1 Operating Instructions Analog or digital sensors with Memosens protocol For inductive measurement of conductivity in liquids BA00182C/07/EN/
2 Documentation information Warnings The structure, signal words and safety colors of the signs comply with the specifications of ANSI Z535.6 ("Product safety information in product manuals, instructions and other collateral materials"). Safety message structure! DANGER Cause (/consequences) Consequences if safety message is not heeded Corrective action! WARNING Cause (/consequences) Consequences if safety message is not heeded Corrective action! CAUTION Cause (/consequences) Consequences if safety message is not heeded Corrective action NOTICE Cause/situation Consequences if safety message is not heeded Action/note Meaning This symbol alerts you to a dangerous situation. Failure to avoid the situation will result in a fatal or serious injury. This symbol alerts you to a dangerous situation. Failure to avoid the situation can result in a fatal or serious injury. This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in minor or medium injury. This symbol alerts you to situations that can result in damage to property and equipment. Symbols used È ä 1 This symbol indicates a cross reference to a defined page (e.g. p. 1). È å 2 This symbol indicates a cross reference to a defined figure (e.g. fig. 2).
3 Table of contents 1 Basic safety instructions Requirements for personnel Designated use Occupational safety Operational safety Product safety Identification Type code for versions with Ex approval Nameplate Temperature classes for hazardous locations Certificates and approvals Installation Installation conditions Installation Post-installation check Wiring Connection to the transmitter Wiring diagram for explosion-hazardous areas 15 5 Commissioning Maintenance Spare parts and return Spare parts Return Technical data Input Performance characteristics Environment Process Mechanical construction Appendix Declaration of conformity for CLS50-G and CLS50D-BA Declaration of conformity for CLS50-V Endress+Hauser
4 Basic safety instructions 1 Basic safety instructions 1.1 Requirements for personnel Installation, commissioning, operation and maintenance of the measuring system must only be carried out by trained technical personnel. The technical personnel must be authorized by the plant operator to carry out the specified activities. The electrical connection may only be performed by an electrical technician. The technical personnel must have read and understood these Operating Instructions and must follow the instructions they contain. Measuring point faults may only be rectified by authorized and specially trained personnel. Repairs not described in the enclosed Operating Instructions may only be carried out directly at the manufacturer's or by the service organization. 1.2 Designated use conductivity sensors are especially suitable for application in the chemical industry and process engineering. The six-decade measuring range and the high chemical resistance of the materials in contact with medium (PFA or PEEK) permit to use this sensor in a number of various applications, e.g.: Concentration measurement of acids and bases Quality monitoring of chemical products in tanks and pipes Phase separation of product/product mixtures Digital CLS50D sensors are used with Liquiline CM44x or Liquiline CM42. Analog CLS50 sensors are used with the transmitters Liquiline CM42, Liquisys CLM223/253 or Mycom CLM153. Any other use than the one described here compromises the safety of persons and the entire measuring system and is not permitted. The manufacturer is not liable for damage caused by improper or non-designated use. NOTICE Use in not specified applications Measurement errors and failures up to the breakdown of the measurement point possible Only use the product acc. to it s specification. Note the technical data of the nameplate. 1.3 Occupational safety As the user, you are responsible for complying with the following safety conditions: Guidelines for explosion protection Installation instructions Local prevailing standards and regulations. 4 Endress+Hauser
5 Basic safety instructions 1.4 Operational safety Before commissioning the entire measuring point, make sure all the connections are correct. Ensure that electrical cables and hose connections are not damaged. Do not operate damaged products, and safeguard them to ensure that they are not operated inadvertently. Mark the damaged product as defective. If faults cannot be rectified, the products must be taken out of service and secured against unintentional commissioning. 1.5 Product safety State of the art The sensor has been designed and tested according to the state of the art and left the factory in perfect functioning order. Relevant regulations and European standards have been met Safety instructions for electrical equipment in hazardous locations Sensors with ATEX approval (CLS50-G and CLS50D-BA for Zone 0 and CLS50-V for Zone 2) have been developed and manufactured in compliance with applicable European standards and guidelines and are suitable for use in hazardous locations. The Declaration of Conformity confirms compliance with the harmonized European standards for using the sensors in hazardous locations. CLS50-G and CLS50D-BA The sensors may be operated in Zone 0 (1G) classified locations. The sensors may be operated in liquid media providing a conductivity of >10 ns/cm, only. The measurement cable must be protected from electrostatic charging when applied in Ex-Zone 0 (1G). CLS50-G The sensor may only be connected to the following transmitters: Mycom type CLM152-Z with transmitter module type FCL1, EC type-examination certificate DMT 99 ATEX E 076 Mycom type CLM153-G, EC type-examination certificate DMT 01 ATEX E 174 Liquiline type CM42-IG, EC type-examination certificate EX The maximum permitted length of the measuring cable is 55 m (180 ft). Endress+Hauser 5
6 Basic safety instructions CLS50D-BA CLS50D is a digital sensor with Memosens protocol. Its connection data are shown in the table below. The sensor may also be connected to the intrinsically safe Memosens connection of the FSDG1 module of the Liquiline type CM42-LG transmitter, EC type-examination certificate EX , provided the transmitter Liquiline M CM42-LG is connected to a galvanically isolated power supply. The maximum permitted length of the measuring cable is 100 m (330 ft). U i I i P i C i L i 5.1 V 130 ma 166 mw 18 μf 72 μh CLS50-V The sensor may be operated in Zone 2 (3G) classified locations. The sensor may only be connected to the following transmitter: Liquiline type CM42-IV, EC type-examination certificate BVS 08 ATEX E 093 The maximum permitted length of the measuring cable is 55 m (180 ft). Sensors with FM or CSA approval (CLS50-O, CLS50-S) Observe the documentation and the control drawings of the transmitter. General notes for all sensor types mentioned above Compliance with the ambient and medium temperature ranges indicated is mandatory to ensure safe operation of the device. Connection and operation of the sensor must be carried out in compliance with the instructions in the Operating Instructions of the sensor and of the transmitter connected. Observe all operating data of the sensor. Metallic process connection parts have to be mounted at the mounting location electrostatically conductive (<1 ΜΩ). Non metallic process connection parts have to be protected from electrostatic charging (even when applied in Ex zone 1 (2G)). Full compliance with regulations for electrical systems in hazardous locations (EN ) is mandatory when using the devices and sensors. Make sure the device is mounted correctly to maintain the degree of protection of the housing (use genuine seals, mount the cable entry correctly, tighten the nut)! The degree of protection only applies when the flange is mounted. 6 Endress+Hauser
7 Identification 2 Identification 2.1 Type code for versions with Ex approval Name Type Version Indumax CLS50 - G x x x for the use in hazardous locations, ATEX II 1G Ex ia IIC T4/T6 Ga Process connections, material no Ex relevance Name Type Version Indumax CLS50 - V x x x for the use in hazardous location, ATEX II 3G Ex nl IIC T4/T6 Gc Process connections, material no Ex relevance Name Type Version Indumax CLS50D - BA x x x x - x x for the use in hazardous location, ATEX II 1G Ex ia IIC T4/T6 Ga Process connections, material, cable connection, calibration, service no Ex relevance 2.2 Nameplate The nameplate can be found on the sensor. The following information is provided on the nameplate: Order code Extended order code Serial number Cell constant (nominal value) Protection class Pressure specification at 20 C Continuous service temperature Compare the data on the nameplate with your order. To find out what CLS50D version you have, enter the order code indicated on the nameplate in the search screen at the following address: Endress+Hauser 7
8 Identification 2.3 Temperature classes for hazardous locations CLS50-G Temperature class Sensor Ambient temperature range T a Medium temperature range T med T4 CLS50-G -20 C to +125 C -20 C to +125 C T6 CLS50-G -20 C to +75 C -20 C to +75 C CLS50D-BA Temperature class Sensor Ambient temperature range T a Medium temperature range T med T4 CLS50D-BA*D CLS50D-BA*C CLS50D-BA*B -20 C to +60 C -20 C to +60 C -20 C to +60 C -20 C to +110 C -20 C to +120 C -20 C to +120 C T6 CLS50D-BA*D CLS50D-BA*C CLS50D-BA*B -20 C to +60 C -20 C to +60 C -20 C to +60 C -20 C to +70 C -20 C to +70 C -20 C to +70 C The plant operator must guarantee the compliance with these temperatures by suitable installation measures CLS50-V Temperature class Sensor Ambient temperature range T a Medium temperature range T med T4 CLS50-V -20 C to +125 C -20 C to +125 C T6 CLS50-V -20 C to +75 C -20 C to +75 C 2.4 Certificates and approvals CE conformity Declaration of conformity The product meets the requirements of the harmonized European standards. It thus complies with the legal requirements of the EC directives. The manufacturer confirms successful testing of the product by affixing the 4 symbol ATEX approvals CLS50-G and CLS50D-BA II 1G Ex ia IIC T4/T6 Ga 8 Endress+Hauser
9 Installation CLS50-V II 3G Ex nl IIC T4/T6 Gc 3 Installation 3.1 Installation conditions Installation position Install the sensor in such a way that the sensor opening is oriented in the flow direction of the medium. The sensor head must be completely immersed in the medium. Fig. 1: Installation position of the sensor 1 Flow direction of medium 2 Minimum water level in the pipe a Sensor distance from the pipe wall a Endress+Hauser 9
10 Installation Installation factor In narrow installation conditions, the conductivity measurement is affected by the pipe walls. This effect is compensated by the so-called installation factor. The transmitter corrects the cell constant by multiplication with the installation factor. The value of the installation factor depends on the diameter and the conductivity of the pipe as well as the sensor s distance from the wall. If the distance from the wall is sufficient (a > 15 mm (0.59"), from DN 80), it is not necessary to consider the installation factor (f = 1.00). If the distance from the wall is smaller, the installation factor increases in case of electrically insulating pipes (f > 1) and decreases in case of electrically conductive pipes (f < 1). The installation factor can be measured using calibration solutions or it can be approximately determined from the following diagram. f a [mm] Fig. 2: Relationship between installation factor f and distance from wall a 1 Electrically conductive pipe wall 2 Insulating pipe wall a [inch] a Air set CLS50 To compensate residual coupling in the cable and between the two sensor coils, you must perform a zero calibration in air ("air set") before installing the sensor. For further information, refer to the Operating Instructions of your transmitter. CLS50D The digital sensor is already adjusted at the factory, an on-side compensation is not necessary. 10 Endress+Hauser
11 Installation 3.2 Installation Flange installation The sensor is suitable for installation in T-pieces DN 80 with the outgoing diameter reduced to DN 50.! WARNING Leakages Danger of injuries by leaking medium Tighten the nut with a torque of 20 Nm. To avoid leakages, regularly check the tightness of the nut Flange, not in contact with medium Fig. 3: Fixed flange, in contact with medium (order option "process connection": 3, 4) 1 2 Flange (stainless steel) Nut 3 4 O-ring Sensor a Endress+Hauser 11
12 Installation Flange, not in contact with medium Fig. 4: Fixed flange, not in contact with medium (order option "process connection": 5, 6, 7) Flange (stainless steel) Nut Sealing disk (GYLON) 4 5 O-ring Sensor a Lap-joint flange, not in contact with medium Fig. 5: Lap-joint flange, not in contact with medium (order option "process connection": A, B, C) Lap-joint flange (PP-GF) Nut (stainless steel) Flange (PVDF) 4 5 O-ring Sensor a Endress+Hauser
13 Installation Assembly installation Fig. 6: Installation of sensor with assembly 1 CLA111 with suspension bracket 2 CLA111 with flange connection 3 4 CLA140 with flange connection CYA611 a Post-installation check Are sensor and cable undamaged? Is the installation position correct? Arrow on the threaded sleeve shows the flow direction = installation position. Is the sensor installed via process connection and not suspended from the cable? Endress+Hauser 13
14 Wiring 4 Wiring! WARNING Device is energized Improper connection can cause injury or death. The electrical connection must only be carried out by a certified electrician. Technical personnel must have read and understood the instructions in this manual and must adhere to them. Prior to beginning any wiring work, make sure voltage is not applied to any of the cables. 4.1 Connection to the transmitter The sensor is supplied with a fixed cable. The connection to the transmitter can be extended using the CYK11 (CLS50D) or the CLK6 (CLS50) special measuring cable (not applicable for use in hazardous locations). Fig. 7: CYK11 for cable extension for CLS50D max. total cable length: 100 m (330 ft) a Fig. 8: CLK6 for cable extension for CLS50 max. total cable length: 55 m (180 ft) Please note that the residual coupling increases when the cable is extended. WH RD RD BU RD BU GN WH YE BN n.c. Pt 100 a Endress+Hauser
15 Wiring 4.2 Wiring diagram for explosion-hazardous areas Sensors for Zone 0 Fig. 9: Connection in hazardous location a Endress+Hauser 15
16 Wiring Fig. 10: Connection in hazardous location a Endress+Hauser
17 Wiring Sensors for Zone 2 EX ma EX II 3G Ex nl Zone 2 Fig. 11: Connection in hazardous location CLS50 a Endress+Hauser 17
18 Commissioning 5 Commissioning Before first commissioning, check if: the sensor is correctly installed the electrical connection is correct. If using an assembly with automatic cleaning, check the correct connection of the cleaning agent (e.g. water or air).! WARNING Incorrect connection of a cleaning unit to an assembly Danger of medium leaking off Before applying compressed air to an assembly with cleaning facility, make sure the connections are correctly fitted. Otherwise, the assembly may not be inserted into the process. 6 Maintenance! WARNING Burning chemicals Danger of chemicals burns to the eyes and skin. Danger of damage to clothing and equipment. It is absolutely essential to protect the eyes and hands properly when working with acids, bases and organic solvents! Wear protective goggles and safety gloves. Clean away splashes on clothes and other objects to prevent any damage. Pay particular attention to the information provided in the safety data sheets for the chemicals used. Clean away fouling on the sensor as follows depending on the particular type of fouling: Oily and greasy films: Clean with grease remover, e.g. alcohol, acetone, as well as hot water and dishwashing detergent if necessary. Lime and metal hydroxide buildup: Dissolve buildup with diluted hydrochloric acid (3 %) and then rinse thoroughly with plenty of clear water. Sulfidic buildup (from flue gas desulfurising or sewage treatment plants): Use a mixture of hydrochloric acid (3 %) and thiocarbamide (commercially available) and then rinse thoroughly with plenty of clear water. Buildup containing proteins (e.g. food industry): Use a mixture of hydrochloric acid (0.5 %) and pepsin (commercially available) and then rinse thoroughly with plenty of clear water. 18 Endress+Hauser
19 Spare parts and return 7 Spare parts and return 7.1 Spare parts Pos. no. Spare part kit Order no. 150, 255 Kit CHEMRAZ seal Nut (Pos. 150) CHEMRAZ O-ring, 2 pcs. (Pos. 255) 150, 250 Kit VITON seal Nut (Pos. 150) VITON O-ring, 3 pcs. (Pos. 250) Kit PTFE sealing disk for DN a Kit PTFE sealing disk for ANSI 2" and JIS 10K 50A 150, 280 Kit fixed flange DN 50, stainless steel (AISI 316L) Nut (Pos. 150) Flange DN 50 (Pos. 280) 150, 285 Kit fixed flange ANSI 2", stainless steel (AISI 316L) Nut (Pos. 150) Flange ANSI 2" (Pos. 285) 150, 286 Kit fixed flange JIS, stainless steel (AISI 316 L)Nut (Pos. 150) Flange DN 50 (Pos. 286) 150, 288, 292 Kit lap-joint flange ANSI 2", PVDF Nut (Pos. 150) Flange ANSI 2", PVDF (Pos. 288) Lap-joint flange, UP-GF (Pos. 292) 150, 287, 291 Kit lap-joint flange DN 50, PVDF Nut (Pos. 150) Flange DN 50, PVDF (Pos. 287) Lap-joint flange, UP-GF (Pos. 291) 150, 289, 293 Kit lap-joint flange JIS, PVDF Nut (Pos. 150) Flange JIS, PVDF (Pos. 289) Lap-joint flange, UP-GF (Pos. 293) a Endress+Hauser 19
20 Spare parts and return 7.2 Return The device must be returned if repairs or a factory calibration are required, or if the wrong device has been ordered or delivered. According to legal regulations, Endress+Hauser, as an ISO-certified company, is required to follow certain procedures when handling returned products that are in contact with medium. To ensure swift, safe and professional device returns, please read the return procedures and conditions on the internet site: 20 Endress+Hauser
21 Technical data 8 Technical data 8.1 Input Measured variable Conductivity Temperature Cell constant k = 1.98 cm Measuring frequency 2 khz Measuring ranges Conductivity: Temperature: 2 μs/cm to 2000 ms/cm (uncompensated) 20 to +180 C (-4 to +350 F) Temperature measurement CLS50D Pt 1000 (class A acc. to IEC 60751) CLS50 Pt 100 (class A acc. to IEC 60751) 8.2 Performance characteristics Maximum measured error -20 to 100 C (-4 to 212 F): ±(5 μs/cm % of the measured value) > 100 C (212 F): ±(10 μs/cm % of the measured value) Temperature response time PEEK versions: approx. 7 min PFA versions: approx. 11 min Endress+Hauser 21
22 Technical data 8.3 Environment Ambient temperature range CLS50D 10 to +60 C (+10 to +140 F) CLS50 10 to +70 C (+10 to +160 F) Storage temperature 20 to +80 C (-4 to +180 F) Ingress protection IP 68 / NEMA 6 (installed with original sealing) 8.4 Process Process temperature CLS50D Without flange DN50, ANSI 2" JIS PVDF lap-joint flange Sensor CLS50D-*1/2 CLS50D-*3/4/5/6/8 CLS50D-*7 CLS50D-*A/B/C PEEK 20 to 125 C (-4 to 260 F) 20 to 125 C (-4 to 260 F) 20 to 125 C (-4 to 260 F) 20 to 125 C (-4 to 260 F) PFA 20 to 110 C (-4 to 230 F) 20 to 125 C (-4 to 260 F) 20 to 125 C (-4 to 260 F) 20 to 125 C (-4 to 260 F) CLS50 Without flange DN50, ANSI 2" JIS PVDF lap-joint flange Sensor CLS50-*1/2 CLS50-*3/4/5/6/8 CLS50-*7 CLS50-*A/B/C PEEK 20 to 180 C (-4 to 360 F) 20 to 180 C (-4 to 360 F) 20 to 180 C (-4 to 360 F) 20 to 125 C (-4 to 260 F) PFA 20 to 125 C (-4 to 260 F) 20 to 125 C (-4 to 260 F) 20 to 125 C (-4 to 260 F) 20 to 125 C (-4 to 260 F) Process pressure max. 20 bar (290 psi), acc. to the sensor version, see pressure-temperature diagram 22 Endress+Hauser
23 Technical data Pressure-temperature diagram Fig. 12: Pressure-temperature-diagram of CLS50D 1 PEEK sensor, without flange 2 PFA sensor, without flange (blue line) 3 PEEK or PFA sensor, with DN50/ANSI 2" flange (red line) 4 PEEK or PFA sensor, with JIS flange 5 PEEK or PFA sensor, with lap-joint flange PVDF (green line) a Fig. 13: Pressure-temperature-diagram of CLS50 1 PEEK sensor, without flange 2 PFA sensor, without flange or with DN50/ANSI 2" flange (blue line) 3 PEEK sensor, with DN50/ANSI 2" flange (red line) 4 PFA sensor, with JIS flange (black line) 5 PEEK or PFA sensor, with lap-joint flange PVDF (green line) 6 PEEK sensor, with JIS flange (grey line) a Endress+Hauser 23
24 Technical data 8.5 Mechanical construction Weight approx. 650 g (1.43 lbs) Material Sensor: Sensor seal: Process connections: G ¾: Chemical durability PEEK, PFA (depending on ordered version) VITON, CHEMRAZ (depending on ordered version) CLS50-**A: stainless steel (AISI 316Ti) CLS50-**B/C: PEEK GF30 CLS50D-**D: stainless steel (AISI 316Ti) CLS50D-**B/C: PEEK GF30 PEEK NPT 1": Fixed flange: stainless steel (AISI 316 L) Sealing disk: GYLON (ceramic-filled PTFE) Lap-joint flange: PP-GF Flange combined with lap-joint flange: PVDF Medium Concentration PEEK PFA CHEMRAZ VITON Sodium hydroxide solution NaOH 0 to 50 % 20 to 100 C (68 to 212 F) not suitable 0 to 150 C (32 to 302 F) not suitable Nitric acid HNO 3 0 to 10 % 20 to 100 C (68 to 212 F) 20 to 80 C (68 to 176 F) 0 to 40 % 20 C (68 F) 20 to 60 C (68 to 140 F) 0 to 150 C (32 to 302 F) 0 to 150 C (32 to 302 F) 0 to 120 C (32 to 248 F) 0 to 120 C (32 to 248 F) Phosphoric acid H 3 PO 4 0 to 80 % 20 to 100 C (68 to 212 F) 20 to 60 C (68 to 140 F) 0 to 150 C (32 to 302 F) 0 to 120 C (32 to 248 F) Sulphuric acid H 2 SO 4 0 to 2.5 % 20 to 80 C (68 to 176 F) 20 to 100 C (68 to 212 F) 0 to 30 % 20 C (68 F) 20 to 100 C (68 to 212 F) 0 to 150 C (32 to 302 F) 0 to 150 C (32 to 302 F) 0 to 120 C (32 to 248 F) 0 to 120 C (32 to 248 F) Hydrochloric acid HCl 0 to 5 % 20 to 100 C (68 to 212 F) 0 to 10 % 20 to 100 C (68 to 212 F) 20 to 80 C (68 to 176 F) 20 to 80 C (68 to 176 F) 0 to 150 C (32 to 302 F) 0 to 150 C (32 to 302 F) 0 to 120 C (32 to 248 F) 0 to 120 C (32 to 248 F) No responsibility is taken for the correctness of this information. 24 Endress+Hauser
25 Appendix 9 Appendix 9.1 Declaration of conformity for CLS50-G and CLS50D-BA a Endress+Hauser 25
26 Appendix a Endress+Hauser
27 Appendix 9.2 Declaration of conformity for CLS50-V a Endress+Hauser 27
28 BA00182C/07/EN/15.12 FM9 / DT
Operating Instructions. Condumax CLS15/16/21. Analog sensors For conductive measurement of conductivity in liquids BA01148C/07/EN/01.
Operating Instructions Analog sensors For conductive measurement of conductivity in liquids BA01148C/07/EN/01.12 71195845 Documentation information Warnings The structure, signal words and safety colors
More informationOperating Instructions Indumax CLS54D
BA00508C/07/EN/02.16 71348190 Products Solutions Services Operating Instructions Conductivity sensor with inductive measurement of conductivity and hygienic design for applications in the food, beverage
More informationOperating Instructions Condumax CLS15/16/21
BA01148C/07/EN/02.14 71267400 Products Solutions Services Operating Instructions Condumax CLS15/16/21 Analog sensors For conductive measurement of conductivity in liquids 0 Table of contents Condumax CLS15/16/21
More informationCondumax CLS16 and CLS16D
Technical Information Condumax CLS16 and CLS16D Hygienic conductivity sensor, analog or digital with Memosens technology, cell constant k = 0.1 cm -1 Application Measurement in pure and ultrapure water:
More informationSECTION INDUCTIVE CONDUCTIVITY ANALYZERS
SECTION 40 75 16 INDUCTIVE CONDUCTIVITY ANALYZERS PART 1 General 1.01 SUMMARY A. Requirements for a digital inductive conductivity sensor for standard, hazardous and hightemperature applications. The sensor
More informationRinse connection adapter CPR 40
Technical Information TI 342C/07/en/10.02 No. 51510059 Rinse connection adapter CPR 40 Rinse connection adapter for use with retractable assembly Applications With the CPR 40, several media can be conveyed
More informationTechnical Data. Dimensions
0102 Model Number Features 5 mm flush Usable up to SIL2 acc. to IEC 61508 Accessories EXG-18 Quick mounting bracket with dead stop BF 18 Mounting flange, 18 mm Technical Data specifications Switching element
More informationTechnical Data. General specifications Switching element function Rated operating distance s n 5 mm
0102 Model Number Features 5 mm flush Usable up to SIL 3 acc. to IEC 61508 Application Danger! In safety-related applications the sensor must be operated with a qualified fail safe interface from Pepperl+Fuchs,
More informationAdditional Operating Instructions SITRANS F. Vortex flowmeters. SITRANS FX330 Ex-i.
Additional Operating Instructions SITRANS F Vortex flowmeters Ex-i Edition 09/2018 CONTENTS 1 Safety instructions 3 1.1 General notes... 3 1.2 EU conformity... 3 1.3 Approval according to the IECEx scheme...
More informationInductive slot sensor
0102 Model Number Features 2 mm slot width Technical Data specifications Switching function Normally closed (NC) Output type NAMUR Slot width 2 mm Depth of immersion (lateral) 5... 7 mm, typ. 6 mm Output
More informationTechnical Data. General specifications Switching element function Rated operating distance s n 5 mm
0102 Model Number Features 5 mm non-flush Usable up to SIL2 acc. to IEC 61508 Technical Data specifications Switching element function NAMUR, NC Rated operating distance s n 5 mm Installation non-flush
More informationVERSAFLOW VORTEX Supplementary instructions
VERSAFLOW VORTEX Supplementary instructions Vortex flowmeter Equipment category II 2G CONTENTS VERSAFLOW VORTEX 1 Safety instructions 3 1.1 General notes... 3 1.2 EC conformity... 3 1.3 Approval according
More informationTechnical Data. General specifications. Rated operating distance s n 5 mm
0102 Model Number Features 5 mm non-flush Usable up to SIL 2 acc. to IEC 61508 Technical Data specifications Switching function Normally closed (NC) Output type NAMUR Rated operating distance s n 5 mm
More informationAdditional Operating Instructions SITRANS F. Vortex flowmeters. SITRANS FX330 Ex-d.
Additional Operating Instructions SITRANS F Vortex flowmeters Ex-d Edition 09/208 CONTENTS Safety instructions 3. General notes... 3.2 EU conformity... 3.3 Approval according to the IECEx scheme... 3.4
More informationAdditional Operating Instructions SITRANS F. Vortex flowmeters. SITRANS FX330 Ex-nA.
Additional Operating Instructions SITRANS F Vortex flowmeters Ex-nA Edition 08/207 CONTENTS Safety instructions 3. General notes... 3.2 EU conformity... 3.3 Approval according to the IECEx scheme... 3.4
More informationOperating Instructions OUSTF10
BA00500C/07/EN/02.16 71316946 Products Solutions Services Operating Instructions OUSTF10 Optical sensor with the OUA260 flow assembly for the measurement of undissolved solids Table of contents Table of
More informationTechnical Data. Dimensions
0102 Model Number Features 8 mm non-flush Stainless steel housing Usable up to SIL2 acc. to IEC 61508 Technical Data specifications Switching element function NAMUR, NC Rated operating distance s n 8 mm
More informationTechnical Data. General specifications Switching element function Rated operating distance s n 8 mm
0102 Model Number Features 8 mm non-flush Usable up to SIL 3 acc. to IEC 61508 Application Danger! In safety-related applications the sensor must be operated with a qualified fail safe interface from Pepperl+Fuchs,
More informationOPTIBAR P 1010/2010 C Supplementary instructions
OPTIBAR P 1010/2010 C Supplementary instructions Pressure transmitter Equipment category II 1G / Ga, II 1D / Da in protection type intrinsic safety Exi KROHNE CONTENTS OPTIBAR P 1010/2010 C 1 Safety instructions
More informationMagnetic-Inductive Flow Velocity Sensor PIT. Installation and Operating Instructions
Magnetic-Inductive Flow Velocity Sensor PIT Installation and Operating Instructions Table of contents 1 IDENTIFICATION... 3 1.1 Manufacturer/supplier... 3 1.2 Product type... 3 1.3 Product name... 3 1.4
More informationRegulated Ex-Heater HEX with internal controller, HEX with external controller II 3 G II 3 D
Regulated Ex-Heater HEX5-1.08 with internal controller, HEX5-2.08 with external controller II 3 G II 3 D Instruction Manual Version 1.02.00 Dear customer, we have made up this operating manual in such
More informationTechnical Information OUSTF10
TI00500C/07/EN/02.16 71316944 Products Solutions Services Technical Information OUSTF10 Optical sensor with the OUA260 flow assembly for the measurement of turbidity and undissolved solids Application
More informationUltrasonic Level Measurement nivopuls FDU 10 S
Technical Information TI 275F/00/en Ultrasonic Level Measurement nivopuls FDU 10 S Level limit switch for liquids with separate electronics unit Non-contact from outside Suitable for use in explosion hazardous
More informationTechnical Data. General specifications Switching element function Rated operating distance s n 4 mm
0102 Model Number Features 4 mm non-flush Accessories BF 12 Mounting flange, 12 mm Technical Data specifications Switching element function NAMUR, NO Rated operating distance s n 4 mm Installation non-flush
More informationTechnical Information OUSAF21
TI00471C/07/EN/02.16 71316927 Products Solutions Services Technical Information Optical sensor for the measurement of low color concentrations Application The sensor is used to measure color concentrations
More informationTechnical Information Liquiline To Go CYM290 and Liquiline To Go Ex CYM291
TI01198C/07/EN/03.16 71326008 Products Solutions Services Technical Information Liquiline To Go CYM290 and Liquiline To Go Ex CYM291 Portable multiparameter device for Memosens ph, conductivity and oxygen
More informationOPTIWAVE X500 Supplementary Instructions
OPTIWAVE X500 Supplementary Instructions OPTIWAVE 3500 C OPTIWAVE 6500 C OPTIWAVE 7500 C Supplementary Instructions for ATEX applications KROHNE CONTENTS OPTIWAVE X500 1 General safety information 4 1.1
More informationInstallation, Operation and Maintenance Instructions for ThermaFlex Gasketed Plate Heat Exchangers
OM010 Installation, Operation and Maintenance Instructions for ThermaFlex Gasketed Plate Heat Exchangers The operating and maintenance instructions contained within this package are for ThermaFlex gasketed
More informationMANUAL Oil Level Sensor
PROCESS AUTOMATION MANUAL Oil Level Sensor KVF-104-PF ISO9001 0102 With regard to the supply of products, the current issue of the following document is applicable: The General Terms of Delivery for Products
More informationOPTIBAR DP 7060 Supplementary Instructions
OPTIBAR DP 7060 Supplementary Instructions Differential pressure transmitter Category ATEX II 1/2G, 2G Ex db ia IIC T6...T1 Ga/Gb, Gb IECEx Ex db ia IIC T6...T1 Ga/Gb, Gb Housing Aluminium: Single chamber,
More informationTechnical Data. General specifications Switching element function DC Dual NC Rated operating distance s n 3 mm
0102 Model Number Features Direct mounting on standard actuators EC-Type Examination Certificate TÜV99 ATEX 1479X Accessories BT32 BT32XS BT32XAS BT33 BT34 V1-G-N4-5M-PUR Female cordset, M12, 4-pin, NAMUR,
More informationTechnical Data. Dimensions
0102 Model Number Features Comfort series 50 mm non-flush Technical Data specifications Switching element function NAMUR, NC Rated operating distance s n 50 mm Installation non-flush Output polarity NAMUR
More informationMANUAL Overflow sensor NVF-104/34-PF
PROCESS AUTOMATION MANUAL Overflow sensor NVF-104/34-PF ISO9001 0102 Overflow sensor NVF-104/34-PF With regard to the supply of products, the current issue of the following document is applicable: The
More informationitemp RTD TMT 187 itemp TC TMT 188 Temperature head transmitter
BA 104R/24/ae/10.03 itemp RTD TMT 187 itemp TC TMT 188 Temperature head transmitter Operating instructions RTD TC TMT 187 TMT 188 Temperature head transmitter Safety message # Warning! Instructions and
More informationInductive Conductivity Measurement. Simultaneous display of conductivity/ resistance/salinity and temperature; large measurement symbol.
Process Analysis Systems Chem Energy Pharm Food Water Stratos Eco 2405 CondI Inductive Conductivity Measurement Simultaneous display of conductivity/ resistance/salinity and temperature; large measurement
More informationContinuous Gas Analyzers AO2000 Series AO2040-CU Ex Central Unit in Category 2G. Operator s Manual 42/24-13 EN Rev. 3
Continuous Gas Analyzers AO2000 Series AO2040-CU Ex Central Unit in Category 2G Operator s Manual 42/24-13 EN Rev. 3 Table of Contents Page Preface 3 General Safety Information 4 Special Safety Instructions
More informationTechnical Information OUSAF22
TI00472C/07/EN/02.16 71316931 Products Solutions Services Technical Information OUSAF22 Optical sensor combined with the OUA260 flow assembly for the measurement of color concentrations Application The
More informationIntrinsically Safe Pressure Transmitters for installation in hazardous locations Models IS-20-S, IS-21-S, IS-20-F, IS-21-F WIKA Datasheet IS-20
Electronic Pressure Measurement Intrinsically Safe Pressure Transmitters for installation in hazardous locations Models IS-20-S, IS-21-S, IS-20-F, IS-21-F WIKA Datasheet IS-20 Applications Chemical, Petrochemical
More informationInductive Conductivity
Inductive Conductivity THORNTON Leading Pure Water Analytics 7100E Non-Fouling Sensors PEEK or PFA Wetted Parts Excellent Chemical Resistance 0/4 to 20mA Outputs for Conductivity & Tempreature 4 Relays
More informationTechnical Information OUSAF44
TI00416C/07/EN/13.16 71316935 Products Solutions Services Technical Information Optical sensor combined with the OUA260 flow assembly for the measurement of UV absorption Application The sensor measures
More informationOperation and Maintenance Manual Sondex All-Welded Heat Exchangers (SAW)
Operation and Maintenance Manual Sondex All-Welded Heat Exchangers (SAW) The contents of this publication are based on the latest information available and the materials that are used at the time of printing.
More informationManual Power relay SR853
Manual Power relay SR853 Product types SR853.8.x.x, Rev. 1 SR853 manual page2 The symbols WARNING, CAUTION, NOTE This symbol warns of a serious hazard. Failure to observe this warning may result in death
More informationMagnetic-Inductive Flow Velocity Sensor PIT. Device Description
Magnetic-Inductive Flow Velocity Sensor PIT Device Description Table of contents 1 IDENTIFICATION...3 1.1 Manufacturer/supplier...3 1.2 Product type...3 1.3 Product name...3 1.4 Issue date...3 1.5 Version
More informationSensor module. Operating instructions. Sensor module, Type 17-51P2-... Document No P2-7D0001 Version: 17 May 2011/Rev. 0
Sensor module Operating instructions Sensor module, Document No. 11-51P2-7D0001 Version: 17 May 2011/Rev. 0 Operating Instructions Sensor-Module Type: 17-51P2-. Document no.: 11-51P2-7D0001 Version: 17.
More informationFlameproof Manual Call Points
Operating instructions Additional languages www.r-stahl.com CS & Clifford & Snell Contents General Information.... Manufacturer.... Information regarding the operating instructions.... Further documents....
More informationLevel Measurement Continuous level measurement Ultrasonic transmitters
Overview Configuration is a short-range integrated ultrasonic level transmitter. This intrinsically safe, 2 wire, 4 to 20 ma loop powered transmitter is ideal for liquids, slurries, and bulk materials
More informationLevel Electrode NRG Original Installation Instructions English
Level Electrode NRG 16-36 EN English Original Installation Instructions 819257-01 Contents Contents Page Important notes Usage for the intended purpose...4 Function...4 Safety note...4 Directives and standards
More informationDruck DPI 705/DPI 705 IS Series
GE Sensing & Inspection Technologies 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.testequipmentdepot.com Druck DPI 705/DPI 705 IS Series Digital Pressure
More informationTechnical Data. Dimensions
0102 Model Number Features 15 mm non-flush Technical Data specifications Switching element function NAMUR, NC Rated operating distance s n 15 mm Installation non-flush Assured operating distance s a 0...
More informationRADAR LEVEL GAUGE SPECIAL SAFETY INSTRUCTION
Special Safety Instruction en RADAR LEVEL GAUGE SPECIAL SAFETY INSTRUCTION Contents TankRadar Pro European ATEX Directive Information............................... 2 ATEX marking and Ex Certification
More informationSECTION ph/orp Sensors
SECTION 40 75 13 ph/orp Sensors PART 1 General 1.01 SUMMARY A. Requirements for a ph/orp sensor for in-process measurement. The sensor will incorporate Memosens digital technology for maximum process and
More informationInstallation, Operation and Maintenance Instructions ThermaFlex Gasketed Plate Heat Exchanger OM010
Installation, Operation and Maintenance Instructions ThermaFlex Gasketed Plate Heat Exchanger OM010 Calorifiers Heat Exchangers Pressurisation Units Sales Tel: 01457 835700 Sales Fax: 01457 832700 E-mail:
More informationMicro/Ultra Filtration System StamoClean CAT 430
Technical Information TI 338C/07/en/10.03 No. 51508729 Micro/Ultra Filtration System StamoClean CAT 430 Long-term Stable Membrane Filter for Immersion Operation in Activated Sludge Basins or Wastewater
More informationPressure sensors. Ex ia I / IIC T6 acc. to ATEX. Features. Description. Measuring ranges. Applications
Force Pressure Temperature Switch Pressure sensors Ex ia I / IIC T6 acc. to ATEX with internal diaphragm with front flush diaphragm Accuracy: 0,25% and 0,5 % Standard output: 4...20 ma; 2-wire system Description
More informationINSTRUCTION MANUAL IS-mB1 Minialite Intrinsically Safe Round LED Beacon
INSTRUCTION MANUAL Minialite Intrinsically Safe Round LED This instruction sheet describes installations which conform to EN60079:Part14:2008 Electrical Installation in Hazardous Areas. When designing
More informationWATLOW IND. FIREBAR Flange Heater Installation & Maintenance Manual I&M NUMBER: Page: 1 Date: 6/11/2008 Rev: 2.00
I&M NUMBER: 316-42-2-1 Page: 1 Pre Installation Check to make sure that heater received is the same as that ordered. Elements may come in contact with each other during shipment. Minor adjustments to elements
More informationV80-H Multistage Pump Instruction Manual
V80-H Multistage Pump Instruction Manual PRODUCT OVERVIEW The VADA horizontal multistage centrifugal pumps combine the functional benefits of centrifugal pumps and the practical benefits of self priming
More informationTechnical Information OUSAF12
TI00497C/07/EN/02.16 71316923 Products Solutions Services Technical Information OUSAF12 Optical sensor combined with the OUA260 flow assembly for absorbance measurement Application The sensor is used to
More informationDissolved Oxygen Measurement liquisys M COM 223 / 253
Technical Information TI 199C/24/ae Dissolved Oxygen Measurement liquisys M COM 223 / 253 Oxygen Controller Benefits at a glance Liquisys M COM 223 Liquisys M COM 253 Due to the modularity of its design,
More informationConductivity Electrode LRG Original Installation Instructions English
Conductivity Electrode LRG 16-4 EN English Original Installation Instructions 818854-02 1 Contents Important notes Page Usage for the intended purpose...4 Function...4 Safety note...5 Directives and standards
More informationRycroft Supapac Plate Heat Exchangers
Rycroft Supapac Plate Heat Exchangers Installation Operation & Maintenance Manual Ormandy Rycroft Duncombe Road Bradford BD8 9TB England TEL +44 (0)1274 490911 FAX +44 (0)1274 498580 www.rycroft.com 1.
More informationDifferential Pressure and Flow Meter. Media 5 Indicator 160 Ø PN 50
Differential Pressure and Flow Meter Media 5 Indicator 160 Ø PN 50 Application Transmitter for measuring and indicating the differential pressure or measured variables derived from it Suitable for gases
More informationFlow switch. Operating Manual. English manual page Page 1 of 15 Fax:
Operating Manual www.jlso-tec-trade.de Flow switch English manual page 1-15 Page 1 of 15 Flow switch Table of Contents Page 1 Device Description and Intended Use... 19 1.1 Flow switch version VH...X...
More informationOperating Instructions for Reka Hot-Melt Glue Applicators
Operating Instructions for Reka Hot-Melt Glue Applicators TR 700 TR 70 TR 80 TR 60 TR 50.4 TR 55 Table of Contents 1. Product Description 3 2. General Safety Instructions 4 3. Safety Measures to be taken
More informationOilSET Installation and Operating Instructions. Oil Separator Alarm Device with SET/DM3AL sensor
Labkotec UK Ltd Adminicle House 1 Lumb Lane Audenshaw Manchester M34 5WH GREAT BRITAIN Tel: 0844 3350 477 Fax: 0161 4281 179 E-mail: info@labkotec.co.uk 10.8.2012 Internet: www.labkotec.co.uk 1/13 OilSET-1000
More informationRosemount 4500 Hygienic Pressure Transmitter. Start
Quick Installation Guide June 2007 Rosemount 4500 Rosemount 4500 Hygienic Pressure Transmitter ProductDiscontinued Start Step 1: Mount the Transmitter Step 2: Set the Switches Step 3: Connect Wiring and
More informationOperating Instructions OUSAF12
BA00497C/07/EN/02.16 71316925 Products Solutions Services Operating Instructions Optical sensor combined with the OUA260 flow assembly for absorbance measurement Document information 1 Document information
More informationAdvance Optima Continuous Gas Analyzers AO2060-Caldos15, -Caldos17, -Magnos106, -Uras14 Category 2G Analyzer Modules
Advance Optima Continuous Gas Analyzers AO2060-Caldos15, -Caldos17, -Magnos106, -Uras14 Category 2G Analyzer Modules Operator s Manual 42/24-12 EN Rev. 4 Contents Page Preface 4 Safety Information General
More informationPEPPERL+FUCHS GmbH
Comfort series 1.5 mm embeddable M8x1 4 25 16 13 LED Switching element function NAMUR NC Rated operating distance s n 1,5 mm Installation embeddable Assured operating distance s a 0... 1,215 mm Reduction
More informationTranslation of the assembly instructions with operating instructions and technical appendix
BA Ex-SBU Switch box type SBU-x(1,2,3)x Translation of the assembly instructions with operating instructions and technical appendix In accordance with EU ATEX Directive 2014/32/EU In accordance with EU
More informationTechnical Information OUSTF10
TI00500C/07/EN/01.13 71157320 Products Solutions Services Technical Information Optical sensor combined with flow assembly OUA260 for measurement of turbidity and non-dissolved solids Application The scattered
More informationWATLOW IND. WATROD Flange Heater Installation & Maintenance Manual I&M NUMBER: Page: 1 Date:6/11/2008 Rev: 2.00
I&M NUMBER: 316-42-8-1 Page: 1 _ Pre Installation Check to make sure that heater received is the same as that ordered. Elements may come in contact with each other during shipment. Minor adjustments to
More informationCenturion Guided Radar HAWK CGR SERIES
Safety Instructions CGR (Cat 3GD) Centurion Guided Radar HAWK CGR SERIES ATEX Category 3GD Intrinsically Safe probe / Flameproof enclosure II 3G Ex ia/d IIC T6 T2 Gc II 3D Ex ia tb IIIC T85 C T255 C Dc
More informationVADA - V80-H PRODUCT OVERVIEW CONSTRUCTION USAGE LIMITATIONS MOTOR WARRANTY
PRODUCT OVERVIEW The VADA horizontal multistage centrifugal pumps combine the functional benefits of centrifugal pumps and the practical benefits of self priming pumps. They warrant high hydraulic and
More informationAdditional Information. Introduction. Temperature transmitter for all communications protocols. Redundancy thanks to two inputs. Measurement made easy
ABB MEASUREMENT & ANALYTICS COMMISSIONING INSTRUCTION TTF300 Field-mount temperature transmitter Temperature transmitter for all communications protocols. Redundancy thanks to two inputs. Measurement made
More informationInstallation instructions
Immersion Heater ROTKAPPE Angular Immersion Heater ROTKAPPE Installation instructions E-MA 10/19 04.13/1 Please supplement the following information from the model plate For the immersion heater ROTKAPPE
More informationSteam Trap BK 45 BK 45-U BK 45-LT BK 46
Steam Trap BK 45 BK 45-U BK 45-LT BK 46 Original Installation Instructions 810437-08 Contents Foreword... 3 Availability... 3 Formatting features in the document... 3 Safety... 3 Use for the intended purpose...
More informationSafety instructions VEGAPULS 64
Safety instructions Intrinsic safety "i" 4 20 ma/hart - two-wire Document ID: 53001 Contents 1 Area of applicability... 3 2 Important specification in the type code... 3 3 Different ignition protection
More informationInstallation guide 864 MTT Multi Thermo sensor Thermometer
Installation guide 864 MTT Multi Thermo sensor Thermometer April 2010 Part no. 4416.234 Revision 3 Honeywell Enraf PO Box 812 2600 AV Delft Netherlands Tel. : +31 15 2701 100 Fax : +31 15 2701 111 E-mail
More informationSurvey of electrical equipment installed in hazardous areas on tankers
(June 2015) Survey of electrical equipment installed in hazardous areas on tankers 1. Application The recommendations in this document apply for survey of electrical installation in hazardous areas on
More informationDecentralised ventilation units
Notes on installation Decentralised ventilation units FSL / SCHOOLAIR GB/en TROX GmbH Heinrich-Trox-Platz 47504 Neukirchen-Vluyn Germany Phone: +49 (0) 2845 2020 Fax: +49 (0) 2845 202265 E-mail: trox@trox.de
More informationDräger Polytron 5100 EC Detection of toxic gases and vapors
Dräger Polytron 5100 EC Detection of toxic gases and vapors The Dräger Polytron 5100 EC is a cost-effective explosion proof transmitter for the detection of toxic gases or oxygen. It uses a high performance
More information! The Caution Symbol (exclamation point) alerts you to a "CAUTION", a safety or
I&M NUMBER: 316-42-10-1 Page: 1 Pre Installation Check to make sure that heater received is the same as that ordered. Elements may come in contact with each other during shipment. Minor adjustments to
More informationFunctional Safety Manual Oil Leak Detector NAR300 System
SD01357G/08/EN/01.14 71255416 Products Solutions Services Functional Safety Manual Oil Leak Detector NAR300 System Application This system is designed to be installed in a pit, dike, plant, or sump pit
More informationWATLOW IND. WATROD Circulation Heater Installation & Maintenance Manual I&M NUMBER: Page: 1 Date: 6/11/2008 Rev: 2.00
I&M NUMBER: 316-42-5-1 Page: 1 Pre Installation Check to make sure that heater received is the same as that ordered. Watlow heaters are built to comply with UL and CSA dielectric requirements, it may be
More informationSmartPower Solutions [ SmartPower Solutions. Quick Installation Guide , Rev AC January 2012
Start Step 1: Physical Installation Step 2: Verify Operation Disposal/Recycling of Depleted Power Modules Product Certifications EC Declaration of Conformity End 00825-0100-4701[ 2012 Rosemount Inc. All
More informationSITRANS T. SITRANS T Explosion protected temperature sensor. General information. Marking of the degree of protection. Range of uses.
General information 1 Marking of the degree of protection 2 SITRANS T SITRANS T Explosion protected temperature sensor Operating Instructions Range of uses 3 Installation 4 Assembly and disassembly 5 Commissioning
More informationWATLOW ELECTRIC MFG CO. FIREBAR Screw Plug Installation & Maintenance Manual I&M NUMBER: Page: 1 Date: 11/25/2013 Rev: 4.
I&M NUMBER: 316-42-3-1 Page: 1 Pre Installation Check to make sure that heater received is the same as that ordered. Elements may come in contact with each other during shipment. Minor adjustments to elements
More informationNRG NRG NRG NRG
GESTRA Steam Systems NRG 16-50 NRG 17-50 NRG 19-50 NRG 111-50 EN English Installation Instructions 818954-04 Level Electrodes NRG 16-50, NRG 17-50, NRG 19-50, NRG 111-50 1 Contents Important Notes Page
More informationUser and maintenance manual
GB User and maintenance manual IMPORTANT SAFETY INSTRUCTIONS These instructions shall also be available on website: docs.whirlpool.eu. YOUR SAFETY AND THAT OF OTHERS IS HIGHLY IMPORTANT. This manual and
More informationModel: 1200 Model: 1200-G Model: 1200-N Float-Controlled Condensate Trap, PN 40
1 Safety instructions Model: 1200 Model: 1200-G Model: 1200-N Float-Controlled Condensate Trap, PN 40 1.1 Proper use Any improper use, intervention in the design and deviation from the design data automatically
More informationOperating instructions Electric agitator RE80Ex. T-Dok-520-GB-Rev.5 Article no Translation of the original operating instructions
GB Operating instructions Electric agitator RE80Ex T-Dok-520-GB-Rev.5 Article no. 200-0332 Translation of the original operating instructions Thank you for selecting a Krautzberger product. This product
More informationOperating Instructions
Operating Instructions MSi-JM0100-USB R. STAHL HMI Systems GmbH Adolf-Grimme-Allee 8 D 50829 Köln Operating Instructions Version: 01.00.03 Issue date: 29.06.2017 Disclaimer Disclaimer Publisher and copyright
More informationDissolved Oxygen Measurement liquisys S COM 223 / 253
Technical Information TI 199C/07/en No. 51500281 Dissolved Oxygen Measurement liquisys S COM 223 / 253 Oxygen Transmitter Benefits at a glance Measuring transmitter in field or panel-mounted housing Universal
More informationCertificate of Compliance
FF0 001 Certificate of Compliance Certificate: 70116279 (LR 025993_0_000) Issued to: General Monitors, Incorporated 26776 Simpatica Circle Lake Forest, California 92630 USA Attention: Larry Vlagea The
More informationOperating Instructions Edition 02/2007
Operating Instructions Edition 02/2007 Temperature Transmitter 7NG3214 with PROFIBUS PA 7NG3215 with FOUNDATION Fieldbus sitrans Introduction 1 General safety notes 2 SITRANS T Temperature transmitter
More informationSITRANS Pressure transmitter SITRANS LH300 (7MF1575..) Transmitter for hydrostatic level Compact Operating Instructions
SITRANS Pressure transmitter Compact Operating Instructions Legal information Warning notice system This manual contains notices you have to observe in order to ensure your personal safety, as well as
More informationWATLOW IND. WATROD Modular Duct Heater Installation & Maintenance Manual I&M NUMBER: Page: 1 Date:6/11/2008 Rev: 2
I&M NUMBER: 316-42-15-1 Page: 1 _ Pre Installation Check to make sure that heater received is the same as that ordered. Elements may come in contact with each other during shipment. Minor adjustments to
More informationElectron Microscopy Sciences
Electron Microscopy Sciences INSTRUCTIONAL MANUAL CAT. 71260-10 through 71260-17 Cabinet Style Lab Companion Vacuum Desiccators sgkcck@aol.com P.O. Box 550, 1560 Industry Road, Hatfield, PA 19440 Safety
More informationOnly for operating with ATEX certificated sensors DMT 99 ATEX E056/N5 Maximum optical input power Maximum radiation intensity Detection pattern
ISO 9001:2000 / ATEX 0158 II 1 GD IIB T135 C (T4) Broad beam optical fibre QW-..-1GD ATEX certificated for Ex zones 0, 1, 2, 20, 21, 22 Applicable for width deviation measurement, detection of small objects
More information