1 Purge and Trap Troubleshooting Made Easy? Tyler Trent- Application Chemist Preston Hoppes- International Tech Support/FSE Tammy Rellar- Application Sales Specialist Teledyne Tekmar

2 Purge and Trap Troubleshooting No Response Carryover/Contamination Elution issue- early, middle, late Water interference Internal Standard recovery variability Linearity/Reproducibility issues System leaks

3 Troubleshooting Tips Take a moment to review manuals Examine the facts and use valid reasoning Identify the root cause of the problem Avoid quick fixes and shotgun approach BE PATIENT!

4 Finding the Root Cause MS,GC, Concentrator, and Autosampler Split the system into 4 separate entities

5 Eliminate the GC/MS Not necessary to disconnect Purge & Trap (P&T) Confirm normal operation of the MS Tune parameters/vacuum Clean source Confirm normal operation of GC Do direct inject of same standard used in P&T Poor chromatography/resolution/reduced response Injection liner/septum Clip column Once GC/MS is eliminated as source of issue, proceed to P&T/ASM

6 Eliminate the Purge and Trap Confirm normal operation of the GC/MS Tune/direct inject Confirm normal operation of the P&T Run diagnostics/leak check Make sure the correct analytical trap is in place Manually load sample into Purge and Trap Eliminates the autosampler

7 No Response Are the analytes getting to the GC/MS Check the trap heater Reaches the set temperature Check 6-port valve Hear the 6-port actuate 6-port rotor installed correctly Correct installation of the 6-port valve

8 No Response Are the analytes getting to the trap Bubbles in the sparge tube during purging Correct flow set in method Mass flow controller working System leak tight Gas tank still have pressure

9 No Response Stratum

10 No Response Check Autosampler Is there sample in sparge vessel? Pressure problems Sample get pulled Valves working correctly Sample and drain valves

11 No Response AQUATek100

12 Carryover/Contamination Carryover for full target list/late eluters and contamination Run GC only Desorb only Check temperature Desorb/Bake Check hot water/meoh rinse

13 Carryover of Full Target List GC/MS Faulty EPC Purge and Trap Solenoid valve leaking cross port Faulty trap heater Sample not draining entirely Dirty glassware Autosampler Check soil vs. water if available System not rinsing correctly Internal standard valve leaking(allows standard to enter vessel) Not transferring entire aliquot to Purge and Trap

14 Carryover of Late Eluters GC Inlet too cold Purge and Trap Not enough bake Bad trap Cold spots Transfer line connection to GC inlet Faulty condensate or trap heater Autosampler Check soils vs. waters if available Hot water heater/rinse failure Not enough bake rinses Sample needle not cleaning up Not transferring bake rinse to purge and trap

15 Carryover of Late Eluters AQUATek100

16 Contamination Issues? No problem! Able to rinse sample pathway with 90 C water Atomx uses methanol to rinse entire pathway Includes syringe, associated transfer lines and valves, sparger and 3-stage needle. User can vary volume and number of rinse times user settable within the software method

17 Data One 90 C Water Sparge Vessel Rinse Two 90 C Water Sparge Vessel Rinses Three 90 C Water Sparge Vessel Rinses 3mL Methanol Sparge Vessel Rinse mL Methanol Sparge Vessel Rinse 3mL Methanol Sparge Vessel Rinse and 3mL Methanol Needle Rinse

18 Elution Issues Tune Auto-tune/BFB tune Direct inject If missing peaks could be a bad EPC Column Top of Trap If missing peaks could be sample tubing/transfer line

19 Early Elution Groups GC Split flow leak Bad column Purge and Trap Small leak Water front too large Too long of dry purge Trap too hot (faulty electronics or heater)

20 Middle Elution Groups (Very Rare) GC Column phase is bad Purge and Trap Desorb preheat problems Condensate trap has too much water Mass flow controller leaking Bad Trap

21 Late Elution Groups GC/MS Source maintenance Carrier flow too low Purge and Trap Poor purge efficiency Restrictive sample tubing Dirty glassware Sample not heated properly Plugged vent valve Bad Trap

22 Water interference MS Vacuum issues GC Split flow too low Bad column Purge and Trap Condensate setting and or failures Dry purge not optimized Sample temperature concerns (too hot) Desorb Time

23 Water Interference Chromatogram showing the water front for a 2min desorb (white) compared to 4min desorb (green)

24 Water Interference Chromatogram showing the water front for a 2min desorb at 1ppb (white) compared to 4min desorb at 1ppb (green)

25 Internal Standard Recoveries GC/MS response increases as Target compound concentration increases Dirty inlet liner or source Validate by preparing 3 increasing target concentrations with Internal Standard concentration fixed in 1mL vials and then direct inject 1uL of the concentrations into the GC If still increasing isolate between source and liner by performing cleaning and or replacement. 6mm draw-out plate

26 Internal Standard Recoveries Purge and Trap Isolated to an active site Sample tubing/condensate trap/analytical trap Autosampler Hand spike vials first Automated delivery system Tubing restrictions Valve malfunction Pressure problems

27 Internal Standard Recoveries AQUATek100

28 Linearity and Reproducibility GC/MS Source maintenance Vacuum problems EPC failures Bad column Method not optimized Oven temperature program Multiplier going bad

29 Linearity and Reproducibility Purge and Trap Method not optimized Trap is failing Drain valve leaking Excess water in system Temperatures not reaching set-points Too much carry over Sample preparation errors Dirty standard syringe or glassware Too much headspace Analyst variability

30 Linearity and Reproducibility Autosampler Not pulling same amount of sample Not transferring sample correctly Improper rinsing between samples Compare soils vs. water if possible Determine if IS is varying and causing everything else to look unstable

31 Purge and Trap System Leaks Perform Leak Checks through Software:

32 Purge and Trap System Leaks

33 Purge and Trap System Leaks Since the Atomx is an Autosampler and P&T concentrator combined, the leak check is more complex If fails the overall leak check it will break down into three subsystems

34 Purge and Trap System Leaks Subsystem A Mass flow controller, Bake valve, Purge valve, Pencil Filter, Purge select valve, Syringe valve, and Pressurize valve

35 Purge and Trap System Leaks Subsystem B All subsystem A, Syringe,7-port valve, Transfer valve, Sample needle, and Soil valve

36 Purge and Trap System Leaks Subsystem C All of subsystem A&B, Sample mount, Sparge vessel, Drain valve, and 6-port Valco Notice traps not included so failure of system with A, B, and C passing indicates leak in trap area

37 Benchmark Make sure system is leak tight first Checks all system controls including solenoid valves, heaters, flows, and 6-port valve Helps isolate a variety of potential issues

38 Miscellaneous Issues Draining sample -Flag on Hamilton Valve -Drain Valve Communication with GC/MS -Standard 63 -User 31 History Log -To check Purge/Bake pressures -To check Errors and Modes they occurred in

39 Tech Support

Simplicity. Reliability. Performance. Stratum. Automated VOC Sample Prep System

Simplicity. Reliability. Performance. Stratum. Automated VOC Sample Prep System Simplicity Reliability Performance Stratum Automated VOC Sample Prep System The Stratum Purge and Trap Concentrator (PTC) is a sample preparation instrument used to remove Volatile Organic Compounds (VOCs)

More information

Tekmar-DOHRMANN The 3100 Sample Concentrator From ManualLib.com

Tekmar-DOHRMANN The 3100 Sample Concentrator   From ManualLib.com Tekmar-DOHRMANN The 3100 Sample Concentrator http://www.manuallib.com/file/2660162 From ManualLib.com ManualLib.com collects and classifies the global product instrunction manuals to help users access

More information

GC Injectors. GC Injectors

GC Injectors. GC Injectors SRI offers a wide variety of GC injectors and injection systems. Up to five injectors may be mounted simultaneously on the Model 8610C or 8610D. The Model 310 will accommodate a single On-column, Heated

More information

TYPES AND APPROPRIATE INJECTORS

TYPES AND APPROPRIATE INJECTORS SRI offers a wide variety of GC injectors and injection systems. Up to five injectors may be mounted simultaneously on the Model 8610C or 8610D. The Model 310 will accommodate a single On-column, Heated

More information

Aromatics and Oxygenates in Gasoline June2018

Aromatics and Oxygenates in Gasoline June2018 SRI has configured in a single GC two common tests performed on gasoline. The first test is a a simplified version of ASTM 4815, oxygenates in gasoline. The second test is a simplified version of ASTM

More information

Automated Solid Phase Extraction (SPE) of EPA Method 1694 for Pharmaceuticals and Personal Care Products in Large Volume Water Samples

Automated Solid Phase Extraction (SPE) of EPA Method 1694 for Pharmaceuticals and Personal Care Products in Large Volume Water Samples Automated Solid Phase Extraction (SPE) of EPA Method 1694 for Pharmaceuticals and Personal Care Products in Large Volume Water Samples Keywords Application Note ENV0212 This collaboration study was performed

More information

MEPS - Micro Extraction by Packed Sorbent Online SPE for GC and LC sample preparation - Extraction to injection in a single process

MEPS - Micro Extraction by Packed Sorbent Online SPE for GC and LC sample preparation - Extraction to injection in a single process - Micro Extraction by Packed Sorbent Online SPE for GC and LC sample preparation - Extraction to injection in a single process Save Hours in Sample Preparation Reduce the time to prepare and inject samples

More information

SCION 436-GC. Dimensions and Weights

SCION 436-GC. Dimensions and Weights SCION 436-GC Specification Sheet Dimensions and Weights Height*: 57 cm (22.5 in.), Width: 32 cm (13.0 in.), Depth: 61 cm (24.0 in.) Weight*: 26.8 kg (59lb) * Typical values Environmental Conditions Operating

More information

Perkin Elmer Clarus 500 Gas Chromatograph

Perkin Elmer Clarus 500 Gas Chromatograph IET International Equipment Trading, Ltd. www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Perkin Elmer Clarus 500 Gas Chromatograph

More information

FIXED GAS ANALYZER (FGA)

FIXED GAS ANALYZER (FGA) Gas Chromatography FIXED GAS ANALYZER (FGA) Productivity remains the primary goal of most work environments. Utilizing a gas chromatographic system capable of performing multiple analyses on a series of

More information

On Line SPE LC/MS/MS Applications

On Line SPE LC/MS/MS Applications On Line SPE LC/MS/MS Applications Dr. Sheher Bano Mohsin Mass Spectrometry Applications March 19, 2014 1 3/21/2014 Introduction Why Online SPE? Sensitivity of the Mass Spectrometers Application Examples

More information

SRI 8610C GC for PCB in Transformer Oil Analysis November 2014

SRI 8610C GC for PCB in Transformer Oil Analysis November 2014 The SRI 8610C Gas Chromatograph ( GC ) is installed on any bench top. When the red lid is raised, you can see the column oven, injector ( on-column type ), detector ( DELCD ) and gas stream dryer. The

More information

On-line SPE Enrichment of Trace Organic Contaminants in Water and Juices

On-line SPE Enrichment of Trace Organic Contaminants in Water and Juices On-line SPE Enrichment of Trace Organic Contaminants in Water and Juices Application Note Environment and Food Authors E. Michael Thurman, Imma Ferrer Center for Environmental Mass Spectrometry Department

More information

High Volume Injection up to 5 ml with the Agilent 1200 Infinity Series Online SPE Solution for Highest Sensitivity

High Volume Injection up to 5 ml with the Agilent 1200 Infinity Series Online SPE Solution for Highest Sensitivity High Volume Injection up to ml with the Agilent 00 Infinity Series Online SPE Solution for Highest Sensitivity Technical Overview Author Bettina Schuhn Agilent Technologies, Inc. Waldbronn, Germany Abstract

More information

Velocity XPT Accelerated Purge and Trap Sample Concentrator

Velocity XPT Accelerated Purge and Trap Sample Concentrator Velocity XPT Accelerated Purge and Trap Sample Concentrator 14-8900-074_vB Copyright 2003 Teledyne Tekmar All rights reserved. Reproduction, adaption, or translation without permission is prohibited, except

More information

INJECTORS Purge & Trap General Operating Procedures continued

INJECTORS Purge & Trap General Operating Procedures continued General Operating Procedures continued 1. Remove the 6 screws that secure the bottom panel to the rest of the GC chassis. Support the panel while you gently rock the GC onto its back, then lower the panel

More information

300 series HPLC autosamplers

300 series HPLC autosamplers 300 series HPLC autosamplers issue 1 Automated HPLC Sampling Manual HPLC sampling can be very difficult and time consuming......hta s range of HPLC autosamplers can perform HPLC sample preparation and

More information

MultiPurpose Sampler MPS robotic series

MultiPurpose Sampler MPS robotic series MultiPurpose Sampler MPS robotic series sample prep for winners The MPS robotic efficiently and reliably automates sample preparation and sample introduction MPS robotic for GC and GC/MS Whether you are

More information

Analysis of Pharmaceuticals and Personal Care Products in Groundwater by Online SPE and LC/MS/MS

Analysis of Pharmaceuticals and Personal Care Products in Groundwater by Online SPE and LC/MS/MS Analysis of Pharmaceuticals and Personal Care Products in Groundwater by Online SPE and LC/MS/MS Application note Water analysis Authors E. Michael Thurman, Imma Ferrer, and Karl Mauch University of Colorado,

More information

HISTORY OF VARIAN CAPILLARY INJECTORS

HISTORY OF VARIAN CAPILLARY INJECTORS HISTORY OF VARIAN CAPILLARY INJECTORS by Randall Bramston-Cook Lotus Consulting 5781 Campo Walk Long Beach, Ca 90803 310/569-0128 714/898-7461 Fax and Herb Neumann Varian, Inc. 2700 Mitchell Drive Walnut

More information

AppNote 6/2008. Automated SPE and Fast GC-ECD Analysis of PCBs in Waste Oil KEYWORDS ABSTRACT INTRODUCTION. MACH, Fast GC-ECD, SPE, PCB, waste oil

AppNote 6/2008. Automated SPE and Fast GC-ECD Analysis of PCBs in Waste Oil KEYWORDS ABSTRACT INTRODUCTION. MACH, Fast GC-ECD, SPE, PCB, waste oil AppNote 6/2008 Automated SPE and Fast GC-ECD Analysis of PCBs in Waste Oil Karine Jacq, Bart Tienpont, Frank David Research Institute for Chromatography, Pres. Kennedypark 26, B-8500 Kortrijk, Belgium

More information

APPLICATION NOTE DETERMINATION OF THE HYDROCARBON-INDEX DIN EN ISO :2000. APPLICATION NOTE As of July 2017, Version: 1.

APPLICATION NOTE DETERMINATION OF THE HYDROCARBON-INDEX DIN EN ISO :2000. APPLICATION NOTE  As of July 2017, Version: 1. DETERMINATION OF THE HYDROCARBON-INDEX DIN EN ISO 9377-2:2000 APPLICATION NOTE www.lctech.de As of July 2017, Version: 1.7 Keywords: DIN EN ISO 9377-2:2000, DEV H53; LAGA-Richtlinie KW/04 2009, Hydrocarbon

More information

Thermo Scientific Rocket Evaporator. Revolutionary solvent evaporation complete laboratory workflow

Thermo Scientific Rocket Evaporator. Revolutionary solvent evaporation complete laboratory workflow Thermo Scientific Rocket Evaporator Revolutionary solvent evaporation complete laboratory workflow Sample evaporation with walkaway capability to produce uncompromised results The Thermo Scientific Rocket

More information

JSB is an authorised partner of DETERMINATION OF THE HYDROCARBON-INDEX DIN EN ISO :2000. i n s p i r a t i o n m e e t s i n n o v a t i o n!

JSB is an authorised partner of DETERMINATION OF THE HYDROCARBON-INDEX DIN EN ISO :2000. i n s p i r a t i o n m e e t s i n n o v a t i o n! JSB is an authorised partner of DETERMINATION OF THE HYDROCARBON-INDEX DIN EN ISO 9377-2:2000 Determination of the Hydrocarbon Oil Index EN ISO 9377-2:2000 (H53-Method) The European method EN ISO 9377-2:2000

More information

Solid Phase Extraction using the 96-well Format ISOLUTE -96 Fixed Well Plates and ISOLUTE Array 96-well Plates

Solid Phase Extraction using the 96-well Format ISOLUTE -96 Fixed Well Plates and ISOLUTE Array 96-well Plates TN114 Solid Phase Extraction using the 96-well Format ISOLUTE -96 Fixed Well Plates and ISOLUTE Array 96-well Plates New approaches to the quantitation of drugs and metabolites in biological fluids have

More information

Introduction to the Fusion TOC Analyzer

Introduction to the Fusion TOC Analyzer Introduction to the Fusion TOC Analyzer Application Note By: Stephen Lawson Abstract Building on nearly 40 years of design and manufacturing experience, Teledyne Tekmar introduces its fifth generation

More information

Agilent GC Sampler 80/120 SPME Option

Agilent GC Sampler 80/120 SPME Option Agilent GC Sampler 80/120 SPME Option User Manual Agilent Technologies Notices Agilent Technologies, Inc. 2011 No part of this manual may be reproduced in any form or by any means (including electronic

More information

Agilent 1290 Infinity II Vialsampler Data Sheet

Agilent 1290 Infinity II Vialsampler Data Sheet Agilent 1290 Infinity II Vialsampler Data Sheet Agilent 1290 Infinity II Vialsampler Data Sheet Product Description The Agilent 1290 Infinity II Vialsampler is designed for UHPLC applications up to 1300

More information

XENOSEP. SPE Systems for EPA Method 1664A. The Easiest, Most Cost Effective Way to Increase Productivity

XENOSEP. SPE Systems for EPA Method 1664A. The Easiest, Most Cost Effective Way to Increase Productivity XENOSEP EPA 1664A Std. Plus 14 g Solids SPE Systems for EPA Method 1664A The Easiest, Most Cost Effective Way to Increase Productivity With Xenosep SPE Filters, we experienced a significant reduction in

More information

Rapid Quantification of Polar and Semipolar Pesticide Metabolites with Combined Online SPE and Direct Injection

Rapid Quantification of Polar and Semipolar Pesticide Metabolites with Combined Online SPE and Direct Injection Application Note Environmental Rapid Quantification of Polar and Semipolar Pesticide Metabolites with Combined Online SPE and Direct Injection Intensity (cps) 5 5 Semipolar metabolites Three polar metabolites

More information

Automação de SPE utilizando sistema robótico Triplus RSH autosampler. Nova Analitica

Automação de SPE utilizando sistema robótico Triplus RSH autosampler. Nova Analitica Automação de SPE utilizando sistema robótico Triplus RSH autosampler Nova Analitica ITSP Introduction The heart of the ITSP system is a patented cartridge containing customer-defined filter and/or SPE

More information

POPULAR CONFIGURATION GCs Cryosulfur GC. System Overview

POPULAR CONFIGURATION GCs Cryosulfur GC. System Overview System Overview The SRI CryoSulfur GC comes with everything you need to detect low-level sulfur compounds in gas samples. Since some sulfur compounds do not trap well, the CryoSulfur GC uses the CryoCooler

More information

Systems. Visit us at

Systems. Visit us at Systems And especially: SPE Module Visit us at www.lctech.de! Systems General e Description Highlights Modules Structure and Availability FREESTYLE TM BASIC FREESTYLE TM SPE 2 FREESTYLE General Description

More information

Enhancing Throughput of Online SPE Analysis of Pesticides in Wastewater The Agilent 1200 Infi nity Series Online SPE Solution

Enhancing Throughput of Online SPE Analysis of Pesticides in Wastewater The Agilent 1200 Infi nity Series Online SPE Solution Enhancing Throughput of Online SPE Analysis of Pesticides in Wastewater The Agilent 00 Infi nity Series Online SPE Solution Technical Overview Author Bettina Schuhn Agilent Technologies, Inc. Waldbronn,

More information

GC DETECTORS Reduction Gas Detector - RGD. Overview

GC DETECTORS Reduction Gas Detector - RGD. Overview The SRI Reduction Gas Detector is the most sensitive detector available for hydrogen, carbon monoxide, and other reducing gases. Low ppb detection limits are achievable for hydrogen. Detection limits in

More information

To request a quote or locate your nearest dealer, contact Customer Care toll-free in the US at or

To request a quote or locate your nearest dealer, contact Customer Care toll-free in the US at or Sample Prep System J2 Scientific introduces the ultimate flexibility and automation for the sample prep lab. The PrepLinc Platform Automated Sample Preparation System offers the ability to combine common

More information

FREESTYLE SPE UNIQUE POSSIBILITIES FOR SOLID PHASE EXTRACTION

FREESTYLE SPE UNIQUE POSSIBILITIES FOR SOLID PHASE EXTRACTION FREESTYLE SPE UNIQUE POSSIBILITIES FOR SOLID PHASE EXTRACTION ONE DEVICE MANY OPTIONS Modular and amazingly independent. The FREESTYLE system is a fully automated sample preparation system for the unsupervised

More information

Standard VistaNET GCC Software: 803V029-(VER)(REV)

Standard VistaNET GCC Software: 803V029-(VER)(REV) This summarizes the software release notes for the standard VistaNET-enabled GCC software using the 8086 SBC. This document will be updated as additional releases are made. Please see similar 803V0xx-

More information

To request a quote or locate your nearest dealer, contact Customer Care toll-free in the US at or

To request a quote or locate your nearest dealer, contact Customer Care toll-free in the US at or Sample Prep System J2 Scientific introduces the ultimate flexibility and automation for the sample prep lab. The PrepLinc Platform Automated Sample Preparation System offers the ability to combine common

More information

Process and Air Monitoring. Online GC systems for continuous monitoring of organic compounds in air and gaseous samples

Process and Air Monitoring. Online GC systems for continuous monitoring of organic compounds in air and gaseous samples Process and Air Monitoring Online GC systems for continuous monitoring of organic compounds in air and gaseous samples For 35 years AMA Instruments has developed and manufactured analytical devices and

More information

ACME DIAGNOSTICS STANDARD OPERATING PROCEDURE

ACME DIAGNOSTICS STANDARD OPERATING PROCEDURE ACME DIAGNOSTICS STANDARD OPERATING PROCEDURE Procedure Number: II.01.200a Effective Date: 09/14/99 Review and Update: 09/2003 TITLE: ADVIA 120 ROUTINE MAINTENANCE PURPOSE AND PRINCIPLE: The ADVIA 120

More information

Recent Improvements in Process GC Instrumentation

Recent Improvements in Process GC Instrumentation ISA Houston, TX Oct., 15 2003 Recent Improvements in Process GC Instrumentation Dr. Jack Driscoll PID Analyzers, LLC copyright 2000-2006 PID Presentation Objectives A new Process GC has been developed

More information

CyTOF Autosampler 5 User Manual

CyTOF Autosampler 5 User Manual CyTOF Autosampler 5 User Manual CyTOF Autosampler 5 User Manual This is Class I, cord-connected equipment. For use in commercial, industrial or business enviroments. For Research Use Only. Not for use

More information

Waters 2695D Separations Module

Waters 2695D Separations Module S U P P L E M E N T A L R E L E A S E N O T E S Waters 2695D Separations Module Firmware Update Version 2.04d Firmware Update Version 2.04d Information This firmware update incorporates improved connectivity

More information

HITACHI CHROMASTER. Clarity Control Module. Code/Rev.: M173/70I Date: 6/23/2018

HITACHI CHROMASTER. Clarity Control Module. Code/Rev.: M173/70I Date: 6/23/2018 HITACHI CHROMASTER Clarity Control Module ENG Code/Rev.: M173/70I Date: 6/23/2018 Phone: +420 251 013 400 DataApex Ltd. Fax: +420 251 013 401 Petrzilkova 2583/13 clarity@dataapex.com 158 00 Prague 5 www.dataapex.com

More information

Automated Sample Preparation: SPE GPC - EVAporation SUPERIORITY BY ADAPTATION. Visit us at

Automated Sample Preparation: SPE GPC - EVAporation SUPERIORITY BY ADAPTATION. Visit us at Automated Sample Preparation: SPE GPC - EVAporation SUPERIORITY BY ADAPTATION Visit us at www.lctech.de! What LCTech does supplies innovative products and methods for the preparation and analysis of food,

More information

Waters 2695 Separations Module

Waters 2695 Separations Module S U P P L E M E N T A L R E L E A S E N O T E S Waters 2695 Separations Module Firmware Update Version 2.04 Firmware Update Version 2.04 Information This firmware update incorporates improved checking

More information

Chemistry Instrumental Analysis Lecture 29. Chem 4631

Chemistry Instrumental Analysis Lecture 29. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 29 Gas Chromatography Detectors Gas Chromatography Monitors the column effluent and produces an electrical signal that is proportional to the amount of analyte

More information

Spark Holland introduces

Spark Holland introduces BETTER SAMPLE CARE Spark Holland introduces ACE Absolutely unique: on-line SPE with Automated Cartridge Exchange Spark Holland introduces ACE Absolutely unique: on-line SPE with Automated Cartridge Exchange

More information

FPD - Flame Photometric Detector

FPD - Flame Photometric Detector FPD - Flame Photometric Detector FPD Overview - The FPD detector is can be configured for 2 modes of operation, one is highly selective for Sulfur compounds and the other mode highly selective for Phosphorus

More information

Customised HPLC Systems for Highest Performance

Customised HPLC Systems for Highest Performance Customised HPLC Systems for Highest Performance SunChrom HPLC - A modular HPLC system with intelligent modules Our new HPLC system is composed of independent "stand alone" modules which contain their own

More information

Troubleshooting. Edwin A. Burgstaler, MT., HP. (ASCP) ASFA 2015 Annual Meeting May, MFMER slide-1

Troubleshooting. Edwin A. Burgstaler, MT., HP. (ASCP) ASFA 2015 Annual Meeting May, MFMER slide-1 Troubleshooting Edwin A. Burgstaler, MT., HP. (ASCP) ASFA 2015 Annual Meeting May, 2015 2011 MFMER slide-1 Troubleshooting Objectives The participant will recognize some problems in apheresis procedures

More information

Genuine Restek Replacement Parts for Agilent GCs

Genuine Restek Replacement Parts for Agilent GCs Genuine Restek Replacement Parts for Agilent GCs Chromatography Products 800-356-688 www.restek.com www.restek.com 800-356-688 84-353-300 From Injector to Detector Restek is your # source for consumables

More information

Application Note. The Agilent 1200 Infinity Series Online SPE Solution for the Highest Sensitivity. Author. Abstract.

Application Note. The Agilent 1200 Infinity Series Online SPE Solution for the Highest Sensitivity. Author. Abstract. Comparison of Online SPE Analysis and Direct Injection of Trace Level Estrogens in Drinking Water with the Agilent 646 and Agilent 649 Triple Quadrupole s The Agilent 1 Infinity Series Online SPE Solution

More information

Agilent 7650A Automatic Liquid Sampler

Agilent 7650A Automatic Liquid Sampler Agilent 7650A Automatic Liquid Sampler Installation, Operation, and Maintenance Agilent Technologies Notices Agilent Technologies, Inc. 2012 No part of this manual may be reproduced in any form or by any

More information

Nexis GC-2030 Specifications

Nexis GC-2030 Specifications Gas Chromatograph C184-E042A Chromatography System Nexis GC-2030 opens up an entirely new chapter in Shimadzu's long history with gas chromatographs. It showcases some of the best specifications on the

More information

FACSCalibur Startup Procedures

FACSCalibur Startup Procedures FACSCalibur Startup Procedures Always turn on the FACSCalibur before turning on the computer when acquiring data. This enables the computer to recognize that the cytometer is connected. When analyzing

More information

RapidTrace + Automated SPE Workstation. Instruments. Accessories Racks Vial Adapters Test Tubes O-Rings Miscellaneous

RapidTrace + Automated SPE Workstation. Instruments. Accessories Racks Vial Adapters Test Tubes O-Rings Miscellaneous RapidTrace + Automated SPE Workstation Instruments RapidTrace + Accessories Racks Vial Adapters Test Tubes O-Rings Miscellaneous Automate Your Extractions Increase your throughput and productivity The

More information

Methods of solid phase extraction. Methods of solid phase extraction

Methods of solid phase extraction. Methods of solid phase extraction Handling of columns For elution either apply pressure at the top of the column or apply vacuum at the column end. For this purpose several procedures are possible as shown in the adjacent figures. The

More information

Agricultural. drugs of abuse. Food & Beverage. Pesticides SOLID PHASE EXTRACTION. Inject. Wash Elute. Cartridges. Concentrate. Silica.

Agricultural. drugs of abuse. Food & Beverage. Pesticides SOLID PHASE EXTRACTION. Inject. Wash Elute. Cartridges. Concentrate. Silica. SOLID PHASE EXTRACTION Reproducible Results - Increased Productivity SOLID PHASE EXTRACTION SOLUTIONS Food & Beverage Drinking Water Clinical Ion Exchange condition Load Mix-Mode drugs of abuse Wash Elute

More information

Agilent Vac Elut SPS 24 Manifold User Guide

Agilent Vac Elut SPS 24 Manifold User Guide Agilent Vac Elut SPS 24 Manifold User Guide A clear revolution in sample processing Agilent offers the most technologically advanced sample processing station for solid phase extraction. Designed for use

More information

E²M. Innovation with Integrity. Enhanced Environmental Mass Spectrometer. Safety & Security

E²M. Innovation with Integrity. Enhanced Environmental Mass Spectrometer. Safety & Security E²M Enhanced Environmental Mass Spectrometer Innovation with Integrity Safety & Security Enhanced GC/MS System The E 2 M is the successor to the highly successful and proven GC/MS system EM 640. It represents

More information

THE RIGHT CONSUMABLES FOR YOUR GC APPLICATIONS. TurboMartix Headspace and Headspace Trap Samplers Consumables and Supplies Reference Guide

THE RIGHT CONSUMABLES FOR YOUR GC APPLICATIONS. TurboMartix Headspace and Headspace Trap Samplers Consumables and Supplies Reference Guide THE RIGHT CONSUMABLES FOR YOUR GC APPLICATIONS TurboMartix Headspace and Headspace Trap Samplers Consumables and Supplies Reference Guide INTRODUCTION AND OVERVIEW Headspace Gas Chromatography for applications

More information

STANDARDD OPERATING PROCEDURE FOR THE LUMINEX

STANDARDD OPERATING PROCEDURE FOR THE LUMINEX The Flow Cytometry Facility STANDARDD OPERATING PROCEDURE FOR THE LUMINEX All users must follow the SOP & fill in the logbook. If you are unsure about anything please do not hesitate to ask for help. Before

More information

Improved Dryer Control

Improved Dryer Control Improved Dryer Control Kenneth C. Hill President Kadant Johnson Systems Division David Vijh Senior Process Control Engineer Technical White Paper Series EXECUTIVE SUMMARY Significant improvements in dryer

More information

Gas Chromatograph. Features. Continuous Analysis of Ultra High Purity Argon to ppb levels.

Gas Chromatograph. Features. Continuous Analysis of Ultra High Purity Argon to ppb levels. Gas Chromatograph Continuous Analysis of Ultra High Purity Argon to ppb levels Features Following on from the success of the AGC Series 100 Argon GC, AGC Instruments introduces the new NovaCHROM 2000 GC.

More information

On-site Detection of Siloxanes and other unwanted VOCs in Landfill/Sewage/Biogas using Gas Chromatograph-Ion Mobility Spectrometer (GC-IMS)

On-site Detection of Siloxanes and other unwanted VOCs in Landfill/Sewage/Biogas using Gas Chromatograph-Ion Mobility Spectrometer (GC-IMS) On-site Detection of Siloxanes and other unwanted VOCs in Landfill/Sewage/Biogas using Gas Chromatograph-Ion Mobility Spectrometer (GC-IMS) by G.A.S. Gesellschaft für analytische Sensorsysteme mbh G.A.S.

More information

DETERMINATION OF PAH In Particulate Matter PM 10 with SPE/EVAporation

DETERMINATION OF PAH In Particulate Matter PM 10 with SPE/EVAporation DETERMINATION OF PAH In Particulate Matter PM 10 with SPE/EVAporation APPLICATION NOTE www.lctech.de As of: February 2017, Version: 1.2 Determination of PAH In particulate matter PM 10 with SPE/EVAporation

More information

Thermo Scientific Dionex AutoTrace 280 SPE Instrument. Automate Large-Volume. Solid-Phase Extractions

Thermo Scientific Dionex AutoTrace 280 SPE Instrument. Automate Large-Volume. Solid-Phase Extractions Thermo Scientific Dionex AutoTrace 280 SPE Instrument Automate Large-Volume Solid-Phase Extractions Thermo Scientific Dionex AutoTrace 280 System features and performance Automate Large Volume Solid-Phase

More information

McNeil Group University of Michigan. MBraun Glovebox. Preparing Items To Be Brought Into the Glovebox

McNeil Group University of Michigan. MBraun Glovebox. Preparing Items To Be Brought Into the Glovebox MBraun Glovebox Preparing Items To Be Brought Into the Glovebox Equipment Glassware, stirbars, needles, spatulas and pretty much anything that can go in the oven: should be dried in the oven for at least

More information

Analyze IT. Process Gas Chromatograph. PGC2000 Series

Analyze IT. Process Gas Chromatograph. PGC2000 Series Process Gas Chromatograph PGC2000 Series PGC2000 Series PGC2002 Process Distillation PGC2003 PNA PGC2005 Temperature Programmed GC PGC2007 Fuel Sulfur Analyzer PGC2008 Olefins Analyzer PGC2009 Fast TP

More information

Single Tower Heat Reactivated Desiccant Air Dryer

Single Tower Heat Reactivated Desiccant Air Dryer Single Tower Heat Reactivated Desiccant Air Dryer iste Dryer Understanding Compressed Air Moisture in compressed air is damaging. Using untreated air results in water and contaminants in your air lines,

More information

HS Dryer. Smart Solutions for Drying Compressed Air. Single Tower Heat Reactivated Desiccant Air Dryer. Engineered Air Products

HS Dryer. Smart Solutions for Drying Compressed Air. Single Tower Heat Reactivated Desiccant Air Dryer. Engineered Air Products Smart Solutions for Drying Compressed Air Single Tower Heat Reactivated Desiccant Air Dryer TM HS Dryer 800.943.794 www.engineeredairproducts.com Engineered Air Products Moisture in compressed air is damaging.

More information

Agilent InfinityLab Online SPE Solution

Agilent InfinityLab Online SPE Solution Agilent InfinityLab Online SPE Solution Instruction Agilent Online SPE Solution - Instruction Technical Information about the Agilent InfinityLab Online SPE Solution. The following possible set-ups are

More information

Reacti-Therm I and III Heating Modules

Reacti-Therm I and III Heating Modules INSTRUCTIONS Reacti-Therm I and III Heating Modules TS-18822 TS-18824 Number TS-18822 TS-18824 Description Reacti-Therm I Heating Module (single block) Reacti-Therm III Heating Module (triple block) 2101.1

More information

SEC-MALS Noise Assessment and Cleaning Guide

SEC-MALS Noise Assessment and Cleaning Guide Page 1 of 10 Technical Note TN3501 SEC-MALS Noise Assessment and Cleaning Guide Summary An important part of obtaining high quality SEC- MALS data is maintaining a clean system. Contaminants in the system

More information

HS Dryer. Smart Solutions for Drying Compressed Air. Single Tower Heat Reactivated Desiccant Air Dryer. Engineered Air Products

HS Dryer. Smart Solutions for Drying Compressed Air. Single Tower Heat Reactivated Desiccant Air Dryer. Engineered Air Products Smart Solutions for Drying Compressed Air Single Tower Heat Reactivated Desiccant Air Dryer TM HS Dryer 800.943.794 www.engineeredairproducts.com Engineered Air Products Understanding Compressed Air Moisture

More information

Online SPE-UHPLC-MS. CHRONOS for Control & System Integration. Flexible autosampler for sample preparation. Superior UHPLC system for analysis

Online SPE-UHPLC-MS. CHRONOS for Control & System Integration. Flexible autosampler for sample preparation. Superior UHPLC system for analysis Online SPE-UHPLC-MS Flexible autosampler for sample preparation Powerful SPE solutions for solid phase extraction Superior UHPLC system for analysis CHRONOS for Control & System Integration www.axel-semrau.de

More information

Sample Preparation Solid Phase Extraction Disks Cartridges. Horizon Technology, Inc. Salem, NH USA

Sample Preparation Solid Phase Extraction Disks Cartridges. Horizon Technology, Inc. Salem, NH USA Sample Preparation Solid Phase Extraction Disks Cartridges Horizon Technology, Inc. Salem, NH USA Extraction Techniques LLE SPE Sep funnel Continuous Manual Automated Extraction The Basics Extracting the

More information

NEOTeCHA Reactor Sampling Type PV

NEOTeCHA Reactor Sampling Type PV A safe and reliable sampling device to meet the demands of the chemical and pharmaceutical industry Features General applications To take a sample from reactors or vessels containing: poisoned media s

More information

AppNote 3/2009. Automation of Solid Phase Extraction Methods using a Robotic X-Y-Z Coordinate Autosampler with Software Control KEYWORDS ABSTRACT

AppNote 3/2009. Automation of Solid Phase Extraction Methods using a Robotic X-Y-Z Coordinate Autosampler with Software Control KEYWORDS ABSTRACT AppNote 3/2009 Automation of Solid Phase Extraction Methods using a Robotic X-Y-Z Coordinate Autosampler with Software Control Fred D. Foster, John R. Stuff, Jacqueline A. Whitecavage, Edward A. Pfannkoch

More information

Polar Air Dryer. Instructions Manual PDRCF PDRCF R E F R I G E R A T E D T Y P E C O M P R E S S E D A I R D R Y E R S

Polar Air Dryer. Instructions Manual PDRCF PDRCF R E F R I G E R A T E D T Y P E C O M P R E S S E D A I R D R Y E R S Polar Air Dryer R E F R I G E R A T E D T Y P E C O M P R E S S E D A I R D R Y E R S PDRCF1150029 PDRCF4602000 Instructions Manual TABLE OF CONTENTS 1. Important safety notes - Please read... 1 1.1. Transportation...

More information

22 The Flame Ionization Detector

22 The Flame Ionization Detector 22 General Information FID pneumatics Special considerations Conditions that prevent the detector from operating Detector shutdown Jets Automatic reignition Lit offset Procedure: Changing the auto-reignite

More information

(PD[ $LU 'U\HU 5 ( ) 5, * ( 5 $ 7 ( ' 7 < 3 ( & ( 6 6 ( ' $, 5 ' 5 < ( 5 6 ('5&) ದ ('5&) Instructions Manual

(PD[ $LU 'U\HU 5 ( ) 5, * ( 5 $ 7 ( ' 7 < 3 ( & ( 6 6 ( ' $, 5 ' 5 < ( 5 6 ('5&) ದ ('5&) Instructions Manual Instructions Manual 1. Important safety notes - Please read... 1 1.1. Transportation... 1 1.2 Positioning... 1 1.3 Installation... 1 1.4 Before operating... 1 1.5 Maintenance by an engineer... 1 1.6 Maintenance

More information

wash system V1.3 Important: Up to slides on one injection. Note: Getting need to turn it samples the next day. and put the slides in foil, put it

wash system V1.3 Important: Up to slides on one injection. Note: Getting need to turn it samples the next day. and put the slides in foil, put it Protocol for hybridization with MAUI hybridization and wash system V1.3 Important: Up to 12 slides can be hybridized in one run. If you hybridize 12 slides on one run, it is recommended that you separate

More information

DYE QUALITY CONTROL REFERENCE MATERIAL KIT

DYE QUALITY CONTROL REFERENCE MATERIAL KIT DYE QUALITY CONTROL REFERENCE MATERIAL KIT CONTENTS I. INTRODUCTION a. MS-Directed Purification b. The Purpose of a QCRM c. What Affects Your QCRM Results II. STORAGE AND STABILITY III. RECOMMENDED USAGE

More information

Training Fees 4,000 US$ per participant for Public Training includes Materials/Handouts, tea/coffee breaks, refreshments & Buffet Lunch.

Training Fees 4,000 US$ per participant for Public Training includes Materials/Handouts, tea/coffee breaks, refreshments & Buffet Lunch. Training Title ADVANCED GC PROCESS Training Duration 5 days Training Venue and Dates Advanced GC Process 5 17 21 February, 2019 $4,000 Dubai, UAE Trainings will be conducted in any of the 5 star hotels.

More information

Auto SPE-06Plus Solid Phase Extraction System. Reeko Instrument USA

Auto SPE-06Plus Solid Phase Extraction System. Reeko Instrument USA Auto SPE-06Plus Solid Phase Extraction System Reeko Instrument USA Reeko Overview Reeko Instrument focuses on the automatic sampler preparation instruments, R&D and manufacture. The company design and

More information

Chromatography. Research Supplies. 2014/2015 Catalog

Chromatography. Research Supplies. 2014/2015 Catalog Chromatography Research Supplies 2014/2015 Catalog The CRS Commitment Service That Rev lves Around You! Our purpose at CRS is to meet our customer s requirements for quality, service, and value. Quality

More information

- Data Brochure Steam Control 279

- Data Brochure Steam Control 279 - Data Brochure Steam Control 279 D 279 12/07 The tekmar Steam Control 279 can operate a single on-off steam boiler or an on-off steam valve using outdoor reset. The control determines the on time of the

More information

Application Note. Authors. Abstract

Application Note. Authors. Abstract Analysis of Four β-lactam Antibiotics in Water Using an Agilent 0 Infinity II LC and an Agilent 0 Triple Quadrupole LC/MS with Direct Injection and Online SPE Application Note Authors Alexander Timm, Rafael

More information

Automated Fractionation System that Increases Recovery and Optimize Precision of Solid Phase Extraction Methodology

Automated Fractionation System that Increases Recovery and Optimize Precision of Solid Phase Extraction Methodology Automated Fractionation System that Increases Recovery and Optimize Precision of Solid Phase Extraction Methodology Joan M. Stevens, Ph.D., and Mark Crawford Gilson, Inc., 3000 Parmenter Street., Middleton,

More information

THE RIGHT CONSUMABLES FOR YOUR GC APPLICATIONS. Clarus 590/690 Gas Chromatography Consumables and Supplies Reference Guide

THE RIGHT CONSUMABLES FOR YOUR GC APPLICATIONS. Clarus 590/690 Gas Chromatography Consumables and Supplies Reference Guide THE RIGHT CONSUMABLES FOR YOUR GC APPLICATIONS Clarus 590/690 Gas Chromatography Consumables and Supplies Reference Guide INTRODUCTION AND OVERVIEW For laboratories looking for a single gas chromatograph

More information

SPME-FFA Field Sampler

SPME-FFA Field Sampler Page 16 SPME-FFA Field Sampler Manual Page 1 Cat.No. 57554-U CHROMLINE Via Anita Garibaldi,40 59100 Prato (PO) Italy Phone: +39 335 1890076 Fax: +39 0574 660136 Email: info@chromlinesrl.com web: www.chromlinesrl.com

More information

Q Trap LC/MS/MS TurboIonSpray Ion Source Manual. About This Manual... 1 Conventions Introduction... 3 Features... 4

Q Trap LC/MS/MS TurboIonSpray Ion Source Manual. About This Manual... 1 Conventions Introduction... 3 Features... 4 Q Trap LC/MS/MS TurboIonSpray Ion Source Manual Table of Contents Table of Contents About This Manual................................................ 1 Conventions.........................................................

More information

Air Server & Multi Channel Samplers (MCS06 & MCS08)

Air Server & Multi Channel Samplers (MCS06 & MCS08) Air Server & Multi Channel Samplers (MCS06 & MCS08) Operators Manual AUGUST 2006 QUI-0004 VERSION 6.1 Ltd. T: +44 (0)1443 230935 F: +44 (0)1443 231531 E: enquiries@markes.com This page left intentionally

More information

Technical specifications of FT-IR Spectrometer

Technical specifications of FT-IR Spectrometer Technical specifications of FT-IR Spectrometer We would like to buy a FT-IR spectrometer which has to be integrated with a Gas Chromatograph with the following technical specifications. 1. The spectrometer

More information

Process Gas Chromatograph

Process Gas Chromatograph Product Data Sheet RAI-PDS-GC-1000A June 2009 Model 1000A Process Gas Chromatograph Emerson Process Management offers a complete line of gas chromatographs specifically designed for process applications.

More information

SystemsQT for Waters Alliance HPLC Systems (Revision G)

SystemsQT for Waters Alliance HPLC Systems (Revision G) R E L E A S E N O T E S SystemsQT for Waters Alliance HPLC Systems (Revision G) Waters has rereleased the Systems Qualification Tool (SystemsQT ) for Alliance HPLC Systems. These release notes detail the

More information

Tubing spares and piping accessories. Gas purification and control 1 Moisture Traps,Carbon traps 2 Gas Purification Panels 3 Gas Control Box

Tubing spares and piping accessories. Gas purification and control 1 Moisture Traps,Carbon traps 2 Gas Purification Panels 3 Gas Control Box Tubing spares and piping accessories 1 Tubing (SS) 2 Teflon tube 3 PU Tube 4 Moduler Piping Gas purification and control 1 Moisture Traps,Carbon traps 2 Gas Purification Panels 3 Gas Control Box Tubing

More information