1 NFPA Overview and 2000 to 2012 Update Sharon S. Gilyeat, P.E. - Principal Koffel Associates, Inc. MD, MA, NC, CT, IN Koffel Associates W.L.L. Qatar Koffel Associates International United Arab Emirates Expertly Engineering Safety From Fire 1
2 Background 6 years Federal Fire Protection Engineer and Safety Manager Koffel Associates, Inc. 27 years Director, Southeast Office Develop and teach for CMS Chairs NFPA 105 Lasers and serves on NFPA 99, 45, 101, and 82 technical committees Past State of MD fire commissioner
3 Introduction and Goal This seminar is focused on the changes to the NFPA 99 Health Care Facilities Code with emphasis on changes since 1999 Not all changes included!
4 CAUTIONS!! Not all changes to NFPA 99 reflected here Slides do not contain full code text only general intent always go to the code for the full text Exceptions may apply
5 History of Development 2005 last edition to 2012 Extended to 5 year cycle for rewrite Clean rewrite Performance-based Code vs Standard June 2009 NFPA rejected and send back to editing June 2012 NFPA accepted CMS adopted as part of their 2012 LSC adoption
6 Chapters Deleted for 2012 All Occupancy chapters Chapter 6 Environmental Systems Chapter 7 Materials Chapter 10 Manufacturers Requirements Chapter 11Laboratories Chapter 21 Freestanding Birthing Centers
7 Annexes Deleted Annex B Nature of Hazards Annex D Safe Use of High-Frequency Electricity Annex F Flammable Anesthetizing Locations
8 2012 Annexes Annex A: Explanatory material Keyed to specific code text Non-mandatory Annex B: Explanatory material Keyed to Chapter Non-Mandatory Annex C Sample Adopting Ordinance Annex D: Informational References
9 New Chapters for 2012 Chapter 4 Fundamentals Establishes Categories based on risk assessment Chapter 7 Information Technology and Communications Systems Chapter 8 Plumbing & Chapter 9 HVAC Chapter 13 Security Management Chapter 15Features of Fire Protection Strike out indicates not adopted by CMS
10 Chapter 12 Emergency Management Totally rewritten Coordinate with requirements of The Joint Commission Lessons learned from recent disasters Better integration with NFPA 1600 Standard on Disaster/Emergency Management and Business Continuity Programs. NOT ADOPTED BY CMS BUT VERY CLOSE TO CMS REQUIREMENTS
11 Navigation Editorial Marks Asterisks Brackets Vertical Lines Not used Bullets Not used NOTE: New NFPA policy on vertical lines and bullets very controversial
12 NFPA Chapter 1 Administration
13 Administration 1.1 Scope Totally rewritten with scopes essentially for each chapter 1.2 Purpose Provide minimum requirements for the installation, inspection, testing, maintenance, performance and safe practice for facilities, material, equipment, and appliances.
14 Application Applies to ALL heath care facilities other than home care Construction and Equipment requirements NEW only Alterations, renovations and modernization meet new An existing system not in compliance SHALL BE PERMITTED to be continued in use, unless the AHJ determines that such use is a DISTINCT hazard to life
15 Patient Care Rooms Governing body NOT AHJ shall determine Critical care rooms General care rooms Basic care rooms Support rooms Anesthesia locations Wet procedure locations (Further discussion later)
16 Equivalency Equivalencies approved by the AHJ shall be recognized as being in compliance with the Code AHJ authorized to grant exceptions to this code
17 NFPA Chapter 3 Definitions
18 Definition Changes Anesthetizing location Administration of general anesthesia Compact storage Defend in place The operational response to an emergency in a building, in which the initial action does not involve evacuation of the building occupants
19 Definition Changes General Anesthesia and Levels of Sedation General Anesthesia Deep Sedation/Analgesia Moderate Sedation/Analgesia (Conscious Sedation) Minimal Sedation (Anxlolysis) Medical Support Gas Support Gas no longer used
20 Definitions Patient Care Room Critical Care Room General Care Room Basic Care Room Support Room
21 Definitions Wet Procedure Location Annex note
22 NFPA Chapter 4 Fundamentals
23 Risk Assessment Building system categories based on the system, not the facility no caregiver intervention Category 1 major injury or death Category 2 minor injury Category 3 no injury but can cause discomfort Category 4 no impact on patient care Risk Assessment Extensive Annex material
24 Annex Material for Chapter 4 Examples of each Category Discussion of major injury Risk assessment ISO/IEC Risk Management Risk Assessment Techniques NFPA 551 Guild for the Evaluation of Fire Risk Assessments SEMI S E Safety Guideline for Risk Assessment and Risk Evaluation
25 NFPA Chapter 5 Medical Gas and Vacuum Systems
26 Medical Gas/Medical Support Gas Oxygen Nitrous oxide Medical air Carbon dioxide Helium Nitrogen Instrument Air
27 Vacuum Systems Medical-surgical vacuum Waste anesthetic gas disposal
28 Medical Gas and Vacuum Systems Hazards Components Category 1 gas and vacuum systems Category 2 gas and vacuum systems Category 3 gas and vacuum systems
29 Medical Gas and Vacuum Systems Hazards Fire Explosions Loss of systems Inadvertent cross-connections
30 Medical Gas and Vacuum Systems 3.32
31 Category 1 Medical Gas and Vacuum Systems Category 1 Medical Piped Gas and Vacuum Systems. Requirements apply to health care facilities that require Category 1 systems as determined using risk assessment as defined in Chapter 4
32 New Vs. Existing through , through existing and new and existing See Tentative Interim Amendment (TIA) TIA 12-4
33 Medical Gas and Vacuum Systems Sources Gas cylinders Cryogenic liquid containers Bulk cryogenic liquids Medical air compressors Medical/surgical vacuum pumps WAGD producers Instrument air compressors Proportioning systems for medical air USP
34 Category 1 Gas and Vacuum Systems Sources - Cylinders
35 Cryogenic Liquid System
36 Bulk Cryogenic Supply
37 Compressor
38 Compressor
39 Vacuum Pump
40 Vacuum Pump
41 Central Supply System - Operations Use of adapter/conversion fittings prohibited Handled in accordance with Chapter 11 Only gas cylinders and accessories are permitted to be stored in central supply room Wooden racks for cylinders permitted
42 Central Supply Room - Construction Lockable doors or gates If outdoors, noncombustible enclosure, with a minimum of two exits If indoors, noncombustible or limited combustible with 1-hour fire resistance rating (see 2015 for clarification) NFPA 70 for ordinary locations
43 Central Supply Room - Construction Indirect heating (steam, hot water) Racks, chains or other fasteners to secure all cylinders. Powered by essential electrical system Racks, shelves or supports made of noncombustible or limited combustible materials Electrical devices at or above 5 feet from floor protected from damage
44 Storage Full or empty, when not connected, shall be stored in locations complying with and (Construction and Ventilation) Cylinders shall not be stored in enclosures containing motor-driven equipment except instrument air reserve headers in same location with instrument air compressors
45 Final Line Pressure Regulators Isolation valves and or check valves to isolate each regulator Pressure indicator downstream of each regulator or immediately downstream of isolating valves for the regulators Piped to allow for indipendant operation Sized for 100 percent of the peak calculated demand Constructed of materials deemed suitable by the manufacturer
46 Category 1 Gas and Vacuum Systems Headers Manifolds Gas Cylinders Cryogenic Liquid Containers Manifolds for Gas Cylinders with Reserve Supply Entire section deleted
47 Bulk Cryogenic Liquid Systems Totally revised and coordinated with NFPA 55, Compressed Gases and Cryogenic Fluids Code
48 Category 1 Gas and Vacuum Systems Emergency Oxygen Supply Connection (EOSC) may be required for maintenance or emergencies Minimum of 3 ft clearance around EOSC Alternative is in-building emergency reserve The word any in may confuse this
49 Emergency Oxygen Supply Connection
50 In-Building Reserve
51 Category 1 Medical Air Supply Systems Medical air quality Be supplied from cylinders, bulk containers, medical air compressor sources, or be reconstituted from oxygen USP and oil-free, dry Nitrogen NF Meet the requirements of medical air USP Have no detectable liquid hydrocarbons Have less than 25 ppm gaseous hydrocarbons Have equal to or less than 51 mg/m3 of permanent particulates sized 1 micron or larger in the air at normal atmospheric pressure
52 Medical Air Compressors Location Indoors in a dedicated mechanical equipment area Adequate ventilation Ambient temperature range as recommended by the manufacturer.
53 Compressor
54 Category 1 Medical Air Supply Systems After coolers Receivers Dryers Filters All component materials are to be determined by the manufacturer
55 Compressor Intake Outdoors above roof level Draw from a clean source of air Located 25 ft. from ventilating system exhausts, vents, vacuum and WAGD discharges Minimum distance 20 ft. above ground Minimum of 10 ft from any door, window or other opening
56 Compressor Intake Intake: Hard-drawn seamless copper ASTM B819, medical gas tube ASTM B88, water tube, type K or L ASTM B280, 280 ACR as permitted for vacuum and WAGD systems. Intake turned down, screened, and noncorroding material
57 Medical Air Quality - Monitoring Dew point monitored and alarmed when dew point exceeds 2 o C (35 o F) CO alarm activated when CO exceeds 10 ppm Dew point and CO monitors activate alarm if power is lost.
58 Medical Air Proportioning System Entire new section and numerous changes throughout the code on medical air proportioning systems Reconstituting medical air from Oxygen USP and Nitrogen NF
59 Category 1 Med-Surg Vacuum Systems Pump Location Indoors in a dedicated mechanical equipment area Adequately ventilated Proper utilities Ambient temperature range per manufacturers recommendation
60 Medical-Surgical Vacuum Pumps
61 Medical-Surgical Vacuum Systems Medical-Surgical Vacuum Systems Vacuum pumps (2 or more) Peak calculated demand met with largest pump out of service Piping per vacuum piping except brass, galvanized or black steel as per manufacturer
62 Vacuum Receivers Materials deemed suitable by the manufacturer ASME Boiler and Pressure Vessel Code 60 PSI and vacuum of 30 in. of Hg Manual drain Capacity based on the technology of the pumps
63 Medical-Surgical Vacuum Systems Medical-Surgical Vacuum Exhaust Located outdoors At least 10 ft. from any door, window, air intake, or other opening or places of assembly Different level than intake Consider prevailing winds, adjacent buildings, topography and other influences End turned down and screened with a noncorroding material
64 Instrument Air Supply Systems Usage Instrument air shall be permitted to be used for any medical support purpose and if appropriate to the procedures to be used in laboratories Cannot be interconnected with medical air Cannot be used for any purpose where the air will be intentionally respired by humans
65 New or Replacement Valves
66 Zone Valves Wall between the valve and outlet/inlet Serves only outlet/inlets on same story Not located in a room that it controls Operable from standing position Pressure/vacuum indicator required Visible and accessible at all times Not in locked rooms, behind doors, open doors, or hidden
67 Zone Valves Immediately outside each vital life-support, critical care, and anesthetizing location of moderate sedation, deep sedation or general anesthesia Readily accessible in an emergency Shutting off the gas to one operating room or anesthetizing area will not affect others
68 Zone Valves ZONE VALVES SHALL BE LABELED IN ACCORDANCE WITH
69 Zone Valve
70 Zone Valve
71 Zone Valves Floor plan indicating areas controlled by valve
72 In-Line Check Valves New or replacement Brass or bronze Brazed extensions In-line serviceability NO threaded connections Threaded purge points on 1/8 in NPT
73 Alarm Requirements - General Separate visual indicators (except on master) Visual non-cancelable Cancelable audible Means to indicate a lamp or LED failure Visible and audible indication that the communication is disconnected. Labeling
74 Alarm Requirements General Reinitiation of the audible signal if another alarm condition occurs Power from life safety branch of the EES Wiring used for communications is protected Communication devices that do not use wire supervised such that failure will initiate alarm LABELING KEPT UP TO DATE!! Alarm switches/sensors removable
75 Alarm Requirements Master Alarm Monitors source supply, reserve source (if any) and pressure in main lines of medical gas and vacuum piping system Two master alarm panels required (on-site maintenance office and continuously staffed location) A centralized computer system can be substituted for one of the master alarm panels
76 Master Alarm Panel Master alarm panel in a facility engineering office
77 Alarm Requirements Master Alarm Master alarm panels required to communicate directly to the alarm-initiating device they monitor Each of the two mandatory alarms must be wired independently to the initiating devices
78 Alarm Requirements Area Alarms Location: 1)Anesthetizing locations where moderate sedation, deep sedation, or general anesthesia 2)Critical Care areas
79 Alarm Requirements Area Alarm Area alarm sensor location Vital life support or critical care area: On patient (use side) of zone valve box assembly Anesthetizing area (moderate sedation, deep sedation or general anesthesia): Either on source side or patient (use side) of zone valve box assembly
80 Alarm Requirements Local Alarm Monitors air compressors, medical/surgical vacuum pump systems, WAGD systems, instrument air, and proportioning systems One signal located on the control panel for the machinery being monitored, or within monitoring device or on separate panel One signal connected to master alarm panel 11 different items to be monitored
81 Flexible Joints Metallic flexible joints permitted where required for expansion, seismic, thermal expansion or vibration and: Bronze, copper or stainless steel Cleaned at factory for oxygen service 300 psi and 1000 o F Brazing extensions Supported as required for their weight
82 Installer Qualifications Installation shall be made by qualified, competent technicians who are experienced in making such installations, including all personnel who actually install the pipe Must meet ANSI/ASSE Standard 6010, Professional Qualification Standard for Medical Gas and Vacuum System Installers
83 Installer Qualifications Installers shall not use their certification to oversee installation by noncertified personnel
84 Installer Performed Tests The 24-hour standing pressure test of positive pressure systems and vacuum systems shall be witnessed by AHJ or designee. Form indicating that this test was performed and witnessed shall be provided to the verifier prior to verifier testing
85 Category 1 Operation and Management Special precautions Piping not used for grounding Medical-Surgical Vacuum and WAGD not used for nonmedical applications Maintenance (new section) General Facilities shall develop and document periodic maintenance programs for these systems and subcomponents
86 Maintenance Programs Inventory all sources, control valves, alarms, manufactured assemblies and I/O Inspection Schedules scheduled inspections shall be established through risk assessment, manufacturers recommendations and requirements of AHJ
87 Maintenance Programs Inspection Procedures inspection procedures and test methods established through risk assessment Maintenance schedules - scheduled maintenance shall be established through risk assessment, manufacturers recommendations and requirements of AHJ
88 Maintenance Programs Qualifications persons maintaining such systems shall be qualified any of the following: Training and certification through the facility ASSE 6040 Professional Qualification Standard for Medical Gas Maintenance Personnel ASSE 6030 Professional Qualification Standard for Medical Gas Systems Verifiers
89 Inspection and Testing Operations Medical air sources Room temperature Room ventilation Shaft seal condition Filter condition Presence of hydrocarbons Water quality if so equipped Intake location Air purity Dew point Carbon monoxide monitor calibration
90 Inspection and Testing Operations Vacuum source - WAGD Exhaust location Instrument air source Filter condition Manifold sources Ventilation Enclosure labeling
91 Inspection and Testing Operations Bulk cryogenic liquid sources per NFPA 55 Final line regulator delivery pressure Valves labels Alarms lamp and audio operation Master, area, local signal operation Station I/O Flow, labeling, latching/delatching/ leaks
92 Category 3 Gas and Vacuum Systems Many similarities with Category 1 but in general less stringent. Existing and New & existing , 5.3.2, & See TIA 12-4 Shall NOT serve more than two adjoining single treatment facilities
93 Category 3 Oxygen and Nitrous Oxide Source storage Total of all gases more than 3000 ft 3 (5000 ft 3 if in DOT 4L cryogenic liquid container) per Category 1 Smaller quantities per Category 3
94 NFPA Chapter 6 Electrical Systems
95 Electrical Systems Moved after gas systems Electrical hazards Electrical distribution system Type 1, 2 and 3 electrical system
96 Electrical Systems New versus existing
97 Electrical Hazards Fire and explosion Electrical shock Thermal energy Interruption of power
98 Distribution Follow NFPA 70 - NEC Critical care areas shall be served by circuits from a critical branch served from a single automatic transfer switch AND A minimum of one circuit by normal power or by a second critical branch automatic transfer switch.
99 Critical Care Areas
100 Critical Care Areas
101 Distribution Access to overcurrent protective devices Only authorized shall have access to those serving Category 1 or Category 2 rooms Shall not be permitted in public access spaces Isolated power panels in critical care areas shall be permitted in those areas.
102 Distribution Receptacles Patient bed general care (category 2)- 4 8 Patient bed critical care (category 1) 6 14 Operating rooms (category 1) - 36 Bathrooms-none Special areas (psychiatric, pediatric, etc) none Pediatric shall have listed tamper resistant or a listed tamper resistant cover (new and existing)
103 Receptacles
104 Quiet Grounds Now referred to as isolated grounds
105 Wet Procedure Locations Special protection against electrical shock Inherently limits ground fault current without interrupting power (IP) or Interrupts power when ground fault current exceeds 6 ma Operating rooms shall be considered wet procedure locations unless risk assessment determines otherwise
106 Wet Procedure Locations Where power interruption under first fault condition is tolerable, the use of a ground-fault circuit interrupter (GFCI) shall be permitted as the protective means that monitors the actual ground-fault current and interrupts the power when that current exceeds 6 ma.
107 Categories Category 1 rooms shall be served only by Type 1 EES Category 2 rooms shall be served by Type 1 or Type 2 EES A Type 1 EES shall be permitted to serve Category 2 rooms in the same facility Basic care rooms shall not be required to be served by EES
108 Unit Lighting Battery-Powered Lighting Units Shall be provided in deep sedation and general anesthesia areas Sufficient to terminate procedures Wired to general lighting branch circuit 1 ½ hour duration Tested monthly for 30 seconds and annually for 30 minutes
109 Receptacle Testing
110 Maintenance and Testing Hospital grade at patient bed locations and where deep sedation or general anesthesia is administered at initial installation, replacement or servicing Additional testing at patient care areas performed at intervals defined by documented performance data Non-hospital grade at patient bed locations and where deep sedation or general anesthesia is administered within 12 months
111 Essential Electrical System
112 Essential Electrical System
113 Type 1 Essential Electrical System Dual sources of normal power shall be considered (not necessary) Minimum two independent sources of power: normal and one or more alternate sources
114 Generator
115 Selective Coordination Overcurrent protective devices serving the EES shall selectively coordinate for the period of time that a fault s duration extends beyond 0.1 seconds Not required as follows: Between transformer primary and secondary where only one or one set of devices exist on the secondary Between devices of the same amp rating in series
116 Type 1 Essential Electrical System Branches Life Safety Branch Critical Branch Equipment Division shall occur at transfer switches when more than one transfer switch is required.
117 Type 1 Essential Electrical System Number of transfer switches shall be based on reliability, design and load Each branch shall have one or more transfer switches One transfer switch shall be permitted to serve one or more branches in a facility with a continuous load on the switch of 150 kva (120 kw) or less
118 Life Safety Branch Elevator lighting, control, communications Automatic doors for egress Fire alarm auxiliary functions Alarm & alerting non-fire(life safety or critical) Generator accessories NO other functions
119 Critical Branch
120 Type 1 Essential Electrical System Critical Branch Blood, bone and tissue banks Telephone equipment Task Illumination and selected receptacles and power in certain areas Additional task illumination, receptacles, and select power circuits needed for effective facility operations, including single-phase fractional horsepower motors.
121 Type 2 Essential Electrical System Emergency Systems Critical Systems Life Safety Branch Equipment Branch One or more transfer switches, single transfer switch permitted under same conditions as Type 1
122 Type 2 Essential Electrical System Equipment Branch Non-delayed-automatic Generator accessories Delayed-automatic connection Task illumination (some areas) HVAC airborne infectious isolation rooms Sump pumps Smoke control systems and stair pressurization Kitchen hood exhaust if required during a fire
123 NFPA Chapter 7 IT and Communication Systems
124 Information Technology and Communication Systems NOT ADOPTED BY CMS Still an ANSI Standard Brand new chapter still some coordination issues Acronyms: Entrance Facility (EF or TEF) Telecommunications Equipment Room (TER) Telecommunications Room (TR) Outside Plant (OSP) Infrastructure
125 NFPA Chapter 8 Plumbing Chapter 9 HVAC
126 Chapter 8 Plumbing New Chapter NOT REFERENCED BY CMS No risk categories Primarily references to plumbing code Some limited information on Grease interceptors Gray waste and clear waste water
127 Chapter 9 - HVAC New Chapter
128 HVAC - General Comply with ASHRAE 170 Laboratories - NFPA 45 Energy conservation ASHRAE 90.1 or local code Commissioned per ASHRAE 90.1, ASHRAE Guidelines 0 or 1
129 HVAC - General Piping per applicable plumbing code Ductwork NFPA 90A or applicable mechanical code
130 Medical Gas Storage & Transfilling All gases other than medical gases ventilation per NFPA 55 Outdoor storage/installations of medical gases and cryogenic fluids Ventilation per NFPA 55 Chapter 11 medical gases no requirements Transfilling ventilation per NFPA 55
131 Medical Gas Storage & Transfilling Indoor storage or manifold areas; storage or manifold buildings, for medical gases and cryogenic fluids shall be provided with natural or mechanical exhaust ventilation
132 Natural Ventilation Two nonclosable louvered openings, each 155cm 2 /35 L (24 in 2 /1000 ft 3 ) of the fluid (liquid or gas) stored, but not less than 465 cm 2 (72 in 2 ) One within 300 mm (1 ft) of the floor, and one within 300 mm (1 ft) of the ceiling Ensure cross ventilation Direct to outside without ductwork Mechanical if natural cannot be met
133 Mechanical Ventilation Mechanical exhaust continuous negative pressure unless approved by the AHJ I L/sec per 300 L (1 cfm per 5 ft 3 ) of fluid, > 24 L/sec (50 cfm) < 235 L/sec (500 cfm) Exhaust inlets within 30 mm (1 ft) of floor Powered by the EES Dedicated exhaust not required but not connected to spaces with flammables OR combustibles
134 Mechanical Ventilation Duct material noncombustible Make-up air (one of the following) From outside or adjacent space with no flammable or combustible materials via noncombustible ductwork From corridors under corridor door up to 24 L/sec (50 cfm) or 15% of the room exhaust per NFPA 90A Ok from building ventilation system
135 Mechanical Ventilation Waste Gas WAGD or by an active or passive scavenging system. Medical Plume Evacuation (such as from lasers) Direct to the outside, or HEPA and direct connection to return or exhaust duct Chemical or thermal sterilization and return to the space
136 EPS Room HVAC NFPA 110 Extracted from NFPA 110 Maintaining engine water jacket temperature Ventilation for cooling room Ventilation for cooling engine Ventilation for combustion
137 NFPA Chapter 10 Electrical Equipment Chapter 11 Gas Equipment
138 Patient Care Vicinity Chapter A space within a location intended for the examination and treatment of patients, extending 1.8 m (6 ft) beyond the normal location of the bed, chair, table, treadmill or other device that supports the patient during examination and treatment and extending vertically to 2.3 m (7 ft 6 in) above the floor.
139 Electrical Equipment Testing Requirements Fixed and Portable Physical integrity new annex note Resistance Leakage current tests Fixed equipment Touch current portable equipment Leakage current tests portable equipment
140 Chapter 11- Gas Equipment Storage of gases and cylinders Performance criteria and testing Administration
141 Gas Cylinders
142 Gas Equipment Storage of cylinders (non-flammable gases) More than 3000 ft 3, use Level 1 storage requirements
143 Gas Equipment Storage of cylinders (non-flammable gases) More than 300 ft 3 but less than 3000 ft 3 Outdoors or secured enclosed interior space of noncombustible or limited combustible construction Secured Can t mix oxidizing gases with flammable liquids, gases or vapors Separate oxidizing gases from combustible materials
144 Gas Equipment Separation Minimum of 20 ft Minimum of 5 ft if AS Fire rated cabinet Several other limitations from Chapter 5
145 Gas Equipment Storage of cylinders (non-flammable gases) Less than 300 ft 3 Patient care area not larger than 22,500 ft 2, individual cylinders not required to be in enclosures Precautions in handling apply Cylinders in use do not count Cylinders available for IMMEDIATE use are not in storage
146 Gas Equipment Performance Criteria and Testing Alterations and repairs require testing Administration Eliminate ignition sources Control flammable substances Service equipment Control transfilling of cylinders and containers Signage
147 Gas Equipment Operation and Management of Cylinders Administration Special precautions for handling Special precautions for making connections Special precautions for safety mechanisms Special precautions for storage
148 Gas Equipment Liquid oxygen equipment New section
149 NFPA Chapter 12 Emergency Management Chapter 13 Security
150 Chapter 12 -Emergency Management Total rewrite of the chapter Coordination with The Joint Commission Coordination with NFPA 1600 Standard on Disaster/Emergency Management and Business Continuity Programs NFPA 1620 Recommended Practice for Pre- Incident Planning Extensive annex notes
151 Chapter 13 Security Management New Chapter Facility shall have a security management plan Security vulnerability analysis (SVA) Responsible person Over 30 item listed in the responsibilities!!!
152 Security Management NOT adopted by CMS Still an ANSI document best practice. Almost laid out as a check list Extensive good annex notes Excellent commentary in NFPA 99 Handbook
153 NFPA Chapter 14 Hyperbaric Chambers
154 Hyperbaric Chambers
155 Hyperbaric Chambers Three classes of chambers Class A Human, multiple occupancy Class B Human, single occupancy Class C Animal no human occupancy Construction and equipment Room containing the chamber shall be AS Piping requirements Totally new requirements
156 Changes for 2012 Time to evacuate the chamber Restriction on Lithium and lithium-ion batteries Chamber gas supply monitoring Use of class I and II lasers Fire protection equipment inside hypobaric chambers
157 NFPA Chapter 15 Fire Protection Features
158 Fire Protection Features New Chapter Heavily based on NFPA 101 Numerous extracts from NFPA 101
159 Fire Protection Features Sprinklers New Closet Exception Hospitals only Patient room closets Under 6 sq ft
160 Fire Alarm Systems Fire Protection Features
161 Fire Protection Features Manual extinguishing equipment Portable fire extinguishers Standpipes Compact storage Maintenance and testing Operating rooms
162 Health Care Facilities Handbook Thank You!
163
164 Future Questions. Join our LinkedIn Discussion Groups Life Safety Code People can join the group and submit their questions or People can and submit anonymously Smoke Management Proposed Changes to the IBC Ask the Expert FAQ.Koffel.com Expertly Engineering Safety From Fire 367
165 Questions? Sharon S. Gilyeat, P.E. Koffel Associates, Inc Ayrsley Town Blvd. / Suite 202 / Charlotte, NC tel Corporate Office/ / fax Follow us on LinkedIn Expertly Engineering Safety From Fire 368
The Impact the Reference Standards in the 2012 Edition of NFPA 101 have on the Operation of Health Care Facilities: NFPA 80 and 99
The Impact the Reference Standards in the 2012 Edition of NFPA 101 have on the Operation of Health Care Facilities: NFPA 80 and 99 Sharon S. Gilyeat, P.E. Lauren Schrumpf August 8, 2018 Expertly Engineering
More information4/18/2017. *2012 edition of NFPA 99 has gone through a major overhaul. *This document is now a Code.
Carsten Lunde, ASSE 6050 Instructor April 20, 2017 Carsten s Yearly Analysis, Inc. Any opinions expressed verbally by the speaker before you and all text in this presentation should be considered the speakers
More information27th Annual AHCA Seminar
27th Annual AHCA Seminar NFPA 99 2012 Edition Overview and Discussion Presented by James Peterkin, PE Sr. Fire Protection Engineer Heery International Monday September 26, 2011 Disclaimer Not Speaking
More informationNFPA 45 NFPA 211 NFPA 1600
NFPA 45 NFPA 211 NFPA 1600 Prepared by: William E. Koffel, P.E., FSFPE Expertly Engineering Safety From Fire 1 Objectives Overview of the NFPA process Understand the scope and application of NFPA 45, NFPA
More informationNFPA 99 for Facility Managers
NFPA 99 for Facility Managers Application to Existing Facilities October 13, 2015 Jonathan Hart: Senior Fire Protection Engineer Learning Objectives Review the layout of NFPA 99 Understand the application
More informationMedical Gas Inspections and Maintenance Programs. HESNI Annual Conference NIU Conference Center May 10-11, 2018 Presented by Mathis Carlson
Medical Gas Inspections and Maintenance Programs HESNI Annual Conference NIU Conference Center May 10-11, 2018 Presented by Mathis Carlson Mathis Carlson I have worked for the last 15+ years in the Medical
More informationChanges to Environment of Care, Equipment Management, and Life Safety Chapters Related to Life Safety Code Updates
Issued September 20, 2017 Changes to Environment of Care, Equipment Management, and Life Safety Chapters Related to Life Safety Code Updates Environment of Care (EC) Chapter Home Care (OME) Accreditation
More informationPrepublication New Text Jan 2018 Changes to Environment of Care and Life Safety Chapters
Prepublication New Text Jan 2018 Changes to Environment of Care and Life Safety Chapters EC.01.01.01 EP3 Related to Life Safety Code Updates Hospital (HAP) Accreditation Program The hospital has a library
More informationEC EP 14 All other electrical K-920 equipment in patient care vicinity NFPA , Chapter 10
MSL Healthcare Partners, Inc. Wild Card EPs Paraphrase K-tag # EC.02.04.03 EP 14 All other electrical K-920 equipment in patient care vicinity NFPA 992012, Chapter 10 K-924 K-928 K-931 K-tag Description
More informationAMENDMENTS TO THE BUILDING AND FIRE CODE FOR THE CITY AND COUNTY OF DENVER
AMENDMENTS TO THE BUILDING AND FIRE CODE FOR THE CITY AND COUNTY OF DENVER The 2016 Denver Building and Fire Code includes the following codes except as amendment herein. 2015 IBC (Building) 2015 IFC (Fire)
More informationNFPA 101, 2012 EDITION IMPACT ON NEW AND EXISTING HEALTH CARE FACILITIES
NFPA 101, 2012 EDITION IMPACT ON NEW AND EXISTING HEALTH CARE FACILITIES Sharon S. Gilyeat, P.E., Principal Koffel Associates, Inc. www.koffel.com sgilyeat@koffel.com Expertly Engineering Safety From Fire
More informationPublic Safety in the Healthcare Facility
CABMET Study Group Public Safety in the Healthcare Facility Matt Baretich Baretich Engineering Fort Collins, Colorado mfb@baretich.com Slide 1 Public Safety: 15% of CBET Exam Electrical Chemical Radiation
More information2018 Joint Commission Standards What s New in the EC/LS/UM GASHE Annual March Meeting Lake Blackshear Resort
2018 Joint Commission Standards What s New in the EC/LS/UM GASHE Annual March Meeting Lake Blackshear Resort Presented March 9, 2018 By Ken Gregory, CHSP President Ken Gregory Consulting 1 2018 Joint Commission
More informationThere are many changes to NFPA 99 from 1993 to 2015, but we want to concentrate on the changes made from 2012 to 2015, and we need to know a little
There are many changes to NFPA 99 from 1993 to 2015, but we want to concentrate on the changes made from 2012 to 2015, and we need to know a little about how all of this developed into the 2015 edition.
More informationCompliance w/ the 2012 Life Safety Code
Compliance w/ the 2012 Life Safety Code Implications of the 2012 NFPA Code Set for Existing & New ASCs William E. Lindeman, AIA http://www.weldesigns.com Financial Interest Disclosure The presenter provides
More informationWHAT S NEW IN NFPA
WHAT S NEW IN NFPA 99-2012 North Carolina Hospital Engineers Association Fall Conference Grove Park Inn Asheville, NC Presented August 24, 2017 By Ken Gregory, CHSP Who Regulates Hospitals? Congress Healthcare
More informationWHAT S NEW IN NFPA
WHAT S NEW IN NFPA 99-2012 North Carolina Hospital Engineers Association Fall Conference Grove Park Inn Asheville, NC Presented August 24, 2017 By Ken Gregory, CHSP Who Regulates Hospitals? Congress Healthcare
More informationSEMICONDUCTOR FABRICATION FACILITIES
CHAPTER 18 SECTION 1801 GENERAL 1801.1 Scope. Semiconductor fabrication facilities and comparable research and development areas classified as Group H-5 shall comply with this chapter and the International
More informationSEMICONDUCTOR FABRICATION FACILITIES
CHAPTER 18 SEMICONDUCTOR FABRICATION FACILITIES SECTION 1801 GENERAL 1801.1 Scope. Semiconductor fabrication facilities and comparable research and development areas classified as Group H-5 shall comply
More informationChanges to Environment of Care and Life Safety Chapters Related to Life Safety Code Updates
Issued June 28, 207 Changes to Environment of Care and Life Safety Chapters Related to Life Safety Code Updates Environment of Care (EC) Chapter Nursing Care Center (NCC) Accreditation Program Standard
More information(h) Pesticide Application.
Fire Code Cannabis Amendment Section 13.20.066 is added to the Marysville Municipal Code to read in its entirety as follows: 13.20.066 Cannabis Growing, Processing, or Extraction Facilities (a) Application.
More informationThe Healthcare Environment Update
2017 The Healthcare Environment Update John Maurer, SASHE, CHFM, CHSP Engineering Department The Joint Commission Engineering Department 2017-1 Most Cited Standards Standard 2015 % Noncompliance 2016 %
More informationChanges to Environment of Care and Life Safety Chapters Related to Life Safety Code Updates
Issued September 20, 2017 Changes to Environment of Care and Life Safety Chapters Related to Life Safety Code Updates Environment of Care (EC) Chapter Ambulatory Health Care (AHC) Accreditation Program
More informationChanges to Environment of Care and Life Safety Chapters Related to Life Safety Code Updates
Issued September 20, 2017 Changes to Environment of Care and Life Safety Chapters Related to Life Safety Code Updates Environment of Care (EC) Chapter Critical Access Hospital (CAH) Accreditation Program
More informationSEMICONDUCTOR FABRICATION FACILITIES
CHAPTER 18 SEMICONDUCTOR FABRICATION FACILITIES SECTION 1801 GENERAL 1801.1 Scope. Semiconductor fabrication facilities and comparable research and development areas classified as Group H-5 shall comply
More informationLife Safety Code NFPA 10 NFPA 13 NFPA 25 NFPA 72 NFPA 90A NFPA 110 NFPA 99 THE 2012 EDITION OF THE LIFE SAFETY CODE HOSPITAL EDITION SESSION #7
THE 2012 EDITION OF THE LIFE SAFETY CODE HOSPITAL EDITION SESSION #7 AGENDA: Life Safety Code Referenced Publications Overview Life Safety Code NFPA 10 NFPA 13 NFPA 25 NFPA 72 NFPA 90A NFPA 110 NFPA 99
More informationChanges to the 2018 Joint Commission Comprehensive Accreditation Manual for Hospitals (CAMH)
Changes to the 2018 Joint Commission Comprehensive Accreditation Manual for Hospitals (CAMH) The following represents a review of the Life Safety chapter of the new 2018 Joint Commission CAMH manual for
More informationCHAPTER 7 ALTERATIONS LEVEL
CHAPTER 7 ALTERATIONS LEVEL 2 SECTION 701 GENERAL 701.1 Scope. Level 2 alterations as described in Section 404 shall comply with the requirements of this chapter. Exception: Buildings in which the reconfiguration
More informationFirst Aspen Federal Regulation Set: K LSC 2000 Health Existing
First Aspen Federal Regulation Set: K 01.01 LSC 2000 Health Existing K 0000 Title CFR Type INITIAL COMMENTS Memo K 0011 If the building has a common wall with a nonconforming building, the common wall
More informationAgency for Health Care Administration
Page 1 of 21 ST - K0000 - INITIAL COMMENTS Title INITIAL COMMENTS Statute or Rule Type Memo Tag These guidelines are meant solely to provide guidance to surveyors in the survey process. ST - K0300 - Other
More informationCourse Outline. Course Introduction Purpose of the course Course Overview Program Outline Instructor Information
The main focus of the ASSE 6040 Medical Gas Systems Maintenance Personnel Course is to provide individuals, who are responsible for the proper inspection, testing, and maintenance of medical gas systems,
More informationIMC Significant Changes Summary Tier I
2009-2015 IMC Significant Changes Summary Tier I 202 Environmental Air The definition of environmental air has been expanded through the addition of parking garage exhaust. 306.1 Access More than just
More informationKey Elements of NFPA 99 (and the other NFPA standards): Perspectives from the Safety Office. April, 2013
Key Elements of NFPA 99 (and the other NFPA standards): Perspectives from the Safety Office April, 2013 creative thinking. custom solutions. creative thinking. custom solutions. Which Version(s) of NFPA
More informationNFPA 101 Code Update from 2012 Edition to the 2015 Edition
NFPA 101 Code Update from 2012 Edition to the 2015 Edition Presented by: Gary Furdell AHCA Fire Protection Specialist 1 2 2012 Update to 2015 Chapter 18 New Health Care Occupancies 4 18.1.3.2 Atrium walls
More informationWhich Version(s) of NFPA 99 Apply?
Key Elements of NFPA 99 (and the other NFPA standards): Perspectives from the Safety Office April, 2013 creative thinking. custom solutions. creative thinking. custom solutions. Which Version(s) of NFPA
More informationTOPICS IN CODE ENFORCEMENT
TOPICS IN CODE ENFORCEMENT CAHED SE DISTRICT MEETING FEBRUARY 17, 2017 Mark Jelinske Cator, Ruma, and Associates mjelinske@catorruma.com 303-462-8475 TALKING POINTS Conducting Surveys on NFPA 2012 as of
More informationSEMICONDUCTOR FABRICATION FACILITIES
CHAPTER 18 SEMICONDUCTOR FABRICATION FACILITIES SECTION FC 1801 GENERAL 1801.1 Scope. This chapter shall govern the design, installation, operation and maintenance of semiconductor fabrication facilities
More informationPROPOSED CODE CHANGES FOR REVIEW BY THE FIRE CODE ACTION COMMITTEE
PROPOSED CODE CHANGES FOR REVIEW BY THE FIRE CODE ACTION COMMITTEE Developed by the Midwestern Regional Work Group CONTENTS Item 15 Limited area sprinkler systems Item 20 Scope of IFC Ch 7 Item 21 Fire-resistance-rated
More informationANNEX AMENDMENTS TO THE INTERNATIONAL CODE FOR FIRE SAFETY SYSTEMS (FSS CODE) CHAPTER 1 GENERAL
Annex 2, page 2 ANNEX AMENDMENTS TO THE INTERNATIONAL CODE FOR FIRE SAFETY SYSTEMS (FSS CODE) CHAPTER 1 GENERAL Section 1 Application 1 The following new sentence is added to the end of paragraph 1.2:
More informationCARBON DIOXIDE SYSTEMS FOR BEVERAGE DISPENSING PERMIT REQUIREMENTS
CARBON DIOXIDE SYSTEMS FOR BEVERAGE DISPENSING PERMIT REQUIREMENTS Town of Brighton Office of the Fire Marshal October 2018 Carbon Dioxide Systems for Beverage Dispensing Requirements Scope The intent
More information2012 International Fire Code Significant Changes. IFC Background. Purpose IBC and 2012 NFPA 101 Significant Means of Egress Changes
2012 International Fire Code Significant Changes Andrew M. Schneider, P.E. Senior Fire Protection Engineer aschneider@koffel.com Koffel Associates, Inc. www.koffel.com Expertly Engineering Safety From
More informationUpdate all extract references to NFPA documents (including NFPA 72) in Chapter 3 and related annex material to the latest editions.
11/16/12 TerraView First Revision No. 15-NFPA 720-2012 [ Global Input ] Update all extract references to NFPA documents (including NFPA 72) in Chapter 3 and related annex material to the latest editions.
More informationWinnipeg Fire Department Fire Prevention Branch
Winnipeg Fire Department Fire Prevention Branch Manitoba Fire Code Life-Safety Equipment Maintenance Requirements The Fire Prevention Branch of the Winnipeg Fire Department has prepared this document of
More informationICC Update - Laboratory Issues Code Changes / Proposals / Trends of Interest 10/2005
ICC Update - Laboratory Issues Code Changes / Proposals / Trends of Interest 10/2005 Kenneth Kretchman, CIH, CSP Mgr, Industrial Hygiene NC State University Ken_Kretchman@ncsu.edu International Building
More informationFIRE SAFETY SURVEY REPORT AMBULATORY SURGICAL CENTERS (ASC) & END STAGE RENAL DISEASE (ESRD) MEDICARE
DEPARTMENT OF HEALTH AND HUMAN SERVICES CENTERS FOR MEDICARE & MEDICA SERVICE Form Approved OMB Exempt FIRE SAFETY SURVEY REPORT AMBULATORY SURGICAL CENTERS (ASC) & END STAGE RENAL DISEASE (ESRD) MEDICARE
More informationChanges in NFPA
Changes in NFPA 72-2002 2002 Oregon Fire Code Committee November 25, 2003 Michael B. Baker, SET Automatic Fire Alarm Association MikeBaker@afaa.org 2003 Michael B. Baker 1 Table of Contents 1. Administration
More informationNEC Requirements for Standby Power Systems. New England Building Officials Education Association Annual Conference October 5, 2015
NEC Requirements for Standby Power Systems New England Building Officials Education Association Annual Conference October 5, 2015 Emergency & Standby Power Systems Alternative power sources are utilized
More informationNORTH CAROLINA. Downloaded (l) For each nursing unit, or fraction thereof on each floor, the following shall be provided:
Housekeeping/Laundry/Maintenance NORTH CAROLINA Downloaded 01.15.11 (l) For each nursing unit, or fraction thereof on each floor, the following shall be provided: (3) a soiled utility room with counter,
More informationAgency for Health Care Administration
Page 1 of 92 ST - K0000 - INITIAL COMMENTS Title INITIAL COMMENTS Type Memo Tag These guidelines are meant solely to provide guidance to surveyors in the survey process. ST - K0100 - General Requirements
More informationRiverside County Fire Department Office of the Fire Marshal Market St., Ste. 150, Riverside, CA Ph. (951) Fax (951)
Riverside County Fire Department Office of the Fire Marshal 2300 Market St., Ste. 150, Riverside, CA 92501 Ph. (951) 955-4777 Fax (951) 955-4886 Technical Policy Title: Plant Processing and Extraction
More informationLaboratories and Hazardous Materials. Company Overview
Laboratories and Hazardous Materials Don E. Contois, P.E. Sam Sinapi Company Overview R.W. Sullivan Engineering (RWS) was established in 1945 and currently has 95 employees. RWS is a full service engineering
More informationFacility Sourcing Practices and Joint Commission Update
Facility Sourcing Practices and Joint Commission Update Presenters: John Brownrigg Director, Pre-Construction Services FX Facility Group Larry Rubin Life Safety Code Surveyor The Joint Commission ACE 2018
More informationFHCA 2014 Annual Conference & Trade Show
FHCA 2014 Annual Conference & Trade Show CE Session #5 Preparing for the Upcoming Code Changes Don t Look Back Tuesday, July 8 8:15 to 9:15 a.m. Canary 3 Operations/Quality Improvement Upon completion
More information2018 IFC Update. Based on the 2018 International Fire Code, (IFC ) Goal. Objectives. Content
Based on the 2018 International Fire Code, (IFC ) The International Fire Code (IFC ), establishes minimum regulations for fire safety. This handout will identify important changes in the IFC from 2015
More information2.07 Permit Application Checklist for Diesel Generators, Diesel Fire Pumps, and Fuel Tanks Serving Generators and Fire Pumps (2016)
2.07 Permit Application Checklist for Diesel Generators, Diesel Fire Pumps, and Fuel Tanks Serving Generators and Fire Pumps (2016) Reference: Documents referenced for this bulletin are as follows: San
More informationAre there any policies regarding outlets that contain USB ports?
Answers to the Questions Mark J. Blackburn, P.E., CFPS February, 2016 Question 1 Are there any policies regarding outlets that contain USB ports? There is no direct reference in code about these types
More information5.07 Air Replenishment Systems
5.07 Air Replenishment Systems Reference: 2010 SFFC, Section 511.2 Definitions: CERTIFIED COMPRESSOR. A compressor used by a contractor for maintenance and testing of air replenishment systems that is
More informationS a n F r a n c i s c o F i r e D e p a r t m e n t B u r e a u o f F i r e P r e v e n t i o n & I n v e s t i g a t i o n 1 of 8 P a g e
5.07 Air Replenishment Systems (2016) Reference: SFFC, Section 511.2. Purpose: The purpose of this bulletin is to describe the requirements for air replenishment systems intended to be used to fill firefighters'
More informationNFPA Edition Review
This is a photographic template your photograph should fit precisely within this rectangle. NFPA 70 & 99 How Recent Code Changes Effect Healthcare Facilities Sam C. Terry, P.E. Application Engineer Eaton
More informationNew DATE OF BLDG. PERMIT OR PLAN APPROVAL: SURVEYOR (SIGNATURE) TITLE OFFICE DATE REVIEW AUTHORITY OFFICIAL (SIGNATURE) TITLE OFFICE DATE
DEPARTMENT OF HEALTH & HUMAN SERVICES CENTERS FOR MEDICARE & MEDICAID SERVICES FIRE SAFETY SURVEY REPORT - AMBULATORY SURGICAL CENTERS Medicare 1. (A) PROVIDER NUMBER Identifying information as shown in
More informationIowa State Fire Marshal
Iowa State Fire Marshal LIFE SAFETY CODE REQUIREMENTS LONG TERM CARE Kyle R. Gorsh, Special Agent in Charge Iowa Department of Public Safety State Fire Marshal Division 2012 LSC & Applicable Standards
More informationGASEOUS HYDROGEN SYSTEMS
CHAPTER 7 GASEOUS HYDROGEN SYSTEMS SECTION 701 GENERAL 701.1 Scope. The installation of gaseous hydrogen systems shall comply with this chapter and Chapters 30 and 35 of the Fire Compressed gases shall
More informationCOURSE SYLLABUS HVAC AND MECHANICAL CONSIDERATIONS FOR PART 3 BUILDINGS CORRESPONDENCE COURSE
COURSE SYLLABUS HVAC AND MECHANICAL CONSIDERATIONS FOR PART 3 BUILDINGS CORRESPONDENCE COURSE 100126 The course is organized into three bands. Band A covers HVACrelated building components and other types
More informationNFPA Impacts Training presentation for: AlaSHE Spring 2017 Conference David Stymiest, Senior Consultant, Smith Seckman Reid, Inc.
NFPA 99-2012 Impacts Training presentation for: AlaSHE Spring 2017 Conference David Stymiest, Senior Consultant, Smith Seckman Reid, Inc. Content Disclaimer - Content Disclaimer: These slides are only
More information2. The Group F occupancy has have an a combined occupant load of 500 or more above or below the lowest level of exit discharge.
2. The Group F occupancy has have an a combined occupant load of 500 or more above or below the lowest level of exit discharge. Exception: Manual fire alarm boxes are not required where the building is
More information2008 NYC Fire Code. New Laboratory Standards. Operational & Maintenance Provisions effective July 2008
2008 NYC Fire Code New Laboratory Standards Operational & Maintenance Provisions effective July 2008 Design & Certificate of Fitness Requirements Effective July 1, 2009 History The new Fire Code legislation
More informationBlack Diamond Fire Department Building Owner s Responsibilities for Fire Safety Equipment
Black Diamond Fire Department Building Owner s Responsibilities for Fire Safety Equipment February 2017 Alberta Fire Code safety equipment maintenance requirements: The Black Diamond Fire Department has
More informationCommittee on NFPA 99
Committee on NFPA 99 M E M O R A N D U M TO: FROM: NFPA Technical Committee on Fundamentals Jeanne Moreau-Correia DATE: March 2, 2010 SUBJECT: NFPA 99 A11 ROP Letter Ballot The ROP letter ballot for NFPA
More informationElement D Services Electrical ODG010107
Element D Services Electrical ODG010107 D5090 Other Electrical Systems Part 1. General This section includes design standards and requirements for other electrical systems. This is a design standard and
More informationIFC Significant Changes from the 2009 to the 2015 Edition Tier 1
IFC Significant Changes from the 2009 to the 2015 Edition Tier 1 Several medical care definitions were created and revised in Chapter 2 of the 2012 IBC, as well as through the occupancy descriptions, and
More informationObjectives. NFPA Dust Specific Documents. Combustible Dust Process Hazard Safety Overview of NFPA Standards for Combustible Dusts
Combustible Dust Process Hazard Safety Overview of NFPA Standards for Combustible Dusts Guy R. Colonna, PE Division Manager, NFPA Combustible Dust Symposium Detroit, MI September 20, 2011 1 Objectives
More informationOsceola County Department of Fire Rescue and Emergency Medical Services
St. Number Dir Street Name Street Type Inspected Time of Day to be Re-inspected Zip Code Occupancy Type Emergency Contact Phone (Inc. Area Code) Building Owner Building Owner Mailing Address Bldg. Owner
More informationRESOLUTION MSC.292(87) (adopted on 21 May 2010) ADOPTION OF AMENDMENTS TO THE INTERNATIONAL CODE FOR FIRE SAFETY SYSTEMS
RESOLUTION MSC.292(87) THE MARITIME SAFETY COMMITTEE, RECALLING Article 28(b) of the Convention on the International Maritime Organization concerning the functions of the Committee, NOTING resolution MSC.98(73)
More informationDanilo B. Concepcion, CBNT, CCHT-A Operations Manager St. Joseph Hospital Renal Services
Life Safety Code Danilo B. Concepcion, CBNT, CCHT-A Operations Manager St. Joseph Hospital Renal Services danilo.concepcion@stjoe.org 714-771-8944 The findings and conclusions in this presentation are
More informationAgency for Health Care Administration
Page 1 of 94 ST - K0000 - Initial Comments Title Initial Comments Statute or Rule Type Memo Tag These guidelines are meant solely to provide guidance to surveyors in the survey process. ST - K0100 - General
More informationPage 1 of 190 Public Input No. 22-NFPA 55-2013 [ Global Input ] Throughout the document, delete ANSI from ANSI/ASME B31.3. Current Text: (multiple locations) ANSI/ASME B31.3 (found in Subsections: 2.3.2
More informationThe Cannabis Industry: Fire Department Requirements
The Cannabis Industry: Fire Department Requirements RVFD Standard 4.1 Rio Vista Fire Department 350 Main St. Rio Vista, CA. 94571 By: Jeff Armstrong, Fire Chief Updated December 2018 1 The purpose of this
More informationAgency for Health Care Administration
Page 1 of 34 ST - K0000 - INITIAL COMMENTS Title INITIAL COMMENTS Statute or Rule Type Memo Tag These guidelines are meant solely to provide guidance to surveyors in the survey process. ST - K0001 - OTHER
More informationStatement of Conditions
Alabama Society for Healthcare Engineering Environment of Care, Life Safety, and Emergency Management Update May 2017 Kathy Tolomeo, CHEM, CHSP, Certified Yellow Belt Engineering Department The Joint Commission
More informationCHAPTER LOCAL AMENDMENTS TO THE INTERNATIONAL MECHANICAL CODE 2000 EDITION
CHAPTER 23.20 LOCAL AMENDMENTS TO THE INTERNATIONAL MECHANICAL CODE 2000 EDITION SECTIONS 23.20.100 Local Amendments To The International Mechanical Code, 2000 Edition 23.20.202 Definitions 23.20.301.14
More informationCMS Announcement 3/27/2017. Adoption of 2012 Life Safety Code Health Care Facilities Code. Kenneth Daily, LNHA
Adoption of 2012 Life Safety Code Health Care Facilities Code Kenneth Daily, LNHA kenn@qissurvey.com CMS Announcement May 2016 Life Safety Code Health Care Code Effective July 5 th Numerous changes many
More informationNFPA 80 Overview January 21, 2016
NFPA 80 Overview January 21, 2016 Kristin Bigda, P.E. Senior Fire Protection Engineer National Fire Protection Association Presenter Kristin Bigda, P.E. Senior Fire Protection Engineer, NFPA Building Fire
More informationCalifornia Fire Code (CFC) Part III contains requirements. Building and Equipment Design Features PART. Chapters 5 through 19
PART 3 Building and Equipment Design Features Chapters 5 through 19 Chapter 5 Fire Service Features No changes addressed Chapter 6 Building Services and Systems Chapter 7 Fire-Resistance-Rated Construction
More informationAgency for Health Care Administration
Page 1 of 41 FED - K0000 - INITIAL COMMENTS Title INITIAL COMMENTS CFR Type Memo Tag FED - K0011 - LIFE SAFETY CODE STANDARD If the building has a common wall with a nonconforming building, the common
More informationRecommended Amendments to the 2015 International Existing Building Code North Central Texas Council of Governments Region
Recommended Amendments to the 2015 International Existing Building Code North Central Texas Council of Governments Region The following sections, paragraphs, and sentences of the 2015 International Existing
More informationBy Authority Of THE UNITED STATES OF AMERICA Legally Binding Document
By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document By the Authority Vested By Part 5 of the United States Code 552(a) and Part 1 of the Code of Regulations 51 the attached document has
More informationNew Facilities. Minimum Requirements
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 Item Reference Violation General Requirements NFPA 1124-7.1.1 Retail sales of consumer fireworks in both new and existing buildings, structures, and
More informationAgency for Health Care Administration
Page 1 of 75 FED - K0000 - INITIAL COMMENTS Title INITIAL COMMENTS Type Memo Tag FED - K0100 - General Requirements - Other Title General Requirements - Other General Requirements - Other List in the REMARKS
More informationElement Z General Design Requirements Existing Facilities Information
PART 1 - INTRODUCTION 1.01 OVERVIEW A. This Section provides general information for building systems and components. B. Refer to the Owner s Design Guideline Elements A though G and Element Z for technical
More informationINTERNATIONAL STANDARD
INTERNATIONAL STANDARD IEC 60079-29-2 Edition 2.0 2015-03 Explosive atmospheres Part 29-2: Gas detectors Selection, installation, use and maintenance of detectors for flammable gases and oxygen INTERNATIONAL
More informationFirst Revision No. 20-NFPA [ Section No ]
of 25 7/15/2015 8:30 AM First Revision No. 20-NFPA 56-2015 [ Section No. 1.1.2 ] 1.1.2 Nonapplication of Standard. This standard shall not apply to the following items: (1)* Piping systems covered by NFPA
More informationMedical Gas Training ASSE 6010 Installer & ASSE 6020 Inspector Courses Available
Medical Gas Training ASSE 6010 Installer & ASSE 6020 Inspector Courses Available Certified Medical Gas Services and Green Mountain Medical Gas Services are offering the required 32-hour, NITC accredited
More informationFIRE SAFETY DURING CONSTRUCTION, ALTERATION AND DEMOLITION
CHAPTER 14 FIRE SAFETY DURING CONSTRUCTION, ALTERATION AND DEMOLITION SECTION FC 1401 GENERAL 1401.1 Scope. This chapter shall govern fire safety measures during the construction, alteration, or demolition
More informationPart I Extraction Process Equipment
745 West Colfax Avenue Denver, CO 80204 p: 720.913.3474 f: 720.913.3596 www.denvergov.org/fire MARIJUANA EXTRACTION GUIDELINE FOR COMMERCIAL / LICENSED FACILITIES The information contained within this
More informationBritish Columbia Building Code 2006 Division B Part 3 Fire Protection, Occupant Safety and Accessibility Section 3.2 Building Fire Safety
Div B 3.2.1.1.(4), (5) & (6) Exceptions in Determining Building Height BCBC1998-3.2.1.1.(4), (5) & (7) replace (harmonize with NBC) - NBC 2005 has significant revisions to provisions for mezzanines. These
More informationAlex Masterton, PE Electrical Engineer Reviewer AHCA Neal Boothe, PE Electrical Engineer/Principal
Alex Masterton, PE Electrical Engineer Reviewer AHCA Neal Boothe, PE Electrical Engineer/Principal FBC 419.3.15.6 All low-voltage system main or branch circuits entering or exiting the structure shall
More informationORDINANCE NO
ORDINANCE NO. 2017-08-031 AN ORDINANCE OF THE CITY OF LANCASTER, TEXAS, REPEALING THE NATIONAL ELECTRICAL CODE 2014 EDITION IN ITS ENTIRETY AND REPLACING IT WITH THE ADOPTION OF THE NATIONAL ELECTRICAL
More informationTaming the Health Care Building Codes NFPA vs. The International and Chicago Building Codes JENSEN & HALSTEAD LTD.
Taming the Health Care Building Codes NFPA vs. The International and Chicago Building Codes Comparing the Codes: NFPA 101 Life Safety Code (2000) (NFPA) International Building Code (2012) (IBC) Chicago
More informationEl Reno Fire Department Fire Alarm Plan Review Worksheet
El Reno Fire Department Fire Alarm Plan Review Worksheet This Fire Alarm Plan Review Checklist is provided as a guide to assist with your Fire Alarm Plan Review Submittal requirements. Please contact us
More information