD-MINI Dry Bath. Instruction Manual MD-MINI Dry Bath Instruction Manual Catalog No. MD-MINI www.majorsci.com service@majorsci.com Version 01B Revision on: 2013.07.15 Packing list MD-MINI: 1x MD-MINI Dry Bath 1x MD-MINI Lid 1x Power Cord

More information

Note To The User: This product has updated features not yet mentioned in the product manual.

Note To The User: This product has updated features not yet mentioned in the product manual. PLEASE READ BEFORE OPERATING THE PRODUCT--- Note To The User: This product has updated features not yet mentioned in the product manual. Celsius / Fahrenheit toggle: To switch the reading between Celsius

More information

Model 5190-E Temperature/Humidity Probe

Model 5190-E Temperature/Humidity Probe Model 5190-E Temperature/Humidity Probe User s Manual Rev. D All Weather Inc. 1165 National Drive Sacramento, CA 95834 USA 800.824.5873 www.allweatherinc.com Copyright 2010, All Weather, Inc. All Rights

More information

Quality Assurance/ Quality Control. Topic 4 - Audio 46. Key Points

Quality Assurance/ Quality Control. Topic 4 - Audio 46. Key Points Topic 4 - Audio 46 Key Points : Measures taken to increase confidence in results. Quality Assurance: Program that assures that measures are being done and in a timely manner. Precision: How close measurements

More information

Using the DewMaster Deluxe Dew Point/ Humidity Hygrometer

Using the DewMaster Deluxe Dew Point/ Humidity Hygrometer JAN 0 ::0 AM SERVOLOCK APPLICATION NOTE 0_0_ Rev Using the Deluxe Dew Point/ Humidity Hygrometer The Model is a versatile, multi-function, optical chilled mirror hygrometer designed to continuously measure

More information

Oxygen Deficiency Monitor and Oxygen Analyzers. Oxygen Sensors for 0-25% and ppm

Oxygen Deficiency Monitor and Oxygen Analyzers. Oxygen Sensors for 0-25% and ppm Oxygen Deficiency Monitor and Oxygen Analyzers Oxygen Sensors for 0-25% and 0-1000ppm ABOUT PUREAIRE MONITORING SYSTEMS PureAire is an experienced safety gas detection manufacturer providing long lasting

More information

HTO-45 HUMIDITY TEMPERATURE PROBE INSTRUCTION MANUAL

HTO-45 HUMIDITY TEMPERATURE PROBE INSTRUCTION MANUAL HTO-45 HUMIDITY TEMPERATURE PROBE INSTRUCTION MANUAL -2- CONTENTS DESCRIPTION...3 OPERATION...3 INSTALLATION (Tips for fixed probe installation)...4 MAINTENANCE...6 CALIBRATION...6 SPECIFICATIONS...8 PLEASE,

More information

Series Temperature Controller Instruction Sheet

Series Temperature Controller Instruction Sheet Series Temperature Controller Instruction Sheet Thank you very much for purchasing DELTA A Series. Please read this instruction sheet before using your A series to ensure proper operation and please keep

More information