DensityPRO+ Gamma Density System with Remote Transmitter Installation Guide P/N Part of Thermo Fisher Scientific.

Size: px
Start display at page:

Download "DensityPRO+ Gamma Density System with Remote Transmitter Installation Guide P/N Part of Thermo Fisher Scientific."

Transcription

1 DensityPRO+ Gamma Density System with Remote Transmitter Installation Guide P/N Revision D Part of Thermo Fisher Scientific

2

3 DensityPRO+ Gamma Density System with Remote Transmitter Installation Guide P/N Revision D

4

5 2001 Thermo Fisher Scientific Inc. All rights reserved. Microsoft and Windows are either trademarks or registered trademarks of Microsoft Corporation in the United States and/or other countries. HART is a registered trademark of the HART Communication Foundation. Fisher-Rosemount is either a trademark or registered trademark of Emerson Electric Company. All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. Thermo Fisher Scientific Inc. (Thermo Fisher) makes every effort to ensure the accuracy and completeness of this manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result of errors or omissions. Thermo Fisher reserves the right to make changes to the manual or improvements to the product at any time without notice. The material in the manual is proprietary and cannot be reproduced in any form without the expressed written consent from Thermo Fisher.

6 This page intentionally left blank.

7 Revision History Revision Level Date Comments Rev Initial release. Rev. A Name change. Rev. B Per ECO Rev. C Per ECO Rev. D Per ECO 5930.

8 This page intentionally left blank.

9 DensityPRO+ Table of Contents Table of Contents Chapter 1 Introduction Description Source Detector Transmitter Features How to Use This Manual Installation Instructions Additional Information Setting Up and Using the Gauge Technical Support Chapter 2 Hardware Installation Licensing FM/CSA Hazardous Location Approvals Installation Guidelines Power Requirements Transmitter-Detector Separation Mounting the Transmitter Dimensional Drawings Transmitter Mounting Instructions Mounting the Gauge Head (source and detector) Mounting Configurations Guidelines for Mounting the Gauge Head Gauge Head Mounting Instructions Single Chain (unichain) Mounting Instructions Dual Chain Mounting Instructions Pipe Saddle (cradle) Mounting Instructions Pipe Spool Mounting Instructions Z-Pipe (axial) Mounting Instructions Chapter 3 Wiring Procedures Wiring Instructions Wiring Drawings Detector Wiring Scintillation Detector Ion Chamber Detector Wiring Transmitter Wiring Transmitter Board Configurations AC Power Wiring DC Power Wiring Transmitter Board-Level Wiring Detector-Transmitter Wiring Flow Input Current Output Relays Contact Closure (Switch) Inputs Serial Communications HART Communications Thermo Fisher Scientific i

10 Table of Contents DensityPRO+ Chapter 4 Troubleshooting and Maintenance Source Housing Transmitter No Display Adding or Replacing Transmitter Boards Replacing the Transmitter AC Power Supply Jumper Settings Current Board Problems Relay Problems Scintillation Detector AC Power Supply Fuse Factory Wiring Jumper Settings Scintillation Detector Troubleshooting Ionization Detector AC Power Supply Fuse Ion Chamber Detector Troubleshooting Appendix A How to Return Equipment for Service...A-1 Appendix B Parts List...B-1 Appendix C Specifications...C-1 Operating Principle...C-1 Model 1400A Transmitter...C-1 Scintillator Detector (Model 9719B/9720B)...C-2 ION Chamber Detector (Model 9701/9702)...C-2 Appendix D Drawings...D-1 ii Thermo Fisher Scientific

11 DensityPRO+ Chapter 1 Introduction Chapter 1 Introduction Description The Thermo Scientific DensityPRO+ gamma density system is designed to provide reliable, accurate process material density measurements for a wide variety of challenging applications. The gauge is mounted outside of the process vessel and never contacts the process material. The gauge can measure the density of almost any liquid, slurry, emulsion, or solution. The gauge can convert the basic density measurement into a variety of output measurements as appropriate for specific applications, e.g., bulk density or solids content per unit volume. Given a temperature input, the gauge can compensate the density measurement relative to a user-specified reference temperature. If a flow input is provided, the gauge can calculate mass flow. The setup menus guide you through the configuration process to help you quickly complete the gauge setup. The gauge has three main parts: a source, a detector, and a transmitter. The source and detector are mounted on opposite sides of the pipe or vessel containing the process material. The transmitter contains the microprocessor (CPU) board and the input/output (I/O) boards and can be mounted some distance away from the detector in a location more convenient for the user. The radioisotope source emits gamma radiation that passes through the process material. The detector measures the energy of the radiation arriving at the detector after passing through the process material (and vessel walls). The gauge determines the density of the process material by measuring the attenuation (energy reduction) of the gamma ray beam as it passes through the process material. Source Detector Transmitter Source A Cesium (Cs-137) radioisotope source is used for most applications. A Cobalt (Co-60) source is available for applications requiring a higher energy source. The radioisotope is bound in ceramic pellets and double-encapsulated in a pair of sealed stainless-steel containers. The resulting source capsule is highly resistant to vibration and mechanical shock. The source capsule is further enclosed in the source head, a lead-filled, welded steel housing. A shaped opening in the lead shielding directs the gamma radiation beam through the process material towards the detector. Outside of the beam path, the energy escaping the source head is very low and well within prescribed limits. Closing the source shutter allows the beam to be turned off (the shutter blocks the radiation) during Thermo Fisher Scientific 1-1

12 Chapter 1 Introduction DensityPRO+ installation or servicing of the gauge. All source housings meet or exceed the safety requirements of the U.S. Nuclear Regulatory Commission (NRC) and Agreement State regulations, see the Gamma Radiation Safety Manual (717904). Detector The gauge uses either an ionization chamber or scintillator type detector to measure the gamma radiation reaching the detector from the source. An ionization chamber is a sealed, temperature-stabilized chamber filled with a relatively dense gas. Radiation entering the chamber ionizes the gas, allowing current to flow between electrodes in the chamber. The detector amplifies this signal and sends it to the transmitter. The ionization chamber detector can withstand vibration and other harsh conditions. The scintillator detector consists of a special plastic scintillator material, a photomultiplier tube, and associated electronics. When radiation strikes the plastic scintillator material, small flashes of light are emitted. As the density of the process material increases, the process material absorbs more of the gamma radiation and the scintillator material generates fewer light pulses. A photomultiplier tube and the associated detector electronics convert the light pulses into electrical pulses that are sent to transmitter for processing to determine the process material density and related measurement values. The plastic scintillator used in the gauge is very efficient at converting the incident gamma radiation into light. The scintillator detector is suitable when more precision or greater sensitivity are required or when a lower-activity source must be used. Transmitter The transmitter uses the signal from the detector to calculate the process material density and related measurement values. These values can be displayed, sent to serial ports, or used to drive current outputs and alarms. The transmitter also monitors system performance and generates system fault and warning alarms. The transmitter is available in either a non-metallic (NEMA 4X) enclosure or a heavy cast aluminum (NEMA 7/NEMA 4) explosion proof enclosure. The NEMA 4X is certified (FM/CSA) for use in Division 2 hazardous locations while the NEMA 7 enclosure is certified (FM/CSA) for use in Division 1 hazardous locations. Note: Refer to the equipment tag on the gauge to verify the hazardous location certifications for your gauge. Measurement Calculation After the transmitter calculates the process material density, it can convert the data into a number of forms. For a slurry consisting of carrier and solid components, the gauge can provide measurements based on the ratio of solids to carrier. Similar measurements can be made for emulsions that consist of two different fluids and for solutions that consist of a material (the solute) dissolved in a fluid (the solvent). 1-2 Thermo Fisher Scientific

13 DensityPRO+ Chapter 1 Introduction If flow data is provided as an input, the gauge can generate mass flow measurements. The gauge can accept a 4-20 ma current input from a magnetic flow sensor or from Thermo Fisher s VersaFLO fixed or portable flow meters. For applications that require temperature compensation, the gauge can accept a temperature input to compensate the density measurement for changes in process temperature. Communications and Measurement Display Inputs and Outputs Communications with the gauge are accomplished via the integral keypad (NEMA 4X transmitter only) or via the RS-232 or RS-485 serial ports. The integral keypad on the NEMA 4X transmitter is normally used as the primary means of communication with the gauge. Menu selections, commands, and parameter values are entered using the keypad. The RS-232 or RS-485 serial ports can be used to communicate with the gauge using a remote terminal, a PC with terminal emulation software, or a Thermo Scientific HHT. Each of these options provides the functional equivalent of the NEMA 4X transmitter's integral keypad and display. Both the NEMA 4X and NEMA 7 transmitters include a four-line by 20-character display. The four-line display (20 characters per line) shows either one menu item or up to eight readouts in alternation (four at a time). The HART communication protocol is supported over the 4-20 ma current output with an optional daughter board. You communicate with the gauge using the standard Fisher Rosemount 275 HART handheld communicator. Refer to the DensityPRO+ HART Operation Manual (P/N ) for detailed instructions. The characteristics of the input and output options for the gauge are summarized in Table 1-1. The DensityPRO+ gauge provides RS-485 and RS-232 serial communication ports, a 4-20 ma current output and two contact closure inputs. The contact closure inputs can be programmed to activate any system command based on a user-provided input. System options include up to six relays, additional 4-20 ma current outputs and contact closure inputs. Thermo Fisher Scientific 1-3

14 Chapter 1 Introduction DensityPRO+ Table 1-1 Input and Output Characteristics Input/Output Characteristics Type Characteristics Comments Input power DC Input Power 12 Vdc option is for safe areas only (no FM/CSA approvals) 24 Vdc (20-28 V), 12 W standard 12 Vdc (9-15 V), 15 W optional Current output Serial communications HART communications Relays Contact closure inputs AC Input Power 110/220 Vac ( V), 50/60 Hz, 25 VA optional ma DC (adjustable range) Can be configured as: 1) Isolated, Self-Powered, or 2) Isolated, Loop-Powered, user must supply the 24 Vdc loop power input. Max Load: 700 ohms RS-232: One terminal block RS-485: One terminal block and one RJ11 (phone) Jack HART protocol supported over the 4-20 ma current output. Two relays (optional) on each I/O Board. Form C SPDT, Isolated, 8 A, 220 Vac. Two contact closure inputs provided on each I/O board. Default range is 4-20 ma DC. One current output is standard; up to two additional current outputs can be provided. Self-Powered configuration is standard. Reconfigure as Loop-Powered by removing a jumper. Full duplex communication with remote terminal or PC. Half-duplex party-line communication to host computer or HHT. Optional daughter board required. Up to six relays maximum, two per I/O board. Assign process alarms to control (open/close) relays. Execute commands based on userprovided contact opening or closure input to the gauge. Flow input 4-20 ma current input Current output from a flow sensor can be input to the gauge. The gauge uses this input to compute mass-flow readouts. Temp. comp. Input Display External display Temp Comp Board ma current input User Temp Sensor ma current input Four lines, backlit, 20 characters per line (NEMA 4X Transmitter only) Optional backlit display of measurement readouts. two lines, 16 characters per line Current output from a temperature sensor can be input to the gauge. This input is used to compensate density measurement for temperature effects. English-language Setup Menus. Up to eight measurement readouts displayed four at a time in alternation. See notes on page 4 2. Up to four measurement readouts can be displayed two at a time in alternation. 1-4 Thermo Fisher Scientific

15 DensityPRO+ Chapter 1 Introduction Features Dynamic Menu System Instantaneous Response Multiple Readouts Extensive Alarms The setup menus guide you through the configuration of the gauge. The Set up density, den. alarms, and flow menu group includes all of the basic parameters and commands required to quickly configure your gauge. Additional menu groups provide specialized parameters and commands allowing the gauge to be tailored to a wide variety of applications. Direct-entry menu shortcuts are also provided, allowing experienced users to access menu items and commands directly, bypassing the menu system. Our Dynamic Process Tracking (DPT) ensures that there is no lag time in the system response to significant changes in process density. When changes occur, the DPT feature reduces the normal averaging time constant by a factor of eight, ensuring a rapid, yet smooth output response. When the process stabilizes, a longer time constant is applied to reduce the fluctuations inherent in radiation-based measurements. In this way, process density changes are immediately reflected in the transmitter output, while the effects of statistical variations in the radiation measurement are greatly reduced. Select up to eight measurement values for display. Available measurement types include density, bulk density, solids concentration, carrier concentration, ratio of solids to carrier, bulk mass flow, bulk volume flow, and the rate of change of any these measurements. You can set up as many as 16 process alarms in addition to system fault alarms and warning alarms. Totalizers and Batch Control Output Signals You can set up four independent totalizers to count time, flow output, or relay signals at any rate. If your gauge has relays installed, each totalizer can close a relay (for an external counter, etc.) at user-specified count intervals. The totalizers can also close relays when particular counts are reached (for batch or sample control). You can assign any measurement to the 4-20 ma current output, or the measurement values can be sent to a remote terminal or host computer as serial data. One current output is provided on the standard I/O board. Two additional I/O boards can be provided as options with one 4-20 ma current output per board. Two relay outputs are optional on each of the I/O boards. Each I/O board also includes two contact closure inputs that can be used to activate any system command based on a user-provided input (open or close). Thermo Fisher Scientific 1-5

16 Chapter 1 Introduction DensityPRO+ How to Use This Manual Installation Instructions Additional Information This guide provides guidelines for planning the installation of your gauge, and instructions for the hardware installation and wiring of the gauge. Chapter 2 Hardware Installation provides guidance on where and how to mount the gauge head and transmitter. Chapter 3 Wiring Procedures explains wiring for the detector and transmitter, including the optional relays, current outputs, contact closure inputs, and serial communications. Chapter 4 Troubleshooting and Maintenance describes maintenance and troubleshooting procedures for gauge. Appendix A How to Return Equipment for Service provides information on returns. Appendix B Parts List contains the reference list needed to order parts. Appendix C Specifications contains the system specifications for the gauge. Appendix D Drawings lists the installation drawings and provides reduced-size copies of drawings. Setting Up and Using the Gauge When installation is complete, refer to the user guide (717819) for detailed instructions for configuring and operating your gauge. Refer to the DensityPRO+ HART Operation Manual (717816) for detailed instructions for the set up and operation of the gauge via the HART protocol. Refer to the Model 9733/9734 Hand Held Terminal Operation Manual (717797) for details on communicating with the gauge using the Thermo Scientific HHT. 1-6 Thermo Fisher Scientific

17 DensityPRO+ Chapter 1 Introduction Technical Support Thermo Fisher is ready to assist you with any installation or setup problems you may have with your gauge. In the United States: Thermo Fisher Scientific 1410 Gillingham Lane Sugar Land, TX Phone: Fax: In Canada: Thermo Fisher Scientific 14 Gormley Industrial Avenue Gormley, Ontario L0H 1G0 Phone: Fax: On the Web: Thermo Fisher Scientific 1-7

18 This page intentionally left blank.

19 DensityPRO+ Chapter 2 Hardware Installation Chapter 2 Hardware Installation This section provides guidance for mounting the gauge. The gauge consists of three components: 1. Model 1400A Transmitter Non-metallic enclosure (NEMA 4X), or Cast aluminum enclosure (NEMA 7/NEMA 4) 2. Detector Model 9701 or 9702 Ionization Chamber Detector, or Model 9719B or 9720B Scintillation Detector 3. Source Head Standard source head model numbers include: 5190, 5191, 5176, 5200, 5201, 5202, 5203, 5204, and The combination of the detector and the source head is referred to as the gauge head. Note: Reduced-size copies of the drawings referenced in this chapter are provided in Appendix D for your convenience. Licensing Warning: The gauge is a nuclear device regulated by federal and/or state authorities. You are responsible for knowing and following the pertinent safety and regulatory requirements. Refer to the Gamma Radiation Safety Manual (717904) for a summary of these requirements. Moving or removing an installed source housing, or any assembly that includes a source housing, requires a person who is specifically licensed to install and commission Thermo Scientific source heads. Your general license permits you to own and install all of the gauge s components, including the source head. However, you may not commission the gauge (remove the shipping bolt and open the source shutter for the first time) without a specific license authorizing radiation commissioning of the gauge. For assistance obtaining a specific license, commissioning, or decommissioning the gauge, contact Thermo Fisher (see page 1-7). FM/CSA Hazardous Location Approvals The FM/CSA hazardous area approvals are listed in the table below. Warning: Do not locate the unit in any hazardous area other than those approved. Refer to the equipment tag for the specific approvals applicable to the configuration of your gauge prior to installing the gauge. Thermo Fisher Scientific 2-1

20 Chapter 2 Hardware Installation DensityPRO+ Model 1400A Transmitter Scintillator Detector Non-XP (Plastic) Enclosure Class I, Div 2, Grp A,B,C,D Class II, Div 2, Grp E,F,G Class III NEMA 4X, Type 4X Model 9719B Class II, Div 1, Grp E, F, G Class III, Div. 1 NEMA 4, Type 4 XP Enclosure Class I, Div 1, Grp C,D Class II, Div 1, Grp E,F,G Class III NEMA 4/7, Type 4/7 Model 9720B Class I, Div 1, Grp B, C, D Class II, Div 1, Grp E, F, G Class III NEMA 4, Type 4 Model 9701 Model 9702 Ion Chamber Detector NEMA 4, Type 4 Class I, Div 1, Grp B, C, D Class II, Div 1, Grp E, F, G Class III NEMA 4, Type 4 Installation Guidelines Warning: Do not apply power to the unit in any hazardous area unless the safety ground is properly wired inside the unit and the cover is properly installed. Power Requirements Refer to the Installation Layout Drawing (868695) for information on installation requirements in hazardous areas. Model 1400A Transmitter AC Power Option: 100 to 240 Vac, 50/60 Hz, 25 VA. 24 V DC Power Standard: Vdc, 12 W 12 V DC Power Option: 9-15 Vdc, 15W, requires Vdc converter option Model 9719B/9720B Scintillation Detector 24 Vdc ±20%, 12VA, at detector input, or 12 Vdc ±30%, 12VA, (w/12-to24 Vdc converter option), or 115/230 Vac (±15%), 50/60 Hz, 12 VA (w/ac power option). Model 9701/9702 Ion Chamber Detector 115/230 Vac (±10%), 50/60 Hz, 120 VA. Transmitter-Detector Separation The maximum cable length between the transmitter and the detector varies with the gauge of the wire, but should not exceed 5000 feet (1500 m) for the ion chamber detector or 1000 feet (300 m) for the scintillation detector. 2-2 Thermo Fisher Scientific

21 DensityPRO+ Chapter 2 Hardware Installation Mounting the Transmitter Selecting a good location to mount the transmitter can ensure the accuracy and ease of operation of your gauge. The operating temperature range is 40º C to 60º C ( 40º F to 140º F). Mount the transmitter so that the display can be easily seen and the keypad (on the NEMA 4 transmitter) is in easy reach. Do not mount the transmitter near high voltage, high current, corrosive vapors, corrosive liquids, or corrosive esters. Dimensional Drawings Refer to the proper dimensional drawing for the transmitter mounting dimensions. See Drawing for the NEMA 4X non-metallic enclosure suitable for Div. 2 hazardous locations and ordinary locations. See Drawing for the NEMA 7/NEMA 4 explosion proof enclosure suitable for Div. 1 hazardous locations. Transmitter Mounting Instructions Mount the transmitter with both four screws, or four nuts, four bolts, and eight washers. Place one screw through each of the four holes in the transmitter enclosure s flange and tighten the screws, or Place one bolt through each of the four holes in the transmitter enclosure s flange, with one washer on each side, screw on the nuts, and tighten the bolts. Mounting the Gauge Head (source and detector) Mounting Configurations The optimum gauge mounting configuration depends on the application; however, the source head and the detector are typically mounted together on the opposite sides of a pipe. The following mounting options are supported for the gauge: Chain mount Single chain (unichain) Dual chain Cradle-type (saddle) clamp Pipe spool with gauge head pre-installed Z-pipe (axial) mount for small diameter pipes The chain mount is the most frequently used mounting configuration this allows the unit to be mounted on a range of pipe sizes using the same mounting hardware. Note: Moving the gauge to a pipe with a different diameter may require changing the radiation source size and/or reconfiguring the detector (ion chamber) electronics. Call Thermo Fisher for assistance. Thermo Fisher Scientific 2-3

22 Chapter 2 Hardware Installation DensityPRO+ Guidelines for Mounting the Gauge Head The operating temperature range is 30º C to 60º C ( 20º F to 140º F) for both the scintillator and ion chamber detector types. Make sure there is enough clearance to install and service the gauge head. (Refer to the appropriate drawings for your type of gauge and mount.) Mount the source housing so the radioactive source identification tag is visible and upright. Avoid locations where process overflow, debris, or other material can collect in the beam path (especially in the source shutter mechanism). Whichever mounting configuration is used, it is important to align the beam path (the centerline of the source housing) as closely as possible with the centerline of the detector housing. Be sure to mount the gauge head securely, since any movement or change in alignment can affect the gauge s calibration. For best performance the beam must pass through a representative cross-section of the process material being measured. It is best to mount the gauge head on a vertical section of pipe. This keeps suspended solids from settling out of the measured area. (The gauge measures only the material that passes through the beam.) To mount the gauge head on a horizontal pipe Position the beam path at a 30 to 45 degree angle from vertical. This position tends to average density variations caused by settling, while reducing the effect of any trapped gases or solids that accumulate in the top or bottom of the pipe. If the process material is a solution, a light slurry, or a single-phase liquid that will not separate, the gauge head can be mounted such that the beam is horizontal. If the process material is a slurry, position the gauge head as far as possible from any elbows, tees, or valves these tend to separate suspended solids. Position the beam path in the plane of the upstream elbow so the measurement includes any uneven distribution caused by the fitting. X - Improper Gauge Head Mounting - Correct Gauge Head Mounting X X X Figure 2.1 Gauge Head Installation Examples 2-4 Thermo Fisher Scientific

23 DensityPRO+ Chapter 2 Hardware Installation Gauge Head Mounting Instructions The following sections provide installation details for the various mounting configurations. Caution: Use proper lifting procedures during installation to avoid injury. Do not over-tighten the bolts. Refer to the following drawings for detector dimensions. Drawing Model 9719B scintillation detector (NEMA 4 housing) Drawing Model 9720B scintillation detector (explosion proof housing) Drawing Model 9701 ion chamber detector (NEMA 4 housing) Drawing Model 9701 ion chamber detector (explosion proof housing) Single Chain (unichain) Mounting Instructions The single chain, or unichain mount is only available with the Model 5211 source housing and the scintillation detector. In this mounting configuration, a mounting chain is inserted through an eye bolt in the rear of the source housing. The two ends of the chain are wrapped around the pipe and engaged in slotted tabs in the detector housing. The following steps describe the installation of the gauge head using a single chain mount. Also refer to the following drawings for mounting details. Drawing Model 9719B detector (NEMA 4 housing), or Drawing Model 9720B detector (explosion proof housing). The mounting hardware kit shipped with the gauge should include: Note: 1 Eyebolt ( x 3 in.), 1 Nut ( ), 1 Belleville Washer 1 Chain 1. Insert the eye bolt through hole in rear of source housing. Place Belleville washer, cupped-side first, onto the eyebolt. Thread the nut on the eyebolt until the bolt just begins to come through the nut. 2. Insert the chain through the eyebolt so that approximately the same length of chain extends on each side of the eyebolt. 3. Position the source housing and detector housing so that the V-shaped mounting features center the housings on the pipe. 4. With the axis of the source housing and detector housings aligned, attach one end of the chain in the slotted tab on one side of the detector housing. Pull the chain tight on the other side and attach it into the slotted tab on the other side of the detector housing. There must be an even number of chain links between the slotted tabs so that there will be an equal number of links on both sides. 5. Maintain alignment of the housings and tighten the nut on the eyebolt ~1/2 turn past point where Belleville washer becomes fully flattened. Thermo Fisher Scientific 2-5

24 Chapter 2 Hardware Installation DensityPRO+ Dual Chain Mounting Instructions In the dual chain mount, two mounting chains are wrapped around the pipe with their ends engaging keyhole-shaped openings in the source head mounting plate. The detector housing is then clamped to the two chains center links (opposite the source mounting plate) by tensioning bolt assemblies as illustrated in the photograph below. Details of the chain mount depend on the detector type (ionization chamber or scintillator), but the installation is similar for both types. Refer to the following drawings for mounting details. Drawing Model 9719B Scintillation Detector Drawing Model 9720B Scintillation Detector Drawing Model 9701/9702 Ion Chamber Detector The following steps describe the installation of the gauge head using a dual chain mount. 1. For either detector type, use the table on page 2-8 to determine the number of chain links to thread through the keyhole slots. Count the links and mark the links that will be engaged at each end of the two chains (four links in all). 2. For Model 9701/9702 ionization chamber detectors, partially assemble the two tensioning bolts (hardware kit ) in the following order: 1 tensioning bolt (large) 1 flat washer (onto bolt next to bolt head) 1 spacer ring 2 Belleville washer (with concave sides together) 1 special nut - thread nut securely onto the bolt end 3. For Model 9719B/9720B scintillation detectors, partially assemble the two tensioning bolts (hardware kit ) in the following order: 1 tensioning bolt (large) 1 chain - place center ring of chain over the bolt head 1 bar nut - position the guard tube towards the bolt head 2-6 Thermo Fisher Scientific

25 DensityPRO+ Chapter 2 Hardware Installation Note: 1 special nut - thread nut securely onto the bolt end. After the special nut is fully threaded, thread the bar nut back down the bolt so that it rests against the special nut 4. From the pipe side of the source mounting plate, thread one end of a chain through the plate s top keyhole opening until you reach one of the four end links you marked in Step 1. Engage this link by sliding it into one of the keyhole slots. 5. Thread one end of the other chain through the bottom keyhole and engage the marked link as you did with the top chain. 6. Position the source housing mounting plate on the pipe. If necessary, use a lift, hoist, or other means to hold it in position. 7. Pass the center part of the upper and lower chains around the pipe and thread their ends through the corresponding keyholes. Engage the marked links in the empty slots. All four chain ends must engage at the same link. If you make an adjustment, adjust all four chain ends by the same amount. Also make sure there are no twists in the chain as you wrap it around the pipe. Otherwise the gauge head will not be aligned correctly. You might have to adjust chain links (thread more or fewer links through each slot) to allow for pipe size tolerance, insulation, and so forth. Start with about 10 cm (4 inches) of play. 8. Position the detector on the pipe, opposite the source mounting plate. If necessary, use a lift, hoist, or other means to hold it in position. 9. To secure the detector and the source head mounting plate on the pipe: Position the upper chain (including the tensioning bolt assembled in Step 1) over the detector s top mounting arm. Place a spacer ring and then two Belleville washers (concave sides together) onto the pilot end of the special nut. Insert the pilot end of the chain tensioning bolt into the hole in the top detector mounting arm and tighten the chain tensioning bolt finger-tight. Repeat the previous three steps for the lower chain. 10. Alternately and uniformly tighten the chain tensioning bolts until the spacers are just touching the mounting arms. 11. Place the loose ends of the upper chain over the top edge of the source mounting plate so they won t interfere with the source housing installation. 12. Position the source housing so its four mounting holes engage the four studs on the mounting plate. If necessary, use a lift, hoist, or other means to hold it in position. 13. Use the provided lock washers and nuts to secure the source housing. Thermo Fisher Scientific 2-7

26 Chapter 2 Hardware Installation DensityPRO+ Pipe Size (inches) Table 2-1 Number of Chain Links to Engage for Dual Chain Mount Chain Links vs. Pipe Size Use Chain Part No Use Chain Part No Models 9719B & 9720B Links From Center Links From End Models 9719B & 9720B Links From Center Links From End Pipe Size (inches) Models 9701 & 9702 Links From Center Links From End Models 9701 & 9702 Links From Center Links From End Pipe Saddle (cradle) Mounting Instructions A pipe saddle mount is a pair of identical mounting plates that you bolt together on opposite sides of the process pipe. 2-8 Thermo Fisher Scientific

27 DensityPRO+ Chapter 2 Hardware Installation Refer to drawing 85726N and use the following steps to install a gauge head with a pipe saddle mount: 1. Assemble the two halves of the saddle mount onto the process pipe using the supplied nuts, bolts, and washers. Do not tighten the nuts and bolts yet, just thread them together. 2. Adjust the saddle mount halves so the mounting plates are parallel and even with each other. Then tighten the bolts evenly so the clamp ends are the same distance apart on both sides of the pipe. 3. Bolt the detector and source housings to the mounting plates on either side of the pipe saddle. 4. Tighten all bolts securely so the gauge head components cannot shift positions. Pipe Spool Mounting Instructions A pipe spool is normally a 30-inch section of pipe with either mounting plates or a complete gauge head already installed. Often, the customer sends a length of the required pipe to Thermo Fisher to be converted into a pipe spool mount. Warning: Use correct pipe fitting techniques suitable for the pipe being used and the process material that the pipe will handle. Use the following steps to install a gauge head with a pipe spool mount: 1. Assemble the pipe spool into the existing pipe. 2. Position the source housing so its identification tag is upright. 3. If the detector and source housings are not already installed, bolt them to the mounting plates on either side of the pipe saddle. 4. Tighten all bolts securely. Z-Pipe (axial) Mounting Instructions A Z-pipe mount is used for small pipes, four inches in diameter or less. It uses a Z-shaped pipe section so the gamma rays can travel along the pipe s axis for several inches. This lets the beam pass through more process material so the material s density can be measured more reliably. Depending on the application, the user typically fabricates the Z-pipe section and then either assembles it into the mounting fixture or sends it to Thermo Fisher for assembly. Fabricating and Assembling the Z-Pipe To fabricate the Z-pipe, refer to the drawings for your specific application. Drawing Z Pipe Fabrication Details (1-3 in. pipe sizes) Drawing Z Pipe Fabrication Details (4 in. pipe) Carefully note the tolerances for each dimension. Any variation in the Z-pipe dimensions, especially in the bracket placement, can lead to misalignment causing inaccurate readings or no readings at all. Thermo Fisher Scientific 2-9

28 Chapter 2 Hardware Installation DensityPRO+ Z-Pipe Installation Refer to the following drawings for mounting dimensions. Drawing Z Pipe Mount (1-3 in. pipe sizes) Drawing Z Pipe Mount (4 in. pipe) Caution: If a lead doughnut is provided with the mounting assembly, make sure its inside diameter matches the pipe s outside diameter. While assembling the mount, clamp the doughnut to the axial portion of the Z-pipe (between the source and detector) as close to the center as possible. Warning: If the process pipe is not strong enough to support the gauge head and mounting assembly, you must provide additional support for the gauge head. After the pipe and mount are assembled, you can install the gauge head using the steps in Pipe Spool Mounting Instructions on page 2-9, with the following additional notes: If your gauge is supplied with a large lead plate with a one- to two-inch hole in the center, the plate should be mounted between the detector and the mounting plate. If your gauge is supplied with a thin steel plate with a small lead disk, the plate should be mounted between the source housing and the mounting plate, with the disk towards the pipe (the disk should fit in the hole in the mounting plate) Thermo Fisher Scientific

29 DensityPRO+ Chapter 3 Wiring Procedures Chapter 3 Wiring Procedures The section describes how to wire the DensityPRO+ detector and transmitter. The required wiring steps consist of: connecting power to the transmitter and the detector wiring the detector to the transmitter Additional wiring steps can include: relay contacts 4-20 ma current output(s) contact closure inputs wiring the optional Temperature Compensation board to the transmitter serial communications Wiring Instructions All wiring must be done in accordance with applicable codes, using approved conduit, boxes, and fittings. Danger: Remove all power from the unit before making any connections. Electrocution can result if power is present. Warning: All wiring must be done by qualified individuals in accordance with applicable codes such as the NEC (National Electric Code) ANSI/NFPA 70 specifications or the Canadian Electrical Code Part 1. Warning: Before wiring the DensityPRO+ gauge, verify that the source shutter is in the closed (OFF) position. Wiring Drawings Model 1400A Transmitter The installation layout drawing (868695) provides general guidance for routing the cables between the detector and the transmitter. Refer to the drawings listed below for instructions on wiring the detector and transmitter supplied with your gauge. Drawing Installation Wiring, Model 1400A Transmitter Scintillation Detector/Transmitter Drawing Installation Wiring, Model 9719B/9720B Scintillation Detector Ion Chamber Detector/Transmitter Drawing Installation Wiring, Model 9701/9702 Ion Chamber Detector Thermo Fisher Scientific 3-1

30 Chapter 3 Wiring Procedures DensityPRO+ Detector Wiring Warning: Do not apply power to the unit in any hazardous area unless the safety ground is properly wired inside the unit and the cover is properly installed. Warning: Hazardous Location Installations The cable entries must be sealed per the Installation Layout Drawing (868695). Non-Hazardous Location Installations The cable entries into the enclosures must be sealed to prevent passage of gas or vapors. The surrounding atmosphere or liquids should not affect the sealing compound. The minimum thickness of the sealing compound should be 5/8 in (16 mm). If metal conduit is used, the conduit must be grounded. Scintillation Detector 1. Use the following procedure for wiring the Model 9719B/9720B scintillation detector: 2. Make sure all source shutters are in the OFF position. 3. Make sure all power to the gauge is turned off. 4. Remove the housing access cover. Model 9719B Remove the bolts that secure the cover to the housing. Model 9720B Loosen the screw on the cover retaining bracket and slide the bracket off of the housing cover. Unscrew the housing access cover (two lugs are provided on the top of the cover to aid in the removal of the cover). 5. Remove the cable conduit plugs only from the hole(s) that will be used. As shown in the Installation Layout drawing (868695), lay one conduit for the signal cables (and DC power input, if applicable), and a second conduit for the AC power input. 6. Pull each cable through the correct conduit fitting and into the enclosure. Leave approximately 150 mm (six inches) for strain relief. Secure the conduit, making sure it is completely sealed. 7. Refer to the Installation Wiring drawing (868577) for detailed instructions. To connect the input power and the signal wires (the data pulse and reference pulse cable and the HV control cable): Remove the plug-in screw-terminal connector from the board connector. The connector should be tightly seated. To remove the connector, brace the board with your hand, being careful not to touch any of the circuits or components. Use the 1/8-inch screwdriver to loosen the screws on the plug-in connector. Insert the wires and make connections as shown in the installation wiring drawing. Tighten the screws in the connector to secure the wires and plug the screwterminal connector back into the corresponding on-board connector. 3-2 Thermo Fisher Scientific

31 DensityPRO+ Chapter 3 Wiring Procedures The temperature compensation board, if provided, uses a spring terminal block connector. For each wire to be connected, use a flat screwdriver to push and hold the appropriate tab down, insert the wire, then release the tab to secure the wire. 8. When the detector wiring is complete, replace and secure the detector housing cover. 9. Refer to the section, Transmitter Wiring on page 3-5 and the appropriate Transmitter Wiring drawings for details of wiring the signal and temperature compensation wires to the transmitter. Power Supply Wiring Scintillation Detector The maximum input power requirement is 6 VA. Note: To meet the requirements of CSA , an external switch or circuit breaker must be installed to allow the power source to be disconnected from the gauge. In addition, protective bonding (grounding) must always be provided, even if a DC power source is used. Scintillator Detector - DC Power! The Model 9719B/9720B scintillation detector is designed to operate on 24 Vdc (± 20%). When the optional 12-to24 Vdc converter board is installed, the detector can operate on 12 Vdc (± 25%). In either case, the input connector for the DC source voltage wiring is located on the Interface/ Adapter board. See installation wiring drawing Note: To meet the requirements of CSA , the input DC terminals shall be supplied from an SELV (Safety Extra Low Voltage) source. Scintillator Detector - AC Power Note: Connect the AC wiring Earth ground to the internal safety ground terminal as shown in the wiring diagram. With the AC Power board installed, the scintillator detector may be operated using either 115 or 230 Vac (± 15%). The AC supply voltage is set by the 115/230 Vac selector switch. The selector switch is located on the AC power board, below the connectors and just above the transformer. To access the selector switch, it is necessary lift the scintillation detector chassis several inches out of the housing. Warning: Applying 230 Vac with the selector switch in the 115 Vac position will damage the equipment. If both AC and DC input power are supplied to the detector, the detector will draw power from whichever source provides the higher DC voltage. Caution: For reliable operation and to maintain safety approval, the F1 fuse on the AC power board must only be replaced with an approved fuse. Refer to the Installation Wiring Diagram, Drawing Thermo Fisher Scientific 3-3

32 Chapter 3 Wiring Procedures DensityPRO+ Ion Chamber Detector Wiring Use the following procedure for wiring the Model 9701/9702 ion chamber detector. You will need a 1/8-inch bladed screwdriver. 1. Make sure all source shutters are in the OFF position. 2. Make sure all power to the gauge is turned off. 3. Remove the housing access cover. Model 9701 Remove the bolts that secure the cover to the housing. Model 9702 Loosen the screw on the cover retaining bracket and slide the bracket off of the housing cover. Unscrew the housing access cover (two lugs are provided on the top of the cover to aid in the removal of the cover). 4. Remove the foam insulator and the two brass-colored centering plates. 5. As shown in the Installation Layout drawing (868695), the AC power and signal wiring should not be run in the same conduit for more than three meters (ten feet). If required, use a conduit splitter and lay one conduit for the signal cables and a second conduit for the AC power wiring. 6. Pull each cable through the conduit and into the enclosure. Leave approximately 150 mm (six inches) for strain relief. Secure the conduit, making sure it is completely sealed. 7. Connect the input power, the ion chamber signal wires, and the temperature compensation board wires, if applicable, as shown in the installation wiring diagram (Drawing ). Note: Connect the AC wiring Earth ground to the internal safety ground terminal (on the inside of the detector housing) as shown in the wiring diagram. Do not connect the shield from the signal cable at the detector end. 8. Position the two brass-colored centering plates over the circuit board with the smaller plate on top of the larger plate. Be careful to not pinch any wiring. Thread in and finger tighten the four screws that secure the centering plates. 9. Insert a medium-sized flat screwdriver blade into the stepped ( jagged ) opening in the larger centering plate. Twist the screwdriver to push the two plates apart until the detector is locked in position, then tighten the four screws to secure the centering plates and the detector. 10. When the wiring is complete, replace and secure the detector housing cover. 11. Refer to the section, Transmitter Wiring on page 3-5 and the appropriate transmitter wiring drawing for details of wiring the signal and temperature compensation wires to the transmitter. Ion Chamber Detector Hi-Meg Value While wiring the ion chamber detector, locate the Hi-Meg label on the detector circuit board. A large resistance value will be written on this label (for example, 4.7 x ). Write down this Hi-Meg value along with the detector s serial number. The Hi-Meg value will be needed during the gauge setup to fine tune the gauge. 3-4 Thermo Fisher Scientific

33 DensityPRO+ Chapter 3 Wiring Procedures Ion Chamber Detector - AC Power The ion chamber detector may be operated using either 115 or 230 Vac (± 10%). The maximum input power requirement is 100 VA. Note: Connect the AC wiring Earth ground to the internal safety ground terminal provided on the inside of the detector housing as shown in the wiring diagram. Warning: Applying 230 Vac with the selector switch in the 115 Vac position will damage the equipment. The ion chamber detector is shipped wired for either 115 or 230 Vac operation. If necessary, the detector s operating voltage can be changed by following the procedure below. The two red wires should be installed on the marked ( RED1 and RED2 ) push-on tabs. Install the four heater wires (two white wires and two black wires) on the pushon tabs as follows: One white wire on the tab marked Wh One black wire on the tab marked Bk For 115 Vac operation: One white wire on the tab marked Wh/115 One black wire on the tab marked Bk/115 For 230 Vac operation: One white wire on the tab marked Wh/230 One black wire on the tab marked Bk/230 Set the 115/230 Vac selection switch (located next to the transformer) to the desired voltage. Caution: For reliable operation and to maintain safety approval, the F1 fuse on the AC power board must only be replaced with an approved fuse. Refer to the Installation Wiring Diagram, Drawing Transmitter Wiring DANGER: Remove all power from the unit before making any connections. Electrocution can result if power is present. Warning: All wiring must be done by qualified individuals in accordance with National Electric Code (NEC) ANSI/NFPA 70 specifications or the Canadian Electrical Code Part 1. If metal conduit is used, it must be grounded. Thermo Fisher Scientific 3-5

34 Chapter 3 Wiring Procedures DensityPRO+ Refer to the following installation wiring drawings when wiring the transmitter to the detector. Model 1400A Transmitter Drawing Installation Wiring Model 1400A Transmitter Scintillation Detector Drawing Installation Wiring, Model 9719B/9720B Scintillation Detector Ion Chamber Detector Drawing Installation Wiring, Model 9701/9702 Ion Chamber Detector Note: To meet the requirements of CSA , an external switch or circuit breaker must be installed to allow the power source to be disconnected from the gauge. In addition, protective bonding (grounding) must always be provided, even if a DC power source is used. Transmitter Board Configurations The Model 1400A transmitter has slots for the following boards (refer to drawing ): AC power supply - optional CPU board - required PCB 1 Standard current I/O / DC power board in one of the following configurations: 24 Vdc input power and current I/O #1, 2 contact inputs, 2 relays (optional) 12 Vdc input power and current I/O #1, 2 contact inputs, 2 relays (optional) PCB 2 Optional current I/O board, 2 contact inputs, 2 relays (optional) PCB 2 Optional current I/O board, 2 contact inputs, 2 relays (optional) PCB 4 The VPI (voltage/current pulse interface) board required. This board provides the wiring interface between the transmitter and the detector and controls the operation of the detector. AC Power Wiring With the AC power option, the system is designed to operate on 100 to 240 Vac, at an input frequency of 50/60 Hz. The same system configuration can be directly wired to either 115 or 230Vac. The maximum input power requirement is 25 VA when all options are present. Refer to Drawing (sheet 1) for wiring requirements. As shown in the wiring diagram, to connect AC power to the transmitter s main board: Connect AC power input (L and N) to the J1 terminal block, and Connect electrical safety ground to a dedicated safety ground terminal. 3-6 Thermo Fisher Scientific

35 DensityPRO+ Chapter 3 Wiring Procedures DC Power Wiring! Note: To meet the requirements of CSA , the input DC terminals shall be supplied from an SELV (Safety Extra Low Voltage) source. 24 Vdc Input Power - Standard With the standard 24 Vdc input power option, the system is designed to operate on Vdc. The maximum input power requirement is 12 W when all options are present. Refer to Drawing (sheet 2) for wiring requirements. As shown in this wiring diagram, to connect 24 Vdc power to the main board: 12 Vdc Input Power - Option Connect +24 Vdc to Pin 8 of the J5 connector on the Current I/O # 1 (PCB 1 board), and Connect the return for the +24 Vdc supply to ground, Pin 9 of the J5 connector. With the 12 Vdc power option installed, the system is designed to operate on 9-15 Vdc. The maximum input power requirement is 15 W when all options are present. Refer to Drawing (sheet 2) for wiring requirements. As shown in this wiring diagram, to connect 12 Vdc power to the main board: Connect +12 Vdc to Pin 2 of the J6 connector on the Current I/O # 1 (PCB 1 board), and Connect the return for the +12 Vdc supply to ground, Pin 1 or Pin 3, of the J6 connector. Note: The J6 connector is only present if the 12 Vdc option is installed. Transmitter Board-Level Wiring Use the procedure below for board-level wiring, to wire RS-232 or RS-485 serial ports, relay contacts, contact switch inputs, 4-20 ma current output(s), and to wire the detector signal cables to the transmitter. You will need a 1/8-inch bladed screwdriver. 1. Verify that power to the transmitter has been shut off. Then open the enclosure cover. 2. Remove the plugs only from the cable conduit holes you will use. The holes are labeled on the transmitter enclosure. For the non-metallic, NEMA 4X enclosure, refer to the dimensional drawing For the cast aluminum, explosion proof enclosure, refer to dimensional drawing Pull the cable through the fitting on the end of the enclosure. Leave approximately 150 mm (six inches) of stress relief. Secure the conduit, making sure it is completely sealed. 4. To connect a cable to a PCB board: Remove the plug-in screw-terminal connector from the board connector. The connector should be tightly seated. To remove the connector, brace the board with your hand, being careful not to touch any of the circuits or components. Use the 1/8-inch screwdriver to loosen the screws on the plug-in connector. Thermo Fisher Scientific 3-7

36 Chapter 3 Wiring Procedures DensityPRO+ Insert the wires and make connections as shown in the appropriate installation wiring drawing. Tighten the screws in the connector to secure the wires. Plug the screw-terminal connector back into the corresponding on-board connector. 5. When all connections are complete, close and secure the transmitter cover. Caution: CSA certified fittings must be used to maintain CSA rating for enclosure. Detector-Transmitter Wiring All signals from the detector are wired to the VPI board that is installed in the transmitter slot PCB 4. Refer to the appropriate installation wiring drawings for the detailed requirements. Caution: Remove power from the transmitter before changing jumper settings. Failure to do so may result in damage to the VPI board. For the scintillator detector, the reference and data pulse lines are wired to pins 15 and 16 and 17 and 18, respectively, of connector J1B of PCB 4. The HV increase and decrease pulse cables are wired to pins 19 and 20 and 22 and 23, respectively, of connector J1B of PCB 4. For the ion chamber detector, the signal cable from the detector is wired to pins 8(+) and 9(-) of connector J1A of PCB 4. The temperature compensation input is wired to connector J1A of PCB 4. The connections depend on the type of temperature compensation input. The transmitter is used in duty cycle mode unless it is wired to a D series scintillation detector. The duty cycle mode is set by connecting jumper p3 pin 1&2 on the scintillation adapter board and shorting JP3 on the VPI board Flow Input Current Output There is also provision on the VPI board for a user-provided 4-20 ma flow input (pins 6/7 of connector J1A on PCB 4). Refer to the DensityPRO+ Operation Manual (717819) for details on how to configure the gauge to use the flow input signal. One current output (I/O) board is supplied standard, installed in the PCB 1 slot. One or two additional I/O boards can be installed as options in the PCB 2 and PCB 3 slots. The current output is programmable between 0 and 20 ma. The minimum output in the self-powered configurations is ~0.4 ma. The minimum output current in the looppowered configuration is ~3 ma. The maximum load is 800 ohms. 3-8 Thermo Fisher Scientific

37 DensityPRO+ Chapter 3 Wiring Procedures The current output is normally shipped from the factory configured as an isolated, selfpowered output. The current output can be reconfigured as isolated loop-powered by removing the jumper from pins 3 & 4 on the J5 connector on the I/O board. In the looppowered configuration, the user must supply +24 Vdc loop power to pin 3 of J5. Pin 2 of J5 is Iout+. Relays Two or four relays are available as options on the I/O board in PCB 1 slot. Refer to the installation wiring drawing (and the label inside the transmitter cover) for wiring details. Relay contacts are Form C SPDT, isolated 8 A, and 220 Vac. Contact Closure (Switch) Inputs Serial Communications RS-485 Wiring Two (user-provided) contact closure inputs are included on each I/O board. The contact closure inputs are dry contact inputs between ground and Switch 1 and ground and Switch 2. The gauge can be configured via software settings to execute a command or other function upon a user-provided contact opening or closing. Refer to the DensityPRO+ Operation Manual (717819) for instructions for assigning commands to the contact closure inputs. The gauge provides both an RS-232 single-drop and an RS-485 multi-drop serial interface. Screw-terminal connectors for both ports are located on the CPU board. The non-metallic, NEMA 4X transmitter includes both a display and a keypad that are typically used as the primary means of communication with the gauge. The explosion proof cast aluminum transmitter does not have a display; therefore, the RS-232 or the RS-485 serial port must be used to communicate with the gauge. The RS-485 connector includes the +8 Vdc required to power the Thermo Scientific HHT as well as the +Data and Data connections. An RJ-11 (phone jack) connector is also provided for the RS-485 port. The HHT can be connected directly to the RJ-11 connector. Both the RS-232 and RS-485 ports provide independent access to the measurement readings and software functions. Both ports are always active and can be used to display measurements. The setup menus, however, can only be accessed by one port at a time. To connect the serial (COM) port on a PC to the RS-485 port on the gauge requires an RS-485-to-RS-232 converter (P/N ). Connect +Data and Data on the Model 1400A transmitter RS-485 connector (located on the CPU board) to the corresponding connections on the RS-485-to- RS-232 converter. Connect the RS-485-to-RS-232 converter to the PC using a standard DB-9 serial cable. Refer to the RS-485 installation wiring drawing (868519) and the Model 1400A transmitter wiring drawings (868515, ). Thermo Fisher Scientific 3-9

38 Chapter 3 Wiring Procedures DensityPRO+ Note: Party-line Communications Initial Setup! To communicate with multiple gauges via an RS-485 party-line, each unit must be assigned a unique unit identification number so it can be addressed individually. All gauges are assigned unit number 0 (zero) by default. To assign a unique unit number to each gauge, you must disconnect each gauge from the party-line in turn and then communicate with the disconnected gauge directly. Alternately, you must remove power from all gauges except one, then assign a unit number to the single powered gauge. Repeat this procedure for the remaining gauges. CPU Board RS-485 Termination RS-232 Wiring Note: If you have trouble using another device on the RS-485 chain, verify that it is properly terminated for its position on the chain. To terminate a device, connect a 120 resistor between its RS-485 +/ data terminals. Never terminate more than the first and last device in the chain. The Model 1400A transmitter includes a three-pin header (W4), located near the bottomfront edge of the CPU board next to the RS-485 connector (J6), to terminate the RS-485 connection. For a single unit installation or for a multiple unit installation where the transmitter is the last unit in line, the W4 jumper should be placed on pins 1 and 2 to terminate the RS-485 line. For a multiple unit installation where the transmitter is the not last unit in line, the W4 jumper should be placed on pins 2 and 3 so that the RS-485 line is not terminated at the transmitter. Refer to the Model 1400A transmitter wiring drawings (868515, ) for RS-232 wiring details. The serial port on a PC (COM1 or COM2) can be connected directly to the RS-232 port on the gauge. This requires five wires ground, transmit (TX), receive (RX), RTS and CTS. The connections between the DensityPRO+ RS-232 port connector (located on the CPU board in the Model 1400A transmitter) and the PC serial port are summarized in the table below for standard DB-9 and DB-25 PC serial port connectors. CPU Board (J8) RS-232 Connector PC Serial Port DB-9 Connector PC Serial Port DB-25 Connector J8-Pin 1 (RTS) Pin 8 Pin 5 J8-Pin 2 (TX) Pin 2 Pin 3 J8-Pin 3 (CTS) Pin 7 Pin 4 J8-Pin 4 (RX) Pin 3 Pin 2 J8-Pin 5 (GND) Pin 5 Pin Thermo Fisher Scientific

39 DensityPRO+ Chapter 3 Wiring Procedures Using Serial Communications To communicate with the DensityPRO+ gauge from a PC requires that the PC be running TMTComm for Windows communications software, or other terminal emulation software. The default communication settings for the RS-232 and RS-485 ports of the DensityPRO+ gauge and for the Thermo Scientific HHT are: 7 data bits, even parity, 1 stop bit, and 9600 baud data rate. HART Communications Refer to the DensityPRO+ Operation manual for additional details about setting up and using serial communications. The HART communication protocol is supported over the 4-20 ma current output with an optional daughter board. You communicate with the gauge using the standard Fisher Rosemount 275 HART handheld communicator. Refer to the DensityPRO+ HART Operation Manual (717816) for detailed instructions. Thermo Fisher Scientific 3-11

40 This page intentionally left blank.

41 DensityPRO+ Chapter 4 Troubleshooting and Maintenance Chapter 4 Troubleshooting and Maintenance Thermo Fisher is ready to assist you with any problems you may have with your gauge. To contact Thermo Fisher: In the United States: Thermo Fisher Scientific 1410 Gillingham Lane Sugar Land, TX Phone: Fax: In Canada: Thermo Fisher Scientific 14 Gormley Industrial Avenue Gormley, Ontario L0H 1G0 Phone: Fax: On the Web: Source Housing Periodically check the source and remove any debris that may have accumulated in the beam path between the source housing and the outer wall of the process vessel. Warning: Do not place your hand between the source and the tank. Use a brush or other tool to remove any accumulated debris. Also check the shutter to make sure it works correctly. For source housings with an exposed shutter lever pivot, you can apply grease to the pivot if necessary to prevent corrosion and jamming. Note: Do not paint or overcoat the source housing without first masking its identification tag and other labeling. All labels on the source housing must remain visible. Thermo Fisher Scientific 4-1

42 Chapter 4 Troubleshooting and Maintenance DensityPRO+ Transmitter DANGER: Remove all power from the unit before opening the transmitter enclosure for servicing. Electrocution can result if power is present. Caution: Avoid mounting the transmitter in direct sunlight. Caution: Store uninstalled transmitter in a facility that maintains the ambient temperature between -30 C and +80 C. No Display When power is applied, the gauge continuously sends the measurement readouts or a setup message to the remote terminal or HHT via the serial ports, except when you access the setup menus. If the display appears to be blank after supplying power to the gauge, it is likely that the display contrast is not set properly. If the display appears to be washed out, that is, if the display characters can be seen faintly or if the display appears to be a solid yellowgreen color, the contrast needs to be increased. If the display appears to consist of dark rectangles (20 per line), the contrast needs to be decreased. To adjust the contrast: On the NEMA 4X transmitter, press or on the keypad several times to increase or decrease the display contrast. On the explosion proof transmitter (no local keypad), the contrast adjustment is performed from the setup menus. See Contrast Adjustment for the Explosion Proof Transmitter below for a detailed description. Note: communications must be established with the gauge via the RS-232 or RS-485 serial port. If adjusting the contrast does not fix the display problem: Verify the power supply at the source. Remove power from the transmitter, then remove the cover and verify that the power cable wires are securely fastened to the correct terminals and that the connectors for RS-485 or RS-232 serial port are correctly wired. If you have display problems when using a remote terminal or the HHT, verify that the communication settings are set properly. The default gauge settings are 7 data bits, even parity, 1 stop bit, and 9600 baud. Contrast Adjustment for the Explosion Proof Transmitter It is assumed that communications have been established with the gauge via the RS-232 or RS-485 serial ports using either a Thermo Scientific HHT or a remote terminal or PC. Refer to the DensityPRO+ Operations Manual (717819) for more details on using the setup menus. 4-2 Thermo Fisher Scientific

43 DensityPRO+ Chapter 4 Troubleshooting and Maintenance Step 1 Find the Adjust Contrast Commands in the Setup Menus The Adjust Contrast commands are located in the Common action items submenu of the Action items top-level menu. From the measurement display, press the right arrow key () to enter the setup menus. Keep pressing to scroll through the top-level menu items until you reach the Action items menu heading shown below. Action items: (Erase memory, Clear alarms Hold outputs, etc) Other functions Press the down arrow key () to enter the Action items menu. The Common action items submenu heading shown below will be displayed. Common action items (clear memory, etc) NEXT Step 2 Increase or Decrease the Contrast Press to enter the Common action items submenu, and then press several to scroll through the menu items until you reach the following screen. Dec LCD Exit this menu. NEXT EXECUTE CMD If the display contrast needs to be decreased (display is too dark), press the right arrow key () several times to decrease the contrast until the desired contrast is obtained. If the display contrast needs to be increased (characters are faint or can not be seen), press the down arrow once () to move to the Inc LCD menu item shown below. Then press the right arrow key () several times to increase the contrast until the desired contrast is obtained. Inc LCD Exit this menu. NEXT EXECUTE CMD Thermo Fisher Scientific 4-3

44 Chapter 4 Troubleshooting and Maintenance DensityPRO+ To return to the previous menu item (Dec LCD), press once or press several times to scroll back to the top of the Common action items submenu and start over. When the contrast is properly adjusted, press the Exit setup key on the HHT or the x key on the terminal keypad to return to the measurement display. 4-4 Thermo Fisher Scientific

45 DensityPRO+ Chapter 4 Troubleshooting and Maintenance Adding or Replacing Transmitter Boards DANGER: Remove all power from the unit before opening the transmitter enclosure for servicing. Electrocution can result if power is present. With the exception of the AC power supply, the transmitter boards can be replaced using the procedure described below. See the next section, Transmitter AC Power Supply for the AC power supply replacement procedure. To install a board in the transmitter: Verify that power has been removed from the transmitter. Remove the screws on the faceplate (cover) and open the faceplate. Align the board with the upper and lower card guide slots and insert the board into the slots. Line up the 3 pin by 16 pin male connector on the board with the corresponding connector on the back of the transmitter. Carefully, but firmly push on the board until the connector is fully seated. Insert and tighten the board retaining screw in the upper card guide. Connect the screw-terminal connector(s) to the corresponding replacement board connectors. Close the transmitter cover and tighten cover screws. To remove a board from the transmitter: Verify that power has been removed from the transmitter. Remove the screws on the faceplate (cover) and open the faceplate. Unplug any screw-in terminal connectors (that have been wired) from the board. Remove the board retaining screw in the upper card guide. Firmly, but carefully push down on the top edge of the board to disengage the board connector from the transmitter, then pull up and out on the bottom edge of the board to remove the board. Replacing the Transmitter AC Power Supply Use the following board replacement procedure for the AC power supply: Caution: If a power supply fails, the input capacitor might still be charged at a high-voltage (up to 400 Volts). Do not remove the four screws that are used to secure the power supply to the mounting bracket. The power supply and the mounting bracket are replaced as a single unit. Note: The old power supply assembly is not field repairable. Thermo Fisher Scientific 4-5

46 Chapter 4 Troubleshooting and Maintenance DensityPRO+ Use the above procedure Adding or Replacing Transmitter Boards to remove all of the other transmitter boards (CPU, PCB 1, PCB 2, PCB 3, and PCB 4). Remove the two hold-down screws from the base of the power supply mounting bracket. These screws secure the power supply mounting bracket to the main board. Disconnect cables from the power supply (AC, DC, and Protective Safety Ground). Cut the tie wrap that secures the toroid to the power supply. Remove and replace the existing power supply and mounting bracket assembly. Reconnect the cables to the power supply (AC, DC, and Protective Safety Ground). Tighten the two hold-down screws to secure the power supply to the main board. (A screw-holding screwdriver will facilitate the installation.) Use a tie wrap to secure the toroid to the power supply mounting bracket. Use the procedure described in Adding or Replacing Boards to replace and reconnect all of the other boards (CPU, PCB 1, PCB 2, PCB 3, and PCB 4). Jumper Settings CPU Board RS-485 Termination For best performance, an RS-485 communication line should be terminated at both ends of the line. The Model 1400A transmitter includes a three-pin header (W4), located near the bottom-front edge of the CPU board next to the RS-485 connector (J6), to terminate the RS-485 connection. For a single unit installation or for a multiple unit installation where the transmitter is the last unit in line, the W4 jumper should be placed on pins 1 and 2 to terminate the RS-485 line. For a multiple unit installation where the transmitter is the not last unit in line, the W4 jumper should be placed on pins 2 and 3 so that the RS-485 line is not terminated at the transmitter. I/O Board Current Output The current output is normally shipped from the factory configured as an isolated, selfpowered output. The current output can be reconfigured as isolated loop-powered by removing the jumper from pins 3 and 4 on the J5 connector on the I/O board. In the looppowered configuration, the user must supply +24 Vdc loop power to pin 3 of J5. Pin 2 of J5 is Iout Thermo Fisher Scientific

47 DensityPRO+ Chapter 4 Troubleshooting and Maintenance VPI Board Detector Type Selection The VPI board is located in the PCB 4 slot. Three, two-pin headers (JP1, JP2, and JP3) are located along the edge of the VPI board near the top of the board. A jumper should be installed on one (and only one) of these headers to indicate the detector type as follows: JP1 ion chamber detector JP2 scintillation detector PRO mode JP3 scintillation detector SGD mode Caution: Remove power from the transmitter before changing jumper settings. Failure to do so may result in damage to the VPI board. Current Board Problems Relay Problems If you suspect a problem with the current output, attach an ammeter in series with the current output load and verify the current output at various levels. Use the commands in the User service & related items submenu, located under the Security, service, and diagnostic functions top-level menu, to hold the current output at specific levels for testing (Refer to Chapter 9 of the DensityPRO+ Operation manual for details). If you suspect a problem with a relay output, you can attach a continuity tester to the suspected relay output and use the Test relays commands to test relay operation. (Refer to Chapter 9 of the DensityPRO+ Operation manual for details.) Scintillation Detector Danger: Remove all power from the unit before servicing. Electrocution can result if power is present. In hazardous locations, ensure that power is removed from the detector before removing the housing cover. Be sure that the housing cover has been replaced and that the grounds are properly connected before reapplying power. Caution: Close the shutter on the source housing before servicing the detector unit. AC Power Supply Fuse Caution: For reliable operation and to maintain safety approvals, the F1 fuse on the AC power board must be replaced with an approved fuse. See the Installation Wiring Diagram (868515) or Appendix B Parts List. Thermo Fisher Scientific 4-7

48 Chapter 4 Troubleshooting and Maintenance DensityPRO+ Factory Wiring Jumper Settings The detector board (positioned horizontally at the bottom of the electronics chassis) mounts on the photomultiplier tube. A factory-installed cable connects the detector board to the J2 connector on the CPU board. If the optional AC power board is installed, a factory-installed cable connects the J3 connector on the AC power board to the J4 connector on the scintillator adapter board. If the optional 12-to-24 Vdc converter board is installed, a factory-installed cable connects the J2 connector on the 12-to-24 Vdc converter board to the J4 connector on the scintillator adapter board. Refer to Drawing sheet 2 for details of the jumper locations and pin numbering. Scintillation Detector Troubleshooting Note: Enable the service-only items in the Special functions menu before proceeding with this section. Refer to Chapter 9 of the DensityPRO+ Operation Manual for details on the service-related menus. If you suspect a detector problem, you can quickly check the detector output using the Internal value of sensor signal item in the Snapshot menu (Direct entry code ). This value (count rate in counts per second) should be much larger with the source shutter ON than with the shutter OFF (typically more than ten times larger). As a further check, if standardization has been completed, set up your standard configuration and compare the Internal value of sensor signal to the STD value in use (standardization value) displayed in the Sensor standardization submenu of the Gauge fine tuning menu (Direct entry code ). If the gauge was working properly when it was last standardized, the two values should be similar. (The match is not exact because the STD value is filtered and corrected for background radiation and source decay.) If the displayed signal value is not behaving correctly, be sure to consider misalignment of the gauge head, buildup or wear, debris in the beam path, or faulty shutter operation before testing the detector signals. If you can eliminate these other factors, the cause is probably on the CPU board. Verify that the connectors on the cable connecting the Detector board to the CPU board are securely seated on both boards. If the problem remains, call Thermo Fisher for assistance. 4-8 Thermo Fisher Scientific

49 DensityPRO+ Chapter 4 Troubleshooting and Maintenance Ionization Detector Caution: Close the shutter on the source housing before servicing the detector unit. AC Power Supply Fuse Caution: For reliable operation and to maintain safety approvals, the F1 fuse on the Ion Chamber detector board must be replaced with an approved fuse. See Appendix B Parts List. Ion Chamber Detector Troubleshooting Note: Select Do display service-only items in the Special functions menu before proceeding with this section. If you suspect a detector problem, you can quickly check the detector output using the Internal value of sensor signal item in the Snapshot menu. The value displayed with the source shutter ON should be much larger than with the shutter OFF (typically more than ten times larger). As a further check, set up your standard configuration and compare the sensor signal to the STD value displayed in the Gauge fine tuning standardization menu. If the gauge was working properly when it was last standardized, the two values should be similar. (The match is not exact because the STD value is filtered and corrected for background radiation and source decay.) If the displayed signal value is not behaving correctly, be sure to consider misalignment of the gauge head, pipe buildup or wear, debris in the beam path, or faulty shutter operation (especially on density probes) before testing the detector signals. If you can eliminate these other factors, the cause is probably in the VPI board or in the detector. To see which is at fault, use an ammeter to measure the detector signal at the ionization detector input of the transmitter s VPI board. (The terminal positions for each connector are labeled inside the transmitter cover or refer to the appropriate wiring diagram.) The current should be between 1 and 10 ma under normal operation, and should drop (typically more than 90 percent) when the source shutter is turned from ON to OFF. If the measured current appears to behave correctly, then the VPI board is probably at fault. Otherwise, check the signal cable between the detector and transmitter for opens or shorts. If the cable is intact, the cause is probably in the detector. Contact Thermo Fisher for assistance. Thermo Fisher Scientific 4-9

50 This page intentionally left blank.

51 DensityPRO+ Appendix A Returning Equipment for Service Appendix A How to Return Equipment for Service Contact Thermo Fisher for specific instructions. Please have the following available: 1. Specific information about the problem. 2. A contact name and phone number, in case we need more information. 3. A purchase order authorizing repairs, or a request for a quote. Thermo Fisher Scientific A-1

52 This page intentionally left blank.

53 DensityPRO+ Appendix B Parts List Appendix B Parts List Use the following list of part numbers to order spare parts for the DensityPRO+. Call Thermo Fisher for pricing and assistance. Part Number Description Model 1400A Transmitter Power Supply Assembly CPU Board I/O Board, 24 Vdc Input Power, No Relays I/O Board, 24 Vdc Input Power, 2 Relays I/O Board, 12 Vdc Input Power, No Relays I/O Board, 12 Vdc Input Power, 2 Relays I/O Board, w/out DC Input Power, 0 Relays I/O Board, w/out DC Input Power, 2 Relays I/O Board Fuse-F1, 250V, 0.75A, Slo-Blow HART Communication Option VPI Board Model 9719B/9720B Scintillation Detector Scintillation Adapter (Transmitter Interface) Board AC Power Supply Board Fuse-F1, 250V, 0.125A (1/8A SB), 3AG to-24 Vdc Converter Board Temperature Compensation Board Model 9701/9702 Ion Chamber Detector Preamp Power Board Fuse-F1, 250V, 5A, Slo-Blow, 3AG Thermo Fisher Scientific B-1

54 This page intentionally left blank.

55 DensityPRO+ Appendix C Specifications Appendix C Specifications This list contains the specifications for the DensityPRO+ detector, transmitter and system capabilities. Operating Principle Gamma radiation from the source passes through the process pipe and material. The amount of radiation reaching detector decreases as the process level increases. Model 1400A Transmitter Power Supply Power input is transient surge protected. DC Power 24 Vdc Standard Vdc, 12 VA 12 Vdc Optional/Custom 9-15 Vdc, 15 VA (No FM/CSA Approvals) AC Power - Optional 100 to 240 Vac, 50/60 Hz, 25 VA FM / CSA Approvals Non-Explosion Proof (plastic) Enclosure: Class I, Div. 2, Groups A, B, C,D Class II, Div. 2, Groups E*,F, G Class III NEMA 4X, Type 4X *Grp E applies to CSA only Explosion Proof Enclosure: Class I, Div 1, Groups C, D Class II, Div 1, Groups E, F, G Class III NEMA 4 / 7, Type 4 / 7 Dimensions Non-Explosion Proof (plastic) Enclosure: 12 in. x 8.5 in. x 8.25 in. (305 mm x 216 mm x 210 mm) Operating Temperature -40 o to +60 o C (-40 o to +140 o F) Circuit Protection Meets IEC 801-1,-2, and 3 Current outputs and relay outputs are isolated Serial Interface RS-232: Full duplex communication with a remote terminal, printer, PC or host computer. RS-485: Half-duplex party-line communication among multiple gauges or between gauge and remote terminal or Thermo Scientific HHT. Current Output One current output (I/O board) standard, second and third current outputs optional. Standard Configuration: Isolated, Self-Powered, 700 ohm max. load. Alternate Configuration: Isolated, Loop-powered, 24 Vdc Nominal Supply Voltage, 800 ohm max. load) Contact Closure Inputs Two contact closure inputs per I/O Board can be programmed to execute commands based on userprovided input. Relays Two optional relays per I/O board (six relays max.); Form C SPDT isolated, 8 amp, 220 VAC. Cable (detector to transmitter) Ion Chamber Single Pair, Foil Shielded wire 22 AWG: maximum 1000 ft (300 m) AWG: maximum 5000 ft (1525 m) Scintillator Detector Individually shielded pairs required AWG: maximum 1000 ft (300 m) Explosion Proof Enclosure: 14 in. x 15.5 in. x 10.5 in. (356 mm x 394 mm x 267 mm) Weight Non-Explosion Proof Enclosure: 3.4 kg (7.5 lbs.) Explosion Proof Enclosure: 27.2 kg (60 lbs.) Thermo Fisher Scientific C-1

56 Appendix C Specifications DensityPRO+ Scintillator Detector (Model 9719B/9720B) Power Supply DC Power 24 Vdc ±20%, 6 VA standard. 12 Vdc ±33%, 6 VA, with 12-to-24 V converter board option. (Not available with AC Power option.) AC Power 115 / 230 Vac ±15%, 50/60 Hz, 6 VA, with AC Power board option. 115 / 230 Vac selection switch located on AC Power Board. (Not available with 12 Vdc Power option.) FM / CSA Approvals Non-Explosion Proof (9719B): Class II, Div. 1, Groups E, F, G Class III, Div. 1 NEMA 4, Type 4 Explosion Proof (9720B): Class I, Div 1, Groups B, C, D Class II, Div 1, Groups E, F, G Class III NEMA 4, Type 4 Dimensions Non-Explosion Proof (9719B): 6.5 in. diameter x 14 in. tall (165 mm diameter x 360 mm tall) Explosion Proof (9720B): 6.5 in. diameter x 15 in. tall (165 mm diameter x 380 mm tall) ION Chamber Detector (Model 9701/9702) Power Supply AC Power 115 / 230 Vac ±10%, 50/60 Hz, 110 VA. FM / CSA Approvals Non-Explosion Proof (9701): NEMA 4, Type 4 Explosion Proof (9702): Class I, Div 1, Groups B, C, D Class II, Div 1, Groups E, F, G Class III NEMA 4, Type 4 Dimensions Non-Explosion Proof (9719B): ~8 in. diameter x 15.5 in. tall (200 mm diameter x 400 mm tall) Explosion Proof (9720B): ~8 in. diameter x 19.5 in. tall (200 mm diameter x 500 mm tall) Weight Non-explosion proof: 58 lb. (26.3 kg) Explosion proof: 100 lb. (45.4 kg) Operating Temperature -30 o to +60 o C (-20 o to +140 o F) Operating Temperature -30 o to +60 o C (-20 o to +140 o F) C-2 Thermo Fisher Scientific

57 DensityPRO+ Appendix D Drawings Appendix D Drawings Information presented in this chapter has been generated from original drawings. Every effort is made to maintain document accuracy. However, in order to enhance legibility, the documents may have been restructured, and some information may have been intentionally excluded. Therefore, the drawings within this guide may not be exact duplicates of the original drawings. Drawings in this manual are included for reference only and may not be the current version. Contact the factory if you need a copy of the latest revision. Table D-1. Installation & Wiring Drawing Rev. Description Page A Installation Layout, Model 1400A Transmitter to 9719B / 9720B / 9701 / 9702 Detectors (2 sheets) D A Installation Wiring, Model 1400A Transmitter (3 sheets) D D Installation Wiring, Model 9719B / 9720B Scintillation Transducer to Model 1400A Transmitter (3 sheets) D A Installation Wiring, Model 9701 / 9702 Ion Chamber Detector (1 sheet) D B RS485 Installation & Wiring (1 sheet) D 12 Table D-2. Mechanical Transmitter Drawing Rev. Description Page F Fitting Installation (CSA/FM), NEMA 4X Enclosure, Model 1400 Transmitter (2 sheets) D D Mounting Dimensions, XP Enclosure, Model 1400 Transmitter (1 sheet) D 15 Table D-3. Mechanical Detectors Drawing Rev. Description Page B Installation Drawing, Detector Assembly, Flat Mount, NEMA 4 Housing (2 sheets) D B Installation Drawing, Detector Assembly, Flat Mount, Explosion Proof Housing (2 sheets) D Mounting Dimensions, Model 9701 (Ion Chamber), NEMA 4 Housing (1 sheet) D Mounting Dimensions, Model 9702 (Ion Chamber), Explosion Proof Housing (1 sheet) D B Installation Drawing, Source/Detector Assembly Single Chain Mount, NEMA 4 Housing (1 sheet) B Installation Drawing, Source/Detector Assembly Single Chain Mount, Explosion Proof Housing (1 sheet) B Installation Drawing, Source/Detector Assembly Two Chain Mount, NEMA 4 Housing (3 sheets) D 22 D 23 D 24 Thermo Fisher Scientific D-1

58 Appendix D Drawings DensityPRO+ Table D-3. Mechanical Detectors, cont. Drawing Rev. Description Page B Installation Drawing, Source/Detector Assembly Two Chain Mount, Explosion Proof Housing (3 sheets) A Installation, Gauge Head, Model 9701/9702 Ion Chamber Detector, Two Chain Mount (1 sheet) D 27 D N E Installation Drawing, Gauge Head, Pipe Saddle Mount (1 sheet) D C Fabrication Details, 1 inch - 3 inch Z-Pipe Section (2 sheets) D B Mounting Dimensions, 1 inch - 3 inch Z-Pipe Mounting Configuration (1 sheet) D A Fabrication Details, 4 inch Z-Pipe Section (1 sheet) D A Mounting Dimensions, 4 inch Z-Pipe Mounting Configuration (1 sheet) D 36 Table D-4. Mechanical Source Heads Drawing Rev. Description Page A Mounting Dimensions, Source Housing, 100 mci or Less (Model 5200) (1 sheet) D B Mounting Dimensions, Source Housing, 100 mci or Less (Model 5201) (1 sheet) D C Mounting Dimensions, Source Housing, 500 mci or Less (Model 5202) (1 sheet) D B Mounting Dimensions, Source Housing, 2000 mci or Less (Model 5203) (2 sheets) D B Mounting Dimensions, Source Housing, 8000 mci or Less (Model 5204) (2 sheets) D 42 Table D-5. Source Housing Options Drawing Rev. Description Page E Mounting Instructions, Remote Manual Actuator, Models Source Housings (2 sheets) D 44 D-2 Thermo Fisher Scientific

59 DensityPRO+ Appendix D Drawings (rev. A): Installation Layout, Model 1400A Transmitter to 9719B/9720B/9701/9702 Detectors (sheet 1 of 2) Thermo Fisher Scientific D-3

60 Appendix D Drawings DensityPRO (rev. A): Installation Layout, Model 1400A Transmitter to 9719B/9720B/9701/9702 Detectors (sheet 2 of 2) D-4 Thermo Fisher Scientific

61 DensityPRO+ Appendix D Drawings (rev. A): Installation Wiring, Model 1400A Transmitter (sheet 1 of 3) Thermo Fisher Scientific D-5

62 Appendix D Drawings DensityPRO (rev. A): Installation Wiring, Model 1400A Transmitter (sheet 2 of 3) D-6 Thermo Fisher Scientific

63 DensityPRO+ Appendix D Drawings (rev. A): Installation Wiring, Model 1400A Transmitter (sheet 3 of 3) Thermo Fisher Scientific D-7

64 Appendix D Drawings DensityPRO (rev. D): Installation Wiring, Model 9719B / 9720B Scintillation Transducer to Model 1400A Transmitter (sheet 1 of 3) D-8 Thermo Fisher Scientific

65 DensityPRO+ Appendix D Drawings (rev. D): Installation Wiring, Model 9719B / 9720B Scintillation Transducer to Model 1400A Transmitter (sheet 2 of 3) Thermo Fisher Scientific D-9

66 Appendix D Drawings DensityPRO (rev. D): Installation Wiring, Model 9719B / 9720B Scintillation Transducer to Model 1400A Transmitter (sheet 3 of 3) D-10 Thermo Fisher Scientific

67 DensityPRO+ Appendix D Drawings (rev. A): Installation Wiring, Model 9701 / 9702 Ion Chamber Detector (sheet 1 of 1) Thermo Fisher Scientific D-11

68 Appendix D Drawings DensityPRO (rev. B): Installation Wiring, RS485 (sheet 1 of 1) D-12 Thermo Fisher Scientific

69 DensityPRO+ Appendix D Drawings (rev. F): Fitting Installation (CSA/FM), NEMA 4X Enclosure, Model 1400 Transmitter (sheet 1 of 2) Thermo Fisher Scientific D-13

70 Appendix D Drawings DensityPRO (rev. F): Fitting Installation (CSA/FM), NEMA 4X Enclosure, Model 1400 Transmitter (sheet 2 of 2) D-14 Thermo Fisher Scientific

71 DensityPRO+ Appendix D Drawings (rev. D): Mounting Dimensions, XP Enclosure, Model 1400 Transmitter (sheet 1 of 1) Thermo Fisher Scientific D-15

72 Appendix D Drawings DensityPRO (rev. B): Installation Drawing, Detector Assembly, Flat Mount, NEMA 4 Housing (sheet 1 of 2) D-16 Thermo Fisher Scientific

73 DensityPRO+ Appendix D Drawings (rev. B): Installation Drawing, Detector Assembly, Flat Mount, NEMA 4 Housing (sheet 2 of 2) Thermo Fisher Scientific D-17

74 Appendix D Drawings DensityPRO (rev. B): Installation Drawing, Detector Assembly, Flat Mount, Explosion Proof Housing (sheet 1 of 2) D-18 Thermo Fisher Scientific

75 DensityPRO+ Appendix D Drawings (rev. B): Installation Drawing, Detector Assembly, Flat Mount, Explosion Proof Housing (sheet 2 of 2) Thermo Fisher Scientific D-19

76 Appendix D Drawings DensityPRO (rev. --): Mounting Dimensions, Model 9701 (Ion Chamber), NEMA 4 Housing (sheet 1 of 1) D-20 Thermo Fisher Scientific

77 DensityPRO+ Appendix D Drawings (rev. --): Mounting Dimensions, Model 9702 (Ion Chamber), Explosion Proof Housing (sheet 1 of 1) Thermo Fisher Scientific D-21

78 Appendix D Drawings DensityPRO (rev. B): Installation Drawing, Source/Detector Assembly Single Chain Mount, NEMA 4 Housing (sheet 1 of 1) D-22 Thermo Fisher Scientific

79 DensityPRO+ Appendix D Drawings (rev. B): Installation Drawing, Source/Detector Assembly Single Chain Mount, Explosion Proof Housing (sheet 1 of 1) Thermo Fisher Scientific D-23

80 Appendix D Drawings DensityPRO (rev. B): Installation Drawing, Source/Detector Assembly Two Chain Mount, NEMA 4 Housing (sheet 1 of 3) D-24 Thermo Fisher Scientific

81 DensityPRO+ Appendix D Drawings (rev. B): Installation Drawing, Source/Detector Assembly Two Chain Mount, NEMA 4 Housing (sheet 2 of 3) Thermo Fisher Scientific D-25

82 Appendix D Drawings DensityPRO (rev. B): Installation Drawing, Source/Detector Assembly Two Chain Mount, NEMA 4 Housing (sheet 3 of 3) D-26 Thermo Fisher Scientific

83 DensityPRO+ Appendix D Drawings (rev. B): Installation Drawing, Source/Detector Assembly Two Chain Mount, Explosion Proof Housing (sheet 1 of 3) Thermo Fisher Scientific D-27

84 Appendix D Drawings DensityPRO (rev. B): Installation Drawing, Source/Detector Assembly Two Chain Mount, Explosion Proof Housing (sheet 2 of 3) D-28 Thermo Fisher Scientific

85 DensityPRO+ Appendix D Drawings (rev. B): Installation Drawing, Source/Detector Assembly Two Chain Mount, Explosion Proof Housing (sheet 3 of 3) Thermo Fisher Scientific D-29

86 Appendix D Drawings DensityPRO (rev. A): Installation, Gauge Head, Model 9701/9702 Ion Chamber Detector, Two Chain Mount (sheet 1 of 1) D-30 Thermo Fisher Scientific

87 DensityPRO+ Appendix D Drawings 85726N (rev. E): Installation Drawing, Gauge Head, Pipe Saddle Mount (sheet 1 of 1) Thermo Fisher Scientific D-31

88 Appendix D Drawings DensityPRO (rev. C): Fabrication Details, 1 inch - 3 inch Z-Pipe Section (sheet 1 of 2) D-32 Thermo Fisher Scientific

DensityPRO+ Gamma Density System with Remote Transmitter Installation Guide P/N Part of Thermo Fisher Scientific.

DensityPRO+ Gamma Density System with Remote Transmitter Installation Guide P/N Part of Thermo Fisher Scientific. DensityPRO+ Gamma Density System with Remote Transmitter Installation Guide P/N 717818 Revision F Part of Thermo Fisher Scientific DensityPRO+ Gamma Density System with Remote Transmitter Installation

More information

DensityPRO+ Gamma Density System with Remote Transmitter User Guide P/N Part of Thermo Fisher Scientific. Revision C

DensityPRO+ Gamma Density System with Remote Transmitter User Guide P/N Part of Thermo Fisher Scientific. Revision C DensityPRO+ Gamma Density System with Remote Transmitter User Guide P/N 717819 Revision C Part of Thermo Fisher Scientific DensityPRO+ Gamma Density System with Remote Transmitter User Guide P/N 717819

More information

Continuous Gamma Level System Installation Guide P/N

Continuous Gamma Level System Installation Guide P/N 4790 Continuous Gamma Level System Installation Guide P/N 717830 Revision F Part of Thermo Fisher Scientific 4790 Continuous Gamma Level System Installation Guide P/N 717830 Revision F 2009 Thermo Fisher

More information

LevelPRO Measurement Systems. Continuous Gamma Level System Installation Guide PN

LevelPRO Measurement Systems. Continuous Gamma Level System Installation Guide PN LevelPRO Measurement Systems Continuous Gamma Level System Installation Guide PN 1-0702-040 LevelPRO Measurement Systems Continuous Gamma Level System Installation Guide PN 1-0702-040 2014 Thermo Fisher

More information

DensityPRO C. Gamma Density System User Guide P/N Part of Thermo Fisher Scientific. Revision D

DensityPRO C. Gamma Density System User Guide P/N Part of Thermo Fisher Scientific. Revision D DensityPRO C Gamma Density System User Guide P/N 900-748300 Revision D Part of Thermo Fisher Scientific DensityPRO C Gamma Density System User Guide P/N 900-748300 Revision D 2007 Thermo Fisher Scientific

More information

Nuclear Density Gauges

Nuclear Density Gauges Nuclear Density Gauges DensityPRO DensityPRO+ DensityPRO-C ConsistencyPRO CutPRO 3680 Authorised Distributor www.transmin.com.au Thermo Scientific Nuclear Density Gauges Energy Attenuation Gauges The expandable

More information

Ronan Model X96S Series Density System

Ronan Model X96S Series Density System Ronan Model X96S Series Density System For more than 27 years Ronan has provided the process control industry with unsurpassed products in the harshest environments. With a philosophy of designing new

More information

Ronan Model X96S Series Weighing System

Ronan Model X96S Series Weighing System Ronan Model X96S Series Weighing System For more than 27 years Ronan has provided the process control industry with unsurpassed products in the harshest environments. With a philosophy of designing new

More information

Troubleshooting Thermo Scientific LevelPRO Gauges

Troubleshooting Thermo Scientific LevelPRO Gauges Troubleshooting Thermo Scientific LevelPRO Gauges Key Words Scintillation detector Current board Relay output Communications Level indication Purpose A chapter in the user guide for the Thermo Scientific

More information

M E A S U R E M E N T S D I V I S I O N

M E A S U R E M E N T S D I V I S I O N M E A S U R E M E N T S D I V I S I O N ACCURATELY MEASURING WHAT YOU DIDN T THINK WAS MEASURABLE Ronan Engineering, founded in 1959, is a privately held Corporation with the exclusive purpose of manufacturing

More information

Model 144H. Product Data Sheet , Rev DB February Content

Model 144H. Product Data Sheet , Rev DB February Content Rosemount 144 PC-Programmable Temperature Transmitter Provides an installation-ready solution for temperature monitoring applications using Complete Point Solutions (CPS) Increases measurement accuracy

More information

Operating Instructions

Operating Instructions Operating Instructions DSGH Radiation-Based Detector with GEN2000 Electronics for Density Measurement Document ID: 31392 Nuclear Revision history Version Description Date 1.0 Initial release. Formerly

More information

Ronan Model X90 Series Point Level Monitor

Ronan Model X90 Series Point Level Monitor Ronan Model X90 Series Point Level Monitor For more than 27 years Ronan has provided the process control industry with unsurpassed products in the harshest environments. With a philosophy of designing

More information

DSGH. Radiation-Based Detector with GEN2000 Electronics for Density Measurement QUICK REFERENCE GUIDE

DSGH. Radiation-Based Detector with GEN2000 Electronics for Density Measurement QUICK REFERENCE GUIDE DSGH Radiation-Based Detector with GEN2000 Electronics for Density Measurement QUICK REFERENCE GUIDE Revision History Revision History Version of manual Description Date 1.0 Initial release 051025 1.1

More information

INSTALLATION, OPERATING & MAINTENANCE INSTRUCTIONS FOR 870 SERIES INDUSTRIAL CONTROL PANELS

INSTALLATION, OPERATING & MAINTENANCE INSTRUCTIONS FOR 870 SERIES INDUSTRIAL CONTROL PANELS INSTALLATION, OPERATING & MAINTENANCE INSTRUCTIONS FOR 870 SERIES INDUSTRIAL CONTROL PANELS GENERAL INDEECO Industrial Control Panels are designed to provide years of trouble free operation if properly

More information

Duct and Rough Service Carbon Monoxide Sensor

Duct and Rough Service Carbon Monoxide Sensor Product Identification and Overview Duct and Rough Service Carbon Monoxide Sensor BAPI s Carbon Monoxide Sensor offers enhanced electrochemical sensing with outstanding accuracy at low concentrations.

More information

THERMAL BUILDING SOLUTIONS EN-TraceTekTTSIM1A-IM-H /16

THERMAL BUILDING SOLUTIONS EN-TraceTekTTSIM1A-IM-H /16 TraceTek TTSIM-1A TraceTek Sensor Interface Module with Relay Installation/OPERATION Instructions Approvals and Certifications TYPE NM General Signaling Equipment 76LJ Only AC versions are UL listed and

More information

TYPE CM-2201 NELSON SINGLE POINT CIRCUIT MANAGEMENT SYSTEM

TYPE CM-2201 NELSON SINGLE POINT CIRCUIT MANAGEMENT SYSTEM 2 Line, 16 Characters/row LCD Display Temperature Input Range -50 C to +500 C -58 F to + 932 F Enclosure NEMA Type 4X Current Rating 30A max (resistive load only) Ambient Temperature -40 C to + 40 C -40

More information

installation and operation manual for Hunter Ceiling Fans

installation and operation manual for Hunter Ceiling Fans For Your Records and Warranty Assistance Model Name: Catalog/Model No.: Serial No.: Date Purchased: Where Purchased: For reference also attach your receipt or a copy of your receipt to the manual. installation

More information

Model D12-IR. Digital Gas Transmitter...

Model D12-IR. Digital Gas Transmitter... Model D12-IR Digital Gas Transmitter... ATI s Series D12 gas transmitter line now includes a versatile Infrared system that can be configured for LEL or select toxic gases. The D12-IR utilizes a compact

More information

Model 144H. Product Data Sheet , Rev DB February Content

Model 144H. Product Data Sheet , Rev DB February Content Rosemount 144 PC-Programmable Temperature Transmitter Provides an installation-ready solution for temperature monitoring applications using Complete Point Solutions (CPS) Increases measurement accuracy

More information

Analog Room Pressure Monitor RPC Series

Analog Room Pressure Monitor RPC Series Description The Room Pressure Monitor is used to measure differential pressure in the range of 0.125 to 1"wc or 30 to 250 Pa. It combines precision high sensitivity silicon sensing capabilities and the

More information

Owner s Guide and Installation Manual

Owner s Guide and Installation Manual For Your Records and Warranty Assistance For reference, also attach your receipt or a copy of your receipt to the manual. Model Name Type 2 Models Owner s Guide and Installation Manual Model No. Catalog

More information

Owner s Guide and Installation Manual

Owner s Guide and Installation Manual For Your Records and Warranty Assistance For reference, also attach your receipt or a copy of your receipt to the manual. Model Name Type 2 Models Owner s Guide and Installation Manual Model No. Date Purchased

More information

SAFETY INFORMATION AND WARNINGS

SAFETY INFORMATION AND WARNINGS This manual refers to the Model SST-3 control panel manufactured since October 31, 2013, which uses a universal (100 277 VAC; 50/60 Hz) power supply. Older units use a voltage-specific power supply and

More information

CP CP

CP CP CP-8161-333 CP-8161-433 Electronic Programmable Controller Six Stage, Dual Setpoint General Instructions APPLICATION Electronic six stage programmable controller with proportional output for heating, cooling

More information

TTSIM-2 TRACETEK SENSOR INTERFACE MODULE WITH LCD AND RELAY INSTALLATION/OPERATION INSTRUCTIONS

TTSIM-2 TRACETEK SENSOR INTERFACE MODULE WITH LCD AND RELAY INSTALLATION/OPERATION INSTRUCTIONS TTSIM-2 TRACETEK SENSOR INTERFACE MODULE WITH LCD AND RELAY INSTALLATION/OPERATION INSTRUCTIONS GENERAL INFORMATION Please read these instructions and keep them in a safe place. These instructions must

More information

INSTALLATION, OPERATING & MAINTENANCE INSTRUCTIONS FOR 350 SERIES CIRCULATION HEATERS

INSTALLATION, OPERATING & MAINTENANCE INSTRUCTIONS FOR 350 SERIES CIRCULATION HEATERS INDEECO Circulation Heaters are designed to provide years of trouble free operation if properly installed and maintained. Please read and follow these instructions for installing and maintaining the heater.

More information

NGC-UIT-ORD User interface terminal for Raychem NGC systems Installation Instructions

NGC-UIT-ORD User interface terminal for Raychem NGC systems Installation Instructions NGC-UIT-ORD User interface terminal for Raychem NGC systems Installation Instructions Description The NGC-UIT-ORD is a panel mounted display used in conjunction with other approved Raychem control and

More information

AM-806. Waste/Door Radiation Monitor OPERATIONS MANUAL

AM-806. Waste/Door Radiation Monitor OPERATIONS MANUAL AM-806 Waste/Door Radiation Monitor OPERATIONS MANUAL Section Contents 1. SPECIFICATIONS 2. MAIN COMPONENTS 3. THEORY OF OPERATION 4. OPERATING INSTRUCTIONS 5. INSTALLATION Page 2 1.0 SPECIFICATIONS SENSITIVITY

More information

SECTION ULTRASONIC TRANSIT TIME CLAMP-ON FLOW MEASURING SYSTEM

SECTION ULTRASONIC TRANSIT TIME CLAMP-ON FLOW MEASURING SYSTEM SECTION 40 71 66 ULTRASONIC TRANSIT TIME CLAMP-ON FLOW MEASURING SYSTEM PART 1-GENERAL 1.01 SUMMARY A. Ultrasonic clamp-on flow measurement system designed for non-intrusive external to pipe measurement

More information

Rev B, 9/2/2009. Kodiak Chiller Overview

Rev B, 9/2/2009. Kodiak Chiller Overview 930-0001 Rev B, 9/2/2009 Kodiak Chiller Overview Presentation Outline Phone: 781-933-7300 Lytron Technical Support Contact Information 3 Introduction 4 Part I: Unpacking 5 Part II: Installation 7 Part

More information

Table of Contents SECTION PAGE SECTION 1 INTRODUCTION Description Features Models... SECTION 2 RS-232 COMMUNICATIONS...

Table of Contents SECTION PAGE SECTION 1 INTRODUCTION Description Features Models... SECTION 2 RS-232 COMMUNICATIONS... SECTION Table of Contents SECTION 1 INTRODUCTION...................... 1.1 Description............................... 1.2 Features................................. 1.3 Models..................................

More information

Eldex Column Heater. Operator s Manual

Eldex Column Heater. Operator s Manual Eldex Eldex Column Heater Operator s Manual Eldex Laboratories, Inc. 30 Executive Court Napa, CA 94558 Tel: (707) 224-8800 Fax: (707) 224-0688 www.eldex.com Rev. B: 081700 2000Eldex Laboratories, Inc.

More information

Greentrol Automation, Inc.

Greentrol Automation, Inc. Model GA-100 Technical Manual GreenAlert 100 Series Greentrol Automation, Inc. Installation, Operation and Maintenance Technical Manual GA-100 Document Name: TM_GA-100_R1C TM_GA-100_R1C Part Number: 960--0000

More information

FLT93 Installation, Operation and Troubleshooting Guide

FLT93 Installation, Operation and Troubleshooting Guide FLT93 Installation, Operation and Troubleshooting Guide Pre-Installation A. To get the best results from the instrument, the instrument should be mounted 20 pipe diameters downstream from any valve, pipe

More information

Carbon Monoxide Transmitter

Carbon Monoxide Transmitter Introduction The CO Transmitter uses an electrochemical sensor to monitor the carbon monoxide level and outputs a field-selectable 4-20 ma or voltage signal. The voltage signal may also be set to 0-5 or

More information

TTSIM-1A. TraceTek Sensor Interface Module with Relay. Installation/Operation Instructions. Installation Items (not supplied) Tools Required.

TTSIM-1A. TraceTek Sensor Interface Module with Relay. Installation/Operation Instructions. Installation Items (not supplied) Tools Required. TTSIM-1A TraceTek Sensor Interface Module with Relay Installation Items (not supplied) General Information Installation/Operation Instructions Please read these instructions and keep them in a safe place.

More information

High Pressure Conductivity Monitor

High Pressure Conductivity Monitor Product Data Sheet PDS 71-EDS880/1/rev.A August 2006 Model EDS880/1 High Pressure Conductivity Monitor COMPLETE SYSTEM FOR THE DETERMINATION OF CONDUCTIVITY IN HIGH PRESSURE SYSTEMS SENSOR INSTALLS DIRECTLY

More information

OPTISENS TUR 2000 Technical Datasheet

OPTISENS TUR 2000 Technical Datasheet OPTISENS TUR 2000 Technical Datasheet Sensor for turbidity measurement in water and wastewater Rugged design for harsh applications Integrated transmitter with direct 4...20 ma output Near infrared light

More information

Owner s Guide and Installation Manual

Owner s Guide and Installation Manual For Your Records and Warranty Assistance For reference, also attach your receipt or a copy of your receipt to the manual. Model Name Type 8 Models Owner s Guide and Installation Manual Model No. Catalog

More information

INSTALLATION INSTRUCTIONS

INSTALLATION INSTRUCTIONS INSTALLATION INSTRUCTIONS The SensorNet I/O Module provides attachment for up to 32 alarms and up to 16 switched devices. An RS-232 or RS-422 communications port is also provided for receiving serial alarms.

More information

Table of Contents. What to Expect with. Mounting Options. Tools Needed

Table of Contents.   What to Expect with. Mounting Options. Tools Needed Table of Contents www.hunterfan.com What to Expect with Your Installation Congratulations on purchasing your new Hunter ceiling fan! It will provide comfort and performance in your home or office for many

More information

Environmental Radiation Monitoring System

Environmental Radiation Monitoring System Environmental Radiation Monitoring System Toshihiro Takagi Osamu Kimura Atsushi Minagoshi 1. Introduction In order to ascertain conditions near the perimeter of a supervised area at a nuclear power plant

More information

OPERATING AND MAINTENANCE MANUAL FOR PLATE HEAT EXCHANGER INDIRECT FIRED WATER HEATER. Electric Heater Company Base Model "BWXP"

OPERATING AND MAINTENANCE MANUAL FOR PLATE HEAT EXCHANGER INDIRECT FIRED WATER HEATER. Electric Heater Company Base Model BWXP OPERATING AND MAINTENANCE MANUAL FOR PLATE HEAT EXCHANGER INDIRECT FIRED WATER HEATER Electric Heater Company Base Model "BWXP" HUBBELL ELECTRIC HEATER COMPANY P.O. BOX 288 STRATFORD, CT 06615 PHONE: (203)

More information

SAFETY CERTIFIED MODEL FP-700 COMBUSTIBLE GAS DETECTOR

SAFETY CERTIFIED MODEL FP-700 COMBUSTIBLE GAS DETECTOR SAFETY MANUAL SIL 2 Certified Model FP-700 Combustible Hydrocarbon Gas Sensor Version 2.0 1 SAFETY CERTIFIED MODEL FP-700 COMBUSTIBLE GAS DETECTOR This manual addresses the specific requirements and recommendations

More information

Isco 3700ZR Sampler. Manual Set

Isco 3700ZR Sampler. Manual Set Isco 3700ZR Sampler Manual Set Assembly #60-2744-045 Copyright 2001. All rights reserved, Teledyne Isco Revision D, December 14, 2012 This set of manuals supports the Isco 3700ZR Refrigerated Sampler.

More information

Technical Reference Manual

Technical Reference Manual Technical Reference Manual DSGH Radiation-Based Detector with GEN2000 Electronics for Density Measurement Document ID: 31399 Nuclear Revision history Table 1: Revision History Version Description Date

More information

Single Point Freeze Protection Heat Trace Control TRACON MODEL FPT 130 Installation and Operation Manual

Single Point Freeze Protection Heat Trace Control TRACON MODEL FPT 130 Installation and Operation Manual We manage heat MANUAL Single Point Freeze Protection Heat Trace Control TRACON MODEL FPT 130 Installation and Operation Manual 1850 N Sheridan Street South Bend, Indiana 46628 (574) 233-1202 or (800) 234-4239

More information

Table of Contents What to Expect with Your Installation. Top Housing. Ceiling Plate. Tools Needed.

Table of Contents What to Expect with Your Installation. Top Housing. Ceiling Plate. Tools Needed. Table of Contents Congratulations on purchasing your new Hunter ceiling fan! It will provide comfort and performance in your home or office for many years. This installation and operation manual contains

More information

SPECIFICATIONS - DETAILED PROVISIONS Section Magnetic Flowmeters C O N T E N T S

SPECIFICATIONS - DETAILED PROVISIONS Section Magnetic Flowmeters C O N T E N T S Revised 04/13/16 SPECIFICATIONS - DETAILED PROVISIONS Section 17210 - Magnetic Flowmeters C O N T E N T S PART 1 - GENERAL... 1 1.01 GENERAL REQUIREMENTS... 1 1.02 PROJECT SPECIFIC REQUIREMENTS... 1 1.03

More information

SEC SIGNATURE ETO MONITOR TRAINING SENSOR ELECTRONICS CORPORATION

SEC SIGNATURE ETO MONITOR TRAINING SENSOR ELECTRONICS CORPORATION SEC SIGNATURE ETO MONITOR TRAINING SENSOR ELECTRONICS CORPORATION SEC SIGNATURE ETO MONITOR SIGETO1420597 PRODUCT INTRODUCED TO ETHYLENE OXIDE STERILIZATION INDUSTRY IN 2001. OVER 850 UNITS IN FIELD WORLDWIDE

More information

PORTABLE ISOTOPE IDENTIFIER Search Tool / Sample Counting System

PORTABLE ISOTOPE IDENTIFIER Search Tool / Sample Counting System FEATURES: HIGH GAMMA SENSITIVITY MULTI-CHANNEL ANALYZER (MCA) WORKS WITH VARIETY OF PROBES RADIONUCLIDE LIBRARY TAILORED TO CUSTOMER REQUEST SIMULTANEOUSLY MEASURES ; DISPLAYS; ALARMS 1-3 SEPARATE DETECTORS

More information

INSTALLATION INSTRUCTIONS FOR 6532 SERIES PACKAGE HEAT PUMP

INSTALLATION INSTRUCTIONS FOR 6532 SERIES PACKAGE HEAT PUMP INSTALLATION INSTRUCTIONS FOR 6532 SERIES PACKAGE HEAT PUMP RV Products A Division of Airxcel, Inc. P.O. Box 4020 Wichita, KS 67204 1976-360 (1-02) PP TABLE OF CONTENTS 1. Warnings......................................................

More information

Rosemount 148 Temperature Transmitter. Reference Manual , Rev CA April 2014

Rosemount 148 Temperature Transmitter. Reference Manual , Rev CA April 2014 Rosemount 148 Temperature Transmitter Reference Manual Reference Manual Rosemount 148 Temperature Transmitter Rosemount 148 Hardware Revision 5 NOTICE Read this manual before working with the product.

More information

CLARITY II TURBIDIMETER

CLARITY II TURBIDIMETER Product Data Sheet PDS 71-T1055/rev.J January 2006 Model T1055 CLARITY II TURBIDIMETER COMPLETE SYSTEM includes single or dual input analyzer, sensor(s), and debubbler assembly CHOOSE U.S. EPA METHOD 180.1

More information

3MAVR60XXD Series Fan. Owner s Guide and Installation Manual. CUL Model NO. : 3MAVR60XXD

3MAVR60XXD Series Fan. Owner s Guide and Installation Manual. CUL Model NO. : 3MAVR60XXD Owner s Guide and Installation Manual 3MAVR60XXD Series Fan CUL Model NO. : 3MAVR60XXD Attach sales receipt to this card and retain as your proof of purchase DATE OF PURCHASE: MODEL NUMBER: RETAILER NAME:

More information

Users Manual. LAURUS Systems, Inc. - Ph: Fax:

Users Manual. LAURUS Systems, Inc. - Ph: Fax: Users Manual LAURUS Systems, Inc. - Ph: 410-465-5558 - Fax: 410-465-5257 - www.laurussystems.com Introduction The rad-d is a security and inspection system that detects emissions from radioactive material.

More information

SEC Signature Process Gas Analyzer

SEC Signature Process Gas Analyzer SEC Signature Process Gas Analyzer Instruction and Operation Manual Sensor Electronics Corporation 5500 Lincoln Drive Minneapolis, Minnesota 55436 USA (952) 938-9486 Fax (952) 938-9617 email sensor@minn.net

More information

Installation and Operating Instructions

Installation and Operating Instructions The Leader of RF and Sonic Level For Assistance Call 1-800-527-6297 Outside North America 1-215-674-1234 Installation and Operating Instructions Series 504-1200 Ultrasonic VeriGAP Switch using 404-1200

More information

Model 6495 Freezing Rain Sensor. User's Manual. All Weather Inc National Drive Sacramento, CA USA

Model 6495 Freezing Rain Sensor. User's Manual. All Weather Inc National Drive Sacramento, CA USA Model 6495 Freezing Rain Sensor User's Manual All Weather Inc. 1165 National Drive Sacramento, CA 95834 USA 800.824.5873 www.allweatherinc.com Contents Technical Specifications...1 Theory of Operation...2

More information

Installation Instructions. Fire Command Center/Fire Alarm Annunciators D1256/D1257

Installation Instructions. Fire Command Center/Fire Alarm Annunciators D1256/D1257 Instructions Fire Command Center/Fire Alarm Annunciators D1256/D1257 74-06925-000-H Page 2 2004 Bosch Security Systems Contents 1.0 Introduction...5 1.1 Before You Begin...5 1.2 Type Styles Used Here...5

More information

TC-5231 Series, TC-5232, & TC-5241 Series

TC-5231 Series, TC-5232, & TC-5241 Series TC-5231 Series, TC-5232, & TC-5241 Series Low Temperature Thermostats General Instructions Application The TC-5231, TC-5232, and TC-5241 low temperature thermostats are used to control temperature in air

More information

SAFETY MANUAL. X2200 UV, X9800 IR, X5200 UVIR SIL 2 Certified Flame Detectors

SAFETY MANUAL. X2200 UV, X9800 IR, X5200 UVIR SIL 2 Certified Flame Detectors SAFETY MANUAL X2200 UV, X9800 IR, X5200 UVIR SIL 2 Certified Flame Detectors SAFETY-CERTIFIED Flame DETECTORs This manual addresses the specific requirements and recommendations applicable to the proper

More information

Owner s Guide and Installation Manual

Owner s Guide and Installation Manual For Your Records and Warranty Assistance For reference, also attach your receipt or a copy of your receipt to the manual. Model Name Type 2A Models Owner s Guide and Installation Manual Model No. Date

More information

CONsOlIDATOR 4 & 8. MulTI- C h ANNEl CONTROllERs. ConsoliDator 4 Model PD940 ConsoliDator 4 Features. ConsoliDator 8 Features.

CONsOlIDATOR 4 & 8. MulTI- C h ANNEl CONTROllERs. ConsoliDator 4 Model PD940 ConsoliDator 4 Features. ConsoliDator 8 Features. CONsOlIDATOR 4 & 8 MulTI- C h ANNEl CONTROllERs ConsoliDator 4 Model PD940 ConsoliDator 4 Features Four 4-20 Four 4-20 Outputs ConsoliDator 8 Features Eight 4-20 Two 4-20 Outputs Common Features Four Pulse

More information

21-light Remote Annunciator. Owner s Manual

21-light Remote Annunciator. Owner s Manual 21-light Remote Annunciator Owner s Manual Annunciator Description... Inside Font Cover Detailed Specifications... 1 Environmental Specifications... 1 Power Supply Requirements... 1 Communication With

More information

Sapphire ULTRASONIC POINT LEVEL SYSTEM

Sapphire ULTRASONIC POINT LEVEL SYSTEM ULTRASONIC POINT LEVEL SYSTEM Introduction Compliance with growing government regulations mandates tighter level controls with advanced diagnostics to avoid releases, overspills and costly environmental

More information

OVEN INDUSTRIES, INC.

OVEN INDUSTRIES, INC. OVEN INDUSTRIES, INC. OPERATING MANUAL Model 5C7-252 TEMPERATURE CONTROLLER With PLC Inputs Introduction Thank you for purchasing our controller. The Model 5C7-252 is an exceptionally versatile unit and

More information

Dual Input ph/orp Analyzer

Dual Input ph/orp Analyzer Instruction Sheet PN 51A-1055pH/rev.I February 2006 Model 1055 SOLU COMP II Dual Input ph/orp Analyzer Model Option 1055-22-32 For additional information, please refer to the Instruction Manuals CD shipped

More information

DENVER PUBLIC SCHOOLS DESIGN AND CONSTRUCTION STANDARDS This Standard is for guidance only. SECTION PUMPS

DENVER PUBLIC SCHOOLS DESIGN AND CONSTRUCTION STANDARDS This Standard is for guidance only. SECTION PUMPS PART 0 A/E INSTRUCTIONS 0.01 Design Requirements A. Pumping system design 1. A primary-secondary pumping system is preferred. Redundant pipes are required for chillers and boilers. 2. Select pumps to operate

More information

SCM. Micro. Streaming Current Monitor. Analyzer. Sensor/Sampler with light shield removed. The Analyzer. Instrument Overview

SCM. Micro. Streaming Current Monitor. Analyzer. Sensor/Sampler with light shield removed. The Analyzer. Instrument Overview Micro SCM Streaming Current Monitor A Watts Water Technologies Company Instrument Overview The instrument consists of the Analyzer and the SCM sensor. Analyzer The Analyzer The monitor has a built in graphic

More information

Snifter ATEX22 VERSION. User Manual. Distributor

Snifter ATEX22 VERSION. User Manual. Distributor Snifter ATEX22 VERSION User Manual Distributor Version 1.4 09/09/2009 Table of Contents 1. INTRODUCTION... 3 1.1. Safety... 3 1.2. Product overview... 4 1.3. How does it work?... 4 2. INSTALLATION... 5

More information

DET-TRONICS SPECIFICATION DATA. Eagle Quantum EQ2200UVHT For Use With C7050B Continuous Duty High Temperature UV Flame Detector DESCRIPTION

DET-TRONICS SPECIFICATION DATA. Eagle Quantum EQ2200UVHT For Use With C7050B Continuous Duty High Temperature UV Flame Detector DESCRIPTION DET-TRONICS SPECIFICATION DATA Eagle Quantum EQ2200UVHT For Use With C7050B Continuous Duty High Temperature UV Flame Detector DESCRIPTION The EQ2200UVHT UV Flame Detector is used with the Eagle Quantum

More information

SAVE THESE INSTRUCTIONS

SAVE THESE INSTRUCTIONS Built-In Dishwasher Dishwashers Write the model and serial numbers here: Model # Serial # You can find them on the tub wall just inside the door or. the lower part of back. Installation Instructions DDW1802W

More information

Flame Rod Manual. Publication Rev A

Flame Rod Manual. Publication Rev A Publication INTRODUCTION This manual contains information for the Flame Rod Flame Detection system from Forney Corporation, 16479 North Dallas Parkway, Suite 600 Addison, TX 75001. All personnel should

More information

Table of Contents. What to Expect with. Tools Needed. Mounting Options. Blades

Table of Contents.   What to Expect with. Tools Needed. Mounting Options. Blades Table of Contents 1.888.830.1326 Ceiling Bracket 30 inches Ladder Downrod 3 Wiring 10 Operation, Maintenance & Cleaning Light Kit 11 Troubleshooting??? 17 15 1 13 5 Blades 9 Switch Housing 12 Canopy 6

More information

STEAMPRO. Steam Generator Troubleshooting and Service Guide

STEAMPRO. Steam Generator Troubleshooting and Service Guide STEAMPRO Steam Generator Troubleshooting and Service Guide TABLE OF CONTENTS Page PREFACE... 1 I. STEAMPRO STEAM GENERATOR SYSTEM...2 II. PLUMBING AND ELECTRICAL...3-4 III. SYSTEM OVERVIEW... 5-10 IV.

More information

ICS Regent. Fire Detector Input Modules PD-6032 (T3419)

ICS Regent. Fire Detector Input Modules PD-6032 (T3419) ICS Regent Fire Detector Input Modules (T3419) Issue 1, March, 06 Fire detector input modules provide interfaces for 16 fire detector inputs such as smoke detectors, flame detectors, temperature detectors,

More information

CMT MANUAL 6. NMC SENSORS

CMT MANUAL 6. NMC SENSORS CMT MANUAL 6. NMC SENSORS TABLE OF CONTENTS INDEMNITY... 5 DOCUMENT INFORMATION... 5 ACKNOWLEDGEMENTS... 5 6.1 INTRODUCTION... 6 6.2 TEMPERATURE AND HUMIDITY SENSORS... 7 6.2.1. TEMPERATURE SENSOR... 7

More information

Lightning Warning System: LWS MK-II. User's Guide. KOREA EMI Technologies Co.,Ltd , Jangam-ri, Majang-myeon, Icheon-si, Gyeonggi-do, Korea

Lightning Warning System: LWS MK-II. User's Guide. KOREA EMI Technologies Co.,Ltd , Jangam-ri, Majang-myeon, Icheon-si, Gyeonggi-do, Korea User's Guide CONTENTS FEATURES WHAT YOU NEED TO GET STARTED INSTALLATION DETAILS Installing the Antenna Installing the External Alert Siren The main console OPERATING PROCEDURE Power On Routine Push Buttons

More information

Installation Instructions. For the 18 Built-In Dishwasher and Front Color Panels

Installation Instructions. For the 18 Built-In Dishwasher and Front Color Panels Installation Instructions For the 18 Built-In Dishwasher and Front Color Panels Printed in USA 154232102 Before You Begin DO NOT INSTALL DISHWASHER UNTIL YOU HAVE READ ALL INSTRUCTIONS. FOR YOUR SAFETY,

More information

MAINTENANCE & TROUBLESHOOTING GUIDE LEAK ALARM CHANNEL DRY OIL WATER AUX ALARM HIGH LOW CRITICAL WATER TANK LEVEL ALARM MODEL LDE-740 ADVANCE PAPER

MAINTENANCE & TROUBLESHOOTING GUIDE LEAK ALARM CHANNEL DRY OIL WATER AUX ALARM HIGH LOW CRITICAL WATER TANK LEVEL ALARM MODEL LDE-740 ADVANCE PAPER PNEUMERCATOR Liquid Level Control Systems Electronic Systems Excluding LC2000 And TMS Series MAINTENANCE & TROUBLESHOOTING GUIDE 400 300 500 FUEL LEVEL LEAK MONITOR 1 200 600 OIL/GAS NORMAL WATER PNEUMERCATOR

More information

Owner s Guide and Installation Manual

Owner s Guide and Installation Manual Tribeca Owner s Guide and Installation Manual English Form# M6001-01 20120416 2012 Casablanca Fan Co. Welcome Your new Casablanca ceiling fan is an addition to your home or office that will provide comfort

More information

Technologies, Solutions, and Applications. Continuous Level Coke Drum Measurement System

Technologies, Solutions, and Applications. Continuous Level Coke Drum Measurement System Technologies, Solutions, and Applications Continuous Level Coke Drum Measurement System Leadership in Continuous Coke Drum Level Refiners are processing heavier crude, which increases the amount of resid

More information

3MAVR60XX Series Fan. Owner s Guide and Installation Manual. UL Model NO. : 3MAVR60XX

3MAVR60XX Series Fan. Owner s Guide and Installation Manual. UL Model NO. : 3MAVR60XX Owner s Guide and Installation Manual 3MAVR60XX Series Fan UL Model NO. : 3MAVR60XX Attach sales receipt to this card and retain as your proof of purchase DATE OF PURCHASE: MODEL NUMBER: RETAILER NAME:

More information

mtc001 Thermocouple Converter

mtc001 Thermocouple Converter THERMOCOUPLE CONVERTER 1-WIRE INTRINSICALLY SAFE mtc001 mtc001 Thermocouple Converter Document No. 10789, Revision A February 2007 2006 CMC Industrial Electronics Ltd. This page intentionally left blank.

More information

3MNLR72XXD Series Fan. Owner s Guide and Installation Manual. UL Model NO. : 3MNLR72

3MNLR72XXD Series Fan. Owner s Guide and Installation Manual. UL Model NO. : 3MNLR72 Owner s Guide and Installation Manual 3MNLR72XXD Series Fan UL Model NO. : 3MNLR72 Attach sales receipt to this card and retain as your proof of purchase DATE OF PURCHASE: MODEL NUMBER: RETAILER NAME:

More information

Continuous and Point Level Measurement. Radiometric Sensors for Liquid and Solid Level Measurement

Continuous and Point Level Measurement. Radiometric Sensors for Liquid and Solid Level Measurement Continuous and Point Level Measurement Radiometric Sensors for Liquid and Solid Level Measurement Contents Leadership in Radiometric Technology 3 plics Easy is Better 4 FiberTrac 31 Continuous Level 6

More information

MEGA. cal electronic gas alarm. Installation, Operation, and Maintenance Instructions. Part No Rev. D Pg.

MEGA. cal electronic gas alarm. Installation, Operation, and Maintenance Instructions. Part No Rev. D Pg. WARD OXYGEN Installation, Operation, and Maintenance Instructions MEGA medical electronic gas alarm MEGA WARD MEGA medical electronic gas alarm WARD POWER SUPPLY Medical Electronic Gas Alarm OXYGEN SYSTEM

More information

Firefly II Multi Voltage RELEASE DEVICES. Firefly II & IIB-MV INSTALLATION MANUAL

Firefly II Multi Voltage RELEASE DEVICES. Firefly II & IIB-MV INSTALLATION MANUAL Firefly II Multi Voltage RELEASE DEVICES Firefly II & IIB-MV INSTALLATION MANUAL GENERAL DESCRIPTION: S/N The The Cookson Company Firefly II-MV time delay release devices are designed for use on rolling

More information

SOLID STATE SECURITIES, INC. THE "SS90" SERIES RELEASE DEVICES MADE IN THE U.S.A. SS90-WPS MODEL A/B INSTALLATION MANUAL

SOLID STATE SECURITIES, INC. THE SS90 SERIES RELEASE DEVICES MADE IN THE U.S.A. SS90-WPS MODEL A/B INSTALLATION MANUAL S S S S S S LISTED L * PATENT PENDING S S 90 OLID TATE ECURITIES, INC SOLID STATE FAIL-SAFE UNIT U 99Y9 RESET RELEASING DEVICE SOLID STATE SECURITIES, INC. THE "SS90" SERIES RELEASE DEVICES MADE IN THE

More information

installation and operation manual for Hunter Ceiling Fans

installation and operation manual for Hunter Ceiling Fans For Your Records and Warranty Assistance Model Name: Catalog/Model No.: Serial No.: Date Purchased: Where Purchased: For reference also attach your receipt or a copy of your receipt to the manual. installation

More information

Operation and Maintenance Manual

Operation and Maintenance Manual Operation and Maintenance Manual GDS-48 Remote Bridge Sensor for Combustible Gases AUTHORIZED DISTRIBUTOR: GasDetectorsUSA.com - Houston, Texas USA sales@gasdetectorsusa.com - 832-615-3588 CAUTION: FOR

More information

GasScanner 4C. Four Channel Monitor. Operator s Manual. MINT-0280-XX Rev. A 02/25/08

GasScanner 4C. Four Channel Monitor. Operator s Manual. MINT-0280-XX Rev. A 02/25/08 GasScanner 4C Four Channel Monitor Operator s Manual MINT-0280-XX Rev. A 02/25/08 Product Warranty Matheson Tri-Gas., warrants gas alarm equipment sold by us to be free from defects in materials, workmanship,

More information

SAMPLE MANUAL - INSTALLATION - FOR REFERENCE ONLY. ChemScan mini Analyzer. Parameter. Method #### Installation, Operation & Maintenance Manual

SAMPLE MANUAL - INSTALLATION - FOR REFERENCE ONLY. ChemScan mini Analyzer. Parameter. Method #### Installation, Operation & Maintenance Manual Parameter Method ####, Operation & Maintenance Manual Rev 3/26/2018 Contact Information Document Information Customer Name: Plant Name/Location: Date: Analyzer Serial #: ASA, Inc. (Applied Spectrometry

More information

Owner s Guide and Installation Manual

Owner s Guide and Installation Manual For Your Records and Warranty Assistance For reference, also attach your receipt or a copy of your receipt to the manual. Model Name Type 2A Models Owner s Guide and Installation Manual Model No. Date

More information

SLATE. Base Module INSTALLATION INSTRUCTIONS R8001A1001

SLATE. Base Module INSTALLATION INSTRUCTIONS R8001A1001 SLATE Base Module R8001A1001 INSTALLATION INSTRUCTIONS Scan for more information Application SLATE brings configurable safety and programmable logic together into one single platform. The platform can

More information

Series 155. Mass Flow Computers TECHNICAL SPECIFICATIONS

Series 155. Mass Flow Computers TECHNICAL SPECIFICATIONS TECHNICAL SPECIFICATIONS Mass Flow Computers Series 155 The Kurz 155 Mass Flow Computer is a versatile system transmitter that integrates the functions of flow and temperature measurement, flow totalization,

More information

5YK60XXX Series Fan. Owner s Guide and Installation Manual. CUL Model NO. : 5YK60XXX

5YK60XXX Series Fan. Owner s Guide and Installation Manual. CUL Model NO. : 5YK60XXX Owner s Guide and Installation Manual 5YK60XXX Series Fan CUL Model NO. : 5YK60XXX Attach sales receipt to this card and retain as your proof of purchase DATE OF PURCHASE: MODEL NUMBER: RETAILER NAME:

More information