DOD FUELS LABORATORY STANDARDS UFC
|
|
- Osborn Lawson
- 5 years ago
- Views:
Transcription
1 DOD FUELS LABORATORY STANDARDS UFC Prepared for: DLA Energy Worldwide Energy Conference April 12, 2017 Dustin Scheuffele, P.E. POL-MCX Omaha District US Army Corps of Engineers The views, opinions and findings c ontained in this report are those of the authors(s) and should not be construed as an offic ial Department of the Army pos ition, polic y or dec ision, unles s so des ignated by other offic ial doc umentation.
2 CONTENT Introduction Basic Lab Type A Lab Type B Lab Type C Lab Type C++ Lab References Questions 2
3 INTRODUCTION Document *DRAFT* UFC , DOD FUELS LABORATORY STANDARDS Dated 19 Sep 2016; Developed by USACE POL-MCX on behalf of FFEP Description Provides general criteria and standard procedures for the design and construction of military land-based Laboratory Facilities which test liquid fuels. Reasons for Document Provide guidance for planning, facility design, and construction documentation involved in the construction of new Fuel Laboratory Facilities. Serve as guidance for modernization or expanding existing Fuel Laboratories. Impact The requirements in this UFC will provide for uniform requirements for construction of DOD Fuels Laboratories based on required fuel testing requirements. 3
4 CONTENT Introduction Basic Lab Type A Lab Type B Lab Type C Lab Type C++ Lab References Questions 4
5 BASIC LAB Architectural Requirements Floors Designed for spill control recessed or other Cabinets solid wood or metal construction Countertops 4 inch backsplash, resistant to fuels and other chemicals. Typical is epoxy resin. Confirm color with customer 5
6 BASIC LAB Mechanical Requirements HVAC Minimum outdoor air IAW ASHRAE 62.1; minimize turbulence around lab hood; controls integration between system and lab hood; slightly negative pressure relative to building Exhaust 24/7 from within 6 inches of lab floor Energy efficiency follow UFC Plumbing UFC ; supply 130 deg F water to sink for cleaning 6
7 BASIC LAB Electrical Requirements Primary power dedicated transformer; step down to 120/208 3-phase Generator MTS at a minimum Lab equipment power provide second transformer with 120/240 single phase secondary Lighting generally non-hazardous, but some locations may require explosion proof ratings Outlets and switches located above counter height, every 4 feet. Two 240V outlets at end of counters. Grounding IAW with NFPA 70. Grounding bar for personnel and equipment must be provided at entrances. Lightning protection IAW UFC and NFPA 780 7
8 BASIC LAB Fire Protection Requirements Lab walls minimum 1-hour fire rated and UL fire-resistant construction Storage rooms Class 1, Division 2, and Group D Lab buildings Industrial occupancy per NFPA 101; fire alarm and mass notification system; and IBC Type I or II. Base Fire Department Coordinate for reviews, inspections, and occupancy permits. 8
9 FUELS LABORATORY EQUIPMENT REFERENCE TABLE
10 CONTENT Introduction Basic Lab Type A Lab Type B Lab Type C Lab Type C++ Lab References Questions 10
11 TYPE A LAB Architectural Lab 2,000 3,200 sq. ft Chem. analysis 260 sq ft Non-hazardous storage 200 sq ft Hazardous storage 100 sq ft Filtration 250 sq ft Balance and flash 150 sq ft Admin 375 sq ft Support sq ft as required 11
12 TYPE A LAB Mechanical Type B requirements Electrical Type B requirements Fire Protection Basic lab requirements Class A per NFPA 45 12
13 TYPE A LAB 13
14 CONTENT Introduction Basic Lab Type A Lab Type B Lab Type C Lab Type C++ Lab References Questions 14
15 TYPE B LAB Architectural Lab 1,200 2,000 sq. ft Chem. analysis 260 sq ft Non-hazardous storage 200 sq ft Hazardous storage 100 sq ft Filtration 250 sq ft Balance and flash 150 sq ft Admin 375 sq ft Support sq ft as required 15
16 TYPE B LAB Mechanical Type C++ requirements Electrical Basic lab requirements Combustible gas monitors with alarms Some areas may require explosion proof treatment Fire Protection Basic lab requirements Class A/B per NFPA 45 16
17 TYPE B LAB 17
18 CONTENT Introduction Basic Lab Type A Lab Type B Lab Type C Lab Type C++ Lab References Questions 18
19 TYPE C LAB Architectural Basic lab requirements Mechanical Basic lab plus science fume hood Emergency eyewash Electrical Basic lab requirements Fire Protection Basic lab requirements Volume limits per NFPA 45 19
20 TYPE C LAB 20
21 CONTENT Introduction Basic Lab Type A Lab Type B Lab Type C Lab Type C++ Lab References Questions 21
22 TYPE C++ LAB Architectural Basic lab requirements Mechanical Basic lab plus lab fume hood Emergency eyewash and shower De-ionized water Electrical Basic lab requirements Fire Protection Basic lab requirements Volume limits per NFPA 45 22
23 CONTENT Introduction Basic Lab Type A Lab Type B Lab Type C Lab Type C++ Lab References Questions 23
24 REFERENCES AASHRAE STD 110, Method of Testing Performance of Laboratory Fume Hoods ASHRAE 62.1, Ventilation for Acceptable Indoor Air Quality ANSI/ASSE Z9.5 Laboratory Ventilation ANSI/ISEA Z358.1, American National Standard for Emergency Eyewash and Shower Equipment NFPA 1, Fire Code NFPA 10, Standard for Portable Fire Extinguishers NFPA 13, Standard for the Installation of Sprinkler Systems NFPA 24, Standard for the Installation of Private Fire Service Mains and Their Appurtenances 24 NFPA 30, Flammable and Combustible Liquids Code NFPA 45, Standard on Fire Protection for Laboratories Using Chemicals NFPA 70, National Electrical Code NFPA 72, National Fire Alarm and Signaling Code NFPA 101, Life Safety Code NFPA 780, Standard for the Installation of Lightning Protection Systems NFPA 400, Hazardous Materials Code 36 CFR 1191, ADA-ABA American with Disabilities Act and Architectural Barriers Act CFR ; Code of Federal Regulations, Personal Protective Equipment (PPE)
25 REFERENCES CFR ; Code of Federal Regulations, Portable Fire Extinguishers CFR 1910, Subpart D; Code of Federal Regulations, Walking-Working Surfaces CFR 1910, Subpart E; Code of Federal Regulations, Means of Egress DOD M, DOD Management of Bulk Petroleum Products, Natural Gas, and Coal MIL-STD-3004, Quality Assurance/Surveillance for Fuels, Lubricants and Related Products MIL-STD-3007, DOD Standard Practice for Unified Facilities Criteria and Unified Facilities Guide Specifications 25 UFC , DoD Building Code UFC , High Performance and Sustainable Building Requirements UFC , Heating, Ventilating, and Air Conditioning Systems UFC , Interior and Exterior Lighting Systems and Controls UFC , Lightning and Static Electricity Protection Systems UFC , Fire Protection Engineering for Facilities UFC , Design, Plumbing Systems UFC , Design: Petroleum Fuel Facilities UFC , Design and O&M: Mass Notification Systems
26 QUESTIONS Dustin Scheuffele, P.E. POL-MCX Omaha District Corps of Engineers
NFPA 45 NFPA 211 NFPA 1600
NFPA 45 NFPA 211 NFPA 1600 Prepared by: William E. Koffel, P.E., FSFPE Expertly Engineering Safety From Fire 1 Objectives Overview of the NFPA process Understand the scope and application of NFPA 45, NFPA
More informationUNIFIED FACILITIES CRITERIA: Criteria Program Security Engineering Criteria
UNIFIED FACILITIES CRITERIA: Security Engineering Criteria NAVFAC Northwest Presented by: Richard Cofer, P.E. Naval Facilities Engineering Command Atlantic Capital Improvements Business Line Engineering
More informationLaboratories and Hazardous Materials. Company Overview
Laboratories and Hazardous Materials Don E. Contois, P.E. Sam Sinapi Company Overview R.W. Sullivan Engineering (RWS) was established in 1945 and currently has 95 employees. RWS is a full service engineering
More informationLABORATORY EQUIPMENT. A. Section Accessibility for People with Disabilities. B. Section Plumbing Basic Materials and Methods
115000 LABORATORY EQUIPMENT PART 1: GENERAL 1.01 RELATED SECTIONS A. Section 013010 Accessibility for People with Disabilities B. Section 220500 Plumbing Basic Materials and Methods C. Section 224500 Safety
More informationSECTION LABORATORY EQUIPMENT CONSTRUCTION STANDARD PART 1: GENERAL Scope of Standard
PART 1: GENERAL 1.01 Scope of Standard A. This standard provides general guidance concerning the specific preferences of the University of Texas at Austin for safety equipment used in or in support of
More informationAsbestos Abate exposed asbestos hazard to personnel
Example of Potential Safety / Health Projects Asbestos Abate exposed asbestos hazard to personnel Applicable Policies, Standards, Instructions and Guidelines 29 CFR 1910.1001; 1926.1101; 1915.1001; 40
More informationFlammable Liquids!!!!!
Sample Written Program For Your Company For Flammable Liquids Provided By: P.O. Box 2136 Slidell, LA 70458 Phone: 985-781-1444 Fax: 985-718-4957 Email: info@se-safety.com Flammable Liquids Purpose Proper
More informationLaboratory Safety Audit
Laboratory Safety Audit Principal Investigator: Building: Room #: Auditor: Date: Lab Safety Plan 1. Lab safety plan available 2. Plan reviewed/evaluated annually 3. Current Lab Safety Guide available 4.
More informationObjectives. NFPA Dust Specific Documents. Combustible Dust Process Hazard Safety Overview of NFPA Standards for Combustible Dusts
Combustible Dust Process Hazard Safety Overview of NFPA Standards for Combustible Dusts Guy R. Colonna, PE Division Manager, NFPA Combustible Dust Symposium Detroit, MI September 20, 2011 1 Objectives
More informationFire Protection Systems Operation and Maintenance 3 PDH / 3 CE Hours
Fire Protection Systems Operation and Maintenance 3 PDH / 3 CE Hours United Facilities Criteria U. S. Department of Defense UFC 3-601-02 PDH Academy PO Box 449 Pewaukee, WI 53072 www.pdhacademy.com pdhacademy@gmail.com
More informationUNIFIED FACILITIES CRITERIA (UFC) HOST NATION FACILITIES IN SUPPORT OF MILITARY OPERATIONS
UNIFIED FACILITIES CRITERIA (UFC) HOST NATION FACILITIES IN SUPPORT OF MILITARY OPERATIONS APPROVED FOR PUBLIC RELEASE; DISTRIBUTION UNLIMITED UNIFIED FACILITIES CRITERIA (UFC) HOST NATION FACILITIES IN
More informationUniversity of California, Davis Laboratory Self-Inspection Checklist
University of California, Davis Laboratory Self-Inspection Checklist Principal Investigator/Laboratory Supervisor: Laboratories Reviewed: Date: Reviewer: Revised 11/2014 I. SAFETY PROGRAM ADMINISTRATON
More informationBUILDING/FIRE CODE REFERENCES 2006 EDITION
BUILDING/FIRE CODE REFERENCES 2006 EDITION This document provides a cross reference with the International Building Code and NFPA 1/Uniform Fire The IBC references the IFC in 97 Code sections. This document
More informationPart I Extraction Process Equipment
745 West Colfax Avenue Denver, CO 80204 p: 720.913.3474 f: 720.913.3596 www.denvergov.org/fire MARIJUANA EXTRACTION GUIDELINE FOR COMMERCIAL / LICENSED FACILITIES The information contained within this
More informationUnion Mission: Restaurant/ Office/ Dorm. Index: Union Mission: Restaurant/ Office/ Dorm. Codes and Project Information:
Union Mission: Restaurant/ Office/ Dorm Index: A1 Title page and Code Info A2 Code Info Interior Designer: A3 First floor plan; Restaurant A4 Second floor plan; Bar and Admin office of Art and Design A5
More informationUNIFIED FACILITIES GUIDE SPECIFICATIONS
USACE / NAVFAC / AFCEC / NASA UFGS-10 44 16 (May 2015) ------------------------ Preparing Activity: NASA Superseding UFGS-10 44 16 (May 2012) UNIFIED FACILITIES GUIDE SPECIFICATIONS References are in agreement
More informationLaboratory Safety Audit
Laboratory Safety Audit Principal Investigator: Building: Room #: Auditor: Date: Lab Safety Plan 1. Lab Safety Plan available 2. Plan reviewed/evaluated annually 3. Current Lab Safety Guide available Personnel
More informationRecommendations Table
Recommendations Table Management Recommendations Requiring Comment No Additional Comments Required Commander, U.S. Army Central C.1 and C.2 A.1, A.2, B.1, B.2, C.3, and C.4 ii DODIG-2016-139 (Project No.
More informationPDH & Professional Training
510 N. Crosslane Rd. Monroe, Georgia 30656 (770) 266-6915 fax (678) 643-1758 PDH & Professional Training PDH Storm, by Engineers Edge, LLC CONTENTS 1. INTRODUCTION 1.1 GENERAL CRITERIA 1.2 DESIGN PROCEDURE
More informationLaboratory Information
Laboratory Information Lab Information Department Room number and function Principal investigator (PI) PI telephone number PI email address Lab Safety contact person Lab Safety contact telephone number
More informationThe Cannabis Industry: Fire Department Requirements
The Cannabis Industry: Fire Department Requirements RVFD Standard 4.1 Rio Vista Fire Department 350 Main St. Rio Vista, CA. 94571 By: Jeff Armstrong, Fire Chief Updated December 2018 1 The purpose of this
More informationRichard B. Alpert, P.E. Senior FPE, Clark Nexsen
Richard B. Alpert, P.E. Senior FPE, Clark Nexsen Clark Nexsen is a Registered Provider with The American Institute of Architects Continuing Education Systems (AIA/CES). Credit(s) earned on completion of
More informationUNIFIED FACILITIES GUIDE SPECIFICATIONS
USACE / NAVFAC / AFCEC / NASA UFGS-21 21 03.00 10 February 2009) ---------------------------------- Preparing Activity: USACE Superseding UFGS-21 21 03.00 10 (April 2006) UNIFIED FACILITIES GUIDE SPECIFICATIONS
More informationUNIFIED FACILITIES GUIDE SPECIFICATIONS
USACE / NAVFAC / AFCEC / NASA UFGS-23 83 00.00 20 (April 2006) -------------------------------- Preparing Activity: NAVFAC Replacing without change UFGS-15768N (September 1999) UNIFIED FACILITIES GUIDE
More informationCHAPTER A-6 FIRE PROTECTION
CHAPTER A-6 FIRE PROTECTION 6.1 GENERAL 6.1.1 Scope 6.1.2 Purpose 6.1.3 Qualifications of Fire Prevention Engineer 6.1.4 Required Design 6.1.5 Critical Projects 6.2 APPLICABLE PUBLICATIONS 6.3 PRECONCEPT
More informationPermitting and Licensing for Hazardous Material Use and Storage
Page #: 1 of 7 Last Reviewed/Update Date: 4/28/11 Permitting and Licensing for Hazardous Material Use and Storage 1. Purpose / Background National Fire Protection Association (NFPA) codes, State of Massachusetts
More informationBUILDING 751-CEH-OFFICE RENOVATION- OVERALL ROOM DATA SHEET
ROOM DATA SHEET FACILITY COMPONENT BUILDING 751-CEH-OFFICE RENOVATION- OVERALL ROOM DATA SHEET Name of Building Organization or Department Net area Critical dimensions Hours of operation Users/Occupancy
More informationCENTRAL STATE HOSPITAL PLAN SAFE USE HANDLING AND STORAGE OF FLAMMABLE AND COMBUSTIBLE LIQUIDS
Page 1 of 6 CENTRAL STATE HOSPITAL PLAN SUBJECT: SAFE USE HANDLING AND STORAGE OF FLAMMABLE AND COMBUSTIBLE LIQUIDS ANNUAL REVIEW MONTH: RESPONSIBLE FOR REVIEW: APRIL CSH SAFETY DIRECTOR LAST REVISION
More informationWORKPLACE INSPECTION CHECKLIST FOR LABORATORY ENVIRONMENTS
WORKPLACE INSPECTION CHECKLIST FOR LABORATORY ENVIRONMENTS Standard OHS Workplace Inspection Checklist Review Date: Next Review: Inspection Site: Inspection Team: Contacts: Written Laboratory Safety Policies/Procedures/Records
More information2018 IFC Update. Based on the 2018 International Fire Code, (IFC ) Goal. Objectives. Content
Based on the 2018 International Fire Code, (IFC ) The International Fire Code (IFC ), establishes minimum regulations for fire safety. This handout will identify important changes in the IFC from 2015
More informationLABORATORY INSPECTION CHECKLIST
LABORATORY INSPECTION CHECKLIST Date of Inspection: Conducted by: Department: Building/Room #: Principal Investigator: PI Present? Yes No Type of Inspection: Semi-Annual (scheduled) Complaint Follow Up
More informationCommon Fire Code Violation Checklist. Listed are some of the most commonly found fire code violations and should not be considered as all inclusive.
Common Fire Code Violation Checklist Listed are some of the most commonly found fire code violations and should not be considered as all inclusive. Exterior Features All fire access roads signed per TFD
More informationNarrative Reports As Regulated by 780 CMR, Section 903.0
EXHIBIT 20 Narrative Reports As Regulated by 780 CMR, Section 903.0 903.1.1 - FIRE PROTECTION CONSTRUCTION DOCUMENTS a. Basis (methodology) of design for the protection of the occupancy and hazard for
More informationUNIFIED FACILITIES GUIDE SPECIFICATIONS
USACE / NAVFAC / AFCEC / NASA UFGS-28 20 02 (November 2008) ----------------------------- Preparing Activity: USACE Superseding UFGS-13729 (April 2006) UNIFIED FACILITIES GUIDE SPECIFICATIONS References
More informationPART 2. Basic Code Requirements - Semiconductor & Microelectronic Facilities using HPM s
PART 2 Basic Code Requirements - Semiconductor & Microelectronic Facilities using HPM s Occupancy Classifications Y A - Assembly Y B - Business Y E - Educational Y F - Factory/Industrial Y H - Hazardous
More informationNFPA 30 FLC-SWC First Revisions Page 1 of 10 11/6/2015 9:21 AM. First Revision No. 208-NFPA [ Section No ]
of 46 11/6/2015 9:21 AM First Revision No. 208-NFPA 30-2015 [ Section No. 9.3.3.1 ] 9.3.3.1 Storage of liquids shall not physically obstruct a means of egress. Submittal Date: Wed Sep 09 18:24:20 EDT 2015
More informationCHAPTER 8 ELECTRICAL
CHAPTER 8 ELECTRICAL 8.1 GENERAL... 3 8.2 REQUEST FOR PROPOSAL DEVELOPMENT... 3 8.3 DESIGN DEVELOPMENT... 3 8.3.1 CONCEPT (30%) DESIGN SUBMITTAL REQUIREMENTS... 3 8.3.1.1 Concept Design Analysis... 3 8.3.1.1.1
More informationSpecial Occupancy Requirements
Special Occupancy Requirements CHAPTER 2 0 AVIATION FACILITIES CHAPTER 2 1 DRY CLEANING CHAPTER 2 2 COMBUSTIBLE DUST-PRODUCING OPERATIONS Special Occupancies The objective of this module is to make the
More informationUNIFIED FACILITIES GUIDE SPECIFICATIONS
USACE / NAVFAC / AFCEC / NASA UFGS-11 44 00 (August 2017) ---------------------------- Preparing Activity: USACE Superseding UFGS-11 44 00 (January 2008) UNIFIED FACILITIES GUIDE SPECIFICATIONS References
More informationConsumer Fireworks Retail Sales (CFRS) Facility Requirements.
Consumer Fireworks Retail Sales (CFRS) Facility Requirements. Site Plan. The site plan for tents shall show the location of the tent on the site and indicate the minimum separation distances required by
More informationICC Update - Laboratory Issues Code Changes / Proposals / Trends of Interest 10/2005
ICC Update - Laboratory Issues Code Changes / Proposals / Trends of Interest 10/2005 Kenneth Kretchman, CIH, CSP Mgr, Industrial Hygiene NC State University Ken_Kretchman@ncsu.edu International Building
More information(h) Pesticide Application.
Fire Code Cannabis Amendment Section 13.20.066 is added to the Marysville Municipal Code to read in its entirety as follows: 13.20.066 Cannabis Growing, Processing, or Extraction Facilities (a) Application.
More informationUNIFIED FACILITIES GUIDE SPECIFICATIONS
USACE / NAVFAC / AFCEC / NASA UFGS-23 54 16.00 10 (April 2008) -------------------------------- Preparing Activity: USACE Superseding UFGS-23 54 16.00 10 (January 2008) UNIFIED FACILITIES GUIDE SPECIFICATIONS
More informationIntroduction. Program Description. Section C.20 University of South Carolina Safety Program Guide FLAMMABLE AND COMBUSTIBLE LIQUIDS
Section C.20 University of South Carolina Safety Program Guide FLAMMABLE AND COMBUSTIBLE LIQUIDS Introduction This Program Guide applies to the storage of flammable and combustible liquids. This guide
More informationUNIFIED FACILITIES GUIDE SPECIFICATIONS
USACE / NAVFAC / AFCEC / NASA UFGS-13 21 26 (February 2016) ----------------------------- Preparing Activity: NAVFAC Superseding UFGS-13 21 26 (April 2006) UNIFIED FACILITIES GUIDE SPECIFICATIONS References
More informationAPPUZO RANGE HOOD INSTALLATION
IMPORTANT SAFETY INFORMATION TO REDUCE THE RISK OF FIRE, ELECTRIC SHOCK, OR PERSONAL INJURY, PLEASE READ THE FOLLOWING: Use the range hood only in the manner intended by the manufacturer. If you have questions,
More informationNFPA 70E Electrical Safety & Arc Flash Training
NFPA 70E Electrical Safety & Arc Flash Training OSHA 1910.332(a) Training, 1910.332(a) Scope. The training requirements contained in this section apply to employees who face a risk of electric shock that
More informationNew Facilities. Minimum Requirements
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 Item Reference Violation General Requirements NFPA 1124-7.1.1 Retail sales of consumer fireworks in both new and existing buildings, structures, and
More information2008 NYC Fire Code. New Laboratory Standards. Operational & Maintenance Provisions effective July 2008
2008 NYC Fire Code New Laboratory Standards Operational & Maintenance Provisions effective July 2008 Design & Certificate of Fitness Requirements Effective July 1, 2009 History The new Fire Code legislation
More information1. Researchers are familiar with the location and contents of: The lab s Chemical Hygiene Plan YES NO
Safety Self-Inspection Checklist for Laboratories One Inspection form per Laboratory Room DATE: BUILDING: ROOM: INSPECTOR: Professor Inspector s Signature Professor s Signature: Answer each question by
More informationLaboratory Safety Laboratory Inspection Checklist
Laboratory Safety Laboratory Inspection Checklist Date Lab Information Department Principal investigator (PI) PI telephone number PI email address Building Lab room numbers Lab phone number Radiation Biosafety
More informationGUIDELINES FOR EMERGENCY PLANNING AND PREPAREDNESS. High-Hazard
GUIDELINES FOR EMERGENCY PLANNING AND PREPAREDNESS High-Hazard This document shall apply to all Group H, High-Hazard occupancies. GENERAL Scope Reporting of emergencies, coordination with emergency response
More informationCommon Building Code Misinterpretations
Common Building Code Misinterpretations Presented by: Ron Melucci, P.E. Technical Manager, FP Division Engineering Planning and Management, Inc. rbm@epm inc.com www.epm inc.com 508 875 2121 Engineering
More informationDESIGN AND CONSTRUCTION STANDARDS SPECIFICATION GUIDELINES
SECTION 10 00 00 - SPECIALTIES 10 28 12 - TOILET PARTITIONS: Owner Preferences: Solid core phenolic, polymer based solid surfacing, and solid plastic toilet compartment and screen systems. No Stainless
More informationWTAMU Academic Research Environmental Health and Safety (AR-EHS) Laboratory Inspection Checklist
WTAMU Academic Research Environmental Health and Safety (AR-EHS) Laboratory Inspection Checklist Date of Inspection: Location (room # & building): Conducted By: Laboratory Title: Principal Investigator/Supervisor:
More informationSmoke Management in High-Rise Structures
Smoke Management in High-Rise Structures Most modern building codes define a high-rise structure as a building greater than 75 feet in height from the lowest level of fire department vehicle access to
More informationHot Work Policy. Department of Risk Management. 1. Purpose. 2. Authority. 3. Scope
1. Purpose The purpose of the University of Denver Hot Work policy is to protect personnel and property from a fire caused by hot work. The DU Hot Work policy defines the responsibilities and requirements
More informationCHAPTER 8 ELECTRICAL
CHAPTER 8 0 8.1 GENERAL... 3 8.2 REQUEST FOR PROPOSAL DEVELOPMENT... 3 8.3 DESIGN DEVELOPMENT... 3 8.3.1 CONCEPT (30%) DESIGN SUBMITTAL REQUIREMENTS... 3 8.3.1.1 Concept Design Analysis... 3 8.3.1.1.1
More informationENVIRONMENTAL HORTICULTURE: AGRISCIENCE AND BIOTECHNOLOGY September 30, 2010
ENVIRONMENTAL HORTICULTURE: AGRISCIENCE AND BIOTECHNOLOGY September 30, 2010 I. PROGRAM REQUIREMENTS Refer to the project-specific Schedule of Spaces for student stations, square footage, and for any requirements
More informationSafety Checklist for Science Instruction, Preparation and Storage Area
Safety Checklist for Science Instruction, Preparation and Storage Area School: Date: Teacher(s): Room or Area: The Science Center of the Boston Public Schools would like to acknowledge both the Science
More informationAdv. 8.4 Guidelines for Biosafety Cabinets Rev
UNIVERSITY OF CINCINNATI Environmental University Health Services 556-4968 Health and Safety ADVISORY NO. 8.4: GUIDELINES FOR THE SELECTION, INSTALLATION AND USE OF BIOSAFETY CABINETS (INCLUDING ISSUES
More informationUFGS (August 2004) UNIFIED FACILITIES GUIDE SPECIFICATIONS
USACE / NAVFAC / AFCEC / NASA UFGS-28 31 49 (April 2006) -------------------------- Preparing Activity: NAVFAC Replacing without change UFGS-13856 (August 2004) UNIFIED FACILITIES GUIDE SPECIFICATIONS
More informationC. EWFs shall deliver the quality and quantity of water that is satisfactory for emergency washing purposes.
22 45 00 PART 1 - GENERAL 1.01 SCOPE A. Emergency Washing Facilities (EWFs) shall be readily available in the immediate work area for employees, students and visitors who may be at risk for exposure to
More informationCombustible Dust Hazard Recognition and Control NFPA Standards for Combustible Dusts
Combustible Dust Hazard Recognition and Control NFPA Standards for Combustible Dusts Guy R. Colonna, PE Division Manager, NFPA Pellet Fuels Institute Conference Asheville, NC July 20, 2010 1 Objectives
More informationOpen to Below: A Discussion of Allowed Floor Openings. Jay A. Ierardi, Ph.D., P.E., LEED AP. Kevin S. Hastings, P.E., LEED AP
Open to Below: A Discussion of Allowed Floor Openings Kevin S. Hastings, P.E., LEED AP Jay A. Ierardi, Ph.D., P.E., LEED AP R.W. Sullivan Engineering (617) 523-8227, Fax (617) 523-8016 http://www.rwsullivan.com
More informationLaboratory Inspection (version 3) Criteria Action Plan
Laboratory Inspection (version 3) Criteria Action Plan Auditor: Jane Knipe Assigned Date: 7 Aug 2015 Audit Date: - Closure Date: - Comments: Category: Entity Workplace Inspections The University of Adelaide
More informationFIRE SAFETY DURING CONSTRUCTION, ALTERATION AND DEMOLITION
CHAPTER 14 FIRE SAFETY DURING CONSTRUCTION, ALTERATION AND DEMOLITION SECTION FC 1401 GENERAL 1401.1 Scope. This chapter shall govern fire safety measures during the construction, alteration, or demolition
More informationFluid Collections at the Smithsonian. Michael Kilby, P.E. Joshua Stewart, P.E. Office of Safety, Health & Environmental Management (OSHEM)
Fluid Collections at the Smithsonian Michael Kilby, P.E. Joshua Stewart, P.E. Office of Safety, Health & Environmental Management (OSHEM) Who We Are OSHEM Authority Having Jurisdiction for Smithsonian
More informationOUTLINE FOR FIRE PREVENTION ACTIVITY
OUTLINE FOR FIRE PREVENTION ACTIVITY Background Buildings under construction present unique fire exposures that require special attention. Fires can result in major losses and long delays in the completion
More informationAMENDMENTS TO THE BUILDING AND FIRE CODE FOR THE CITY AND COUNTY OF DENVER
AMENDMENTS TO THE BUILDING AND FIRE CODE FOR THE CITY AND COUNTY OF DENVER The 2016 Denver Building and Fire Code includes the following codes except as amendment herein. 2015 IBC (Building) 2015 IFC (Fire)
More informationSafety Tips for the Office
Safety Tips for the Office An office fire can be devastating - resulting in business interruption or permanent closure, loss of lives, jobs, property and information. Offices are loaded with combustible
More informationChapter 3: Emergency Planning for Laboratories
Chapter 3: Emergency Planning for Laboratories Every laboratory on campus must have an emergency preparedness plan for their specific laboratory. Many emergencies can happen from power outages to floods,
More informationMOBILE FOOD VENDING Food Trucks & Trailers
MCKINNEY FIRE DEPARTMENT FIRE MARSHAL S OFFICE MOBILE FOOD VENDING Food Trucks & Trailers These guidelines are to be followed when a business, facility or organization plans to operate a food truck or
More informationBuilding Self-Inspection Program
Building Self-Inspection Program www.tridentinsurance.net Lines of Business: Property, General Liability, Worker s Compensation, Public Official Liability Risk Control Strategy/Key Issues: Reduce physical
More informationMinimum Standards for Engineers Practicing Fire Protection Engineering in the State of Oklahoma September 14, 2016
1. Introduction The intent of this document is to establish minimum standards for the practice of fire protection engineering within the State of Oklahoma as conducted by licensed professional engineers.
More information2010 FIRE CODE OF NEW YORK STATE
2010 FIRE CODE OF NEW YORK STATE CHAPTER 14 FIRE SAFETY DURING CONSTRUCTION AND DEMOLITION SECTION 1401 GENERAL 1401.1 Scope. This chapter shall apply to structures in the course of construction, alteration,
More informationFIRE PREVENTION STANDARDS
DATE: 04/11/02 APPROVED: NUMBER: FP0-010 REVISED: FIRE PREVENTION STANDARDS TITLE: Large Family Day Cares (9-14 kids) PURPOSE: The purpose of this standard is to acquaint the applicant with the applicable
More informationFirst Aspen Federal Regulation Set: K LSC 2000 Health Existing
First Aspen Federal Regulation Set: K 01.01 LSC 2000 Health Existing K 0000 Title CFR Type INITIAL COMMENTS Memo K 0011 If the building has a common wall with a nonconforming building, the common wall
More informationHealth and Safety Plans Health and Safety Policy Statement Fire Prevention Plan Objective
Health and Safety Plans Health and Safety Policy Statement It is the policy of Prospect College to provide and maintain a safe environment free of injuries and other impairments to the health of the students,
More informationDistribution Restriction Statement
DEPARTMENT OF THE ARMY ETL 1110-3-485 U.S. Army Corps of Engineers CEMP-ET Washington, D.C. 20314-1000 Technical Letter No. 1110-3-485 15 October 1997 Engineering and Design FIRE PROTECTION FOR HELICOPTER
More information2012 International Fire Code Significant Changes. IFC Background. Purpose IBC and 2012 NFPA 101 Significant Means of Egress Changes
2012 International Fire Code Significant Changes Andrew M. Schneider, P.E. Senior Fire Protection Engineer aschneider@koffel.com Koffel Associates, Inc. www.koffel.com Expertly Engineering Safety From
More informationPlant Processing and Extraction Facilities
Orange County Fire Authority Community Risk Reduction 1 Fire Authority Road, Building A, Irvine, CA. 92602 www.ocfa.org 714-573-6180 Plant Processing and Extraction Facilities Guideline G-16 Serving the
More informationFIRE PREVENTION OFFICE Placer Hills Fire Protection District NEW PROJECT GUIDELINES
Established 1949 FIRE PREVENTION OFFICE Placer Hills Fire Protection District NEW PROJECT GUIDELINES The following are guidelines for new development. Any submitted plans must comply with these guidelines,
More informationLaser Class 4 Laser Type Nd-YAG Laser Manufacturer New Wave Research Laser Model SoloII-30 PIV Laser Serial / Id # Serial number 16582
THIS IS AN EXAMPLE OF A STANDARD OPERATING PROCEDURES RECOMMENDED OR REQUIRED FOR YOUR LABORATORY. YOU CAN USE THIS DOCUMENT AS A GUIDELINE. IF YOU HAVE ANY QUESTIONS, CONTACT CARINE VANPETEGHEM AT CVANPETE@UOTTAWA.CA
More informationQuestions/Comments for Richard Roux from Webinar on 7/11/13
Questions/Comments for Richard Roux from Webinar on 7/11/13 Below are the questions and comments from the July 11, 2013 webinar 2013 NFPA 72: Design and Installation Requirements for Mass Notification
More informationPurpose of the Act. Learning Objectives. Latest with DOJ. Latest with DOJ. Texas Registered Accessibility Specialists Association
Texas Registered Accessibility Specialists Association Mission Statement To enhance accessibility in the built environment throughout Texas To raise the level of service provided by our members by presenting
More informationHAZARDOUS CHEMICALS CHANGES TO HAZARD COMMUNICATION & CODE REQUIREMENTS FOR HEALTH CARE
HAZARDOUS CHEMICALS CHANGES TO HAZARD COMMUNICATION & CODE REQUIREMENTS FOR HEALTH CARE AGENDA 2 Global Harmonized System of Classification and Labelling of Chemicals (GHS) OSHA Communication Standard
More informationGUIDELINES FOR INSPECTING GROUND BASED SPARKLING DEVICE (GBSD) RETAIL SALES FACILITIES
OFFICE OF THE FIRE MARSHAL Worcester County GOVERNMENT CENTER - ONE WEST MARKET STREET - ROOM 1003 SNOW HILL, MARYLAND 21863-1194 TEL: 410-632-5666 FAX: 410-632-5664 GUIDELINES FOR INSPECTING GROUND BASED
More informationScientific Equipment & Furniture Association SEFA Recommended Practices For Liquid Chemical Storage Cabinets
Scientific Equipment & Furniture Association SEFA 11-2010 Recommended Practices For Liquid Chemical Storage Cabinets SEFA World Headquarters 1205 Franklin Avenue Ste 320 Garden City, NY 11530 Tel: 516-294-5424
More information1.2 To ensure the safety of all persons working in or occupying a building where a hot work operation is performed.
REQUIREMENT 2. HOT WORK OPERATIONS 1.0 PURPOSE 1.1 To establish a requirement with responsibilities for any hot work operations during all phases of construction, renovation and routine maintenance work
More informationThis policy is applicable to all DMS-owned, -leased and -managed facilities.
DEPARTMENT OF MANAGEMENT SERVICES Real Estate Development and Management TITLE: Open Flames & Flammable Materials in DMS Owned, -Leased and -Managed Facilities EFFECTIVE: March 9, 2009 REVISED: June 17,
More informationOBC INDUSTRIAL CLASSIFICATION OBOA ANNUAL MEETING AND TRAINING SESSIONS LONDON ON OCTOBER 4, 2011
OBC INDUSTRIAL MAJOR OCCUPANCY CLASSIFICATION OBOA ANNUAL MEETING AND TRAINING SESSIONS LONDON ON OCTOBER 4, 2011 Randal Brown & Associates Ltd. 2011 1 PRESENTERS RANDAL BROWN P.ENG., PE, BDS, FSPFE PRESIDENT
More informationLaboratory Safety: Engaging 600+ research groups
Laboratory Safety: Engaging 600+ research groups Stephanie Tumidajski Chemical Safety Professional Division of Research Safety sthess@illinois.edu 217-244-1360 University Overview University of Illinois
More informationNFPA Marine Fire Protection Codes and Standards An Overview. Guy R. Colonna, P.E. National Fire Protection Association. Quincy, MA USA
Proceedings of the ASME/USCG 2013 3rd Workshop on Marine Technology and Standards MTS2013 July 24-25, 2013, Arlington, VA, USA MTS2013-0321 NFPA Marine Fire Protection Codes and Standards An Overview Guy
More informationFIRE PROTECTION PROGRAM
Fire Protection Program Page 1 of 13 FIRE PROTECTION PROGRAM 1.0 Objective The objective of this performance assessment is to evaluate the effectiveness of the laboratory's fire protection program as implemented
More informationSECTION SPECIAL EXHAUST SYSTEMS. A. Laboratory Ventilation and Hazardous Exhaust Systems. 3. Chemical Fume Hoods and Storage Cabinets
SECTION 15852 SPECIAL EXHAUST SYSTEMS PART 1 GENERAL 1.01 SUMMARY A. Laboratory Ventilation and Hazardous Exhaust Systems 1. Pre-Design Considerations 2. General Design Criteria 3. Chemical Fume Hoods
More informationCHAPTER 6 [CE] REFERENCED STANDARDS
CHAPTER 6 [CE] REFERENCED STANDARDS User note: About this chapter: Chapter 6 lists the full title, edition year and address of the promulgator for all standards that are referenced in the code. The section
More informationUFGS (June 2006) UNIFIED FACILITIES GUIDE SPECIFICATIONS
USACE / NAVFAC / AFCESA / NASA UFGS-10 44 16 (May 2009) ------------------------- Preparing Activity: NASA Superseding UFGS-10 44 16 (June 2006) UNIFIED FACILITIES GUIDE SPECIFICATIONS References are in
More informationSTATE FIRE MARSHAL S REQUIRED FIRE PROTECTION SYSTEMS NARRATIVE REPORT
STATE FIRE MARSHAL S REQUIRED FIRE PROTECTION SYSTEMS NARRATIVE REPORT The interaction of fire protection systems can be extremely complex. The detection of a fire in a building by an automatic device
More informationHEALTH/LIFE SAFETY ANNUAL INSPECTION CHECKLIST (ROE)
HEALTH/LIFE SAFETY ANNUAL INSPECTION CHECKLIST (ROE) Administrative and General Building Requirements Safety reference plans (1) Fire alarm audibility (10) Alcohol hand-rub-dispensers (19) School safety
More information